Professional Documents
Culture Documents
Presio 769 N
Presio 769 N
Presio 769 N
•
Differential Pressure monitoring
Transmitter •
analysing
PAD
www.kobold.com
1
Differential Pressure Transmitter Model PAD
Features
Superior performance
High reference accuracy:
±0.075 % of calibrated span
(optional: ±0.04 % of calibrated span)
Long-term stability (0.125 % URL for 3 years)
High rangeability (100:1) for range 4-0
Flexibility
Data configuration with HART® configurator
Zero point adjustment
Reliability
Continuous self-diagnostic function
Automatic ambient temperature compensation
Description
EEPROM write protection
The Kobold Differential Pressure Transmitter model PAD is a
micro processor-based high performance transmitter, which Fail-mode process function
has flexible pressure calibration and output, automatic com- CE EMC conformity standards (EN 50081-2, EN 50082-2)
pensation of ambient temperature and process variable, con- Transmitter Description
figuration of various parameters, communication with HART® Electronics module
protocol. The application is very various, as measuring pres- The Electronics module consists of a circuit board sealed
sure, flow and level by application method. All data of sensor in an enclosure. There are a MCU module. an analog mo-
is to be input, modified and stored in EEPROM. dule, a LCD module and a terminal module in a transmitter.
As an option the Kobold Pressure Transmitter is also available The MCU module acquires the digital value from the ana-
as a flow meter. This flowmeter model PAD-F has added the log module and applies correction coefficients selected from
totalizing function in the PAD transmitter. So it is available to EEPROM. The output section of the MCU module converts
check the flow rate and totalizing flow. It measures the flow the digital signal to a 4…20 mA output. The MCU module
rate by using differential pressure without compensation of communicates with the HART®-based configurator or control
temperature and static pressure. The shape of the PAD-F is systems such as DCS. The Power section of MCU module
the same as the standard device and it is only the terminal has a DC-to-DC power conversion circuit and an input/out-
block which is different since there are two more terminals for put isolation circuit. The LCD module plugs into the MCU mo-
the read-out of the pulse output. dule and displays the digital output in a user-configured unit.
0 = without
C = engineering unit (must be
S = standard
chosen when using the
N = ½" NPT (waterproof 0 = without
differential transmitter as a
epoxy- IP67) 2 = manifold
flowmeter)
2 = ¼" NPT polyester F = ATEX, 2-ways
D = teflon o-ring (wetted part)
S = silicone female painted flameproof, Ex d (st. steel)
E = oil free finish
I = inert filling (standard) aluminium E* = ATEX, intrinsi- 3 = manifold
F = side vent / drain bottom
liquid 4 = ½" NPT G = G ½ epoxy- cally safe, Ex i 3-ways
G = side vent / drain top
X2) = special female (adapter) polyester (st. steel)
H = multi-planar
X2) = special painted 5 = manifold
process connection
aluminium 5-ways
M = housing in stainless steel
X2) = special *option E (st. steel)
N5) = mounting of PAD onto
in preparation
diaphragm seal
Y2) = special
2)
1
specify flow rate engineering unit, Δ p and flow rate Order code X and Y must be specified in writing
3)
at URV (Upper Range Value), Δ p and flow rate (generally '0') not for PAD-H
4)
at LRV (Lower Range Value) not for PAD-D and PAD-F
5)
pulse scale (choose only one value from Diaphragm seal model and application data to be speicified in clear text.
0.001, 0.01, 0.1, 1, 10, 100, 1 000, 10 000 m³/pulse) and Application Index on pages 15, 16 to be filled out. For summary of
pulse width (choose only one value from diaphragm seal models and possible ranges, see following page 11 on
10 ms, 50 ms, 100 ms), while ordering so that max. wards. For dimensional details see DRM data sheet.
6)
duty cycle is 49 pulses/sec Remote type. Option '2' possible only with PAD-G/A...
