Professional Documents
Culture Documents
Horizon Service Manual
Horizon Service Manual
Service Manual
MAN-03310 Revision 004
Horizon™
Bone Densitometry System
Service Manual
Part Number: MAN-03310
Revision 004
Customer Support
Toll Free in USA: +1.800.321.4659
Email: SkeletalHealth.Support@hologic.com
In Europe, South America, or Asia, contact your local dealer or distributor.
© 2017 Hologic, Inc. Printed in the USA. This manual was originally written in English.
Hologic, APEX, Discovery, Horizon, QDR, and associated logos are trademarks and/or registered trademarks of Hologic, Inc. and/or
its subsidiaries in the U.S. and/or other countries. All other trademarks, registered trademarks, and product names are the property of
their respective owners.
This product may be protected by one or more U.S. or foreign patents as identified at www.Hologic.com/patents.
Horizon Service Manual
Table of Contents
Table of Contents
1 Introduction ...............................................................................................................................1
1.1 System Overview...............................................................................................................................................1
1.1.1 Purpose ......................................................................................................................................................1
1.1.2 Operation...................................................................................................................................................3
Patient Positioning and Scanning..............................................................................................................3
Technique Factors........................................................................................................................................3
1.1.3 BMC Results ..............................................................................................................................................4
1.1.4 Shielding ....................................................................................................................................................4
1.1.5 Automatic Internal Reference System for Daily Calibration .............................................................4
1.1.6 X-ray Scans ................................................................................................................................................4
1.2 X-ray Scanning Principles ................................................................................................................................4
1.2.1 Mounting ...................................................................................................................................................4
1.2.2 Beam Filter.................................................................................................................................................4
1.2.3 Program .....................................................................................................................................................4
1.3 Functional Overview ........................................................................................................................................5
1.3.1 Operator Console .....................................................................................................................................6
1.3.2 Distribution Board....................................................................................................................................7
Interface.........................................................................................................................................................7
Motor Control Subsystem ..........................................................................................................................8
Routing..........................................................................................................................................................8
C-arm Interface Board.................................................................................................................................8
C-arm Subsystem.........................................................................................................................................9
Detector Board ...........................................................................................................................................10
Control Panel Board ..................................................................................................................................10
1.4 Leakage Technique Factors............................................................................................................................15
1.4.1 Beam Filtration........................................................................................................................................16
1.4.2 Beam Quality...........................................................................................................................................16
1.4.3 Line Voltage and Maximum Line Current .........................................................................................16
1.4.4 Maximum Deviation from Indicated Values......................................................................................16
1.4.5 Measurement Criteria for Technique Factors.....................................................................................16
8.17.7 Calibration (Area, BMD, and BMC) with ACF and BCF Correction Factors ............................378
8.17.8 RECALYZE and Add Array AP Scans to the QC Database.........................................................379
15 Troubleshooting..................................................................................................................492
15.1 Hardware and Software Issues ...................................................................................................................492
15.1.1 Motion Errors......................................................................................................................................492
15.2 System Index Errors ......................................................................................................................................500
15.2.1 How to Read QDR Error Messages .................................................................................................500
Log File Example .....................................................................................................................................500
Index Error 32, 40, 42, 50, or 605 ............................................................................................................500
Index Error 184.........................................................................................................................................501
Index Error 194 or 195 .............................................................................................................................501
Index Error 201 or 202 .............................................................................................................................501
Index Error 204.........................................................................................................................................501
Index Error 215.........................................................................................................................................501
Index Error 218.........................................................................................................................................501
Index Error 258, 297 or 418.....................................................................................................................501
Index Error 260.........................................................................................................................................501
Index Error 272 Can be caused by:........................................................................................................501
Index Error 275.........................................................................................................................................502
Index Error 278 can be caused by..........................................................................................................502
Index Error 294.........................................................................................................................................502
Index Error 298.........................................................................................................................................502
Index Error 307 or 309 .............................................................................................................................502
Index Error 313 on Whole Body Scans .................................................................................................502
Error Code Index 318 during QC. X-ray Hi/lo Does Not Match Filter Drum Position.................502
Index Error Code 347 or Code 184 during Hip Scans ........................................................................503
Index Error 358.........................................................................................................................................504
Index Error 360 Motor Reversed during Stepping .............................................................................504
Index Error 365 or 367 during QC.........................................................................................................504
Index Error 368.........................................................................................................................................504
Index Error Code 309 or 371...................................................................................................................504
Index Error 373.........................................................................................................................................505
Index Error 389 Motor at Limit..............................................................................................................505
Index Error 413.........................................................................................................................................505
Index Error 472.........................................................................................................................................505
Index Error 548.........................................................................................................................................505
Index Error 557.........................................................................................................................................505
Index Error 585.........................................................................................................................................505
Index Error 587.........................................................................................................................................505
Index Error 599 No Motion Detected during Stepping......................................................................505
List of Figures
Figure 1: Horizon C/Ci System.......................................................................................................................... 1
Figure 2: Horizon W/Wi System ....................................................................................................................... 2
Figure 3: Horizon A System............................................................................................................................... 2
Figure 4: Q Scan Plot........................................................................................................................................... 5
Figure 5: Horizon System Block Diagram ....................................................................................................... 6
Figure 6: Horizon Operator Console Block Diagram..................................................................................... 7
Figure 7: Horizon C-arm Block Diagram......................................................................................................... 9
Figure 8: Horizon Distribution and Motor Control Block Diagram .......................................................... 10
Figure 9: PCIe Communication Controller Board/Distribution Board Interconnection Diagram ........ 19
Figure 10: Distribution Board High Level Interconnection Diagram.......................................................... 21
Figure 11: Distribution Board/Motor Controller Board Interconnection Diagram ................................... 22
Figure 12: Distribution Board/TZ Control Board Interconnection Diagram ............................................. 24
Figure 13: Control Panel Controller Interconnection Diagram .................................................................... 30
Figure 14: Distribution Board/C-arm Interface Board Interconnection Diagram...................................... 33
Figure 15: C-arm Interface Board High Level Interconnection Diagram.................................................... 35
Figure 16: C-arm Interface Board/X-ray Controller Assembly Interconnection Diagram........................ 36
Figure 17: C-arm Interface Board/ Detector Assembly Interconnection Diagram .................................... 40
Figure 18: Analog Digital Converter Board/Integrator Multiplexor Subsection
Interconnection Diagram................................................................................................................. 43
Figure 19: C-arm Interface Board/Analog/Digital Converter Board Interconnection Diagram .............. 45
Figure 20: Horizon A Space Requirements ..................................................................................................... 49
Figure 21: Horizon A Space Requirements for an External Console........................................................... 50
Figure 22: Horizon C, Ci Space Requirements................................................................................................ 51
Figure 23: Horizon W, Wi Space Requirements ............................................................................................. 52
Figure 24: Horizon A Dimensions and Clearances ........................................................................................ 53
Figure 25: Horizon C, Ci Dimensions and Clearances .................................................................................. 54
Figure 26: Horizon W, Wi Dimensions and Clearances ................................................................................ 55
Figure 27: Crated Unit (Horizon A).................................................................................................................. 56
Figure 28: Crated Unit (Horizon C, Ci, W, and Wi) ....................................................................................... 57
Figure 29: Table X Drive..................................................................................................................................... 59
Figure 30: Shipping Bracket Location (Horizon C, Ci, W, Wi) ..................................................................... 60
Figure 31: Shipping Bracket Location (Horizon A) ........................................................................................ 60
Figure 32: Table End Stop (Horizon A, W, Wi)............................................................................................... 61
Figure 33: Shipping Bracket............................................................................................................................... 62
Figure 34: Re-tighten Caster Bar Screws .......................................................................................................... 62
Figure 35: Off-load Ramps (as packed)............................................................................................................ 63
List of Tables
Table 1: Horizon Models ....................................................................................................................................1
Table 2: Horizon Models ....................................................................................................................................3
Table 3: Horizon Scan Options ..........................................................................................................................3
Table 4: Pentium PC Components ....................................................................................................................7
Table 5: System Circuit Board Function .........................................................................................................10
Table 6: Product Specifications ........................................................................................................................12
Table 7: System Operating Parameters...........................................................................................................13
Table 8: System Dimensions ............................................................................................................................13
Table 9: Exam Mode Performance ..................................................................................................................13
Table 10: Duty Cycle ...........................................................................................................................................15
Table 11: X-ray Beam Filtration .........................................................................................................................16
Table 12: Typical Measured Values for Half Value Layer in a Horizon System........................................16
Table 14: Maximum Deviation ..........................................................................................................................16
Table 15: Measurement Criteria for Technique Factors .................................................................................17
Table 16: PCIe Communications Controller Board/Distribution Board Interconnection Descriptions..18
Table 17: Distribution Board/Motor Controller Board Interconnection Descriptions...............................23
Table 18: Motor Controller Board/Stepper Motor and Position Encoder Interconnection Descriptions23
Table 19: TZ Drive Service Switches.................................................................................................................23
Table 20: Distribution Board/TZ Control Board Interconnection Descriptions .........................................25
Table 21: Motor Power Input .............................................................................................................................25
Table 22: Power Supply Cable DIN Rail Connections ...................................................................................26
Table 23: 24VDC Output Power Supply Connections, 1 to 6........................................................................26
Table 24: 120VAC Input to Power Supply Connections, 7 to 9.....................................................................26
Table 25: Right Position Potentiometer J9........................................................................................................26
Table 26: Left Position Potentiometer J7 ..........................................................................................................27
Table 27: Right Column Motor J5......................................................................................................................27
Table 28: Left Column Motor J3 ........................................................................................................................28
Table 29: Tz Drive Board LED Indicators ........................................................................................................28
Table 30: Jumpers ................................................................................................................................................29
Table 31: Distribution Board/Control Panel Controller Interconnection Descriptions .............................31
Table 32: Control Panel Controller/Control Panel Interconnection Descriptions ......................................31
Table 33: Distribution Board/C-arm Interface Board Interconnections.......................................................34
Table 34: C-arm Board Connectors ...................................................................................................................36
Table 35: C-arm Interface Board/ Detector Assembly
Interconnection Descriptions ............................................................................................................40
Table 36: Integrator/Multiplexor Board/Solid State Detector Boards Interconnection Descriptions ......42
1 Introduction
1.1 System Overview
1.1.1 Purpose
The Hologic Horizon™ X-ray Bone Densitometer estimates the bone mineral content (BMC) and bone
mineral density (BMD) of:
• selected areas of the body, or
• the entire skeleton.
It does so using X-rays of two different energy levels. This dual-energy scheme scans and estimates bone
by eliminating soft tissue within the selected area from the scan.
Table 1 Horizon Models
Model Figure
Horizon C/Ci Figure 1: Horizon C/Ci System on page 1
Horizon W/Wi Figure 2: Horizon W/Wi System on page 2
Horizon A Figure 3: Horizon A System on page 2
Note In this manual, Horizon refers to all models in the Horizon series. Information
that applies to only one model or group of models will be clearly identified.
Horizon W and Wi models have a larger table with greater extension than the Horizon C and Ci.
Horizon W and Wi Table moves:
• in and out
• left and right
Horizon W and Wi, Arm moves:
• left and right
Scan A W Wi C Ci
AP Lumbar Spine
Supine Lateral Spine
Hip
Dual Hip
SE Femur
Forearm
Whole Body
AP and Lateral Imaging
Single Energy
Dual Energy
Visualization of Abdominal
Aortic Calcification
1.1.2 Operation
Patient Positioning and Scanning
The patient lies face up on the table.
The operator:
1. Uses a cross-hair laser to position the scanning arm over the region of interest
2. Enters patient data
3. Selects the type and size of scan
4. Initiates the scan with a single keystroke.
Technique Factors
Technique factors, such as tube current and voltage, are preselected and fixed. Since testing is performed
by a fan beam method (in contrast to conventional radiography that floods the area), scanning time is a
function of the:
• dimensions of the area to be measured
• desired resolution
• desired precision
1.1.4 Shielding
In most cases, no additional shielding is necessary for patient, operator, or room. The system can be placed
in any convenient non-shielded examination room. Contact your state regulatory agency for details about
any additional shielding requirements.
1.2.3 Program
The program works in the following manner:
1. Load preliminary scan and obtain regions of interest from operator.
2. Estimate k as an average value of:
k = [Ltissue - Lair] / [Htissue - Hair]
where Ltissue indicates a low-energy measurement with tissue-equivalent material interposed by the filter
drum, and Lair, Htissue and Hair are similarly defined.
Note The subscript air designates the filter drum segment that is empty (i.e.,
contains neither bone, nor tissue-equivalent material).
3. Using this value of k, calculate Q for each point scanned using the formula given above (Q = L -
kH). This array of Q values constitutes a Q scan. Displays the Q scan.
4. Compile a histogram of the Q values.
Note Because a large portion of the scan contains soft tissue only, this histogram will
have a large peak.
Choose a threshold value just above this peak, and apply that value to
discriminate, point by point in the Q scan, between bone points (whose Q is
above the threshold) and non-bone points (whose Q is below the threshold).
5. Use the non-bone points to calculate a baseline value for each scan line. Using these points, form a
new histogram and repeat steps 4 and 5 until the results converge.
6. Smooth the segment boundaries to eliminate isolated noise-generated bone points.
7. Display the bone and non-bone points for operator approval.
8. Determine the constant of proportionality (d0) that relates the Q values to actual BMC (grams). This
constant is determined by measuring how much Q shifts when bone-equivalent material is
interposed by the filter drum.
9. Calculate the total bone mineral values by adding up the Q values for all bone points in each region
of interest (e.g., each vertebra), and multiplying by d0.
10. Determine the bone areas by counting the number of bone points in each region of interest.
11. Calculate bone mineral density as:
BMD = BMC / area
12. Display the calculated results and print the report.
• Console
The Horizon Operator’s Console consists of a Pentium PC on a convenient, roll-around computer cart
designed specifically for Horizon models. AC power from the scanner is fed to the Operator’s Console via
a power strip attached to the PC cart.
Table 4 Pentium PC Components
Component Function
Video Controller Chipset drives the monitor and the PCIe Communications Controller Board to
interface the PC to the Scanner
On-board Network Interface controls communications with the site’s network or a DICOM®
Card (NIC) network.
Mouse gives the operator easy control over the Windows 7-based software
Keyboard used for data input to the software
Color Inkjet Printer color printing
Optional Laser Printer substituted for the inkjet printer
The Distribution Board receives command inputs from the PCIe Communications Controller Board in the
PC and voltages from the DIN rail and Low Voltage Power Supply.
Commands and voltages dealing with the generation of X-rays or the acquisition of scan data are
transferred to the C-arm Interface Board, which controls both functions. The Distribution Board also
receives Table and C-arm motion commands from the PCIe Communications Controller Board in the PC.
Routing
When a motion command is received, the Distribution Board routes the command along with a board
address to the Motor Driver Boards.
The addressed Motor Driver Board converts these digital commands into analog signals that are strong
enough to drive the motors.
A Position Encoder attached to the idler pulley moved by the motor, senses changes in the position of the
arm or table. This device provides a voltage feedback to the Motor Driver Board that is translated into a
motor position, which in turn, is fed back to software in the Operator’s Console PC.
The C-arm Interface Board collects data from most of the circuits mentioned above for transmission back
to the Distribution Board and, finally, the software.
The Aperture Motor Assembly sends back position information indicating which of the aperture slots is in
position.
The Filter Drum Assembly sends back filter and reference phase information.
The XRC sends back information concerning fault conditions, beam conditions, and an AC Line signal for
generating AC Line Interrupts to the software.
C-arm Subsystem
The C-arm Subsystem controls the generation and detection of X-rays. It processes the commands received
from the Distribution Board and passes them to the X-ray Controller or the Detector Assembly. The X-ray
Controller, as its name implies, controls the generation of the X-rays by the X-ray Source Assembly (tank).
