Professional Documents
Culture Documents
تعديل مناقشة
تعديل مناقشة
Scientific Research
Al-Nahrain University
College of Engineering
A Thesis
Submitted to the College of Engineering of
Al-Nahrain University in Partial Fulfillment
of the Requirements for the Degree of Master of Science
in
Biomedical Engineering
by
Hamza Abbas Fadhel
)B.Sc. Biomedical Engineering 2015(
We certify that this thesis entitled "Gait Analysis of below Knee Amputees
Wearing Different Prosthetic Feet" was prepared under our supervision at
the College of Engineering / Al-Nahrain University in partial fulfillment of
the requirement for the degree of Master of Science in Biomedical
Engineering.
Asst. Prof. Dr. Sadiq J. Hamandi Asst. Prof. Dr. Muhammed Abdulsattar
Muhammed
Supervisor Supervisor
/12/2020 /12/2020
Asst. Prof. Dr. Sadiq J. Hamandi Asst. Prof. Dr. Muhammed Abdulsattar
Muhammed
)Supervisor and Member) (Supervisor and Member)
12/2020/ 12/2020/
The suitability of the prosthetic limb of the amputee is a key factor for
rehabilitation, as well as the type of prosthetic foot, which is one of the
factors affecting the gait parameter. The study aims to assess subject
preference for prosthetic foot based of dynamic performance using different
prosthetic feet. Six male subjects with unilateral below-knee amputation were
chosen to fit with four types of prosthetic feet (1S66 SACH foot Ottobock
inc., 1C30 Trias carbon Ottobock inc., Dynastep Proteor group and Dynacity
Proteor group), asked them to walk with their usual speed on the straight
pathway perpendicularly with the camera. The Spatio-temporal parameters for
both amputated leg and the sound leg measured manually by using the
Kinovea software. Also, Skillspector was used for calculating the joint angles
in the sagittal plane, while the ground reaction forces (GRFs) were acquired
from the Advanced Mechanical Technology Incorporation (AMTI) force
platform. The results showed there are no significant differences in the
spatiotemporal parameters for p=0.05 between the four types of prosthetic
feet on the sound and amputee leg. The comparison between the amputee and
sound leg showed a significant difference p<0.05 for the single support time
when the patients fit with Trias 1C30 (18% lower than the sound side) and
Dynastep (14% lower than sound side). For joint kinematics the study found a
little effect in joint angles in the lower limb for both the amputated side and
the sound side all joint range of motion ROM at the normal value with hip
ROM 50 deg, knee ROM 50 deg and ankle ROM 35 deg. An increase in the
sound side ground reaction force when using SACH foot, Dynastep and Trias
1C30 by 6%, 12% and 19% respectively over the amputated side except for
the Dynacity foot the first peak in the sound side is lower than the amputated
side by 1.7%. With all types of used prosthetic feet.
List of Contents
Contents
Page
Abstract I
List of Contents II
Notations V
List of Tables VI
List of Figures VII
CHAPTER ONE: Introduction
1.1 Introduction 1
1.2 Literature Review 3
1.3 Aims of the Research 10
1.4 Thesis Outline 10
CHAPTER TWO: Theory
2.1 Introduction 12
2.2 Activity level 12
2.3 Prosthetic foot 13
2.4 Gait Cycle Classification 15
2.4.1 The Main Tasks of the Gait Cycle 16
2.5 Gait Analysis 17
2.5.1 Spatiotemporal Parameters 17
2.5.1.1 Normative Values 18
2.5.2 Gait Kinematics 19
2.5.2.1 Measurements of Gait Kinematics 19
2.5.2.2 Calculation of Segment and Joint Angles 21
2.5.2.3 Normal Gait Kinematics 22
2.5.3 Gait Kinetics 24
2.5.3.1 Ground Reaction Forces 24
2.5.3.2 Normal Kinetics 26
2.6 Force Platform 27
CHAPTER THREE: Subjects and Methodology
3.1 Introduction 28
3.2 Subjects 28
3.3 Ethical Approval 28
3.4 Prosthetic Foot 29
3.5 Methodology 32
3.5.1 Research Procedure 32
3.5.2 Markers and their Placements 33
3.5.3 The Gait Analysis System 35
3.5.4 Measurement of the General Gait Parameters 36
3.5.4.1 Statistical Analysis 39
3.5.5 Measurement of Kinematics 39
3.5.5.1 Digitization Process of Marker 39
3.5.5.2 Filtering of Gait Data 42
3.5.6 Measurement of Kinetics 43
CHAPTER FOUR :Results and Discussion
4.1 Introduction 47
4.2 Spatiotemporal Parameters 47
4.3 Kinematic Analysis 53
4.4 Ground Reaction Forces 77
CHAPTER FIVE : Conclusions and Recommendations for Future
Works
5.1 Conclusions 102
5.2 Recommendations for Future Works 103
REFERENCES 104
APPENDISES A-1
Appendix A A-1
Abbreviations
3D = Three Dimension
AMTI = Advanced Mechanical Technology Incorporation
AVI = Audio Video Interleaved
BH = Body Height
BM = Body Mass
BW = Body Weight
COG = Center of Gravity
COM = Center of Mass
COP = Center of Pressure
ESAR = Energy Storing and Return
GC = Gait Cycle
GRF = Ground Reaction Force
IC = Initial Contact
ISW = Initial Swing
LA = Limb Advancement
LR = Loading Response
MATLAB = MATrix LABoratory
MFCL = Medicare Functional Classification Level
MST = Mid-Stance
MSW = Mid-Swing
PSW = Pre-Swing
SACH = Solid-Ankle Cushion-Heel
SLS = Single Limb Support
TST = Terminal Stance
TSW = Terminal Swing
WA = Weight Acceptance
Notations
Symbols Notations
fs = Sampling frequency
= Medial-lateral ground reaction
Fx
force
= Forward-backward ground reaction
Fy
force
Fz = Vertical ground reaction force
FYb = Braking force
FYp = Propulsive force
Fxm = Medial force
Fxl = Lateral force
g = Gravitational acceleration
Xi, Yj = Cartesian position of the center of
joint in x-y direction
𝜃 = Angular displacement
List of Tables
(2-2) Energy storing and release feet. (a) Single axis (b) multi 14
axis
(2-3) Left extremity initial contact takes place while the right 16
extremity is at 50% of the gait cycle
(4-1) Single support time of the sound and amputee legs for six 51
subjects
(4-2) Case-1 Hip joint angle (a) sound side (b) amputated side 56
(4-3) Case-1 knee joint angle (a) sound side (b) amputated side 57
(4-4) Case-1ankle joint angles (a) sound side (b) amputated side 58
(4-5) Case-2 hip joint angle (a) sound side (b) amputated side 60
(4-6) Case-2 knee joint angle (a) sound side (b) amputated side 61
(4-7) Case-2 ankle joint angle (a) sound side (b) amputated side 62
(4-8) Case-3 hip joint angle (a) sound side (b) amputated side 63
(4-9) Case-3 knee joint angle (a) sound side (b) amputated side 64
(4-10) Case-3 ankle joint angle (a) sound side (b) amputated side 65
(4-11) Case-4 hip joint angle (a) sound side (b) amputated side 67
(4-12) Case-4 knee joint angle (a) sound side (b) amputated side 68
(4-13) Case-4 ankle joint angle (a) sound side (b) amputated side 69
(4-14) Case-5 hip joint angle (a) sound side (b) amputated side 70
(4-15) Case-5 knee joint angle (a) sound side (b) amputated side 71
(4-16) Case-5 ankle joint angle (a) sound side (b) amputated side 73
(4-17) Case-6 hip joint angle (a) sound side (b) amputated side 74
(4-18) Case-6 knee joint angle (a) sound side (b) amputated side 75
(4-19) Case-6 ankle joint angle (a) sound side (b) amputated side 76
(4-20) Case-1 vertical ground reaction force (a) sound side (b) 79
amputated side
(4-23) Case-2 vertical ground reaction force (a) sound side (b) 83
amputated side
(4-26) Case-3 vertical ground reaction force (a) sound side (b) 87
amputated side
(4-29) Case-4 vertical ground reaction force (a) sound side (b) 91
amputated side
(4-32) Case-5 vertical ground reaction force (a) sound side (b) 94
amputated side
(4-35) Case-6 vertical ground reaction force (a) sound side (b) 98
amputated side
(4-36) Case-6 forward-backward ground reaction force (a) sound 99
side (b) amputated side
(4-37) Case-6 medial-lateral ground reaction force (a) sound side 100
(b) amputated side
CHAPTER ONE
Introduction
1.1 Introduction
As a result of increased accidents and wars in Iraq, the number of
amputees reached 160 thousand to 180 thousand, according to the estimates
for the Iraqi Ministry of Health and civil society organizations for the year
2019 only. In the Nineveh Governorate, the number of amputations exceeds
four thousand patients with different types of amputations for the last four
years due to the war against ISIS [1].
Estimated statistics indicate that the number of lower-limb amputations
in the United States ranges from 2.8 to 43.9 per 10,000 people, most of whom
are below the knee amputation [2].
Despite scientific progress in all fields, the issue of amputees remains a
challenge for all countries [3]. Amputations may increase in the coming years
with increasing life expectancy and increasing diabetes cases, and it is
expected to be twice more when people reach the age of 65 [4].
Also, the chance of amputation is greater in accidents in the case of
vascular patients. By the middle of the 21 st century, amputations are expected
to double, according to a study done in 2008 [5]. In Korea, below-knee
amputees are the most of limb amputations [6]. Therefore, consideration
should be focused on the below knee amputations, to aid amputees to
effectively back to their earlier activities of life [7]. The suitability of the
prosthetic limb of the amputee is a key factor for rehabilitation, as well as the
type of prosthetic foot, which is one of the factors affecting the gait parameter
[8].
