Professional Documents
Culture Documents
Dialyzer Type
Dialyzer Type
Insights
A hollow fiber dialyzer bundle comprises 7–17 x 103 semipermeable hollow fibers
that allow solute and fluid transfer between blood and dialysate. Typical fibers have
an internal diameter of 180–200 microns and wall thickness of 30–40 microns,
yielding 1.0–2.5 m2 of surface area. Fibers may have features, such as undulations,
to distribute dialysate flow evenly through the fiber bundle.
The fiber bundle is enclosed in an outer housing that forms the dialysate
compartment. The header is the space enclosed by the end cap and the
polyurethane potting material that holds the hollow fibers and forms a barrier
between the blood and dialysate compartments. Headers channel blood from the
dialyzer inlet into the membrane fibers and from the membrane fibers into the
dialyzer outlet. The end caps can be removed from some dialyzers. In those cases,
an O-ring is used to form a seal between the end cap and the potting material. Blood
and dialysate flow in opposite directions (countercurrent flow) to maximize diffusive
solute transfer.
MEMBRANE MATERIALS AND BIOCOMPATIBILITY
Non-synthetic membranes are derived from natural materials such as cotton and are
less biocompatible than synthetic membranes. Biocompatibility can be improved by
substituting hydroxyl groups, which reduces the ability of cellulose membranes to
activate complement and cause leukopenia. Moieties that have been substituted for
cellulose include acetate, diethylaminoethyl (DEAE), benzyl, polyethyleneglycolic,
and vitamin E. The resultant membranes are referred to as modified cellulose
membranes. Only cellulose diacetate and cellulose triacetate membranes remain in
widespread clinical use in the U.S. Modified cellulose membranes can be either high
flux or low flux.
PSf (polysulfone and a family of polysulfone blends) offer good biocompatibility and
protection from endotoxins
PES (polyethersulfone) which is a blend of hydrophobic base polymers with good
biocompatibility and less albumin loss
CTA (cellulose triacetate) with increased solute permeability especially for beta-2
microglobulin
PMMA (polymethylmethacrylate) with increased adsorption properties to enhance
removal of some inflammatory molecules
PEPA (PES plus polyarylate) with enhanced endotoxin protection
EVAL (ethylene vinyl alcohol copolymers) with low inflammatory impact
PAN (polyacrylonitrile) with good biocompatibility and enhanced fluid removal due to
hydrophilic properties
DIALYZER PERFORMANCE CHARACTERISTICS
The primary mode for removal of small solutes (e.g., urea) by hemodialysis is
diffusion down the concentration gradient between plasma water and dialysate.
Transfer of small solutes (e.g., HCO3-) from dialysate to plasma water also occurs
primarily by diffusion. The rate of diffusion is a function of the thickness and porosity
of the membrane and the diffusivity of the solute in the membrane. It is expressed as
the diffusion coefficient of the membrane for a given solute. The rate of diffusion is
greatest for small molecules and the diffusivity of a solute in a membrane decreases
logarithmically as solute size increases. The rate of diffusion also decreases as
membrane thickness increases and porosity decreases.
Convection provides better removal of large solutes than diffusion because the
decrease in sieving coefficient with increasing solute size is less marked than the
decrease in diffusion coefficient. Dialyzer manufacturers usually provide sieving
coefficients of albumin, beta-2 microglobulin, myoglobin, and lysozyme as
parameters of albumin leak and convective performance.
During hemodialysis, proteins may adsorb to both the planar surface of the
membrane and the inner surface of its pores. The adsorbed protein may decrease
diffusive and convective removal of other solutes by effectively reducing the pore size
of the membrane. Hydrophobic surfaces adsorb serum proteins, which can contribute
significantly to low-molecular-weight protein removal by some membranes.
DIALYZER CLEARANCE
DIALYZER SELECTION
For example, Mr. Doe has a V of 40L and the target Kt/V is 1.4:
Kxt/V or Kxt/40 = 1.4
Let's assume you have a dialyzer that has in-vitro Kurea of 250 ml/min at Qb/Qd of
300/500 ml per minute. This clearance should be multiplied by 0.85, which is about a
15% adjustment from in-vitro to in-vivo clearance. Therefore, K would be 250 x 0.85
= 212.5 ml/min.
212.5 x t = 56,000
On the other hand, if you know the duration of treatment that you are going to
prescribe (let's assume 4 hours or 240 minutes), you can solve for the Kurea needed
to deliver Kt/Vurea of 1.4.
K x 240 = 56,000
So, you will need a dialyzer with published in-vitro urea clearance of 274 ml/min at a
Qb/Qd of 300/500 ml/min for this patient.
1. You can have different Qb/Qd, such as 400/600 or 500/800 ml/min, and should use
them in these calculations.
2. Anthropometric TBW calculations overestimate V, therefore, the delivered
Kt/Vurea may be higher than the calculated one.
3. Total delivered dose may have a contribution from residual kidney function, which is
not accounted for in these calculations.
4. At best, this calculation provides you with a starting point, and adjustments might be
needed based on the measured Kt/Vurea lab values.
5. Currently, there is no evidence to suggest that adequacy targets should be modified
for different etiologies of ESRD, except in special circumstances such as pregnant
patients who need more frequent or daily HD with additional adjustment to the
dialysate composition given their special physiological needs.
6. It would be prudent to avoid delivering high clearances in patients with liver failure or
others who may be at higher risk of increased intracranial pressure such brain trauma
or surgery.