Fume Hood (United Version)

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 34

FUME HOOD Product operation manual

A-PR / VA / BY type

※ Product operation manual must be fully informed before the customer use.
※ Product Operation Manual must be fully informed before the customer use.
This operating manual of Fume hood is a precaution for safety to keep the appropriate usage and the user’s safety as well as the laboratory safety, and to prevent any loss of property etc.
FUME HOOD Product operation manual
A-PR / VA / BY type

Thank you for purchasing Chemical Fume hood.


This product provides a safe and pleasant working environment by connecting the blower, which exhausts the polluted air and poisonous
gas inside laboratory through the duct system.

03
Contents
1. Preparation section
1.1 Warning and caution for safety 06
1.2 Installation environment 12
1.3 Transportation 13
1.4 Product installation 14

2. Use section
2.1 Buttons and features of the controller 16
2.2 Use of Controller 19
2.3 Use of valve 22

3. Other sections
3.1 Cleaning and repair 24 3.5 Product warranty contents and scope 30
3.2 Checklist before claim report 25 3.6 Product warranty
3.3 Product dimension 27
3.4 Power distribution box power distribution 29
diagram
04
FUME HOOD
Product operation manual

Preparation section
1.1 Warning and caution for safety 06
1.2 Installation environment 12
1.3 Transportation 13
1.4 Product installation 14

CHEMICAL FUME HOOD


Preparation section Warning and caution for safety

Warning
If not observed, user may be injured.

Caution
If not observed, property loss may be caused.

This is a sign of compulsory order, which shows the items to be surely observed.

This is a sign of prohibition of an action, which stresses the items not to be done.

This is a sign of disassembly prohibition.

06
Preparation section Warning and caution for safety

◎ Warning and Caution for Safety – installation


Installation should be asked to an expert engineer. It should be installed at a place not influenced by wind.
▶ It may be a cause of electric shock, fire, explosion, and Injury. ▶ Product performance may be influenced if air-conditioner,
door etc. are affected by air stream.

For power supply, a socket of AC 220 V / 60 Hz / rated over 15 A


should be solely used for product. And, do not arbitrarily extend Do not give excessive strength or impact to the backside of product.
the length of power cord.(International receptacle optional)
▶ There is a danger of electric shock, fire, failure, and explosion .
▶ In case multiple-tab is used in parallel with other equipment, or
extendedpower cord is used, electric shock, fire may be caused. ▶ It may cause a damage to a floor.

▶ In case a transformer is used, electric shock, fire may be caused.


▶ If any conditions of voltage/frequency/rated current are different, When a product is installed, it should be installed at a place where
electric shock, fire may be caused. the power cord is easily unplugged.
▶ Electric shock or fire may be caused by short circuit.

Do not install near the heating equipment and the object sensible
to heat, moistly/oily/dusty place, where direct rays and water (rain It should be installed where the bottom is strong and horizontal.
water etc.) is reached, or where there is a possibility of gas leak.
▶ Unusual vibration, noise, and failure may be caused.
▶ There is a danger of electric shock, fire, failure, and explosion.
▶ A socket with earth must be used.

07
Preparation section Warning and caution for safety

◎ Warning and caution for safety – related to power supply


Periodically wipe the foreign materials (dusts, water etc.) on the When the power plug is unplugged, do not unplug the power
terminal and connection area of the power plug well with a dry cloth. cord (cable) by pulling.
▶ Electric shock or fire may be caused. ▶ Hold the power plug and unplug it.
▶ Do not wipe with a wet cloth. ▶ Electric shock or fire may be caused.

Be sure to connect the power plug to the end of socket, and do Connect the power plug with its cord directed downward.
not use any damaged power plug, power cord, and loose socket.
▶ If connected in the opposite direction, lines may be cut at inside
▶ Electric shock or fire may be caused. of the cord, which may cause electric shock or fire.

