CYW-201 0434

Endoscope Washer & Disinfector

Operator’s Manual

EU Representative Document No.: CUM-02-E


MEDISERV TECHNOLOGIES ITALIA March 25, 2013 (rev.2)
Plazza A. Graf, 132-10126
Tel: +39.011.6670443

Address: 5Fl., Joongil Eines Platz III, 519, Dunchon-daero, Jungwon-gu, Seongnam-si, Gyeonggi-do 462-807, Korea

Tel: +82.31.747.9900 http://www.choyangmed.com

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 1


Introduction

Chapter 1 System Overview

Chapter 2 Safety Precautions

Chapter 3 System Description

Chapter 4 Installation

Chapter 5 Operation

Chapter 6 Cleaning & Maintaining

Chapter 7 Disinfectant & Detergent

Chapter 8 Troubleshooting

Chapter 9 Specification

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 2


CHAPTER

1
System Overview

1.1 Introduction to the system

CYW- 201 is an automatic disinfection device for flexible endoscopes, which has been designed
to help you with your daily routine work.

The installation of the CYW-201 will be done in a few minutes.


Insert manually pre-cleaned endoscopes into the basin, connect the channels of the endoscope,
and turn on the device.

At the end of the program, a perfectly cleaned, disinfected, rinsed and dried endoscope is
ready for the examination of your next patient.

1.2 Intended Use

Only properly trained individuals may operate. Never use the device for any purpose other
than the manufacturer’s specific intended purpose.

CYW-201 is a device which automatically washes and disinfects the surface of endoscopes and
the inside of its channel using the disinfectant.

Follow the instructions advising from the manufacturer as the way to wash and disinfect
depends on the manufacturers.

This is a medical device and must be used in a hospital and medical institution.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 3


1.3 Product list

Thank you for purchasing an Endoscope washer-disinfector of Choyang Medical Industry LTD.
Before you start to use this model, Please make sure that the following items are included.
If any items are missing, contact the manufacturer or suppliers.

Main equipment Grid Accessory basket

Channel tube Residual drain tube Suction tube

Leakage tester’s Channel connector adapter


Leakage tester’s coil tube
connector cap (ONE) (ONE)
`

Olympus Pentax Fujinon Olympus Pentax Fujinon

Depending on the endoscope Depending on the endoscope


you can choose one cap. you can choose one adapter.

Water supply & Drain


User manual Service manual
Adapter

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 4


CHAPTER

2
Safety Precautions

2.1 General Labeling


Various icons are used in this manual to indicate correct device use, and help prevent property
damage and injury to you and others. The meanings of these icons are described below.

Icon Description
Failure to follow instructions indicated by this icon may lead to
serious injury or death.
Failure to follow instructions indicated by this icon may lead to
injury or property damage.

Precautionary notations are indicated by the following symbols.


Icon Description

Do not perform this action.

You are strongly encouraged to perform this action.


`

Unplug the device from the power source.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 5


■ Only properly trained individuals may operate or service the device. Never use
the device for any purpose other than the manufacturer’s specific intended
purpose.
■ For service or repair, contact your authorized CHOYANG MEDICAL service
representative

■ Do not disassemble or modify the device.

■ Do not contact main board and electric parts with water.

■ To avoid risk of electric shock, this equipment must only be connected to a


supply mains with protective earth.

■ If this device acts abnormally, unplug it.

■ If the power cord is damaged, unplug it.

■ Do not damage the power cord.

■ Provide correct voltage.

■ Do not use the provided AC adapter or power cord with any other devices.

■ Do not hold the power cord plug with wet hands.

■ Make sure that excessive dust and debris do not accumulate on the power cord.

■ Do not place heavy objects on the device.

■ If you need to move the device, all tanks should be empty.

■ Do not modify this equipment without authorization of the manufacturer.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 6


■ Do not move the device when the power cord is taut.

■ Unplug the device for servicing.

■ Do not touch the inside of the device.

■ Do not operate when water tank is empty.

