BC300

Semi-auto Biochemistry Analyzer

Contec Medical Systems Co.,Ltd.

1
Announce

Statement
This service manual contains proprietary information, which is protected by copyright. All rights reserved.
Photocopy, reproduction, adaption or translation, for any part of the manual without prior written permission,
is prohibited.
Our company owns the final explanation right to this manual.
Our company assumes no responsibility for any error that may appear in this document, or for incidental or
consequential damage in connection with the furnishing, performance or use of this material.
Please contact us in time for applying this manual.
The content in the manual is subject to change according to actual use or product upgrades and we apologize
sincerely for inconvenience caused by version updates.
Service Policy:
All of our products have passed strict production guard, quality inspection and got high-normal quality
certification. For the problems about payment, consignment and quality in the production and management
process, our company will try our best to give proper solution actively with the attitude that "Be serious and
responsible to seek long cooperation".
Product Quality Promise:
1.Our company will give free repair for product's quality problems during warranty period, which do not
contain improper use or human damage, or directly replace by customers.
2.For products beyond warranty period, our company will repair, but charge for accessories and upkeep,
taking the circumstances into consideration.
3.For product failure caused by improper use or human damage, we will try our best to repair, but charge for
accessories and upkeep, taking the circumstances into consideration.
4.When it is necessary to replace parts during repair and adding part cost, we will give user prior notice and
ask for agreement. After the failure disappears, we will return you the parts replaced.
Convention
Warning Points some information you should know to avoid injury to patient and medical staff.

Caution Points some information you should know to avoid damage to the device.

Note Points some important information you should pay attention to.

This manual is used with related User Manual, and the same contents are not repeated. To be familiar
with the device is prerequisite to repair, so before servicing, read the device's User Manual carefully.

1
Announce

General Safety Information


Warning
 Before disassembling, ensure the external power supply have been cut off.
 There may be electrical shock when opening the device enclosure. All servicing to
the device must be carried out by personnel trained and authorized by our
company.
 Servicing personnel must wear corresponding apparatus (such as static bangle,
etc.).
 After disconnecting the power cable, turn the power on, and turn off in one second
to set free the power stored in internal DC circuits.

Caution
 Before disassembling, please turn off the power, and disassemble all the
peripheral equipments, such as power cable and other cables, etc.; for some
circuits and equipments are still working in the instance of power off and direct
disassembling will cause damage to some circuitry.
 When disassembling the plug wires (electric wires), avoid directly pulling and
drawing , first make sure how they connected and then pull out without strong
force.
 Use proper tools, such as screwdriver and tweezer, etc., and be careful during
operation to avoid any human damage to the device.
 After repair, do necessary maintenance to the device.

Note
 For most components or accessories of this device are tiny, such
as screw and washer, it is very important to strictly record the location, size and
specification for each component and accessory, and put the components
disassembled in sorts, which will help improve service efficiency.
 Disassemble the device strictly according to correct steps described in the manual.
 Please refer to the user manual for safety information.
 Other important safety information is located in this manual where is appropriate.

2
Contents

Contents
1 Overview.......................................................................................................................................................5
1.1 Brief Introduction...................................................................................................................................5
1.2 Functions and Performance...................................................................................................................5
1.3 Appearance.............................................................................................................................................6
1.4 Operation Environment..........................................................................................................................6
2 Hardware Principle.......................................................................................................................................7
2.1 Module Composing................................................................................................................................7
2.2 Module and Interface.............................................................................................................................7
3 Device Inspection.........................................................................................................................................9
3.1 Appearance Check and Main Components Inspection..........................................................................9
3.2 Basic Functions Inspection....................................................................................................................9
4 Device Disassembling/Assembling............................................................................................................10
4.1 Disassembling......................................................................................................................................10
4.1.1 Disassembling of Shells........................................................................................................................10
4.1.2 Disassembling of Core Board, Bridge Board and LCD.......................................................................11
4.1.3 Disassembling of Printer and Print Board............................................................................................12
4.1.4 Disassembling of Key Board................................................................................................................13
4.1.5 Disassembling of Shortcut Keys and Microswitch..............................................................................14
4.1.6 Disassembling of Main Control Board.................................................................................................14
4.1.7 Disassembling of Optical System.........................................................................................................14
4.1.8 Disassembling of Peristaltic Pump.......................................................................................................20
4.2 Assembling..........................................................................................................................................21
4.3 Maintenance.........................................................................................................................................25
4.4 Updating...............................................................................................................................................26
5 Troubleshooting..........................................................................................................................................27
AppendixⅠ Parameter Specification...............................................................................................................29
AppendixⅡ Internal Wiring Diagram............................................................................................................31
AppendixⅢ Product Failure Feedback Table...............................................................................................32