1/09 - 2011
* Special Measuring span with adequate lower and upper range limits on request
Unit Conversion
Range code bar kg/cm² KPa psi in H2O at 4 °C mm H2O at 4 °C in Hg at 0 °C
2 0.015 0.015 1.5 0.217 6 152 0.422
3 0.075 0.076 7.5 1.087 30 765 2.215
4 0.373 0.38 37.3 5.410 149 3804 11.014
5 1.865 1.902 186.5 27.049 749 19018 55.072
6 6.900 7.036 690 100.073 2773 70361 203.750
7 20.681 21.088 2068 299.930 8310 210878 610.660
8 68.950 70.309 6895 1000.009 27708 703097 2036.025
9 206.800 210.876 20680 2999.303 83105 2108781 6106.597
0 413.700 421.856 41370 6000.211 166085 4218566 12216.550
1/09 - 2011
Dimensions
PAD standard* PAD multi planar process connection*
112 112
86 73 39 86 73 39
s s
U T I ON - U T I ON -
- CA IN - CA IN
D D
E
E
EX
EX
GI Z
GI Z
PLO
PLO
E NE R
E NE R
TERMINALS
TERMINALS
S I VE A TM
S I VE A TM
FIELD
FIELD
ILE
ILE
113
WH
WH
113
OS
O
N
SP
P
E HE E HE
OP RE OP RE
D O N OT D O N OT
LOCK LOCK 192.5 LOCK LOCK
202.5
79.5
XXXX CF8M
89.5
42 42
316
54 54
PAD with multi planar flange and angle type bracket* PAD standard with flat type bracket (vertical mounted)*
86 112
86
112
s
U T I ON - s - CA
U T I ON -
- CA IN IN
D D
I ZE
EX
E
EX
IZ
PL O
I L E E NE R G
PLO
I L E E NE R G
TERMINALS
S I V E A TM
FIELD
S I V E A TM
TERMINALS
FIELD
WH
WH
OS
N
OS
HE
P
E
N
OP
P
E HE RE
OP RE D O N OT
D O N OT
LOCK LOCK
LOCK LOCK
323.5
XXXX CF8M
324.5
316
210
122
47
94
110
210
120 112
s
U T I ON -
- CA IN
D
I ZE
EX
PL O
I L E E NE R G
TERMINALS
S I V E A TM
FIELD
WH
OS
N
HE
P
E
OP RE
D O N OT
LOCK LOCK
205.5
47
Bleed valve
28.5
51.0
84.0 (open, approx.)
Instrument
Process side Block valve side
Sied
24.0
54.0
86.0
31.0
33.3
78.0
½" NPT
4 places
¼" NPT
Static vent
2 places
Block valve
85.0 (open, approx.)
Block valve
Process Vent port Process
24.0
¼" NPT 54.0 ¼" NPT
Static port Vent port
2 places 228.0 (open, approx.)
1/09 - 2011
Process side
welded
Fig. 1
Dimensions (mm): Examples for DN50 / DN 80/DN100 / 2" ANSI / 3" ANSI/4"ANSI
Flange type D k d2 b f d4 X d5 RL
DN50 PN16 165 125 18 18 2 4 48
102
DN50 PN40 165 125 18 20 2 4 48
2" ANSI Cl. 150 152.4 120.6 19 19.1 2 4 48
92
2" ANSI Cl. 300 165.1 127 19 22.3 2 8 48 50 mm (2")/
DN80 PN16 200 160 18 20 2 8 76 100 mm (4")/
138
DN80 PN40 200 160 18 24 2 8 76 150 mm (6")/
3" ANSI Cl. 150 190.5 152.4 19 23.9 1.6 4 76 200 mm (8")/
127 (customer
3" ANSI Cl. 300 209.5 168.3 22 28.4 1.6 8 76
DN100 PN16 220 180 18 20 2 149 8 89 specified)
DN100 PN40 235 190 22 24 2 149 8 89
4" ANSI Cl. 150 228.6 190.5 19 24 1.6 157.2 8 89
4" ANSI Cl. 300 254 200 22 32 1.6 157.2 8 89
Fig. 2
1/09 - 2011
Example of PAD-G remote assembed with (extended) diaphragm seal and capillary
(for dimensional details, see DRM data sheet)
Fig. 3
Dimensions (mm): Examples for DN50 / DN 80 /DN100 / 2" ANSI / 3" ANSI/4"ANSI
Flange type D k d2 b f d4 X d5 RL
