Ultima X5000 Gas Monitor Data Sheet-En

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

ULTIMA® X5000 Gas Monitor

The future looks bright.

Simple retrofits have identical Intuitive display features


footprint and wiring to ULTIMA X new design equipped with
Gas Monitor series. organic LED (OLED) display,
with full word text in 9
languages. Bright green,
Bluetooth® wireless yellow and red status LEDs
technology allows mobile for extreme visibility.
device to act as HMI
screen and controller.
Industry-first,
touch-button interface
provides intuitive,
tool-free user experience.

Reduce setup time by at


least 50% with the X/S
Instrument status indicators
Connect App
illuminate power, fault and
alarm conditions.

Advanced Sensor Technology

• Patented XCell Sensors with TruCal technology* extend calibration cycles


for as long as 18 months, actively monitor sensor integrity and compensate
for environmental factors and electrochemical sensor drift.
• Diffusion Supervision sends acoustic signal every 6 hours to check that
sensor inlet isn't obstructed so gas can reach the sensor.
• Worry-free operation; automatically self-checks four times per day.
• Three-year warranty and five-year expected life for XCell Sensors.
• Dual sensor capability doubles sensing power
with half the footprint of a single gas sensor transmitter.
• SafeSwap enables safe and quick XCell Sensor replacement without
powering off gas detector. *Only CO & H2S
XCell Sensors

Applications
• Chemical • Utilities
• Oil and gas • Wastewater
• Petrochemical • General industry
ULTIMA® X5000 Gas Monitor
Specifications

PRODUCT SPECIFICATIONS ENVIRONMENTAL SPECIFICATIONS *


COMBUSTIBLE GAS Catalytic Bead (XCell Comb) OPERATING XCell: -40°C to +60°C
SENSOR TYPE Infrared (XIR PLUS) TEMPERATURE RANGE XIR PLUS: -40°C to +60°C
TOXIC GAS & OXYGEN Carbon Dioxide CO2 (XIR Plus) (may differ by gas, see manual)
SENSOR TYPE Carbon Monoxide CO (XCell Toxic) STORAGE -40°C to +60°C
Carbon Monoxide CO H2-Resistant (XCell Toxic) TEMPERATURE RANGE
Hydrogen Sulfide H2S (XCell Toxic)
Oxygen O2 (XCell O2) RELATIVE HUMIDITY 10-95% (XCell Toxics & O2)
Sulfur Dioxide SO2 (XCell SO2) (NON-CONDENSING) 0-95% (XCell Comb)
SENSOR MEASURING Combustible: 0-100% LEL 15-95% (XIR PLUS)
RANGES CO2: 0-2%, 0-5% Vol
MECHANICAL SPECIFICATIONS
CO: 0-100, 0-500, 0-1000 ppm
INPUT POWER 11 to 30 VDC, 3 wire, <5W nominal
CO-H2Resist: 0-100 ppm
H2S: 0-10, 0-50, 0-100 ppm SIGNAL OUTPUT Dual 4-20 mA current source, HART
O2: 0-25% BLUETOOTH Bluetooth Low Energy (BLE) v4.3 or higher
SO2: 0-25 ppm (OPTIONAL)
TYPICAL SENSOR 5 years – XCell Sensors RELAY RATINGS 5A @ 30VDC; 5A @220 VAC
LIFE 10 years - infrared (3X) SPDT – fault, warn, alarm
SENSOR T90 RELAY MODES Common, discrete, horn
ZERO DRIFT
PERFORMANCE* (TYPICAL) REPEATABILITY NOMINAL MAX XIR PLUS 6.7W
(PER YEAR)
POWER XCell Comb 4.9W
XIR PLUS COMB < 2 sec. < +/- 1% LEL n/a
(W/ RELAYS) XCell Toxic & O2 2.8W
XIR PLUS CO2 < 6 sec. < +/- 1% Vol n/a XIR PLUS and XCell Comb 10.9W
XIR PLUS and XCell Toxic & O2 7.0W
XCELL COMB < 22 sec. < +/- 3% LEL ≤ 5% LEL
Dual XIR PLUS 11.6W
XCELL CO < 9 sec. < +/- 1% ≤ 1% full scale Dual XCell Toxic & O2 3.6W
XCELL H2S < 23 sec. < +/- 1% ≤ 1% full scale Dual XCell Comb 10.6W
Dual XCell Comb and XCell Toxic & O2 5.3W
XCELL O2 < 11 sec. < +/- 0.1% Vol ≤ 0.2% Vol
EMC DIRECTIVE Complies with EN 50270, EN 61000-6-4,
XCELL SO2 < 6 sec. +/- 1% ≤ 1%FS per mo. EN 61000-6-3
APPROVALS Divisions (US/Canada) DISPLAY Organic LED (multi-lingual) with Contrast Ratio of
CLASSIFICATION Class I, II, III; Div 1&2, T4/T5/T6; 2000:1 and View Angle of 160°
Zones (Global) HART HART 7, HART device description language
Ex db nA IIC T5 Gb (Class I, Zone 1/ Zone 2) available
Ex tb IIIC T85°C Db (Class II, Zone21) FAULTS MONITORED Low supply voltage, RAM checksum error, flash
Enclosure rating checksum error, EEPROM error, internal circuit
Type 4X, IP66 error, relay, invalid sensor configuration, sensor
faults, calibration faults, general system
Markings vary by component. See Manual for
WIRING 3-wire (single sensor) or 4-wire (dual sensor)
specific component markings.
REQUIREMENTS shielded cable.
WARRANTY X5000 Transmitter: 2 years Refer to manual for mounting distances and wire
XIR PLUS: 10 years source, 5 years electronics gauge.
XCell Sensors: 3 years
DIMENSIONS
APPROVALS CSA, ATEX, IECEx, Inmetro, DNV Marine, CE
Marking. HOUSING (W X H) 5.88’’x 5.71’’ [150 mm x 145mm]
Complies with C22.2 No. 152, FM 6320 w/ XCell Sensor 5.88’’ x 10.15’’ [150 mm x 258 mm]
RED, FCC, Suitable for SIL 2 w/ XCell Sensor & 13.42’’ x 10.15’’ [341 mm x 258 mm]
XIR Sensor
* Typical response at standard temperature and pressure test conditions LID (DEPTH)
w/ Relay Board 4.86’’ [123mm]
w/o Relay board 3.86’’ [98 mm]

WEIGHT 8.8 lbs (4 kg) - 316 SS

Note: This Bulletin contains only a general description of the products shown. While product uses and performance capabilities MSA operates in over 40 countries
are generally described, the products shall not, under any circumstances, be used by untrained or unqualified individuals. The
products shall not be used until the product instructions/user manual, which contains detailed information concerning the proper worldwide. To find an MSA office near you,
use and care of the products, including any warnings or cautions, have been thoroughly read and understood. Specifications are
subject to change without prior notice.
please visit MSAsafety.com/offices.

ID 0720-185-MC / 10.2019
© MSA 2019 MSAsafety.com/detection

You might also like