Professional Documents
Culture Documents
Laboratory Evaluation of Heat and Moistu
Laboratory Evaluation of Heat and Moistu
moisture Exchangers
Toyoki KUGIMIYA, Tran Gock PHUC* and Katsuo NUMATA
Humidification of gases during mechani- Of these, heated humidifiers are the most
cal ventilation and, recently, during endotra- widely used because of their effectiveness.
cheal general anesthesia is widely accepted But they are not without some disadvan-
and practiced", The benefits of humidifi- tages and complications, which are listed in
cation include the preservation of mucocil- table 16 - 10 . Heat and moisture exchangers
iary function, the maintenance of pulmonary (HME) have been used since the early 1960s
mechanics, and the conservation of body to warm and humidify gases in patients with
temperature'<". Failure to humidify in- tracheostomies!". The HME is used during
spiratory gases may incur alteration of pul- mechanical ventilation and general anesthe-
monary mechanics and body temperature, sia by placing it between the endotracheal
and may lead to increases in pulmonary tube and the ventilator or anesthesia circuit.
arteriovenous shunting, with systemic arte-
rial oxygen desaturatiorr'v", and decreases Table 1. Problems associated with Heated
in compliances", surfactant 2 ,28 , and ciliary Humidifiers
transport/P".
Immobile
Many devices have been manufactured to
Electricity required
supply heat and humidity to inspired gases". Electrical shock
Department of Anesthesiology, Faculty of "Rain out" in respiratory circuit
Medicine University of Tokyo, Tokyo, Japan Temperature monitoring required
*Metran Ltd. Bacterial contamination
Address reprint requests to Dr. Kugimiya: De- Overhydration
partment of Anesthesiology, Faculty of Medicine, Hyperthermia
University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tracheal burns
Tokyo, 113 Japan Incorrect connection
The HME captures heat and moisture from hygroscopic condenser humidifiers but Pall
the patient's respiratory tract carried in the HME 15-22 and Terumo Breathaid did not
expiratory gases and returns collected heat incorporate that principle. To evaluate the
and moisture to the cool, dry gases subse- efficiency of each HME, we constructed the
quently inspired by the patient. Thus, the test rig shown in figure 1. The humidity of
HME is also called the "artificial nose". the inspiratory gas was reduced to zero by
Although there have been many studies passing the gas through an inspiratory limb
comparing the effectiveness of HME in dif- drying agent chamber. The dried gas then
ferent clinical and laboratory settings 12 - 26 , passed through the HME being tested and
we conducted our laboratory study on six entered an artificial lung consisting of a s-
commercially available heat and moisture ex- liter anesthesia bag and water bath (Bennett
changers in order to determine and compare, Cascade Humidifier).
first, their efficiency to retain humidity from The temperature of the water bath was
the respiratory tract and, second, their con- maintained at a heat level necessary to keep
tribution"to airway resistance. the temperature at 31°C immediately proxi-
mal to the HME (at temperature point "c").
Materials and Methods In this way, we controlled the "humidity
The six different heat and moisture ex- load" presented to the HME. This humidity
changers we evaluated are listed in table 2 load was identical for each HME and can
with their physical characteristics. Four were be calculated as the arithmetic product of
82 Kugimiya et al J Anesth 1989
the water content of the saturated gas at determined as the difference in the weight
the prevailing temperature of 31 DC (33mg/L) of the artificial lung before and after each
and the total flow (400L /hr). Thus, the experiment run. (6wt 1 = initial weight of
humidity load was identical for each HME artificial lung minus final weight of artificial
(13.2g/hr). lung).
Copper wire mesh was incorporated in Water loss was also calculated by record-
the water bath and anesthesia bag to stabi- ing the changes in the weights of the HME
lize temperature and humidity. Expired gases ("e" in figure 1) and expiratory drying agent
passed through each HME where heat and chamber (enclosed in the double-dotted line
humidity could be reclaimed before these in figure 1 and labeled "B"): 6wt2 = 6wt
gases entered the distal expiratory limb. of HME plus 6 wt of expiratory drying agent
Moisture that the HME failed to reclaim chamber. We difined the "water discrep-
was captured within an expiratory limb dry- ancy" as the arithmetic difference between
ing agent chamber (labeled "B" in figure 1 these two figures: water discrepancy = 6wtl
and enclosed by a double-dotted line). Each munus 6wt2. The calculated water discrep-
HME was tested at a tidal volume of 666 ml ancy was less than 10% of the numerical
at a ventilation rate of 10 per minute (hourly value of 6 wtl for each experimental run.
ventilation = 400 L/hr). Peak inspiratory The observed hourly water loss during each
flow rate was adjusted to 40 L/min. The test run was divided by hourly flow (400
amount of water lost from the artificial lung L/hr) to generate a figure for water loss per
(that portion of the test rig labeled "D" in liter. The "moisture output" of each HME
figure 1 and enclosed by the dotted line) was can be calculated as the difference between
Vol 3, No 1 Laboratory evaluation of HME 83
its "humidity load" and its "water loss" increase in resistance after water immersion.