3
Service Manual

1 Overview

1.1 Brief Introduction

The Semi-auto Biochemistry Analyzer measures biochemical indexes by analyzing blood and other body fluid,
then combines with other clinical information, to help diagnose disease, evaluate organs function, identify
disease gene and determine the norm for future therapy.

1.2 Functions and Performance

1) Based on the Linux operating system, color graphical interfaces, large and color LCD. Operating with
touch screen, function keys and external USB mouse or keyboard, which is simple and fast.
2) Multiple measurement methods (End point, Two-point, Speed(kinetics), double wavelength
(Bichromatic)) are optional, and multiple calculation methods (factor, linear regression, curvilinear
regression, etc.) are also optional.
3) Store 500 test items and 10000 test results; supply many methods of printing and many formats of report.
4) Each item can be set for two quality-control, automatically statistic and draw QC curve, store one year's
QC data and QC chart, also can query and print.
5) Auto-dormancy for the lamp to prolong its life.
6) Power-off protective function to automatically save measurement results.
7) Self-check function for alarming to light route, liquid route and machine components, etc.
8) Standard data communication interface, be convenient for data transmission and processing.
9) Absorbency range: -0.3-3.0Abs
10) Resolution: 0.001Abs(display), 0.0001 Abs((inner calculation)
11) Light source: Halogen lamp
12) Wavelength: 340,405,510,546,578,620nm+ 2 free positions
13) Stability: ≤0.005Abs/30min
14) Half-bandwidth: ≤12nm
15) Temperature control: Room Temperature, 25℃, 30℃, 37℃
16) Temperature accuracy: ±0.1℃
17) Colorimetric cell: 30μl quartz flow colorimetric cell
18) Injection: 0~6000ul(adjustable)
19) Cross Contamination: ≤1.0%
20) Storage: 500 test items and 10000 test results
21) Interface: Standard RS232 interface, 4 USB interfaces, Ethernet interface
22) Display:7″color LCD (800*480)
23) Print: Built-in high-speed thermal printer, and connect external printer.
24) Processor: Embedded high-speed processor
25) Power Supply: 100VAC~240VAC, 50/60Hz

4
Service Manual

1.3 Appearance

Printer cover

Colorimetric cell bracket

Touch tablet and LCD

Operation panel

Colorimetric cell cover


Colorimetric cell

Figure 1.3-a Front View

Earth wire:protective ground


Fuse socket: pull out the fuse socket to
replace the fuse
Figure
Power socket: the inserted interface of
1.3-b the power cord
Rear Power switch: switch on or off the
View power
Fan: used for giving out heat
1.4 USB interface: communicating with
external equipment
Operation Environment USB interface: communicating with
external equipment
USB interface: communicating with
Temperature: 10℃-30℃ PC
Relative humidity: ≤80% RJ45 interface: connecting with
network cable for remote network
Atmospheric pressure: 86kPa~106kPa RS-232 interface:communicating with
PC
Drain port:used for discharging the
waste water

5
Service Manual

2 Hardware Principle

2.1 Module Composing

The device consists of main control board, PC, peristaltic pump control module, temperature control module,
light collection module, filter wheel control module, printer and key module.
The connection among each module is represented as Figure2.1:
Peristaltic pump Temperature Light collection
control module control module module Figure 2.1

2.2 Module and

Mian control board PC Interface

1 Main control
board
Filter wheel control Printer module Key module Communicate with PC,
module and control peristaltic
pump, filter wheel, light collection, Peltier temperature control, printer and key module.
2 Peristaltic pump control module
Used for controlling the peristaltic pump to aspirate liquid, rinse and calibrate. The PC sends the
communication protocols of each task to the lower computer for controlling peristaltic pump motor for
finishing relative operations.
3 Filter wheel control module
Control the position of filter wheel. The light rate is different for different wavelength, so select proper
wavelength is very important. When test different samples, select relative wavelength for selecting the filter in
different positions.
4 Light collection module
Used for signal collecting and processing. The photocell transforms the light signal to electric signal, after
magnifying by amplifier, finally, sends the digital signal which is transformed by AD conversion to singlechip
for processing.
5 Peltier temperature control module
Real-time collection for the temperature of the colorimetric cell, and control the temperature in set range by
PID adjustment.
6 Printer module
Print trend graph, test data, quality-control report and patient integrative report, etc.
7 PC module
Control the LCD for displaying information of test items. Parameters can be set, and it finishes relative
operations with main control board by protocols.
8 Key module
As the input equipment, it is used for user to input information.

6
Service Manual

3 Device Inspection

3.1 Appearance Check and Main Components Inspection

1)Products' shell should meet the requirement of smooth and bright surface, even color and luster, without any
defect such as mechanical damage and crackle, etc.
2)Products' panel should be no coating come off or corrosion. The text and mark on the panel should be clearly
visible, firm and smooth.
3)The plastic parts on the product should be no gas mark, cracking, deformation or phenomenon of instillations
overflow.
4)The biohazard warning sign on the rear side of the upper shell should be clear and visible.
5)The shells meet tight junction, well fixed.
6)The adjusting mechanism of product should be flexible and reliable, the fasteners should be no loose
phenomenon.
7)The product's accessories are complete, tagged correctly.

3.2 Basic Functions Inspection

Install the printing paper, insert the three-wire power cord into the power socket of the device, turn the power
switch to "ON".
1 Inspection requirement
There is not any fault alarm from turning on the power switch to entering system standby interface.
2 Inspection methods
After the buzzer sounds, the device starts working. The power indicator lights, LCD displays "loading", it will
enter the application after the startup of core board is complete, the source lamp of optical system lights. If the
printer has paper, it will make a paper walk test. After completing self-check, enter the interface which is
waiting for the stable light source, shown as Figure3.2-a. Press "CANCEL" to cancel the waiting, the
peristaltic pump will running after pressing "ASP", the rotational filter wheel switch each filter, collect light
intensity values of each wavelength. If there is a intensity value lower than alarm value, the screen will
displays error prompt. If filter wheel rotate abnormally, the screen will displays error prompt too. After
entering the standby interface, the bottom right corner of the screen displays that the temperature of the
colorimetric cell is 36-38℃. Because of the different of environment temperature, after entering the standby
interface, the temperature of the colorimetric cell need some time to be stabilized at 37℃, shown as
Figure3.2-b.

7
Service Manual

Figure3.2-a Figure3.2-b

8
Service Manual

4 Device Disassembling/Assembling

4.1 Disassembling

4.1.1 Disassembling of Shells

Use the screwdriver to disassemble the enclosures as following steps.(note: please turn off the device before

disassembling.)

1 Eliminate all residual liquid in the pipeline, and fix the flowing colorimetric cell.
2 Place a soft cloth on a flat table, then place the device on the cloth reversedly.
3 Take out the seven screws and washer assemblies(ST2.9*9.5) connected the metal pedestal and upper
enclosure, shown as Figure 4.1.1-a.

ST2.9*9.5

ST2.9*9.5

Figure 4.1.1-a
4 Place the device on the table, and open the light source cover, then disassemble the pipette and
colorimetric cell.
5 Disassemble the colorimetric cell cover and light source cover, then take the upper enclosure out
carefully.
6 Pull out the connection cable from the socket to separate the upper and nether enclosure. Pay special
attention to the cables between upper and nether enclosures, to avoid damaging the sockets for excessive force.
The structure for nether enclosure is shown as Figure 4.1.1-b.

9
Service Manual

Figure 4.1.1-b

4.1.2 Disassembling of Core Board, Bridge Board and LCD

1 Pull out FFC cable and key cable, then disassemble eight screws and washer assemblies(ST2.9*8) fixed
the LCD sustaining board, shown as Figure 4.1.2-a.

10
Service Manual

ST2.9*8

LCD sustaining
board

Figure 4.1.2-a
2 Pull out the FFC cable, then remove the LCD sustaining board, and disassemble the four screws (M3*5)
fixed the core board to take out the core board, shown as Figure 4.1.2-b.

M3*5

Figure 4.1.2-b
3 Pull out the core board, bridge board , LCD and touch tablet cable, and take out the bridge board and
LCD, shown as Figure 4.1.2-c.
FFC cable(50P) FFC cable(40P)

11
Service Manual

TFT LCD

Bridge board

Touch tablet cable

Figure 4.1.2-c

4.1.3 Disassembling of Printer and Print Board

1 Pull out the lines connected to printer, and disassemble the four screws (ST2.9*6.5) fixed printer, then
take out the two printer pressing boards and printer, shown as Figure 4.1.3-a.

Lines connected
to printer
Printer pressing
board

Figure 4.1.3-a
2 Pull out the lines connected to print board, and disassemble the four screws (ST2.9*8) fixed print board,
then take out the print board, shown as Figure 4.1.3-b.

ST2.9*8

Figure 4.1.3-b

12
Service Manual

4.1.4 Disassembling of Key Board

Pull out the lines connected to shortcut key, disassemble the four screws and washer assemblies( ST2.9*8) and
four chamfered washers to take the key board out, shown as Figure 4.1.4.

Figure 4.1.4

4.1.5 Disassembling of Shortcut Keys and Microswitch

Disassemble the two screws and washer assemblies(ST2.9*8) fixed microswitch board to take the microswitch
and shortcut key out, shown as Figure 4.1.5.

13
Service Manual

Fixed board of
microswitch
ST3*8W7.0

Microswitch

Shortcut key

Figure 4.1.5

4.1.6 Disassembling of Main Control Board

Disassemble the connection lines and the four screws (M3*6) fixed the main control board to take out the main
control board.
Note Button battery locates on the main control board, when the following problems appear, such
as time flutters, the date can't save after changing, please replace the battery in time.

4.1.7 Disassembling of Optical System

Pull out the lines connected to optical system and main control board, disassemble the four screws ( M4*16)

14
Service Manual

fixed the nether enclosure and and optical system to take out the optical system, shown as Figure 4.1.7.

Figure 4.1.7

4.1.7.1. Disassembling of Filter Chamber

Disassemble the four screws (M4*8) fixed the pedestal of the optical system to take it out, shown as Figure
4.1.7.1.

Figure 4.1.7.1

15
Service Manual

4.1.7.1.1. Disassembling of Light Collection Board

Disassemble the four screws (M2*6) fixed the light collection board to take it out,shown as Figure 4.1.7.1.1.

Figure 4.1.7.1.1

4.1.7.1.2. Disassembling of Position Checking Board

Disassemble the two screws (M2*6) fixed the position checking board to take it out, shown as Figure
4.1.7.1.2.

16
Service Manual

Position checking board

4.1.7.1.2

17
Service Manual

4.1.7.1.3. Disassembling of Filter and Stepping Motor

1 Disassemble the four screws (M3*6) fixed the cover board of the filter chamber to take it out, shown as
Figure 4.1.7.1.3-a.

Cover board of the


filter chamber

Figure 4.1.7.1.3-a
2 Screw out the screw (M3*6) fixed filter wheel to take it out, shown as Figure 4.1.7.1.3-b.

M3*6

Figure 4.1.7.1.3-b
3 Filter for each wavelength is shown as Figure 4.1.7.1.3-c, the wavelength for gap is 340nm, and each
wavelength deasil arranges from 340nm, 405nm, 510nm, 546nm, 578nm, 620nm(the screen-printing of filter
signed with the corresponding wavelength, install it in accordance with the above order).

18
Service Manual

340nm

Gap

Figure 4.1.7.1.3-c
4 Firstly remove the copper loop, then take out the filter, shown as Figure 4.1.7.1.3-d.

19
Service Manual

Figure 4.1.7.1.3-d.
5 Disassemble the four screws (M3*6) and four spring washers(φ3) fixed the stepping motor in filter
chamber to take the stepping motor out, shown as Figure 4.1.7.1.3-e.

Figure 4.1.7.1.3-e
Stepping
motor
4.1.7.2. Disassembling of Temperature Control Module Screw(M3*6) and
spring washers(φ3)

Disassemble the screw (M4*10) fixed on the pedestal of the optical system to take out the temperature control
module, shown as Figure 4.1.7.2.

20
Service Manual

Figure 4.1.7.2

4.1.7.2.1. Disassembling of Temperature Sensor

Cut off the ribbon, take out the temperature sensor from the bottom of the colorimetric cell pedestal, shown as
Figure 4.1.7.2.1.

21
Service Manual

Thermal silica gel

Ribbon
Temperature sensor

22
Service Manual

Figure 4.1.7.2.1

4.1.7.2.2. Disassembling of Peltier

Disassemble the two screws (M3*40) and the washer(Φ3) fixed the cooling fin to take out the cooling fin and
Peltier, shown as Figure 4.1.7.2.2.

23
Service Manual

The white matter is


the thermal silica
gel

Peltier

Figure 4.1.7.2.2

24
Service Manual

4.1.7.3. Disassembling of Light Source Chamber

Disassemble the three screws (M4*10) fixed on the pedestal of the optical system to take out the light source
chamber, shown as Figure 4.1.7.3.

Adjusting screw A
Fixed screw
Fixed screw B

Figure 4.1.7.3

4.1.7.3.1. Disassembling of Lamp

1 Disassemble the screw (M3*10) and the washer(Φ3) fixed lamp socket to take it out, shown as Figure
4.1.7.3.1.

25
Service Manual

Figure 4.1.7.3.1
2 Screw out the two screws fixed lamp to take it out.

4.1.7.3.2. Disassembling of Convex Lens

1 Disassemble the two screws(M3*6) fixed the convex lens sheath and the screw(M3*12) on the left side
to take out the convex lens sheath, shown as Figure 4.1.7.3.2-a.

M3*6

M3*12

Figure 4.1.7.3.2-a

26
Service Manual

2 Screw out the round nut in convex sheath to take out the bush and double convex lens, shown as Figure
4.1.7.3.2-b.
Sheath Double convex lens Bush Round nut

Figure 4.1.7.3.2-b

4.1.8 Disassembling of Peristaltic Pump

As shown in Figure 4.1.8-a, disassemble the four screws (M3*6), and screw out the two luer connectors,
separate the peristaltic pump from waste tube 1(silicone tube CA0034, length: 200mm), waste tube 2(silicone
tube CA0034, length: 600mm) and peristaltic pump pipe (external diameter: 5mm, internal diameter: 1.8mm,
length: 100mm), then the peristaltic pump can be taken out. The structure for peristaltic pump is shown as
Figure 4.1.8-b.

Waste tube 2

Right luer connector

Left luer connector

Waste tube 1 BC300 pipette

Figure 4.1.8-a

27
Service Manual

Pump Clamping plate

Luer connector

Screw(M5*20)

Pump pipe
Locknut

Figure 4.1.8-b

4.1.8.1. Disassembling of Pump Pipe

Disassemble the pump pipe as following steps:


1 Loosely screw the screw(M5*20), and take out the clamping plate, then the pump pipe can be seen,
shown as Figure 4.1.8.1.

Figure
4.1.8.1
2 Cut
off the
ribbon, take
out the
pump pipe,
then the
pump pipe
will be
separated from plastic interface.
3 Disassemble the dial with a steel ruler.

