léÉê~íáåÖ=fåëíêìÅíáçåë
båÖäáëÜ
Operating Instructions
=
Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS XG 3
Table of contents
1 General data............................................................................................................ 6
båÖäáëÜ
1.9 Structure of the document ............................................................................ 10
1.9.1 Identification of the danger levels...................................................... 10
1.9.2 Formats and symbols used ............................................................... 10
2 Safety information.................................................................................................... 11
60 50 681 D3352
D3352.201.05.16.02 08.2013 3
Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS XG 3
3 Unit description......................................................................................................... 17
4.1 Replacing the bite block, contact segment, or chin rest ................................ 29
4.2 Changing the temple supports and temporomandibular joint supports ......... 30
4.3 Removing/inserting the sensor...................................................................... 31
5 Operation.................................................................................................................. 32
60 50 681 D3352
4 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS XG 3
båÖäáëÜ
5.8 Using the remote control ............................................................................... 52
5.9 Canceling an exposure ................................................................................. 53
5.10 Reprogramming the kV/mA values ............................................................... 54
5.11 Activating the Info menu................................................................................ 54
5.12 Activating the service menu .......................................................................... 55
6 Maintenance............................................................................................................. 56
7 Malfunctions ............................................................................................................. 60
60 50 681 D3352
D3352.201.05.16.02 08.2013 5
1 General data Sirona Dental Systems GmbH
1.1 Dear Customer, Operating Instructions ORTHOPHOS XG 3
1 General data
General data
systems, and since 1996 has been a pioneer of digital X-ray technology.
You benefit from the vast experience we have gained through the
thousands of digital panoramic X-ray systems with CCD sensors installed
worldwide. This device is characterized by many features including
outstanding image quality, simple operation, and a high day-to-day
reliability.
This device enables you to take the following digital images:
Standard and temporomandibular joint exposures
60 50 681 D3352
6 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 1 General data
Operating Instructions ORTHOPHOS XG 3 1.3 Contact information
Fabrikstrasse 31
64625 Bensheim
Germany
Phone: +49 (0) 6251/16-0
Trademark
båÖäáëÜ
The X-ray system includes other components, such as PC software,
Other valid documents
60 50 681 D3352
D3352.201.05.16.02 08.2013 7
1 General data Sirona Dental Systems GmbH
1.5 Warranty and liability Operating Instructions ORTHOPHOS XG 3
60 50 681 D3352
8 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 1 General data
Operating Instructions ORTHOPHOS XG 3 1.7 Intended use
● Conservative dentistry
● Endodontics
● Periodontology
● Prosthodontics
● Functional diagnosis and therapy of craniomandibular dysfunctions
båÖäáëÜ
● Surgical dentistry
● Implantology
● Oral and maxillofacial surgery
● Orthodontics
Contraindications:
● Display of cartilage structures
● Display of soft tissue
60 50 681 D3352
D3352.201.05.16.02 08.2013 9
1 General data Sirona Dental Systems GmbH
1.9 Structure of the document Operating Instructions ORTHOPHOS XG 3
DANGER
Imminent danger that could result in serious bodily injury or death.
WARNING
Potentially dangerous situation that could result in serious bodily injury
or death.
CAUTION
Potentially dangerous situation that could result in slight bodily injury.
NOTICE
Potentially harmful situation which could lead to damage of the product
or an object in its environment.
IMPORTANT
Instructions for use and other important information.
The formats and symbols used in this document have the following
meaning:
60 50 681 D3352
10 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 2 Safety information
Operating Instructions ORTHOPHOS XG 3 2.1 Information on the unit
2 Safety information
Safety information
Accompanying documents
Accompanying documents
båÖäáëÜ
Connector pins or sockets bearing ESD warning labels must not be
touched or interconnected without ESD protective measures. See also
"Electrostatic Discharge" and "Electromagnetic Compatibility" [ → 15].
since otherwise the air circulation will be obstructed. This can cause the
unit to overheat.
Do not spray into the ventilation slots Do not spray liquids such as disinfectants into the ventilation slots. This
may lead to malfunctions. Use wipe disinfection only in the vicinity of the
ventilation slots.
2.3 Condensation
Extreme fluctuations of temperature may cause condensation inside the
Safety information for condensation: Customer
Condensation
unit. Do not switch the unit on before it has reached normal room
temperature. See the chapter on “Technical data” [ → 19].
60 50 681 D3352
D3352.201.05.16.02 08.2013 11
2 Safety information Sirona Dental Systems GmbH
2.4 Qualifications of operating personnel Operating Instructions ORTHOPHOS XG 3
personnel.
Personnel undergoing education or training, or who are using the device
Qualification
as part of general training may only operate the unit under the constant
supervision of properly trained personnel.
To operate the unit, the operating personnel must:
● have read and understood the Operating Instructions
● be familiar with the fundamental structure and functions of the unit
● be able to recognize irregularities in the functioning of the unit and
implement the appropriate measures where necessary
If any parts of the unit touch the patient during the rotary movement, let
go of the exposure release button (X-Ray) immediately or stop the unit at
once by actuating the unit main switch or an Emergency Stop switch!
60 50 681 D3352
12 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 2 Safety information
Operating Instructions ORTHOPHOS XG 3 2.8 Laser light localizer
The light localizers are intended for correct patient positioning. They must
not be used for any other purposes.
A minimum distance of 10 cm (4") is required between the eye and the
laser. Do not stare into the beam.
The light localizers may be switched on only when functioning perfectly.
Repair work must be carried out by authorized staff only.
Do not use the system with any other lasers, and do not make any
changes to settings or processes that are not described in these
operating instructions. This may lead to a dangerous exposure to
radiation.
2.9 Hygiene
båÖäáëÜ
The protective sleeves must be exchanged and all auxiliary exposure
Hygiene
free.