Do not excessively bend, pull, twist or tie the power cord. And In case power cord or power plug is damaged, let them be
do not hang it up on the metal part, nor put heavy object on it, exchanged by the manufacturer or service agent or qualified
nor insert in between or into the backside of product. engineer in order to prevent any danger.
▶ Electric shock or fire may be caused.

08
Preparation section Warning and caution for safety

◎ Warning and caution for safety – related with Use


Never disassemble, repair, and modify it arbitrarily. Do not apply excessive force for the door glass.
▶ In case repair and reinstallation is required, please contact ▶ Electric shock or fire may be caused.
service center.
▶ It may be a cause of electric shock, fire, explosion, and injury.

When the interior, exterior of product and door glasses are


If any strange sound or burning smell, smoke is occurred from cleaned, use of neutral detergent is recommended.
the product, immediately shut off the power and contact service
center.
▶ Electric shock or fire may be caused.
When the interior lighting is replaced, be sure to pull out the plug.
Do not step on top of product, nor put any object (drug, ▶ Electric shock or fire may be caused.
equipment, materials etc.) on it.
▶ If the glass is damaged, there is a danger of injury.

Do not use the explosive inflammable materials.


▶ It may be a cause of electric shock, fire, explosion, and Injury.

09
Preparation section Warning and caution for safety

◎ Warnings and cautions for safety – related with use

150mm
or more

Wear the personal protective equipment In case equipment is stored, backside and floor
Do not use it with a head inside the hood.
before use. side should be spaced for a smooth air flow.

PERCHLORIC PERCHLORIC
ACID ACID

Do not directly throw away the waste


Do not store reagent and other articles. Do not use perchloric acid (strong acid). solution into the cup sink.

10
Preparation section Warning and caution for safety

◎ Additional cautions

[Other cautions]
▶ Use the fume hood after checking if it is normally operated or not ▶ Do not use the chemicals of which use is prohibited in the fume hood.
before test.
▶ Product performance test should be taken more than one time per
▶ Push and turn the handle of the locking device for the gas valve. one year.
(Test should be implemented by an expert in a proven method.)
▶ Do not evaporate any chemicals in the hood with disposal of
chemicals. ▶ All test materials and equipment should be put inside for use within
150mm from the door.
▶ In case test solution is spilt or poured, it should be cleaned
immediately. ▶ In case any accident is occurred, close the fume hood door in
use and contact an expert and a person in charge of equipment.
▶ In case any horizontal window is installed, use the horizontal
window with closed.(Do not open the horizontal window with the
door lifted up.)

▶ Do not use the fume hood as a storage

▶ In case any chemicals are stored in the space of power distribution box,
there is a danger of explosion and fire.

※ Use this product for laboratory only.


- Please get the service through the qualified expert.

11
Preparation section Installation environment

[Installation environment]
Installation environment conditions Safety measure
▶ Fume hood(A-PR/ A-VA/ A-BY) is designed in accordance with the ▶ Required conditions : It should be handled by an authorized personnel.
following conditions.

Should be used in clean room. Protection against water ① The product should be treated in
infiltration is not available.
accordance with the laws and
regulations for the human health and
safety of each countries.

Should be used at a Within the maximum ② When the product is handled or


temperature between humidity of 80% accessories are disassembled, it
5℃ - 40℃ should be performed by a trained
person only.

③ The product should be handled by


use of a forklift equipped with a
proper load capacity.

※ Please be careful when the


forklift is moved as the product
height is high.

12
Preparation section Transport method

[Transport method] [Grounding]


▶ When the product is moved, do not pull it forcibly! Please be sure to earth to prevent any accident of electric shock
by leakage current.
- It may be a cause of Failure or Injury.
▶ In case there is an earth terminal with a socket.
▶ In case of installation, repair, and cleaning, it should be moved by use of a cart. - Separate earth is not required.
- In case product is dragged or pulled without cart, the fixed foot may be broken.
▶ In case there is no earth terminal with a socket.
▶ When the product is moved, do not lay down and move it. - Please purchase an earth wire separately and connect to the
backside of product.
- It may be a cause of failure or injury.
- Please connect the earth wire with the metal water pipe or
earth lines.
▶ It should be installed on the flat and strong floor.
- If the floor is uneven when the product is installed, it may cause a vibration grounding
and noise, and the door may not be leveled. The floor should be a hard place
to withstand the weight of product.