(moisture output = humidity load minus wa- Both the Pall and the Gambro-Engstrom de-
ter loss). "Efficiency" was calculated as the vices showed a small increase in resistance
ratio of moisture output to humidity load after water immersion. The Terumo device
(%efficiency = [moisture output j humidity did not show any increase in resistance. Be-
load] x 100). Water content figures were de- cause we had noted that the NMI device
rived from a standard water vapor content showed wide variances in initial resistance,
table. we measured a total of five samples to as-
Water loss and efficiency figures were also certain this variance which is reflected in the
generated in the absence of any HME under greater standard deviations. We found that
the test conditions described earlier. the color of the insert affected resistance.
An electronic balance (ALSEP model EX- The deeper the color, the higher the resis-
8000A) was used to determine weights of tance. Quality control problems seem to exist
various test rig components. Three samples in the manufacturing processes for the NMI.
Of each HME were evaluated. Resistances
Discussion
to airflow were measured by recording the
pressure drop across each HME in a dry Chalon et al. B found that the rmmmum
and a wet state at a flowrate of 50 L j min humidity level necessary to prevent ciliary
and 60 Ljmin. Wet testing was performed by cellular morphologic changes during anesthe-
immersing each HME in water at 31°C for 10 sia was 14 mgjL. However, it has been shown
min. Wet testing was performed in order to that physiologic value of 25-28 mgjL of hu-
simulate .prolonged use of each HME under midity is desirable to prevent alteration in
actual clinical conditions. pulmonary mechanics and body temperature
as the larynx is normally exposed to this
Results
level of moisture'". Another study showed
Table 3 shows the inspiratory gas temper- a minimum level of 23 mgjL (100% RH at
ature, the expiratory gas temperature, the 25°C) to be acceptablef". Recently, ECRI
water loss from the artificial lung in gH 20 set a minimum requirement for the output
per hour, the water loss from the artificial of HME to be 21-24 mgH 2 0 at 27-29°C to
lung in mgH 20 per liter of ventilatory gas, provide humidity for long term ventilation
the .moisture output of the HME in mgH 20 needs of patients'P.
per liter of gas at the expiratory gas tem- Our aim in this study was to evaluate
perature and the water retention efficiency the water retaining capacity of each HME by
in %. All parameters are expressed in Mean loading them with the same level of absolute
± Standard Deviations. The Pall and all the humidity at an identical temperature. Thus,
hygroscopic HME showed satisfactory results we were able to avoid the measurement of
with respect to their water retaining effi- humidity with a hygrometer, the response
ciency. The Terumo device functioned very time of which is not fast enough to suit
poorly and its water retaining capacity was our purposes. The average water output and
only half as much as the other HME. Table 4 water retaining efficiency was derived from
shows the pressure drop through each HME. the water loss from the artificial lung and
The NMI device showed higher resistance the moisture contet was read from the table.
even under dry conditions. The resistance of We believe our test methods enabled us to
NMI increased greatly after immersing them obtain an objective assessment of HME.
in the water for 10 min, showing the highest Our results confirmed that the Pall, Por-
resistance of all HME tested. The Siemens tex, Siemens, NMI and Cambro-Engstrom
device also showed an appreciable increase devices were acceptable for clinical use. Of
after water immersion. In fact, its absolute these, the NMI and Cambro-Engstrom de-
increase was even larger than the NMI de- vices were the most efficient in moisture
vice. The Portex device also showed some retaining capacity. However, the NMI device
84 K ugimiya et al J Anesth 1989
tracheal anesthesia. Anesth Analg 62:758- 25. Shelly M, Bethune DW, Latimer RD: A
763, 1983 comparison of five heat and moisture ex-
20. Chalon J, Markham JP, Ali MM, Ra- changers. Anaesthesia 41:527-532, 1986
mananthan S, Turndorf H: The Pall Ulti- 26. Branson RD, Hurst JM: Laboratory eval-
pore breathing circuit filter - An effecient uation of moisture output of seven airway
heat and moisture exchanger. Anesth Analg heat and moisture exchangers. Resp Care
63:366-370, 1984 32:741-747,1987
21. Leigh JM, White MG: A new condenser 27. Dery R: The evolution of heat and moisture
humidifier. Anaesthesia 39:492-493, 1984 in the respiratory tract during anesthe-
22. Mebius C: A comparative study of dis- sia with a non-rebreathing system. Can
posable humidifiers. Acta Anaesth Scand Anaesth Soc J 20:296-309, 1973
27:403-409, 1983 28. Tsuda T, Noguchi H, Takumi Y, Aochi 0:
23. MacIntyre NR, Anderson HR, Silver RM, Optimum humidification of air administered
Schuler FR, Coleman RE: Pulmonary func- to a tracheostomy in dogs. Scanning elec-
tion in mechanically ventilated patients dur- tron microscopy and surfactant studies. Brit
ing 24 hour use of a hygroscopic condenser J Anaesth 49:965-977, 1977
humidifier. Chest 84:560-564, 1983 29. Rashad K, Nilson K, Hunt HH Jr, Graff
24. Ogino M, Koplic R, Mannino FL: Moisture TD, Benson DW: Effect of humidification
conserving efficiency of condenser humidi- of anesthetic gases on static compliance.
fiers. Anaesthesia 40:990-995, 1985 Anesth Analg 46:127-133, 1967