Professional Documents
Culture Documents
Building Agilent GC-MSD Deconvolution Reporting Libraries For Any Application Technical Overview
Building Agilent GC-MSD Deconvolution Reporting Libraries For Any Application Technical Overview
Building Agilent GC-MSD Deconvolution Reporting Libraries For Any Application Technical Overview
2
Procedure for Building Databases and
Libraries*
Create the *.TAB File
1. Load a data file (for this example, oxymix_2.d,
see Figure below) which consists of the peaks
(spectra) of standard compounds for the library
entries. Make sure no peaks were overlapped in
the data file causing contaminated spectra. In
situations where clean spectra of target com-
pounds cannot be obtained (for example, over-
lapping peaks), refer to the DRS User
Information section “Adding Compounds to
Existing AMDIS_32 Libraries” for instructions
to deconvolute the spectra before adding them
to the library.
*A manual procedure for building the library is also available; please refer to applica-
tion note “Retention Time Locking: Creating Custom Retention Time Locked
Screener Libraries” [5] for further information.
3
2. Create a *.tab file using Microsoft Excel. Refer
to Table 1 for explanation of columns in the
*.tab file [4]. For column F, sort all entries in
*.tab file into ascending retention time (RT)
order before saving, this is especially important
if the library is to be created from multiple
datafiles. See the Figure below.
Note : Sort all entries in .tab file into ascending RT order before saving – especially
important if the library is to be created from multiple datafiles
4
3. Save the file in *.tab type in C:\database.
Go to File | Save as | and choose Text (Tab
delimited) (*.txt). The file name must be
enclosed in “ as shown.
Path C:\DATABASE
Type in “oxy_original.tab”
5
5. Close Excel, otherwise you will get an error
message (Could not open file
C:\database\oxy_original.tab) from MSD
ChemStation.
6. Click No.
6
3. Select oxy_original.tab in C:\database, click
Open.
Type in oxymix.L
Type in C:\DATABASE\oxymix.SAV
7
The figure below shows the new library, oxymix.L,
oxymix.scd, oxy_original.tab files and the *.sav
directory.
8
The *.scd file, found in the database folder, could
be used to create Calibration Table (quantitation
database) automatically; refer to Appendix A for
further information.
9
3. Click No to the detailed graphics question, as
shown below.
10
Create *.MSP from *.L
1. Go to Windows Start, Programs, NIST Mass Spectral
Database and click Lib2NIST Converter.
11
3. Select folder “C:\NIST05\MSSEARCH” for NIST
Library and Output, select Text File [.MSP] for
Output Format, highlight the library in the left
panel, then click Convert.
12
Convert *.MSP to *.MSL
1. As shown below, go to Windows Start,
Programs, NIST Mass Spectral Database and
then click AMDIS_32.
13
4. You will be prompted with a Transfer from
Library to Library window, click Files. A
subsequent Files window will display.
14
5. In Files window,
click Load Library… ,
change the file type to *.MSP,
select OXYMIX.MSP in C:\NIST05\MSSEARCH,
then click Open.
Back to Files window,
click Create New Library… ,
select folder C:\NIST05\AMDIS32\LIB,
save files as type Target Library(*.MSL),
type in oxymix.MSL for File name,
click OK.
Path
C:\NIST05\MSSEARCH
Select oxymix.MSP
Path
C:\NIST05\AMDIS32\LIB
6. Click Close.
15
7. Highlight all compounds on left and click
Transfer->. The *.MSL file is saved in
C:\NIST05\AMDIS32\LIB.
16
2. In Files window,
click Load Library…,
select “oxymix.MSL in C:\NIST05\AMDIS32\LIB,
then click Open.
Back to Files window,
click Create New Library…,
select folder C:\NIST05\AMDIS32\LIB,
save files as type RI Calib Library(*.CSL),
type in oxymix.CSL for File name,
click OK.
Path
C:\NIST05\AMDIS32\LIB
Select oxymix.MSL
Path
C:\NIST05\AMDIS32\LIB
3. Click Close.
17
4. Highlight all compounds on left and click
Transfer->.
18
Using AMDIS Interactively
1. Open AMDIS, click Analyze and then
Settings… shown below.
19
Select oxymix.MSL in
C:\NIST05\AMDIS32\LIB and then click Open.
Path
C:\NIST05\AMDIS32\LIB
20
3. Highlight RI Calibration Library and then click
Select New.
Path
C:\NIST05\AMDIS32\LIB
21
Confirm the proper *.CSL is selected
22
Do not pick one of the existing file names. Type in
a unique name for a *.cal file to be created.
Path
C:\NIST05\AMDIS32\LIB
23
5. Click the “Identif.” tab, type in Minimum Match
Factor suitable for your analysis (for example,
60) and highlight RI Calibration/Performance
in Type of analysis. This setting is used to
create the *.CAL file.
6. Click Save.
24
7. Click Yes.
8. Close AMDIS.
25
2. Click Method Associate Settings, and then
New Method Associate Settings ….
26
4. Select oxymix.MSL as AMDIS target library,
and click Open.
27
6. Select oxymix.CAL” as RI Calibration Data,
and click Open.
28
8. Select onsite.ini as AMDIS Initialization
Settings File.
29
10.Click Exit, and then Exit and Save.
Note : AMDIS only recognizes and uses the onsite.ini settings file when it runs automatically. The DRS software copies
the settings file specified here to onsite.ini before it deconvolutes the chromatogram
References
1. V. Giarocco, B. Quimby, and M. Klee, “Retention
Time Locking: Concepts and Applications”,
Agilent Technologies, publication 5966–2469E,
www.agilent.com/chem
2. H. Prest, P. Wylie, K. Weiner, and D. Agnew,
“Efficient Screening for Pesticides and
Endocrine Disrupters Using the 6890/5973
GC/MSD System”, Agilent Technologies,
publication 5968–4884E, www.agilent.com/chem
3. P. Wylie, M. Szelewski, and C.K. Meng, “Compre-
hensive Pesticide Screening by GC/MSD Using
Deconvolution Reporting Software”, Agilent
Technologies, publication 5989–1157EN,
www.agilent.com/chem
4. Michael J. Szelewski, Kenneth R. Weiner, and
Chin-Kai Meng, “Building and Editing
RTL/Screener/Quant Databases and Libraries”,
Agilent Technologies, publication 5989–0916EN,
www.agilent.com/chem
5. Kenneth R. Weiner and Harry F. Prest, “Reten-
tion Time Locking: Creating Custom Retention
Time Locked Screener Libraries”, Agilent
Technologies, publication 5968–8657E,
www.agilent.com/chem
30
Appendix A
31
www.agilent.com/chem
3. Copy oxymix.SCD from the C:\database
directory into relevant *.M (oxyquant.M)
directory.
Reload the method from MSD DataAnalysis pro- Printed in the USA
gram to review the calibration table. Make sure October 3, 2005
5989-2249EN
that the CAS number is entered on page 2 for each
compound - no leading zeroes, no dashes, etc.