▶ Please avoid the place where the sun is beating down, where there is much
heat, where it is severely cold.
- It the temperature of an installation place is too high or low, abnormality
of function may be caused.

13
Preparation section Product installation

[Product installation]
① Please install the product after being fully informed of the
cautions before the product installation.

② Remove the packing of product.

③ Please make sure if the components are all right.

④ Combine with the lower part according to the product composition.

⑤ Put the product on the installation place and connect the power
cord of the upper part.

⑥ Push the power switch and operate.

[First user checklist]


Please check if the product is horizontally installed.

Please push the power button after power is connected, and check if there is any abnormality with the blower operation.

14
FUME HOOD
Product operation manual

Use Section
2.1 Buttons and features of the controller 16
2.2 Use of Controller 19
2.3 Use of valve 22

CHEMICAL FUME HOOD


Use section Part Names and Functions

[Buttons and feature of the controller]


The static touch controller can only operate with a light touch.

Blower
Power <Setting>

Airflow Enter
check

Up Menu
<Move>

Down
<Move> Alarm off

Lamp Display

16
Use section Part Names and Functions

① [Power button] ② [Air flow monitor]


You can turn the power ON / OFF. - Normal operation: Yellow LED
- Un normal operation: In case of abnormal status, Red LED turns on
/off and Alarm sound activation.
(When equipped with wind speed sensor)

<Normal operation> <Un normal operation>

③ [Up, Down button] ④ [LAMP button]


You can move the menu. You can turn the Lamp ON / OFF.

<Up> <Down> <OFF> <ON>

17
Use section Part Names and Functions

⑤ [Blower] ⑥ [Enter]
You can turn the ON / OFF. - You can choice the menu.
(When Using only the Blower)

<OFF> <ON>

⑧ [Alarm]
It can be On / Off when an alarm occurs.
⑦ [Menu]
You can launch the menu.

<OFF> <ON>

Inverter vol Mute


<비활성화> <활성화>
40%
<Mute message>
18
Use section Use the Controller

[Type of the Fume Hood Control]


※ Basic type screen
(No air flow sensor)

<Inverter Type> <Blower Type> <Damper Type>

Inverter type Frequency Blower type Frequency Damper type Frequency

Inverter vol 60Hz Blower speed 60Hz Damper post 60Hz


40% 83% 75%

Control value Inverter Control value Blower Control value Damper


graph Control value graph Control value graph Control value

Inverter vol 60Hz Blower speed 60Hz Damper post 60Hz


INV FAN DMP

Inverter mode Blower mode Damper mode

19
Use section Use the Controller

[Type of the Fume Hood Control]


※ It is an option type screen.
(Mounting a Air Flow Sensor)

<Inverter Type> <Blower Type> <Damper Type>


Show current
air flow speed Frequency Blower type Frequency Damper type Frequency
value

Vel:0.40ms 60Hz Vel:0.40ms 60Hz Vel:0.40ms 60Hz


A:0.20-0.50m 40% A:0.20-0.50m 40% A:0.20-0.50m 40%

Air flow speed Inverter Air flow speed Blower Air flow speed Damper
range setting Control value range setting Control value range setting Control value

Vel:0.40ms 60Hz Vel:0.40ms 60Hz Vel:0.40ms 60Hz


A:0.20-0.50m INV A:0.20-0.50m Fan A:0.20-0.50m DMP

Inverter mode Blower mode Damper mode

20
Use section

[ Error message]
※ Mounting a Air Flow Sensor

Screen Contents

Vel:Sensor Err 60Hz A warning sound and a message will occur when the
A:0.20-0.50m 40% air flow sensor is out of order.