DN50 PN16 165 125 18 18 2 4 48
102
DN50 PN40 165 125 18 20 2 4 48
2" ANSI Cl. 150 152.4 120.6 19 19.1 2 4 48
92
2" ANSI Cl. 300 165.1 127 19 22.3 2 8 48 50 mm (2")/
DN80 PN16 200 160 18 20 2 8 76 100 mm (4")/
138
DN80 PN40 200 160 18 24 2 8 76 150 mm (6")/
3" ANSI Cl. 150 190.5 152.4 19 23.9 1.6 4 76 200 mm (8")/
127 (customer
3" ANSI Cl. 300 209.5 168.3 22 28.4 1.6 8 76
DN100 PN16 220 180 18 20 2 149 8 89 specified)
DN100 PN40 235 190 22 24 2 149 8 89
4" ANSI Cl. 150 228.6 190.5 19 24 1.6 157.2 8 89
4" ANSI Cl. 300 254 200 22 32 1.6 157.2 8 89
1/09 - 2011
DRM-606
R20 G¾ 0 ...10 600
Capsule seal with rotable
short capsule +350 °C
male, capillary
R28 M 28 x 1,5 0 ...10 600
DRM-611
SMS R40 1 ½" Ø 34.5 0 ... 1.6 40
SMS socket with union nut,
+200 °C
capillary
R50 2" Ø 45.5 0 ... 0.6 40
1/09 - 2011
* Note: Threaded diaphragm seal only available with PAD-G/A. For PAD-D only flange, rotating connection or clamp diaphragm seals are possible.
DRM-615
APV-RJT R20 1" Ø 29.5 0 ... 2.5 100
DRM-617
DRM-620
DRM-620/1
DRM-622/1
F48 Ø 48 mm Ø 48 0 ...0.6 40
F48 2 Ø 48 mm Ø 48 0 ...0.6 40
DRM-624/1
F1H Ø 100 mm Flange, capillary Ø 63.5 +250 °C 0 ... 0.4 40
DRM-625/1
R15 G½ 0 ... 0.4 40
* Note: Threaded diaphragm seal only available with PAD-G/A. For PAD-D only flange, rotating connection or clamp diaphragm seals are possible.
* Note: Threaded diaphragm seal only available with PAD-G/A. For PAD-D only flange, rotating connection or clamp diaphragm seals are possible.
* Note: Threaded diaphragm seal only available with PAD-G/A. For PAD-D only flange, rotating connection or clamp diaphragm seals are possible.
Application Index
Please fill out the following Application Data Sheet while inquiring/ordering model PAD assembled with diaphragm
seal model DRM
Order/ Inquiry Ref./ Item No.
Medium:
Operating density g/cm²
Operating viscosity cSt
...continued...
1/09 - 2011
Capillary:
length in 'mm' at port 'H' (if other units, please specify) mm
length in 'mm' at port 'L' (if other units, please specify) mm
protection hose required (yes/no)
Height Adjustment:
No
1.) Factory calibration for height adjustment required
Yes, choose from the following
option(s) 2 … 5
2.) PAD higher than DRM at port 'H' (Fig. 2 or 3), specify +Δh(1) mm
3.) PAD lower than DRM at port 'H' (Fig. 2 or 3), specify -Δh(1) mm
Following options "4" and "5" not valid while ordering PAD-G…
4.) PAD higher than DRM at port 'L' (Fig. 2), specify +Δh(2) mm
5.) PAD lower than DRM at port 'L' (Fig. 2), specify -Δh(2) mm
Options: Extended diaphragm seal at both ports (see Fig. '1' or Fig. '3' for dimension RL,
tick only if option needed)
No (extended diaphragm seal only at port 'H')
If No, length 'RL' of extended diaphragm seal at port 'H'
Yes
If Yes, length 'RL' of extended diaphragm seal at port 'H' mm