Analytical measurement
Transmitters
Liquiline CM14 168
Liquiline M CM42 170
Liquiline CM44x
Thiết bị thu, phân tích dữ liệu sensor 172
Memobase Plus CYZ71D 174

Pressure
Pressure
Smartec CLD18 176
Smartec CLD134 178
Assemblies
Flexdip CYA112/CYH112 180
Flowfit CPA240/CPA250 182
Flowfit CYA251 pH/ORP, oxygen, NIR 184
Cleanfit CPA450 186
Cleanfit CPA871 188
Cleanfit CPA875 190
Automatic cleaning and calibration

Flow
Flow
Topcal CPC310 192
pH sensors
Glass pH/redox electrodes (analogue and digital) 194
Tophit CPS441D/471D/491D 196
Ceramax CPS341D 198
Conductivity sensors
Condumax CLS15D/16D/21D 200
phân tích độ dẩn điện
Indumax CLS50D/54D 202

Temperature
Temperature
Memosens CLS82D 4-electrode conductivity sensor
đánh giá độ tinh khiết
204
nước
Turbidity and sludge blanket sensors
Turbimax CUS51D/52D 206
Turbimax CUS71D đo độ đục 208
Dissolved oxygen sensors
Oxymax COS22D 210
Oxymax COS51D/61D phân tích oxy trong nước chỉ tiêu DO 212
Analysers
Liquiline System CA80xx 214

Analytics
Analysis
TOCII CA72TOC 216
phân tích màu, đúng thì nó sẽ ra COD, BOD
Samplers
Liquistation CSF48 218
Liquiport CSP44 220
quỉ này để lấy mẫu tự động System Components
Recorders/&
Recorders

For further information, visit our website at: www.uk.endress.com


Theory&
Solutions
Services
168 Analytics
Niveau
Level

Liquiline CM14
Single-channel analytical transmitter for pH, conductivity or
amperometric dissolved oxygen.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Memosens digital technology The Liquiline CM14 single-channel,


• Easy to operate panel-mounted transmitter simplifies
Monstername

• Two contacts for limit contactor pH, conductivity or dissolved


• Second current output for oxygen monitoring by allowing
temperature advanced Memosens digital sensor
• Compact panel-mounted design technology to be incorporated
into straightforward applications.
Available as a complete system,
including sensors and cables, this
Analytics

great value digital transmitter is the


perfect solution for OEM applications
and simple monitoring solutions.
Componenten
Registratie/
System Components
Recorders &

Technical data
E-Direct

Measured variables : pH, conductivity, amperometric dissolved oxgen


Sensors : Digital
Housing : Panel mounted
Housing body : Polycarbonate
Front membrane : Polyester (UV resistant)
Reference temperature : 25°C
Protection : Front: IP65/NEMA 4X
Communication : 2 x 0/4…20mA active
Services &
Solutions
Theorie
Liquiline CM14 169

Level
Level
Pressure
Pressure
Dimensions
43.9 (1.73)

mm (inch)

41.9 (1.65)

48 (1.89)

Flow
Flow
16.7 118.6 (4.67) 96 (3.78)
(0.66) 151.8 (5.98)

Temperature
Temperature
90.9 (3.56)

88.9 (3.50)

Display and operating elements

Analytics
Analysis
Display and operating elements 1
1. LC display for measuring values and configuration
data
2. Status LED power connected
3. Status LED alarm function 2
4. Status LED limit contactor relay 1
5. Status LED limit contactor relay 2 3
6. Dot matrix display for measuring units and menu 4
positions
5
7. Operating keys 7
System Components
Recorders/&
Recorders

6 6 6

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
170 Analytics
Niveau
Level

Liquiline M CM42
Transmitters for pH/redox, conductivity and dissolved
oxygen.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• For all types of sensor, digital or Liquiline CM42 is easy to use and Available with either a corrosion-
analogue features a large plain text display resistant robust plastic housing
Monstername

• ATEX, FM, CSA and NEPSI showing the measured value in or stainless steel housing,
certified figures 28mm high for easy reading. Liquiline CM42 is suitable for both
• Predictive maintenance system With step by step menu-driven plant use and for integration into
detects when sensors must be operation, user error is virtually panels, in both safe and hazardous
cleaned, calibrated or replaced eliminated. Operation is simplified areas. Better still, it is developed in
• HART, PROFIBUS and with the ‘navigator’ button – a accordance with IEC 61508 and the
FOUNDATION Fieldbus twisting dial with an integrated international safety standard SIL2
Analytics

compatible ‘enter’ key. Parameter setting, for simple integration into safety
servicing, diagnosis and predictive instrumented systems.
maintenance are in plain text format
Componenten
Registratie/

– no bulky manuals required!


System Components

Technical data
Recorders &

Measured variable : pH/redox, conductivity, concentration, resistivity,


dissolved oxygen, temperature
Sensors : Analogue or digital: glass, ISFET and Pfaudler
E-Direct

Housing : Plastic or stainless steel


Temperature sensors : Pt100, Pt1000, NTC 30K
Reference temperature : 25°C
Repeatability : < 0.01
Signal on alarm : ≥ 21.5mA or ≤ 3.6mA
Protection : IP67 (NEMA 4X)
Communication : HART, PROFIBUS or FOUNDATION Fieldbus
Services &
Solutions
Theorie
Liquiline M CM42 171

Level
Level
Pressure
Pressure
Dimensions (mm)

165
150

110
134

165
150

110
134
177

Plastic housing Stainless steel housing


148

177
0 11 164
148

0 11
164

26
26
132.5
132.5 134.5
134.5
136.5
136.5 136.5
177

177

220 130 136.5

165
165
220

150
148

110
150

110
134
134
148
164

220
0 164

220 130

2
11

138

2
2
0 11

138
2

Flow
Flow
26 26

132.5 132.5
134.5 134.5
220 220 130 130 136.5 136.5
136.5 136.5
144 138 174 220
220
2

138
2

174

2
144

2
164 144

144
95
177

144

95
144

95
148

174
95

174

174 174
0 11

Temperature
Temperature
mm
144

144

95

mm130 mm
95

220
22

mm
174 10

10

95
22

95

138
174
2

mm mm
22

22

mm mm
144
10
10

Electrical connections
CM42 CPK10 Sensor
CM42 CPK9 / CYK71

Analytics
144

Analysis
BN GN
95

187
112 BN
131 187

221 222 220 219 111 113 112

WH 187 pH
188 WT
188

113 GN GN
197132

197
YE YE
112
134 133

111
198
198

WH
232 231

219 188
233187

220 BK WT
188

113 ϑ
198 197 234

mm 196 GN
22

YE
111
YE
197

219
187
System Components

BK
220
188
Recorders/&
Recorders

222 196

221 197

# BN: n.c. PE

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
172 Analytics
Niveau
Level

Liquiline CM44x
Multi-parameter analytical transmitter for use with
Memosens digital sensors.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• For safe area use Liquiline CM44x is suitable for use All sensors use the same Memosens-
• Quick and easy commissioning with pH, ORP, conductivity, DO, Liquiline standardised protocol for
Monstername

• Standardised intuitive operation turbidity, nitrate and chlorine sensors. total data and mechanical interface
• Easily expandable Available in its most basic form as compatibility and increased process
• Full diagnostic capability a single sensor input, single alarm safety. When used together, any
• Datalogger relay safe area solution, you can easily combination of different types of
• Expandable up to 8 channels expand the system with additional digital sensors, cables, holders and
inputs and alarms to something much assemblies results in a measurement
more powerful. Choose as little or as solution perfectly tailored to your
Analytics

much functionality as you need – it’s requirements. This modular concept


that simple. means that our devices are easily
adapted to different applications
Componenten
Registratie/

and requirements, giving you more


flexibility across your plant.
System Components
Recorders &

Technical data
Measured variable : pH/ORP, conductivity, oxygen, turbidity, nitrate,
chlorine, sludge blanket, SAC
E-Direct

Sensors : Digital
Protection : IP66/67 (NEMA 4X)
Communication : HART, PROFIBUS, Modbus RS485/TCP, Ethernet IP
Output : 0/4...20mA, HART
Response : t90 = max 500ms for 0...20mA increase
Ambient temperature : -20…+60°C
Housing : Plastic
Services &
Solutions
Theorie
Liquiline CM44x 173

Level
Level
Pressure
Pressure
Dimensions (mm)

H 5 6 7 8 I

14
C E

Ø4
D F G

162
A

1 2 3 4 B

Flow
Flow
Ø9
199 128

95
76

194

Temperature
Temperature
237 mm (inch)

Installation

300 (11.8)
170 (6.69)

Analytics
Analysis
80 (3.15)
270 (10.6)
170 (6.69)

170 (6.69)

170 (6.69)
80 (3.15)

320 (12.6)

System Components
Recorders/&
Recorders

8 x Ø 7 (0.28) mm (inch)

80 (3.15) 80 (3.15)
? (?)

? (?)

mm (inch) mm (inch)

Mounting plate Weather protection cover

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
174 Analytics
Niveau
Level

Memobase Plus CYZ71D


Measure, calibrate and document Memosens sensors with
one single tool.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Calibrate and measure up to four Memobase Plus is a unique software Operation is simple and intuitive, as
sensors simultaneously tool that allows simultaneous the user is guided through calibration
Monstername

• Fully traceable and documented measurement and calibration of and maintenance procedures and the
• Simple identification of sensor multiple pH/ORP, conductivity, software automatically identifies the
deterioration dissolved oxygen and chlorine sensors.
• Fully FDA 21 CFR Part 11 sensors in a controlled environment
compliant (optional) using your laptop or PC. This results Full traceability and documentation
in a simplified and optimised ensures you will comply with quality
maintenance and calibration assurance procedures. Slope and
Analytics

effort which is fully traceable and calibration history is included in


documented. the calibration document, enabling
simple identification of any sensor
Componenten
Registratie/

Calibrate and measure up to four deterioration. Finally a real answer


sensors simultaneously can be had for the old question of,
Memobase Plus effectively turns “How long will my sensor last?”
your computer into a calibration and By tracking your sensors via the
measuring station by connecting Memobase Plus database, the
the sensors to your PC via USB. Up calibration history will quickly show
System Components

to four sensors can be managed you in graphical format whether the


simultaneously in any combination. slope and zero points are still within
Recorders &

acceptable limits that you can set.


