BS-480 Operation Card V1.0 en

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Operation Card of BS-480 Chemistry Analyzer

1 Check before Powering On cover is installed correctly.

Note: Every day before analysis, remove the


5 Calibration
plug on the reagent carousel in order to prevent Run calibrators when needed.
mechanical reset failure and bending reagent
Select Reagent-Cal F5 to request calibrations.
probe.
According to the printed calibrator list, load calibrators
a. Check the water supply and make sure that water in assigned positions on the sample carousel.
can be supplied continuously based on the Select to start the analysis.
consumption of no greater than 20L/H”. Select Reagent-Biochemistry Calibration or ISE
b. Check if the connections between the water supply, Calibration to recall calibration results.
water supply module, drainage module, and 6 Reagent Blank
analyzer are correct and tight.
c. Check if the power supply is available and can Run reagent blanks when needed.
provide correct voltage. Select Reagent-Cal F5-Rgt Blk to request reagent
d. Check the connections between operation unit and blanks.
printer. Make sure the connections are correct and Select to start the analysis.
secure. Check the power cords of the analyzing Select Reagent-Biochemistry Calibration to recall
unit, operation unit and printer and make sure they reagent blank results.
are well connected to the power sockets. 7 Quality Control
e. Check if sufficient printing paper is prepared in the
printer. If not, refill the printing paper. Select Program-Control to program control samples.
f. Check if the high-concentration waste tank has Load controls in assigned positions on the sample
been emptied. If not, empty it. carousel.
High-concentration waste: 1.5L/H. Select to start the analysis.
g. Check if the low-concentration waste tubing is not Select QC-Levey-Jennings or Twin-Plot or Results to
bent and the sewer opening is lower than the waste recall control results.
outlet of the system. 8 Programming Routine Samples
h. Check the sample probe, reagent probe, sample
 Sample programming
mixer and reagent mixer for dirt and bend. If they
are polluted, clean them; if they are bent, replace 1) Manually programming samples: (allowed in any
them. conditions)
i. Check if the reagent carousel cover is installed Select Program-Sample to program samples.
correctly, no sample cup is left in the sample
carousel. Check the concentrated/diluted wash Load samples in assigned positions on the sample
solution and sample probe/reagent probe wash carousel.
solution. If needed, refill more. 2) Program samples with LIS: (allowed only with LIS
j. Check for sufficient concentrated and diluted wash host)
solution. If necessary, fill more. Load samples on the sample carousel.
k Check positions D/D2, and W/W for reagent probe Select Program-Sample-List F5, and then select
wash solution, sample probe wash solution and Download F3 to obtain samples from the LIS host.
physiological saline. If needed, refill more or 3) Automatically programming samples: (allowed only
replace them. with LIS host and sample bar code reader)
l If an ISE module is used, check for ISE wash Load bar-coded samples to the sample carousel,
solution in position D1 on sample carousel. If 4) Batch program:
needed, refill more. Select Program-Sample, enter the first sample ID, select
2 Powering On Batch F3, enter the last sample ID, and then choose
chemistries to run.
Switch on the water/power supply in the following
order: water supply, water supply module, drainage  Select to start the analysis.
module and main power switch, analyzing unit power,
 Select Result-Current Results or History Results
printer, display monitor and computer.
to recall the current test results or those obtained
When the Windows system is started up, the BS-480 before the current day.
Operating Software will be run automatically.
9 Programming STAT Samples
3 Checking System Status  Sample programming
Check the system status (printer, biochemistry system,
ISE module and LIS host), analyzing unit status, alarm 1) Quickly programming STAT samples
status, reagent/calibration status, maintenance status,
and smart module status. Select to show the quick STAT sample
4 Preparing Reagents programming window, enter program information, and
then load the STAT samples.
Select Reagent-Load F1, set up biochemical reagents,
2) Programming common STAT samples
load them to the assigned positions on the reagent
carousel, and then remove the bottle caps. Make sure Select Program-Sample, choose chemistries to be run,
that correct bottle types are chosen so that reagent and mark the STAT checkbox with a tick.
inventory check can be done correctly. Load samples in assigned positions on the sample
After loading reagents, ensure that the reagent carousel

-1-
Operation Card of BS-480 Chemistry Analyzer

carousel. 12 Transmitting/Printing Sample Results


3) Batch program: Transmitting sample results to LIS host:
Select Program-Sample, enter the first sample ID, mark Select Result-Current Results or History Results,
the STAT checkbox with a tick, select Batch F3, enter choose desired results, and then select Host F8 to send
the last sample ID, and then choose chemistries to run. the selected results to the LIS host.
 Select to start the analysis. Printing results:
Select Result-Current Results or History Results,
 Select Result-Current Results or History Results choose desired results, and then select Print F7 to print
to recall the current test results or those obtained out the selected results.
before the current day. To print multiple samples on one report, select the
samples in the result list, and then select Options
10 Adding/Removing Samples and Chemistries F2-Print Multi-Sample Report.
Chemistries can be added to samples of any status.
 Adding samples to sample carousel
13 Reagent Status and Test Status
During analysis, you are allowed to check the reagent
Select Program-Sample, and then program added status on the Reagent/Calibration screen and to check
samples. the measuring status on the Sample-Status screen.
Select Sample Load F1 to request sample stop. When 14 Daily Maintenance
the corresponding sample carousel indicator is off, load
samples in assigned positions on the sample carousel, By the end of each day, maintain the instrument
according to the scheduled daily maintenance
and then select to start analysis. procedures and those indicated in yellow.
 Adding/Removing chemistries
15 Powering Off
Select Program-Sample, enter the sample ID, and then After exiting the Windows operating system, switch off
modify the programmed chemistries. the power supply in the following order: printer, display
Select to start the analysis. monitor, analyzing unit power, computer, water supply
module and drainage module (Optional).
11 Rerunning Samples When the analyzing unit power is switched off, the
Finished samples can be rerun in manual or auto mode. refrigeration system is still running. If you are going to
Only finished chemistries can be rerun. If a chemistry is store the system for over 7 days, switch off the main
run for more than one replicate, it cannot be rerun until power.
all replicates are finished.
The system allows samples to be rerun manually or 16 Check after Powering Off
automatically in certain conditions. 1. Remove the calibrators, controls and patient
Manual rerun: samples from the sample carousel.
 Select Result-Current Results, choose desired 2. Check the analyzer panel for stains and wipe them
samples and results, and then select Rerun F5; or off with clean gauze if any.
3. Check the high-concentration waste tank. Clear it
 Select Program-Sample-List F5,and then select if necessary.
Rerun F4.

 Select to start the analysis.

Auto rerun:
Set up auto rerun conditions on the Define/Edit
Chemistries and Reference/Critical Range windows.
When the test is finished, the system will compare the
test results with the defined conditions. If a result is
beyond the defined linearity range or critical range, the
sample will be rerun automatically.
Recalling rerun results:
Select Result-Current Results or History Results,
choose desired sample to view the relevant results, or
select Options F2-Recall Rerun Results to view all
replicate results of the selected sample.

-2-