4.2 Assembling

Assemble according to the opposite steps of disassembling. And pay attention to the following operations:
1 Installation for pump pipe: after installing pump pipe, when use screw to adjust the clamping plate,
adjusting from tight to loose, to make sure moderate impaction and smooth work in the pump, after absorbing,
the liquid in the pipette should not flow out when the peristaltic pump is stopping.
2 Installation for halogen lamp: insert the lamp entirely. When tightening the screws of amphenol connector,
use proper force to avoid damaging to the Halogen lamp. Finally, wipe it.
3 Position adjustment for the halogen lamp: refer to OPERATION METHOD in User Manual for details.
 Enter "System---Machine set(password:111111)" interface, Click "Rinse" to aspirate the distilled water
into colorimetric cell;

28
Service Manual

 Input number 1-6 in "Filter check" item, click "Turn" to circumrotate the filter to 340-620nm;
 Click "Start" in "Lamp check" item to display real-time AD value, click "Stop" to stop AD updating;
 Insert the light socket into light source chamber, switch to each filter and record the AD value;
 Select the filter whose AD value is maximal, adjust the fixed screw on the light source socket to change
its height; adjust the adjusting screw A and B to change the pitching angle of the light source socket(as Figure
4.1.7.3), then change the AD value. When AD value is stable in about 26000, lock the fixed screw in the
middle of the light source socket;
 After adjusting, check the AD value in each filter, its range should be 14000~29000.
4 Assembling and installation for lens: when assembling lens, put the double convex lens, bush, round nut in
the sheath one by one; place the assembled lens into the light source chamber, one side of round nut inside of
light source chamber, the lens pedestal should exceed the lamp plane for 2mm.
5 Assembling and installation for temperature sensor: weld temperature sensor onto wire, then put on a heat
shrinkable tube, shown as Figure 4.2-a.

Figure 4.2-a
 Coat the temperature sensor with thermal silica gel, through the round hole on the pedestal, insert the
sensor into the deepest of the colorimetric cell pedestal, shown as Figure 4.2-b.

29
Service Manual

Thermal silica gel

Colorimetric
Temperature sensor
cell pedestal

30
Service Manual

Figure 4.2-b
6 Installation for Peltier and cooling fan: stick the backing plate onto the colorimetric cell pedestal with 3M
double-sided foam tape(width is 3mm), reserve 2mm gap with the bottom of the pedestal, shown as Figure
4.2-c.

Figure 4.2-c
 Spread and smooth the thermal silica gel evenly over the side of the colorimetric cell pedestal, make the P
eltier flat on the silica gel, then press it tightly, shown as Figure 4.2-d. Note: the silica gel should be spread in
proper quantity, otherwise it will spill over the adhesive surface; the direction and position of the Peltier should
be placed in the middle and abut on the pedestal.

Figure 4.2-d
 Spread and smooth the silica gel evenly over the Peltier, shown as Figure 4.2-e. Note: the silica gel
should be spread in proper quantity, otherwise it will spill over the adhesive surface.

31
Service Manual

Figure 4.2-e
 Stick the cooling fin on the Peltier, adjust its position to align with the hole on the pedestal, then press it
tightly, fix it with two screw(M3*40) and washer(Φ3), shown as Figure 4.2-f. Note: the top of the screw
should through the cooling fin.
§

Screw(M3*40) Washer

Figure 4.2-f
7 Installation for touch tablet
 Stick Poron single-sided adhesive on the around of the window inside the upper shell of the device, the
paste position of foam should next to stopper groove, shown as Figure 4.2-g.

32
Service Manual

Poron single-sided
adhesive

Figure 4.2-g
 Tear the inner protective film of touch tablet and protective paper of double-sided foam tape, make the
touch tablet cable through the shell, align to display window of the screen, stick the touch tablet on the shell,
shown as Figure 4.2-h.

Touch tablet

Figure 4.2-h
8 Installation for core board, bridge board, LCD
 Install the core board on the LCD sustaining board with four screws(M3*5), shown as Figure 4.2-i.