■ Do not operate if door is open.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 7


2.2 Box labeling

2.3 Safety labeling

Label Position : Near the Water inlet of rear

Label Position : Cover of the Main board in the device

Label Position : The Lid

Label Position : Near the Air-drain button

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 8


Label Position : Near the disinfectant tank level gauge

Label Position : Near the detergent tank level gauge.

Label Position : Near the leakage air outlet button (yellow button)

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 9


2.4 Symbol of Device

Symbol Description

CE Marking of conformity

Manufacturer information

Date of manufacture

Serial Number

Manufacturer’s EU representative information

Follow instructions for use

Dangerous voltage

Protective earth

Electricity specification
220-230Vac, 50/60Hz

Fuse specification

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 10


CHAPTER

3
System Description

3.1 Front view

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 11


3.2 Rear view

3.3 Basin view

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 12


3.4 Control Panel

No Name Function

1 WASHING - Set the time of washing in “MR.Set” PROGRAM


- Operate the process of washing in “MANUAL” PROGRAM
2 DETERGENT - Set the time of detergent in “MR.Set” PROGRAM
- Operate the process of detergent in “MANUAL” PROGRAM
3 DISINFECTANT - Set the time of disinfectant in “MR.Set” PROGRAM
- Operate the process of disinfectant in “MANUAL” PROGRAM
4 AIR - Air : Supply air to each channel in the endoscope
5 DRAIN - Drain : Drain existing liquid from the basin
Related with disinfectant
- Retank: Fill disinfectant in the disinfectant tank (Push the button
6 ReTANK 2times)
7 DISCHARGE - Discharge : Discharge old disinfectant in the disinfectant tank
outside of the machine (Push the button for 2~3
seconds)
8 LEAKAGE Process for leakage test of an endoscope
9 MODE SET Choose the A,B,C mode
- Auto : Operate “AUTO” PROGRAM
10 PROGRAM - Manual : Operate each process(No.1~8 in this chart) regardless of
“AUTO” PROGRAM
- MR.Set : Set the time of each process
Operation start and stop
11 START/STOP Stop temporarily if you push the button during operation.
Reset if you push the button for 2~3 seconds.
Lamp for taking action when
12 Alarm lamp
DIS.LOW – there is no disinfectant, DOOR – lid opens
DRAIN – there is some liquid in the basin, END – the process is completed.
Showing setting and remaining time for washing, detergent and
13 LED DISPLAY disinfectant and condition of tank indicating “Lo” when tank is
empty.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 13


3.5 System specifications

► LEAKAGE TESTER

► 2 ENDOSCOPES CAN BE WASHED AT ONCE.

► OPEARTING PROCEDURE
TIME : WATER – 10 sec ~ 120 sec (every 10 sec), 2 min ~ 5 min (every 30 sec)
DETERGENT – 10 sec ~ 120 sec (every 10 sec), 2 min ~ 5 min (every 30 sec)
DISINFECTANT – 1 min ~ 10 min (every 1 min), 10 min ~ 20 min (every 5 min),
20 min ~ 60 min (every 10 min), 1 hr ~ 9 hr (every 1 hr)
AUTO PROGRAM
A MODE : WASHING 1 – DETERGENT – WASHING 2 – DISINFECTANT – WASHING 3- AIR
B MODE : WASHING 1 – DETERGENT – WASHING 2 – AIR
C MODE : WASHING 1 – DISINFECTANT – WASHING 2 - AIR

► TANK CAPACITY
- WATER : 18 Liter
- DETERGENT : 1.5 Liter
- DISINFECTANT : 20 Liter

► DISINFECTANT FILTERING SYSTEM

► COUPLERS : 6 OUTLETS IN CLEANING BASE(2 Scopes)

► DISCHARGE : 1 OUTLET (Compulsory Discharge on rear panel)


3 OUTLETS (In the front door)

► POWER : AC 220-230V, 50/60Hz

► DIMENSION : 620 X 620 X 1030 (W X D X H)

► WEIGHT : 65 Kg

► Protective device : There are two fuses for preventing over current.
There are alarm lamps in the control panel.
Stop operation if door is open while working.

