Part Number: SP00223
Rev: A
Date: 19 January 2004
© 2004 Draeger Medical, Inc.
NOTE: Table 1 is for reference purposes 2. Install the gauge mount (3) in the
only. It is provided for use in the bracket, oriented with the side port
identification of replacement parts. facing the machine. Secure the gauge
The item numbers listed in the mount with the 7/8 jam nut (5) and lock
“ITEM NUMBERS’ column washer (4) supplied in the kit.
corresponds to the item numbers
used in Figures 1 through 4. 3. Install the 1/8 NPT x 1/8 hose barb
fitting (7) in the side port of the gauge
1. Attach the gauge bracket (2) to the mount. Use a small amount of Loctite
underside of the NM6000 cockpit as #271 (red) on the threads.
shown in Figure 1. Remove the existing
screws and use the 8-32 x ½ in. btn hd
screws (13) supplied in the kit, with the
original lock washers.
SP22301
GAUGE MOUNT
7/8 LOCKWASHER 4
*R 2 GAUGE BRACKET
1
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
4. Assemble the chrome T-fitting (10), 1/8 hose barb fitting (7) as shown in Figure
MPT Qdisc-M fitting (12), 1/8 NPT 2. Use a small amount of Loctite #271
Qdisc-F fitting (11), and 1/8 NPT x 1/8 (red) on the threads of all fittings.
*R
Q DISC W/VALVE-F
11
*R X 1/8 NPT FITTING
*R
10 CHROME T-FITTING
FIGURE 2. T-Assembly
2
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
5. Place a hose clamp (6) on each end of the 7. Install two stick-on cable clamps (8) in
23 in. long piece of clear hose (9) the approximate locations shown in the
supplied in the kit. illustration. Route the tubing as shown.
SP22303
HOSE CLAMP 6
STICK-ON
CABLE CLAMP 8
6 HOSE CLAMP
23 IN. LONG 9
CLEAR HOSE
T-ASSEMBLY
3
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
8. Install the breathing pressure gauge 10. Connect the T assembly to the patient
assembly (1) on the gauge mount. Orient interface panel as shown in the
the gauge as desired and tighten its illustration.
locking ring.
11. Locate the Operator’s Instruction
9. Disconnect the existing airway pressure Manual binder and insert the
hose from the patient interface panel Supplement (P/N 4117388) supplied in
and connect it to the T assembly as the kit for the breathing pressure gauge
shown in Figure 4. option.
1 BREATHING PRESSURE
GAUGE
AIRWAY PRESSURE
HOSE
SP22304 T-ASSEMBLY
FIGURE 4. Machine Connections
4
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Table 1: Kit, Breathing Pressure Gauge Asm Option (P/N 4117271) Replacement Parts List
5
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO SERVICE PROCEDURE TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS