Download as pdf or txt
Download as pdf or txt
You are on page 1of 31

Brilliance CT 16, 40, 64,

Big Bore & DualVexp


3810C, 505C

High Voltage System Troubleshooting


Proprietary Notice:
This document and the information contained in it is proprietary and confidential information of Philips Medical Systems ("Philips") and
may not be reproduced, copied in whole or in part, adapted, modified, disclosed to others, or disseminated without the prior written
permission of the Philips Legal Department. Use of this document and the information contained in it is strictly reserved for current
Philips personnel and Philips customers who have a current and valid license from Philips for use by the customer’s designated in-
house service employee on equipment located at the customer’s designated site. Use of this document by unauthorized persons is
strictly prohibited. Report violation of these requirements to the Philips Legal Department. This document must be returned to Philips
when the user is no longer licensed and in any event upon Philips’ first written request.
PHILIPS

Objectives
After this lecture you will be able to:

• Identify Diagnostic tools used for Troubleshooting

• Use troubleshooting and diagnostic tools for the high voltage


system

• Identify expected results from Error Logs

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
2
PHILIPS

High Voltage Overview


• Spellman parts:
ƒ APM, Anode Power Module
ƒ KPM, Kathode Power Module
ƒ High Speed Starter
ƒ System Control Board
• MRC Tube
• RHOST
ƒ Heat Calculator K
ƒ Dose Modulation
• FRC
• Safety
• Service Tools
• Cautions and Precautions
• Troubleshooting

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
3
PHILIPS

Troubleshooting High Voltage System


1) System Control Board LEDs
2) RHOST test points and LEDs
3) FRC Fuses and LEDs
4) Manual Exposure: Test High Voltage without the scanner
application:
5) No Load test, In case of ARCs
6) View Logs Info Column ( include Info in filter list ! )
7) CANSENDER Commands; RHOST_DEVEL_REQ.
8) CRX-500L Theory document Detailed generator information.

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
4
PHILIPS

System Control Board LED’s


Example: Leds at power up,
state Estop_init

Explanation on metal cover RHOST and SCB

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
5
PHILIPS

System Control Board LED’s

Example: APM Phase Fault

Use Analysis Tool Doc

Open System T/S Flowchart,


Estop Init,
Generator T/S Flowchart

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
6
PHILIPS

RHOST Test Points and LEDs (Rhost service data.xls)

Click Hyperlink:
RHOSTHi Res.pdf
and use zoom

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
7
PHILIPS

FRC LED’s and Fuses (explanations on top and front cover)


FRC LEDS

Note: Led 13 may be Note: F6, F7 are 3 AMP


frequently RED since in BR40, BR64 and Big
Release 2.2.1 Bore FRC

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
8
PHILIPS

Manual Exposure
• Test x-ray generation without the scanner application nor the IRS
• Allows manual specification of X-ray shot

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
9
PHILIPS

No Load Test
• To determine if source of arc is in tube or generator. Filament is disabled -> no mA, only kV
• Run with case A. tube connected or case B. tube disconnected from generators
1. If arc occurs in case A and not in case B: source of arcing was in tube
2. If arc only occurs in both cases A and B: source of arcing was in generator

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
10
PHILIPS

Add INFO
to the Filter
settings !!!

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
11
PHILIPS

Example 1 : Reading Exposure Count, Arc Count, Filament


• CANSEND : send RHOST_DEVEL_REQ parameter 1010
1010 = Request Tube exposure data
• CANTRACE : read info in messages GANTRY_INFO_ERROR_BKST, or
• VIEWLOGS : shows the messages in the Info Column

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
12
PHILIPS

Example 2A : Readout with CANTRACE


CANSEND : send RHOST_DEVEL_REQ parameter 1161
1161 = request ALL generator registers
CANTRACE : read info in messages GANTRY_INFO_ERROR_BKST

BKST answers,
RHOST_DEVEL_REQ click on each
CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
13
PHILIPS

Example 2B : Readout with Viewlogs


CANSEND : send RHOST_DEVEL_REQ parameter 1161
VIEWLOGS : shows the same messages in the Info Column
Add INFO to
the Filter
settings !!!

Fault Registers

CPM Registers

CPM registers: DVD – Training – Training Documents – XRay decoding


Tip: Make a snapshot (Snagit) from a known good status for reference
CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
14
PHILIPS

Example 2C : Decode Generator Status Hex Codes


• Use xray decode.xls ( Incenter / DVD, Training documents )
• or use lookup table with additional info ( link in course schedule )
• Example:

Hex 9 =
Binair 1001

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
15
PHILIPS

Example 3A : Setting Grid Voltages

• CANSEND : RHOST_DEVEL_REQ 1150, 5 parameters


• CANTRACE : Here the request can also be seen after sending

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
16
PHILIPS

Example 3B : Reading Grid Voltages

• CANSEND : RHOST_DEVEL_REQ 1159


• CANTRACE : RHOST_DEVEL_ANS
• Not visible in viewlogs !

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
17
PHILIPS

List of RHOST_DEVEL_REQ and parameters

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
18
PHILIPS

CPM Failure

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
19
PHILIPS

Tube Arc

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
20
PHILIPS

CPM Arc

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
21
PHILIPS

CPM Arc

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
22
PHILIPS

APM Arc

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
23
PHILIPS

Inadequate Rail Voltage

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
24
PHILIPS

Error Log – ERROR_BLOB - GantryEalReportAns


GantryEalReportAns
provides information on
the system status of
system control boards.
i.e., Ghost, RHost, DMS,
Couch

You must select “Show


Additional Info….” to
reveal system status

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
25
PHILIPS

Error Log – ERROR_BLOB Series Load Structure


Series * Load Structure
is an outgoing message
that provides the system
control boards the with
all the parameters
necessary to run a scan
sequence

There is NO information
that is useful for
troubleshooting

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
26
PHILIPS

Error Log – ERROR_BLOB GantryEalLogsAns


GantryEalLogsAns
reveals information on
the system logs

There is NO information
that is useful for
troubleshooting

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
27
PHILIPS

X Ray Registers In Error Log Blob

If “S_EAL_NO_ERROR” is
reported by the RHost there
may not be a register dump

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
28
PHILIPS

Show BLOB Info in HEX


Register values must be displayed
in HEX for the values to be
decoded

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
29
PHILIPS

XRay Decode

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
30
PHILIPS

HV Decoding

CSIP Level 1 © 2005 Koninklijke Philips Electronics N.V. All Rights Reserved. Refer to Front Cover.
Medical Systems ACADEMY - CTC - CT 3810C, 505C 23 April 2008 – Image Calibration
31

You might also like