E-Direct

Technical data
Measured variables : pH/ORP, conductivity, dissolved oxygen, chlorine
Software functions : Measure, calibrate, sensors, reports
Services &
Solutions
Theorie
Memobase Plus CYZ71D 175

Level
Level
Fully FDA 21 CFR Part 11 compliant, protection ensure the integrity of be carried out easily and documented
Memobase Plus is even ideal for the calibration. ‘As found - as left’ as an optional part of all calibration
use in the highly controlled and measurement is also possible, to look procedures. All the information is
regulated life sciences industry. User at the difference in measuring quality stored in a local or central database,
administration levels with password before and after calibration. This can making it easy to share.

Measuring system

Pressure
Pressure
1. PC (not supplied) 1 2 3 4
2. USB hub (optional, not supplied)
3. 1 to 4 USB cables
4. 1 to 4 Memolink boxes
5. 1 to 4 CYK20 Memosens cables or CYK10 Memosens
process cables
6. 1 to 4 Memosens sensors

Flow
Flow
Temperature
Temperature
Memobase Plus connection diagram

1. Cable with mini USB plug


2. Cable with M12 plug

Analytics
Analysis
Memobase Plus dimensions (mm)

Dimensions of Memolink boxes (mm)

The Memolink boxes can be stacked on top of one 37 (1.46)


another with the ‘Power/Data’ LED still easily visible. 97 (3.82)
31.5 (1.24)
System Components
Recorders/&
Recorders
77.5 (3.05)

ata
/D

MEMO SENS
r
we
Po

Power / Data

mm (inch)

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
176 Analytics
Niveau
Level

Smartec CLD18
Hygienic conductivity measurement system for food &
beverage applications.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Compact design allows Designed for use in the food & beverage IP69K protection
installation in small pipes industry, Endress+Hauser’s Smartec Compact yet robust, the new
Monstername

• Made from highly durable PEEK CLD18 conductivity measurement Smartec offers a simple solution
• Suitable for CIP system offers real value for money. to straightforward conductivity
• IP69K protection With 3-A approval and a range of measurement. It features IP69K
• 3-A and FDA compliant hygienic process connections, the protection for high temperature and
CLD18 has been cleverly designed high pressure washdowns and can
in compliance with FDA regulations also withstand process temperatures
to provide reliable, cost-effective from -10ºC up to +130ºC. It’s great for
Analytics

conductivity measurement in various smaller pipe diameters making it the


food & beverage applications including most cost-effective choice for a range
product monitoring, phase separation of food & beverage applications such as
Componenten
Registratie/

and simple CIP applications. bottling plants, dairies and breweries.

Compact and cost-effective


Whilst the Smartec CLD18 has been
Technical data designed to be compact and cost-
effective, there’s been no skimping
System Components

Measured variables : Conductivity, temperature on features and functionality. Set-


Conductivity range : 200μS/cm...1000mS/cm (uncompensated) up couldn’t be easier as it is factory
Recorders &

Temperature : –10°C...+130°C calibrated; there’s also a choice of


Temperature measurement : Pt1000 plastic or stainless steel housings to
Output signal : 0/4…20mA (galvanically isolated)
E-Direct

suit your particular requirements.


Response time : Conductivity: t95 < 1.5s, temperature: t90 < 50s Two outputs, for conductivity and
Repeatability : Max 0.5% of measured value ± 5μS/cm ± 2 digits temperature, come as standard. Best of
Housing : Plastic or stainless steel all, Smartec CLD18 has surface mount
Sensor : PEEK technology for immunity against plant
Process connections : Stainless steel (316L) or PVC-U vibration – so even if your application
Seal : EPDM is susceptible to vibrations, your CLD18
Protection : IP69K will continue giving clear, reliable
readings throughout.
Services &
Solutions
Theorie
Smartec CLD18 177

Level
Level
Pressure
Pressure
Dimensions (mm)

A B
85 85

Flow
Flow
123

127
17.1

17.1
32.8
32.8

114.5 112

Temperature
Temperature
C D
123

127

5.2
53.3

53.3
69

69

Analytics
Analysis
20.5
E F
28 28
73.5

78

System Components

23
Recorders/&
Recorders

mm

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
178 Analytics
Niveau
Level

Smartec S CLD134
Inductive conductivity measurement with compact or
separate transmitter for the food industry.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Fast response time enabling safe The Smartec S CLD134 transmitter cleanliness is key, its robust stainless
and efficient phase separations including the CLS54 sensor conforms steel housing is easy to clean. It will
Monstername

• Hygienic stainless steel housing to EHEDG and 3-A requirements, even slot into your existing hardware:
• Designed according to ASME offering the highest level of sanitary the conductivity system is compatible
BPE guidelines safety. For maximum versatility, the with HART and PROFIBUS for simple
• Food-grade virgin PEEK body Smartec S transmitter is available integration.
• HART and PROFIBUS compatible with both compact and remote
housings for simple installation. Since
Analytics
Componenten
Registratie/

Technical data
Measured variable : Conductivity, concentration, temperature
Conductivity measuring
System Components

Range (recommended) : 100µS/cm…2000mS/cm (uncompensated)


Concentration measuring range : NaOH: 0…15%
Recorders &

HNO3: 0…25%
H2SO4: 0…30%
H3PO4: 0…15%
E-Direct

User 1 (to 4): 4 tables available with remote


parameter set switching
Temperature measuring range : -10...+125°C (sterilisation at 150°C, 60
mins maximum)
Temperature sensors : Pt1000 switchable to Pt100
Output signal range : 0/4…20mA, galvanically isolated
Temperature response time : t90 ≤ 26s
Services &
Solutions
Theorie
Smartec S CLD134 179

Level
Level
Pressure
Pressure
142

160 / 6.30
Ø7
225

Dimensions (mm)
225 / 8.86

CLD134 CLD134

Flow
Flow
Remote version 175 Compact version
142 225
95

160 / 6.30

176.6
Ø7 207
mm /inch
225

9
Ø

***
225 / 8.86 33
142

Temperature
Temperature
10.2

175

A B
95

CLD134, dairy pipe


95

fitting DIN11851,
DN50
mm /inch

11 4 Ø 51
0 0 Ø 65
28.5 mm
62.6
9 9
Ø Ø
58 58
78.5 78.5
10.2

Analytics
10.2

Analysis
C D

Pressure-temperature load curve

55.8
2.85 64
0 0 66.85

9 9
Ø Ø
System Components

58 58
78.5 78.5
Recorders/&

10.2 mm 10.2
Recorders

A = sterilisation time (max 60 min)

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
180 Analytics
Niveau
Level

Flexdip CYA112 and CYH112


Assembly and holder system for sensors in open basins,
channels and tanks.
PressureDruk
Flow

CYA12 CYH12
Flow
Analyse/ Temperature Temperatuur

• Optimum configuration of any The Flexdip CYA112 assembly and In addition to the standard G1 sensor
measuring point CYH112 holder forms a modular connection thread, adapters can be
Monstername

• Easy to install and maintain holding system for sensors used in used to match to other, different
• Stainless steel or PVC versions open basins, channels and tanks. For sensor threads and designs. A quick
in-situ measurement in water and screw connector allows sensors to be
wastewater plants, the assembly assembled without the need for cable
and associated sensor are directly twisting meaning that sensors with
immersed in the medium. The flexible an inductive Memosens plug-in head
CYH112 holder is suitable for fixing can be easily installed and replaced.
Analytics

on the floor, wall or directly on a rail.


Componenten
Registratie/

Technical data
CYA112 assembly
Material : Stainless steel or PVC
Connection angle : 0, 45°, 90°
System Components

Temperature : -20...+60°C
Pressure : Unpressurised
Recorders &

Length : 600...3600mm (in 600mm steps)


Sensor connection : NPT¾’’, G1’’, Pg 13.5, G¾’’, G1½”
E-Direct

Immersion pipe : Standard, chain or floater


CYH112 holder
Material : Stainless steel or plastic
Installation : Floor, wall or rail
Length : 500, 1000 and 1800mm
Immersion pipe connection : Chain (plastic), chain (stainless steel), cross
clamp, pendulum holder
Services &
Solutions
Theorie
Flexdip CYA112 and CYH112 181

Level
Level
Pressure
Pressure
Installation

max. 1200 (47.2)

Flow
Flow
max. 1200 (47.2)

Temperature
Temperature
mm (inch)

Maximum extension

lbs kg

Analytics
Analysis
22 10

14.3 6.5

11 5
8.8 4
System Components
Recorders/&
Recorders

0 mm
0 500 1000 1200
inch
19.7 39.4 47.2

The given values are valid for floor, wall and rail mounting. The maximum extension depends on the load
(weight of immersion pipe, assembly and cable).