Screw(M3*5)

Figure 4.2-i
 As shown in Figure 4.2.-j, tear the protective film of the TFT LCD, then reversedly place it inside of the
upper shell, stick the back of LCD bridge board on TFT screen with 3M double-sided foam, connect touch
tablet cable to LCD bridge board. Note: the interface of resistive touch tablet and capacitive touch tablet are
different, the following is the connection of resistive touch tablet.

33
Service Manual

TFT LCD

Capacitive touch tablet interface


Bridge board
Resistive touch tablet interface

Touch tablet cable

Figure 4.2.-j
 Connect the TFT LCD to LCD bridge board with a FFC cable, and connect LCD bridge board to core
board with another FFC cable, shown as Figure 4.2-k. Note: the connection direction of FFC cables should
according to the figure.
FFC cable(50P) FFC cable(40P)

Figure 4.2-k

4.3 Maintenance

 Operate the device following USER MANUAL strictly. When a test item switches to the other one,
rinsing the colorimetric cell is necessary. Rinse all residual reagent to avoid inaccurate result.
 The device has good repeatability for using fixed colorimetric cell, but because the reagent and sample
needed to be added are little, pay special attention to the accuracy of the sample and reagent.
 The reason for the device doesn't aspirate liquid is that it hasn't been used for a long time, which causes
the pipette staved, jammed or pump pipe conglutination, then it needed to be rinsed or replaced
 During working process, it is strictly prohibited to aspirate in strong acid reagent to avoid damaging the
metal head of the pipette. After each using, use the distilled water to rinse the colorimetric cell and
pipette. If the device can't calibrate zero after aspirating in the distilled water for many times, use the
detergent to soak the colorimetric cell. Place it in 37℃ environment, use 10% detergent to fill with the
colorimetric cell for soaking about 15 minutes, then use water to rinse.

4.4 Updating

Refer to the operation methods in USER MANUAL, insert the flash disk contained updating package, click

34
Service Manual

USB updating button, password is 888888, then the system will update automatically, the message "update
successful" appears after finishing updating, then the device will re-boot automatically, finally remove the
flash disk.

35
Service Manual

5 Troubleshooting

No. Trouble Analysis Solution


1.No power.
1.Power on the device.
The analyzer can't turn 2.The plug isn't inserted into the power
1 2.Properly insert the power cord.
on. socket properly.
3.Replace the fuse.
3.The fuse is broken.
Unstable time and
2 date, and can't be No power. Replace the battery.
saved after modifying.
1. There is air bubble or pollution in
Distilled water blank is 1.Rinse the colorimetric cell.
the colorimetric cell.
3 unstable, high or 2.Replace the pipette.
2. The pipette is broken.
negative value. 3.Replace the photocell.
3. Photocell is bad.
1.Blow with the rubber pipette bulb or remove t
1.Dust appears in the light route.
he light route for wiping.
2.The location of lamp is improper.
4 Low light intensity 2.Re-adjust the location.
3.Lamp aging.
3.Replace the lamp.
4.Filter aging.
4.Replace the filter.
1.Aspirate the air.
1.Re-aspirate the distilled water.
2.Lamp aging.
Big absorbency 2.Replace the lamp.
5 3.The colorimetric cell is broken.
deviation. 3.Replace the colorimetric cell.
4.The colorimetric cell isn't inserted to
4.Insert the colorimetric cell into the bottom.
the bottom.
1.False parameters settings.
1.Re-input.
2. There is air bubble or pollution in
2.After rinse the colorimetric cell, test again.
Inexact result or big the colorimetric cell.
6 3.Replace the lamp.
deviation. 3. Lamp aging.
4.Replace the filter.
4. Filter aging.
5.Replace the photocell.
5. Photocell is bad.
1.The standard value or the factor isn't 1.Adjust reagent parameter according to reagent
The result is zero or inputted. manual.
7
incorrect. 2.Without saving after testing the 2.Re-make the standard, several times are
standard. recommended, then click"Save".
1.The pump pipe isn't installed
properly.
1. Check and properly install the tube.
2.The pipeline or colorimetric cell is
2. Smooth the pipeline with a syringe, or replace
Non aspiration sample blocked.
the blocked pipeline or colorimetric cell.
8 or Aspirate volume is 3.The device doesn't work for a long
3. Remove the pump tube and smooth it, or repla
inconstant . time and doesn't release the pipeline, so
ce it.
the pump pipe may be partial
4. Re-calibrate the peristaltic pump.
conglutination or aging.
4. Inaccurate steps for aspirating liquid.
9 Quality control is out 1.Reagent invalidation. 1. Check the reagent validity.