The X-ray Controller unit rather than individual components. The X-rays pass through the patient and are
sensed by the Detector Assembly.
Low Voltage Power Supply (+/- 15V) (+ 24V) JP6 & JP9
TY
JP5 TY Stepper Motor Driver
TY
TX
JP7 TX Stepper Motor Driver
AR TX
AR Stepper Motor Driver JP8
AR
AY
JP11 AY Stepper Motor Driver
AY
Detector Board
The Detector Board sends data collected from scans back to the software through the C-arm Interface
Board and the Distribution Board. Mathematical algorithms applied to this data produce the scans seen by
the operator and to calculate the BMD, BMC, and area of the bone matter scanned.
Unit Range
Peak Potential +/- 15%
Current: +/- 40%
Time: +/- 10%
Criteria Description
Peak Potential The voltage peak is measured with an oscilloscope. Voltage is a squarewave
pulse.
Peak is defined as the peak voltage of the 4 millisecond pulse, discounting any
initial overshoot.
Current Current is measured with an oscilloscope on the last millisecond of the
4 millisecond pulse.
Time Time of each pulse is measured with an oscilloscope and defined as the time
between 50% rise and fall times of the peak potential pulse.
Time of the scan is measured by counting the number of AC line pulses from
the start to the end. X-ray pulses are synchronous with the AC line.
2 Functional Description
This section provides a detailed functional description along with interconnection diagrams. Refer to
Section 2.3.2 Interface Connections on Page 20 for a block diagram and a brief functional description of each
block.
2.1 Computer
The computer is a Pentium 4-based (or higher) PC that comes equipped with a hard disk drive, a
keyboard, a mouse, a flat panel monitor, and a DVD R/W for archiving scans and backups of the system
database. The computer interfaces to the scanner via a PCIe Communications Controller Board.
PCI
Communication DISTRIBUTION
Command BOARD
Board
KP1085_004-0201
2.3.1 Power
The Distribution Board receives +24 and +/-15VDC from the Multi-voltage DC Switching Power Module.
The +24VDC is applied through four individual circuit breakers to the Table X, Table Y, Arm R, and Arm Y
Motor Controller Boards. The +24 and +/-15VDC are passed to the C-arm Interface Board. The +24VDC is
also reduced to +7 and +5VDC by regulators to power op-amps and analog switches located on this board.
The +7VDC is passed to the Control Panel Controller Board. The +5VDC powers the digital section of the
Distribution Board.
Note +7VDC may measure anywhere from +6.25VDC to +7.25VDC. This is true
everywhere +7VDC is shown in this manual.
Six (6) green LEDs indicate the status of the +24 (motors), +24 (control), +15, -15, +7 and +5VDC (ON
indicates the respective voltage is present). Five (5) red LEDs indicate the status of the five circuit breakers
applying voltage to the motor drivers/controller. ON indicates the circuit breaker has been tripped by an
over-current condition.
The Motor Controller Boards receive movement commands from the system computer via the
communications bus. Each Motor Controller Board contains an ID switch and four status indicators
(LEDs). The ID switch is a 16-position rotary encoded switch (SW1) that is read during system
initialization to determine the Motor Controller Board address for communicating with the computer. ID
switch settings for the Motor Controller Boards are as follows:
Table X drive 4
Table Y drive 5
C-arm Rotation 6
C-arm Y drive 7
The four red status LEDs provide visual indications of motor drive power on, Stepper CPU active, motor
drive direction, and motor step pulses.
The Motor Controller Board also has provision (JP2) for connecting limit switches to inhibit motor
operation when the mechanism goes beyond established mechanical limits. This feature is not used in the
system and, therefore, no cable is connected to JP2.
2.4.1 Power
Power input to the Motor Controller Board is +24VDC. This voltage provides the motor drive power and is
converted down to +5VDC for use by logic circuits on the board. The +5VDC is also converted to -5VDC.
Two green LEDs provide a visual indication of the power present on the Motor Controller Board.
Switch Function
Mode (Normal/Service) Determines whether the TZ Drive is in Normal or Service operation.
Left Up/Down When the TZ drive is in Service mode, moves the left pedestal up or
down.
Right Up/Down When the TZ drive is in Service mode, moves the right pedestal in the
direction specified by the Direction switch (not active in normal mode).
Reset Resets the board after manual operation. The TZ Control Board must
be reset after any manual operation.
2.5.2 Power
Power input to the TZ Control Board is + 24VDC from the standalone power supply to drive the pedestals
powers circuitry located on this board.
Logic power input to this Motor Controller Board is +24VDC from the distribution board. This voltage
provides the motor drive logic power and is converted down to +5VDC for use by logic circuits on the
board. The + 5VDC is converted to ± 5 VDC and ± 3VDC. The ± 5VDC powers logic circuitry on this board,
while the ± 3VDC provides the reference voltage for the position sensors. Two green LEDs provide visual
indication of the DC status (ON indicates the respective voltage is present).
Logic power input to this Motor Controller Board is +24VDC from the distribution board. This voltage
provides the motor drive logic power and is converted down to +5VDC for use by logic circuits on the
board. The + 5VDC is converted to +/- 5 VDC and +/� 3VDC. The +/�5VDC powers logic circuitry on
this board, while the +/� 3VDC provides the reference voltage for the position sensors. Two green LEDs
provide visual indication of the DC status (ON indicates the respective voltage is present).
-24VDC 24V return/reference for right and left pedestal motors JP2-2
-24VDC 24V return/reference for right and left pedestal motors JP2-3
+24VDC +24V Power for right and left pedestal motors JP2-4
+24VDC +24V Power for right and left pedestal motors JP2-5
+24VDC +24V Power for right and left pedestal motors JP2-6
Connector
Pin Signal Name Type Description
J9-1 POT_VREF Power Buffered output of 3 Volt reference to the high side of the
right column position potentiometer through a 100 ohm
series resistor. Referenced to AGND.
J9-2 AGND Power Shield connection point for right column position pot
cable.
J9-3 POT_IN Analog Input from the wiper connection of the right column
position potentiometer. Referenced to AGND.
J9- POT_RTN Power Return/reference for the right column position
potentiometer. Connected to AGND through a 100 ohm
resistor. Output to low side of the right column position
potentiometer.
Connector
Pin Signal Name Type Description
J7-1 POT_VREF Power Buffered output of 3 Volt reference to the high side of the
left column position potentiometer through a 100 ohm
series resistor. Referenced to AGND.
J7-2 AGND Power Shield connection point for the left column position pot
cable.
J7-3 POT_IN Analog Input from the wiper connection of the left column
position potentiometer. Referenced to AGND.
J7-4 POT_RTN Power Return/reference for the left column position
potentiometer. Connected to AGND through a 100 ohm
resistor. Output to low side of the left column position
potentiometer.
Connector
Pin Signal Name Type Description
J5-1 MOTOR_OUT_A Power Armature connection to right column motor.
0 – 24VDC during motoring. Referenced to
24VRTN.
J5-2 — — No Connection
J5-3 +5 VDC Power +5VDC power output to right column position
encoder. Not used for this application.
J5-4 S1 Input Active-high input from right column position
encoder Ch. A. Referenced to 0 VDC.
Not used for this application.
J5-5 S2 Input Active-high input from right column position
encoder Ch. B. Referenced to 0VDC.
Not used for this application.
J5-6 — — No Connection
J5-7 0 VDC Power Return/reference for +5VDC to right column
encoder. Not used for this application.
J5-8 MOTOR_OUT_B Power Armature connection to right column motor. 0 –
24VDC during motoring. Referenced to 24VRTN
Connector
Pin Signal Name Type Description
J3-1 MOTOR_OUT_A Power Armature connection to right column motor.
0 – 24VDC during motoring. Referenced to
24VRTN.
J3-2 — — No Connection
J3-3 +5 VDC Power +5VDC power output to right column position
encoder. Not used for this application.
J3-4 S1 Input Active-high input from right column position
encoder Ch. A. Referenced to 0VDC.
Not used for this application.
J5-5 S2 Input Active-high input from right column position
encoder Ch. B. Referenced to 0VDC.
Not used for this application.
J3-6 — — No Connection
J3-7 0 VDC Power Return/reference for +5VDC to right column
encoder. Not used for this application.
J3-8 MOTOR_OUT_B Power Armature connection to right column motor. 0 –
24VDC during motoring. Referenced to 24VRTN
Indicators
There are 12 LED indicators on the Tz drive board. Eight (8) of the LEDs are driven from a MCP23008 I/O
expander IC that is controlled by the MCU through an I2C serial connection. Other indicators are
connected between power and ground and indicate the presence of voltage at the power inputs to the
board or the outputs of on board regulators. Table 29 describes the function of each of the indicators.
Table 29 Tz Drive Board LED Indicators
LED1 PSON Enable output to 24V motor power Power supply enable output is not
supply is ON used for this application
LED6 MTR The MCU has detected a fault in the Overcurrent condition or no motion
FAULT operation of the left lifting column detected during attempted re-
LEFT motor position
LED7 MTR The MCU has detected a fault in the Overcurrent condition or no motion
FAULT operation of the right lifting column detected during attempted re-
RIGHT motor position
LED8 SYS TICK Slow blink (2/sec): MCU operating,
Normal mode
Fast blink (4/sec): MCU operating,
Service mode
LED9 24V 24VDC logic input power is present
at J1
LED10 5V +5VDC is present at output of
voltage regulator U14
LED11 3.3V +3.3VDC is present at output of
voltage regulator U17
LED12 MOTOR 24VDC motor input power is
PWR present at J2
Jumpers
There are three 2-position jumpers on the Tz drive board that function as described in Table 30.
Table 30 Jumpers
2.6.1 Power
The Control Panel Controller Board receives +7VDC from the Distribution Board. The +7 is reduced to
+5VDC to power the circuitry on this board and is applied to the Operator Control Panel to power the
LEDs. A green LED, on this board, provides a visual indication of the +5VDC power (ON indicates the
voltage is present).
Signal Description 1 2
CPC Pin CP Pin
Signal Description 1 2
CPC Pin CP Pin
SR0 Returns the state of the C-arm switch LEFT JP6-1 JP1-1
position when SW0 is active. Signals the
state of the TABLE switch IN position when
SW1 is active. Signals the state of the LOAD
switch when SW2 is active.
SR1 Returns the state of the C-arm switch JP6-2 JP1-2
RIGHT position when SW0 is active.
Signals the state of the TABLE switch OUT
position when SW1 is active. Signals the
state of the HOME switch when SW2 is
active.
SR2 Returns the state of the LASER and JP6-3 JP1-3
ENABLE switches when SW2 is active.
EMERGENCY_CPANEL Returns the state of the STOP switch. (Part JP6-23 JP1-23
HW_EMERGENCY_RET of the safety daisy chain.) JP6-24 JP1-24
+5V Provides power for the Control Panel LEDs. JP6-21 JP1-21
JP6-22 JP1-22
1. CPC = Control Panel Controller
2. CP=Control Panel
Note On the Horizon A model, the Data Acquisition System (DAS) consists of the
Analog to Digital Converter, Integrator/Multiplexor, and Solid State Detector
Boards. On the Horizon C and W models, a single-board, 128-Channel Detector
Assembly is used. On the Horizon Ci and Wi models, a single-board,
64-Channel Detector Assembly is used.
The C-arm Interface provides circuitry to:
• Control and monitor operation of the X-ray Controller Board. Four LEDs (two red, one green, and
one yellow) on this board, provide a visual indication of the status of the X-ray Controller and the
X-ray control circuitry of this board.
• Generate timing references to the DAS and PCIe Communication Command Board.
• Control power to the Positioning Laser.
• Command the Aperture stepper motor and monitor its mechanically linked position-sensing
device.
• Command the stepper motor of the inner and outer filter drums and monitor the encoded signals
returned from each drum of the device. Two green LEDs, on this board, provide a visual indication
of the Reference Drum operation.
• Generate +7VDC for the DAS system
2.7.2 Power
The C-arm Interface Board receives + 24 and +/-15VDC from the Distribution Board. The +/-15V is passed
through this board to the DAS. The + 24VDC is also regulated to +5 and +7VDC. The + 5VDC powers the
digital section of this board and the laser while the +7VDC is applied to the DAS.
Three green LEDs, on this board, provide a visual indication of the + 24VDC and +5VDC power (ON
indicates the respective voltage is present). The third LED monitors the +28VDC used on the QDR-4500
models.
Figure 15 shows the boards and assemblies that connect to the C-arm Interface Board
Figure 15 C-arm Interface Board High Level Interconnection Diagram
2.7.5 Interface
Figure 16 shows the interface connections between the C-arm Interface Board and the X-ray Controller
Assembly. Table 34 describes the interface signals and identifies the connectors and their pin assignments.
Note that the AC input power comes directly from the DIN Rail Power Module and connects to the X-ray
Controller Assembly.
C-arm Connectors
Table 34 details the C-arm board connectors. The operation of the communication and debug connectors
are described in the subsequent sections. Connector pin signal descriptions are shown only for those
connectors that add understanding to the design.
Table 34 C-arm Board Connectors
Note A thermal overload protector is installed in the source unit which will open
when the temperature limit is reached. The circuit resets automatically when
X-ray Source Assembly temperature returns to normal levels.
Switched signals charge integrating capacitors in this subsection during a given charging time. After the
charging time, the switched signals are turned off and the charges are held on the capacitors. Each
integrator is sampled, in photodiode order, by a multiplexor and the output sent to the Analog to Digital
subsection. At the end of sampling, all the integrating capacitors are discharged in parallel by shorting
them out with analog switches.
Each integrator has an additional input into which a test signal can be applied when there are no X-rays
present. This test signal is used to verify the operation of the integrators and multiplexors by the System
Diagnostics program on APEX V5.5 and higher, systems (D.A.S. Tab - Test Signal).
2.10.2 Power
The Solid State Detector subsection receives +/-15V from the Integrator/Multiplexor subsection. Voltage
regulators, located on this board, convert this voltage to +/- 12V to power the amplifiers and switches.
Note 1. Refer to schematic drawing 140-0048-SD for connector and pin assignments
(cannot scope these signals).
2. Refer to schematic drawing 140-0050-SD for connector and pin assignments
(cannot scope these signals).
3. Note that the CHANNEL hardware numbers are reversed from the software
numbers. e.g. Hardware
CHANNEL0 = software CHANNEL215.
2.11.1 Power
The Integrator/Multiplexor board receives +/-15V and +5V from the Analog to Digital board. The +/-15V is
passed through this board to the Solid State Detector boards. Voltage regulators, located on this board,
convert this voltage to +/-12V to power circuitry contained on this board. Analog and digital returns are
kept separate.
Note Jumper JP5, on the ADC board, must always be present on Horizon A models.
AC power from the Toroid Power Module is sent to the Scanner on a four-wire cable providing split 240
volts to the DIN Rail. From the DIN Rail, unfiltered AC is fed to the line filter and, from there, distributed
to various sections of the Scanner. Unfiltered AC is also fed to an AC outlet connector at the rear of the
Electronics Tray. This outlet provides AC power to the Operator’s Console power strip. A second
AC outlet provides unfiltered AC only while X-rays are being produced. This outlet is used in the factory
to give a visual indication of X-ray production during testing, and also for customers located in localities
where an external X-ray annunciator is required.
Table 40 Jumpers
3 System Installation
3.1 Preparing for an Installation
3.1.1 Site Considerations
For Hologic bone densitometer systems, the required installation procedures vary according to customer
site-specific existing conditions. For example, is the:
• new system the first and only bone densitometer on site?
• new system replacing another bone densitometer?
• new system being installed where one or more bone densitometer systems are already present?