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
182 Analytics
Niveau
Level

Flowfit CPA240/250
pH/redox flow-through assemblies with a maximum of
three electrode locations.
PressureDruk
Flow

CPA240-2 CPA240-3 CPA250-A


Flow
Analyse/ Temperature Temperatuur

• Bypass installation Flow-through assemblies CPA240 Benefits at a glance


• Direct installation in pipes and CPA250 are intended for direct • Simple installation and removal of
Monstername

• Calibration without removing mounting in/between process electrode holder


electrodes from holder pipelines. The assemblies can be used • Three electrode locations for
• Various process connections for a wide range of temperatures and electrodes and cleaning
• Vertical inflow option prevents applications. Vertical inflow prevents • Integrated potential matching
sedimentation of medium solids from being deposited in the pin in Hastelloy C4 or tantalum
assembly. They are used in many (analogue only)
applications, such as: • Flexible process connections
Analytics

• water treatment plants • Usable at high pressures and


• artificial fertiliser production temperatures (up to 10 bar and
• sugar industry 150ºC max.)
Componenten
Registratie/

• gas scrubbers • Drain screw for sampling


• petrochemical plants • Electrode connection area
protected by a cover
• Chemical cleaning may also be
retrofitted
System Components
Recorders &

Technical data
E-Direct

CPA240-2x CPA240-3x CPA250-A


Material : PVDF Stainless steel 1.4404 PP
Wet sections (PM) : HC4 or tantalum HC4 or tantalum 1.4539
O-rings : EPDM, Viton Chemraz or Fluoraz EPDM, Viton Chemraz or Fluoraz EPDM
Temperature : 0…120ºC -15…150ºC Unpressurised/80ºC
Pressure : 8 bar/50ºC 10 bar/150ºC 6 bar/20ºC
Electrode length : 120mm 120mm 120mm
Services &
Solutions
Theorie
Flowfit CPA240/250 183

Level
Level
Pressure
Pressure
Flow
Flow
Dimensions (mm)
CPA240 (-2 or -3) CPA250-A

Version
Version Version
Version
AA
AA
Version BA
BA
Version
AA BA
321)
321)

Temperature
Temperature
381)
381)
321)
cover

381)
177
cover

136.5
177

cover
136.5
(with cover
cover
177

136.5
cover
(with
(with

234
(with
(with

111.5
52

(with
111.5
52 52

PG 13.5
111.5

238
238 PG 9
238 119
119
119

22
179
179 PG 16
80

179

Version
Version Version
Version 38 94 90
AB.
AB.AC,
AC,AD
Version AD BB,
BB,BC
BCBD
Version BD 110
125
321)

AB. AC, AD BB, BC BD


321)

Analytics
Analysis
387)
321)

387)
cover
177
cover

387)
177

136.5
cover
136.5
cover

cover
177
(with

136.5
cover
(with

(with
(with

(with
52

111.5
52 52

111.5

(with
111.5

250
250
119
119
250
179 119
179
234

179
System Components

PG 13.5
Version
Version Version
Version
AE
AE
Version BA
BA
Version
Recorders/&
321)

PG 9 PG 13.5
321)

337)

AE BA
337)

Recorders
321)

337)
cover
177
(with cover

cover
192.5
177

22
(with cover
192.5

PG 16
cover
177

cover
192.5
80
(with

(with
(with

(with
67.5
67.5

38 94 90
67.5

110
250
250 75
75 125
250 135 75
135
135

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
184 Analytics
Niveau
Level

Flowfit CYA251
Flow assembly for nitrate/SAC, turbidity and oxygen sensors.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Simple pipe or wall mounting Flowfit CYA251 is a flow assembly


• Easy adaptation to your process: for nitrate/SAC, turbidity and oxygen
Monstername

process connections for almost sensors with a diameter of 40mm.


all pipe diameters available It allows optimum adaption to your
• Short sensor response times piping system thanks to numerous
thanks to low assembly volume available process connections. The
• Automatic or manual cleaning ruggedized material withstands
even aggressive media and harsh
environmental conditions and a back-
Analytics

pressure valve protects your process


from blowing out and increases
operational safety. Better still, with
Componenten
Registratie/

the automatic cleaning nozzle, you’ll


benefit from extended maintenance
intervals.
System Components
Recorders &

Technical data
E-Direct

Compatible sensors : CAS51D (SAC/nitrate); COS31, COS41, COS51D,


COS61(D) (oxygen); CUS31, CUS41, CUS51D (turbidity)
Temperature : 0°C…+50°C
Pressure : Max 6 bar at 20°C, 4 bar at 50°C
Typical flow values : 200l/h (oxygen), 100l/h (turbidity/UV)
Seal : EPDM
Weight : 1.5kg or 1.8kg (depends on version)
Services &
Solutions
Theorie
Flowfit CYA251 185

Level
Level
Dimensions (mm)

mm (inch)

Pressure
Pressure
Z
83 (3.27)

42
(1.65)

218 (8.58)
(3.11)
79

35 (1.38)
X

144 (5.67)

Flow
Flow
Sensor CAS51D CAS51D CAS51D
COS61D CUS31 COS51D
CUS51D
2 mm 8 mm 40 mm
COS61 CUS41 COS41
COS31
Z mm (inch) 468 (18.43) 471 (18.54) 477 (18.78) 284 (11.18) 326 (12.83) 625 (12.80) 332 (13.07)

Connections NPT ¾” Rp ¾ DN 25 ANSI 1” Hose D20 G1¼


X mm (inch) 83 (3.27) 83 (3.27) 83 (3.27) 50.5 (1.97) 137 (5.39) 61 (2.40)
Y mm (inch) 50.5 (1.99 50.5 (1.99) 50.5 (1.99) 115 (4.53) 50.5 (1.99 44.5 (1.75)

Temperature
Temperature
Process connections

A. External thread G1¼, PVC


B. Flange ANSI 1”, PVC A
C. Glue-in port DN 25, PVC
D. Hose D20, PVC
E. Internal thread Rp ¾”, PVC
F. Internal thread NPT ¾”, PVC

Analytics
B

C Analysis

D
System Components
Recorders/&
Recorders

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
186 Analytics
Niveau
Level

Cleanfit CPA450
Retractable assembly with ball valve for use with
pH/ORP and DO sensors.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• No process downtime Perfect for use across industry, the


• Simplified maintenance and Cleanfit CPA450 retractable assembly
Monstername

calibration simplifies maintenance procedures


• Integrated rinse and calibration by allowing the replacement of pH/
connections ORP or oxygen sensors while the tank
is full or under process conditions.
Sensor cleaning and calibration
can take place without any process
interruption!
Analytics
Componenten
Registratie/

Technical data
Immersion depths : Max 100mm, 250mm, 700mm (depending on
System Components

process connection)
Process connection : Full range including threaded and flanged
Recorders &

connections
Process pressure : Max 12 bar at 100°C (max 2 bar for manual
retraction, max 6 bar at 130°C)
E-Direct

Process temperature : -15...130°C


Weight : 2-10kg depending on the version
Immersion tube : Stainless steel
Rinse connection plugs : PVDF
O-rings : EPDM, Viton, Kalrez
Ball valve : Stainless steel
Ball valve sealings : PTFE
Services &
Solutions
Theorie
Cleanfit CPA450 187

Level
Level
Pressure
Pressure
Flow
Flow
Temperature
Temperature
Pressure-temperature load curve

p [psi ] p [bar]
A

Analytics
Analysis
174 12
145 10
11 6 8 B
87 6
58 4
29 2
System Components

T [°C ]
130
Recorders/&

-15 0 20 40 60 80 100 120 140


Recorders

T [°F]
266
5 32 68 104 140 176 212 248 284

Pressure-temperature load curve


A. Maximum process operating pressure (static), only for
completely installed assembly
B. Advised upper insertion/retraction pressure (functional)

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
188 Analytics
Niveau
Level

Cleanfit CPA871
Retractable assembly for pH/ORP, oxygen and NIR sensors.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Clean, sterilise or calibrate/ The Cleanfit CPA871 retractable Its intelligent functions prevent
adjust the sensors without assembly provides reliable pH/ any leakage of medium during
Monstername

interrupting the process. ORP, oxygen and NIR measurement operation, cleaning or calibration,
• Optional immersion chamber in sensors with a 12mm diameter offering optimum protection of the
eliminates problems due to and 225/360mm length. Manual or process and operating personnel.
sticky media. pneumatic, the assembly is easily The retractable assembly flexibly
• Wide variety of process installed in both vessels and pipes adapts to your application. Be it
connections and wetted and you can remove, clean, sterilise or long immersion depths in sticky
materials, even for corrosive calibrate/adjust the sensors without media, aggressive environments or
Analytics

media or hazardous areas. interrupting the process. hazardous areas, you choose the right
Cleanfit CPA871 guarantees the material and specification to suit your
highest operational safety in both application.
Componenten
Registratie/

standard and demanding applications.