► Expected service life : 5 years

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 14


CHAPTER

4
Installation

4.1 Unpacking and Inspection

The CYW-201 automatic endoscope washer & disinfector system has passed full quality
assurance testing before shipment. Thus, the unit should be operated upon delivery. The unit
should be unpacked, installed and tested only by an authorized Choyang Medical
representative.

4.2 Space Requirements

The working area for CYW-201 should be prepared according


to the dimensions shown in figure on the right.
- Device size : 620mm(L) x 620mm(W) x 1030mm(H)

4.3 Environmental requirements

Do not install the device in the following places


Unstable or inclined surfaces.
Outdoors or places that exposed the gases(ex. Sulfurous
acid gas, chlorine etc.) or high humidity.

The range of environmental conditions


To ensure that the CYW-201 can perform optimally, it is recommended to keep the
environmental conditions as below.
Operating environment Transport and storage environment
Room temperature 15°C ~ 40°C -20°C ~ 70°C
Relative humidity 30% ~ 75%. 10% ~90%
Atmospheric pressure 80kPa ~ 106kPa 50kPa ~ 106kPa

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 15


4.4 Line Power Considerations

The unit is delivered factory-set for the proper ac line voltage rating of the country to which it
is shipped. The ac line voltage appears on the label affixed to the rear panel.

The unit is equipped with a power cord – one end for connection at the rear panel of the unit,
the other end with a 2-pin plug for connection to a single-phase ac receptacle, suited to the
labeled power rating.

Electric power cable 3m long (supplied) with plug for local power source

4.5 Other Requirement

The device is gravity-drained it is essential that the distant drain is lower than the water at the
rear part of device.

The best drainage conditions are achieved by installing the distant drain connection less than
20cm from ground level and not more than 1m distant

4.6 Installation

1. Connect water hose to both ends of water tab and water inlet.
2. Make sure to insert the drain hose into a floor drain positioned lower than the drain out
on the CYW-201
3. For user’s safety, connect earth cable to water pipe.

Never connect the earth cable to plastic


plumbing lines, gas lines, or hot water pipes.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 16


CHAPTER

5
Operation

5.1 Power Supply

1. Plug in the power cable to the power supply.


2. Turn on the circuit breaker on the back and on/off switch on the front of the device.

* The LED indicators on the control display show that the power is switched on.

Never use an extension cord.

5.2 Water Supply

1. Turn on the water supply valve.


2. Turn off the water supply valve when the device is not in use.

* Water can be supplied by pressing on/off switch. Level sensor that prevents liquid in basin
from overflowing is installed.
* It is required to use under 3kgf/cm2 water pressure.
* Potable water is the minimum standard. Incoming water must be pre-filtered to minimum of
0.45 microns. The water filter is included with the device. The routine maintenance schedule
recommends every 6 months or sooner, depending on the pre-filtration system and the
quality of the incoming water.

Use under 3kg/㎠ water pressure

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 17


5.3 Detergent Supply

1. Open the detergent hole cap in the basin and pour detergent by using that funnel.
2. Close the detergent hole cap in the basin.

* You can check the level on the tank level in front of the device.
* Select a detergent solution that has bacteriostatic properties to inhibit bacterial growth in the
detergent reservoir and detergent line. The dual enzymes in the detergent rapidly attack
soils and the low foam properties aid effective recirculation. Follow the manufacturer’s
dilution and temperature recommendations.

Never use household detergent in the device.

Do Not be re-used.

5.4 Disinfectant Supply

1. Press the <PROGRAM> button to change “manual” in order to


pour disinfectant onto basin.
2. Seeing the floating sensor on the corner of up left, pour
disinfectant till the full position, and press <RE-TANK> button on
the control panel 2 times continuously. In short, disinfectant goes
down to the disinfectant tank.
3. Buzzer will sound, when finished.

* Select a low-foaming, high level disinfectant specifically manufactured for high-level medical
instrument disinfection. The product must be capable of destroying M. tuberculosis.
* Consult the product label for appropriate contact time and temperature when programming
the disinfection cycle.

