ALIFAX - Roller 20 PN

Preventative Maintenance Check List


Customer:__________________________ Location: ____________________Date:_________________
Model:_____________________________SN:_________________________Report:________________
PM Activities Completed Remarks
1. Check for any damage of the needle and replace it if damaged.
2. Check and replace the Pump tubes at every 30,000 Tests.
3. Check the status of PTFE tubing from syringe to the reading block and
from the pump to waste tank, replace the tubing if deformed or
damaged
4. Withdrawal Piston: Remove grease residuals in the inner part of piston on
the needle support and Lubricate the needle support, clean the steel flag on the
side facing the sensor.
5. Check the status and the tension of belt of Syringe( Z Axis ), in particular the
point of lockage to the syringe carriage, replace it if damaged

6. Clean and lubricate as explained for carriage (X axis).


7. Check if blood flow is normal and regular.
8.Check the rotation of pump, pay attention to rolls are regularly rolling
9.Verify the regularity of pressure of rolls due to springs. If some
squeaking is heard while the rolls operate, lubricate the rolls spraying
them with a bit of silicon oil.

10. Check the status of PTFE tubing from syringe to the reading block and
from the pump to waste tank, replace the tubing if deformed or
damaged .
11. Check the correct open/close of front door (open and close the doors),
Technical door check if the movement on the hinges is not altering the correct
detection of closed door by the sensor.

12. Check if the Home sensor is distant less than 2 mm from the magnet
(the optimal distance is 1 mm) and check if it is detected. Adjust, if necessary
the distance moving the sensor toward the rotor

13. Check the status of belt tension of motor eventually adjust the
tension of belt . Remember that an excessive tension of belt enhances
the motor noise level during rotation

14. Check if the magnets stuck on the head of pump are present, check
the sensor, and look if the sensor, on the commands panel, is lighted
when the magnets face the sensor during rotation.
during rotation.
15. Check the calibration of the CPS Module.
16. When all the described checks are performed at the level of 30.000
test, reset the Maintenance Counter
NOTE: If an inspection step necessitates part(s) to be replaced record activity within Incident Report.
Yes – indicates mandatory Preventive Maintenance Activity has been completed. NR – indicates that elective Preventive
Maintenance Activity was not required.
Inspect – visually inspected, no action required. Clean – based on inspection clean procedure was performed. Replace – based
on inspection, item replaced

Engineer Name: __________________________________Signature: _________________________