System Components
Recorders &

Technical data
Parameters : pH/ORP, oxygen, NIR
E-Direct

Sensors : Gel, ISFET, KCl


Version : Standard or immersion
Drive : Manual or pneumatic
Housing : Stainless steel
Temperature : -10°C...+140°C (for all materials except PVDF and conductive
PVDF)
Pressure : Manual: 8 bar up to 140°C, pnuematic: 16 bar up to 140°C
Services &
Solutions
Theorie
Cleanfit CPA871 189

Level
Level
Pressure
Pressure
Flow
Flow
Temperature
Temperature
Measuring system
Measuring system (example)

Analytics
Analysis
1. Cleanfit CPA871 assembly
2. Measuring cable
3. Liquiline CM44x transmitter
4. Sensor 3

1 2
System Components
Recorders/&
Recorders

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
190 Analytics
Niveau
Level

Cleanfit CPA875
Retractable assembly to keep your process 100% sterile.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Dynamic sealing prevents Designed to meet the food & beverage installed in both vessels and pipes
contamination. and life science industry standards, and you can remove, clean, sterilise
Monstername

• Remove, sterilize and calibrate the Cleanfit CPA875 retractable or calibrate/adjust the sensors
the sensor while process is assembly provides reliable pH/ORP, without interrupting the process.
running. oxygen and NIR measurement in Fully certified (EHEDG and ASME
• EHEDG-certified assembly: sensors with a diameter of 12mm BPE) and made of FDA-recommended
process connection and service and 225/360mm in length. Manual materials, the CPA875 is the right
chamber. or pneumatic, the assembly is easily choice where hygiene counts!
Analytics
Componenten
Registratie/

Technical data
System Components

Parameters : pH/ORP, oxygen, NIR


Recorders &

Sensors : Gel, ISFET, KCl


Chamber : Single or double
E-Direct

Drive : Manual or pneumatic


Housing : Stainless steel
Surface roughness : Ra < 0.76μm or Ra < 0.38μm (optional)
Temperature : -10°C...+140°C
Pressure : Manual: 8 bar up to 140°C, pnuematic: 16 bar up
to 140°C
Certification : Pharmaceutics CoC, EHEDG, ASME BPE, FDA,
USP Class VI (optional)
Services &
Solutions
Theorie
Cleanfit CPA875 191

Level
Level
Measuring system

Single chamber measuring system


1. Cleanfit CPA875 assembly
2. Measuring cable
3. Liquiline CM44x transmitter
3
4. Sensor

Pressure
Pressure
1 2

Double chamber measuring system


1 2 3
1. Control unit
2. Measuring cable
3. Liquiline CM44x transmitter

Flow
Flow
4. Cleanfit CPA875 assembly
5. Sensor
6. Valve manifold
7. Limit position switch relay signal
8. Control signals (electric/pneumatic)
11/12. Inlet/outlet of inner service chamber
21/22. Inlet/outlet of front service chamber
31/32. Drive control
7
8
4

Temperature
Temperature
6 31 32
12 6
11
22
21
5

Sealing principle

Analytics
Analysis
Sealing principle
A. Double chamber in service position
B. Double chamber in measuring position
C. Single chamber in measuring position
D. Single chamber in service position
1. ‘Moving’ seals in the double chamber
2. ‘Fixed’ seals in the single chamber System Components
Recorders/&
Recorders

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
192 Analytics
Niveau
Level

Topcal CPC310
Fully automatic pH/ORP measuring, cleaning and
calibration system. Suitable for hazardous area use.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Low maintenance, even with Endress+Hauser’s Topcal S additional activation of two external
heavily soiled or aggressive measuring, cleaning and valves for steam sterilisation or for
Monstername

media calibration system makes fully using a spray flange at the assembly,
• In-process cleaning and automatic, economical pH and ORP is now available as standard, with
calibration: no need to remove measurement possible even in the multi-hoses easily connectable to
electrode aggressive process environments. the retractable assembly. However,
• High accuracy and With the CPC310, there is a the real cost-saving bonus is the
reproducibility significant reduction in the use of fully automatic inline cleaning and
• System status messages with cleaning agents and buffers as these calibration without the need for
Analytics

feedback to control room are pumped through individual pipes electrode removal so that minimal
• HART and PROFIBUS compatible to the new rinsing block. In this way, maintenance is required.
all media is present front end, directly
Componenten
Registratie/

at the retractable assembly. The

Technical data
System Components

Measuring system : CPG310 control unit, Mycom S CPM153


transmitter, rinse block, membrane pumps
Recorders &

Cleaning and calibration : Clean, clean sterilise, clean calibrate, clean sterilise
Programs calibrate
Measured variables : pH (analogue or digital),
E-Direct

ORP (analogue or digital),


temperature
Measuring range : pH: -2…16pH,
ORP: -1500…+1500mV/-300…+300%,
temperature: -50…+200°C
Current output : 0/4…20mA, max load 600Ω
Repeatability : 0.1% of measuring range
Services &
Solutions
Theorie
Topcal CPC310 193

Level
Level
Topcal S comes complete with both Three independent, industry- Automatic cleaning and calibration
fixed and user-definable programmes tested double-membrane pumps functions eliminate the problem of
for process optimisation. Early replace the revolver pump from the glass breakage and contamination,
warning functions guarantee previous generation, increasing the whilst the high reproducibility of
reliability and system performance, unit’s performance particularly for the calibration increases product
even in hazardous areas, thanks heavy-duty applications and where quality. The hygienic condition of the
to the predictive maintenance aggressive cleaning agents are used, system is maintained, as the system
sensor condition check and e.g. concentrated acids and alkalis. remains closed during calibration and
medium monitoring functions. sterilisation.

Pressure
Pressure
Flow
Flow
Connections

Temperature
Temperature
6
MEAS CAL

DIAG PARAM

1 2 3 7
?

4 5 8

Analytics
Analysis
12

13
14
15 11
9
10
System Components
Recorders/&
Recorders

Complete measuring system (non-Ex)


1. pH/ORP sensor 9. Membrane pumps with cleaner and buffer bottles
2. Cleanfit assembly 10. Superheated steam, water, cleaner (optional)
3. Mycom S CPM153 transmitter 11. Rinse block
4. Special pH measuring cable 12. Valve for rinse water control
5. Power supply/control cable 13. Electrical wiring
6. Power supply for Mycom S CPM153 14. Compressed air
7. Power supply for CPG310 control unit 15. Media (cleaner, buffer, superheated steam etc)
8. CPG310 control unit

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
194 Analytics
Niveau
Level

Glass pH/redox electrodes


A wide range of digital and analogue pH, redox,
reference and temperature electrodes.
PressureDruk
Flow

Digital Digital Compact CPS71D CPS16D


electrode electrode electrode
CPS11D CPS41D CPF81(D)
Flow
Analyse/ Temperature Temperatuur

• Memosens digital pH sensor The pH or redox electrode and the of combination electrodes with
saves process data directly in the reference electrode are at the heart integrated Pt100, Pt1000 or NTC.
Monstername

sensor of every pH or redox measurement.


• pH combination electrodes with The right choice is therefore very Application
integrated Pt100/Pt1000/NTC important in achieving reliable Our range of electrodes covers
make assembly and calibration measurement with acceptable virtually every pH application. There
simple operating time and minimum are applications in all industries;
• A wide range of electrodes maintenance. Information is provided from pH measurements in yoghurt, to
ensures the right choice for each to help you to select the right wastewater from flue gas scrubbing,
Analytics

application electrode. If in doubt, we would advise from demin water to extremely


you to seek advice from us when acid or alkaline products with high
making the selection. Endress+Hauser conductivity. There is a pressure
Componenten
Registratie/

electrodes are equipped with limit (15 bar max) and temperature
Memosens digital technology with limit (135ºC max). A complete set-
inductive hermetically-sealed up consists of a pH/redox electrode,
connectors (TOP68 also available). (usually) a temperature sensor, pH
The benefit of this connector is that measuring cable, assembly and a
it is watertight so that failures due measurement transmitter. Depending
System Components

to moisture no longer occur. In order on the application, it is worth


to simplify installation (wiring), considering an automatic cleaning
Recorders &

maintenance and calibration as much system.


as possible, we would recommend
E-Direct

in almost every case that you use Installation guidelines


a combination electrode (pH or In order to prevent the presence of
redox electrode and reference air in front of the glass diaphragm
electrode combined). Temperature of a pH electrode, it is necessary for
compensation is necessary for pH the electrode to be built in with the
measurement in order to guarantee small glass sphere at the bottom,
accuracy. This is why we have taken at a minimum angle of 15° relative
a further step towards increasing to the horizon (in most cases).
ease of use, by offering a range Inflow velocity should preferably
Services &
Solutions
Theorie
Glass pH/redox electrodes 195

Level
Level
be a maximum of approximately sensor and transmitter guarantees operation, and minimum/maximum
0.5m/s. Higher velocities can result interference-free measuring signals pH value and temperature.
in discrepancies, particularly at low and a dramatic simplification of Memosens sensors can be pre-
conductivities or with contaminated process technology. Memosens is the calibrated under laboratory
electrodes. (Process) assemblies first to treat the problems of plug-in conditions and then quickly
are required for simple and correct systems - not just the symptoms. It replaced at the measuring point –
installation. successfully eliminates thus eliminating the problems of
Electrodes must always remain issues such as moisture ingress, measuring points that are difficult
‘wet’. ‘Standing dry’ can lead to a leakages, corrosion, salt bridges, to access. The sensors identify

Pressure
Pressure
(temporary) loss of function. ground loops and handling via themselves within the application;
Calibration is carried out using an inductive non-contact plug-in each measuring point recognises its
buffers. Calibration frequency varies connection that offers complete individual sensor, which significantly
depending on the application and galvanic isolation. reduces downtime. Memosens also
the desired accuracy. The same offers a longer life because sensors
installation guidelines apply to redox The Memosens integrated can be checked and regenerated
electrodes as apply to pH electrodes. circuit board also offers a wealth by cleaning in the laboratory. The
of possibilities. For example, pre-calibration also means that
Memosens manufacturing data, current sensor calibration under soiled or moist
The Memosens digital pH sensors properties and sensor history can all conditions is unnecessary and
break new ground by enabling be recorded and used as the basis for cuts work in hazardous areas to a

Flow
Flow
important process data to be saved predictive maintenance. Calibration minimum.
directly in the sensor. The inductive, data is stored, along with other key
two-way data transfer between information such as overall time in

Temperature
Temperature
Glass electrode Pg13.5 NPT ¾” pH Redox Combined rH
CPS11D X X

Analytics
Analysis
CPS12D X X
CPS16D X X X
CPS41D X X
CPS42D X X
CPS71D X X
CPS72D X X
System Components

