Quick Reference Guide

1. Power button 2. A.Power/Volume


3. Soft menu key 4. E-ink key
5. PW mode 6. Color mode
7. M mode 8. CW mode
9. Power mode 10.B mode
11.E-ink key 12.4D mode
13.Dual Window 14.E-ink key
15.E-ink key 16.Depth
17.Zoom 18.iTouch
19.E-ink key 20.Freeze
21.E-ink key 22.Angle
23.Body Mark 24.Comment
25.Clear 26.Cursor
27.Set 28.Set
29.Measurement 30.Caliper
31.Update 32.Trackball
33.Battery indicator

Note: The E-ink key is a user-defined key


1. Getting Started V Flow
New Patient 1. Under B/Color mode of Carotid exam mode with a linear probe, tap [V Flow] on the touch
1. Tap [Info] on the touch screen, fill in basic patient information. screen to enter V Flow mode.
2. Tap [Probe] on the touch screen to choose probe and exam mode. 2. Use <Trackball> and <Set> key to adjust the position and size of the ROI.
3. Tap [Start Capture] or press <Update>. The image shows the change of the blood flow over a
2. Scanning period of time.
B/M Mode: Press <B> / <M> knob to enter B / M mode and rotate the knob to adjust the gain. RIMT/R-VQS
CDFI/Power Mode 1. Under B mode of carotid exam mode with a linear probe, tap [RIMT]/[R-VQS] on the touch
screen to enter RIMT/R-VQS mode.
1. Press <C> / <PD> key to enter Color/Power mode.
2. Use <Trackball> and <Set> key to adjust the position and size of the ROI. Tap [Start Calc] to
2. Use <Trackball> and <Set> key to adjust the position and size of the ROI.
measure RIMT/R-VQS on left and right carotid. Tap [Accept Result] and save results.
PW/CW Mode
1. Press <PW> / <CW> key to enter PW/CW mode. Move <Trackball> to adjust the position of the 3. Measurement
sample volume. Press <Measure> / <Caliper> to enter the application/general measurement status. Move the cursor
2. Rotate the corresponding soft menu key to change the size of the sample volume and rotate to choose the measurement tool, and then go to the desired position to measure.
<Angle> knob to change the angle of the sample volume. 4. Post Scanning
3. Press <Update> or <PW> / <CW> key to obtain the Pulse Wave Doppler.
Comment
Free Xros M mode (Anatomical M mode)
Press <ABC> to choose the comment setting position and then add the comment to the image.
1. Under B mode, tap [Free Xros M] on the touch screen to enter Free Xros M mode.
2. Use <Set> key to switch the sample line, move <Trackball> to set the position of the sample line, Body Mark
and rotate <Angle> knob to change the angle of the sample line. Press <Body Mark> and use the <Cursor> and <Trackball> to choose the desired one. Press <Set> to
Tissue Doppler Imaging confirm adding, the added one will be highlighted in green solid frame at the bottom-left corner. Move
<Trackball> to place the probe marker, and rotate <Angle> knob to adjust the direction of the probe.
1. Under B/Color mode with phased array probe, tap [TDI] on the touch screen to enter TDI mode.
2. Use <Trackball> and <Set> key to adjust the position and size of the ROI. Save Images or Cine
3. Press <M> key to enter TVM mode; press <PW> key to enter TVD mode; press <PD> key to enter 1. Press the user defined key of <Save> to save a single-frame image or a cine to the system.
TEI mode. 2. Tap [Review] on the touch screen, or select an exam of a patient in the [iStation] screen and click
iScape [Review Image] to see the image or the cine.
1. Under B mode, tap [iScape] on the touch screen to enter iScape mode. Report and Print
2. Press <Update> to start the acquisition; press <Freeze> to stop the acquisition. 1. Tap [Report] on the touch screen and move the cursor to the comments box and type the text.
Elasto Imaging 2. Click [Add Picture] on the report page. After selecting images, click [Exit] to add images. Click
[Preview] to preview the report, and then click [Print] to print the report.
1. Under B mode with a proper probe, tap [Elasto]->[STE]/[StrainE]/[STQ] on the touch screen to
enter Elasto imaging mode. End Exam
2. Adjust ROI according to the actual situation. Tap [End] on the touch screen to end an examination. You may start a new exam by repeating the
instructions above.
3. Adjust the image parameters to obtain optimized image and necessary information.
5. Image Management
3D/4D Imaging
Press <iStation> to enter the patient data management system and choose the exams. Click [Send
1. Under B mode with a volume probe, press <3D> user-defined key or <4D> key to enter 3D/4D
Exam] to send the patient data to USB or DICOM, etc. Or, select images in Review screen, and click
mode.
[Send to] on the upper right corner to send images to USB or DICOM, etc.
2. Use <Trackball> and <Set> key to adjust the ROI and curved VOI.
3. Press <Update> to start the image acquisition; press <Freeze> to stop the image acquisition. 6. Touch Screen Gestures
Fusion Imaging Press <F10 Setup> -> [System] -> [iConsole&Footswitch] / [Key Board] / [Gesture] to assign user
1. Connect the magnetic sensor to the probe and patient. Place the magnetic generator near the frequently used function to the E-ink key (key function), foot switch, F3-16, F12, or two finger gesture.
patient. Switch the magnetic navigation controller on. 6. Touch Screen Gestures
2. Load CT/MR PET/Freehand image data in iStation screen. Tap [Fusion] on the touch screen to 1. Swipe the touch screen from top to bottom to enter the mapping mode, and then swipe the
enter Fusion mode, tap [Load Volume Data] on the touch screen to import the image data. touch screen from left to right to show the mapping menu.
3. Tap [Capture US & Resp] on the touch screen. Tap [Registration]→[Confirm Registration] on the 2. You can adjust image parameters or perform measurement through the menu.
touch screen. The plane of volume data coincides with ultrasound image plane by moving or
rotating CT/MR/PET/Freehand volume data. Tap [Mark on Volume Data] on the touch screen. Note: For detailed information, please refer to the operator’s manual.

P/N: 046-020256-00 (1.0)