At present, there are many types of prosthetic feet and are described
according to the patient's condition [9]. Numerous single-axis and multi-axis
prosthetic feet exist. In the late 1950s, an inactive foot consists of a foot and
ankle connected in one set such as solid ankle-cushin heel (SACH) [10].
Because the weight of the SACH foot is light and its toughness, it turns out to
be the most recommended in the United States for trans-tibial and trans-
femoral amputation [11].
The new version of SACH is designed to be lighter than the old one,
durable, and include an elastic heel that wrappings at heel strike, make the
patient walk in a more dynamically sense [8]. The mechanical disadvantage in
SACH outcome in variations in the gait of below-knee amputees [2].
Other inactive feet offer slight energy-saving and released during the gait
because of the mechanical properties of material made from it such as energy
storing and return (ESAR), therefore, they offer some biomechanical gains
[12]. These properties offer the energy to push the amputation forward such
as that supplied by the calf muscle at the time of late stance [13]. In many
cases, the prosthetic limbs do not compensate for the amputated limb,
sometimes it affects the speed of the walk, and a difference in spatiotemporal
parameters and joint angles had been occurred compared to the non-
amputated person during gait [7]. It causes a decrease in the swing phase of
the sound side and an increase of amputated side stride-length, as a result of a
mismatch in the adaptation of the patient to the prosthetic foot is asymmetries
in gait [14,15]. Differences between a sound and amputated limb can be
recognized by measuring gait parameters, kinetics, and kinematics, and these
differences are used as an indication for the rehabilitation process [16,17]. By
comparing the Non-amputated individuals with the amputated individuals, the
first group moves with a higher pace and cadence and the stride length is
longer [18].
The research reported a quite longer step length, a longer swing time,
and a smaller stance phase for the amputated limb [19]. The patient relies on
the sound side to compensate for the performance deficiency on the prosthesis
[20]. This will lead to pain in joints and deterioration. Struyf et al. established
that impaired persons suffer from osteoarthritis in the sound side are three
times higher compared to non-amputated persons [2].
Gait analysis is an essential way in the assessment of operational events,
rehabilitation development, kinematics of joint, spatiotemporal parameters,
and many others are factors of concern such as ground reaction forces [21].
Some specialists, such as the prosthetist, measure the spatiotemporal
parameters such as stride rate, cadence, and step length to know the progress
of the rehabilitation process [22]. Currently, numerous viable motion systems
for two-dimension analysis based on the camera. Kinematic gait analysis can
be categorized into two main classes [23]. The first one contains inertial
components. The second is a visual technique using inactive makers, which
were established to do actual-time gait analysis [22].
2.1 Introduction
This chapter contains five main parts illustrate the amputations and their
types, defining the prosthesis and its contents, summarize the gait and its
classifications, define the gait analysis techniques, and give a short
explanation about the devices used in gait analysis.
This study focused on the types of the prosthetic feet and their effects on
gait, so the prosthetic feet will be clarified and classified according to their
axis of motion and the materials made from them into three main sections.
a) SACH foot:
It is the oldest type of artificial feet known and the most widespread due
to its lightweight and simplicity of manufacture. It is available in two types: a
hard heel and flexible keel as shown in Figure 2-1 [53]. It is used for
amputees with little mobility and who are classified as K1 level. This foot is
considered without articulation [53].
(a) (b)
Figure 2-1 SACH foot (a) solid keel (b) flexible keel.[52]
b) Energy storing and release feet (ESAR).
Allows single-axis motion (plantarflexion) to happen at heel strike to
midstance to make certain stability for the wearer. One of the characteristics
of this foot is that it adds more balance, so it is used for amputation above the
knee and the possibility of using it for various terrains as shown in Figure 2-2
[53].
Finally, in the 1990s the dynamic response foot was produced to “lighten
the load” on the amputee using a material called carbon fiber to offer a light-
weight and strong prosthetic feet. The design provides absorption of the
impact at heel strike and releases that energy at toe-off [54].
Another design of dynamic response foot is the “split blade shape”
which allows multi-axis movements, medial and lateral ground slope, and
have a more natural transition from heel strike stage to late stance as shown in
Figure 2-4 [53].
(a) (b)
Figure 2-2 Energy storing and release feet. (a) Single axis (b) multi axis. [52]
c) Computerized feet
Powered and computerized feet similar to the intelligent knee joint the
ankle joint in this type of prosthetic feet have the ability to do the articulation
dorsiflexion and plantarflexion under microprocessor control to mimic the
activity of the calf muscle. The main challenges to this foot were to create
energy and how to store and release it [55].
Normally, the gait cycle consists of two major phases: the stance phase
and the swing phase as shown in Figure 2-3. This cataloging is identified as
the Ranchos cataloging and an event that is isolated in each period. The
names of these events are created from the movement of foot [57].
The stance phase is that portion of the gait cycle when the foot is in
contact with the ground, and typically represents approximately 62% of the
total healthy adult walking gait cycle. Three-foot rockers (heel, ankle, and
forefoot) occurring during stance phase, serve to control the forward fall of
the body during normal ambulation [57].
The swing phase is defined as the period when the foot no longer
contacts the ground and the limb advances in preparation for subsequent foot
contact. The swing phase occupies the residual 38% of gait cycle [57].
The gait cycle can also have characterized by eight distinct events; each
event has its time period measured as a percentage of the gait cycle. The
stance phase consists of five events: initial contact (IC) about (0-2) %, loading
response (LR) (0-10) %, mid-stance (MST) (10-30) %, terminal stance (TST)
(30-50) %, and pre-swing (PSW) (50-60) %. The swing phase, on the other
hand, consists of the other three events: initial swing (ISW) (60-73) %, mid-
swing (MSW) (73-87) %, and terminal swing (TSW) (87-100) % [58].
The two main phases of the gait cycle are the stance and swing phases.
The body is alternating during walking from one limb to another, meaning the
events of the gait of one limb are shifted from the other by 50%, e.g. while
one limb is swinging in the air, the other limb is supporting the body on the
ground. There is a time period the two limbs supporting the bodyweight on
the ground, which is named as the double limb support as shown in Figure 2-3
[58].
The cycle time is the interval of a complete gait cycle, which is divided
into stance time and swing time [58].
Parameters used in gait analysis and related to time and distance are
known as temporal and spatial parameters, respectively. Typically, the
measured spatial parameters are stride and step length, while temporal
parameters are walking speed, gait time, single support time and cadence
[59]:
1- Walking Speed
Although everyone has a usual (free or self-selected) walking speed,
the actual speed is continuously adjusted according to the conditions. The
walking speed can be calculated from the equation [59]:
2- Cadence
It can be defined as a number of steps per minute. The cadence can be
calculated from equation following [59]:
Since the way each person walks is different from the other and therefore
it depends on the nature of the person himself/herself and is affected by
internal and external factors, there is no fixed normal value for the
spatiotemporal parameters, but some scientists have studied them on natural
people and estimated as a range for each of the temporal-spatial factors for
different ages as shown in Tables 2-1 [59].
Table 2-1 Approximate ranges of the general gait parameters in free speed walking for
normal male subjects of different ages.[62]
The major motion capture sensors which are categorized in Table 2-2.
The two main families of sensors which have been commonly used in human
motion capture for rehabilitation engineering are optoelectronics and non-
optoelectronics sensors. The first groups may or may not use markers to track
movements. If they use, markers are attached to the body to represent major
skeletal segments and joints while the optical system (a camera and post-
processing vision system) tracks the markers and obtains the body segments
and joints’ position and orientation. In markerless systems, the image features
such as colors, edges, shapes, and/or depth are used to interpret the motions.
The non-optoelectronics sensors include inertial, mechanical, and magnetic
systems [59].
Table 2-2 Types of motion capture system.[59]
Non-optoelectronics Motion Capture Optoelectronics Motion Capture
(i) Inertial sensors (i) Marker trigonometry with IR
(ii) Magnetic systems cameras
(iii)Wearable systems (ii) Contrast-based
(iv) Mechanical systems (a)With color markers
(b)With skin detection
(iii) Depth-based
Figure 2-5 Gait analysis system for measuring kinematic and kinetic variables.[60]
2.5.2.2 Calculation of Segment and Joint Angles
The absolute segments angles can be computed from the data of the
marker trajectories by the use of simple arctangent functions as shown in
Figure 2-6 (a), the angle is always measured counterclockwise from the right
horizontal.
The general equation utilized to calculate the absolute segment angle is [61]:
Yj−Yi
𝜃ij = tan-1( Xj−Xi ) (2.4)
The joints' kinematics of a normal gait are shown in Figure 2-7. For the
hip joint, the angular displacement curve pattern is almost sinusoidal, starting
in approximately 30° flexion at the initial contact of the foot. Then extend
when the other side is at the initial contact about (50% cycle). It reaches a
maximum extension of 10° at the end of terminal stance and back to flexion
before the foot leaves contact with the ground when the next initial contact of
the ipsilateral happens [53].
Figure 2-7 Sagittal plane joint angles measured in degrees during a single gait cycle of the
right limb, the flexion of the hip, knee and ankle dorsiflexion considered positive.[62]
While the knee joint curve shows two flexion peaks at the stance and
swing phase. The flexion peak is much smaller in the stance phase than that in
the swing phase [63].
The knee angle approximately 5° of flexion in the initial contact. During
the loading response period, the knee went from 10° to 15° of flexion. At the
mid-stance period, the knee influences 5° of full extension just before the end
of stance then beginning a period of rapid flexion going to initial swing phase,
during mid-swing, the knee influences 60° of maximum flexion before
reaching a full extension at the end of swing phase then began to flex again
prior to initial contact [63].
The angle of the ankle joint, at the initial contact, is in zero-degree angle.
After that, during the loading response, there is a momentary plantarflexion
from 5° to 12° occurs. Following by moderate dorsiflexion that continues to
the mid-stance and terminal stance reaching 10° prior to pre-swing, then when
the foot loosening contact with the ground begins to rapidly plantar-flex this
occurs during the pre-swing remains into early swing reaches a maximum of
20° for the rest swing phase. Finally, at the end of the swing phase, it backs to
the zero-degree angle [53].
center of mass occurs and the vertical force is decreased under the
bodyweight forming a concave with its bottom ( F z ) during mid-stance. At
min
stance, F z falls down as the other limb starts supporting the bodyweight.