CPS76D X X X
Recorders/&

CPS91D X X
Recorders

CPS92D X X
CPS96D X X X
CPF81D X X
CPF82D X X

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
196 Analytics
Niveau
Level

Tophit CPS441D/471D/491D
Glass-free IsFET pH electrodes.
PressureDruk
Flow

CPS441 CPS471D CPS491


Flow
Analyse/ Temperature Temperatuur

• Long-term, stable pH Following the success of horizontally, allowing flush fitting


measurement Endress+Hauser’s Tophit glass- with the vessel wall for easy cleaning.
Monstername

• ATEX and FM approval free IsFET pH sensor in a wealth of All versions feature an integrated
• EHEDG certified applications, the range has been temperature sensor providing effective
• PEEK sensor body extended to include sensors specially temperature compensation and Tophit
• With liquid KCl reference optimised for extreme processes is even suitable for use in hazardous
electrode (CPS441), hygienic applications areas (ATEX and FM approval).
• Available with Memosens digital (CPS471) and for applications that are
technology prone to build-up (CPS491). The new CPS441 electrode features a
Analytics

liquid KCl reference, so it is ideal for


Endress+Hauser’s Tophit simplifies use in extreme processes e.g. products
inline pH measurement and achieves that have a high content of organic
Componenten
Registratie/

cost savings due to its high stability solvent and/or low conductivity
and consistently fast response times, applications.
even low temperatures. Unlike most
glass electrodes, Tophit with its The CPS471 is fully sterilisable. It is
robust PEEK design can be installed ideal for CIP processes when combined
with a Cleanfit retractable assembly,
System Components

offering a long service life. It conforms


to 3-A, EHEDG and FDA standards and
Recorders &

Technical data
is fully autoclavable - it withstands
Measuring range : 0…14 pH, -15°C…100°C (sterilisable 135°C for 1 hour) process conditions from -15°C to
E-Direct

Pressure : 10 bar at 100°C 135°C, for the maximum in hygienic


3 bar at 135°C performance.
Sensor shaft : PEEK, FDA approved
Seals : EPDM or perfluoro elastomer For media that is prone to build-
Diaphragm : ceramic up, the CPS491(D) is the ideal
Process connection : Pg 13.5 choice. As the first IsFET sensor on
Temperature sensor : Pt1000 the market with an open aperture
Surface roughness : Ra < 0.8µm reference junction, it is perfect for
water purification and wastewater
Services &
Solutions
Theorie
Tophit CPS441D/471D/491D 197

Level
Level
applications or in processes with corrosion, salt bridges, ground loops conditions and then quickly
quickly changing pH values and/ and handling via an inductive non- replaced at the measuring point –
or alternating temperatures and contact plug-in connection that thus eliminating the problems of
pressures. offers complete galvanic isolation. measuring points that are difficult
The Memosens integrated to access. The sensors identify
The Memosens digital pH sensors circuit board also offers a wealth themselves within the application;
break new ground by enabling of possibilities. For example, each measuring point recognises its
important process data to be saved manufacturing data, current sensor individual sensor, which significantly
directly in the sensor. The inductive, properties and sensor history can all reduces downtime. Memosens also

Pressure
Pressure
two-way data transfer between be recorded and used as the basis for offers a longer life because sensors
sensor and transmitter guarantees predictive maintenance. Calibration can be checked and regenerated
interference-free measuring signals data is stored, along with other key by cleaning in the laboratory. The
and a dramatic simplification of information such as overall time in pre-calibration also means that
process technology. Memosens is the operation, and minimum/maximum calibration under soiled or moist
first to treat the problems of plug- pH value and temperature. conditions is unnecessary and
in systems - not just the symptoms. cuts work in hazardous areas to a
It successfully eliminates issues Memosens sensors can be minimum.
such as moisture ingress, leakages, pre‑calibrated under laboratory

Flow
Flow
Temperature
Temperature
Dimensions (mm)
Tophit CPS441 Tophit CPS471/491
pH combination electrode
connects to CPK12 cable

1 2

Analytics
3 Analysis

6
120 / 225 / 360 / 425 *

4
System Components

Length:
Sensor head
120mm/
Recorders/&

225/ 1 KCI electrolyte connection


Recorders

2 TOP 68 plug-in head


3
360/ 3 Reference electrode
425 4 IsFET chip
5 Seal (EPDM or perfluoro elastomer)
6 Sensor shaft

12 4

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
198 Analytics
Niveau
Level

Ceramax CPS341D
Memosens digital sensor with pH-sensitive enamel.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Direct installation in container Designed for use in the food &


nozzles or piping beverage and life science sectors, the
Monstername

• Self-cleaning via the flowing Ceramax CPS341D digital pH sensor


medium meets highest demands of measuring
• Long-term stability accuracy, pressure, temperature,
• High corrosion resistance for sterility and durability. Offering
harsh applications continuous online measurement, it
can withstand the rigours of inline
CIP and SIP processes. What’s more,
Analytics

its steel substrate means it is ideal in


applications with solids or agitator
turbulence and the protective enamel
Componenten
Registratie/

coating offers corrosion protection –


even in acids.
System Components

Technical data
Recorders &

pH range : -2…+14 pH (depending on application)


Minimum conductivity : 50µs/cm
Process temperature : 0...140°C
E-Direct

Temperature sensor : NTC 30KΩ


Process pressure : 0…6 bar
Sensor shaft : Porcelain enamel metal substrate (PEMS),
chemically resistant and shock resistant
Other materials : Stainless steel
Process connection : Wide range including Triclamp and dairy fittings
Protection : IP68
Services &
Solutions
Theorie
Ceramax CPS341D 199

Level
Level
Pressure
Pressure
Flow
Flow
Temperature
Temperature
Dimensions (mm) Installation
112 (4.41)

Analytics
Analysis
295 (11.6)

0 ... 360°
System Components
Recorders/&
Recorders
120 (4.72)

12 (0.47) CPS341D can be installed in any position.

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
200 Analytics
Niveau
Level

Condumax CLS15D/16D/21D
Digital twin-electrode conductivity sensors.
PressureDruk
Flow

CLS15D CLS16D CLS21D


Flow
Analyse/ Temperature Temperatuur

• Hygienic process connections Condumax conductivity sensors are Memosens digital technology
for installation in pipes or flow the ideal solution for conductivity All sensors feature Memosens
Monstername

chambers measurement in both water and digital technology for improved


• Available with plug-in head or wastewater applications. measurement integrity and efficiency.
fixed cable The inductive non-contact measured
• Easy-to-clean electropolished CLS15D sensors are specially value transmission of Memosens
surfaces designed for the measurement of pure guarantees maximum process
and ultrapure water e.g. monitoring safety. An automatic error message
ion exchangers, reverse osmosis, chip is generated if the sensor fails or
Analytics

cleaning, electro-deionising, WFI the connection between sensor


(water for injection) and distillation. and transmitter is interrupted. As
a direct result, the availability of
Componenten
Registratie/

CLS16D has been specially designed the measuring point is dramatically


for hygienic applications (particularly increased.
the life science sector) offering
3-A, EHEDG, FDA and ASME BPE
approvals.
System Components

CLS21D features two coaxial


electrodes made of graphite offering a
Recorders &

large measuring range. The graphite


guarantees high chemical stability
E-Direct

and low polarisation.

All Condumax sensors are designed


for use with our Liquiline CM14,
CM42 and CM44x.transmitters.
Services &
Solutions
Theorie
Condumax CLS15D/16D/21D 201

Level
Level
Pressure
Pressure
Flow
Flow
Temperature
Temperature
Technical data
CLS15D CLS16D CLS21D
Measured variables : Conductivity, temperature Conductivity, temperature Conductivity, temperature
Measuring range : k = 0.01cm–1: 0.04...20μS/cm, k = 0.1cm–1: 0.04…500μS/cm k = 1cm–1: 10μS/cm…20mS/
k = 0.1cm–1: 0.1...200μS/cm cm
Temperature sensor : Pt100 (CLS15), NTC (CLS15D) Pt100 or Pt1000 (CLS16), NTC Pt100 (CLS21), NTC (CLS21D)
(CLS16D)
Temperature : -20…+100°C (CLS15, fixed -5°C...+120°C -20°C…+135°C
cable), -20...120°C (CLS15,

Analytics
Analysis
clamp and CLS15D)
Pressure : 12 bar at 20°C 12 bar at 20°C 16 bar at 20°C
Process connection : NPT ¾” thread and clamp Clamp 1”, 1½”, 2” according G1 or NPT 1” thread, DIN
1½” (CLS15, fixed cable), NPT to ISO 2852 (also suitable 11851, SMS, clamp 2”
½” thread and clamp 1½” for TriClamp, DIN 32676),
(CLS15, plug-in), NPT ¾” and Tuchenhagen Varivent N
½” thread and clamp 1½” DN50...125, Neumo BioControl
(CLS15D) D50
System Components

Material : Stainless steel Stainless steel Graphite


Surface roughness : Ra ≤ 0.8μm (Ra ≤ 0.4μm Ra ≤ 0.8μm (Ra ≤ 0.4μm Not stated
Recorders/&
Recorders

optional) optional)
Protection : IP67 (CLS15), IP68 (CLS15D) IP67 (CLS16, fixed cable), IP68 IP65 (CLS21, 4-pole plug-in),
(CLS16, plug-in and CLS16D) IP67 (CLS21, fixed cable), IP68
(CLS21D)
Certification : ATEX, FM, CSA ATEX, FM, CSA ATEX, FM
Hygienic approvals : None 3-A, EHEDG, FDA, ASME BPE None

