Contemporary digital ultrasonic flaw detectors calculate commonly used parameters for reflector position in angle beam testing, typically including sound path length, flaw depth, surface
distance, and beam leg. These calculations are based on a measured pulse transit time, calibrated material sound velocity and probe zero offset, and entered values for test piece
thickness, beam angle, and beam index point. By default, calculations assume a flat test piece. In the case of circumferential inspection of tubular parts, the calculations must be
adjusted for the effects of test piece curvature. This is done by means of a mathematical correction based on entered outside and inside diameters of the test piece. The following topics
will be reviewed.

1. Mathematical Theory
2. Previous Approaches
3. Implementation of Curved Surface Correction in EPOCH series flaw detectors
4. Measurement examples
5. Use of contoured wedges

1. Mathematical theory
In the case of common flat plate tests and axially-oriented tests of pipes and similar products, leg length and reflector depth and surface distance can be calculated through simple
trigonometry through the formulas below. But when testing a curved part in the circumferential direction, these formulas become inaccurate due to the curvature that effectively causes
reflectors to drop away from the probe, which lengthens the sound path and changes the reflection angle at the inside surface of the part. For example, compare similar conditions for a
45° beam in flat and curved plates both 20 mm thick. The signal obtained at a 35mm sound path length in a flat plate corresponds to a reflector located at a depth of 15.2mm, 24.8mm
from the exit point. But the same 35 mm sound path in a 100mm diameter pipe corresponds to a reflector 19.2 mm deep from the OD surface and 31.7mm along the test surface from the
exit point to the point over the reflector.

https://w w w .olympus-ims.com/es/resources/w hite-papers/theory-and-use-csc/ 1/10


11/10/23, 11:19 Theory and Use of Curved Surface Correction (CSC) Softw are in Olympus NDT EPOCH Series Flaw Detectors | Olympus IMS

To correct for curvature, a number of mathematical relationships can be applied, based on the angles, measured sound path and dimensions defined below. Utilizing these formulas with
leg management enables proper calculation of surface projection and vertical depth of flaws.

Legend for equations:


S = Surface Distance
Β = Angle of Refraction
Α = Sound path distance
t = Thickness
R1 = Inside Radius
R2 = Outside Radius
Βmax = Maximum angle of refraction
Ε = Angle of reflection at ID
Φ = Projection angle
D = Depth of reflector
1. Projection angle at ID (first leg)
Φ = sin-1(R2/R1 sinΒ)

2. Sound path distance Α, half skip


Α = R1∗sin(Φ_Β)/sin(Β)

3. Depth of reflector D
D = R2 – √Α2 + R22 - 2R2ΑcosΒ

4. Surface distance S
S = ΠR2/180[sin-1(R2/R1 sinΒ)-Β] Note: for first leg test, R1 = t-D

5. Maximum angle of refraction for a given wall thickness (angle of impingement)


Βmax = sin-1(R1/R2)

2. Previous approaches

https://w w w .olympus-ims.com/es/resources/w hite-papers/theory-and-use-csc/ 2/10


11/10/23, 11:19 Theory and Use of Curved Surface Correction (CSC) Softw are in Olympus NDT EPOCH Series Flaw Detectors | Olympus IMS

Traditional analog flaw detectors measured sound path length only and could not automatically calculate reflector depth or surface distance, or count skip legs. Live automatic displays of
these additional dimensional parameters, based on trigonometric calculations, were introduced with the development of digital instruments. However early model digital flaw detectors
calculated depth for flat test pieces only. Both with these instruments and analog flaw detectors, operators performing circumferential testing on pipes and similar curved test pieces used
graphs and nomograms such as the ones below to calculate dimensions.

The graph at left is used to determine a correction factor k that is based on refracted angle and the ratio of wall thickness and outside diameter. The nomogram at right is used to
determine the actual reflected angle at the inside of the curved surface, which because of the curvature is less than the refracted beam angle at the sound entry point.