When the limb is swinging in the air, the ground reaction forces have zero
value because there is no touching to the ground [57].
Figure 2-8 Force plate data obtained from a normal individual while walking at normal
speed plotted against time as a percent of the stance phase.
(A) Vertical forces; (B) Anterior-posterior forces; (C) Medial-lateral forces.[57]
The forward-backward force ( F y) has an initial negative component
(braking force ( F y )), that represents the backward horizontal friction between
B
the foot and (floor). The second positive F y components, directed forward,
represent the plantar-flexor muscle activity in pushing the foot against the
floor, a propulsion force ( F y ) creating the movement [57].
P
foot abducts for toe-off, it turns into positive (medial force ( F x )) [57].
M
During the gait cycle, the joint moment and GRF direction and location
are changed. At heel strike, the GRF position posteriorly to the ankle joint
which makes the knee extensor muscles (i.e., triceps surae) to contract to
compensate the force. At the beginning of the gait, the bodyweight does not
expose to the limb, so the force generated by the muscle was low to overcome
the ground reaction force [64]. The vertical force must equal the body’s
weight to oppose the gravity force to keep the center of mass height above the
ground, so it has been studied [48]. When the vertical force is less than the
bodyweight, the center of mass moves downward and vice versa [48].
The force plate measures the ground reaction forces exerted by a person
steps on it during gait. These devices consist of a top plate (mounted level
with the surrounding floor) separated from a bottom frame by force
transducers near each corner. Its principle of work is that the applied force on
the top surface causes a certain amount of strain within the transducer. Force
plates enable one to measure not only the vertical and shear (anterior-
posterior and medial-lateral) forces but also the "center of pressure" during
gait. Modern video motion systems have made it possible to determine
ground reaction forces and moments through their kinetics software, which
requires ground reaction forces [57].
CHAPTER THREE
Subjects and Methodology
3.1 Introduction
*Mobility level giving by the United States Medicare Functional Classification Level (MFCL)..
** Mean and standard deviation for the subjects had complete the test.
The ethical approval for the protocol was obtained from the ethical
committee at Al-Nahrain University and data were collected at Al-Warith
Derman center for prosthetic and orthosis with the supervision of the therapist
physician of the center after the subjects sign up the formal list of the ethical
committee.
The most popular four types of prosthetic feet in Iraq were used in this
study, these feet are the most common used by the amputees and because of
its availability at Al-Warith center they are as follows:
o The upper blade brings the energy return needed for the push-off.
Comfort at heel strike is provided by the elastomer heel wedge.
Foot weight: 445 g.
Figure 3-3 Dynacity foot blades functions.[66]
3.5 Methodology
3.5.1 Research Procedure
If possible, the passive markers are placed directly on the subject's skin.
If it was not possible, the markers were placed on tight fixed clothes to
minimize the errors come from marker movements during the gait [67].
Subject information such as weight and height were measured at the Al-
Warith prosthetic center. The subjects were asked about their amputation
causes, time laps and their trained prosthetic foot.
The subjects were fitted to the four types of the prosthetic feet and the
alignment procedure between their own sockets and the new prosthetic foot
was done. The alignment procedure consist three stages.
After that, the passive markers had been put on the seven joint landmarks
of the lower limb of each subject.
Some barefooted walking trials had been done on the walking path by
the subjects with their self-selected walking speed. The data were acquired for
the trials had a well-fitted foot on the force plate.
Finally, three walking trials data were collected for each subject and the
average of these data was calculated to minimize the errors.
Table 3-2 Passive markers specific locations in the left and right limbs.[61]
To compute the gait time, a stopwatch within the software was used
starting at the first heel strike and record the time at the second heel strike at
the same side, as shown in Figure 3-6.
Also, the step length can be measured, the step counts in a specific time
to find the cadence and stride frequency.
The kinematics system was used in gait analysis to record the angles of
the joint and location of and orientation of the body segments by placing
markers on the skin surface in locations that accurately represent the action of
the underlying joints. A camera registers these markers, and their sites
converted into motion data via motion program (Skill Spector).
One camera (25 frames per second) was used and positioned at right
angles with the plane of motion (2.8 m) afar, fitted on a tripod stand for the
height of the subject hip from the floor, it was used to make two-dimensional
measurements.
In the Skill Spector, the human lower limb is simplified into a model
composed of seven body positions hip, knee (left and right), ankle (left and
right), fifth metatarsal for the right foot, and first metatarsal for the left foot.
One of the digitized models for which side needed to digitize was chosen.
After that, the place of the moving markers in every single frame was marked.
Finally, the digitizing points were exported as a .txt format, so it’s ready to be
imported to MATLAB. Figure 3-8 showing marker digitizing procedure.
Figure 3-8 Skill Spector program operation: (a) selection of digitization mode; (b)
selection of markers' position; (c) export digitizing point.
Figure 3-9 Flow chart of kinematics parameter calculations.
3.5.5.2 Filtering of Gait Data
Figure 3-10 Padding procedure used on a signal before it enters Butterworth filtering.[61]
The AMTI force plate has six transducers component to measure the
upper surface displacement at all three axes when there is an applied force.
Figure 3-12 illustrates the force plate axes. The force plate’s six outputs
included the three force vector magnitudes (Fx, Fy, Fz), and three moments of
force based on the center of the platform (Mx, My, Mz) [57].
Then, it is passed to the computer unit where the Net force software
permits the user to observing and save data in real-time [68].
Figure 3-12 Coordinate system of a force platform.[57]
4.1 Introduction
When talking about the kinematics part, the lower limb joint angles (hip,
knee, and ankle) also plotted for both limbs with a percentage of gait after
taking the average of trials for each joint angle for both limbs.
Similarly, for the kinetic part, the average ground reaction forces (Fx, Fy,
and Fz) for the sound and the amputated limb had been plotted as normalized
values to the bodyweight with the percent of stance phase and compared with
the normal graphs.
Because this study assesses the difference in gait when the subjects
fitted to different types of feet, the spatiotemporal parameters of the gait had
been measured using the optical two-dimension way [22]. By using the
Kinovea software (Kinovea-0.8.15), spatiotemporal parameters were measured
manually from the videos recorded for both the amputated leg and the sound
leg.
As shown in Table 4-1, the Dynacity is the best one to give the highest
symmetry in the gait time between the amputated leg and the sound leg. While
the highest asymmetry was found in SACH foot, with 0.2 (s) difference for the
sound leg greater than the amputated leg. The lowest gait time was recorded
when the subjects use Dynastep foot with 1.4 (s) for each leg. This makes the
subjects to walk with greater cadence and speed comparing with other types of
feet.
For the stance time, the greatest difference between sound leg and
amputated leg is found at Dynastep foot with 0.12 (s) for the sound leg, in all
feet types the stance time for the sound leg is greater than the stance time for
the amputated leg.
Equal single support times between the legs were observed in Case-5 when
fitted to SACH foot, due to a long time has passed since the amputation (10
years), as shown in Table 3-1, and being well-trained over five years on using
SACH foot, before being fitted with his Echelon foot.
In Case-6, SACH foot also shows equal single support time for both legs.
Because Case-6 has a stiff ankle joint in the sound leg, Case-6 cannot move
his ankle joint in the sound side with its normal movement during gait (Dorsi-
flexion and Planter-flexion). For that, the sound foot had transferred through
the gait like a SACH foot.
For the SACH foot, despite Case-1 had 8 years of donning prosthesis with
SACH foot, more difference in the single support time was found in Case-1,
because of the long stump, the elder one between subjects, his level of
movement K1 and the rigidity of the SACH foot, as shown in Figure 4-1. So
Case-1 trained to no longer depend on the amputated leg during the gait and
tried to transfer the bodyweight to the sound leg more time during gait to get
more stable walking.
0.9
0.8
single support time (s)
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
1 2 3 4 5 6
sach foot sound side Dynastep sound side Trias1c30 sound side Dynacity sound side
.Figure 4-1 Single support time of the sound and amputee legs for six subjects
A significant difference between the amputated leg and sound leg in the
single support time appears when the subjects fitted to Dynastep foot, this
difference came from the greatest difference observed in Case-1 and Case-2.
That is because these two subjects had the greatest body masses, 107 and 105
kg respectively. So, the rubber heel compressed rapidly and released the
energy at the earlier time of gait making the subjects to transfer the
bodyweight to the sound leg for a longer time.
The greatest difference in single support time between the sound side
and the amputated side recorded for Case-1 when fitted to Trias 1C30.
The results suggested the Dynastep foot to be used for Case-6. Because
it is perfect for his situation with a sound limb ankle defect. The records had
shown a low single supporting time on the sound leg than the amputated leg.
This will help Case-6 in lowering the time of weight support on the defected
sound leg.
The Dynacity had the lowest difference in single support time between
the amputated leg and sound leg for Case-1. So, the foot is the best choice for
K1 level subjects, Case-1 needed some extra training time to be a K2 level
with a more stable walk without the need for a crutch.
The results of this study were consistent with previous studies where
below knee amputation subjects walking with reduced speed [69, 70].
High symmetry of step length between legs is found in SACH foot with
sound leg step length shorter than the amputated leg, matched with the finding
of Isakov e. et al [24]. Subjects had to depend on the curved shape of the
SACH planter during gait and on the hip flexor muscle to transfer the
amputated leg forward for a long step because SACH is an inactive foot and
can’t store energy that brings the body forward during the late stance. The
highest asymmetry of step length was found in Trias 1C30 of 44.23±5.14 cm
for amputee leg and 49.28±4.37 cm for sound leg [71].
