TEST REPORT NUMBER: BGDP21000172

DATE: 29-Sep-2021

Page 1 of 7
ITS Labtest Bangladesh Ltd
Kabirpur, Ashulia, Savar, Dhaka-1349, Bangladesh, www.intertek.com
TEST REPORT NUMBER: BGDP21000172
DATE: 29-Sep-2021

APPLICANT : JMI Hospital Requisite Mfg. Ltd.


Factory: JMI Industrial Park, Vitikandi,
Gazaria-1510, Munshiganj

ATTN : Dr. Md. Zakir Hossain

Sample Description : KN95 (JMI Respirator) Mask In White


Applicant Provided Care Instructions:
-
:
Date Received/Date Test Started 18-Sep-2021
Date received Test Request Form :
18-Sep-2021
:
Testing Period 18-Sep-2021To 29-Sep-2021
Rating :
-
:
Sample Name KN 95 Mask
:
Size -
Colour :
White
:
Standard As Per Local Government Guideline
Order No :
-
Style No :
KN95, GB2626-2006
:
Buyer Name Self Reference
Manufacturer's Name :
-
:
Agent Name -
No Of Sample :
100 Pieces
:
Product End Use -
Brand Name :
-
:
Batch No JH-SP-060-21
:
Origin -
Country Of Destination :
-
:
Fabric Type -
Order Quantity :
-
:
Ref. -

AUTHORIZED BY
FOR ITS Labtest Bangladesh Ltd. [Textiles - Dhaka]

NEYAMUL HASAN
COUNTRY BUSINESS LINE LEADER, SOFTLINES

Page 2 of 7
ITS Labtest Bangladesh Ltd
Kabirpur, Ashulia, Savar, Dhaka-1349, Bangladesh, www.intertek.com
TEST REPORT NUMBER: BGDP21000172
DATE: 29-Sep-2021

TEST CONDUCTED : AS PER THE REQUEST OF THE APPLICANT. FOR FURTHER DETAILS PLEASE REFER
TO ENCLOSED PAGE(S)

CONCLUSION:
Tensile Strength Of Head Harness M
Inward Leakage M
Inhalation Resistance M
Exhalation Resistance M
Nacl Particles Filter Efficiency M
Dead Space M
Sub Micron Particulate Filtration M

Note : M = Meet Buyer's Requirement F = Below Buyer's Requirement


# = No Specified Requirement - = Did Not Perform
N/A = Not Applicable * = See Remark
C = Conform Label

Page 3 of 7
ITS Labtest Bangladesh Ltd
Kabirpur, Ashulia, Savar, Dhaka-1349, Bangladesh, www.intertek.com
TEST REPORT NUMBER: BGDP21000172
DATE: 29-Sep-2021

(TEST CONDUCTED (AS REQUESTED BY THE APPLICANT)


1. Tensile Strength Of Head Harness:
(GB 2626-2019, Clause 6.11)

As Received After Pretreatment Requirement


# (Pass) # (Pass) 10N for 10 Seconds
At Up Left
# (Pass) # (Pass)
At Up Right
# (Pass) # (Pass)
At Low Left
# (Pass) # (Pass)
At Low Right
Remark:
* = Fail With 10N For 10 Seconds And Detachment/Breakage
# = Passed With 10 N For 10 Seconds And No Detachment Or Breakage.
N = Newton

2. Inward Leakage
(GB 2626-2019 Clause 6.4)
Particles In Aerosol: NaCl
Temperature: 22.0ºC
Relative Humidity: 58%
Total Inward Leakage
TIL exercise TIL wearer
Subjects (2) (3) (4) (5) (%)
(1)
#1 2.46 3.22 4.04 2.69 3.64 3.21
As Received
(A.R) 7.90 4.06 2.39 3.30 3.69 4.27
#2
3.83 4.22 6.29 4.80 3.69 4.57
#3
4.63 2.89 3.64 3.73 4.88 3.95
#4
7.12 6.44 4.21 6.69 4.36 5.76
#5
#6 2.46 9.82 3.89 5.76 2.47 4.28
After
Temperature 3.69 5.44 4.40 4.89 3.69 4.42
#7
Control (T.C)
3.48 6.89 3.29 6.76 4.89 5.06
#8
4.76 3.60 6.44 4.89 5.44 5.03
#9
6.44 2.88 12.88 3.63 7.84 6.73
#10
• Reading1 = Waling For 2 Minutes Without Head Movement And Talking
• Reading2 = Walking For 2 Minutes Turing Head From Side To Side.
• Reading3 = Moving The Head Up & Down For 2 Minutes.
• Reading4 = Reciting The Alphabet, Agreed Test Out Loud for 2 Minutes.
• Reading5 = Walking For 2 Minutes Without Head Movement And Talking.