The movements of the unit must not be obstructed by physical
Obstruction of unit movements
60 50 681 D3352
D3352.201.05.16.02 08.2013 13
2 Safety information Sirona Dental Systems GmbH
2.13 Combination with other equipment Operating Instructions ORTHOPHOS XG 3
60 50 681 D3352
14 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 2 Safety information
Operating Instructions ORTHOPHOS XG 3 2.16 Electromagnetic compatibility
båÖäáëÜ
Protective measures
Electrostatic discharge
60 50 681 D3352
D3352.201.05.16.02 08.2013 15
2 Safety information Sirona Dental Systems GmbH
2.17 Electrostatic discharge Operating Instructions ORTHOPHOS XG 3
60 50 681 D3352
16 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 3 Unit description
Operating Instructions ORTHOPHOS XG 3 3.1 Certification
3 Unit description
Unit description
3.1 Certification
ORTHOPHOS XG 3 complies with:
with AS/NZS 3200.1.0
Certification
● IEC 60601-1
● IEC 60601-1-3 / 2008
● IEC 60601-2-63
● AS/NZS 3200.1.0
The dental X-ray equipment for extraoral radiography ORTHOPHOS XG
D3352 complies with IEC 60601-2-63: 2012
Original language: German
This product bears the CE marking in accordance with the provisions of
CE mark, general
båÖäáëÜ
0123
60 50 681 D3352
D3352.201.05.16.02 08.2013 17
3 Unit description Sirona Dental Systems GmbH
3.2 Technical data Operating Instructions ORTHOPHOS XG 3
Image acquisition scale: For P1, normal dental arch (slice center)
approx. 1:1.19, i.e. the acquired image is
magnified by approx. 19% on average
compared to reality.
Exposure time for a 14.9 s max.
cephalometric image:
Image acquisition scale for a approx. 1:1.1, i.e. the acquired image is
cephalometric image: magnified by approx. 10% on average
compared to reality.
assembly:
Focal spot size acc. to IEC 0.5 mm
60336, measured in the
central X-ray beam:
Marking of focal spot:
Class I device
Operating unit data, 100 W long-term power output
60 50 681 D3352
18 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 3 Unit description
Operating Instructions ORTHOPHOS XG 3 3.2 Technical data
temperature:
Air humidity: 10% – 95%
Admissible operating Acc. to IEC 60601-1 between +10°C and
temperature: +40°C (50°F – 104°F)
Operating altitude: ≤ 3000 m
X-ray tube
X-ray tube and PAN sensor
Siemens SR 90/15 FN or
CEI OCX 100
PAN Sensor
Digital CCD line sensor, repluggable for panoramic exposure technique
båÖäáëÜ
Detail resolution: 0.027 mm pixel size
Focus-sensor distance: 497 mm
Diagrams, 2D
60 50 681 D3352
D3352.201.05.16.02 08.2013 19
3 Unit description Sirona Dental Systems GmbH
3.2 Technical data Operating Instructions ORTHOPHOS XG 3
60 50 681 D3352
20 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 3 Unit description
Operating Instructions ORTHOPHOS XG 3 3.2 Technical data
båÖäáëÜ
Network card: Network RJ45, 100 MBit/s
USB port: For Version 1.1 and higher, required for USB
components only
Software: Internet Explorer 6.0, SP1
Acrobat Reader 8.0, contained on CD, required for
the PDF test report function
60 50 681 D3352
D3352.201.05.16.02 08.2013 21
3 Unit description Sirona Dental Systems GmbH
3.3 Main components of the product Operating Instructions ORTHOPHOS XG 3
A B C D E F G H I
PAN CEPH TS TS
P11260 14.1 64 8
Prog. P1S kV mA
64
14,2 s
64kV
AEC 8mA
T R
T R ORTHOPHOS ist aufnahmebereit ?
Q P O N M L K J
A Main switch
B Light localizer with height adjustment of light line (Frankfurt
horizontal plane)
C Light localizer central light beam for face center
D Control mirror for patient positioning
E Tray for jewelry, etc.
F Forehead support
G Temple supports
H Pushbutton for sensor removal
I Sensor
J Primary diaphragm field on the X-ray tube assembly
K Bite block, contact segment or chin rest
L Holder for chin rest, bite blocks, or contact segments etc.
M Handle for patient
N Drawer for accessories
O Multipad (swiveling control panel)
P Touch bar for swiveling the control mirror in and out
Q Release button
60 50 681 D3352
22 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 3 Unit description
Operating Instructions ORTHOPHOS XG 3 3.3 Main components of the product
3.3.2 Multipad
Multipad
A B C D E F G H I J K L
P1 14.1 64 8
Prog. S kV mA
båÖäáëÜ
T R
T S R Q P O N M
60 50 681 D3352
D3352.201.05.16.02 08.2013 23
3 Unit description Sirona Dental Systems GmbH
3.3 Main components of the product Operating Instructions ORTHOPHOS XG 3
Q "R" key for unit return with Ready LED (flashes if the unit is not
ready for an exposure).
R "T" key for test cycle without radiation with LED display
S "Close temple supports" key
T "Open temple supports" key
60 50 681 D3352
24 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 3 Unit description
Operating Instructions ORTHOPHOS XG 3 3.3 Main components of the product
A B A B
P 1 14.1 64 8 P 1 14.1 64 8
Prog. s kV mA Prog. s kV mA
båÖäáëÜ
R E R
F
D
A Radiation indicator
B "Unit ON" LED display
C Display field
D Exposure release button
E "R" key for return of unit
F Exposure release button with coiled cable
60 50 681 D3352
D3352.201.05.16.02 08.2013 25
3 Unit description Sirona Dental Systems GmbH
3.4 Spare parts and consumables Operating Instructions ORTHOPHOS XG 3
A F
B C
E D
60 50 681 D3352
26 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 3 Unit description
Operating Instructions ORTHOPHOS XG 3 3.4 Spare parts and consumables
D
2
1
C
A
båÖäáëÜ
F
60 50 681 D3352
D3352.201.05.16.02 08.2013 27
3 Unit description Sirona Dental Systems GmbH
3.4 Spare parts and consumables Operating Instructions ORTHOPHOS XG 3
60 50 681 D3352
28 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 4 Installation and start-up
Operating Instructions ORTHOPHOS XG 3 4.1 Replacing the bite block, contact segment, or chin rest
exposure program.
1. Pull the accessories upwards and out of the holder.
ª The accessory part disengages.
2. Insert the bite block, contact segment, or chin rest.
ª The accessory part engages.
The chin rest can be combined with the bite block rod or the bar.
➢ Insert the rod for the bite block or the bar into the chin rest from
above.
båÖäáëÜ
60 50 681 D3352
D3352.201.05.16.02 08.2013 29
4 Installation and start-up Sirona Dental Systems GmbH
4.2 Changing the temple supports and temporomandibular joint supports Operating Instructions ORTHOPHOS XG 3
(A) "1" right and (C) "2" left must be inserted in place of the temple
supports (B).
✔ Temple supports are inserted in the unit.
Procedure
B 1. Press the respective locking button and remove the temple supports
B.
ª Both temple supports are removed.
2. Insert a sterile ear holder D into temporomandibular supports A and
1
2 C.
ª The ear holders snap into the temporomandibular joint supports.
A C 3. Insert temporomandibular joint supports A and C into the holders on
the device.
ª The temporomandibular joint supports snap into place.
ª The unit is converted for temporomandibular joint exposures.
D
60 50 681 D3352
30 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 4 Installation and start-up
Operating Instructions ORTHOPHOS XG 3 4.3 Removing/inserting the sensor
NOTICE
Electrostatic charges from persons are discharged on the unit.
Electrical components of the unit are destroyed.
➢ Do not touch any electrical components or unprotected plug
contacts.
➢ Discharge yourself by touching a conductive grounded object.
båÖäáëÜ
Removing the sensor
Removing/inserting the sensor
60 50 681 D3352
D3352.201.05.16.02 08.2013 31
5 Operation Sirona Dental Systems GmbH
5.1 Switching the unit on Operating Instructions ORTHOPHOS XG 3
5 Operation
Operation
CAUTION
Switching the unit on
NOTICE
Condensation: Customer
D T R
60 50 681 D3352
32 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.2 Readings on the digital display
CAUTION
båÖäáëÜ
Pressing the R key moves the unit to the starting position.
H 301 14.1 64 8
Prog. S kV mA
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it to
T R the starting position.
If the Ready LED above the R key starts flashing and error message H
301 appears on the display, briefly press return key R to bring the rotating
element into the starting position. The Ready LED then switches off and
the help message disappears.
The system is now ready to operate.
60 50 681 D3352
D3352.201.05.16.02 08.2013 33
5 Operation Sirona Dental Systems GmbH
5.4 Panoramic exposure Operating Instructions ORTHOPHOS XG 3
60 50 681 D3352
34 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.4 Panoramic exposure
båÖäáëÜ
P1 C – Panoramic exposure, constant 1.25x magnification
IMPORTANT
Please ensure that the 1.25 x enlargement is only guaranteed at the
horizontal level. As patient positioning can vary, a reference object is to
be used at the point where a measurement is to be performed.
60 50 681 D3352
D3352.201.05.16.02 08.2013 35
5 Operation Sirona Dental Systems GmbH
5.4 Panoramic exposure Operating Instructions ORTHOPHOS XG 3
60 50 681 D3352
36 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.4 Panoramic exposure
båÖäáëÜ
5.4.3 Selecting an exposure program
The exposure programs appear in the sequence P1, P1 L, P1 R, P1 A,
Selectingsequence
Program an exposure
withprogram
panoramic and temporomandibular exposures only
A B
➢ Select the exposure program. Push the program selection key
forward (A) and backward (B).
ª The program number, the appropriate exposure time and the
P1 14.1 64 8
programmed kV/mA values for the second patient symbol from
Prog. S kV mA the left appear on the digital display.
ª The exposure program is selected.
T R
60 50 681 D3352
D3352.201.05.16.02 08.2013 37
5 Operation Sirona Dental Systems GmbH
5.4 Panoramic exposure Operating Instructions ORTHOPHOS XG 3
A
Manual setting of kV/mA values
If the preset kV/mA combinations do not provide a satisfactory result, you
can also set the kV/mA values manually in all programs.
➢ Select another kV/mA value. Push the kV/mA keys forward (B) and
B backward (C).
ª The selected kV/mA value is shown on the digital display. If the
new value happens to agree with the value programmed for
another patient symbol key, its LED then lights up.
P1 14.1 64 8
Prog. S kV mA
ª The kV/mA value is set.
T R
60 50 681 D3352
38 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.4 Panoramic exposure
CAUTION
The height adjustment motor starts slowly and then increases its speed.
A patient positioned in the unit may be injured by moving parts.
➢ Monitor the patient and the movement of the unit during height
adjustment.
➢ To make minor corrections, press and immediately release the
keys.
Laser light localizer
CAUTION
The light localizer consists of one Class 1 laser.
Patients and users can be blinded by the laser light localizer.
båÖäáëÜ
➢ Do not stare directly into the laser beam. Make sure that the laser
beam does not meet the eyes of the patient.
➢ A distance of at least 10 cm must be maintained between the eye
and the laser.
IMPORTANT
Metallic objects
Ask the patient to take off all metallic objects such as glasses and
jewelry in the head and neck area as well as all removable dental
prostheses. The tray in front of the control mirror is used for depositing
jewelry.
Tip: As long as a height adjustment key is pressed, the digital display
Multipad reference values
shows a reference value for the height setting which is saved in the
1233 additional information area of the SIDEXIS software.
Prog. S kV mA
60 50 681 D3352
D3352.201.05.16.02 08.2013 39
5 Operation Sirona Dental Systems GmbH
5.4 Panoramic exposure Operating Instructions ORTHOPHOS XG 3
5.4.5.1 Positioning with chin rest and rod for bite block
Positioning with chin rest and rod for bite block
✔ The chin rest and bite block segment, as well forehead support and
temple supports are inserted in the unit.
✔ The relevant hygienic protective sleeves are pulled over the
accessories.
1. Guide the patient in front of the control mirror.
2. Set the unit height using the up (A) and down (B) keys. CAUTION!
A The height adjustment motor starts slowly and then increases its
speed.Press and hold the until the desired height is reached. The
motor movement is accompanied by an acoustic signal.
ª The patient's chin and the chin rest on the unit are at the same
B height.
3. Turn the bite block away from the patient.
ª The bite block is pointing towards the control mirror.
4. Instruct the patient to place his chin on the chin rest and place both
hands on the hand grips.
5. Turn the bite block towards the patient and instruct him to bite on the
bite block.
ª The patient's anterior teeth are positioned in the indentation in the
bite block. If necessary, push the lower anterior teeth forwards
until they reach a stop.
60 50 681 D3352
40 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.4 Panoramic exposure
6. Check the patient's occlusal plane (C). Correct the height of the unit
A using the up (A) and down (B) keys.
ª The occlusal plane is slightly inclined toward the front.
7. Check the position of the patient's spine.
ª Ensure that the patient's spine is slightly inclined, as shown in the
diagram.
Tip: To achieve the correct positioning of the spine, ask the
patient to take a small step towards the column of the unit. The
patient's cervical vertebrae are thus stretched. This prevents
C B regions of diminished density in the anterior tooth region.
8. Swivel the control mirror outwards by pressing the left recess on the
touchbar (D).
ª You can see the patient in the control mirror.
båÖäáëÜ
light localizer, press the key again. It switches off automatically
after approx. 100 seconds.
10. Align the patient with the central light line (G).
ª The light beam reflects in the center of the patient's anterior teeth
or the middle of his face (mid-sagittal).
G
F
E
D
11. Align the patient's head according to the Frankfurt horizontal plane
A E (E).
Tip: The Frankfurt horizontal is used as a reference plane. It runs
between the upper edge of the ear canal and the deepest point of the
lower eye socket edge.
12. Adjust the height of the light localizer using the slider (F).
ª The light beam reflects on the upper edge of the outer ear canal.
13. Correct the patient's head inclination as necessary. Briefly press the
up (A) and down (B) height adjustment keys.
C B ª The light beam reflects on the lowest point of the lower eye
socket edge.
60 50 681 D3352
D3352.201.05.16.02 08.2013 41
5 Operation Sirona Dental Systems GmbH
5.4 Panoramic exposure Operating Instructions ORTHOPHOS XG 3
14. Press the forehead support adjustment key (H) and the temple
support adjustment key (I).
ª The forehead and temple supports stop moving automatically
when they come into contact with the patient's head.
Ensure that the patient's head does not move backward when the
forehead support is put in place.
15. Check the patient's position and make any final corrections as
necessary.
16. Swivel the control mirror back into place by pressing the right recess
on the touchbar (D).
ª The patient can see himself in the control mirror.
17. Instruct the patient to exhale, place his tongue against the roof of his
mouth, and to hold this position until the end of the exposure.
H I ª The patient is positioned in the unit.
60 50 681 D3352
42 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.4 Panoramic exposure
✔ The chin support and bar, and the forehead support and temple
supports are inserted in the unit.
✔ The relevant hygienic protective sleeves are pulled over the
accessories.
1. Instruct the patient to place his chin on the chin rest and place both
hands on the hand grips.
2. Instruct the patient to place his subnasale (the base of his nose)
against the bar. If the patient's lower jaw contains anterior teeth, place
the bar between his chin and his lower lip.
3. Place a cotton roll between the patient's upper and lower jaw.
ª The patient's upper and lower jaw are aligned.
4. Proceed as described under "Positioning with chin rest and rod for
bite block" [ → 40] from step 6.
båÖäáëÜ
5.4.5.3 Positioning with bite block
✔ The yellow bite block, forehead support and temple supports are
Positioning with bite block
60 50 681 D3352
D3352.201.05.16.02 08.2013 43
5 Operation Sirona Dental Systems GmbH
5.5 Temporomandibular joint exposure Operating Instructions ORTHOPHOS XG 3
aspect with the mouth open and closed and provides 4 views in one
image.
A B
➢ Select the exposure program. Push the program selection key
forward (A) and backward (B).
ª The program number, the appropriate exposure time and the
P1 14.1 64 8
programmed kV/mA values for the second patient symbol from
Prog. S kV mA the left appear on the digital display.
ª The exposure program is selected.
T R
60 50 681 D3352
44 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.5 Temporomandibular joint exposure
båÖäáëÜ
A
Manual setting of kV/mA values
If the preset kV/mA combinations do not provide a satisfactory result, you
can also set the kV/mA values manually in all programs.
➢ Select another kV/mA value. Push the kV/mA keys forward (B) and
B backward (C).
ª The selected kV/mA value is shown on the digital display. If the
new value happens to agree with the value programmed for
another patient symbol key, its LED then lights up.
TM1.1 12.8 71 8
Prog. S kV mA
ª The kV/mA value is set.
T R
60 50 681 D3352
D3352.201.05.16.02 08.2013 45
5 Operation Sirona Dental Systems GmbH
5.5 Temporomandibular joint exposure Operating Instructions ORTHOPHOS XG 3
CAUTION
The height adjustment motor starts slowly and then increases its speed.
A patient positioned in the unit may be injured by moving parts.
➢ Monitor the patient and the movement of the unit during height
adjustment.
➢ To make minor corrections, press and immediately release the
keys.
Laser light localizer
CAUTION
The light localizer consists of one Class 1 laser.
Patients and users can be blinded by the laser light localizer.
➢ Do not stare directly into the laser beam. Make sure that the laser
beam does not meet the eyes of the patient.
➢ A distance of at least 10 cm must be maintained between the eye
and the laser.
IMPORTANT
Metallic objects
Ask the patient to take off all metallic objects such as glasses and
jewelry in the head and neck area as well as all removable dental
prostheses. The tray in front of the control mirror is used for depositing
jewelry.
Tip: As long as a height adjustment key is pressed, the digital display
Multipad reference values
shows a reference value for the height setting which is saved in the
1233 additional information area of the SIDEXIS software.
✔ The forehead support and temporomandibular joint supports with ear
Prog. S kV mA Positioning for a lateral temporomandibular joint exposure
2. Set the unit height using the up (A) and down (B) keys. CAUTION!
A The height adjustment motor starts slowly and then increases its
speed.Press and hold the until the desired height is reached. The
motor movement is accompanied by an acoustic signal.
ª The ear holders for the temporomandibular joint support are at
B the same height as the patient's ears.
60 50 681 D3352
46 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.5 Temporomandibular joint exposure
båÖäáëÜ
I
6. Switch on the light localizer. CAUTION! Risk of dazzle
ª Two red light lines reflect on the patient's head. To switch off the
light localizer, press the key again. It switches off automatically
after approx. 100 seconds.
7. Align the patient with the central light line (G).
ª The light beam reflects in the center of the patient's anterior teeth
or the middle of his face (mid-sagittal).
G
F
E
D
60 50 681 D3352
D3352.201.05.16.02 08.2013 47
5 Operation Sirona Dental Systems GmbH
5.5 Temporomandibular joint exposure Operating Instructions ORTHOPHOS XG 3
12. Check the patient's position and make any final corrections as
necessary.
13. Swivel the control mirror back into place by pressing the right recess
on the touchbar (D).
ª The patient can see himself in the control mirror.
14. Instruct the patient to exhale, place his tongue against the roof of his
mouth, and to hold this position until the end of the exposure.
ª The patient is positioned in the unit.
H
60 50 681 D3352
48 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.6 Starting the test cycle
check that the unit is functioning correctly and to ensure that a complete,
uninterrupted cycle is possible. The rotating unit stops automatically if the
resistance increases.
1. Press the T key.
ª The program enters test cycle mode. The LED above the T key
lights up. Only the exposure program number appears on the
P1
Prog. S kV mA
digital display. None of the LEDs above the patient symbol keys
light up.
2. Press the release button.
T R
ª The test cycle is started.
3. Wait until the test cycle has been completed.
4. Press the T key again.
båÖäáëÜ
ª The program exits test cycle mode.
If the unit is positioned in an x-ray room that is equipped with a door lock
and visual contact with the patient can be guaranteed, the exposure
should be released using the remote control, see "Using the remote
control" [ → 52].
Radiation protection measures
WARNING
The unit emits X-ray radiation.
Excess exposure to X-rays is detrimental to health.
➢ Use the prescribed accessories for radiation protection.
➢ Do not stay in the X-ray room during exposure. Move as far away
from the unit as the coiled cable for the release button allows you to.
Unit movement
CAUTION
The movement of the system may be adversely affected by the patient's
physical constitution, clothing, or dressings, or by wheelchairs or
hospital beds.
The exposure is automatically terminated if the movement of the unit is
inhibited. The exposure must be repeated.
➢ Ensure that the movement of the unit is not impaired when
positioning the patient. Before triggering the exposure, perform a
test cycle using the T key.
Letting go of the release button prematurely
CAUTION
Prematurely letting go of the release button cancels the exposure
immediately.
The exposure must be repeated.
➢ Take care not to let go of the exposure release button prematurely.
Press the release button until the end of the exposure. Note that
radiation may be released several times during an exposure cycle.
60 50 681 D3352
D3352.201.05.16.02 08.2013 49
5 Operation Sirona Dental Systems GmbH
5.7 Releasing the exposure Operating Instructions ORTHOPHOS XG 3
CAUTION
The exposure memory of the unit is cleared when the unit is switched
off.
Images that have not been transferred to SIDEXIS are irretrievably lost.
The exposure must be repeated.
➢ Wait until the exposure data have be completely transferred. Do not
switch the unit off before the X-ray exposure is displayed on the
SIDEXIS screen.
Moving into the starting position
CAUTION
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it to
the starting position.
Automatic exposure blocking
IMPORTANT
23 Premature release of a new exposure is prevented by the automatic
R Prog. S kV mA
exposure blocking function. This function is used for thermal protection
of the X-ray tube.
After the release button is pressed, the expiring cooling time in seconds
appears on the digital display. If you let go of the release button before
the cooling time has expired, the Ready LED above the "R" key also
starts flashing. After you press the "R" key, the program data again
appears on the digital display.
Another exposure can be released only after the cooling period has
elapsed.
✔ The help message H... must not appear alternating on the digital
display of the Multipad.
Tip: If the doors of the X-ray room are not properly shut, the message
"H321" (Close the door) appears on the digital display of the Multipad
and on the remote control.
✔ The exposure program, radiation time, and kV/mA values are
selected and appear on the digital display C.
✔ The patient is positioned in the unit.
PAN
1260 CEP
H
TS
64
LS
T
R P6.1
12,1s
Filmk
asset
te einra 62kV
sten 8mA
IMPORTANT
The patient's arms must hang down freely at his/her sides, He/she
should not tense or hunch his/her shoulders.
Advise the patient not to move his/her head in any way during the
exposure and check to make sure that this does not happen.
60 50 681 D3352
50 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.7 Releasing the exposure
1. Press release button (A) and hold it down until the end of the
exposure.
A
ª The exposure is released. The rotary movement of the selected
exposure program is performed automatically. During radiation,
the optical radiation indicator B lights up on the Multipad. In
addition, an acoustic signal sounds throughout the entire
B radiation time. Radiation can be released several times during
the exposure.
2. Press and hold release button A. Wait until a short pulsed tone
sequence sounds (can be deactivated by a service engineer). The
P1
Prog.
14.1 64 8
S kV mA
C exposure is finished when a row of dots and the program number are
alternately displayed on the Multipad. The end of the exposure cycle
T R
can also be seen on the SIDEXIS monitor, namely when the progress
indicator shows 100% and the preview image starts to build up.
ª The X-ray image is displayed on the PC monitor in SIDEXIS. The
forehead and temple supports open automatically.
båÖäáëÜ
3. Let go of release button (A).
P1 14.1 64 8 ª The exposure is completed. The confirmation of the exposure
Prog. kV mA
data appears on the digital display of the Multipad. The program
S
R
number, the elapsed radiation time, the tube voltage, the tube
H 320 14.1 64 8
Prog. S kV mA current, error/help messages and the dose area product are
displayed.
H 320 1234mGycm
2
The Ready LED above the R key flashes.
Prog. S kV mA
4. Guide the patient out of the unit.
5. Press the Return key R on the Multipad.
ª Confirmation of the exposure data is acknowledged.
6. WARNING! The patient may be injured by moving parts. Press the
Return key R on the Multipad again.
ª The rotating unit moves to the starting position. The Ready LED
switches off.
ª The device is now ready for the next exposure.
With the two-part exposure program TM 1 two exposures (TM 1.1 and TM
1.2) are prepared.
TM1.2 12.8 71 8 ✔ The first temporomandibular joint exposure has been released as
Prog. S kV mA described above. The exposure program has changed from TM 1.1
to TM 1.2 on the digital display of the Multipad. The rotating unit has
automatically moved to the starting position.
1. Instruct the patient to open his/her mouth.
ª The patient has opened his/her mouth without changing his/her
position.
2. Press the release button A and hold it down until the end of the
second exposure.
ª The second exposure is released.
3. Wait until a short pulsed tone sequence sounds (can be deactivated
by a service engineer).
60 50 681 D3352
D3352.201.05.16.02 08.2013 51
5 Operation Sirona Dental Systems GmbH
5.8 Using the remote control Operating Instructions ORTHOPHOS XG 3
R
E R
To test the function of the radiation display (A), it is briefly illuminated for
D approx. one second.
F
Use the Return key (E) to acknowledge exposures and move the rotating
unit to the starting position.
If a row of dots appears in the Prog. field on the digital display (C), this
means the device is currently in a preparatory phase (e.g. device
movements, parameter settings, program loading times, etc.). Wait until
the dots disappear automatically and the system signals that it is ready
again.
.......... 8.1
Prog. s
60 50 681 D3352
52 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.9 Canceling an exposure
CAUTION
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it to
the starting position.
båÖäáëÜ
Prog. S kV mA displayed. The LED above the R key flashes.
:
:
2. Guide the patient out of the unit.
3. Press the R key.
2
H 320 1234mGycm
Prog. S kV mA
ª The rotating unit moves to the starting position.
ª The device is now ready for the next exposure.
IMPORTANT
The program settings must be checked before the exposure is repeated.
Any changed program settings must be preselected again.
60 50 681 D3352
D3352.201.05.16.02 08.2013 53
5 Operation Sirona Dental Systems GmbH
5.10 Reprogramming the kV/mA values Operating Instructions ORTHOPHOS XG 3
exposure program and for each individual patient symbol key can be
overwritten (reprogrammed).
IMPORTANT
Termination of the programming mode
1. Select the exposure program for which the values are to be changed.
A B
Press the program selection keys (A) forward/backward.
ª The desired program is displayed.
2. Briefly press the Memory key (C).
P1 14.1 64 8 ª The LED above the Memory key lights up.
Prog. S kV mA
3. Select the patient symbol corresponding to the kV/mA value you want
to change. Press the desired patient symbol key (D).
ª The LED above the patient symbol key lights up.
T R
4. Set the new kV/mA value. Push the kV/mA keys (B) forward and
backward.
D C ª The desired kV/mA value is displayed.
5. Save the setting. Briefly press the Memory key (C).
ª The LED above the Memory key briefly flashes and then switches
off. The program display jumps back to P1.
ª The new kV/mA value is programmed.
P1 14.1 64 8 Systemsoftware
2. Select the individual parameters from the list. Press the program
Prog. S kV mA
>2 s
Prog. S kV mA
selection keys (B) forward/backward.
ª The desired parameter is displayed.
V02 .12 . 00 Systemsoftware
Prog. S kV mA Prog. S kV mA
3. Briefly press the Memory key (A).
ª The value selected for the parameter is displayed.
P1
Prog.
14.1 64 8
S kV mA 4. Press the service menu key (D) briefly.
ª The list of parameters is displayed.
D
5. Press the service menu key (D) briefly.
ª The Info menu is terminated. The unit displays exposure
readiness.
60 50 681 D3352
54 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 5 Operation
Operating Instructions ORTHOPHOS XG 3 5.12 Activating the service menu
1. Press and hold the service menu key (B) for at least 2 seconds.
A
ª The LED above the service menu key (B) lights up, followed by
the LEDs above all patient symbol keys (C). The operational
readiness LED (A) flashes.
2. Enter the PIN.
Prog. S kV mA
ª The service menu is opened.
NOTE: If an incorrect PIN is entered or no entry is made within 5
T R
seconds, the program then returns to exposure readiness.
3. Press the service menu key briefly (B).
ª The service menu is terminated. The unit displays exposure
readiness.
C B
båÖäáëÜ
60 50 681 D3352
D3352.201.05.16.02 08.2013 55
6 Maintenance Sirona Dental Systems GmbH
6.1 Cleaning and care Operating Instructions ORTHOPHOS XG 3
6 Maintenance
Maintenance
6.1.1 Cleaning
Cleaning
NOTICE
Cleaning the ventilation slots
6.1.2 Disinfecting
Disinfecting
CAUTION
Cleaning and care agents may contain aggressive ingredients.
Unsuitable cleaning and care agents are detrimental to health and
attack the surface of the unit.
➢ Do NOT use: Substances containing phenol, peracetic acid,
peroxide or any other oxygen-splitting agents, sodium hypochlorite
or iodine-splitting agents.
➢ Use only cleaning and disinfecting agents approved by Sirona!
60 50 681 D3352
56 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 6 Maintenance
Operating Instructions ORTHOPHOS XG 3 6.1 Cleaning and care
båÖäáëÜ
60 50 681 D3352
D3352.201.05.16.02 08.2013 57
6 Maintenance Sirona Dental Systems GmbH
6.1 Cleaning and care Operating Instructions ORTHOPHOS XG 3
6.1.3 Sterilization
Sterilization
Sterilization of accessories
WARNING
Infections can be transmitted from patient to patient.
Accessories that are not sterilized correctly can cause illness in patients.
➢ All accessories that are suitable for sterilization should only be
sterilized in an autoclave at 134 °C (273° F), with at least 3 minutes
holding time and at 2.1 bar (30.5 psi) overpressure.
134°C 273°F
WARNING
The hygienic protective sleeves single use devices.
Unsterile hygienic protective sleeves can cause illness in patients.
➢ Replace the hygienic protective sleeves after each patient.
60 50 681 D3352
58 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 6 Maintenance
Operating Instructions ORTHOPHOS XG 3 6.2 Inspection and maintenance
intervals to protect the health and safety of patients, users and other
persons.
Annual inspection
In order to ensure the operational safety and functional reliability of your
product, you as the system owner should check the equipment at regular
intervals (at least once a year) or commission your dental depot to do so.
The information provided in the supplied document "Inspection and
maintenance" should be helpful here.
båÖäáëÜ
The information provided in the supplied document "Inspection and
maintenance" should be helpful here.
Country-specific requirements
Country-specific requirements
60 50 681 D3352
D3352.201.05.16.02 08.2013 59
7 Malfunctions Sirona Dental Systems GmbH
7.1 Help messages Operating Instructions ORTHOPHOS XG 3
7 Malfunctions
Malfunctions
CAUTION
Pressing the R key moves the unit to the starting position.
A patient positioned in the unit may be injured by moving parts.
➢ Check that a patient is not positioned in the unit before moving it to
the starting position.
The sensor is not in the appropriate slot for the selected exposure type.
➢ Plug the sensor into the PAN sensor slot.
ª The program sequence is continued.
60 50 681 D3352
60 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 7 Malfunctions
Operating Instructions ORTHOPHOS XG 3 7.2 Error message structure
CAUTION
The exposure memory of the unit is cleared when the unit is switched
off.
Images that have not been transferred to SIDEXIS cannot be restored
using SiRescue.
➢ Never switch off the unit before transferring the images to SIDEXIS
with SiRescue.
båÖäáëÜ
The error messages are displayed on the device in the form of an error
Error message structure
code. The display does not show any plain text error output.
The error codes are structured according to the following pattern: Ex yy zz
Explanation of abbreviations:
Ex – Error type
The x character provides a foundation for making quick decisions as to
how serious the error is and how to handle the error.
yy – Locality
Describes the impaired function of the device.
zz – Identification
Further specification of the error with a consecutive number.
60 50 681 D3352
D3352.201.05.16.02 08.2013 61
7 Malfunctions Sirona Dental Systems GmbH
7.3 Error description Operating Instructions ORTHOPHOS XG 3
E1 – System warning/message
The error is in an acceptable tolerance range. Device operation is not
directly impaired.
1. Acknowledge the error message.
2. Contact your Customer Service.
ª Continued device operation is ensured.
E2 – Overload
The error can be traced back to temporary overheating or something
similar.
1. Acknowledge the error message.
2. Wait for a moment and repeat the procedure step. If the error
reappears, extend the waiting time.
ª The error no longer occurs after a certain waiting period.
3. If the error persists, contact your Customer Service.
E4 – Unassigned
60 50 681 D3352
62 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 7 Malfunctions
Operating Instructions ORTHOPHOS XG 3 7.3 Error description
E6 – Self-check
The error occurs spontaneously and without a corresponding operation.
1. Acknowledge the error message.
ª The error no longer occurs.
2. If the error remains, switch the unit off and on again. NOTICE!
Observe waiting period!
ª The error no longer occurs.
3. If the error persists, contact your Customer Service.
båÖäáëÜ
7.3.2 yy – Locality
yy – Locality
81 – Sensor
42 – Remote
61 – Diaphragm control
60 50 681 D3352
D3352.201.05.16.02 08.2013 63
8 Program values Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS XG 3
8 Program values
Index 2A
This level series is factory programmed for the Federal Republic of
Germany. It guarantees that the applicable legal regulations which must
be complied with since January 1, 1999 are strictly observed.
Furthermore, this level series can also be applied worldwide. National
regulations must be complied with. The table lists the maximum exposure
times.
Level series for index 2A
kV 60 60 60 60 60 62 64 66 68 71 73 77 80 85 90
mA 3 5 6 7 8 8 8 8 8 8 8 7 7 6 6
60 50 681 D3352
64 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 8 Program values
Operating Instructions ORTHOPHOS XG 3
Index 4A
Index 4A
Index 4A, which specifies a reduced level series for children and
adolescents, must be complied with as a minimum requirement in all new
installations and systems moved from one location to another in the
Federal Republic of Germany since January 1, 1999. Furthermore, this
level series can also be applied worldwide. National regulations must be
complied with. The table lists the maximum exposure times.
Level series for index 4A
båÖäáëÜ
P1L 12.9s 8.0s 62/8 64/8 69/15 73/15
P1R 12.9s 8.0s 62/8 64/8 69/15 73/15
P1A 21.8s 14.1s 62/8 64/8 69/15 73/15
P1C 20.1s 14.1s 62/8 64/8 69/15 73/15
P20 16.4s 11.3s 62/8 64/8 69/15 73/15
BW10 23.0s 8.8s 62/8 64/8 69/15 73/15
TM1.1+ 16.1+16.1s 6.4+6.4s 68/8 71/8 73/15 77/14
TM1.2
kV 60 60 60 60 60 62 64 66 68 71 73 77 80 85 90
mA 3 5 6 7 8 8 8 8 8 8 8 7 7 6 6
kV 60 60 60 60 60 62 64 66 69 71 73 77 80 84 90
mA 9 10 12 14 16 16 16 16 15 15 15 14 14 13 12
60 50 681 D3352
D3352.201.05.16.02 08.2013 65
8 Program values Sirona Dental Systems GmbH
Operating Instructions ORTHOPHOS XG 3
Index 1A
Index 1A
kV 60 60 60 60 60 62 64 66 69 71 73 77 80 84 90
mA 9 10 12 14 16 16 16 16 15 15 15 14 14 13 12
60 50 681 D3352
66 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 8 Program values
Operating Instructions ORTHOPHOS XG 3 8.1 Dose information
Sirona has been calculated already. The DAP value can be used without
any further calculations.
Specification of the dose-area product (DAP/absorbed dose) for
DAP ORTHOPHOS XG 3
båÖäáëÜ
P1A 14.2s 62/8 45 64/8 47 68/8 53 73/8 60
P1C 14.2s 62/8 45 64/8 47 68/8 53 73/8 60
P20 11.3s 62/8 36 64/8 38 68/8 43 73/8 49
BW10 8.8s 62/8 28 64/8 30 68/8 33 73/8 38
TM1.1+TM1.2 6.4+6.4s 68/8 48 71/8 52 73/8 55 77/7 53
calculate the value using the kV/DAP lists; see sample calculation:
Explanation
Art. 3.3 of the X-Ray Ordinance requires that the system must include
either devices that show the
DAP display for the radiation exposure of the patient, or that this
information can be derived e.g. from tables.
The manufacturers of dental equipment have agreed on using the same
measurement method. To compensate for measuring errors as well as for
system and instrument variations, a tolerance of 20 % must be taken into
account.
The radiation exposure is specified as the dose area product (DAP) of the
absorbed dose (Gy x cm2 per mAs) for each unit as well as each
selectable kV level and diaphragm.
Calculation
The values indicated below have been calculated for the parameter
combinations proposed by Sirona. If you use other setting values,
proceed as follows using the kV/DAP lists as an aid:
1. Select the set kV level from the table of the respective X-ray system
and note down the DAP factor.
2. Multiply the DAP factor by the actually used mA (as indicated on the
X-ray system).
3. Multiply the result by the actual exposure time (see Multitimer or
table).
60 50 681 D3352
D3352.201.05.16.02 08.2013 67
8 Program values Sirona Dental Systems GmbH
8.1 Dose information Operating Instructions ORTHOPHOS XG 3
Sample calculation
X-ray with program P1 and a kV/mA parameter combination of 64 kV/
10 mA
Ad 1. 64 kV has a DAP factor of 0.41 with diaphragm 10
Ad 2. 10 mA displayed
Ad 3. The exposure time is 14.1 s
60 50 681 D3352
68 D3352.201.05.16.02 08.2013
Sirona Dental Systems GmbH 9 Dismantling and disposal
Operating Instructions ORTHOPHOS XG 3 9.1 Dismantling and reinstallation
9.2 Disposal
Your product is marked with the adjacent symbol. Within the European
Economic Area, this product is subject to Directive 2002/96/EC as well as
the corresponding national laws. This directive requires environmentally
sound recycling/disposal of the product. The product must not be
disposed of as domestic refuse!
båÖäáëÜ
Please observe the disposal regulations applicable in your country.
Disposal procedure
We advise that this product is subject to the stipulations in EC guideline
2002/96 governing waste electrical and electronic equipment and must
be disposed of in line with these special requirements within the
European Union (EU).
Prior to disassembly / disposal of the product, it must be fully prepared
(cleaned / disinfected / sterilized).
When disposing of equipment permanently, please proceed as follows:
In Germany:
To initiate return of the electrical device, please send a disposal order to
"enretec GmbH".
1. You can find a form for placing a disposal order on the company's
homepage (www.enretec.de) under the menu item "Entsorgung
elektrischer und elektronischer Geräte" (Disposal of electric and
electronic devices). The form can either be downloaded or completed
online.
2. Fill out the form with the corresponding details and send it either as
an online order or fax it to enretec GmbH at +49(0)3304 3919 590.
You can also get in touch with the following contacts for disposal
orders and any questions relating to this you may have:
Phone: +49(0)3304 3919 500;
E-mail: pickup@eomRECYCLING.com
Mailing address: enretec GmbH, Geschäftsbereich eomRECYCLING
Kanalstrasse 17, 16727 Velten
ª Your nonpermanently installed unit will be collected in the practice
and your permanently installed unit will be picked up curbside at your
address by appointment.
All disassembly, transport and packaging costs are to be borne by the
owner/operator of the equipment. The disposal itself is free of charge.
60 50 681 D3352
D3352.201.05.16.02 08.2013 69
9 Dismantling and disposal Sirona Dental Systems GmbH
9.2 Disposal Operating Instructions ORTHOPHOS XG 3
60 50 681 D3352
70 D3352.201.05.16.02 08.2013