The sagittal plane angular displacement for the sound and the amputated side
of each subject were plotted for the hip joint, the knee joint, and the ankle
joint. These joint angles were computed with equations 2.5, 2.6 and 2.7 for
ankle, knee, and hip joints, respectively.
For the subject case-1, Figures 4-2, 4-3 & 4-5 show the hip, knee, and
ankle angles for the amputated side and the sound side, respectively.
As shown in Figure 4-2, the sound side hip angle in the case of using
Trias1C30 had a flexion angle with 27º at the heel strike greater than that
when using the other feet types, creating an extension in the sound side-step
length.
At about 55% of the gait, at the pre-swing, the sound side hip angle shows the
maximum extension angle of 22º, then it reflexed gradually through the
terminal swing until the second heel strike.
This hyperextension 22º in the sound side causes lower back pain to Case-1
during the gait.
For both Dynastep and Dynacity feet, the sound hip angle curve shown
in Figure 4-2 has a symmetry pattern with 20º of flexion at heel strike. A
slower decrease in angle and extension at 55% for Dynastep, at 59% it
reaches 13º of extension, then reflexed again to the terminal swing.
When using the feet Trias 1C30, Dynastep, and Dynacity, the sound hip angle
curve has higher symmetry compared to the normal curve in Figure 2-7, but
in lower flexion angle peaks except for Trias 1C30 which has the greater
flexion peak.
Only SACH foot has the lowest sound hip flexion angle at the heel strike of 9º
and a lower extension angle of 6º.
For the amputated side, the hip angle pattern exhibit symmetry to the
normal pattern when using Dynastep and Dynacity feet.
The amputated side hip had flexed when the SACH foot was used,
which means Case-1 elongated his amputated step length.
Trias 1C30 shows late stance maximum extension of gait. Dynacity has
the highest symmetry to the normal pattern over the other type of feet but in
less flexion and extension angles.
For the knee joint angle, Figure 4-3 shows the angles during the gait for
the amputated side and the sound side.
For the sound side, the knee joint angles curve exhibited a normal curve
like when the subject uses Dynacity and Dynastep, with a double flexion
curve of 3º flexion at the beginning of heel strike and maximum knee flexion
10º at mid-stance, then it has fully extended at 45% of the gait (terminal
stance) finally to transfer the leg for the next gait, it is back to maximum
flexion at the mid-swing.
For the sound side knee angle, when using the SACH foot, the knee
began the gait with 5º of extension and at the mid-stance, the sound knee had
lower peak flexion with 5º, the maximum peak flexion at the mid-swing about
40º.
Trias1C30 had affected the sound knee angles with 16º of knee flexion at the
heel strike then extended slowly throughout the stance phase to the pre-swing
reaching 5º, and reflexed at the mid-swing with 30º.
For the amputated side, Dynastep, Dynacity, and SACH feet appear to
have normal curve patterns as shown in Figure 4-3 with initial contact of 11º
of flexion. While the knee is fully extended during the stance when the Trias
1C30 was used. This had appeared to cause pain in the knee joint during
ambulation because the foot was aligned in the plantar-flexion position. This
situation can be corrected by realigning the foot position.
Also, the prosthetic foot had affected the ankle angles during the gait.
Figure 4-4 shows the sound side and the amputated side ankle joint angles.
At the sound side, the more familiar angular displacement curve
compared to the normal curve was that when using Trias 1C30, which shows
small plantar-flexion at the loading response, dorsal flexion at the terminal
stance, and small plantar-flexion at the initial swing.
The other three feet had asymmetry in the sound side curves compared to
the normal curve, with no plantar-flexion, because these feet are more
flexible, which makes the sound side ankle do more dorsiflexion until transfer
the amputated side to the second heel strike.
For the SACH foot, the sound side appears to have 14º dorsiflexion
which is more than the normal value due to the shape of the SACH which
makes the subject to make longer step length.
For the amputated side shown in Figure 4-4, the Dynastep and Trias
1C30 had been aligned in plantar-flexion position which appears at the initial
contact, while SACH, which is a fixed foot, appears to has little dorsiflexion
at the mid-stance and it still throughout the gait.
Trias 1C30 has maximum dorsiflexion at the terminal stance for storing
the energy then reposition to the neutral at the swing phase.
Dynastep and Dynacity have the same kinematic curve pattern for the
amputated side.
Since the amputated side had no leg muscle function in the planter and
dorsal flexion, the recorded ankle angles result from the flexibility of the
prosthetic feet during the gait phases.
sound carbon sound sach
30 sound dynastep sound dynacity
sound hip angle (θ) in
20 Flexion
10
degrees
0
Extension
-10
-20
-30
(a)
25
20
amputee hip angle (θ) in
15
Flexion
10
degrees
-5
Extension
-10
-15
-20
(b)
Figure 4-2 Case-1 Hip joint angle (a) sound side (b) amputated side.
sound carbon sound sach
60
sound dynastep sound dynacity
50
sound knee angle (θ) in
40 Flexion
degrees
30
20
10
0
Extension
-10
(a)
amputee carbon amputee sach
60 amputee dynastep amputee dynacity
50
Amputee knee angle (θ)
40
Flexion
30
in degrees
20
10
0
Extension
-10
(b)
Figure 4-3 Case-1 knee joint angle (a) sound side (b) amputated side.
16
sound carbon sound sach
14
sound dynastep sound dynacity
12
Dorsalflexion
10
sound ankle angle (θ) in
6
degrees
0
Plantarflexion
-2
-4
(a)
amputee carbon amputee sach
20
amputee dynastep amputee dynacity
amputee ankle angle (θ)
Dorsalflexion
15
in degrees
10
0
Plantarflexion
-5
(b)
Figure 4-4 Case-1 ankle joint angles (a) sound side (b) amputated side.
For Case-2, the amputated side and the sound side hip angles shown in
Figure 4-5. He had a more symmetrical curve between the two sides when
using the Dynastep foot. Also, they are similar to the normal curve.
The Dynacity foot has both side symmetrical curve with slower hip extension
throughout the stance phase to the mid-swing until it re-flexed when reaches
the terminal swing.
For Case-2, the knee had osteoarthritis on the amputated side. Figure 4-
6 shows the knee joint angle for both sides. The Trias 1C30 shows rapid knee
extension at 45% of gait at the amputated side. This caused knee joint pain to
the subject when using the foot.
Also, Trias 1C30 affects the sound side knee to stay in a fully extended
position through the stance phase.
The more normal curve appearance is at the Dynastep foot with
maximum knee flexion 42º at the sound side and 52º at the amputated side.
Because Dynastep foot has an articular joint, so it makes 5 º planter-flexion at
heel strike at the amputated side and shows symmetrical ankle joint angles in
both sides as shown in Figure 4-7.
Dynacity show dorsiflexion through the gait because of
For subject Case-3, who has a sound leg ankle fixation plate, hence
unable to walk barefoot and needs to use shoes while walking to help him in
the process of attaching the joint and to relieve the pressure on the foot instep
while walking. Figures 4-8, 4-9 and 4-10 show the joint angles of the hip,
knee, and ankle, respectively.
Dynacity had a symmetrical pattern for the sound side and the
amputated side with 28º flexion at the initial contact in the amputated side and
24º at the contralateral side, 5º extension at the pre-swing in the amputated
side and 7º at the contralateral side, then back to the flexion position in the
swing phase.
Dynacity has the longest period of hip flexion-extension through the
stance extend from 0-77 % of gait.
40
sound carbon sound sach
sound dynastep sound dynacity
30
sound hip angle (θ) in
20
Flexion
degrees
10
0
Extension
-10
-20
(a)
35
amputee carbon amputee sach
30
amputee dynastep amputee dynacity
25
amputee hip angle (θ) in
20
Flexion
15
degrees
10
0
Extension
-5
-10
-15
(b)
Figure 4-5 Case-2 hip joint angle (a) sound side (b) amputated side.
80
sound carbon sound sach
70
sound dynastep sound dynacity
60
sound knee angle (θ) in
50
Flexion
40
degrees
30
20
Extension
10
-10
(a)
60
amputee carbon amputee sach
amputee dynastep amputee dynacity
50
amputee knee angle (θ)
40
Flexion
30
in degrees
20
10
Extension
-10
(
b)
Figure 4-6 Case-2 knee joint angle (a) sound side (b) amputated side.
The same thing for the Dynastep except for the sound side which has a
regular pattern in companion with Trias1C30 with maximum extension angle
at 55% equal to 7º. SACH foot had the maximum sound hip extension angle
at 55% equal to 15º and 11º at the amputated side.
For the knee joint angles shown in Figure 4-9, the sound side knee angle,
when using Dynastep and Dynacity, shows 20º knee flexion at the loading
response to mid-stance which is good for subject Case-3 to absorb the forces
at the initial contact. The second peek flexion is with 66º more than the
normal curve shown in Figure 2-7 at the mid-swing with Dynastep and
Trias1C30.
35
sound carbon sound sach
30 sound dynastep sound dynacity
25
Dorsalflexion
sound ankle angle (θ) in
20
15
degrees
10
0
Plantarflexion
-5
-10
(a)
12 amputee carbon amputee sach
amputee dynastep amputee dynacity
10
8
Dorsalflexion
amputee ankle angle (θ)
4
in degrees
-2
Plantarflexion
-4
-6
(b)
Figure 4-7 Case-2 ankle joint angle (a) sound side (b) amputated side.
SACH foot has the normal pattern at the sound side with lowest knee
flexion at the heel strike, but in the amputated side begin with knee flexion
flowed by extension at the terminal stance and reflexed at the mid-swing.
For the amputated side, the knee stays flexed with all the feet when the
subject wants to absorb the force during the gait.
40
sound carbon sound sach
sound dynastep sound dynacity
30
sound hip angle (θ) in
20
degrees
Flexion
10
0
Extension
-10
-20
(a)
35
amputee carbon amputee sach
30 amputee dynastep amputee dynacity
25
amputee hip angle (θ) in
20
Flexion
15
10
degrees
0
Extension
-5
-10
-15
(b)
Figure 4-8 Case-3 hip joint angle (a) sound side (b) amputated side.