To determine which procedures apply for any given site, refer to Hologic documents:
• CSD-0042-F19 Horizon Installation Checklist
The Operator's Console may be up to 45 ft (13.72 m) away from the Scanner, but since leakage and scatter
radiation levels are extremely low, it can be safely located in the same room with the Horizon unit.
When the X-ray is on, all personnel other than the patient should remain about 6 ft (2 m) away from the
scanner to avoid unnecessary exposure to scatter radiation. At the same time, never leave the system
unattended whenever a patient is on the table.
Short Doorway
If a Horizon system must be moved through a doorway that is not at least 81 in. (206 cm) high, the tabletop
can be removed. This allows the unit to fit through a doorway 79 in. (201 cm) high (inside dimension).
See Section 3.2.6 Remove Tabletop (If Necessary) on Page 58.
Narrow Hallway
If a Horizon unit must be moved through a hallway that is not at least 45 in. (114 cm) wide, the C-arm can
be removed. This allows the unit to fit through a hallway 29 in. (74 cm) wide (inside dimension). See the
instructions for removing the C-arm below.
The tabletop can be removed before taking the unit off the pallet. Follow the procedure below to remove
the tabletop:
1. Remove the right-side table rail end cover Phillips screws (2) located on the right end). This cover
slides straight out the side.
2. Remove the table locking bracket located on the left side of the scanner (Figure 33).
3. Remove the head end pedestal covers. You will need to move the table to its extreme forward
position to access the center screws on each panel.
4. Remove the two screws (Phillips) that secure the tabletop in place (located on top of the table near
the rear center) and slide the table to the left, far enough to access the Table X-bearing blocks.
5. Remove the right-side, front endcap — 3 Phillips screws (Figure 29).
6. Remove the back Phillips screw from the right-side cover (of the table X-drive assembly) and slide
the cover out from the front.
7. Unplug the cable to the Motor Controller and Control Panel boards.
8. Disconnect the stainless steel cable guard (2 screws) and remove the Table X-drive attachment
bracket (4 bolts and 2 nuts) (Figure 29).
9. Remove the left rail stop (1 counter sunk screw located next to the left-side, front endcap) to allow
the tabletop to be rolled off from the front.
10. Remove the 8 Allen screws (6mm) holding the table X-drive bearing blocks.
11. Remove the tabletop by carefully sliding it off the front of the unit (requires two people). Be careful
not to slide the bearing blocks off the rail.
12. Install the right angle bracket (found in the miscellaneous hardware kit) to hold the bearing blocks
and X-drive bracket in place while the scanner is moved (Figure 29). If this bracket is not available,
tape the bearing blocks, and X-drive bracket in place.
Figure 29 Table X Drive
7. Disconnect the ribbon cable from the Arm R Motor Controller board.
8. Disconnect the Y bearing blocks at the rear of the C-arm.
9. Remove the bracket that connects the C-arm carriage to the AY belt.
10. Lift the C-arm carriage up and out of the scanner unit.
11. Tape, or tie wrap, the bearing blocks in place to prevent them from sliding off the end of the rail.
Before the Horizon can be lifted to the vertical transport position, you must slide the table top to the right
and clamp it so it clears the floor when the unit is lifted to the vertical transport position.
To accomplish this, do the following:
1. Slide the tabletop to the right until it clears the table base panel.
2. Adjust the table end stop as pictured in Figure 32.
Figure 32 Table End Stop (Horizon A, W, Wi)
1. Remove the four shipping brackets (Figure 33) as follows: loosen the screws that hold the auxiliary
caster bars (located under each bracket), remove the lag bolts and remove the bracket. Then,
re-tighten the screws that hold the auxiliary caster bar to the frame.
Figure 33 Shipping Bracket
Caution The screws that hold the auxiliary caster bars must be re-tightened (Figure 34)
to prevent the unit from dropping when moved.
2. Locate the two off-load ramps supplied with the unit. Ramps are attached to the shipping skid base
(Figure 35).
3. Install the two off-load ramps. Ramps may be installed at the end of the skid (Figure 36), or the side
of the skid.
Figure 36 Ramps Installed at End of Skid
4. Move the unit down the ramp and away from the skid.
5. If hallways to the destination room are wide enough to move the unit in the horizontal position,
move the unit and everything that shipped with it to the destination room. Continue the install by
following the procedure starting at Step 2 of Set the System in Place on Page 65.
If the unit must be placed in the vertical position, continue with the procedure in Stand System on Foot End
to Move in a Narrow Environment (A, W, Wi).
Before the Horizon can be stood up to vertical transport position, you must slide the table top to the right
and clamp it so it clears the floor when the unit is stood up to vertical transport position.
To accomplish this, do the following:
1. Slide the tabletop to the right until it clears the table base panel.
2. Adjust the table end stop as pictured in Figure 32.
3. Attach the caster bars and cross brace located in the separate shipping box (Figure 37).
4. Disconnect and move the C-Arm assembly from the head end of the table to the foot end of the table
and re-attach the shipping bracket to secure the X-ray Source Assembly to the frame (Figure 33)
Warning: Cross-brace must be installed before tipping the unit on its end.
5. With two people, tip the unit on end, Figure 38 and Figure 39 (cardboard, or equivalent, can be used
to protect the floor).
Warning: The unit will feel heavier when putting it down than it felt when tipping it up.
This is because the pivot point is different. For safety purposes, two people must
be used to tilt the unit down.
2. Remove the caster bars and cross brace. Position the unit in its final location in the room.
3. Level the unit.
Caution This step is very important that the base of the system, the C-arm and table top
must be level to 0.1 degrees
3. Remove the counterweights (Figure 41) before installing the upper C-arm.
Note Models W and Wi have a box frame that must be installed to the C-arm
assembly. The C-arm is then attached to the box.
1. Carefully place the C-arm on the bottom bolts. Tilt the arm back slightly and slip the cables through
the openings and then tilt the arm forward until the front bolts are in place.
2. Put nuts and washers on all the bolts that attach to the scanner and tighten.
3. Install the cables.
4. Install T-beam end caps.
5. Using a digital level insure that the scanner is level (front to back and left to right).
Install Upper C-arm (A)
Follow the procedure below to install a Horizon A upper C-arm assembly:
6. Remove the four (4) 1/2-in. bolts that lock the upper C-arm in place (left in place during shipment).
7. Remove the screws for the C-arm shoulder cover (Figure 45) that were left in place during
shipment.
8. Remove the cover from the upper C-arm.
9. Install the upper C-arm onto the lower C-arm (tilt the front of the upper C-arm up to slide it in
place).
10. Install the four bolts that secure the upper C-arm to the lower C-arm.
11. Re-install the detector assembly cover with all counterweights.
12. Install the two large counter weights in the back of the upper C-arm and the A/D board assembly.
13. Connect the ribbon cable and laser power cables from the Lower C-arm to the upper C-arm, then
connect the cable from the A/D board assembly to the Detector assembly (Figure 46).
Figure 46 Upper C-arm Cables
Note Measure the potential voltage between ground and neutral at the wall outlet
that is being used to supply the system power. If the measured potential voltage
is greater than 250 mV (0.25 VAC) call Hologic Technical Support before
continuing.
Caution Ensure all toroid power cable wires are clamped onto the conductors and not the
insulation jacket.
8. Install the system cables (Section 3.3.3 Install Power Connection Jumpers with InRush Limiter ).
Figure 56 Main Power Wiring Configuration
Note These diagrams can also be found on the inside of the Toroid Assembly Cover.
Wire From To
Blue Circuit Breaker — Load Pin 1A — TB2
Brown TB2 Limiter Pin 7A — TB2
Brown Jumper Pin 3A — TB2 Pin 4A — TB2
Wire From To
Blue Circuit Breaker — Load Pin 1A — TB2
Brown Inrush Limiter Pin 6A — TB2
Brown Jumper Pin 3A — TB2 Pin 4A — TB2
Wire From To
Blue Circuit Breaker — Load Pin 1A — TB2
Brown Inrush Limiter Pin 7A — TB2
Brown Jumper Pin 3A — TB2 Pin 4A — TB2
Wire From To
Blue Circuit Breaker — Load Pin 1A — TB2
Brown Inrush Limiter Pin 5A — TB2
Brown Jumper Pin 2A — TB2 Pin 4A — TB2
Wire From To
Blue Circuit Breaker — Load Pin 1 — TB2
Brown Inrush Limiter Pin 7 — TB2
Blue Jumper Pin 1A — TB2 Pin 4A — TB2
Brown Jumper Pin 3A —TB2 Pin 7A — TB2
Wire From To
Blue Circuit Breaker — Load Pin 1A — TB2
Brown Inrush Limiter Pin 5A — TB2
Blue Jumper Pin 1A — TB2 Pin 4A — TB2
Brown Jumper Pin 2A — TB2 Pin 5A — TB2
For more detailed wiring diagrams, see ID-00123 (Horizon C/W/Ci/Wi) and ID-00128 (Horizon A).
8. Route the Operator Console Communications Cable through the cable clamp block on the rear of
the Scanner.
9. Route the Operator Console AC Supply Cable through the cable clamp block on the rear of the
Scanner.
10. Route the system input A/C Supply Cable through the cable clamp block on the rear of the Scanner.
11. Replace the Foot End Pedestal base frame cover.
12. Replace the Foot End Pedestal middle telescoping and upper covers.
RED 22-3
RED JP6-1
BLK JP6-6
17 10
GRY 21-B
GRY JP6-4
WHT JP6-3
BLU JP6-9
RED JP6-2
18 11
System Installation
19 12
20 13
21 14
22 15
23 16
24 17
25 18
24V -
26 19
24V +
15(A) -
15(B) -
15(A) +
15(B) +
27 20
28 21
Multi DC 29 22
30 23
Power Supply 31 24
32 25
33 26
PWR-00071 34 27
35 28
Horizon Service Manual
36 29
37 ASSY, SIGNAL DISTRIBUTION BOARD 30
030-3129 38 31
GND
NEUTRAL
HOT
39 32
40 33
41 34
42 35
ASSY, DIN RAIL 43 36
44 37
45 38
NTC 46 39
RED 22-3 47 40
3 4 48 41
YEL 22-1 49 42
1 2 50 43
R22 44
VIO 22-A1 VIO ASSY P/N 140-0086 45
A1 A2 RED +24 46
GRY 21B GRY RED +24 1 JP6 SCHEM P/N 140-0086-SD 47
VIO TB21 WHT +15 (A) 2 48
GRY 24 RET 3 To Din Rail 49
ORN VIO 22-A1 4 50
GRY TB20 BLK 15 RET 5
6 OPNL[1..26]
BRN BLK1 ORN 22-A1 7
3 4 BLU -15 (B) 8
BLK BLK2 9
1 2 JP4 26
R19 25 CBL-01736
VIO GRY YEL 22-1 1 JP9 24
A1 A2 2 To Table 23 CABLE, OPER PNL, EMI SHEILDED
GRN\YEL 3 To Din Rail 22
GB18 GRN\YEL 4 21
5 20
BLU 6 19
NB17 BLK3 18
17
BLK4 16
NB16 BLU 15
14
JP11 13
BLK TB15 BLK JP7 JP5 12
1 11
BRN BRN 2 1 1 10
TB14 BRN 3 2 2 9
BLK1 4 3 3 8
BRN 5 4 4 7
F13 8 AMP BLK2 6 5 5 6
7 6 6 5
BRN
BLK3 8 7 7 4
3 4 9 8 8 3
BRN BLK4 10 9 9 2
BLU 1 2 11 10 10 1
GRN\YEL R12 GRN/YEL 12 11 11
AUX_LIGHT_AC[1..3] 13 12 12
ORN 12-A1 GRY
A1 A2 14 13 13
15 14 14
GB11 16 15 15
17 16 16
BLU 18 17 17
NB10 BLU 19 18 18
20 19 19
BLK BLK 20 20
BLU FH9 8 AMP
GRN\YEL
CONSOLE_AC[1..3] AY Motion TX Motion TY Motion W AND Wi MODELS ONLY
GB8
CBL-01734 CABLE, TABLE YY DRIVE SHIELDED
TY[1..20]
NB7
BLU
BLU NB6
BLK BLK
FH5 8 AMP CBL-01664 CABLE, TABLE XX DRIVE SHIELDED
TX[1..20]
Horizon C/Ci and W/Wi DIN Rail AC Input/Output Wiring Diagram
BRN BRN
FH4 8 AMP
FH3 8 AMP
CBL-01846 CABLE, ARM AY DRIVE
AY[1..20]
BLK BLK
BLU H SUPRESSOR H
N N
G SSB G
BRN BRN
H SUPRESSOR H BLU
N N
G SSA G GRN/YEL
AC_IN[1..4]
PE GND STUD
RED 22-3
RED JP6-1
BLK JP6-6
17 10
GRY 21-B
GRY JP6-4
WHT JP6-3
BLU JP6-9
RED JP6-2
18 11
19 12
20 13
21 14
22 15
23 16
24 17
24V -
26 19
24V +
15(A) -
15(B) -
15(A) +
15(B) +
27 20
28 21
Multi DC 29 22
30 23
Power Supply 31 24
32 25
33 26
PWR-00071 34 27
35 28
36 29
37 ASSY, SIGNAL DISTRIBUTION BOARD 30
030-3129 38 31
GND
NEUTRAL
HOT
39 32
40 33
41 34
42 35
ASSY, DIN RAIL 43 36
44 37
45 38
NTC 46 39
RED 22-3 47 40
3 4 48 41
YEL 22-1 49 42
1 2 50 43
R22 44
VIO 22-A1 VIO ASSY P/N 140-0086 45
A1 A2 RED +24 46
GRY 21B GRY RED +24 1 JP6 SCHEM P/N 140-0086-SD 47
VIO TB21 WHT +15 (A) 2 48
GRY 24 RET 3 To Din Rail 49
ORN VIO 22-A1 4 50
GRY TB20 BLK 15 RET 5
6 OPNL[1..26]
BRN BLK1 ORN 22-A1 7
3 4 BLU -15 (B) 8
BLK BLK2 9
1 2 JP4 26
R19 25 CBL-01736
VIO GRY YEL 22-1 1 JP9 24
A1 A2 2 To Table 23 CABLE, OPER PNL, EMI SHEILDED
GRN\YEL 3 To Din Rail 22
GB18 GRN\YEL 4 21
5 20
BLU 6 19
NB17 BLK3 18
17
BLK4 16
NB16 BLU 15
14
JP11 13
BLK TB15 BLK JP7 JP5 12
1 11
BRN BRN 2 1 1 10
TB14 BRN 3 2 2 9
BLK1 4 3 3 8
BRN 5 4 4 7
F13 8 AMP BLK2 6 5 5 6
7 6 6 5
BRN
BLK3 8 7 7 4
3 4 9 8 8 3
BRN BLK4 10 9 9 2
BLU 1 2 11 10 10 1
GRN\YEL R12 GRN/YEL 12 11 11
AUX_LIGHT_AC[1..3] 13 12 12
ORN 12-A1 GRY
A1 A2 14 13 13
15 14 14
Horizon A DIN Rail AC Input/Output Wiring Diagram
GB11 16 15 15
17 16 16
BLU 18 17 17
NB10 BLU 19 18 18
20 19 19
BLK BLK 20 20
BLU FH9 8 AMP
GRN\YEL
CONSOLE_AC[1..3] AY Motion TX Motion TY Motion W AND Wi MODELS ONLY
GB8
CBL-01734 CABLE, TABLE YY DRIVE SHIELDED
TY[1..20]
NB7
BLU
BLU NB6
BLK BLK
FH5 8 AMP CBL-01664 CABLE, TABLE XX DRIVE SHIELDED
TX[1..20]
BRN BRN
FH4 8 AMP
FH3 8 AMP
CBL-01846 CABLE, ARM AY DRIVE
AY[1..20]
BLK BLK
BLU H SUPRESSOR H
N N
G SSB G
BRN BRN
H SUPRESSOR H BLU
N N
G SSA G GRN/YEL
AC_IN[1..4]
PE GND STUD
System Installation
83
Horizon Service Manual
Horizon Service Manual
System Installation
Attach the Communications Cable to JP10 on the Distribution Board and the ground strap to the
grounding stud.