Vel:AIR Fail 60Hz If the air flow value is overflow,


A:0.20-0.50m 40% "Air Fail" message will be generated.
Use section Use the Valve

[Use the Valve]

▶ Use the Gas valve(Yellow) ▶ Use the Water valve(Green)


① Push the valve. ① Turn the valve left and right.
※ Excessive force can cause malfunction.

② Turn left to open and Turn it to the right to close it.


※ Be sure to close the valve after use.

At the time of each valve opening, the internal P. P. cock


※ When closing the valve, do not push forward, but turn it right.
(nozzle) with valve attached Gas / Water / Air / Vacuum is released.
※ If you open the valve without pushing it, the valve will be damaged.

22
FUME HOOD
Product operation manual

Other section
3.1 Cleaning and repair 24
3.2 Check list before claim report 25
3.3 Product dimension 27
3.4 Power distribution box power distribution 29
diagram
3.5 Product warranty contents and scope 30
3.6 Product warranty 32

CHEMICAL FUME HOOD


Other section Cleaning and Keep

[Cleaning and repair] [How to keep]

1 2

Power should be disconnected from the Clean inside of work table with a sponge In case the fume hood should not be used inevitably for more than one month,
socket before cleaning. and soft cloth wet with neutral detergent. or should be kept in warehouse, the front door should be closed completely,
and should be sealed completely by use of vinyl.

3 4

Clean the outside surface, and remove Cleaning is completed. Please shut off the power of the product. Do not move, nor touch it while being
dusts etc. partially from the upper part and kept in a warehouse.
work table.

When the inside and outside of this product is cleaned, do not use brush, sponge, rough cloth etc. of which surface is rough.
Do not spray the product directly with water, nor clean it with Benzene, Thinner etc.
- Discoloration, Deformation, Damage, Electric Shock, Fire may be caused.

24
Other section Checklist before claim report

1. Power of the blower is not switched on.


Checklist Actions

1-1. Isn't the voltage different? 1-1. This product is used only for 220V.
1-2. Isn't the power plug removed? 1-2. After checking the power plug, insert it again.
1-3. Isn't the circuit breaker away? 1-3. After checking the distribution box, operate it again.

2. The Lamp is not operated.


Checklist Actions

2-1. Isn't the power plug removed? 2-1. After checking the power plug, insert it again.
2-2. Isn't the circuit breaker away? 2-2. After checking the distribution box, operate it again.
2-3. Isn't the life time of lamp expired? 2-3. Replace the lamp.

3. The Blower is not operated, and the Lamp is operated.


Checklist Actions

3-1. Isn't there any abnormality for the blower wiring? 3-1. Please check the condition of blower wiring.
3-2. Isn't the blower wiring out of order? 3-2. Please replace the blower wiring.

25
Other section Checklist before claim report

4. The Lamp is flickering.

4-1. Isn't the life time of Lamp expired? 4-1. Please replace the lamp.
4-2. Isn't there any abnormality for the wiring? 4-2. Please check the condition of wiring.

Please refer to the above situations if any problem is occurred for the product.
If any problem is not solved with the solutions described as above, please contact CHCLAB Co., Ltd.

26
Other section Product demension

[TECHNICAL DATA]
A-PR / VA / BY A-PR / VA / BY
Dimension Dimension
1200 1500 1800 2100 2400 1200 1500 1800 2100 2400

Width 1200mm 1500mm 1800mm 2100mm 2400mm


110V X 1EA 110V X 2EA
Receptacle
220V X 1EA 220V X 2EA
Depth 950mm(A-PR) / 815mm (Other)

Height 2360mm Blower Option

Steel plate 1.6t


Air foil Lamp 1EA 2EA
with epoxy powder coated

Cup sink 1EA 2EA LCD Display A-PR / A-VA

Outside Material Phenolic laminate 6t Lamp Switch A-BY

Inner Material Phenolic laminate 4t Blower Switch A-BY

Worktop Epoxy 25t (A-PR,A-VA)/ Ceramite 30t (A-BY) Electrical Box 220V Single phase

Gas valve 1EA 2EA Duct (Φ) 200 / 250 / 300

Water valve 1EA 2EA Weight (Kg) - - - - -

Vacuum,Air
Option
Valve ※ For more information, please visit www.chclab.com.