80
sound carbon sound sach
70 sound dynastep sound dynacity
60
sound knee angle (θ) in
50
Flexion
40
degrees
30
20
Extension
10
-10
(a)
70
amputee carbon amputee sach
60 amputee dynastep amputee dynacity
50
amputee knee angle (θ)
Flexion
40
in degrees
30
20
10
Extension
-10
(b)
Figure 4-9 Case-3 knee joint angle (a) sound side (b) amputated side.
At the ankle joint, the sound side angle shown in Figure 4-10 has the
maximum plantar-flexion with 2º at Dynacity in the loading response, then
prolonged dorsiflexion until reaching 13º maximum peak at 47% of gait with
SACH. When using SACH, the sound side kept on dorsiflexion through the
swing phase.
The good characteristic of sound side ankle joints appears at Dynacity, it
has prolonged dorsiflexion during the stance phase, and this helps the subject
to slowly do the dorsiflexion motion at the sound foot without making it
painful for him.
15 sound carbon sound sach
sound dynastep sound dynacity
10
Dorsalflexion
sound ankle angle (θ) in
5
degrees
0
Plantarflexion
-5
-10
(a)
14
amputee carbon amputee sach
amputee dynastep amputee dynacity
12
amputee ankle angle (θ)
Dorsalflexion
10
8
in degrees
2
Plantarflexion
-2
(b)
Figure 4-10 Case-3 ankle joint angle (a) sound side (b) amputated side.
In the amputated side shown in Figure 4-10, the Trias1C30 (carbon foot)
and SACH have no plantar-flexion. The SACH has the lowest dorsiflexion at
the terminal stance because of its stiffness. At the loading response, the
maximum plantar-flexion is for Dynastep, because it has an articular joint.
The maximum dorsiflexion is for Dynastep in the terminal stance phase, the
energy-storing period, more than the normal by 2º.
Dynacity has a prolonged and ideal dorsiflexion angle 10º similar to the
normal curve in Figure 2-7.
At the terminal stance, the amputated side does not record any planter
flexion because of the absence of the leg muscle.
The hip angles, for the sound side and amputated side, for subject Case-4
is shown in Figure 4-11.
The sound side was affected according to the type of foot, similar to the
normal curve that appeared when using the SACH foot. For the remaining
three feet, the effect was the same. It appeared to be in the flexion position at
25º and gradually decrease during the stance period with maximum hip
extension at 15º with Trias1C30 at the mid-swing.
For the amputated side, the similar to normal curve of hip angle was
found at the Trias1C30 with initial contact flexion angle 22º, terminal stance
extension angle 12º, and flexion with 25º at the mid-swing.
The rest prosthetic feet shows the same hip angles during the gait, with
initial contact flexion angle 15º-17º then extend gradually until reaching 10º
of hip extension during the mid-swing.
The amputated side and sound side for subject Case-4 knee joint angles
are plotted in Figure 4-12.
As shown in Figure 4-12 (a) and (b), Trias 1C30 shows knee angle
symmetry for the amputated side and sound side, with maximum knee flexion
at the sound side reaching 53º and 48º at the amputated side.
The Dynacity foot was the only foot at the amputated side has to show
knee flexion in the mid-stance equal to 15º, this knee flexion helps the subject
to absorb the forces during the initial to the mid-stance, so the gait with this
foot is more comfortable for him.
25
20
Flexion
sound hip angle (θ) in
15
10
degrees
-5
Extension
-10
-15
-20
(a)
Flexion
20
15
degrees
10
5
0
-5
Extension
-10
-15
-20
(b)
Figure 4-11 Case-4 hip joint angle (a) sound side (b) amputated side.
sound carbon sound sach
60
sound dynastep sound dynacity
50
sound knee angle (θ) in
40 Flexion
30
degrees
20
10
0
Extension
-10
(a)
50
amputee knee angle (θ)
40
Flexion
in degrees
30
20
10
0
Extension
(b)
Figure 4-12 Case-4 knee joint angle (a) sound side (b) amputated side.
The ankle joint angles for subject Case-4, for the amputated side and the
sound side, are shown in Figure 4-13. The sound side ankle shows plantar-
flexion with SACH and Trias1C30 at the mid-stance. While Dynastep and
Dynacity do not show sound side plantar-flexion because of the high forward
pushing force from the amputated side.
On the amputated side, the only foot has a plantarflexion during the
initial contact is the Dynastep foot about 3º because it has an articular joint.
When the subject at the heel strike phase, the foot was plantar flexed then
dorsiflexed to 5º at the terminal stance.
Dynacity has the maximum dorsiflexion angle of 9º at the terminal
stance which is the phase for energy storing. Because of that, the subject
changes his foot to the Dynacity at the end of the test.
25
Dorsalflexion
sound ankle angle (θ) in
20
15
degrees
10
0
Plantarflexion
-5
-10
(a)
amputee carbon amputee sach
amputee dynastep amputee dynacity
10
Dorsalflexion
amputee ankle angle (θ)
6
in degrees
0
Plantarflexion
-2
-4
(b)
Figure 4-13 Case-4 ankle joint angle (a) sound side (b) amputated side
Hip joint angles for the sound side and the amputated side for the subject
Case-5 were plotted in Figure 4-14.
The graphs show a normal pattern but with lower flexion angles at the
sound side with 12º-17º at initial contact while the amputated side with 20º-
28º.
The two curves in Figure 4-14 do not show any abnormal pattern during
the gait.
15
Flexion
10
degrees
0
Extension
-5
-10
-15
-20
(a)
30
amputee carbon amputee sach
amputee dynastep amputee dynacity
20
amputee hip angle (θ) in
Flexion
10
degrees
0
Extension
-10
-20
-30
(b)
Figure 4-14 Case-5 hip joint angle (a) sound side (b) amputated side.
For the knee joint as shown in Figure 4-15 of Case-5, the maximum
flexion angle on the sound side appears when using Dynastep, but in general,
the sound side shows high symmetry to the normal curve especially when
using Dynastep.
At the amputated side, as shown in Figure 4-15; all feet begin the gait
with knee flexion because the subject needs to start the gait with foot flat not
with heel strike to get more balance during the ambulation over the amputated
side.
60 sound carbon sound sach
sound dynastep sound dynacity
50
sound knee angle(θ) in
40
Flexion
30
degrees
20
10
Extension
-10
(a)
50 amputee carbon amputee sach
amputee dynastep amputee dynacity
40
amputee knee angle (θ)
Flexion
30
in degrees
20
10
Extension
-10
(b)
Figure 4-15 Case-5 knee joint angle (a) sound side (b) amputated side.
At the ankle, the joint angles show different curves when using different
feet for the sound and amputated side as shown in Figure 4-16.
In the sound ankle, the curves when using Dynacity and Trias1C30 are the
closest curves to the normal with small plantar-flexion at the mid-stance 3º-5º,
dorsal flexion 12º-13º at the terminal stance, and a plantar-flexion at heel off
and pre-swing phase. That is because of the action of the calf muscle for
pushing the body forward.
At the amputated side, Dynastep has the maximum first planter flexion
because it has an articular joint, so it gives the closer initial contact pattern to
the normal gait. The second planter flexion occurs with Trias1C30 and
Dynacity at the pre-swing phase because of the alignment and fixation of the
foot in the plantar-flexion position.
The Maximum dorsiflexion occurs with Trias1C30 of 7º at 40 % of gait.
For subject Case-6, sound and amputated side; hip, knee, and ankle joint
angles are shown in Figures 4-17, 4-18 and 4-19, respectively.
The hip joint curve at the sound side shows a normal pattern when using
Trias 1C30 and Dynacity feet. SACH and Dynastep have prolong decreasing
in hip flexion angle because of the subject with these two feet used the crutch,
which makes him walk in flex hip position.
On the amputated side only Dynacity shows a normal pattern.
Knee joint angles are also recorded for the sound side and amputated
side for the subject Case-6 as shown in Figure 4-18a. Trias 1C30 and
Dynacity have the normal pattern on the sound side, but Dynastep and SACH
feet have flexed sound knee through the stance phase because of the use of a
crutch.
For the amputated side, shown in Figure 4-18b, knee contracture has
been diagnosed because all feet appear knee flexion throughout the gait.
Dynacity gives a good sound side ankle joint angles curve as shown in
Figure 4-19a. On the amputated side, the foot exhibits some dorsiflexion
through the terminal stance.
20
Dorsalflexion
sound ankle angle (θ) in
15
10
degrees
0
Plantarflexion
-5
-10
(a)
amputee carbon amputee sach
14 amputee dynastep amputee dynacity
12
Dorsalflexion
amputee ankle angle (θ)
10
8
in degrees
6
4
2
0
Plantarflexion
-2
-4
-6
(b)
Figure 4-16 Case-5 ankle joint angle (a) sound side (b) amputated side.
30
Flexion
20
degrees
10
0
Extension
-10
-20
(a)
35
amputee carbon amputee sach
30 amputee dynastep amputee dynacity
25
amputee hip angle (θ) in
20
Flexion
15
10
degrees
-5
Extension
-10
-15
(b)
Figure 4-17 Case-6 hip joint angle (a) sound side (b) amputated side.
60
40
Flexion
30
degrees
20
10
Extension
-10
-20
(a)
amputee carbon amputee sach
amputee dynastep amputee dynacity
60
amputee knee angle (θ)
50
Flexion
40
in degrees
30
20
10
0
Extension
(b)
Figure 4-18 Case-6 knee joint angle (a) sound side (b) amputated side.
25
sound carbon sound sach
sound dynastep sound dynacity
20
Dorsalflexion
15
sound ankle angle (θ) in
10
degrees
5
Plantarflexion
-5
-10
(a)
30
amputee carbon amputee sach
25 amputee dynastep amputee dynacity
amputee ankle angle (θ)
20
Dorsalflexion
15
in degrees
10
5
Plantarflexion
-5
-10
(b)
Figure 4-19 Case-6 ankle joint angle (a) sound side (b) amputated side.
