8/15/23, 10:35 PM COM Port Mapping with Keyspan USB Serial Adapter

COM Port Mapping with Keyspan USB Serial Adapter


In the event that a customer has purchased a serial device and is attempting to connect it to a PC that
does not have a serial port but does have a USB port, a USB to serial adapter should work. X-Rite
recommends the "Keyspan model USA-19HS" USB to serial adapter. It ships with a software
installation CD that will install drivers, a users manual, a useful Help file, as well as the "Keyspan USB
Serial Adapter Assistant". It is this "Keyspan USB Serial Adapter Assistant" that will allow a user to
change port mappings on the PC to select the required "COM1", "COM2", "COM3", or "COM4".
Some X-Riteproducts, such as PREVU* and Cholesterol 1,2,3, must be connected to ports mapped to
"COM1", "COM2", "COM3", or "COM4" in order to communicate with the spectrophotometer. If
the customer has installed a USB to serial adapter, properly connected the spectrophotometer, and still
receive a communications error when opening the PREVU* and Cholesterol 1,2,3 software; the adapter
could very well be selecting a communications port other than the required "COM1", "COM2",
"COM3", or "COM4". If the "Keyspan model USA-19HS" has been installed, follow these directions:

COM Port Settings/Changes -

Go to Windows "Start" > "All Programs" > "Keyspan USB Serial Adapter" > "Keyspan Serial
Assistant" to access the following window that will default to the "Adapter Status" tab indicating
which communications port the adapter is mapped to. The example below indicates that "COM4" is
selected and the MD22 spectrophotometer should have no problem communicating with the PREVU*
and Cholesterol 1,2,3 software providing that "COM4" is selected in that software's "Options"
window.

✉ Contact  Call

https://www.xrite.com/service-support/com_port_mapping_with_keyspan_usb_serial_adapter 1/6
8/15/23, 10:35 PM COM Port Mapping with Keyspan USB Serial Adapter

The "Properties" tab states the current configuration and allows the user to test it by clicking on the
"Test This Port". "PASSED" or "FAILED" will appear in the "Serial Test Status" box.

✉ 

https://www.xrite.com/service-support/com_port_mapping_with_keyspan_usb_serial_adapter 2/6
8/15/23, 10:35 PM COM Port Mapping with Keyspan USB Serial Adapter

The "Port Mapping" tab allows you to configure a different communications port for the "Keyspan
USB Serial Adapter". It would be advisable to select "Show All COM Assignments" before
considering a making any changes. In the example below "COM2" is not in use and will be
selected. Having the mode selected as "Dynamic" will allow the user to successfully connect to each
of the PC's USB ports.

✉ 

https://www.xrite.com/service-support/com_port_mapping_with_keyspan_usb_serial_adapter 3/6
8/15/23, 10:35 PM COM Port Mapping with Keyspan USB Serial Adapter

After selecting another COM port ("COM2" in this example), confirmation window will appear.
Make sure that you are selecting the proper COM port (in this example: "COM2") and then select
"Yes" to continue.

A second confirmation window will indicate a successful remapping of the port. Select "OK".

✉ 

https://www.xrite.com/service-support/com_port_mapping_with_keyspan_usb_serial_adapter 4/6
8/15/23, 10:35 PM COM Port Mapping with Keyspan USB Serial Adapter

Restart the computer for these settings to take place.

More detailed information on the "Keyspan USB Serial Adapter Assistant" can be accessed by
selecting the "Help" tab and then selecting the "User's Manual" as shown below:

Cholesterol 1,2,3 COM Port Setup -

If the customer is using PREVU* software, the spectrophotometer will automatically select the proper
COM port as long as it is "COM1", "COM2", "COM3", or "COM4". If the customer is using
Cholesterol 1,2,3 software and they still receive the following error while trying to open the software:

✉ 

https://www.xrite.com/service-support/com_port_mapping_with_keyspan_usb_serial_adapter 5/6
8/15/23, 10:35 PM COM Port Mapping with Keyspan USB Serial Adapter

Select "OK" to the "Error" window and the Cholesterol 1,2,3 home screen will appear. Click on the
"Options" icon as circled in red below:

In the resulting window, select the proper COM port ("COM2" in this example)and then select
"OK".

Close out of the software and then reopen it to test. No communications error should appear if
properly configured.