The ComCons on the Horizon use a new style of connector on both the ComCon and Signal Distribution
Board ends of the cable. The cable plugs directly into the ComCon on the PC end and the Signal
Distribution Board in the scanner.
Note Horizon systems currently ship with distribution board PCB-01057, which has
a PCIe connector for the umbilical cable. Some older Horizon systems shipped
with distribution board 140-0086, which requires Umbilical Distribution Board
Adaptor PCB-00683.
A P-Clamp is used as a strain relief for the cable on the back of the PC.
Figure 65 Umbilical Distribution Board Adapter Orientation and Umbilical Cable Routing
Caution Do not misalign the Umbilical Distribution Board Adapter in the Signal
Distribution board slot as in the picture below. It can be inserted in the board off
by one set of pins in either direction. It will cause either non-operational system
or system that produces errors.
3.4.4 Hardware
Figure 70 Horizon Control Desk Components
3.4.5 Assembly
Base
1. From the top side, attach the Base (Figure 71) to the left side of the Frame with two (2) of the six (6)
1/4-20 X 0.50 Long, Pan PH Screws (Figure 72).
Figure 71 Base
Note Assemble the console from bottom to top. Hand-tighten all mounting hardware
to facilitate alignment.
2. Attach the Base to the right side of the Frame with two (2) of the six (6) 1/4-20 X 0.50 Long, Pan PH
Screws (Figure 73).
Monitor Shelf
1. Place the Monitor Shelf on the frame of the computer stand (Figure 75).
Figure 74 Monitor Shelf
2. From the bottom, attach two (2) 1/4-20 X 1.50 Long, Pan PH Screws in the rear of the Monitor Shelf
(Figure 75).
3. Use the two (2) 1/4-20 X 2.25 Long, Pan PH Screws (Bottom) to attach the front Shelf Support
through the Monitor Shelf (Figure 76).
Figure 76 Shelf Supports
Rear Brace
4. With the 1/4-20 X 0.50 Long, Pan PH Screws, attach the Rear Brace to the frame (Figure 77).
Printer Shelf
1. Attach the printer shelf supports to the frame with the two (2) black hex bolts (1/4-20 X 1.625 Long,
SHC Screw [Top]) (Figure 78).
Figure 78 Alignment Bolts
2. Place the Printer Shelf on the frame, aligning the holes over the hex bolts.
3. From the bottom, attach the shelf with four 1/4-20 X 1.50 Long, Pan PH Screws (Figure 79).
Attach Casters
1. Turn the control desk frame on its back.
2. Attach the fixed casters (MME-00459) to the rear of the stand (Figure 80).
3. Attach the locking casters (MME-00458) to the front (Figure 80).
Figure 80 Attach Casters
2. Attach the power strip to the console with the Scr, 4-40 X.375 in., Pan HD, SS and Split Lock
Washers.
Figure 84 Power Strip
3. Cover the screw holes with the plugs that were removed from the strip.
4. To the right and left of the power strip, attach the # 6-32 X .38 Long, Pan PH Screws and # 6 Internal
Tooth L’Washers to the rear brace (Figure 85).
5. Plug the power strip to the power cable from the Horizon system.
6. Place the PC on the console stand and attach cables.
7. Plug in the printer and computer to the power strip.
Figure 86 Plug in Printer and Monitor
1. Connect the signal cable from the umbilical cable to the computer communication card (Comm
Con).
2. Plug in the input power from the umbilical cable to the power strip on the control desk.
3. Attach the system power cable to the facility power at the wall.
4. Apply power to the system, place CB1 power switch in the up (ON) position.
5. Turn on the printer and install ink printer heads and or cartridges with paper.
6. Turn on computer and verify it boots normally and stops at the login screen.
7. Log in as User QDR and verify that the system turns on with the launch of the APEX application.
2. Exit the software by clicking the Exit button and selecting the Exit QDR without shut down?
option and then clicking the OK button.
Figure 91 Exit the Software
3. Insert the dongle (if necessary), click My Computer and then double-click on Default (C:).
4. In the Search Results window, double-click on the QDR folder. Double-click on the Utilities folder.
Locate the icon for the Service Utility and drag it to the desktop. Close the Search window and then
double-click the Service icon on the desktop. The software will now be restarted in service mode.
+/- .25”
“A” “A”
Figure 93 Pedestal
2. Loosen the four bolts fastening the top of the pedestal and the four (4) bolts fastening the bottom of
the pedestal.
Note Before adjusting the distance between the rails, make sure the upper and lower
brackets are securely fastened to the pedestal. Refer to Steps 3 - 5 of the previous
section (Aligning Table) and then go to the next step.
3. Adjust the distance between the T-rails and the table edge-to-rail gap and tighten the bolts.
4. Install the pedestal covers that were removed in Step 1.
Note If making an adjustment, make sure that the detector also stays angled at 0.1
degrees or less from head to foot on the detector cover, and that no twist is being
applied to the C-arm.
5. Rotate the C-arm until the center of gravity for the upper C-arm is over the lower C-arm
(approximately 60 degrees rotated).
6. Locate the six (6) 1/4 in. upper C-arm attachment bolts (three (3) on each side). Loosen only two on
each side so that the C-arm can pivot on the remaining screw. Move the C-arm until it is parallel to
the lower C-arm assembly using the digital level to measure the front-to-back and head-to-foot
angles.
7. Tighten all the bolts when done.
8. Adjust the table top in the A dimension (Figure 92). SeeTable Edge to T-Rail (A Dimension)
Adjustment on Page 104.
2. Turn off the Horizon system computer power and main circuit breaker on the rear of the foot end
pedestal.
3. Remove the Electronics Tray covers.
4. Loosen the two mounting bolts holding the belt tension block.
5. Adjust the tension nut so that the spring is compressed to 7/8 in.
6. Tighten the two mounting bolts holding the tension block.
Note Proper calibration of the motors depends upon other motor calibrations. You
must calibrate motors in the following order: TZ, AY, TY, TX, AR, and
Aperture. Proper motor calibration depends on all motors always being
calibrated in the order shown.
The motor controller board control and drive the table and C-arm movements. Calibrate motors in the
following order: AY, TY, TX.
Use Table 54 to determine the calibration procedures you need to perform for each Horizon model.
Perform the calibration procedures identified for each motor (√) in order from left to right.
Table 54 Calibration Procedures
Model TZ AY TY TX AR
A √ √ √ √ √
1
W NA √ √ √ NA1
Wi NA1 √ √ √ NA1
C NA1 √ NA1 √ NA1
Ci NA1 √ NA1 √ NA1
1. Not applicable to the system.
Note Perform the motor calibration in the order shown below. In the Tests section
click Calibrate. See the instructions for each motor.
Some motor calibration changes are not applied until the system is re-booted.
Caution A fully power cycle of the system is required after performing TZ motor
calibration and after performing aperture motor calibration.
Calibrate TZ
Check Calibration Values
1. From the Hologic Main Window, click Utilities.
2. Click Service Utilities.
3. Select System Diagnostics. Select the Tz Motor tab.
4. Place SW6 (Figure 1) on the Tz motor driver board in Service mode.
Use the up/down service switches to move the pedestals to the lowest position.
Caution Drive both pedestals at the same time. If the pedestals are not synchronized, the
board will stop functioning.
Caution The expected travel of the motors is 22.9 cm. Enter the actual measure travel
distance on the calibration screen.
Use a metric ruler to carefully measure the travel distance of the table. This is a
critical measure for proper system function.
Adjust the Encoders
1. Observe the encoder count on the diagnostic page for the right encoder.
2. Adjust the right string encoder until the position reads 325 +/- 2.
3. On the Tz PCB, disconnect the right string encoder connector (J-9).
4. Move the left string encoder connector from J-7 to J-9.
5. Observe the detector counts on the diagnostics display.
6. Adjust the left string encoder until the value is 325 ± 2 of the set value the right string encoder was
set to in Step 5.
7. Move the left string encoder connector back to J-7.
8. Reconnect the right string encoder to J-9.
9. Use the up/down service switches (SW2 and SW3) to move the pedestal to a position at least one 1
in. above the lowest position.
calibrate_position=10,1000,1000,819,50000,500,500,3494
pos_limit_position=182784
neg_limit_position=0
The four values for the set table calibration are, respectively:
• the left pedestal lower and upper encoder limits
• the right pedestal lower and upper encoder limits
the left lower and right lower values need to be 325 ± 2 counts.
Adjust the Encoder if it Exceeds Specifications
Caution Be sure to restart the software in Service Mode after calibrating the aperture, so
the software can access the new data file. If the system is not rebooted, the
software will not be able to position the pedestals correctly.
Turn off the Table and Restart
1. Exit the application with shutdown.
2. Cycle the power at the rear of the table on CB1 Off and then back On.
3. Restart the computer.
4. Log in as Field Service.
Calibrate AY (Horizon A, W)
Caution To ensure correct C-arm travel, ensure that the lower carriage frame and AR
Motor cover is in place before performing AY Calibration.
The program moves the C-arm and draws an encoder versus distance calibration grid. The positive
and negative linear fit parameters are displayed at the top left and top right of the plot in the form Y
= Intercept + Slope * X.
5. If the two slopes are within 0.3 of each other (Figure 6), click Continue.
The Table (AY Motor) Calibration dialog box displays the calibration parameter values.
6. Check that the parameters meet specifications. For more information, see MOTOR$AY Calibration
Parameters.
The last value on the calibrate_position line in the dialog box must be 3750 ± 5 ticks. To adjust this
value, refer to AY Calibration Parameters on Page 112.
7. Click Save.
Note Because the encoder is underneath the X-ray Source Assembly, adjusting it can
be counter-intuitive.
1. For example, if a value of 3770 appears (20 ticks above the correct calibration
position at the left physical stop), move the C-arm all the way to the right.
2. If the value on the right is 500, adjust it to 480 to reduce the corresponding
value on the left stop.
3. Hit Calibrate to move the C-arm to the left stop. The value should be
reduced by 20 ticks.
AY Calibration Parameters
Values are separated by commas within calibration fields.motor_direction determines whether the
positive step direction is the reverse of the direction of increasing encoder values. The value is 0 if the
direction is not reversed and 1 if the direction is reversed. This value is set from the corresponding
parameter in the MOTOR_XX.PRO file and is a constant for each motor. The direction of increasing
encoder values must agree with the system coordinates. The positive step direction is set by the wiring
harness for the stepper motor.
calibrate_position has eight values:
(1)Tolerance. The position tolerance for absolute moves in encoder ticks. The calibration program sets
this field to the value found in the corresponding calibrate_position field in the MOTOR_XX.PRO file.
(2,3)NumberOfSteps,StepDistance. The number of motors steps corresponding to the step distance in
µm. The ratio of these two numbers determines the step size (41187/2288 18 µm). The calibration
program sets these fields to the values found in the corresponding calibrate_position fields in the
MOTOR_XX.PRO file. These values are a property of the mechanical design of the system and should
never change.
Note Because the encoder is underneath the X-ray Source Assembly, adjustment can
seem counter-intuitive. When we perform calibration, pressing Esc as the
C-arm travels left should stop it at 3750 ± 5.
If, for example, the C-arm stops at 3770, 20 stops greater than it should be, we
need to send the C-arm all the way to the right to determine where it stops.
If the C-arm stops at 500 on the right, we need to reduce the travel by 20 to 480.
We then must recalibrate, press Esc as the C-arm travels left, and then check the
left stop once again to see if the C-arm has stopped at 3750 ± 5.
3. When TY reaches the left mechanical stop (Figure 101), click OK.
Figure 101 Click OK when TY Motor Reaches Left Mechanical Stop
4. When TY reaches the right mechanical stop (Figure 102), click OK.
Figure 102 Right Mechanical Stop
Calibrate TX
The TX motor centers the table and C-arm.
1. Select the TX Motor tab.
2. Click Calibrate.
3. A dialog box prompts moving the C-arm to 0° degrees (Figure 13).
Figure 104 Motor$TX Calibration
Caution The arm must be at 0°.
Note This message only applies to Horizon A systems. Use a digital level to set the C-
arm to 0° ± 0° for the A systems.
Ignore this message for other models.
4. Click OK.
5. When the TX motor reaches the outer mechanical limit, click Cancel (or press Esc) and verify that
the starting position is correct (1000 ±5).
6. Start Calibrate again and when TX reaches the outer mechanical limit (Figure 13), click OK.
Figure 106 Outer Mechanical Limit
The program moves the table and draws an encoder versus distance calibration grid. The positive
and negative linear fit parameters are displayed at the top left and top right of the plot in the form
Y = Intercept + Slope * X.
7. If the two slopes are within 0.3 of each other (Figure 14), click Continue.
Calibrate AR (Horizon A)
Confirm that the table is level front-to-back.
Caution Confirm that the table is level front-to-back before calibrating MOTOR$AR.
Unlevel tables cause positioning tolerance errors on whole body, lateral, and
IVA scans.
1. Remove the table pad.
2. Place the digital level on the table at the foot end. The digital level
3. value must be 0.0°.
4. Move the level to the head of the table.
5. Confirm that the level is still at 0.0°.
If the level results check is not 0.0 degrees, check that the:
a. system base frame, head-to-foot rails are at 0.0 degrees
b. C-arm cross table arms at the foot end on the left side measure 0.0 degrees
c. head end on the right side measures 0.0 degrees.
Continue the Calibration Procedure
The program moves the TY and AY motors to center positions, then moves the TZ motor to topmost
position.
1. Remove the C-arm X-ray Source Assembly cover.
b. b. Use the Table IN/OUT switch on the operator panel to move the C-arm until the level is
0.0° ± 0.1°
3. Click OK.
4. When the AR motor reaches the AP mechanical limit (Figure 109), click Cancel.
Figure 109 AR Motor AP Mechanical Limit
7. When the AR motor reaches the lateral mechanical limit, click OK.
8. Level the C-arm to 83° +0.1° (Figure 112).
a. Place the digital level on the top of the tank assembly.
b. Use the Table IN/OUT switch on the operator panel to move the C-arm until the level is 83° +0.1°.
c. Remove the level.
9. Click OK.
The program moves the C-arm and draws an encoder versus distance calibration
grid.
The positive and negative linear fit parameters are displayed at the top left and top
right of the plot in the form Y = Intercept + Slope * X.
10. If the two slopes are within 0.1 of each other, click Continue.
11. If the slopes are not within the specified range, see Adjust encoder versus distance calibration slopes on
Page 122.
The MOTOR$AR Calibration dialog displays calibration parameter values.
12. Check that the values meet specifications. For more information see Calibration Parameters for
MOTOR$AR on Page 121.
The MOTOR$AR Calibration dialog box displays the position value. The next-to-
last value on the calibrate_position line in the dialog box position value must be
250 ± 5 ticks.
13. To adjust the position value, see Adjust the Position Value on Page 122.
14. Click Save.
Figure 112 Level the C-arm to 83°
1. Place the tool X-ray Leakage Test Tool (TLS-02693) over the aperture assembly.
2. On the System Diagnostics screen, Detectors tab, turn on X-rays by clicking Xrays (F2) or by
pressing the <F2> key.
3. Slowly move the meter probe (Victoreen Model 450P or equivalent) over all accessible surfaces (see
table above for meter response times) to detect any leaks.
4. If a leak is detected (a spike in the reading), measure a point at a one meter (approx. 40 in.) distance
from the leak and ensure that the reading is 1.0 mrad/h (10 µGy/h) or less.
