Automated Clinical Chemistry Analyzer: Parameters
Automated Clinical Chemistry Analyzer: Parameters
Calculations
Calculated Parameter Indication Unit Equation
mg/dL LDL-C = TCHO value - (HDL-C vlaue + TG value/5)
LDL Cholesterol LDL-C
mmol/L LDL-C = TCHO value - (HDL-C value + TG value/2.2)
non-HDL Cholesterol non-HDL-C mg/dL or mmol/L non-HDL = TCHO value - HDL-C value
Globulin GLOB g/dL or g/L GLOB = TP value - ALB value
Albumin/Globulin ratio ALB/GLOB - ALB/GLOB = ALB value / (TP value - ALB value)
BUN/Creatinine ratio BUN/CRE - BUN/CRE = BUN value / CRE value
GOT/GPT ratio (AST/ALT ratio) GOT/GPT (AST/ALT) - GOT/GPT=GOT value / GPT value (AST/ALT=AST value / ALT value)
Sodium/Potassium ratio Na/K - Na/K=Na value / K value
Anion Gap Anion Gap mEq/L or mmol/L Anion Gap = Na value - (Cl value + TCO2 value)
3
Improved workflow for setting consumables
with the Consumable Disk.
Large, 7-inch, intuitive LCD touch panel 3-Step easy measurement Set the slide and the specimen, then press the Start key.
All the processes hereafter are fully automated.
The simple and easy-to-use operation screen provides a swift transition
speed of 0.5 seconds or less for stress-free operation.
500
❸ Press
❶ Set slide ❷ Set specimen
the Start key
Continuous measurement
470
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17
As soon as the pipetting procedure for 1st sample Preparation for 2nd sample
Width
the first sample is completed, the slides Auto pipetting Auto pipetting
for the next sample can already be set
Measurement Measurement
and the test started. This minimizes
450
Results are shown Results are shown
450
Setup screen Display of results Help screen waiting times for patients.
NX600 scale for installation (mm) Rotate the catalog to perform measurement.
0
0 50 100
100 150 200
200 250 300
300 Depth 360
350 370 Height 400
400
4 5
Dilute solution
Place dedicated
tube here
Results
print out.
Mixing cup
Tips
Up to 12 tips (3 to 12 samples)
can be set at once
More compact.
Set consumables for multiple samples at once
A sliding front cover eliminates the Previous The layout is simpler than our previous models, and this makes it easier to set consumables.
need for a large space above the model 420 NX600 370
(NX500)
mm mm Multiple tips can be set at once, which eliminates the need to set reference solution for each
device to open and close the cover.
measurement. This reduces the frequency of consumable replacement.
Functions
Patient-friendly testing Easy lot compensation Wide range of usable blood Minimize the risk of
by QC card collection tubes biological hazard
Each test needs only 10µL of sample.
10 μL
/ test
(CRP needs 5µL/test, ISE needs 50µL/3
tests). Manual pipetting can be also
performed when the sample is less.
Corrections are made simply by
reading the QC card included in
In addition to Fuji tubes (0.5/1.5 mL),
commercially available blood collection
Slide reagents after measurement are
automatically discarded to the disposal
box, minimizing the risk of contamination.
the slide package into the main unit. tubes can be used as specimen tubes.
Less invasive for newborns at NICU. *CRP: Calibration is required. *Please select suitable sample racks for your
*50 µL are used for the simultaneous I S E : Calibration and correction are not work flow. (refer last page:Option
measurement of 3 electrolyte items. necessary. QC card is not attached. items/Sample Rack)
Plasma Filter
Reliable technology ensures test quality. Aspiration
Plasma
accumulation
Blood separation in 1 minute 1 Read PF card.
C
Blood cell separating
membrane
2 Set PF on the top of B Glass fiber layer for
Plasma Filter (PF) can cut the turn around time sample tube. (A) blood cell separating
The FUJI DRI-CHEM slide reagent has high reliability and stability thanks to fine chemical technology and the pre-treatment process of the sample.
3 Press Start.
cultivated through FUJIFILM’s long history in photographic film manufacturing. It can generate a plasma sample by aspirating
and separating the whole blood inside the PF 4 Suction unit move to be
Less variation of results between operators, high result reproducibility and daily precision, connected on PF and start
within 1 minute. Just set the PF on top of the
and excellent correlation with wet chemistry are its remarkable features. aspiration of whole blood. (B ) Sample Tube
sample tube and press START.
5 Whole blood is separated A
* This function is not available in the NX600i at the glass fiber layer in
PF to sample plasma. (C)
FUJI DRI-CHEM SLIDE Two types of slides are provided for different test items.
* TCO2: not applicable
500
FUJI DRI-CHEM (U/L)
800 8.0
5.0
400 Series name FUJI DRI-CHEM NX600 FUJI DRI-CHEM NX700
600 6.0 4.0
300 Width ✕ depth ✕ height 470(W) ✕ 360(D) ✕ 370(H) mm 500(W) ✕ 380(D) ✕ 410(H) mm
400 4.0 3.0
r = 0.999 200 r = 0.999 r = 0.999 Weight 28 kg 33 kg
2.0 r = 0.994
200 Syx = 7.518 Syx = 8.143 2.0 Syx = 0.071 Syx = 0.199
100 1.0 Number of measurement cells 13 colorimetric + 1 electrolyte (independent) 13 colorimetric + 1 electrolyte (independent)
n = 89 n = 55 n = 57 n = 33
0 0 0.0 0.0 Throughput (1) (only colorimetry) 120 tests/hour 180 tests/hour
0 200 400 600 800 1000 0 100 200 300 400 500 600 0.0 2.0 4.0 6.0 8.0 10.0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0
Throughput (2) (colorimetry + electrolytes) 128 tests/hour 190 tests/hour
IFCC standardization method (no PALP) (U/L) Hexokinase method (mg/dL) Ion-selective electrode method/ Latex immunoturbidimetry (mg/dL)
dilution method (mEq/L) Throughput (3) (15-item measurements) approx. 9 minutes approx. 9 minutes
Number of specimen that can be set 1 specimen 5 specimens at the same time