M229 Lecture5 PulseSeqRAREandSim
M229 Lecture5 PulseSeqRAREandSim
• Final project
- start thinking
- come to office hours
- discussion in class next Tue 4/23
- presentation on 6/13 Thu
• Office hours
• Homework 1 due 4/26
• Homework 2 due 5/3
Outline
• Rapid GRE
- gradient and RF-spoiled GRE
τ τ
T2 decay
TE = 2τ
TR
SE
Spin Echo
• Image contrast
- Based on TE, TR
- T1w, T2w, PDw
- Can augment with prep pulses
• Scan time
- TSE = Npe x TR
- TR = 1000 ms, Npe = 256: TSE = 4+ min
- usually combined with 2D multislice acq
Multi-echo Spin Echo
90o 180o 180o 180o
...
τ τ 2τ 2τ
T2 decay
TE1 = 2τ TE2 = 4τ TE3 = 6τ
...
TR
SE1 SE2 SE3
Can perform T2 mapping.
RARE (Turbo Spin Echo)
90o 180o 180o 180o
...
PE
TEeff
...
ky ky ky TR
kx kx kx
CPMG Conditions
• Carr-Purcell-Meiboom-Gill conditions
- ensure echoes only occur at desired
positions in the sequence, and
- signals at each position have the same
phase
• When satisfied
- SE and STE coincide (same phase)
- secondary SE and FID are crushed
90ox FP, T2
x y x y x y
z z
180ox FP, T2
x y x y
Spin Echo: 90 x
o - 180 y
o
z z z
90ox FP, T2
x y x y x y
z z
180oy FP, T2
x y x y
TSE Sequence Params
90ox 180oy 180oy 180oy
...
ESP = 2τ
TEeff ETL
...
TR
x Nshot
Echo train length (ETL) Number of shots (Nshot)
Echo spacing (ESP) Effective TE (TEeff)
TSE Sequence Params
• Scan time
- Recall TSE = Npe x TRSE
- Nshot = Npe / ETL
- TTSE = Nshot x TRTSE = (TSE / ETL) x (TRTSE/TRSE)
• Example: 3D volume
- Npe = 256*256; ETL = 32; Nshot = 2048
- TR = 1000 ms: TTSE = 34 min
TSE Image Contrast
• TE eff, TR
- T1w, T2w, PDw
- PE ordering affects TEeff
• Bright fat
- J-coupling of protons in lipids (CH3-CH2-);
fCS ~ 25 Hz, fJ ~ 7 Hz @ 1.5 T
- S = S0 * exp(-t/T2) * cos(nech π fJ ESP)
- Shortening of apparent T2 (in SE)
Bright Fat
• Magnetization transfer
- MT effect
• Partial echo
• Multi-echo
• Mag-prep
Extensions and Variations
• Partial Fourier
- Sample ~half of k-space data, reconstruct
assuming Hermitian symmetry (real-valued
MR images)
- reduce refocusing pulses, reduce SAR
- better control of TEeff
• Parallel imaging
- Undersample k-space data, reconstruct
using information from multiple coils
- reduce refocusing pulses, reduce SAR
Related Sequences
• Bloch Equations
- RF excitation
- T1, T2 decay
- free precession
- gradient pulse
Bloch Simulation
Rotation:
2 3 2 3
1 0 0 cos ✓ 0 sin ✓
Rx (✓) = 4 0 cos ✓ sin ✓ 5 Ry (✓) = 4 0 1 0 5
0 sin ✓ cos ✓ sin ✓ 0 cos ✓
2 3
cos ✓ sin ✓ 0
Rz (✓) = 4 sin ✓ cos ✓ 0 5
0 0 1
2 3
cos !0 t sin !0 t 0
Rz (!0 t) = 4 sin !0 t cos !0 t 0 5
0 0 1
'
x y
• Homework 2, part 1A
- Steady state for bSSFP, SSFP-FID and
SSFP-Echo
Gradient-spoiled GRE
SS signal as a function of off-resonance:
bSSFP GRE (SSFP-FID)
• Homework 2, part 1B
- Transition to steady state for bSSFP
- catalyzation schemes
Balanced SSFP
Transition to steady state:
TR = 5 ms
Δϕ = π
θ = 60o
• Web resources
- ISMRM 2010 Edu: Miller, Weigel
- ISMRM 2011 Edu: Miller, Weigel
• Further reading
- Bernstein et al., Handbook of MRI Sequences
- Haacke et al., Magnetic Resonance Imaging
- Scheffler, Concepts in MR 1999; 11:291-304
- Hennig, JMR 1988; 78:397-407
Thanks!
• Acknowledgments
- Brian Hargreaves
• Next lecture
- EPG and MATLAB demo