36
Service Manual

2.False settings. 2. Check parameter settings.


of the target value 3. Check the flowing colorimetric cell. 3. Check whether there is pollution or air
bubble.
1.The connection plug of temperature 1.Plug and unplug or replace the connection
Failure for sensor is looseness. line.
10
temperature-control 2.Temperature sensor is broken. 2.Replace the temperature sensor.
3.Peltier is bad. 3.Replace the Peltier.
1.The settings for print mode is false. 1.Reset the print mode.
11 The printer can't print. 2.The print wire falls off. 2.Plug and unplug the print wire.
3.The print wire is broken. 3.Replace the print wire.
1.False voltage. 1.Test the output voltage and replace the power
2.Software is broken. supply.
Blank screen, blurred
12 3.Cable is broken. 2.Refresh the software, refer to 4.3.
screen, crash fault
4.Insufficient solder of LCD bridge 3.Replace the cable.
board. 4.Replace the LCD bridge board.

37
Service Manual

AppendixⅠ Parameter Specification

Dimensions: 410(L)×340(W)×150(H)mm
Weight: about 6kg
Operation environment:
Temperature: 10℃-30℃
Relative Humidity: ≤80%
Atmospheric pressure: 86kPa~106kPa
Transportation and storage:
Temperature: -40℃~55℃
Relative Humidity: ≤95%
Atmospheric pressure: 86kPa~106kPa
Test item:
Liver function
Albumin ALB
Total protein TP
globulin GLB
Total bilirubin T BILI
Direct bilirubin D BILI
Indirect bilirubin I BILI
Glutamic-oxalacetic transaminase AST
Glutamic-pyruvic transaminase ALT
gamma-glutamyltransferase GGT
Alkaline phosphatase ALP
Lactic dehydrogenase LDH
Total bile acid TBA
Prealbumin PALB
Glucose(OXI) GLU

Renal function
Urea UREA
Creatinine CR
Uric acid UA
Total protein TP
Albumin ALB
Glucose GLU

Myocardial enzyme
Creatine kinase CK
Creatine kinase isoenzyme CK-MB
Glutamic-oxalacetic transaminase AST
Lactic dehydrogenase LDH
Blood lipid
Cholesterol CHOL

38
Service Manual

Triglyceride TRIG
Lipoprotein cholesterol HDL-C
Apolipoprotein A ApoA
Apolipoprotein B ApoB
Apolipoprotein E ApoE

Blood iron
Iron IRON
Total iron binding capacity TIBC

Ions and others


Potassium K
Sodium Na
Chlorine Cl
Calcium Ca
Magnesium Mg
Phosphorus(inorganic) P
Copper Cu
Zinc Zn
Glucose GLU
Amylase AMS
Lipase LPS

39
Service Manual

AppendixⅡ Internal Wiring Diagram

40
Service Manual

AppendixⅢ Product Failure Feedback Table

Product Failure Feedback Table

Client Name Linkman Telephone

Product Name Product Model Purchase Date


Product Number

No. Failure Description (attached figure) Frequency Date


1

Note This table is used for failure information feedback to help our company deal with new
failures in time and give good solutions. Please fill in carefully.

41
Service Manual

Contec Medical
Systems Co., Ltd.
Address:No.112 Qinhuang West Street, Economic & Technical Development Zone, Qinhuangdao,
Hebei Province, PEOPLE’S REPUBLIC OF CHINA
Tel: +86-335-8508888
Fax: +86-335-8015422
E-mail: cms@contecmed.com.cn
Website: http://www.contecmed.com

File No.:CMS2.782.140EWS/1.0
File Ver.:1.0
Release Date: July 2015

42

You might also like