Once k and the refracted angle Β have been determined, it is possible to calculate the actual leg length AB from the formula:

AB = tk cosΒ

Once the leg length is known, a reflector can be accurately placed within the first or second leg of the sound path based on the measured sound path length. Further calculations of flaw
depth and surface position must be done manually. As an alternative to calculations, look-up tables have been published that relate these geometrical corrections to test piece diameter,
wall thickness, and inspection angles.
https://w w w .olympus-ims.com/es/resources/w hite-papers/theory-and-use-csc/ 3/10
11/10/23, 11:19 Theory and Use of Curved Surface Correction (CSC) Softw are in Olympus NDT EPOCH Series Flaw Detectors | Olympus IMS

In the absence of software for curved surface correction, another alternative to manual calculations or look-up tables was to empirically set up flaw detectors with the aid of a reference
standard representing the tube wall in question and incorporating a series of side drilled holes. Reflections from the multiple reference holes can be identified with respect to horizontal
position and a screen overlay can be used to mark their vertical depth and surface projected positions on the display. The resulting screen display is horizontally non-linear, but reflector
positions can be extrapolated from it by referencing the overlay.

3. Implementation of Curved Surface Correction in EPOCH series flaw detectors


The Curved Surface Correction (CSC) software in the EPOCH series flaw detectors performs calculations automatically based on setup information input by the user.

These calculations assume that initial probe calibration for sound velocity and zero offset have been performed as usual on an IIW-type block or similar flat reference block. Following that
calibration, the operator enters the probe angle, the thickness of the test piece, the “X value” (the distance between the beam index point and front edge of the wedge), and the part

https://w w w .olympus-ims.com/es/resources/w hite-papers/theory-and-use-csc/ 4/10


11/10/23, 11:19 Theory and Use of Curved Surface Correction (CSC) Softw are in Olympus NDT EPOCH Series Flaw Detectors | Olympus IMS

diameter. The EPOCH software then uses that information plus the measured sound path length to calculate and display the following, based on the equations in Section 1.

sound path length

Projected surface distance from front of probe (or beam index point if no X-value is input) along outside of the test piece

depth of reflector

Assuming proper calibration and setup, these measurements will be as accurate as those on a flat test piece in a given setup. The example screens below show a typical measurement
of a first leg reflector in a thick wall reference sample, shown with and without CSC correction. When the correction is applied, the flaw depth measurement is shallower (1.13” versus
1.31”) and the surface distance or projection is farther from the probe (1.39” versus 1.07”).

CSC correction off CSC correction on


Additional notes:

In all tests that involve circumferential probe alignment it is very important to hold the probe normal to the surface rather than tilting it. Tilting will skew the sound beam and
effectively change the refracted angle in the test piece.
In cases where the ratio of wall thickness to diameter is large, the sound beam at a given angle may not reflect off the ID, instead continuing straight across to the OD without a
skip. (See section 1, equation 6.) In these cases CSC calculations are not valid and the EPOCH will not report dimensional information.

4. Example measurements using EPOCH 600


First leg measurements of pipe section with 0.1” side drilled holes as shown in photograph, using EPOCH 600 flaw detector and A543S-RM transducer with ABWM-4T-45 wedge.
Diameter is 5.0”, wall thickness 0.470”. Reflectors are 0.1” diameter side drilled holes with centerline depths of 0.160” and 0.310” from OD (1/3 and 2/3 thickness).

The SDH diameter factor causes a small offset in the reading. When establishing the physical location of the reflector within the part it is important to measure depth and surface
projection to the normal of the outer radius tangent. The measured sound path displayed is not affected by CSC correction status. Proper calibration of zero offset and material velocity
are required. In the examples below:

Physical measurement of sound path to hole OD


Surface projection is measured from the beam index point (X Value=0)

https://w w w .olympus-ims.com/es/resources/w hite-papers/theory-and-use-csc/ 5/10


11/10/23, 11:19 Theory and Use of Curved Surface Correction (CSC) Softw are in Olympus NDT EPOCH Series Flaw Detectors | Olympus IMS

First leg measurements

FIRST LEG CSC Off CSC ON Physical Measurement

Sound Path 0.47 0.47 0.46

Depth 0.33 0.31 0.31

Surface distance 0.33 0.38 0.40

CSC off CSC on

Second leg measurements

https://w w w .olympus-ims.com/es/resources/w hite-papers/theory-and-use-csc/ 6/10