Do not mix with detergents or other chemicals

* Disinfectants must be monitored for potency on a daily basis. Consult the disinfectant
manufacturer recommendations for monitoring guidelines.
* Use the manufacturer’s test strips to test the potency of the solution on a daily basis. If the
potency of the solution is below its minimum recommended concentration discard and
replace it with fresh solution.
CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 18
* Never use disinfectant beyond the manufacturer’s recommended reuse life, even if the
potency levels are acceptable.
* Never use disinfectant with unacceptable potency levels, even if the reuse date is unexpired.

5.5 Endoscope Pre-cleaning

All endoscopes must be pre-cleaned prior to disinfection.


Before connection the channels of the endoscopes to adaptors on the device, manual pre-
cleaning is required to minimize contamination.
Follow the endoscope manufacturer instructions and established professional guidelines to
properly pre-clean the endoscope.

The following are the usual method.


1. Washing a secretion or blood on the surface of endoscope right after examination.
2. Brushing inside of biopsy channel (use a brush supplied by endoscope manufacturer)

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 19


5.6 Leakage Test

Leakage test an endoscope prior to disinfection procedures.

This automated test is not a substitute for the endoscope


manufacturer’s manual leakage test. Follow the
manufacturer’s instructions when performing a manual test.

Use the following procedure to leakage test an endoscope.


1. Install the water-resistant cap.
2. Connect leakage tester’s coil tube and connector to device and endoscope as the below
picture.

3. Load the endoscope into the basin.


4. Press <PROGRAM> button to change “AUTO” or “MANUAL”
5. Press <LEAKAGE> button for more than 2 seconds to fill up water in the basin
6. Water supplied after 5 seconds, supplying air inside of scope.
7. Check any bubble coming out, moving insertion tube up and down, right and left.
8. If any bubble found, first of all, the scope must be taken out immediately and repaired.
9. After 5 minutes, water will be drained out automatically without supplying air anymore.
10. Detach leakage tester adapter for next procedure.
11. Press yellow button to blow out air in endoscope after leakage test.

Air supply Air / Water supply The end

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 20


Water-resistant cap on the endoscope should be
covered.

Pentax Olympus

Before
connecting
the water-
resistant
cap

After
connecting
the water-
resistant
cap

[Connect & Disconnect Method of Channel tube]


1. Assemble the channel tubes and the channel connector adapter as the below picture.

2. Connect the endoscope’s channel and device’s channel.


(Refer to each endoscope manufacture’s manual for the
connection.)

3. Put all of accessories in the accessory basket.

4. Push the channel button in the basin to disconnect the


channel tube.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 21


5.7 Time set

User can set up ‘washing’ ‘detergent’ ‘disinfectant’ times.


Press <PROGRAM> button to change “MR.set”, and set up ‘washing’, ‘detergent’ and
‘disinfectant’ times you want, pressing <WASHING> ,<DETERGENT> and <DISINFECTANT>
button one by one.

Users can set up ‘washing’ ‘detergent’ and ‘disinfectant’ times they want, as following limits:

* Washing / detergent : 10 sec ~ 120 sec (every 10 sec), 2 min ~ 5 min (every 30 sec)
* Disinfection : 1 min ~ 10 min (every 1 min), 10 min ~ 20 min (every 5 min),
20 min ~ 60 min (every 10 min), 1 hr ~ 9 hr (every 1 hr)

Disinfection time : Follow the specification of


disinfectant manufacturer.

Disinfectant expiration : Follow the specification


of disinfectant manufacturer

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 22


5.8 Auto mode

After time set, transfer to ‘AUTO’ program again and select one of modes you want.
And then the device is ready to operate as set-times, upon pressing <START/STOP> button.

# If you press <STOP> button in the middle of operation, you must press <STOP> button
once more to go back to normal modes.
# In auto or manual, if any failures of operation even after pressing <START/STOP> button,
press <DRAIN> button for all water out then press <START/STOP> button again.