5. Click Xrays (F2) or press the <F2> key to turn off X-rays.
6. Remove the TLS-02693 Leakage Test Tool from the Aperture assembly.
3. Set the following parameters in the Device Control Screen (Figure 115 and Figure 116) to the values
shown:
• High Gain = 2
• Low Gain = 1
• X-Ray Mode = 3
• Aperture = 7
4. Verify that the screen shows Pulse equal to 1.
5. Select Filter Wheel to ON to ensure the filter drum is spinning.
Note To turn X-rays Off and On you can press the F2 Function key on the keyboard
or by using the mouse toggle the F2 button on the screen.
7. Insert the X-ray beam alignment tool, TLS-05376, into the slot on the C-arm (Figure 118).
Note The handle of the alignment tool protrudes from the C-arm to prevent
inadvertently leaving the tool in the C-arm after performing an alignment.
8. Press the keyboard F2 function key or with the mouse toggle the F2 button on the screen to turn X-
rays On.
9. Check the voltage output on the Device Control Screen (Figure 119).
Note Regardless of the initial voltage readings, the goal of the adjustment is to center
the filter drum and aperture assembly in the middle of the X-rays beam to
achieve maximum amplitude with a uniform flat signal across the top of the
graph.
If the arm is deflected in either direction, when the filter drum assembly is
properly aligned, the voltage drop should be the same in both directions.
On Horizon A systems use Emergency Motion to raise the table (Figure 6) to
access the area underneath the table top.
Caution Do not drive the table to its maximum raised position. Raising the table to its
maximum height at abnormal speed can damage hardware.
2. From the Hologic main menu, select Utilities > Service Utilities >System Diagnostics > Detectors
tab and select aperture 7 on Horizon C, W, and A models only.
Note Do not use the arrow up/down function to select Aperture 7. Instead, use the
mouse to highlight the displayed aperture number, type 7 and press Enter.
3. Use the alignment block TLS-02698, and alignment pin 099-0111 to align the hole in the filter drum
base plate to the hole in the aperture plate and to the aperture base plate assembly.
The pin should drop through easily. If the pin does not drop in, select the C-Arm
tab on the System Diagnostics screen.
4. In the Commands section at the lower right, type 10 or -10 in the Move Relative field to re-align the
aperture plate. You may need to visually check that you are moving the aperture plate in the correct
direction relative to the base plate hole.
The pin should now drop down through the hole in the back left of the aperture
plate down through the alignment block and into the base plate hole with
minimum force. If the pin still does not align with the base plate hole, continue to
move the aperture plate ten steps at a time until alignment is achieved.
5. Press or click F2 to turn on X-rays.
6. Locate the (2) 10-32 x .37L Hex should bolt screws on the bottom sled bracket of the Filter Drum
Sled assembly, and loosen them by 1/2 turn.
7. Adjust the rear sled adjustment to move the whole assembly left or right in the beam (figure 119).
The goal is to achieve the maximum voltage on the left side of the survey screen.
8. Press or click F2 to turn off x-rays.
9. Tighten the (2) 10-32 x .37L Hex bottom sled screws.
10. Perform skew alignment (See Perform Skew Filter Drum Base Alignment).
11. Repeat the lateral and skew process until beam is flat and at maximum amplitude.
12. Remove the pin and block when done.
2. Ensure that all components can freely adjust by turning the sled Skew adjustment screw. Loosen
filter drum base screws slightly if necessary (Figure 124).
Note Deflect the arm slightly to the left and right. The voltage should decrease when
the outside the beam path.
5. Adjust the Filter Drum Assembly Skew (Figure 126) until the amplitude displayed on the Device
Control Screen is flat.
Figure 126 Voltage Amplitude on Device Control Screen
If a detector at the end of the charts (e.g., Detector 44, Figure 127) has on an amplitude less than 6.0V and
less than the detector on the opposite side (e.g., Detector 171, Figure 128), the detector must be adjusted
front to back, so that both detectors display the same voltage.
Figure 127 Details Detector 44
8. Tighten screws.
Caution Make sure the pin and block have been removed before performing Aperture
Calibration.
1. From the Hologic main window, click Utilities > Service Utilities > System Diagnostics.
2. Click the C-Arm tab.
3. In the upper right corner of the screen, click Calibrate Aperture (Figure 131) to move the aperture to
its right and left physical limits.
Figure 131 Select Calibrate Aperture
The x-rays turn on as the aperture plate moves left to right. As the aperture moves through the beam path,
the signal is captured and displayed. The chart will identify the signal levels for each aperture specific
encoder positions.
The vertical lines on the final output should correspond to the center of the aperture openings. The data
contributes to software positioning and aperture identification.
Slope Value
4. Click Continue.
As the potentiometer moves in both directions, a linear change should be displayed in the Aperture
Calibration Plot (Figure 133).
5. Click Continue to open the Aperture Calibration Peaks screen (Figure 134)
Figure 134 Aperture Calibration Peaks
• The green lines should be centered at the peak of the signal for each aperture if the aperture is
moving correctly. This screen also gives an indication of whether the aperture plate is moving
erratically or binding.
• The red lines are typically displayed in pairs to reflect the pairing of the apertures. In between the
pairs, the beam should be off (maximum dark).
6. Click Continue to open the Aperture Calibration Data screen (Figure 135) and review the offset
numbers.
Figure 135 Aperture Calibration Data
Caution Be sure to restart the software in Service Mode after calibrating the aperture, so
the software can access the new data file. If the system is not rebooted, the
software will not be able to find a selected aperture correctly.
3.8.1 Principles
Radiation leakage from the HVPS/S results from a defect in the lead liner of the tank or its cover. If there is
a hole in the lead liner, the resulting leakage will produce a pencil beam of X-rays. If there is a crack in the
lead liner, the resulting leakage will produce a fan beam. In either case, the initial survey must be taken
close to and almost touching the tank with the survey meter probe. If a leak is found, the meter indicates a
spike in the reading. If this occurs, you must then move the probe one-meter (approximately 40 in.) along
the beam from that spot to determine if the leakage is within acceptable levels.
2. Start the Utility by selecting Utilities > Service Utilities > X-Ray Survey.
3. Change X-Ray Mode to 3 and Aperture to 7 to lock in the changes.
Warning: X-rays are generated during this procedure. Keep hands, head and other body
parts out of beam.
1. Position the survey meter one meter (approx. 40 in.) from the tank and move it all around the
outside of the instrument.
2. The reading must be 1.0 mrad/h (10 µGy/h)or less. Record the highest reading in the service report.
3. Press the <F2> key to turn off X-rays, remove the tool, and replace the tank cover.
Note Enter this information into the Radiation Measurement Report (CSD-0042-
F07).
Note Watch for leading edge of L4 in the scan image window. It should start between
scan line counts of 30 to 35 as indicated at the bottom left hand corner of the
screen.
The Phantom image should be in the center of the scan screen left to right.If the Phantom is not in the
center or starts before or after the line counts of 30 to 35;
7. Click Reposition.
The table and arm return to the start positions.
8. Using the vertical blue line as a guide (Figure 139), move the scan image left to right in the window
with your mouse until it is centered.
Figure 139 Repositioning
9. Place the starting edge of L4 in the scan image 5 clicks (use arrow keys on keyboard) below the
horizontal blue line.
10. Click Restart Scan to move the table and arm to new start position.
11. DO NOT move anything on the table.
12. Now adjust the Laser beam back to the laser alignment mark on the Phantom.
Caution Any adjustment of this potentiometer affects the QC highs and lows. You must
run detector flattening and recalibrate the machine after adjusting the A/D
gain. DO NOT ADJUST THE A/D GAIN UNLESS ABSOLUTELY
NECESSARY AS DEFINED in Step 9 in the section that follows (Verify and
Adjust the A/D Gain).
This procedure flattens the X-ray beam for each scan mode and uses the metal edges of the table to
calculate the center table position in the x-axis.
Note All covers, table mat, etc., normally in the X-ray path, MUST be on the Scanner
before running Beam Flattening.
1. Restart the Horizon software in service mode (if not already). Press the Center Table button and
turn on the laser.
2. Using a DXA Quality Control Phantom (Part Number ASY-01564), place the phantom on end with
the flattening target facing up. Orient the phantom so the metal edge is to the right (head end) and
tissue to the left (foot end) as viewed from the front of the table. Align the phantom flattening target
to the laser crosshair.
3. Start the X-Ray Survey Utility by selecting Utilities > Service Utilities > X-Ray Survey.
4. Set Pulse to 1, High gain to 2, Low gain to 1, X-ray Mode to 3 and Aperture to 7 (for A, C, and W
models).
5. For Display, click Graph.
6. Click X-Rays (F2) to turn X-Rays on.
7. Check that the phantom covers the entire beam. This is critical. If phantom does not cover the entire
beam, move it until it does. Keep the phantom as straight as possible.
8. Turn off X-rays with the <F2> key and press <Esc> or click Close to exit X-Ray Survey.
9. Reboot the PC, logon as Field Service, and restart the software in Service Mode to return to Horizon
Main Menu screen.
10. Select Utilities > Service Utilities > Detector Flattening.
11. Click Flattening –Basic First.
Figure 142 Flattening –Basic First
7. In the Patient window (Figure 145), enter Flat/Lat Verification in Last Name.
8. Click OK.
Figure 145 Patient Window
9. Enter a height and weight for the patient in the Patient Confirmation Window (Figure 146).
10. Click OK.
11. At the Select Scan Type screen, select AP/Lateral (Figure 147).
12. Deselect Use Default Scan Mode.
Figure 147 Select AP/Lateral
17. At the Select Lateral Scan Mode for AP/Lateral Exam screen, select Fast Array (f).
18. Click Next.
Figure 149 Select Fast Array (f)
19. Follow the instructions (Figure 150) regarding patient (spine phantom) alignment with the laser.
20. Click Continue.
Figure 150 Alignment Warning
21. Place two sheets of paper against the left and right edges of the phantom (Figure 151).
Figure 151 Paper against Right Edges of Phantom
25. After the scan has been completed, on the Analyze Setup screen (Figure 153), select Choose
Analysis Method.
26. Select Lumbar Spine from the pull-down.
27. Click Next.
Note If Lumbar Spine is not available from the drop down menu, you must load base
analysis protocols from the application disk first.
32. If image adjustment is necessary to visualize the bone, select the SUN symbol to
open the Image Toolbox (Figure 156).
33. When done, click the Sun symbol to switch back to the Global ROI Toolbox.
Figure 156 Global ROI Toolbox
35. Select the top line of the Global ROI and move it to the vertebral space that separates L1 and L2
(Figure 157)
Figure 157 Move Top Line of Global ROI to Vertebral Space
39. Verify that the angle displayed under the image (near the bottom of the screen) is less than ± 2
degrees (Figure 162).
Figure 162 Bone Map Start Position
40. Using the sheets of paper as a guide, carefully rotate the phantom on its back edge towards the C-
arm so that the bone segments are facing the rear of the table as shown (Figure 163).
Except for the rotation, do not move the phantom in any other direction.
45. From the Exit Exam screen, select Analyze Scan (Figure 165).
46. From the Analysis Setup screen, select Choose Analysis Method.
47. Select Supine Lateral from the pull-down menu (Figure 166).
Figure 166 Supine Lateral Analysis Method
50. Select Line Mode. Select the bottom line of the Global ROI and make the sides vertical (Figure 168).
Figure 168 Line Mode
51. Verify the left edge of the ROI is at the left edge of the image.
52. Select Whole Mode (Figure 169).
53. Using the right arrow key, count the number of steps required to align the left edge of the Vertebral
Boundaries ROI (the middle vertical line in Figure 169) to the left edge of spine.
54. Record the number of steps taken here: ______________.
55. The steps will be less than 6 when properly calibrated.
56. If the number of steps taken was 6 or less, click Cancel on the scan analysis screen, click Yes and Exit
to confirm analysis exit and proceed to Step 66.
57. If the number of steps exceeds 6 steps, you must adjust the table height.
58. Click Cancel on the scan analysis screen,
59. Click Yes and
60. Exit to confirm analysis exit and proceed to Step 61.
61. Click Utilities > Service Utilities > Table Height.
62. Refer to Figure 170.
a. Subtract the actual number of steps recorded in Step 53 from the number 10.
b. Enter the difference into the text box as a positive number to lower the table (Example for this
image: +4).
c. Record value entered into the text box ______________. Click Next when done.
63. From the APEX or QDR main screen, select Exit, Exit QDR without shut down and click OK.
64. Re-boot the system and login as user Field Service. When the APEX or QDR main screen displays,
select Exit, Exit QDR without shut down and click OK.
65. Start APEX or QDR software in Service Mode.
66. Detector Flattening (Big Flat) must now be run since the table height has been changed. Click
Utilities > Service Utilities > Detector Flattening and follow the on-screen instructions.
Note The Table Height Utility parameter limits are ±5 to raise or lower the table;
however, actual limits may be less for any given system. If the limits are
exceeded, the Lateral Position setup cannot function and a Table Height Out
of Limits or similar error message will display. If this is the case, enter
countering numbers until values are within limits (indicated by no error
message).
For Example: you entered +5 and the Table Height Out of Limits or similar
error message displayed. To counter, enter -2. This will result in you having
input the value +3. Any time you run the Table Height utility, you must always
re-run Detector Flattening (Big Flat).
67. Select Perform Exam and select patient Flat/Lat Verification in the last name field and
68. Click OK.
69. Click OK.
70. Repeat Steps 14 through 54.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam
system or if the customer has multiple systems, then a cross calibration
must first be performed to ensure that any longitudinal studies begun on
the QDR being removed can safely be continued on the new Horizon. Refer
to the cross-calibration procedure in Section 3.12 Fan Beam Cross-
Calibration Gold Standard System on Page 167 before uninstalling the
existing QDR. The Cross-Calibration procedure applies to fan beam
systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
3. Compare values displayed with actual values printed on the phantom. If there is a discrepancy,
check to make sure you have the phantom that shipped with the unit. Check with Hologic Technical
Support if needed.
3.11.3 Perform BCA Calibration Scans and Table Top Radiographic Uniformity (A, W
and Wi)
This procedure verifies proper X-Ray beam alignment and the uniformity of the X-Ray transparency of
entire tabletop in a manual process (for Horizon A, W and Wi models). BCA Calibration Scans must be
performed on the system prior to verifying the uniformity test scan.
Note On Models A, W, and Wi, after Flattening is performed and 30 minute cool
down is expired, perform a Daily QC scan. Automatically the system will
perform BCA calibration scans on the QC Phantom. After the calibration scans
are complete the system will inform you to remove the QC Phantom and clear
the table to perform the Table Top Uniformity Scan test. This test will report the
results automatically and must pass with the Hi and Low Air SD of less than
2.0.
If the system is passing normally, skip to next step.
If the system is not passing with less than SD 2.0, verify that table alignments
and beam alignments are correct, adjust as necessary and run Flattening again.
1. If necessary, restart the software in Service Mode.
2. Type WBAIRQC in the Patient Biography. Enter the serial number of the unit in the Patient ID field.
Make sure that only the pad is on the table. Clear the table of anything else.
3. Perform one Whole Body air scan (a scan with nothing on the table). Do NOT interrupt the scan for
any reason. When the scan completes, select Exit Exam. Do not analyze the scan that was just
performed.
4. From the Main Menu screen, select Utilities > Service Utilities > Table Top Radiographic
Uniformity.
5. Select the Whole Body scan and click the Next>> button.
6. Table Top Radiographic Uniformity results tabs are now displayed. Select the Low Air Tab.
7. If the Global Stats S.D. (2) is less than 2.0, the instrument is properly aligned. If the Global Stats S.D.
is greater than or equal to 2.0, then the machine is not aligned properly. Check the following:
• If the image appears “streaky” or “banded”, check for loose wires or other debris between the table
and the tank.
