ECNF19008 TS-CSA-Detector Sensing Function, - Auto Tracking Function, LCD Rotation Speed (3 Bug Issues)

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

Engineering Change Notice

ECNF19008

Engineering Change Notice

1. Information
TS-CSA : Detector sensing function, Auto Tracking function, LCD Rotation
Subject
Speed (3 Bug Issues)
Registration Number ECNF19008

Revision Number 0

Department R&D

Originator Sangwook SEO

Reviewed by Youngmin SEO

Approved by Kinam YANG

ECN issued Date 2019.07.08

Applicable due Date 2019.07.08


TS-CSA, TS-CSA Source Tilting, WBS-TA, WBS-TM (Motorized),
Applicable Model
WBS (Motorized)
Affected Products FDR-SMART-FGXR

In TS-CSA System of EMC 3rd case, if the firmware version of Ceiling Main
Control Board is V1.00.14 or V1.00.15, it must be changed to V1.00.16
firmware.

Action It is also necessary to upgrade the previous version firmware of Ceiling


Switch Control Board to V1.00.05 version.

If the firmware version of Communication Board is V1.00.08, it must be


changed to V1.00.09 firmware.

Change Information

Item Model Before After Comment

F/W Ceiling Main Control Board Ver 1.00.15 Ver 1.00.16 Function Change
Ceiling Switch Control
F/W Ver 1.00.04 Ver 1.00.05 Function Change
Board
F/W Communication Board Ver 1.00.08 Ver 1.00.09 Function Change

DRGEM Corporation

7F, E-B/D Gwangmyeong Techno-Park, 60 Haan-ro, Gwangmyeong-si, Gyeonggi-do, Korea 1/4


Tel: +82-2-869-8566, Fax: +82-2+869+8567, E-mail: drgem@drgem.co.kr
Engineering Change Notice

ECNF19008

2. Description of change
1. Detector sensing issue (PBT)
TS-CSA has function for sensing direction of 17x14, 17x14 detector in PBT
(EMC 3rd)
We prepared firmware for EMC 4TH. This firmware can be installed for EMC
3rd and EMC 4TH both System.
But when we installed this firmware on EMC 3rd System, TS-CSA cannot
detect whether detector is mounted in the Bucky tray of PBT or not and
Reason for change detector direction.

2. Auto tracking function issue


In WBS Right direction case, the Bucky tray in PBT moved opposite
direction of Tube angle.

3. LCD Rotation speed was slow intermittently.

Software N/A
Ceiling_Main_Control_Board_V1.00.16
Firmware Communication_Board_V1.00.09
Ceiling Switch Control Board_V1.00.05
Hardware N/A

Document N/A

Version Date Subject


In EMC 3 case, TS-CSA cannot
rd

detect whether detector is mounted


in the Bucky tray of PBT or not and
detector direction.
This has been modified.
Ceiling_Main_Control_Board_V1.00.15
2019.06.26
Communication_Board_V1.00.09 In right direction of WBS case, the
Bucky tray in PBT move opposite
direction of Tube angle.
This has been modified.

DRGEM Corporation

7F, E-B/D Gwangmyeong Techno-Park, 60 Haan-ro, Gwangmyeong-si, Gyeonggi-do, Korea 2/4


Tel: +82-2-869-8566, Fax: +82-2+869+8567, E-mail: drgem@drgem.co.kr
Engineering Change Notice

ECNF19008

LCD Rotation speed is slow

Ceiling Switch Control Board _V1.00.05 2019.06.26 intermittently.


This bug has been modified.

Modified contents

1. Detector sensing issue (PBT)

- TS-CSA checks direction of 14x17, 17x14, 17x17 Detector via limit switch.

- In EMC 3rd case, TS-CSA cannot detect whether detector is mounted in the Bucky tray of PBT
or not and detector direction.

- This bug has been modified.

- Application method of EMC 3rd and EMC 4th equipment is different. The Application method is
as follows.

 In case of EMC 3rd equipment, Turn ‘off’ Table Communication button of ETC tab in
System Setting as shown below.

 In case of EMC 4th equipment, Turn ‘on’ Table Communication button of ETC tab in
System Setting as shown below.

Current Firmware Version of FGXR Current Firmware Version of FGXR


Ceiling_Main_Control_Board_V1.00.15 Ceiling_Main_Control_Board_V1.00.16
(TS-CSA) (TS-CSA)

DRGEM Corporation

7F, E-B/D Gwangmyeong Techno-Park, 60 Haan-ro, Gwangmyeong-si, Gyeonggi-do, Korea 3/4


Tel: +82-2-869-8566, Fax: +82-2+869+8567, E-mail: drgem@drgem.co.kr
Engineering Change Notice

ECNF19008

2. Auto tracking function issue

- In right direction of WBS case, the Bucky tray in PBT move opposite direction of Tube angle.

- This bug has been modified.

Current Firmware Version of FGXR Change Firmware Version of FGXR


Communication_Board_V1.00.08 Communication_Board_V1.00.09

3. LCD Rotation issue

- LCD Rotation speed is slow intermittently.

- This bug has been modified.

Current Firmware Version of FGXR Change Firmware Version of FGXR


Switch Control Board _V1.00.04 Switch Control Board _V1.00.05

3. Contact Point
 Customer Support Team

E-mail: cs@drgem.co.kr

Tel: +82-70-8230-8541

DRGEM Corporation

7F, E-B/D Gwangmyeong Techno-Park, 60 Haan-ro, Gwangmyeong-si, Gyeonggi-do, Korea 4/4


Tel: +82-2-869-8566, Fax: +82-2+869+8567, E-mail: drgem@drgem.co.kr