Professional Documents
Culture Documents
Good Laboratory Practices (GLP) Part-1 - Pharmaguideline
Good Laboratory Practices (GLP) Part-1 - Pharmaguideline
• Dilutions:
Ø Use appropriate, cleaned & dried volumetric apparatus.
Ø Label the glassware, which is in use.
Ø Ensure that the pipettes and volumetric flasks are of Class A grade.
Ø Ensure that the tip of pipettes is proper and not broken.
Ø Use appropriate solvents for dilution.
Ø Wipe the pipette before inserting into the volumetric flask.
Ø Make the dilutions carefully & accurately.
Ø Do not use contaminated sucker/rubber bulb.
Ø Make up the volume of the solution when it is at room temperature.
Ø Follow the methods properly & crosscheck once.
• Weighing:
Ø Ensure that the balances are clean and calibrated.
Ø Check the spirit level of the balance before use.
Ø Balance weighing range & capacity should be considered before weighing.
Ø Take the weight once the balance is stabilized.
Ø During weighing the door of the balance & balance room should be closed properly
Ø Follow the SOP correctly.
Ø Avoid spillage.
Ø Do not forget to check cleanliness before & after weighing.
• Column care:
Ø Avoid any jerks or vibrations to the column.
Ø Always replace the column ferrules whenever the column is to be kept idle & fix end caps to avoid silica drying.
Ø Store the column in appropriate solvents suggested by manufacturer.
Ø Wherever specified, use proper guard columns.
Ø Use the column for particular Raw material/ product/ test whenever required.
Ø Before and after analysis wash the column with prescribed solvent.
Ø Use/ Attach the column in the same direction of flow provided by the manufacturer.
• HPLC
Ø Flush the complete system (without column) with warm water (40°) once in a week using a union instead of a column.
Ø If the system suitability is not achieved by the column then regenerate the column as per manufacturer’s instructions / SOP for column
generation.
Ø Purge seal wash line, injector and all the inlets of the pump with water before use.
Ø Sonicate all ports inlet filter with [water: methanol (50:50)] once in a week.
Ø Check the tubing/Syringe for air bubbles of all ports.
Ø Check the valves for precipitation & clogging.
Ø Clean all the inline filters, frits & check valve cartridges at least twice in a month.
Ø Before converting the system to normal phase wash with water/ Methanol/ IPA/ Diluent serially.
Ø After completion of the analysis, follow the reverse procedure to convert to reverse phase.
Ø Do not store system in solvents.
Ø For purging of system start with the lower flow and then stabilize system at the required condition.
Ø After analysis take backup of the data in the relevant location.
Ø Ensure that the system is calibrated before use.
Ø Before the start, ensure that respective solvents & inlets inserted are as per the requirements.
Ø After analysis, do not leave the system idle in harsh solvents like ACN, store the system in ACN: Water mixture (20:80).
Ø Avoid spillage on the system.
Ø Do not interchange any of the modules or parts among the systems even of the same make.
Also see:
Ankur Choudhary is India's first professional pharmaceutical blogger, author and founder of Pharmaceutical Guidelines, a widely-read
pharmaceutical blog since 2008. Sign-up for the free email updates for your daily dose of pharmaceutical tips.
Need Help: Ask Question