OPERATOR'S MANUAL
Baker Thimble
Exhaust Connection
SterilGARD® SG-400: — SG4-TEC
SterilGARD® SG-600: — SG6-TEC
Note: Thimble connections are not a substitute for
hard connections. Your biosafety officer may find it
desirable to hard connect Class I Type A/B3
cabinets under certain conditions. Do not attempt
to use a TEC with a biological safety cabinet other
than the two specified above. Class II Type BI and
B2 cabinets must be hard connected to function
properly. If there are any questions regarding
thimble or hard connections, please call the Baker
Company at (207) 324-8773.
THe BAKER COMPANY
"Creating immaculate atmospheres"Installation Instructions
STEP 1
Remove existing
hardware around the |
perimeter of the
exhaust opening
located on top of the |
Biological Safety 1
Cabinet. (BSC)
If the unit is equipped
with an internal exhaust colfar, it must be
removed.
STEP 2 e
Place the supplied
silicon gasket over the
studs which surround
the perimeter of the
exhaust opening. The i
studs should be forced |
through the gasket
material. The comers i
should be partially cut at 90° angles so that one
piece of gasket encompasses all of the studs.
‘This will minimize the number of gasket joints.
STEP 3
Remove Decon Seal
Plate from the thimble
by unscrewing the
four (4 wingnuts on
top of the thimble.
STEP 4
Insert four (4)
rubber bumpers and
two (2) thumbscrews,
(supplied in cloth bag)
into the holes left open
as a result of removing
the hardware from the
Decon Seal Plate.
STEP 5
Place thimble exhaust
connection on to
gasketing and studs
on top of BSC.
Replace hardware
removed during
Step 1. Thimble label
must be facing
towards the front of the cabinet.
| exhaust system to the
STEP 6
Attach the inhouse
cight inch (8") dia-
meter collar at the top
of the thimble, The
connection must not
cover or interfere with
the two (2) airflow
measuring ports
Tue Bakr
CoMmPANy
Neen eee eeDecontamination Procedure
STEP 1
Use a 5/16" nut driver
(or wrench) to remove
the five (5) cap nuts
on the front access
panel. Remove the
panel to allow access
for installing Decon
Seal Plate
STEP 2
Remove the four (4)
rubber bumpers and fh
the nwo (2)
thumbserews from the
top of the thimble.
STEP 3
Insert the Decon Seal
Plate with the studs
facing upwards
through the ace:
door opening. Pu:
the four (4) studs
through the holes left
open by the rubber
bumpers,
STEP 4
Place flat washer and
wingnut over each
stud and screw tightly.
This will seal the plate
against the top of the
thimble,
STEP 5
Seal off the front and
rear slots with a
non-porous tape.
WARNING:
‘This step is critical
to contain
decontaminant.
Thimble Exhaust Connection portion of BSC
is now ready for decontamination procedure.
Ma ee
STEP 6
After the
decontarnination
process, loosen all
four (4) wingnuts to i
the full height of the :
stud. Now turn
thumbscrews (2) into | 2
the weld nuts ontop Lo
of the thimble. This will break the seal (suction) to
allow for cabinet ventilation,
Tue BAKER COMPANY
eeMeasuring Cabinet Exhaust
STEP 2
Measure the airflow
through the front and
rear slots, again
following the Thimble
Airflow Measuring
Procedure. Calculate
STEP 1
Using the wo (2)
airflow ports, take a
ten-point traverse
using the directions
provided by the
Thimble Airflow
Measuring Procedure, Total Slot CFM
located on the front through both stots.
access panel. Calculate the Total Duct CFM The actual cabinet exhaust equals Total Stot
through the thimble connection. subtracted from Total Duet
STEP 1
See STEP | Decontamination Procedure. This will allow access through the
front panel to locate the leak checking device over the exhaust HEPA filter
without having to remove the thimble.
STEP 1
Remove the four (4) large rubber grommets at the base of each comer of the
thimble.
STEP 2
Insert a 3/8" nut driver* through the resulting hole to Joosen the four (4) filter
clamps.
STEP 3
The HEPA filter can now be removed through the front of the BSC following
the instructions provided on the BSC owner operator's manual
Sor 7/16" 12 point socket
Tue Baker Company