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
202 Analytics
Niveau
Level

Indumax CLS50D/54D
Digital conductivity sensors for the chemical, food and life
science industries.
PressureDruk
Flow

CLS50D CLS54D
Flow
Analyse/ Temperature Temperatuur

• Non-contact: insensitive to CLS50D: The chemical resistance separation applications.


electrode soiling and polarisation of the PEEK or PFA sensor material CLS54D: The CLS54D conductivity
Monstername

• Fast response time for efficient means that the sensor can be used sensor is specifically intended for use
phase separation in virtually any chemical process in hygienic applications in the food,
• CLS50D for high temperatures up to 180°C, e.g. concentration pharmaceutical and biotechnology
• CLS54D for hygienic applications measurement of acids and alkalis, industries. Thanks to its food-grade
• ATEX, FM, CSA and TIIS certified product quality monitoring and phase virgin PEEK body, seamless design
and hygienic certificates, it meets the
exacting demands of these industries.
Analytics

Technical data The CLS54D is especially suitable for


phase separation of product/water
CLS50D
and product/product mixtures in
Componenten
Registratie/

Measuring range : 2µS/cm…2000mS/cm pipe systems, control of CIP (cleaning


Temperature : -20…180°C (CLS50), in place) processes in the return
-20...+125°C (CLS50D) channel, concentration control in
Pressure : 20 bar max (20°C) remaking of CIP cleaning agents,
Temperature sensor : Pt100, class A product monitoring in pipes and
Min pipe diameter : Minimum DN80 bottling plants and, leakage detection.
System Components

Material : PFA or PEEK


Protection : IP67/NEMA6 (installed with original sealing)
Recorders &

CLS54D
Measuring range (recommended) : 100µS/cm…2000mS/cm (uncompensated)
E-Direct

Temperature : -10…150°C
Pressure : 12 bar max (up to 90°C)
Temperature sensor : Pt1000, class A complies with IEC 60751
Temperature response time : t90 ≤ 26s
Min pipe diameter : minimum DN80
Material : virgin PEEK
Surface finish : Ra ≤ 0.8µm
Protection : IP68/NEMA6P
Services &
Solutions
Theorie
Indumax CLS50D/54D 203

Level
Level
Dimensions (mm)

CLS50D - A
G¾” B
CLS54D A
CLS54D

Pressure
Pressure
133.7 133.7 133.7 133.7/5.26

Ø 89.5
Ø 84
17.7
12.5 11 4/0.16
0 0
0 0
16 16 62.6
36.5 36.5

Flow
Flow
Ø6 mm Ø6
58 58/2.28
78.5 78.5/3.09

Varivent N Dairy pipe fitting


DN40-125 DIN118851, DN50

C
133.7/5.26 133.7/5.26
CLS54D: pressure-temperature load curve

Temperature
Temperature
55.8/2.20
2.85/0.11 64/2.52
0 0

58/2.28 58/2.28
78.5/3.09 mm/inch 78.5/3.09

Analytics
Analysis
A = sterilisation time (max 60 mins)

Chemical durability
Medium Concentration PEEK PFA Chemraz Viton
Sodium hydroxide 0 to 50% 20 to 100°C 20 to 80°C 0 to 150°C Not suitable
System Components

solution NaOH
Nitric acid HNO3 0 to 5% 20 to 60°C 20 to 60°C 0 to 150°C 0 to 120°C
Recorders/&
Recorders

0 to 40% 20°C 20 to 60°C 0 to 150°C 0 to 120°C


Phosphoric acid H3PO4 0 to 50% 20 to 60°C 20 to 60°C 0 to 150°C 0 to 120°C
Sulphuric acid H2SO4 0 to 2.5% 20 to 80°C 20 to 100°C 0 to 150°C 0 to 120°C
0 to 30% 20 20 to 100°C 0 to 150°C 0 to 120°C
Hydrochloric acid HCI 0 to 5% 20 to 100°C 20 to 50°C 0 to 150°C 0 to 120°C
0 to 10% 20 to 100°C 20°C 0 to 150°C 0 to 120°C

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
204 Analytics
Niveau
Level

Memosens CLS82D
Digital four-electrode conductivity sensor for hygienic
applications.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Wide measurement range Memosens CLS82D is a hygienic, with simple operation. It resists
• EHEDG-qualified aseptic digital sensor for applications corrosion and moisture, enables lab
Monstername

hygienic design, FDA-approved with widely varying conductivity calibration and facilitates predictive
materials, certified in accordance values such as phase separation, maintenance.
with 3-A. chromatography, CIP monitoring
• Fully sterilizable and in small pipes, fermentation and Ideal for life sciences
autoclavable ultrafiltration. Its wide range of Particularly in the life sciences
• ATEX certified for hazardous process connections and a small industry, conductivity must be
area use compact design ensures the perfect measured over a wide range; from
Analytics

• Logging of sensor-specific fit to your process. With Memosens average to high values in the process
data for easy traceability and digital technology, CLS82D combines to low values in rinsing processes
predictive maintenance maximum process and data integrity with ultrapure water. Memosens
Componenten
Registratie/

CLS82D four-electrode conductivity


sensor allows you to perform these
measurements reliably with a single
sensor. With its certified aseptic
design, this sensor boasts exceptional
Technical data reliability and precision and is also
System Components

suitable for installation positions


Measured variables : Conductivity, temperature where space is restricted.
Recorders &

Conductivity range : 1µS/cm...500,000µS/cm


Temperature : –5°C...+120°C Reliable products and processes
Sterilization (max 45 min) : Max 140°C at 6 bar
E-Direct

A unique feature of the sensor is its


Pressure : 17 bar at 20°C, 9 bar at 120°C innovative sensor element made from
Surface roughness : Ra < 0.38μm ceramic with platinum electrodes.
Repeatability : 0.2% of reading The advantage of these materials is
Sensor : Platinum and ceramic their similar temperature expansion
Process connections : Stainless steel 1.4435 (AISI 316L) behaviour. Even during extreme
Seal : EDPM temperature changes the material
Protection : IP68/NEMA 6P (connector) bond stays tight so that no gaps occur
at any time.
Services &
Solutions
Theorie
Memosens CLS82D 205

Level
Level
Increased plant availability
The sensor’s plug & play capability
means that you can integrate the
Memosens CLS82D in your process in
the shortest possible time, simplifying
commissioning and maintenance.
In conjunction with the Memobase
Plus software, you can calibrate your
sensors in the laboratory under ideal

Pressure
Pressure
conditions for improved sensor and
quality management

Flow
Flow
Temperature
Temperature
Measuring system

Analytics
Analysis
A complete measuring system consists of the
following components:
1. Memosens CLS82D conductivity sensor
2. Liquiline CM44x transmitter
3. Memosens CYK10 measuring cable
1

2
System Components
Recorders/&
Recorders

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
206 Analytics
Niveau
Level

Turbimax CUS51D/52D
Turbidity sensors for the measurement of drinking
water, process water and wastewater.
PressureDruk
Flow

CUS51D CUS52D
Flow
Analyse/ Temperature Temperatuur

• Simple installation Turbimax turbidity sensors are Thanks to the large measuring
• Factory calibrated the ideal solution for accurate and range, however, the sensor can
Monstername

reliable turbidity measurement also be used in wastewater flow.


in both water and wastewater It is installed using an immersion
applications. assembly, a flow-through assembly
or directly in the pipe using an inline
Using the 90° scattered light method, retractable assembly and can be used
the CUS31 is used mainly for drinking effectively in applications such as
water and process water applications. checking filters in drinking water
Analytics

Technical data
Componenten
Registratie/

CUS51D CUS52D
Measuring principle : 90° nephelometric; 135° back scattered light; Nephelometric turbidity sensor (90°
four beam pulsed scattering) according to ISO7027

Measuring range : 0...4000 FNU/0…150g/l 0.000...4000 FNU/NTU


System Components
Recorders &

Wavelength : 860 ± 30nm 860nm


Optical compensation : Using reference photodiodes Using reference photodiodes
Accuracy : <2% of the measured value or 0.1 FNU < 0.5% of measured value (measuring range:
E-Direct

0 to 10 FNU)
Temperature/pressure : -5…50°C (max 80°C short term) / 0.5 to 10 bar -20...85°C/0.5 to 10 bar
Process connection : G1’’ / NPT ¾’’ G1” and NPT ¾”; 2” clamp (depends on
sensor version)/DIN 32676

Material : Stainless steel Stainless steel


Protection IP68 IP68
Services &
Solutions
Theorie
Turbimax CUS51D/52D 207

Level
Level
production, effluent and sediment The sensor is insensitive to medium avoid product loss. With Memosens
checking in physicochemical colour and daylight, thanks to digital technology, the CUS52D
purification processes (e.g. the modulating IR light sources. combines maximum process and data
flocculation), filter, cooling and Intelligent signal processing integrity with simple operation. It
surface water monitoring etc. compensates for contamination on enables lab calibration and simplifies
the windows and reduces the effect predictive maintenance.
The CUS51D is used mainly for of air bubbles and coarse particles.
wastewater and sludge applications
such as active sludge tank, The CUS52D is a smart sensor with

Pressure
Pressure
sludge return from district and lab accuracy that allows unattended
industrial wastewater purification, operation in all measuring points of
effluent and sediment checking your water production. Thanks to
in physicochemical purification its hygienic, self-cleaning design,
processes and sludge concentration you can mount it directly into your
measuring etc. pipeline. This way you will save on
extensive bypass installations and

Flow
Flow
Temperature
Temperature
Dimensions CUS51D (mm) Cleaning system: CUS51D
(0.79)
20
12 (0.47)
69 (2.72)