• Check the Aperture Calibration.
• Check AC Line voltage.
• Check +/- 15 V Power Supply.
• Check for a dedicated line.
• Check Detector Flattening.
• Re-run daily QC (detailed in MAN-03644 Horizon QDR Series User Guide).
Print a copy of the test results and include it with the other paperwork being returned to Hologic.
After realignment, repeat the Table Top Radiographic Uniformity test. If the global S.D. is less than 2.0,
you are done. If the global S.D. if equal or greater than 2.0, print a copy of the test results and include it
with the other paperwork being returned to Hologic.
If the global is equal to or greater than 2.0, the non-uniformity may be attributed to the table itself. If so,
continue with this procedure. Make sure that the machine has body composition loaded. If it does, then
follow the appropriate procedure in Section 3.11.3 Perform BCA Calibration Scans and Table Top Radiographic
Uniformity (A, W and Wi) on Page 158.
4. Be sure the spine phantom has been entered in the patient database in the form of “Spine Phantom
#nnnn” (where nnnn is the number of the phantom) so that the QC plot program can identify it
properly.
5. From the Main Menu screen, select Perform Exam, select the Spine Phantom, and click the OK
button at Patient Information dialog box.
6. Select AP Lumbar Spine, uncheck default scan check box, and click the Next>> button.
7. Select Array, click the Next>> button, set the scan length to 6 in., and start the scan.
8. Analyze the scan and click the Close button to exit the Analysis screen.
9. Start the Field Service Calibration program by selecting Utilities > Service Utilities > Field Service
Calibration from the Main Menu screen.
10. Select the Spine Highs/Lows for Array Spine phantom option to automatically calculate the new
thickness indicators for array spine.
11. Select the Spine Phantom scanned in step 7 from the list of Analyzed Scans and click the Next>>
button. Click the Continue button on the Successfully Calculated message box.
12. Click the Temp Write button and click the Continue button on the Successfully Wrote message box.
Exit Field Service Calibration.
3. Center the table, place the phantom on the table and align with the laser. The sides of the phantom
should be parallel to the table. Use a ruler if necessary.
4. Click Perform Exam from the application main screen.
5. At the Select a Patient for this Exam window, select the phantom, and click OK.
6. At the Patient Confirmation screen, click OK.
7. At Select Scan Type screen, select AP/Lateral, uncheck Use Default Scan Mode, and click Next>>.
8. At the Select AP Lumbar Spine Scan Mode for AP/Lateral Exam screen, select Array (a), and click
Next>>.
9. At the Select Lateral Scan Mode for AP/Lateral Exam screen, select Fast Array (f), and click Next.
10. Recheck phantom alignment with laser.
Note Do not move the arm or table, otherwise the test will have to be started over.
11. Place two sheets of paper against the left and right edges of the phantom (Figure 151) When done,
click Continue.
12. On the Scan Parameters screen, set the scan length to 6.0 in., click Enter and click Start Scan.
13. When scan has finished, select Spine as the analysis method, and click Next>>.
14. Analyze the scan and click Start Position.
15. Refer to Figure 172 Verify that the angle displayed under the image (near the bottom of the screen)
is less than ± 2 degrees.
16. Using the sheets of paper as a guide, carefully rotate the phantom on its back edge towards the
C-arm so that the bone segments are facing the rear of the table as shown. Except for the rotation,
do not move the phantom in any other direction.
17. Click Close. When the Continuously Press Enable message appears, press and hold the Lateral
Enable switch until the C-Arm and table move to the Lateral position.
18. On the Scan Parameters screen, click Start Scan.
19. When the “Continuously Press Enable” message appears, press and hold the Lateral Enable switch
until the C-Arm and table move back to the AP position.
20. From the Exit Exam screen, select Analyze Scan.
21. Analyze the scan (refer to user guide for Lateral analysis) and close the screen.
22. Click Exit to close the Exit Analysis screen.
23. Restart APEX software in service mode (if not already). From the APEX Main window, select
Utilities > Service Utilities > Field Service Calibration.
24. Select Highs/Lows for Array Spine Phantom.
25. Select the AP scan just performed and click Next.
26. Click Continue on the Calculated highs and lows page.
27. Select Highs/Lows for Array Lateral Phantom.
28. Click Continue on the Calculated highs and lows page.
29. Click Temp Write to write temporary results and click Continue to exit the write dialog box.
Note For Horizon system with version APEX 5.6 and higher, only 10 reproducibility
scans for calibrations are required. Earlier versions required 20 scans. Acquire 5
scans with a period of 15 between groups.
14. Analyze the first scan manually selecting Lumbar Spine for the analysis method.
15. Note the scan #.
16. Click the Close button to exit analysis.
17. From the Main Menu screen, select Utilities > Service Utilities > Auto Analyze.
18. Click unanalyzed Scans and select the 10 scans, click Compare radio button and then click Next.
19. Now, select the QC Scan performed in Perform QC Scan on Page 186.
20. Click Next.
21. The 10 scans will now be compared with the auto-analyzed QC Scan.
3.11.7 Calibration (Area, BMD, and BMC) with ACF and BCF Correction Factors
Calibration for Area, BMD, and BMC is accomplished in three stages:
• Array Scan Thickness Measurement & Calibration
• Calibration of Area and BMC for Array Scan Modes
• Adding Array AP scans to the QC database
Follow the procedure, in order, and exactly as shown, for each stage of the calibration.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam system
or if the customer has multiple systems, then a cross-calibration must be
performed to ensure that any longitudinal studies begun on the QDR being
removed can safely be continued on the new Horizon. Refer to the cross-
calibration procedure in Section 3.12 Fan Beam Cross-Calibration Gold
Standard System on Page 167 before de-installing the existing QDR. The Cross
Calibration procedure applies to fan beam systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
1. From the Main Menu screen, select Utilities > Service Utilities > Field Service Calibration.
2. Click the ACF/BCF button.
3. Select the 11 reproducibility scans performed in previous step.
4. Click Next
5. Click Continue at the Successfully Calculated dialog box.
6. At the Field Service Calibration dialog box, click the Write button to write the values to arrc.txt file.
7. Click Continue to return to the Main Menu.
Using a test patient, perform at least one scan in each of the scan modes to verify that machine is fully
functional in all modes of operation. Verify that there are no electrical or mechanical problems during each
scan mode.
[ ] Perform an AP Spine Array, Fast Array, and Express scan.
[ ] Perform a Left Hip Array, Fast Array, and Express scan.
[ ] Perform a Whole Body scan (A, W, and Wi only).
[ ] Perform a Forearm scan.
[ ] Perform the IVA Scan QC Procedure described below.
20. Count the 12 steps associated with the Step Wedge Phantom.
The number of steps visible on the Step Phantom should be 11 and 1 more for the table.
21. Skip this step on models Ci and Wi. By clicking the + sign to increase the magnification and using
the scroll bar to examine the x-ray test pattern image on the screen, you should be able to see three
lines for the 0.8 test pattern image.
22. When completed, click Cancel.
A radiation dose measurement must be performed at installation time, and done routinely whenever the
machine is serviced. This test requires the use of a Victoreen Model 450P, Unfors, or equivalent measuring
device set to measure dose over time (mR/Hr).
Note The dose measurements must be done with the machine fully assembled and the
table mat in place.
1. If using a Victoreen, place the meter in the center of the table approximately ½ in. to the right of the
laser crosshair. If using an Unfors, place the detector on the crosshair and raised 2.0 in. above the
table surface.
2. Click the Perform Exam button, create a test patient biography, and select the Array spine scan.
3. Set scan length for 9.5 in. and click the Start scan button.
4. The meter is now scanned.
5. Observe the scan to verify that the entire chamber of the meter is included in the scan.
6. Record the result from the meter.
Note The arrayExplorer spine scan dose should be less than 20mrad (200µGy).
A radiation scatter measurement must be performed at the time of installation. You must use a survey-
type radiation meter (Victoreen 450P or equivalent).
Perform the following:
1. Center the table and C-arm.
2. Place the spine phantom in the center of the table and align using the laser.
3. Start an AP Spine scan.
4. Using the <Esc> key, stop the scan when it is between the two middle vertebrae.
5. Start the X-ray Survey Utility by selecting Utilities > Service Utilities > X-ray Survey.
6. Using the survey meter, measure the scatter radiation at a distance of 2 meters from the phantom on
a horizontal plane all around the unit. Move the meter slowly.
Note If you cannot measure the scatter radiation at a distance of 2 meters because of
space restrictions, you can measure at 1 meter and divide the result by 4. This
reading must be less than 1.0 mrad/h (10 µGy/h) at 2 meters (approx. 80 in.) in
array AP scan mode.
7. Record the highest reading in the service report and then turn off X-rays by clicking on X-ray (F2).
Note Enter this information into the Radiation Measurement Report (Section 0.1.10
Radiation Measurement Report on Page 5).
3.18 Perform QC
Once the machine has been fully assembled and calibrated, at least one QC scan should be performed.
• Perform the daily QC procedure and enter the scan into the QC database.
Caution The new system must be properly calibrated before performing cross-calibration
Warning: Verify the selection with the site supervisor before proceeding with this
procedure.
If the Gold Standard System cannot readily be identified, use the system that is
producing the most consistent scans as the Gold Standard.
Cross-Calibration Calculator
This document contains sample Cross-Calibration Calculator output (Figure 213 on Page 196). You must
record the values obtained during this procedure in worksheet TLS-05948.
• Include a printed copy of the completed worksheet with your install or service call paperwork, or
• send the worksheet as an email attachment.
Refer to TLS-05948 for the Excel Cross-Calibration Calculator for worksheet details.
Warning: If data from the old system cannot be recovered, contact Technical Service to
arrange for systematic recovery of data from the unit that will be replaced.
3. In the Scan Information popup, select each of the scans to be removed from the plot (Figure 177).
Figure 177 Removing Outliers from the Scan Information Popup Window
Note Be sure to select each scan in the popup that you intend to remove. Outliers that
are overlooked may skew results.
Scan Averaging
5. On the Scan Averaging report, confirm that the Coefficient of Variation (CV) % for the selected
scans are:
BMD < 0.6% Area < 0.8%
BMC < 0.8%
6. If the CV values exceed these percentages, locate and eliminate any outlier scans from the QC plots.
Caution If your CV percentages for the scans exceed specifications, check for an outlier
(Red X at the top or bottom of the plot) that might have been overlooked. A
missed outlier will skew values and produce unacceptable CV percentages.
8. Print each of the nine (9) tabs for all of the plot results.
9. Fax/email the plots and values to Hologic Skeletal Health Technical Support, BED-tsppt@hologic.com,
before proceeding.
10. Technical Support will tell you which values to record.
In some cases, you may be required to perform 20 scans of the spine phantom to obtain the values.
11. Record the QC plot values obtained/provided by Technical Support in Section 2 on the
Cross-Calibration Calculator Spreadsheet (Figure 179).
Result Average % CV
BMD 1.053 0.41%
BMC 58.05 0.36%
Area 55.12 0.32%
12. Print this report.
Caution The patient must be called Phantom #XXXXX, where #XXXXX equals the
serial number from the phantom.
Do Not use Spine Phantom #XXXXX.
2. Select Utilities > Service Utilities > AutoScan.
3. Left-click on Select a Patient.
Figure 181 Select a Patient
Note Performing these scans as the customer would use the system produces results
that are far more accurate for this calibration procedure.
Caution Record the values for the first scan to identify a start point.
AutoScan the gold standard phantom 5 scans at a time.
Allow 20 min between the groups of 5 scans for the HVPS X-ray tube anode
to cool down. This process replicates customer use of the system and minimizes
deviation in the scan results.
14. Perform a total of 20 scans on the Gold Standard spine phantom.
Analyze the First Scan
Lumbar Spine Analysis
1. From the QDR/APEX desktop, click Analyze Scan and manually analyze the first scan acquired.
2. Choose Lumbar Spine from the Analysis Method drop down.
12. Compare the values given to Figure 185 Conditions Range Diagram to determine your next step.
Figure 185 Conditions Range Diagram
5. Record the values for Area and BMC (from Section 2) on the worksheet in the area labeled
Section 6 (Figure 190).
6. Click Continue.
7. Click Yes.
Figure 190 Cross-Calibration Calculator Section 6
3. Click OK.
4. Click OK in Area Within Limits (Figure 192).
Figure 192 Area within Limits
6. Click Exit.
7. In Windows Explorer, browse to C:\QDR\Data folder.
Figure 194 Arrc File Location
4. Select the 20 gold standard phantom scans to re-calculate the correction factors.
5. Click Write (Figure 197) to finalize the update in arrc.txt.
6. Click Exit.
Figure 198 Successful Calculation of ACF and BCF
7. Click Continue.
8. Click Continue.
9. Select Exit the QDR application with shutdown.
10. Reboot the PC.
Apply New Correction Factors to Gold Standard Spine Scans
1. Log on as Field Service.
2. Re-launch the application in Service Mode.
3. Select Utilities > Service Utilities > Auto Analyze.
4. Select the 20 gold standard phantom scans on the All Scans tab for patient Phantom # XXXXX.
5. Click RECALYZE.
6. Click Next to apply correction factors to scans.
Scan Averaging
1. Go to Utilities > Service Utilities > Scan Averaging.
2. Select the 20 scans (Figure 200) for the patient Phantom # XXXXX just recalyzed.
3. Click Next.
Figure 201 Scan Averaging Results
Figure 201 shows one scan (second from the last on the list) that is out of range. The CV (%) is out of range
as a result.
4. Exclude this scan from the list.
Note These values should match the BMD, BMC, and Area in Section B of the TLS
spreadsheet. If these numbers DO NOT match, repeat steps 8a through 8i one
time. If they still do not match contact Technical Support.
5. Click Print.
6. Click OK.
7. Validate that the New System is calibrated with the Gold Standard Spine Phantom.
8. Recalibrate New System With Gold Standard Spine Phantom.
a. Use 20 good scans on the Gold Standard Spine Phantom.
b. Go to Utilities > Service Utilities > Field Service Calibration.
c. Click the ACF/BCF button to re-calculate the correction factors.
d. Click the Write button to finalize the update in arrc.txt and then click Exit.
e. Exit the QDR application with shutdown and then reboot the PC.
f. Log on as Field Service and re-launch the application in Service Mode.
g. Select Utilities > Service Utilities > Auto Analyze click the RECALYZE radio button and the 20
scans on the All Scans Tab for patient Phantom # XXXXX.
h. Go to Utilities > Service Utilities > Scan Averaging, select the 20 scans for patient Phantom #
XXXXX just recalyzed and print the results.
i. Record the averages for BMD, BMC, and Area in Figure G on the worksheet.
Figure 203 shows one scan (second from the last on the list) that is out of range. The CV (%) is out of range
as a result.
11. Exclude this scan from the list.
Note The BMD, BMC, and Area average numbers are the numbers that will be used
to re-label the new phantom. They must be from a tight grouping of scans
without outliers.
Note When requesting labels, be sure to provide the determined values to the Parts
Group, so that the values can be added to the label.
2. Re-label the new Spine Phantom with the averages of BMD, BMC and AREA from Step 9 (Section
8, Figure 204, on the worksheet).
3. On the new spine phantom label, indicate that the new values resulted from a cross-calibration by
writing the following:
Cross-calibrated to S/N XXXXX using Spine Phantom #YYYYY
Service Log
4. In Utilities > Service Utilities > Service Log, record the following information (Figure 205):
• This system was cross-calibrated with phantom # xxxxx from DXA S/N XXXXX to BMD XX.XX,
BMC XX.XX, and AREA X.XXX values and the Month/Day/Year on which the cross-calibration
was performed.
