Calibration Certificate
Date of Issue: 20/12/2021 Certificate No.RT-1742
Customer Details M/S RUDRA NDT SERVICES(OPC)PRIVATE.LIMITED.
€201,B WingSun Residency, Anand Mahal Road, Adajan,
Surat Gujrat-395009
Instrument Type : Radiation Survey Meter
Instrument Manufacturer : Pulsecho systems.
Model inirad
Serial Number(s) L818
Instrument Received on + 15/12/2021
Environmental Conditions 22543 °C
Calibration Date 2 20/12/2021
Calibration Due Date 2 19/12/2022
Standard reference value was measured with the help of a 100cec reference graphite ionisation,
chamberin conjuction with a electrometer, traceable to Radiation Standards Section, RSSD, Bhabha
Atomic Research Centre, Mumbai
(Cettificate No, - BARC/RSSD/RSS/CAL/C-120/2017)
Method of Calibration : Placing Radiation Survey meter in front of
collimated beam from Co-60 source and checking,
reading at various distances & durations.
# This certificate shall not be reproduced in full without the approval of the calibration laboratory
Hot No B.G.A. %, Bhosari General Block, M.LD.C., Bhosari, Pune- 411026. Maharashtra
Contact No. (020) 27121843, 27119490, 8600100708,
E-mail :tanmaypangare@gmail.com / bkp_andt@ yahoo,in
Visit us at : wny.qualityndt.org
Page 1 of 2LOBAL
i.
i
——
a
QUALITY GLOBAL CALIBRATION AND INSPECTION SERVICES
‘AERB RECOGNISED CALIBRATION LABORATORY FOR RADIATION MONITORING INSTRUMENTS
NABL ACCREDITED CALIBRATION LABORATORY AS PER ISOIIEC 17025:2017
=
oon, S ‘3107
? wor
Calibration Certificate
Date of Issue : 20/12/2021 Certificate No.RT-1742
Instrument Type : Radiation Survey Meter
Instrument Manufacturer : Pulsecho systems.
Instrument Model. :Minirad
Instrument Serial.No. L818
‘Measurement Details:
Range | Expected Value [ Mean Observed | Uncertanity ]
Reading in% |
0-5 R/hr [2.50 R/hr 2.55 Rhhr 2.00 |
0-500 mR/hr [250 mR/hr 252 mR/hr, 0.80 |
0-50 mR/hr——_—(| 25 mRV/hr 25.3 mR/hr ie |
0-5 mR/hr [2.50 mR/hr 2.52 mR/¢hr 0.80
Procedure No : PRQ-O1 Reference Standard : IEC 61584
Note :
# The Calibration factor is defined as the ratio of true value to observed value.
+# This Certificate applies only to the items being calibrated. It shall not be reproduced except in full, without the
vwaitten approval of the calibration laboratory
# The reported expanded uncertainty in measurement is based on the standard uncertainty in measurement
multiplied by the coverage factor k~2, which for a normal distribution corresponds to a coverage probability
of approximately 95%
Calibrated By : SEE So
j f M7 saan
re 3 1.0.6, s
Bh angare (8 BHOSARI
RSO QD SH
2
(The Head) OFS