NP T ¾“

G1

Analytics
Analysis
14.5 (0.57)

40
(1.57)
194 (7.64)

System Components
Recorders/&
Recorders

mm (inch)

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
208 Analytics
Niveau
Level

Turbimax CUS71D
Ultrasonic immersion sensor for interface measurement.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Three different sensor models In many instances in process Turbimax CUS71D is a sensor for
allow optimal adaption to the engineering, suspensions are many applications of the interface
Monstername

measurement task. separated into their solid and liquid measurement:


• Simple commissioning thanks to components via sedimentation. To do • Wastewater treatment: primary
predefined calculation models. this economically and efficiently, it is clarifier, sludge thickener,
• Intelligent sensor: all necessary to continuously monitor the secondary clarifier
characteristics and calibration separation and transition zones of the • Water purification: settling basin
values are stored in the sensor. clarification and settling phases. after flocculant dosage, sludge
height in contact sludge process
Analytics

• Chemical industry: static


separation process
Componenten
Registratie/
System Components

Technical data
Recorders &

Measured variable : Interface (turbidity optional)


Measuring range : 0.3...10m (turbidity: 0...50 (200) NTU)
Max measured error : 35mm at 3m (turbidity: 1% of the measuring range
E-Direct

at 50 NTU)
Temperature : +1°C…+50°C
Pressure : 0...6 bar abs
Protection : IP68
Sensor material : ABS and epoxy plastic
Wiper material : Rubber
Optical window : Sapphire
Services &
Solutions
Theorie
Turbimax CUS71D 209

Level
Level
Dimensions (mm)
Standard sensor Sensor with wiper

Pressure
Pressure
mm (inch)

20 (0.79)
mm (inch)
20 (0.79)

NPT 3/4"

46 (1.81)
G1

NPT 3/4"

12 (0.47)
46 (1.81)

G1 34.7 (1.37)

Flow
Flow
12 (0.47)

34.7 (1.37)

110 (4.33)
135.5 (5.33)
68.8 (2.71)
43.3 (1.70)

19 (0.75)
74.3 (2.93)

Temperature
Temperature
74.3 (2.93)

Installation

A B

Analytics
Analysis
C
D

System Components

F

Recorders/&

Basin configuration
Recorders

A. Sensor
B. Minimum distance of sensor to
basin wall: 45cm
C. Reference point e.g. water surface
D. Zero point
E. Basin depth
F. Opening angle of ultrasonic cone: 6°

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
210 Analytics
Niveau
Level

Oxymax COS22D
Digital amperometric oxygen sensor for standard,
hygienic and sterile applications.
PressureDruk
Flow

COS22D
Flow
Analyse/ Temperature Temperatuur

• Maximum process safety through Oxymax COS22D digital sensor and life science industries in addition
non-contact inductive signal is specially designed for the to general process applications.
Monstername

transmission requirements of dissolved oxygen The COS22D also has application-


• Easy handling thanks to storage of measurement, offering long-term specific versions e.g. carbon dioxide
sensor-specific data stability and excellent repeatability. It compatible trace sensor for the
• Predictive maintenance possible is robust enough to withstand frequent beverage industry and a trace sensor
thanks to registration of sensor sterilisation and autoclaving so is for the power industry.
load data ideal for use in the food & beverage
Analytics
Componenten
Registratie/

Technical data
Measured variable : Dissolved oxygen, temperature
System Components

Response time : t90 < 30s, t98 < 60s


Repeatability : ±1%
Recorders &

Process temperature : -5°C...+135°C


Process pressure : 0...12 bar
Min flow rate : 0.02m/s (COS22D-*1), 0.10m/s (COS22D-*3)
E-Direct

Sensor shaft : Stainless steel, titanium, C22 Alloy


Electrode : silver/platinum
Sealing ring : Fluorelastomer (FDA), perfluorelastomer (USP)
Membrane : Silicone, PTFE
Process connection : Pg 13.5
Protection : IP68
Certification : ATEX, FM and CSA
Services &
Solutions
Theorie
Oxymax COS22D 211

Level
Level
Pressure
Pressure
Flow
Flow
Temperature
Temperature
Dimensions (mm)

COS22D
50 (1.97)

Analytics
Analysis
Pg 13,5

120 /225 /360 /420 (4.72 /8.86 /14.2 /16.5)*

System Components
Recorders/&
Recorders

12 (0.47)

mm (inch)

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
212 Analytics
Niveau
Level

Oxymax COS51D/61D
Digital oxygen sensors for water, wastewater and utilities.
PressureDruk
Flow

COS51D COS61D
Flow
Analyse/ Temperature Temperatuur

• Maximum process safety The Oxymax range of dissolved hazardous areas). Designed to
through non-contact inductive oxygen sensors provide continuous be low-maintenance with a long
Monstername

signal transmission measurement of the concentration of operating life, the sensors offer
• Easy handling thanks to storage dissolved oxygen in water to maintain outstanding value for money. Thanks
of sensor-specific data quality and ensure effective process to Memosens digital technology,
• Predictive maintenance possible control. Oxymax COS51D/61D combine
thanks to registration of sensor maximum process and data integrity
load data Oxymax COS51D/61D are reliable with simple operation. They resist
and highly accurate oxygen corrosion and moisture, enable lab
Analytics

sensors for all kinds of water & calibration and facilitate predictive
wastewater applications (including maintenance.
Componenten
Registratie/

Technical data
COS51D COS61D
Measuring principle : Amperometric sensor Fluorescence quenching optical sensor
Measured variable : Dissolved oxygen (mg/l, μg/l, ppm, ppb, % SAT, Dissolved oxygen (mg/l, % SAT, hPa),
System Components

hPa) temperature
Recorders &

Measuring range : 0.01...100mg/l, 0.00...1000 %SAT, 0...20mg/l (0...20ppm), 0...200 %SAT,


0...2000hPa 0...400hPa
Repeatability : ±1% ±5%
E-Direct

Response time : t90 < 3 minutes, t98 < 8 minutes (normal t90: 60s
response), t90 < 30s, t98 < 90s (fast response)
Process temperature : -5…50°C -5…60°C
Process pressure : Up to 10 bar Up to 10 bar
Process connection : G1 and NPT¾” G1 and NPT¾”
Protection : IP68 IP68
Certification : ATEX, FM and CSA Safe area only
Services &
Solutions
Theorie
Oxymax COS51D/61D 213

Level
Level
Pressure
Pressure
Dimensions (mm)
COS51D

Flow
Flow
190 (7.48)
171 (6.73)
144 (5.67)

Temperature
Temperature
Ø40 (1.57) mm (inch)

COS61D dimensions with optional M12 plug COS61D dimensions with optional cleaning unit
* depending on cleaning unit version

Analytics
Analysis
21 (0.83)
15 (0.59)
97 (3.82)

NPT 3/4"
47 (1.85)

G1
12 (0.47)

11 (0.43)
230 (9.06)

242 (9.53)

System Components
186 (7.32)

196 (7.72)

Recorders/&
Recorders
77 (3.04) / 66 (2,59)*

Ø40 (1.57)

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
214 Analytics
Niveau
Level

Liquiline System CA80xx


Colourimetric analysers for online monitoring in water
applications.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Easy upgrade to measuring The Liquiline System CA80 range Liquiline System CA80xx range
station with up to four digital of analysers offers high-precision includes:
Monstername

Memosens sensors. measurement in all critical control • Liquiline System CA80AM


• Advanced diagnostics with points. As member of the Liquiline ammonium analyser
remote access for increased platform, it enables plug & play and • Liquiline System CA80PH
process safety. easy upgrade to a measuring station orthophosphate analyser
• Reduced operating costs - minimising the installation effort. • Liquiline System CA80NO nitrate
through automatic, configurable Automatic calibration and cleaning analyser
calibration and cleaning as well as the low consumption of • Liquiline System CA80CR chromate
Analytics

• Long reagent shelf life with reagents save on operating and analyser
optional cooling module. maintenance costs. Advanced • Liquiline System CA80FE iron
diagnostics with remote access analyser
Componenten

• Digital communication for


Registratie/

remote access. ensure process safety and support you • Liquiline System CA80TP total
in providing process documentation phosphate analyser
to the authorities. • Liquiline System CA80COD COD
analyser
Technical data
Highly precise measurement at
System Components

Input : 1 or 2 measuring channels, 1-4 digital sensor inputs all critical control points demands
for sensors with Memosens protocol (optional) suitable sample preparation. The
Recorders &

Output : 2 x 0/4...20mA, MODBUS RS485/TCP (optional), Liquiline System CAT810, CAT820


Ethernet (optional), alarm relay and CAT860 are perfectly adapted to
E-Direct

Process temperature : 4...40°C the respective sampling spots.


Process pressure : < 0.2 bar absolute
Sample requirement : 22ml/measurement
Sample temperature : 4...40°C
Sample flowrate : Min 5ml/min
Sample consistency : Low solid content : TS < 50mg/l (ppm)
Sample supply : Unpressurised
Sample pH value : 5…9
Services &
Solutions
Theorie
Liquiline System CA80AM 215

Level
Level
Liquiline System CAT860 is the basin or the outlet. It can be equipped with a cross-flow filter to
perfect choice for demanding equipped with various filters and avoid blocking.
conditions in the inlet. Its automatic cleaning options to tailor it to the
backflush function with cleaning sampling situation. All sample preparations are fully
solution and automatic air cleaning controlled by the Liquiline System
clean off fat and proteins and Liquiline System CAT810 prepares CA80AM analyser to guarantee
prevent blocking of the ceramic samples in the outlet or from a perfect synchronization of the
filter. pressurized pipes. Thanks to its measuring point. The filters of
low dead volume, it reflects process Liquiline System CAT860 and

Pressure
Pressure
Liquiline System CAT820 is the changes promptly and shortens the CAT820 can be combined with the
flexible solution for bacteria-free response time of your downstream Flexdip CYH112 holder system to fit
sample preparation in the aeration measuring devices. The system is them into any installation situation.