Page 4 of 7
ITS Labtest Bangladesh Ltd
Kabirpur, Ashulia, Savar, Dhaka-1349, Bangladesh, www.intertek.com
TEST REPORT NUMBER: BGDP21000172
DATE: 29-Sep-2021

Item Requirement Result Verdict


Individual Exercise (TIL exercise ≤11%) ≥46 49 Pass
Individual Wearer (TIL Wearer ≤8%) ≥8 10 Pass

3. Inhalation Resistance:
(GB 2626-2019, Clause 6.5)

Inhalation Resistance As Received Requirement


(Pa)
Specimen 1 157.0 ≤350 Pa
Specimen 1 152.6
Overall Result Pass

Inhalation Resistance After Temperature and Humidity Pretreatment Requirement


(Pa)
Specimen 1 147.6 ≤350 Pa
Specimen 1 153.4
Overall Result Pass

4. Exhalation Resistance:
(GB 2626-2019, Clause 6.6)

Exhalation Resistance As Received Requirement


(Pa)
Specimen 1 247.7 ≤250 Pa
Specimen 1 244.9
Overall Result Pass

Exhalation Resistance After Temperature and Humidity Pretreatment Requirement


(Pa)
Specimen 1 244.2 ≤250 Pa
Specimen 1 249.2
Overall Result Pass

Page 5 of 7
ITS Labtest Bangladesh Ltd
Kabirpur, Ashulia, Savar, Dhaka-1349, Bangladesh, www.intertek.com
TEST REPORT NUMBER: BGDP21000172
DATE: 29-Sep-2021

(TEST CONDUCTED (AS REQUESTED BY THE APPLICANT)


5. NaCI Particles Filter Efficiency:
(GB 2626-2019, Clause 6.3)

Teste Condition
Air Flow : 85 L/min
Particles Aerosol : NaCl Concentration Of Aerosol : 37.0 mg/m3
Temperature : 21.0ºC Relative Humidity: : 40.0%

Filter Efficiency (%) As Received Requirement

Specimen 1 99.6 ≥95%


Specimen 2 99.9
Specimen 3 97.3
Specimen 4 96.9
Specimen 5 97.0
Specimen 6 98.6
Specimen 7 99.5
Specimen 8 99.6
Specimen 9 99.2
Specimen 10 99.9
Specimen 11 99.7
Specimen 12 99.9
Specimen 13 99.6
Specimen 14 99.9
Specimen 15 99.9

Filter Efficiency (%) After Temperature and Humidity Pretreatment Requirement

Specimen 1 99.9 ≥95%


Specimen 2 99.9
Specimen 3 99.0
Specimen 4 98.5
Specimen 5 98.3

Page 6 of 7
ITS Labtest Bangladesh Ltd
Kabirpur, Ashulia, Savar, Dhaka-1349, Bangladesh, www.intertek.com
TEST REPORT NUMBER: BGDP21000172
DATE: 29-Sep-2021

(TEST CONDUCTED (AS REQUESTED BY THE APPLICANT)


6. Dead Space:
(GB 2626-2019, Clause 6.9)

Facepiece Category: Disposable Facepiece


Requirement
Concentration Of CO2 (%) 0.81% ≤1%

7. Sub Micron Particulate Filtration


(ASTM F2100-19ε1, Section 9.3, Testing Refer to ASTM F2299/F2299M-17): Particle size in
aerosol: 0.1 µm, Aerosol: PSL, Test area: 60 cm2, Airflow Test: 37.4 L/Min, Sampling
time: 1 min.
Tested Particle Filtration Efficiency (%) Requirement
Sample/Component
Specimen 1 98.4 ≥95%
Specimen 2 95.2
Specimen 3 97.5
Specimen 4 98.6
Specimen 5 97.5
Average 97.4

## END OF THE TEST REPORT ##


This report is made solely on the basis of your instructions and/or information and materials supplied by you. It is not intended to be a recommendation for any
particular course of action. Intertek does not accept a duty of care or any other responsibility to any person other than the Addressee in respect of this report and
only accepts liability to the Addressee insofar as is expressly contained in the terms and conditions governing Intertek's provision of services to you. Intertek makes
no warranties or representations either express or implied with respect to this report save as provided for in those terms and conditions. We have aimed to conduct
the Review on a diligent and careful basis and we do not accept any liability to you for any loss arising out of or in connection with this report, in contract, tort, by
statute.For Terms and condition follow the link: http://www.intertek.com/terms.

Page 7 of 7
ITS Labtest Bangladesh Ltd
Kabirpur, Ashulia, Savar, Dhaka-1349, Bangladesh, www.intertek.com

You might also like