AUTO MODE
A MODE W-1 DET W-2 DIS W-3 AIR
B MODE W-1 DET W-2 AIR
C MODE W-1 DIS W-2 AIR

Warning Indicators Description


○DIS.LOW ○DOOR ○DRAIN ○END Normal operation.
●DIS.LOW ○DOOR ○DRAIN ○END Disinfectant tank is empty.
●DIS.LOW ○DOOR ●DRAIN ○END Some liquid exists in the basin.
○DIS.LOW ●DOOR ○DRAIN ○END Door is open.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 23


5.9 Manual mode

Manual mode is for each operation. Press <PROGRAM> button to change


“MANUAL”, and then press each button to perform what you want to operate.

- WASHING : When you push the washing button, washing process

operates as much time as you set. It is completed with alarm after automatic

draining.

- DISINFECTANT: When you push the DISINFECTANT button, disinfectant process operates

as much time as you set. it is completed with alarm after used disinfectant goes back to the

disinfectant tank.

- AIR : When you push the air button, the air is provided to channel connector with pressure

for 15 seconds.

- DRAIN : When you push the drain button for 2~3 seconds, it is completed with alarm after

liquid in basin is drained outside of the machine.

- ReTANK : When you push the button two times after pouring the disinfectant in basin, it is

filled in disinfectant tank. It is recommended to check that there is no liquid in basin before

filling with the disinfectant.

- DISCHARGE : When you push the discharge button for 2~3 seconds, the disinfectant in its

tank is discharged after pulling the disinfectant in the tank up to the basin.

It is recommended to drain the used disinfectant at most using the outlet which is located

lower side of front door.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 24


CHAPTER

6
Cleaning & Maintaining

6.1 Cleaning

- Keeping the device clean improves its performance, reduces unnecessary repairs, and
lengthens its life.
- You should keep basin dry after use.
- It’s best to clean spills on the outside of device as they occur. Use a soft damp cloth to gently
wipe away spills and dust. Do not use benzene and thinner. It can be cleaned with a neutral
detergent or alcohol using a clean, soft cloth.
- If you decide to dispose of this device, please observe local waste disposal regulations.

6.2 Tank Liquid Discharge / Supplement

1. Disinfectant discharge / supplement


1) Disinfectant replacement
The disinfectant life cycle depends on many different kinds. After a month past at least,
it must be replaced normally. In the shortage of disinfectant, ‘DIS. LOW’ warning lamp
appears of the control panel.
2) Disinfectant discharge
Press <DISCHARGE> button on the control panel more than 3 seconds.
Open the door located in the left front of this device and find discharging holes with
faucet.
3) Disinfectant supplement
Refer to disinfectant supply on chapter 5.

2. Detergent Supplement
- No need to discharge detergent (in case of the shortage, refill some from time to time,
checking the tank level indicator)

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 25


3. Water Discharge
- No need to discharge water except any case of moving or not using for a long time.
- Discharge water through water discharging hole (water faucet) in the bottom left on the
front this device.

6.3 Replacement of components & Cleaning method

1. Basin cleaning
- Open the lid after using the device to keep dry in the basin. (Except when long term
disused.)
- Periodically wipe the inside of the lid and the basin with a dry soft cloth to remove any dirt
or liquid.
- Level sensor in the basin : Open the cap of the level sensor and softly clean the sensor with
a endoscope brush. Carefully screw on the cap again.
- Channel connection(PMCD) in the basin : Clean the inside of channel connection(PMCD)
with a endoscope brush.

2. Disinfectant filter cleaning


① The cleaning procedure is needed when debris is in the disinfectant filter.
② Open the disinfectant filter cap to the right which is located in front of the frame.
③ After taking out the filter net and removing the debris, start to wash with water.
⑤ The cleaned filter net is in place and screwed the disinfectant filter cap.

Unless the disinfectant filter cap is not properly closed, the


leak might occur during disinfectant procedure.

3. Fuse (replacement)
① Open the fuse holder. (If necessary, use the tool such as a flat-head screwdriver.)
② There are two 250V 5A fuses. Close the fuse holder after replacing fuse if needed.
- Fuse is necessarily recommended to use the certificated product.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 26


CHAPTER

7
Disinfectant & Detergent

The efficacy of a disinfection procedure is directly related to the disinfectant solution used and
the amount of time the endoscope is exposed to that solution.

CHOYANG MEDICAL INDUSTRY LTD. provides information about the disinfectant & detergent to
be used, as established during type testing.

[DISINFECTANT]
AIDAL PLUS made in Australia (Whiteley Medical)

For HIGH LEVEL DISINFECTION, Soak instruments in undiluted solution 35℃ for 5
minutes or 20℃ for 10 minutes.

AIDAL PLUS is a stable, odour suppressed, mildly acidic 2% glutaraldehyde based sterilant
suitable for use in hospitals, dental/medical surgeries and veterinary Hospitals for the
sterilisation and high level disinfection of critical and semicritical medical devices.
AIDAL PLUS shows several major improvements over alkaline activated Glutaraldehyde.
Improved biocidal performance is achieved with reduced glutaraldehyde odour and reduced
propensity for irritation and yellowing to human skin compared with alkaline glutaraldehyde
preparations.
AIDAL PLUS has been tested in a wide array of tests and various conditions, including testing
at the end of the shelf life, at the end of the re-use period, diluted to the minimum
recommended concentration of active material(1.5% glutaraldehyde), and then soiled and
tested in hard water against hardest to kill organisms.
AIDAL PLUS will not damage endoscopes or other complex medical equipment when used in
accordance with directions and the instrument manufacturer’s instructions.
AIDAL PLUS has demonstrated performance as a bactericidal, fungicidal, virucidal,
mycobactericidal and sporicidal high level disinfectant with complete efficacy under these
conditions as a Class IIb medical device.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 27


[DETERGENT]
MATRIX made in Australia (Whiteley Medical)

MATRIX has been shown in internationally peer reviewed published clinical data to provide
excellent soil and biofilm removal from materials used in endoscopes and other medical
devices.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 28


CHAPTER

8
Troubleshooting
PROBLEM POSSIBLE CAUSE SOLUTION
* The power is not connected. * Connect the power cable.

* A ELB switch is in the “OFF” position. * Turn on the ELB switch “ON”.
Does not supply the power.
* The fuse is broken or the circuit * Replace the fuse or repair the

breaker is out of order. Circuit breaker.

* Check that the water supply


The water supply does not * Water supply valve is close.
valve is open.
work. * The water supply pump is out of order.
* Contact a service representative.

* There is no disinfectant in the tank. * Check the tank level and add
The disinfectant supply does
* Break down the disinfectant tank or the disinfectant.
not work.
level valve. * Contact a service representative.

* The Single use solution switch is in the


The disinfectant is * Turn on the ELB switch “O”.
“I” position.
discharged automatically. * Contact a service representative.
* The drain valve is out of order.

The disinfectant is flowing * The main board has problems.


* Contact a service representative.
over the basin. * Failure in the level sensor

* Press the drain button.


* DRAIN lamp is on. * Check that the door is closed
Does not working. * DOOR lamp is on. completely.

* “Lo” message is displayed. * Add disinfectant or supply the

water.

* The drain hose is folded or clogged. * Straighten the hose.


Does not drain well.
* The drain pump is out of order. * Contact a service representative.

Does not retank after the

disinfectant phase has * The level sensor is out of order. * Contact a service representative.

finished.

Does not supply the air * Failure in the air valve.


* Contact a service representative.
during the Leakage test. * Failure in the air valve..

* Please refer to the service manual for further details and contact your authorized
CHOYANG MEDICAL service representative.

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 29


CHAPTER

9
Specification for CYW-201

Product Endoscope Washer-Disinfector


Model CYW-201
Dimension 620(L) x 620(W) x 1030(H)mm
Weight 65kg
Electricity AC 220-230V, 50/60Hz
Power 250VA
Fuse 250V 5A, 2ea
Protection against electric shock Class 1
Capacity Basin : 16L
Water tank : 18L
Disinfectant tank : 20L
Detergent tank : 1.5L
Software version C3e12 (2013.01.12)

CUM-02-E(rev.2) CHOYANG MEDICAL INDUSTRY LTD. 30