Dimensions (mm)

417 (16.42)
530 (20.87)

Flow
Flow
793 (31.22)
793 (31.22)

Temperature
Temperature
417 (16.42)
530 (20.87)
793 (31.22)
793 (31.22)

Analytics
Analysis
900 (35.43)

System Components
Recorders/&
Recorders

634 (24.96)

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
216 Analytics
Niveau
Level

TOCII CA72TOC
Online TOC analyser using thermal catalytic combustion.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Exchangeable furnace The TOCII analyser determines the The heatable salt trap simplifies
• Externally triggered TOC (total organic carbon) content maintenance in applications with
Monstername

self-testing of an aqueous sample. It uses high salt loads. When samples


• Heatable salt trap for increased thermal catalytic combustion with with high salt loads are combusted,
service life subsequent NDIR (non-dispersive deposits form on the catalyst and
• pH-controlled acid dosing infrared) detection of the produced the furnace that lead to inaccurate
• Double-batch operation carbon dioxide. As TOC indicates the measurements and can result in
total organic load of water, it is used congestion. With the heatable salt
to assess the water quality and can trap, salts can be removed from
Analytics

serve as a basis for the calculation of the furnace quickly and efficiently,
wastewater charges. significantly increasing the
availability of the measuring point. It
Componenten
Registratie/

also reduces operating costs thanks


Technical data to longer maintenance intervals and
easy handling of the salt trap.
Measuring range : 0.25…600mg TOC/l, 1...2400mg TOC/l,
2.5...6000mg TOC/l, 5...12000mg TOC/l
(depending on version)
System Components

Signal inputs : 8 signal inputs 24 V DC, active, load max. 500 Ω


Output signals : 0/4...20mA, galvanically separated
Recorders &

Inorganic carbon removal : > 95% using a pH controlled TIC scrubber


Sample temperature : 0...40°C
E-Direct

Sample flow rate : 20ml/min


Sample volume : 90ml
Sample consistency : Aqueous (flammable substances not allowed
in ignitable concentrations - a sample
preparation is required)
Protection : IP54
Accuracy : 0.4% for 20% of full scale, 2.4% for 80% of full scale
Repeatability : 0.4% for 20% of full scale, 1.6% for 80% of ful scale
Services &
Solutions
Theorie
TOCII CA72TOC 217

Level
Level
Dimensions (mm)
785 (30.9)
400 (15.8)

600 (23.6) 50
(1.97)

Pressure
Pressure
1010 (39.8)
1558 (61.3)

Flow
Flow
mm (inch)

Temperature
Temperature
Operation
1 = Main switch
2 = Computer
3 = Pump P2 sample analysis
4 = Junction box
5 = Compressor switch
6 = Separation chamber
7 = Dosing valve
8 = Injection unit
9 = Acid filter
10 = Tube furnace with catalyst
11 = Combined filter
12 = Heated salt trap (optional)

Analytics
Analysis
13 = Circuit gas flowmeter
14 = Dilution water pump (optional)
15 = Acid hose connector
16 = MV4 calibration standard changeover C1/C2
17 = Valve for online sample/manual sample
18 = MV1 wastewater/calibration standard changeover
19 = Pump P3, acid dosing
20 = Pump P4, sample – dilution (optional)
21 = Mixing chamber (optional)
22 = Pump P1, sample – strip chamber/condensate
extraction
23 = Condensate hose connector
24 = Stripping gas dosing valve
25 = Carrier gas dosing valve
System Components

26 = Strip chamber with pH electrode


Recorders/&
Recorders

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
218 Analytics
Niveau
Level

Liquistation CSF48
Automatic stationary sampler for liquid media.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Easily modified into a measuring Liquistation CSF48 stationary Thanks to the flexible electronics
station sampler offers the maximum in and expansion slots, inputs for digital
Monstername

• Integrated data logger simple sampling. Based on the sensors with Memosens protocol
• Two bottle trays for simple Liquiline-Memosens platform, the can be quickly connected – with
transportation Liquistation goes above and beyond automatic detection software, they
• Can connect one or two the call of duty with the option to be can be operated with immediate
Memosens digital sensors used as a complete measuring station effect.
• MCERTS certified - without the need for additional
• Flow assembly option for transmitters. The modular housing consists of
Analytics

sampling from pressurised pipes two parts with separately lockable


doors to ensure that samples are
always stored securely and the
Componenten
Registratie/

innovative power supply and cooling


Technical data system ensure that the samples are
not exposed to any fluctuations in
Housing : Plastic and stainless steel temperature. Further protection is
Repeatability : Vacuum pump: 5%, peristaltic pump: 5% provided by the clever design - with
Communication : One service interface accessible via front panel no screws on the outside of the
System Components

connection (optional); Commubox FXA291 housing, once locked the sampler


(accessory) required for communication with the PC cannot be broken open.
Recorders &

Certification : MCERTS
Inputs : 2 x analogue; 2 x binary; 2 x optional Memosens
E-Direct

digital inputs
Outputs : 2 x binary outputs; 2 x 4...20mA optional;
2 x relay (optional)
Power supply : 100...240V AC ±10%, 50/60Hz
Datalogging capability : SD card
Protection : IP68
Suction height : 6m or 8m
Sample detection : Conductive or capacitance (for heavily fouling media)
Services &
Solutions
Theorie
Liquistation CSF48 219

Level
Level
625 625
471 471 753 753

A A A A
1258
1258

Pressure
Pressure
903
903

Dimensions (mm)
355 355 125 125 816 816
Liquistation CSF48 with plastic housing (without and
with stand)
625 625 625 625
471 471 753 753 471 471 753 753

Flow
Flow
A A A A A A A A
1258

1258
1758
1758

903

903
1403
1403

355 355 125 125 816 816

Temperature
Temperature
625 625
530
530

471 471 753 753

mm mm 355 355 125 125 816 816

A A A A

Installation
1758

1758

Foundation plan 730 (28.74)

Analytics
A = Fasteners (4 x M10)

Analysis
1403

1403

110 (4.33)
B = Cable inlet
311 (12.24)
61 (2.4)

C = Outlet for condensate and overflow > DN50


D = Sample supply from below > DN80 60 (2.36)

Dotted line = Dimensions of Liquistation


B D
268.9 (10.59)

A A
530

530
235 (9.25)

mm mm 355 355 125 125 816 816


355 (13.98)
605 (23.82)

System Components
165.1 (6.5)

C C
302.5 (11.91)

Recorders/&
107.1 (4.22)

Recorders

A A

151 (5.94) 151 (5.94)

393 (15.47)
mm (inch)
786 (30.94)

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com
• For product selection help, try our online Application tool: www.uk.endress.com/applicator
Theory&
Solutions
Services
220 Analytics
Niveau
Level

Liquiport 2010 CSP44


Portable sampler for fully automatic sampling and
distribution of liquid media.
PressureDruk
Flow
Flow
Analyse/ Temperature Temperatuur

• Flexible programming: time or Liquiport 2010 CSP44 portable • Authorities and water conservation
event-controlled water sampler offers fully automated boards:
Monstername

• Uniform operation: same sampling and defined distribution of - Water protection and water
interface as Liquiline liquid media and is ideal the following quality
transmitters applications: - Monitoring of indirect/direct
• Lockable sample base to prevent • Municipal and industrial sewage dischargers
sample manipulation treatment plants: - Labs and hydrological institutes
• Integrated data logger for - Self-monitoring
recording measured values - Process monitoring Depending on the version ordered,
Analytics

• Simple cleaning and - Monitoring of indirect dischargers one or two Memosens digital sensors
maintenance - Manhole monitoring can be connected to the CSP44.
Furthermore, two 0/4…20mA
Componenten
Registratie/

analogue outputs are available, as


well as a cleaning function and an
alarm relay.
Technical data
Sampling methods : Event sampling, single and multiple samples, sampling
table, time proportional sampling (CTCV), flow
System Components

proportional sampling (VTCV), flow proportional


Recorders &

sampling/time override (CTVV)


Input types : Up to 2 analogue inputs, up to 2 binary inputs, 1 or 2
digital inputs for Memosens sensors
E-Direct

Dosing volume : 10ml...10000ml


Dosing accuracy : ±5ml or 5% of the set volume
Repeatability : 5%
Process temperature : +2°C…+50°C
Process pressure : Unpressurised
Protection : Entire device with closed cover: IP54, controller: IP65
Interface : Back-lit LCD with menu-guided operation
Services &
Solutions
Theorie
Liquiport 2010 CSP44 221

Level
Level
System design

System design
1. Controller cover
2. Cover of battery compartment 1
3. Upper carrying handles
4. Unit upper compartment

Pressure
Pressure
5. Peristaltic pump with pump tubing
6. Bottle retaining cover
7. Lockable latches
8. Lower carrying handles
9. Unit lower compartment
10. Bottle distribution
11. Lockable latches
2
12. Suction line connection 15
13. Medium detection
14. Electrical connections 3
15. Controller
14

4
13

12 5

Flow
Flow
11 6

7
10
8

Temperature
Temperature
9

Dimensions (mm)
480 (18.9)

Analytics
Analysis
320 (12.6)

675 (26.6)

System Components
Recorders/&
Recorders
355 (14.0)

• For more information, specific application advice or to order, please contact us on 0161 286 5000 or email: info@uk.endress.com
• To download technical documentation, please visit our website: www.uk.endress.com