The document provides technical specifications for a cooling cabinet, including dimensions of 7x8x8.5 feet for the external chiller room. The cabinet is required to maintain temperatures between +2 to +8°C while storing chemical or biological agents, with a maximum ambient temperature of 43°C. It has one hinged door measuring 34x78 inches and uses a split refrigeration system with a 15,000 BTU/hr air-cooled compressor unit to provide cooling.
The document provides technical specifications for a cooling cabinet, including dimensions of 7x8x8.5 feet for the external chiller room. The cabinet is required to maintain temperatures between +2 to +8°C while storing chemical or biological agents, with a maximum ambient temperature of 43°C. It has one hinged door measuring 34x78 inches and uses a split refrigeration system with a 15,000 BTU/hr air-cooled compressor unit to provide cooling.
The document provides technical specifications for a cooling cabinet, including dimensions of 7x8x8.5 feet for the external chiller room. The cabinet is required to maintain temperatures between +2 to +8°C while storing chemical or biological agents, with a maximum ambient temperature of 43°C. It has one hinged door measuring 34x78 inches and uses a split refrigeration system with a 15,000 BTU/hr air-cooled compressor unit to provide cooling.
Sr. No. Description A Technical Details of Cooling Cabinet 1 External Room Size - Chiller Room - Inches 84'' x 96'' x 102'' 2 External Room Size - Chiller Room – Feet 7.00 x 8.00 x 8.50 3 External Room Size - Chiller Room – Meter 2.13 x 2.44 x 2.59 4 Room Temperature required +2 to +8 °C 5 Max Ambient Temperature 43°C 6 Internal Volume in CFT 389 cubic feet 7 Product to be stored Chemical/ biological agent 8 Incoming Product Temperature +20 to +30 °C range 9 Cooling Time 24 Hours 10 Door Openings 5-10 times per day 11 Lighting Load 2 Watts/sqmt B PUF Panel Details 1 Material Of Insulation Polyurethane Foam (PUF) with Cyclopentane blowing agent 2 Density of panels 40 +/- 2 Kg/cubic meter 3 Thickness Of Insulation for walls/Ceiling 60 mm 4 Thickness Of Insulation for floor 60 mm 5 Lamination for Walls SS 304 Sheet inside and outside 6 Lamination for Ceiling SS 304 Sheet inside and outside 7 Lamination for floor FLOOR - P P OUTER - 18G STAINLESS STEEL INNER C Door Details (Blue Star make) 1 Door Clear Opening 34" x 78" 2 Insulation Thickness 60 mm 3 Type Of Door Hinged Door with Manual Swing 4 Number of doors 01 Nos 5 Lamination for door SS 304 Sheet inside and outside D Refrigeration System Details 1 Type of System Split 2 Type of condenser Air cooled 3 Type of compressor Hermetic 4 Total Capacity Of Units 15000 BTU/Hr per unit 5 No. of Units 1 No. (Fully SS Indoor unit with 100% copper inside) 6 Power consumption 3.1 KWH (3.1 units per hour) – Typical value 7 Refrigerant R22 8 Compressor Run Time 16 – 18 hrs. if unit runs 24 Hours 9 Electricity Power Supply 415 V/ 3 Ph /50 Hz 10 Temperature indicator Digital Temperature indicator cum controller 11 Defrost Cycle Air Defrost