@iyleoJlg peas auiaill aie] auuugall
TS Inspection & Technical Survey Est.
C.R. 2050060083 - C.C.No. 120676
Page: 12
180 9001 Certifeation
CERTIFICATE OF CALIBRATION
Certificate No. : 20190526-05N
Customer: AL SYED COMPANY
Address: Dammam, Kingdom of Saudi Arabia
‘This Cerificate of Calibration certifies tha he instrument below has been tasted/calibrated using equipment traceable to NPL, NIST,
‘SASO or other Natonalntemational standards or verfod by ratio type measurements using self calling techniques. Information on
the callraon procass & the results are produced below or are gven inthe fellowing pages of tis document. Al related traceable
documents are on feat ITS (IC Division) and may be examined on request
Instrument Details
Instrument Name : Pressure Gauge
Manufacturer/Brand : Wika
Model/Type : EN837-1
Range/Capacity :0- 360 psi
Resolution 40 psi
Serial No. 10526
Identification No. : Not indicated
Remarks : Passed
Calibration Date : 26 May 2019
Recommended Due Date: 25 November 2019
@)
Nasir Shari
Calibrated By:
QVALITY CONTR
Checked By: CHECKED
QAIac
Signed for and on behatf of
Inspection and Technical Suwvey Est
“Te Measaemnlrevuli parain ory fo the calibaied ineraent athe tine of alban
‘This certiieae & the attached measurement result sheets shall not be reproduced exept in fll without the prio: pemission of TS
P.O. Ba: 8116 - DAMMAM 32436 - Kingdom of Saudi Arabia - Tel.: +966 138376604 - Fak S681 38376605,
www. its-arbia.comPage? 2
‘Manufacturer: Wika Certificate No, : 20190526-05N
Device Serial Nos 10526 Identification No.: Not indicated
Test Result:
Gauge Set Point | Standard Reading (inereaeing) Standard Reading (Decreasing) | Average! Ear
psi Se Xt a= [= DSi si
0.0 ‘0.0 0.0 0.0, 0.0. 0.0 00) 0.000, 0.000
20.0 30.0 90.0 20.0, 30.0 90.0_| 90.0 90.000 0.000,
180.0 1810 | 181.0 | 180.0 | 180.0_| 179.0 | 190.0 [180.167 0.167
270.0) 271.0 | 2720 | 2700 | 268.0 | 268.0 | 269.0 | 269.833 0.167
360.0, 358.0 | 3590 | 3580 | 368.0 | 359.0 | 3580 | 358.333, =1.667
Uncertainty of Measurement: + 2.0 % of full scale reading at 96% confidence level (k=2)
Condition of Tested Unit
[Within the tolerance on receipt (as found) (Within the tolerance but there is sighi crack on the glass.
[The instrument was adjusted No]
[The instrument was repaired No]
[Within the tolerance on dispatch (as left) Full scale was tested using CM-33 Master with the scope of 0.08%|
Calibration Description and Information
[Test method [The measuring instrument was checked and calibrated atITS Laboratory by using he
reference standards traceable to NIST and/or S|. The test was performed in accordance with
lapplicable formulated procedure based from below standard
[Calibration Procedure |ITS-CLP-O1 EURAMET CG-17,
[Environment's conditions |Temperature 20° 22°C Relative Humidity 60% 210%
[other ‘The readings given are the resulls at the time of testing and do not carry any Implication
regarding the long term stability ofthe instrument.
Reference Standard Used
(Description its > [Cal Due Date [Traceability (Certificate
[Pressure Calibrator icw-30 26 October 2020 | Primary Measurement [2017013621
[Pressure Calibrator icw-s3 28 October 2020_|- lpov70138522
Calibration Technician
-End of certiticate-
FATS-t-002