27
Other section Product demension

[A-PR] [A-VA] [A-BY] [Airflow]

Chemical fume hood operation manual CHC LAB Customer Center

※ Product appearance, specification etc. may be changed without ※ Using Inconvenience and failure receive
prior notice for product improvement. TEL) +82-42-933-0036
FAX) +82-42-933-0039
※ Pictures, photos in the manual may not be matched up with
※ You can get fast service if you inform us correct Model No.,
the product.
Contact No. before you call us.
※ Please be careful not to lose the warranty in the operation manual. ※ Inter homepage
www.chclab.com

28
Other section Electric control box, electric diagram

[220V Single phase electric control box, electric diagram] Basic type

① Earth leakage breaker (110V)

② Earth leakage breaker (220V)

③ Terminal block

220V

110V

110V 220V

BLACK (S)
WHITE (W)
RED (R)
BLUE (B)
GREEN (G)
29
Other section Electric control box, electric diagram

[220V 3-Phase 3-Line type electric control box, electric diagram]

① Circuit breaker

② AC electromagnetic switch

③ Overload relay

④ Earth leakage breaker (220V)

⑤ Earth leakage breaker (110V)

⑥ Terminal block

BLACK (S)
WHITE (W)
RED (R)
BLUE (B)
GREEN (G)
30
Other section 제품 보증 내용 및 보증 범위

▣ The warranty period of A-PR/ VA/ BY is 1 year after the first installation, and 1 year from the dispatch
date by CHC LAB Co., Ltd. Consumables and accessories are not included, and details are as follows.

Warranties Exceptional items of warranty

① Malfunction or abnormality of electric device such as ① Entire consumables and accessories


power supply, lamp etc.
ㆍ Consumables
② Abnormality of front door - Lamp (fluorescent lamp) and other consumables
③ Crack and damage by the defect of product. ㆍ Accessories attached at time of order manufacturing
④ In case use of product is difficult due to other product defects. - Consent, utility valve and other accessories attached at time of
order manufacturing

② In case damaged due to falling or user's carelessness.

③ Physical impact or accidental damage.


(Body and tempered glass of door etc.)

④ In case used after arbitrarily disassembled and modified(changed) by a user.

⑤ In case a detergent contained an ingredient other than neutral


detergent is used.

If any problem is occurred, please contact the dealer (Purchase place) or CHC LAB.

31
Other section Product warranty

PRODUCT WARRANTY
1. CHC LAB Co., Ltd. enforces warranty for product based on the notification of the fair
trade commission as below

2. The warranty for quality of this product is received as per the contents written in the warranty.

3. Please call to the purchase place or +82-42-933-0036 if any failure occurred for the product.

Product name Chemical fume hood


Model No. A-PR / A-VA / A-BY
Serial No. Serial No.
Date of purchase 20 . .
Purchasing office

Warranty 1 year after installation

CHC LAB Co.,Ltd.


www.chclab.com

※ In case product purchase date is not confirmed, the warranty period is dated from the
manufacture date or the date that has passed 3 months from the import customs
clearance date.

32
FUME HOOD
A-PR/ A-VA/ A-BY Product operation manual
Ver.01612

※ The Intellectual property right included in this operation manual (Contents, Image etc.) is owned by CHC LAB Co., Ltd., and any
unauthozized use without the consent of our company is prohibited.

※ The colors of the products included in this operation manual may be different with those of acual products, and the design
and performance may be changed without prior notice for the improvement of performance.

CHC LAB Co.,Ltd. Web Site face book E-mail


139, Techno 2-ro, Yuseong-gu, Daejeon, Korea / Tel.+82-42-933-0036 / Fax.+82-42-933-0039 chclab.com facebook.com/chclab marketing@chclab.com

You might also like