Table 4-2 Range of motion at lower limb joints for the amputated side and sound side.
range of motion
sound side amputated side
(deg)
subjects joints Trias1C30 SACH Dynastep Dynacity Trias1C30 SACH Dynastep Dynacity
hip 49.3 28.3 32.9 34.3 29.6 36 26.9 28.3
case-1 knee 32.9 44.5 50.6 53.9 45 40.3 44.5 46.6
ankle 9.4 14.6 12.8 12.8 17.8 15.9 11.9 9.3
hip 22.3 36 35.4 28.1 39.5 31.2 41.7 32.1
case-2 knee 31.5 57.4 57.9 41.3 33.5 42.4 52 38
ankle 18.8 14.9 17.1 31.2 11.8 9.1 12.2 10.9
hip 35 35 36.1 32.4 32.6 42.5 34.9 33.4
case-3 knee 55.5 52.9 56.6 52.7 39.8 50.8 43.5 46.5
ankle 16.1 13.4 15.7 14.2 9.8 6.3 12.2 10.6
hip 37.8 26.5 27.7 29.4 38.3 41.8 36.3 40.2
case-4 knee 49.7 36.2 43.6 48.4 42.9 33.9 45 38.1
ankle 13.4 15.4 19.8 24.4 8.8 1.8 7.5 9.2
hip 29.6 25.7 31.8 29.8 49.3 35.4 31.6 40.4
case-5 knee 45 45.8 47.5 45.9 32.9 38.1 29.1 32.6
ankle 17.8 14.3 18.6 17.2 9.4 13.7 14.1 10
case-6 hip 39 24.2 28.5 39.4 24.2 20.1 20.2 36.1
knee 36.3 27.9 33.2 58.1 36 19.3 24.7 26
ankle 13.7 17.4 21.1 28 26 10.9 8.9 10.5
120
Fz % of BW
100
80
60
40
20
-20
(a)
Amputated leg SACH Amputated leg Dynacity
Amputated leg Dynastep Amputated leg Trias1c30
120
100
Fz % of BW
80
60
40
20
(b)
Figure 4-20 Case-1 vertical ground reaction force (a) sound side (b) amputated side.
Sound leg SACH Sound leg Dynacity
Sound leg Dynastep Sound leg Trias1c30
15
10
Fy % of BW
-5
-10
-15
(a)
Amputated leg SACH Amputated leg Dynacity
Amputated leg Dynastep Amputated leg Trias1c30
15
10
Fy % of BW
-5
-10
-15
-20
(b)
Figure 4-21 Case-1 anterior-posterior ground reaction force (a) sound side (b) amputated
side.
-2
-4
-6
(a)
Amputated leg SACH Amputated leg Dynacity
Amputated leg Dynastep Amputated leg Trias1c30
4
3
Fx % of BW
2
1
0
-1
-2
-3
-4
-5
(b)
Figure 4-22 Case-1 medial-lateral ground reaction force (a) sound side (b) amputated side.
The ground reaction forces for the subject Case-2 had been plotted as a
percentage of the bodyweight with the percentage of stance phase for the
sound side and amputated side in Figures 4-23, 4-24 and 4-25.
The sound side vertical GRF in Figure 4-23 (a) shows a higher first peak
with SACH about 118 % of BW, but a lower extended peak with Dynacity
and a similar second peak when using different feet.
At the amputated side, the vertical GRF in Figure 4-23 (b) shows similar
curves to the normal for Dynastep and Trias1C30, but the two peaks are lower
in Trias1C30 with a late first peak at about 30% of stance.
Dynacity has a stepped first peak with some period of steady then rising
upward to make the first peak of about 107% of the BW then decrease to 90%
of BW and suddenly increase upward at late stance to give the push-off force,
the higher ones, and 115-117 % of BW, with Dynacity and Dynastep but
earlier in Dynastep than Dynacity.
SACH foot shows mid stance hump about 100% of BW, because, as
shown previously, the subject had no knee flexion with this foot at the mid-
stance to absorb the force which causes the middle peak appearance at the
mid-stance.
At the sound side, all feet have a clear shape of forward-backward force
with 10% BW braking and 15% BW push-off force greater than the
amputated side, as shown in Figure 4-24 (a).
The forward-backward force shows a maximum braking force of 10% of
BW at the amputated side with Dynastep foot and 12% of BW forward force,
as shown in Figure 4-24 (b). It has the most curve symmetry between the
sound side and the amputated side.
Trias 1C30 has the lowest braking force with 3% BW on the amputated
side as shown in Figure 4-24 (b).
The sound side medial-lateral forces shown in Figure 4-25 (a) have
appeared some lateral force at initial contact and small medial force in the
mid-stance, 1% of BW, then 4-6% of BW at toe-off.
At the amputated side in Figure 4-25 (b), all feet have appeared only
medial force with two expanded peaks during the gait, the maximum one with
SACH and the minimum with Trias1C30.
sound leg SACH sound leg Dynacity
sound leg Dynastep sound leg Trias1c30
140
120
Fz % of BW
100
80
60
40
20
(a)
140 amputated leg SACH amputated leg Dynacity
amputated leg Dynastep amputated leg Trias1c30
120
100
Fz % of BW
80
60
40
20
-20
(b)
Figure 4-23 Case-2 vertical ground reaction force (a) sound side (b) amputated side.
sound leg SACH sound leg Dynacity
20
sound leg Dynastep sound leg Trias1c30
15
Fy % of BW
10
-5
-10
-15
(a)
10
Fy % of BW
-5
-10
-15
(b)
Figure 4-24 Case-2 anterior-posterior ground reaction force (a) sound side (b) amputated
side.
sound leg SACH sound leg Dynacity
Fx % of BW
sound leg Dynastep sound leg Trias1c30
8
-2
-4
-6
(a)
4
Fx % of BW
-1
(b)
Figure 4-25 Case-2 medial-lateral ground reaction force (a) sound side (b) amputated side.
The subject Case-3 had done the test on the force plate with shoes at the
sound leg because he has an ankle joint fixation plate, the shoes have helped
him to tight the ankle joint during walking.
The vertical GRF for the sound side shown in Figure 4-26 (a) has the
maximum initial contact peak with Trias1C30 about 124% of BW, while the
rest feet have about 115% of BW which is lower than normal, and the keel off
hump is similar at all feet with a range of 114-119% of the BW, which is in
the normal range.
For the amputated side shown in Figure 4-26 (b), Dynacity and Dynastep
have the same vertical GRF, but the increasing of force slop in Dynastep is
more than the slop in Dynacity at the initial contact. This causes pressure in
the socket.
Trias1C30 has a maximum initial peak with 120% of BW, while SACH
has a spike rising about 122% of BW at the beginning of the mid-stance, these
two feet have greater mid stance force, which caused lower back pain for the
subject.
The forward-backward forces for the amputated and sound side, for all
types of feet, are represented in Figure 4-27.
On the sound side shown in Figure 4-27 (a), Case-3 shows 8% of BW braking
force and 8 % of BW propulsive force with Dynastep foot. The clear shape
was recorded at the sound side with all feet except SACH, which has the
lowest braking force.
At the amputated side in Figure 4-27 (b), the feet have shown a
propulsive peak force that came from the energy stored in the feet which was
released at the keel off phase, but with irregular braking force at mid-stance
with a small hump for all feet which represent the backward force between
the foot and the ground, the lesser one with Dynacity is 5% of BW.
The medial-lateral forces at the sound side with different types of feet of
Case-3 are shown in Figure 4-28 (a). A little effect on the pattern was
recognized, a maximum lateral force at the initial contact with SACH is 6%
of BW, the lowest with Trias1C30 is 1% of BW, and a maximum first medial
peak with Trias1C30 is 5% of BW. The second medial peak, with all feet,
shows regular shape and value.
At the amputated side, all feet have only a medial single peak through
the stance. The maximum one with Trias1C30 is about 9% of BW and a
smaller one with Dynacity is below 6% of BW.
100
Fz % of BW
80
60
40
20
(a)
100
Fz % of BW
80
60
40
20
-20
(b)
Figure 4-26 Case-3vertical ground reaction force (a) sound side (b) amputated side.
Sound leg SACH Sound leg Dynacity
10 Sound leg Dynastep Sound leg Trias1c30
8
Fy % of BW
-2
-4
-6
-8
-10
(a)
-5
-10
-15
-20
(b)
Figure 4-27Case-3 anterior-posterior ground reaction force (a) sound side (b) amputated
side.
The study suggests that subject Case-3 to use the Dynacity foot because the
forces graphs have shown a clear pattern with less effects on the sound side.
8 Sound leg SACH Sound leg Dynacity
Sound leg Dynastep Sound leg Trias1c30
6
4
Fx % of BW
-2
-4
-6
-8
(a)
8
Fx % of BW
-2
(b)
Figure 4-28 Case-3 medial-lateral ground reaction force (a) sound side (b) amputated side.
For the subject Case-4, at the end of the test, the subject decided to
change his prosthetic foot with Dynacity foot, the following graphs of the
forces will clarify the causes.
The vertical GRF shown in Figure 4-29 shows a clear pattern and high
symmetrical between the amputated side and the sound side with Dynacity
foot, with a lower mid stance force 80% of BW and a higher push-off force in
the amputated side comparing with other types.
The sound side push-off force has a high similar shape to the normal
with all feet because it depends on the forces come from the plantar-flexion
muscle.
The first small hump at initial contact in the sound side, with the use of
Dynacity, comes from the high pushing energy in the amputated side, which
causes some impact force at the heel strike of the sound leg.
100
80
Fz % of BW
60
40
20
(a)
100
Fz % of BW
80
60
40
20
(b)
Figure 4-29 Case-4 vertical ground reaction force (a) sound side (b) amputated side.