• The recalibration results of the spine phantom # XXXXX are now: BMD xx.xx, BMC xx.xx, and
Area x.xxx Phantom re-labeled Month/Day/Year.
Figure 205 Utilities Service Log
Note The new system QC Phantom should be the only phantom in the list.
4. Select the 21 scans (or fewer if any were excluded) for the Spine Phantom performed in step Step 9
on page 187, (Section 8Section 8 on the worksheet).
Figure 207 Scan Selection
5. Click Finish.
6. Go to QC > QC Data Management > Select Scans.
Figure 208 Select Quality Control Scans
7. In the Select Quality Control Scans (Figure 208), select the scans used in step Step 9 on page 187,
(Section 8 on the worksheet).
Note The BMD, BMC, and Area average numbers must be from a tight grouping of
scans (no outliers).
3. Enter the Area and BMC values that were recorded in Section 8 on the worksheet).
Caution DO NOT change the Phantom Number or Height values (Figure 210).
Caution Enter Area and BMC data carefully on screen. The values appear in a different
order in the Cross-Calibration Calculator Spreadsheet.
AREA and BMC are displayed sequentially in Arrc.txt file.
BMD, BMC, and Area are in sequence in the Cross-Calibration Calculator
Spreadsheet.
4. If asked, click Continue > Yes to Accept the new ACF and BCF values from the Field Service
Calibration program.
5. Click Write to update the Arrc.txt file.
Figure 211 Click Write
6. Select Exit.
7. In Windows Explorer, browse to the C:\QDR\data folder.
8. Double-click on Arrc.txt to open the text file.
9. Verify that the information for your system matches the numbers in the file for:
• Serial Number
• Phantom Serial Number
• Area
• BMC
10. Record the results from the Arrc.txt file in Section I on Worksheet (Figure 212).
Note For additional details about System Backup, refer to MAN-04725 DXA
Replacement CPU Upgrade Checklist or to Technical Bulletin TB-00540.
Note Place a copy of this cross calibration backup on the initial system backup disk
included with the system software.
Updated Label
• Client configurations Programs > Hologic > HL7 Client - Capture Screen shots
• Base line scans. Does the customer have base line scans on the system?
• Copy from C:\QDR\Data folder all the *.axx, *.pxx, and *.rxx files to the CDRW or USB
removable media in a folder named Data.
• Saved DXReports
• Copy from C:\Program Files \Hologic\Physician's Viewer\ Options\DxReport\Output\
folder all of the *.dot files to the CDRW or USB removable media in a folder named
DXOutput.
Note PV7 now has a Backup and Recover function that will allow you to backup and
restore report rules, templates, CannedText and HL7 configuration files.
• Saved HXReports
• Copy from C:\QDR\HxQuestionnaire\Results all of the files to the CDRW or USB
removable media in a folder named HXOutput.
• Express Workflow configurations
• Go to Utilities > System Configuration > Study and capture a screenshot.
• Configurations on the Report tab: - Screen shots of each tab.
• Configurations on the Analyze tab: - Screen shot of tab.
• Configurations for Analyze and Reports: - Screen shot of tab.
• Configured Archive locations:
• Go to Utilities > System Configuration > Archive and capture screenshots of all the archive
location pool configurations.
• Verify the archive configuration:
• Did the customer have “Do Not Delete Scans” checked?
• Did the customer DICOM Store to PACS?
• Did the customer archive to a mapped archive location?
• Copy the canned text.mdb from the folder DXReport Temporary on the CDRW or USB
removable media into the Program Files\Hologic\Physician’s Viewer\Options\DX
Report\Temporary folder.
• In the Program Files\Hologic\ Physician’s Viewer\Options\DX Report \Templates folder,
rename the *.dot files to *ORIG.dot.
• Copy the saved files from the folder DXReport Templates on the CDRW or USB removable
media into the Program Files\Hologic\ Physician’s Viewer\Options\DX Report \Templates
folder.
• Install HL7 Report option. If necessary. – QDR login
• Copy the HL7report.mdb file in the folder DXReport Temporary on the CDRW or USB
removable media to the Program Files\Hologic\Physician’s Viewer\Options\DX Report
\Temporary folder.
• Copy the custom named file xxxxx.dot or the HL7report.dot from the CDRW or USB
removable media into the Program Files\Hologic\Physician’s Viewer\Options\DX Report
\Templates folder.
• Copy the HologicHL7.dot from the folder named OfficeXX Startup on the CDRW or USB
removable media.into the C:\Program Files \Microsoft Office\OfficeXX\ STARTUP.
• Client configurations Programs > Hologic > HL7 Client - from the Screen shots
• Install FRAX. – QDR login
• Install BCA Reporting options. If necessary – QDR login
• Install Hx Questionnaire. – QDR login
• Once Hx Questionnaire has been reinstalled, rename the HologicHx.xml in the Template
folder
• Copy the HologicHx.xml from the folder named HXQuestionnaire Template on the CDRW
or USB removable media into the C:\QDR\HxQuestionnaire\Template\ folder.
• Install all the option keys. – QDR login
• Base line scans. Did the customer have base line scans on the system?
• Copy from the CDRW or USB removable media in the folder named Data to the
C:\QDR\Data folder all the *.axx, *.pxx, and *.rxx files.
• Saved DXReports. Did the customer have DXReports on the system?
• Copy from the CDRW or USB removable media in the folder named DXOutput to the
C:\Program Files \Hologic\Physician's Viewer\ Options\DxReport\Output\ folder all of
the *.dot files.
Note PV7 now has a Backup and Recover function that will allow you to back and
restore the saved reports, templates, CannedText and HL7 configuration files.
default default
Subnet Mask
Gateway
Modality Worklist □
Provider
MPPS □
Provider
PACS Store □
Provider
Storage Commit Provider □
Network Archiving
Table 60 PACS Vendor and Outbound reporting to RIS or EMR Provider Information
Name Vendor:
Table 60 PACS Vendor and Outbound reporting to RIS or EMR Provider Information
Version Version:
Caution Changing the backup filename makes restoring the correct file difficult.
4. Click OK.
This procedure restores Windows 7 on a QDR system with computer part number ASY-08135.
Set Up Windows
1. Accept the default settings for the country, time and currency, and keyboard layout. Click Next.
2. For the user name enter Admin and for the computer name enter HOLOGIC-PC. Click Next.
3. Enter password for all three fields (password, retype password, and password hint). Click Next.
4. Enter the Product Key from the Windows COA. Click Next.
5. Check I accept the license terms. Click Next.
6. Select Ask me later.
7. Set the local date, time, and time zone. Click Next.
8. Select Work network. Windows setup finalizes the settings and then Windows opens under the
Admin account.
9. Select Log Off. Login as Field Service.
10. Open the control panel and select Windows Update > Change settings.
11. Set the Important updates frequency setting to Never check for updates (not recommended). Click
OK.
12. Go back to the control panel home and select Action Center > Change Action Center settings.
13. Uncheck all messages except Network firewall and User Account Control. Click OK.
14. Log out of the Field Service account and log in with the QDR account.
15. Open the control panel and select Action Center > Change Action Center settings.
16. Uncheck all messages except Network firewall and User Account Control. Click OK.
17. Close the control panel and log out.
Re-Install Applications
The ASY-08135 Image Disk only contains the Windows 7 operating system. No pre-installed Hologic
applications are included on the disk.
1. Log in as QDR user.
2. Manually install Hologic applications such as Apex/QDR and Physician’s Viewer.
3. Reinstall any required non-Hologic software or files such as MS Word and custom DxReport
templates.
Note The DXReport backup captures the following reporting information listed in the
.zip file.
4.
Note The installation program begins. The progress of the SQL Server 2012 Express
installation will be displayed in the command window. A message will be
displayed when it has completed. This may take several minutes so please be
patient.
11. The Select Horizon Machine Type is displayed. Select the appropriate Horizon Model, click Next.
12. The Serial Number screen is displayed. Enter the serial number and Install key. Click Next.
Figure 223 Serial Number
13. The Choose Destination Location is displayed. Click Next to accept the C:\QDR folder as the
destination location. A progress bar displays while the program files are being installed.
14. The InstallShield Wizard Complete screen appears. Yes, I want to restart my computer now option
is selected.
Figure 224 Restart Computer
15. Remove the APEX applications DVD from the DVD drive. Click Finish.
16. The system reboots.
Note When Hologic Connect is started and connection is made to the system by the
Help Desk, the APEX application will turn off the table communications. After
connection is terminated the Apex application must be restarted to use the table.
DICOM Configuration
Use the following table to configure DICOM and related network parameters on a QDR bone densitometer
(DXA) system. Click on any parameter to see its configuration screen and how to navigate to it.
The example shown below is for a local network using Conquest on the FE laptop as a PACS. All
parameter values shown are examples; actual values at any given site may vary. You should use the actual
values from the Connectivity Site Survey for your site.
Table 62 Local network Sample
Device Check if
IP Address AE Title Port Number Not Used
Gateway
DICOM Testing
Always test all DICOM services purchased on the customer network and PACS. Failure to do so may
result in loss of customer data.
If the customer network or PACS is not available for testing, use the following procedures to locally setup
and test DICOM services to verify proper operation of the Hologic modality. NOTE: Local testing is not a
replacement for testing on the customer network and PACS.
The example shown below is for a local network using Conquest on the FE laptop as a PACS. All
parameter values are examples; actual values at any given site may vary.
Click on any DICOM Service or linked Device in the table below for a detailed test procedure.
Gateway
6. In the Select Scan Type window (Figure 233), verify that the Patient Name is Worklist, Test Patient
and that the Patient ID is 1111.
This indicates a successful test.
7. When done, click Cancel.
Figure 233 Select Scan Type
Tool Size/Type
Hex driver 3/32-in.
Hex driver or wrench 5/32-in.
Nut driver 1/4-in.
Nut driver 3/8-in.
Nut driver 5/16-in.
Nut driver 7/16-in.
Screwdriver Narrow slotted
Screwdriver Phillips head
Screwdriver Slotted
Wrench 3/8-in.
Note The two bottom nuts are accessible with a ratchet wrench and extension.
Note When starting this procedure, make sure the encoder coupling setscrew is not
tightened.
Note When replacing the TY Motor Controller board, ensure that the ID switch is set
to 5.
1. Center the table and remove the two screws that secure the tabletop in place.
2. Shut down the computer and then turn off the main circuit breaker.
3. Pull the tabletop to the left far enough to remove the right table rail end cover, and remove the
cover. Mark this as the right cover so that it is not confused later with the left cover (they are not
interchangeable).
4. Pull the tabletop to the right far enough to remove the left table rail end cover, and remove the
cover. Mark this as the left cover so that it is not confused later with the right cover (they are not
interchangeable).
5. At the encoder end of the belt, loosen the 2 bolts holding the tension block, remove the tensioning
nut, and remove and replace the belt.
6. Install the tension spring and tensioning nut.
7. Tighten the tension nut so that the spring is compressed to 7/8 in. The bracket cutout can be used as
a measuring guide (the inside of the washer should be flush with the bracket cut).
8. Tighten the two mounting bolts holding the tension block.
9. Restore power, boot the computer and login as Service.
10. Perform the procedure detailed inCalibrate TY (Horizon Models A, W, Wi) on Page 122.
Note When starting this procedure, make sure the encoder coupling setscrew is not
tightened.
Note When starting this procedure, make sure the encoder coupling setscrew is not
tightened.
Note When starting this procedure, make sure the encoder coupling setscrew is not
tightened.
Note When replacing the TX Motor Controller board, ensure that the ID switch is set
to 4.
Table X Belt
To remove and replace the Table X Belt located within the X Table Drive Assembly:
1. Before removing power from the system, move the table to the far left using the Motor Control Pad.
The Motor Control Pad can be accessed from the Main Menu screen by choosing Utilities >
Emergency Motion.
2. Shut down the computer and then turn off the main circuit breaker.
3. Remove 2 screws from the right table rail end cover, and slide the cover off from the end.
4. Remove the front endcap from the table X drive assembly (3 Phillips screws).
5. Remove the back Phillips screw from the right side cover (of the table X drive assembly) and slide
the cover out from the front. This provides access to the belt.
6. At the encoder end of the belt, loosen the 2 bolts holding the tension block, remove the tensioning
nut, and remove and replace the belt.
7. Install the tension spring and tensioning nut.
8. Tighten the tension nut so that the spring is compressed to 7/8 in. The bracket cutout can be used as
a measuring guide (the inside of the washer should be flush with the bracket cut).
9. Tighten the two mounting bolts holding the tension block.
10. Restore power, boot the computer and login as Service.
11. Perform the procedure detailed inCalibrate TX on Page 124.
Note When starting the MOTOR$TX calibration procedure, make sure the coupling
setscrew is not tightened.
Note When starting the MOTOR$TX calibration procedure, make sure the coupling
setscrew is not tightened.
Table X Encoder
To remove and replace the Table X Encoder located within the X Table Drive Assembly:
1. Before removing power from the system, move the table to the far left using the Motor Control Pad.
The Motor Control Pad can be accessed from the Main Menu screen by choosing Utilities >
Emergency Motion.
2. Shut down the computer and then turn off the main circuit breaker.
3. Remove 2 screws from the right table rail end cover, and slide the cover off the end.
4. Remove the front endcap from the X table drive assembly. The cover is held on with 2 hex screws.
5. Remove 6 Phillips screws from the right side cover (of the X table drive assembly) and slide the
cover out from the front. This provides access to the encoder and belt.
6. Remove 5 flat head Phillips screws from the upper pedestal cover and remove the cover. This
provides access to the Motor Controller Board.
7. Unplug the encoder from the Motor Controller Board and pull out the cable.
8. Remove the coupling holding the encoder to the belt drive.
9. Remove the encoder from the bracket assembly.
10. Replace the encoder on the bracket, install the coupling but do not tighten the screws.
11. Restore power, boot the computer and login as Service.
12. Perform the procedure detailed inCalibrate TX on Page 124.
Note When starting the MOTOR$TX calibration procedure, make sure the encoder
coupling setscrew is not tightened.
TZ Pedestal
To remove and replace either pedestal, refer to Figure 244 and complete the procedure that follows. The
procedure requires two pieces of 2 x 4 lumber, approximately 3 ft long, and two pieces of foam padding.
Figure 244 Installing the Rotary String Encoder
Note If removing the left pedestal, the left T-rail will become loose. Be careful not to
drop this T-rail.
If removing the right pedestal, the table’s foot end is floating and not captured
and can fall off the rail if the table is pulled toward the head-end.
10. Press the appropriate pedestal switch on the TZ Motor Controller board to lower the pedestal
completely.
11. Remove the pedestal power plug, the lower four pedestal bolts, and remove the pedestal.
12. Replace the pedestal and install the pedestal lower bolts and or nuts, but leave them loose for now.
13. Replace the power plug.
14. On the TZ Motor Controller board, set the Direction switch to Up.
15. Press the appropriate pedestal switch on the TZ Motor Controller board to raise the pedestal until it
just touches the T-rail.
16. Install the upper pedestal bolts leaving the bolts loose for now.
17. Using the switches on the TZ Motor Controller board, raise both pedestals and remove the 2 x 4s
and foam.
18. Measure from the inside of one T-rail to the inside of the other. It must be 65 in. at both the front and
back. If it is not, move the pedestal until the measurements are correct.
19. On the TZ Motor Controller board, set the Direction switch to Down.
20. Using the switches on the TZ Motor Controller board, lower both pedestals all the way down.
21. On the TZ Motor Controller board, set the Service switch to Normal. Press Reset (large black
button).
22. Remove the table pad.
23. Launch Emergency Motion from the Utilities Menu in the application.
24. Move the C-arm to the right (head-end). Select Arm Right.
25. On the right side of the C-Arm, use a metric tape measure and measure from the front edge of the
table across the top of the table to the back edge of the right side of the C-arm.