11/10/23, 11:19 Theory and Use of Curved Surface Correction (CSC) Softw are in Olympus NDT EPOCH Series Flaw Detectors | Olympus IMS

SECOND LEG CSC Off CSC ON Physical Measurement

Sound Path 1.14 1.14 NA

Depth 0.14 0.27 0.22

Surface distance 0.8 1.04 1.03

CSC off CSC on

Third leg measurements of pipe section with 0.1” side drilled holes.

https://w w w .olympus-ims.com/es/resources/w hite-papers/theory-and-use-csc/ 7/10


11/10/23, 11:19 Theory and Use of Curved Surface Correction (CSC) Softw are in Olympus NDT EPOCH Series Flaw Detectors | Olympus IMS

THIRD LEG CSC Off CSC ON Physical Measurement

Sound Path 1.78 1.78 NA

Depth 0.32 0.16 0.160

Surface distance 1.26 1.54 1.50

CSC off CSC on

5. Use of contoured wedges


As the diameter of a cylindrical test piece become smaller, circumferentially oriented coupling of conventional wedges becomes progressively more challenging, due to reduced contact
area and the increased possibility of rocking. For that reason, special cylindrically contoured wedges are sometimes used on smaller diameter test pieces. As a guideline, the
International Institute of Welding recommends contouring whenever test piece radius R is less than the length of the wedge W squared divided by four:

https://w w w .olympus-ims.com/es/resources/w hite-papers/theory-and-use-csc/ 8/10


11/10/23, 11:19 Theory and Use of Curved Surface Correction (CSC) Softw are in Olympus NDT EPOCH Series Flaw Detectors | Olympus IMS

The SDH diameter factor causes a small offset in the reading. When establishing the physical location of the reflector within the part it is important to measure depth and surface
projection to the normal of the outer radius tangent. The measured sound path displayed is not affected by CSC correction status. Proper calibration of zero offset and material velocity
are required. In the examples below:

R W2 / 4

ASTM standard E-164, Standard Practice for Ultrasonic Contact Examination of Weldments, calls for use of radiused wedges whenever the part diameter is less than 20 inches or 500
mm. Olympus NDT experience suggests that this is a conservative recommendation, especially when using small angle beam assemblies, but it is up to the user to determine
applicability to specific test.

The minimum outside diameter for which a wedge of a given size can be cylindrically contoured depends on mechanical considerations related to wedge size and angle. The following
chart lists the approximate minimum contouring diameter for three common Panametrics-NDT wedge types.

Minimum OD for cyl. contour 45 degree 60 degree 70 degree

ABWM-4T 0.8” or 20 mm 1.1” or 28 mm/td> 1.5” or 38 mm

ABWM-5T 1.5” or 38 mm 2.2” or 56 mm 2.8” or 70 mm

ABWM-6T 1.2” or 30 mm 1.7” or 43 mm 2.4” or 60 mm

Panametrics-NDT contoured wedges maintain a beam index point at the BIP location marked on the wedge, and the nominal wedge angle is calculated based on beam refraction at that
point. However it is important to note that cylindrical wedge contouring affects the incident angle and thus the refracted angle from front to back across the diameter of the beam, resulting
in a range of refracted angles and effectively a defocusing and broadening of the beam in that axis.

When using contoured wedges, appropriately contoured reference blocks must also be used to calibrate zero offset and verify beam index point and refracted angle. Conventional IIW and
AWS-type reference blocks cannot be used because of the gap between the block and the concave surface of the wedge.

Suscríbase a los boletines de noticias

Imprimir

Soluciones para inspeccionar defectos y espesores

Videoscopios y boroscopios

Microscopios industriales

https://w w w .olympus-ims.com/es/resources/w hite-papers/theory-and-use-csc/ 9/10


11/10/23, 11:19 Theory and Use of Curved Surface Correction (CSC) Softw are in Olympus NDT EPOCH Series Flaw Detectors | Olympus IMS

Analizadores XRF y XRD

Servicios y asistencia

Material didáctico

Copyright EVIDENT, All rights reserved.

Terms Of Use | Privacy Notice | Cookies | Configuración de cookies | About Us | Mapa del sitio

https://w w w .olympus-ims.com/es/resources/w hite-papers/theory-and-use-csc/ 10/10

You might also like