Sound leg SACH Sound leg Dynacity
Sound leg Dynastep Sound leg Trias1c30
10
8
Fy % of BW
-2
-4
-6
(a)
15
10
-5
-10
-15
-20
(b)
Figure 4-30 Case-4 anterior-posterior ground reaction force (a) sound side (b) amputated
side.
The sound and amputated side medial-lateral forces for Case-4 are
shown in Figure 4-31. Which indicates a low lateral force than the normal
about 3% of BW and two medial forces at the mid-stance and the terminal
stance.
Also, the amputated side shows one extended medial force through the stance.
Sound leg SACH Sound leg Dynacity
Sound leg Dynastep Sound leg Trias1c30
4
3
Fx % of BW
-1
-2
-3
-4
(a)
7
6
Fx % of BW
5
4
3
2
1
0
-1
-2
(b)
Figure 4-31 Case-4 medial-lateral ground reaction force (a) sound side (b) amputated side.
Vertical GRFs of subject Case-5, for the amputated and sound legs, were
plotted in Figure 4-32. The amputated side has initial contact effects
concerning the types of feet with a maximum hump at 13% of stance with
Dynastep equal to 66% of BW which is more than the normal value of 40% of
BW that had been shown in Figure 2-8 (a).
On the sound side, the vertical GRFs are nearly similar to each other.
The lowest one is when using Dynacity foot. This is favorable to decrease the
load on the sound limp joints. Also, the Dynacity has symmetrical vertical
GRFs for the sound and amputated side.
Sound leg SACH Sound leg Dynacity
Sound leg Dynastep Sound leg Trias1c30
160
140
Fz % of BW
120
100
80
60
40
20
(a)
Amputated leg SACH Amputated leg Dynacity
Amputated leg Dynastep Amputated leg Trias1c30
140
120
Fz % of BW
100
80
60
40
20
(b)
Figure 4-32 Case-5 vertical ground reaction force (a) sound side (b) amputated side.
The forward-backward forces, for the subject Case-5, are shown in
Figure 4-33. For the sound side, it shows a clear shape with lower braking and
propulsive forces than the normal shape.
For the amputated side, the maximum horizontal propulsive force of
about 12% of BW is with Dynacity and irregular braking force because of less
friction effect between the ground and the prosthetic feet at the heel strike.
-2
-4
-6
-8
(a)
Amputated leg SACH Amputated leg Dynacity
15 Amputated leg Dynastep Amputated leg Trias1c30
Fy % of BW
10
-5
-10
-15
(b)
Figure 4-33 Case-5 anterior-posterior ground reaction force (a) sound side (b) amputated
side.
The medial-lateral GRFs, of the subject Case-5, are shown in Figure 4-
34. For the sound side, the curve shows a normal pattern for all types of feet.
But at the amputated side, it is only a medial force with one extended hump
without lateral force at the initial contact. This means the subject moves
through the stance without foot adduction.
Sound leg SACH Sound leg Dynacity
Sound leg Dynastep Sound leg Trias1c30
4
3
Fx % of BW
-1
-2
-3
-4
(a)
Amputated leg SACH Amputated leg Dynacity
Amputated leg Dynastep Amputated leg Trias1c30
7
6
Fx % of BW
-1
-2
(b)
Figure 4-34 Case-5 medial-lateral ground reaction force (a) sound side (b) amputated side.
As mentioned previously, subject Case-6 is a K1 level amputee. He used
a crutch during the test. It affects the GRFs recorded as shown in Figures 4-
35, 4-36 and 4-37.
The sound side vertical forces, when using the four types of feet, have
irregular patterns with a very high hump at heel strike when using SACH foot
(the subject's foot). The clear sound side vertical GRF appears when using
Dynastep foot.
A small rise in the sound side force at the mid-stance has appeared with
all feet except Dynastep. This made this foot type is more compatible with the
subject by reducing the lower back pain and knee pain.
At the amputated side, the forces showed a big and single convexity
through the stance because the subject has knee contracture so he can't flex
the knee more to absorb the vertical force during the mid-stance.
Also, Dynastep has a regular rise in the vertical force at the initial
contact and a small decrease at mid-stance because of the subject with a long
stump. Dynastep is the only one having an articular joint, so it has a little
effect on the sound side and smooth rise in the amputated side.
All the vertical forces recorded are below the bodyweight, because some
of the bodyweight was transferred to the crutch during the stance phase.
For the forward-backward forces, the curve shows a symmetrical pattern
between the amputated and sound side for all types of feet. With high braking
force for SACH and Trias1C30 is 10% BW at the sound side and is 13% BW
at the amputated side as shown in Figure 4-36.
The horizontal pushing forward force at the sound side is similar, with
6% of BW, despite the foot type in the contralateral leg.
The medial-lateral forces of the sound side, shown in Figure 4-37 (a),
have no lateral forces because of the use of the crutch which supports the
body medially during walking causing an increase in the medial force in the
amputated side over the sound side.
120
100
Fz % of BW
80
60
40
20
(a)
Amputated leg SACH Amputated leg Dynacity
Amputated leg Dynastep Amputated leg Trias1c30
90
80
70
Fz % of BW
60
50
40
30
20
10
(b)
Figure 4-35 Case-6 vertical ground reaction force (a) sound side (b) amputated side.
Sound leg SACH Sound leg Dynacity
Sound leg Dynastep Sound leg Trias1c30
10
Fy % of BW
-5
-10
-15
(a)
10
-5
-10
-15
(b)
Figure 4-36 Case-6anterior-posterior ground reaction force (a) sound side (b) amputated
side.
Sound leg SACH Sound leg Dynacity
Sound leg Dynastep Sound leg Trias1c30
5
4
Fx % of BW
-1
-2
(a)
Amputated leg SACH Amputated leg Dynacity
Amputated leg Dynastep Amputated leg Trias1c30
10
Fx % of BW
-2
-4
(b)
Figure 4-37 Case-6 medial-lateral ground reaction force (a) sound side (b) amputated side.
The results summarized in Table 4-3 for the Fz1 and Fz2 show an increase at
Fz1 in the vertical GRF at the sound side when using SACH foot, Dynastep
and Trias 1C30 by 6%, 12% and 19% respectively over the amputated side
except for the Dynacity foot the first peak in the sound side is lower than the
amputated side by 1.7%.
At Fz2 the high asymmetry between the amputated side and sound side found
at SACH foot by 43% lower than the sound side.
And high symmetry with Dynacity with 3% lower than the sound side which
means that Dynacity has the ability to store and release the energy to mimic
the sound side calf muscle to give the body the push up force for ambulation.
Table 4-3 Vertical ground reaction forces at Fz1 and Fz2 as % of BW.
Sound side Amputated side
Forces Dyna- Dyna- Trias Dyna- Dyna- Trias
Subject SACH SACH
(% BW) city step 1C30 city step 1C30
Fz1 111 105 107 103 111 104 105 105
case-1
Fz2 120 107 106 107 106 111 108 108
Fz1 118 100 104 97 109 107 112 104
case-2
Fz2 119 114 119 116 106 115 117 109
Fz1 111 115 114 124 123 111 109 119
case-3
Fz2 109 117 112 113 113 114 115 109
Fz1 97 108 97 10 103 110 94 106
case-4
Fz2 100 105 105 63 100 107 100 95
Fz1 122 112 127 133 112 110 115 106
case-5
Fz2 12 114 119 59 105 107 106 111
Fz1 98 76 72 79 75 75 80 80
case-6
Fz2 85 75 72 77 70 74 74 78
CHAPTER FIVE
Conclusions and Recommendations for Future Works
5.1 Conclusions
2- Also, the study obtained that subjects' weight affected the single support
time parameter when fitted to Dynastep foot. The difference is greater at
heavy subjects and is lower with lightweight subjects.
3- For joint angles, the types of prosthetic feet have no effect on the hip angle
for the amputated side and sound side. All feet have symmetrical hip angles
curves. Trias 1C30 foot only affects the sound side knee angles curve by
reducing the knee flexion angle at the initial stance. Ankle joint angles
appear high asymmetrical curves especially for SACH foot and Dynacity
foot, while Trias1c30 had less effect on the sound side ankle joint angles.
4- The results show an increase in the sound side vertical GRFs, especially at
the period of initial contact, when using SACH foot, Dynastep and Trias
1C30 by 6%, 12% and 19% respectively over the amputated side except for
the Dynacity foot the first peak in the sound side is lower than the
amputated side by 1.7%. Because of the fear of the amputee to support his
body on the amputated side for that Dynacity is the best foot in lowering the
forces at the sound side.
5- Prosthetic feet had fewer effects on the forward-backward component of
the GRFs on the sound side, but at the amputated side, the propulsive force
depends on the energy stored in the prosthetic foot materials.
Appendix A
Statistical analysis
T-Test
Independent statistical analysis for two group (amputated and sound side) for
each prosthetic foot.
As shown in figure 1 the gait time in V2 was selected as test variable and the
limb side in V2 was selected as grouping variable.
Notes
Comments
Filter <none>
Weight <none>
A- 1
N of Rows in Working Data
12
File
Missing Value Handling Definition of Missing User defined missing values are treated
as missing.
/MISSING=ANALYSIS
/VARIABLES=V2
/CRITERIA=CI(.95).
[sach foot]
Group Statistics
Equality of Equality of
Variances Means
F Sig. t df
assumed
A- 2
Equal variances
.153 9.909
not assumed
95%
Confidence
Interval of the
Difference
Mean Std. Error
Upper
A- 3
T-Test
Notes
Comments
Filter <none>
Weight <none>
Missing Value Handling Definition of Missing User defined missing values are treated
as missing.
/MISSING=ANALYSIS
/VARIABLES=V2
/CRITERIA=CI(.95).
A- 4
[dynastep]
Group Statistics
F Sig. t df
V2 Equal
variances .281 .608 .025 10
assumed
Equal
9.84
variances not .025
0
assumed
95%
Confidence
Interval of the
Difference
Mean Std. Error
Sig. (2-tailed) Difference Difference Lower
A- 5
V2 Equal variances .0022222222 .0903928326 -.1991855601
.981
assumed 22222 66815 83803-
Upper
T-Test
Notes
Comments
Filter <none>
Weight <none>
Missing Value Handling Definition of Missing User defined missing values are treated
as missing.
A- 6
Syntax T-TEST GROUPS=V1(1 2)
/MISSING=ANALYSIS
/VARIABLES=V2
/CRITERIA=CI(.95).
[dynacity]
Group Statistics
F Sig. t df
V2 Equal
-.017
variances .015 .906 10
-
assumed
Equal
-.017 10.0
variances not
- 00
assumed
A- 7
Independent Samples Test
95%
Confidence
Interval of the
Difference
Mean Std. Error
Sig. (2-tailed) Difference Difference Lower
Upper
T-Test
Notes
Comments
Filter <none>
Weight <none>
A- 8
N of Rows in Working Data
12
File
Missing Value Handling Definition of Missing User defined missing values are treated
as missing.
/MISSING=ANALYSIS
/VARIABLES=V2
/CRITERIA=CI(.95).
[carbon]
Group Statistics
F Sig. t df
A- 9
V2 Equal
-.023
variances .010 .924 10
-
assumed
Equal
-.023 9.72
variances not
- 3
assumed
95%
Confidence
Interval of the
Difference
Mean Std. Error
Sig. (2-tailed) Difference Difference Lower
Upper
Explore
Notes
A- 10
Comments
Filter <none>
Weight <none>
/COMPARE GROUPS
/STATISTICS DESCRIPTIVES
/CINTERVAL 95
/MISSING LISTWISE
/NOTOTAL.
A- 11
Cases
Descriptives
Median 1.54666666666
6667
Variance .048
Minimum 1.22666666666
66666
Maximum 1.83999999999
99999
Range .613333333333
3333
A- 12
Interquartile Range .373333333333
3357
Median 1.33333333333
3333
Variance .028
Minimum 1.25333333333
33332
Maximum 1.62666666666
66667
Range .373333333333
3335
A- 13
Upper Bound 1.60487868671
1773
Median 1.40000000000
0000
Variance .034
Minimum 1.22666666666
66666
Maximum 1.73333333333
33334
Range .506666666666
6668
Median 1.50666666666
6667
Variance .030
A- 14
Std. Deviation .173521265640
928
Minimum 1.28000000000
00000
Maximum 1.73333333333
33334
Range .453333333333
3334
Tests of Normality
Kolmogorov-Smirnova Shapiro-Wilk
Gait time
A- 15
Normal Q-Q Plots
A- 16
A- 17
Detrended Normal Q-Q Plots
A- 18
A- 19
ONEWAY gait time BY foottype
/STATISTICS DESCRIPTIVES
/PLOT MEANS
/MISSING ANALYSIS
/POSTHOC=TUKEY ALPHA(0.05).
Oneway
Notes
Comments
Filter <none>
Weight <none>
A- 20
Split File <none>
Missing Value Handling Definition of Missing User-defined missing values are treated
as missing.
/STATISTICS DESCRIPTIVES
/PLOT MEANS
/MISSING ANALYSIS
/POSTHOC=TUKEY ALPHA(0.05).
A- 21
Descriptives
gait time
Minimum Maximum
1 1.2266666666666666 1.8399999999999999
2 1.2533333333333332 1.6266666666666667
3 1.2266666666666666 1.7333333333333334
4 1.2800000000000000 1.7333333333333334
ANOVA
gait time
Total .754 23
95%
Confidence
Mean
Interval
(I) foot (J) foot Difference (I-
type type J) Std. Error Sig. Lower Bound
A- 22
4 .0288888888 .1077583029 -.272719728
.993
88889 36952 093867-
Multiple Comparisons
Tukey HSD
1 2 .412719728093868
3 .410497505871645
4 .330497505871646
A- 23
2 1 .190497505871645
3 .299386394760534
4 .219386394760534
3 1 .192719728093868
2 .303830839204979
4 .221608616982757
4 1 .272719728093867
2 .383830839204979
3 .381608616982756
Homogeneous Subsets
gait time
Tukey HSDa
foot type N 1
2 1.408888888888
6
889
3 1.411111111111
6
112
4 1.491111111111
6
111
1 1.520000000000
6
001
Sig. .734
A- 24
a. Uses Harmonic Mean Sample Size =
6.000.
Means Plots
A- 25
A- 26
شكر و تقدير
ُ
ك الحم ُد رّبي على ه ِ
ذه َ إلنج ِاز هـذا َ
الب ْحث َفل َ َ َ ق ْ والعزيم ةَ والتوفي َ
اإلرادةَ َ
َ نحي
أح َم دهُ تَعالـــى َعلى َم َ ْ
النِ َعم أبداً مـ ــا َبـ ـ ـ ّ
ـقيت.
كما اتقدم بجزيل الشكر لالستاذ المساعد الدكتور محمد عبد الستار محمد لتقديمه النصح ومشاركته
ــذت َعلــى ْأي دي ِهم
تلمـ ُ ِّ ِ ِ ِ
ين تَ َ
كر َم ْوصول ل ُك ل أســات َذتـــي األفاض ل الذ َ الش ُ
بعد فَ ُ
الفاعلة التمام هذا العملَ .و ُ
رحل ِة
علم َعلى ْأي دي ِهم فــي َم َ رفت بِ التَ ِ
ين تَ َش ُ ِّ ِ راحـــ َل ِدراسـَــتي ِّ
الس يما أس ات َذتي الك رام ال ذ َ
فــي جمي ع م ِ
َ َ
ِ ناء واإلمتِ ِ
مات الثَّ ِ
أسمى َكِل ُ
األفاض ل في مركز الوارث فين
ظ َ كر المو َ
نانَ .وال ْأنسى ُش َ المــا ِجستير َفل ُكم ْ
وكذلك شكر خاص للمتطوعين الذين لوالهم لما تم هذا العمل.
زو َجتيَ ،ع َملي هـذا لَم ِ تزاز واإل ْف ِ ِ ِ واإلع ِ مل ِع ِ
بارات ُّ
الش ِ أج ِ
تقدم بِ ْ
تخار لوالدي َووال َدتي َو ْ ْ كر أود الَ َ
َكذلك ُ
ودعاِئ ُكم فَ ُشكراً لَ ُكم ِمن القَ ـ ـ ـْلب. يكتَمل لَوال مساعدتِكم و ِ
دعم ُكم ُ َ ُ َ َ
تعتبر مالءمة الطرف الصناعي للمبتور ع``امالً رئيس``يا ً في إع``ادة التأهي``ل ،وك``ذلك ن``وع الق``دم
االصطناعية ،وهي أحد العوام``ل ال``تي ت``ؤثر على عوام``ل المش``ي .ته``دف الدراس``ة إلى قي``اس العوام``ل
الزمانية-المكانية والحركية وردود افعال القوى لحاالت البتر تحت الركبة وكيف تت``أثر عن``دما يس``تعمل
األشخاص أقدام صناعية مختلفة .تم اختيار ستة ذك``ور` ب``ترت أط``رافهم من ج``انب واح``د أس``فل الركب``ة
ليستخدمو̀ا أربعة أنواع من األقدا̀م االص``طناعية (1S66 SACH foot Ottobock inc., 1C30
Trias carbon Ottobock inc., Dynastep Proteor group and Dynacity Proteor
)groupطلب منهم المش``ي بالس``رعة المعت``ادة على المس``ار المس``تقيم عموديً``ا م``ع الك``اميرا ،تم قي``اس
العوامل الزمانية-المكانية لكل من الساق المبتورة والساق الس``ليمة ي``دويً̀ا باس``تخدا̀م برن``امج Kinovea
أيض`ا ،تم اس``تخدام برن`امج تتب``ع الحرك``ة (
(برنامج )Kinovea-0.8.15من مقاطع الفيديو المس`جلةً .
)Skillspectorلقياس زواي̀ا مفاصل االطراف` السفلية ،بينما تم تس``جيل ق``وى رد الفع``ل األرض``ي من
منصة قوة شركة التكنولوجيا الميكانيكية المتقدمة ( .)AMTIأظه``رت النت``ائج ع``دم وج``ود ف``روق` ذات
داللة إحصائية في العوامل الزمانية-المكانية لـ( ( p = 0.05بين األن``واع األربع``ة لألق``دام الص``ناعية
للساق السليمة والمبت``ورة والمقارن``ة بين الس``اق الس``ليمة والمبت``ورة .تظه`ر` ف``رق ذو دالل``ة احص``ائية p
(Trias 1C30 <0.05ل```وقت االس```تناد على ط```رف واح```د عن```د االش```خاص عن```د اس```تعمال
.)Dynastepتمت مناقشة ك``ل زواي``̀ا االط``راف` الس``فلى عن``د المش``ي وق``وى رد فع``ل أرض``ية بش``كل
منفصل .لخصت هذه الدراسة إلى أن هناك تأثيرًا طفيفًا لنوع القدم االصطناعية على العوامل الزمانية-
المكانية لمس``توى` نش``اط األجس``ام ،K1-K3وك``ذلك` األجس``ام المدرب``ة جي`دًا ك``انت أق``ل ت``أثراً ،بالنس``بة
لحركية المفاصل وجدت الدراسة تأثيرًا طفيفًا في زوايا المفاصل في الط``رف الس``فلي لك``ل من الج``انب
المبتور والجانب السليم وزيادة قوة رد فعل األرض على الجانب الس``ليم اعلى من الج``انب المبت``ور م``ع
جميع أنواع االقدم الصناعية-المستخدمة.
وزارة التعليم العالي والبحث العلمي
جامعة النهــــــــرين /كلية الهندسة
(تحليل المشي لحاالت بتر أسفل الركبة باستخدام انواع مختلفة من االقدام الصناعية)
رسالة مقدمة
في
من قبل