Download as pdf or txt
Download as pdf or txt
You are on page 1of 1

Method: Flame Sensitivity Page 1 Date: 2/18/2021 11:13:39 AM

====================================================================================================
Analysis Begun

Logged In Analyst: Biochemistry Project Technique: AA Flame


Spectrometer: AAnalyst 800, S/N 600S11030201 Autosampler: S10

Sample Information File: E:\Documents and Settings\All Users\PerkinElmer\AA\Data\Sample Information\


IPV Flame.sif
Batch ID:
Results Data Set: PM 18Feb2021
Results Library: E:\Documents and Settings\All Users\PerkinElmer\AA\SystemFiles\Service\Results.mdb

====================================================================================================
Method Loaded
Method Name: Flame Sensitivity Method Last Saved: 2/21/2019 2:26:47 PM
Method Description: Using 2 mg/L Cu Standard

Analyte Calibration Equation Wavelength Slit Lamp Energy


Width Current
Cu 324.8 Non Lin Thru 0 324.8 0.7H 15 71

====================================================================================================
Sequence No.: 1 Autosampler Location:
Sample ID: Calib Blank 1 Date Collected: 2/18/2021 11:08:47 AM
Analyst: Data Type: Original

----------------------------------------------------------------------------------------------------
Replicate Data: Calib Blank 1 Analyte: Cu 324.8
Repl SampleConc StndConc BlnkCorr Time Signal
# mg/L mg/L Signal Stored
1 [0.00] -0.000 11:08:51 AM Yes
2 [0.00] -0.000 11:09:02 AM Yes
3 [0.00] -0.000 11:09:14 AM Yes
4 [0.00] -0.000 11:09:27 AM Yes
5 [0.00] -0.000 11:09:39 AM Yes
6 [0.00] -0.000 11:09:51 AM Yes
7 [0.00] -0.000 11:10:03 AM Yes
8 [0.00] -0.000 11:10:15 AM Yes
9 [0.00] 0.000 11:10:27 AM Yes
10 [0.00] 0.000 11:10:39 AM Yes
Mean: [0.00] -0.000
SD: 0.0000 0.0001
%RSD: 0.00% 83.59
Auto-zero performed.

====================================================================================================
Sequence No.: 2 Autosampler Location:
Sample ID: 2mg/L Cu Std Date Collected: 2/18/2021 11:11:05 AM
Analyst: Service PKI Data Type: Original

----------------------------------------------------------------------------------------------------
Replicate Data: 2mg/L Cu Std Analyte: Cu 324.8
Repl SampleConc StndConc BlnkCorr Time Signal
# mg/L mg/L Signal Stored
1 0.355 11:11:06 AM Yes
2 0.355 11:11:18 AM Yes
3 0.355 11:11:30 AM Yes
4 0.353 11:11:42 AM Yes
5 0.352 11:11:54 AM Yes
6 0.352 11:12:06 AM Yes
7 0.357 11:12:18 AM Yes
8 0.356 11:12:30 AM Yes
9 0.355 11:12:42 AM Yes
10 0.355 11:12:55 AM Yes
Mean: 0.354
SD: 0.0017
%RSD: 0.48

Cu 324.8 - Measured Characteristic Concentration: 0.025 mg/L

You might also like