26. Move the C-arm to the left (foot-end). Select Arm Left.
27. On the left side of the C-Arm, use a metric tape measure and measure from the front edge of the
table across the top of the table to the back edge of the left side of the C-arm.
28. With your hand force the foot-end of the table in or out to match the measurement taken at the head
end of the table.
29. Tighten the lower and upper pedestal bolts and nuts or the right (head-end) pedestal.
30. Repeat the measurement steps again to ensure these measurements are equal to one another from
the back of the C-arm to the front edge of the table.
Note If additional adjustment is needed, loosen the lower pedestal bolts on the right
(head-end) to make the adjustment.
31. Use the Emergency Motion to move the table out. Select Table Out.
32. Look at the clearance between the table rollers and top T-Rail on the foot end.
33. Use the Emergency Motion to move the table out. Select Table In.
34. Look at the clearance between the table rollers and top T-Rail on the foot end.
35. Adjust the foot end T-Rail until equal clearance is achieved at the table in and out positions.
36. Tighten the lower and upper pedestal bolts the left (foot-end) pedestal. Refer toCalibrate TZ on
Page 107 for details.
37. After you complete the calibration process, replace all covers and table pad.
Refer to Figure 243 and Figure 244 and follow the procedure below:
1. Remove the electronics tray covers.
2. Push in the Emergency Stop switch, then press and hold the Table Up switch to move the table all
the way up.
Warning: Shut down the computer, then turn off the main circuit breaker on the foot-end
pedestal.
3. Remove the flat head Phillips screws from the upper pedestal cover and remove the cover.
4. Remove the flat head Phillips screws from the lower pedestal cover and remove the cover.
5. Remove the linear encoder top screw (Phillips).
6. Remove the linear encoder bottom screw (access to the bottom Phillips screw is through the frame
hole).
7. Unplug the linear encoder cable from the TZ Drive board and pull the cable out.
Caution Do not allow the string to snap back into the encoder after being extended. This
can permanently damage the unit.
8. Route the new encoder cable through the path of the old encoder and plug it into the TZ Drive
board.
9. Install the rotary encoder at the bottom using the screw and two washers provided.
Note The bracket must sit on the top surface of the base frame before tightening.
10. Replace the linear encoder top screw by extending the string and fastening to the top using screw
and washers provided.
11. Turn on the instrument power, boot the computer and log in as Service.
12. Refer to the procedure inCalibrate TZ on Page 117 for details. When calibration and adjustment is
done, replace the pedestal covers.
Adjustment
Note Perform the following adjustment steps ONLY when the pedestal encoders are
NOT within the +/-2 counts specified (target value is 325 +/- 2).
Note Upon recalibration, the 'Lower Right' and 'Lower Left' count values should
become virtually equal. If the 'Lower Left' and 'Lower Right' count values differ
by more than +/- 2 counts, readjust the encoder by rotating the center hub until
the 'Encoder Position' count value equals the 'Lower Left' count value +/- 2
count difference. Secure the three adjustment screws on the face of the encoder.
Return to Step 1 of theCalibrate TZ on Page 117 to recalibrate at the new
encoder setting.
Warning: The X-ray Source Assembly weighs 130 lb (59 kg). To prevent personal injury
two people are needed to safely remove the assembly.
Caution Do not use the Table Lift function to raise the table. Use Emergency Motion or
pedestal switches on TZ control board only.
2. Select the Utilities > Service Utilities > System Diagnostics > Detectors tab.
3. Set Aperture to 7.
4. Shut down the computer and then turn off the main circuit breaker.
5. Remove the lower C-arm cover by removing 2 Phillips screws and sliding it out from the front.
6. Remove the 4 Phillips screws holding the Lead Shield Carriage Mounting Assembly and remove
the shield.
Remove the Filter Drum Assembly
7. Remove the Filter Drum Assembly.
Figure 247 Filter Drum Assembly
Tool Size/Type
Hex driver 3/32-inch
Hex driver or wrench 5/32-inch
Nut driver 1/4-inch
Nut driver 3/8-inch
Nut driver 5/16-inch
Nut driver 7/16-inch
Screwdriver Slotted (1/4 in.)
Screwdriver Narrow slotted (1/8 in.)
Screwdriver Phillips head, 0
Screwdriver Phillips head, 1
Screwdriver Phillips head, 2
Wrench 3/8-inch
Note In areas with limited physical access to the aperture and filter drum assembly,
on the Horizon C, lift the table insert to access the assembly. For W and A
systems, detach the table from its drive belt and push the table to the side
If the room doe not have enough clearance to push the table to one side, remove
the X-ray Source Assembly to access the aperture/filter drum assemblies.
6. From the front of the unit, loosen the mounting screws (4) to remove the lead shield (Figure 251).
7. Unplug connectors J9 and J15 (labeled) from the aperture/filter drum controller board.
Figure 252 Removing Connectors
8. Remove the three (3) 3/32 in. Allen head screws that secure the filter drum assembly (Figure 253).
Note Although the filter drum can be disassembled while attached to the Horizon
Unit, it is usually best, especially if the assembly will be replaced, to remove the
filter drum assembly and place it on the scanner table, a workbench, or other
convenient work area.
2. Remove the filter drum assembly from the Horizon unit (Figure 254).
Figure 254 Filter Drum Assembly Removed
Part Description
ASY-07159 ASSEMBLY, APERTURE/FILTER DRUM, SINGLE SLOT
Part Description
ASY-07557 KIT,HORIZON-CI/WI APERTURE REPLACEMENT
Figure 255 ASY-07159 Aperture/Filter Drum Assembly Top View
Aperture Removal
1. Remove the four (4) 6-32 x .75 in. aperture screws (Figure 255).
Aperture Replacement
1. Place the aperture assembly on top of the filter drum.
2. Orient the plate so that the short end is at the back of the filter drum assembly (Figure 256). The gap
at the back edge is 2.21 in (5.61 cm).
Caution The orientation of the plate is specific. The measurement from the slot to the
back edge of the plate is 2.21 in (5.61 cm). The distance from the front edge to
the slot is larger. If not, the table front to back alignment will be off. The back
edge of the beam will be cut off because the slot would be too far forward in the
beam path.
Part Description
ASY-07116 FILTER DRUM ASSEMBLY, HORIZON
Table 71 Field Replacement Kit for the Horizon Aperture/Filter Drum Assembly
Part Description
ASY-08683 FILTER DRUM, HORIZON, FRU
ASY-07098 ASSEMBLY, APERTURE/FILTER DRUM, NINE SLOT
Figure 257 ASY-07116 Aperture/Filter Drum Assembly Top View
Note For Horizon A systems, it may be easier to position the table as far up and as far
back as it will go.
2. Shut down the computer and then turn off the main circuit breaker on the foot end pedestal.
3. Remove the lower C-arm cover by removing two Phillips dress screws on either side of the C-arm
cover and sliding it out from the front.
4. Remove the four (4) Phillips screws holding the Carriage Front Lead Shield and remove the shield.
2. Disconnect the Aperture Position Cable from J2 on the Filter Drum Controller Board (Figure 260).
3. Remove the two screws that secure the aperture motor shaft to the aperture assembly (Figure 261).
Note This step frees the assembly and allow access to the screws below.
Caution Upon completion of this step, the aperture assembly is no longer secured to its
mounting plate.
4. Refer to Figure 262 and remove the two screws (aperture belt side) and two screws (aperture motor
side) that secure the aperture assembly to the drum assembly.
Note You will need to slide the aperture assembly slightly to the left to access one of
the screws on the aperture motor side.
Note The calibration key should be perpendicular to and flush with the end plate. The
end of the key with the flat surface should be towards the motor and the stepped
side facing down, so that it engages the slot in the end plate and slots in both
drums
Figure 266 Alignment Tool Placement
16. Remove the segment and place it after segment 1 on the work table
17. Continue in order until each remaining segment (3 – 6) has been removed, then placed in sequence
on the work table.
Number the Inner Drum Segments (1 – 4)
18. With a marker, identify the first brass segment of the inner filter drum with a number 1.
19. Rotate the drum with the segments in place, number each remaining segment and its location on
the end plate.
20. Loosen the screws and remove the segment (Figure 21).
21. Removing Inner Segment
22. Rotate the outer drum counter-clockwise to expose the screws that secure the next segment to the
drum.
23. Number that segment 2 and remove it.
24. Continue in order until each remaining segment (3 – 4) has been numbered and removed.
2. Align screw holes in the unmarked end plate to the approximate location of the screw holes in the
marked end plate (Figure 268).
3. Starting with segment number 1 (Figure 269), tighten the segments in order.
Figure 271 Filter Drum Assembly Base Plate Serial Number Location
4. Write the original aperture/filter drum serial numbers on the new labels.
5. Apply clear laminate 540-0111 over the serial numbers.
6. Restore power, boot the computer and login as Service.
7. Perform the Calibrate Aperture Procedure detailed in the Horizon Service Manual, MAN-03310.
8. Replace all covers that were removed.
Label Placement
If an aperture assembly is being replaced, labels must be placed on the:
• Filter drum base plate.
• Back of the C-arm
If a drum assembly is being replaced, labels must be placed on the:
• Filter Drum base plate
• Back of the C-arm
Remove Covers
1. Shut down the computer
2. Turn off the main circuit breaker on the foot end pedestal.
3. Lift and remove the top C-arm cover.
4. Lift it carefully from the system.
5. Remove the 5/32 in. hex head screws (4) on both sides of the detector cover/shield (Figure 2).
Figure 274 Remove Screws on Either Side of Detector Cover/Shield
6. Remove the weight from the front of the C-arm (Figure 276).
Figure 275 Remove Weight
7. Slide the Detector Cover forward to clear obstructions and remove it (Figure 276).
Remove Cables
9. Disconnect the cables (Figure 279) from A/D board:
a. three (3) from the Detector Board
b. one (1) ground to the chassis
c. one (1) from the A/D Board (Figure 280)
Figure 279 Detector Cables
2. With a Philips screwdriver, start to loosen the Detector Assembly screws (Figure 282).
Figure 282 Loosen Detector Screws
3. With a needle-nose pliers, grip and remove the screws (4) that secure the Detector Assembly to the
upper C-arm base plate (Figure 283).
Figure 283 Use Pliers to Hold and Remove Screws from Detector
Caution Be sure to maintain a firm grip on the flat washer, lock washer and mounting
screw.
5. Slide the Detector Assembly forward towards the front of the C-arm (Figure 284).
Figure 284 Slide and Lift Detector Assembly
2. Remove the screws from the cover of the Detector Mounting Base Plate (Figure 286) to access the
laser assembly (Figure 287).
Figure 286 Remove Screws from the Detector Mounting Base Plate
Caution Small springs apply load from the laser to the cover. Do not lose them when
removing the adjustment screws.
Do not use power tools to remove the screws,
2. Remove the laser assembly (Figure 289).
Figure 289 Laser Assembly
Bottom Top
Caution Bring the base plate to the laser assembly to avoid losing the three (3) springs
mounted on the assembly.
2. Align the openings in the base plate to the screws in the assembly (Figure 290).
Caution Always start the adjustment screws by hand. Do not use power tools to tighten
the screws.
4. Remount the detector assembly onto the base plate (Figure 19).
5. Align the laser detector base plate to the detector assembly cover.
Figure 292 Align the Laser Detector Base Plate to the Detector Cover
7. Place the board end of the Detector Assembly in the end of the C-arm (Figure 294).
9. Align the Detector Assembly to the pencil marks drawn before removal of the Detector Assembly.
10. Use a needle-nose pliers to start the screws holding the assembly.
Caution Be sure not to grip the lock and flat washers when starting the screws.
Mount Covers
1. Attach the weight to the front of the C-arm.
2. Replace the Detector Cover.
3. Replace the C-arm upper and lower covers.
Removal
1. Shut down the computer and then turn off the main circuit breaker on the foot end pedestal.
2. Unplug the cables and remove the screws on the detector assembly and remove it from the system.
3. Lift off and remove top C-arm cover.
4. Remove the upper C-arm back cover.
5. Remove 4 5/32 in. hex head screws on either side of the detector cover/shield.
6. Remove the 4 Phillips screws that secure the detector assembly to the upper C-arm base plate.
Installation
1. Install the new detector assembly using the screws and cables removed in the previous step.
2. Turn on the Horizon power, boot the computer and login as Service.
3. Perform X-ray Beam Alignment (Section 22 Aperture Alignment and Calibration on Page 255).
4. Perform the Section 21.2 A/D Gain Control Adjustment on Page 226.
5. Perform the Laser Positioning Offset Adjustment (Section 21.1 Check Laser Positioning Offset on
Page 226).
6. Replace any covers that were removed and perform Detector Flattening (Section 21.6 Detector
Flattening on Page 240).
7. Perform the Field Service Calibration procedure (Section 21.4 Check Phantom Values on Page 228).
8. Perform a System Backup.
9. Perform a System Recover. Verify that the System Backup recovers without error.
10. Verify proper system operation.
1. Shut down the computer and then turn off the main circuit breaker on the foot end pedestal.
2. Remove the top C-arm cover.
3. Remove the cables at the Detector Assembly.
4. Remove the 4 bolts holding the Detector Assembly (on rubber grommets) to the C-arm.
5. Remove the Detector Assembly.
6. To replace the laser only, loosen the laser clamping screw. To replace the laser assembly, remove the
three mounting screws.
7. Restore power, boot the computer and login as User.
8. To adjust the laser, loosen the laser clamping screw, turn the laser on and rotate it until the correct
alignment is seen.
9. Perform procedure in Section 21.1 Check Laser Positioning Offset on Page 226.
Table 81 Cables
Table 82 Miscellaneous
Table 83 Mobile
3. Set the following parameters in the Device Control Screen (Figure 359 and Figure 360) to the values
shown:
• High Gain = 2
• Low Gain = 1
• X-Ray Mode = 3
• Aperture = 7
4. Verify that the screen shows Pulse equal to 1.
5. Select Filter Wheel to ON to ensure the filter drum is spinning.
Figure 301 X-Ray Survey Screen (X-rays OFF)
Note To turn X-rays Off and On you can press the F2 Function key on the keyboard
or by using the mouse toggle the F2 button on the screen.
7. Insert the X-ray beam alignment tool, TLS-05376, into the slot on the C-arm (Figure 362).
Note The handle of the alignment tool protrudes from the C-arm to prevent
inadvertently leaving the tool in the C-arm after performing an alignment.
8. Press the keyboard F2 function key or with the mouse toggle the F2 button on the screen to turn X-
rays On.
9. Check the voltage output on the Device Control Screen (Figure 363).
Note Regardless of the initial voltage readings, the goal of the adjustment is to center
the filter drum and aperture assembly in the middle of the X-rays beam to
achieve maximum amplitude with a uniform flat signal across the top of the
graph.
If the arm is deflected in either direction, when the filter drum assembly is
properly aligned, the voltage drop should be the same in both directions.
On Horizon A systems use Emergency Motion to raise the table (Figure 6) to
access the area underneath the table top.
Caution Do not drive the table to its maximum raised position. Raising the table to its
maximum height at abnormal speed can damage hardware.
2. From the Hologic main menu, select Utilities > Service Utilities >System Diagnostics > Detectors
tab and select aperture 7 on Horizon C, W, and A models only.
Note Do not use the arrow up/down function to select Aperture 7. Instead, use the
mouse to highlight the displayed aperture number, type 7 and press Enter.
3. Use the alignment block TLS-02698, and alignment pin 099-0111 to align the hole in the filter drum
base plate to the hole in the aperture plate and to the aperture base plate assembly.
The pin should drop through easily. If the pin does not drop in, select the C-Arm
tab on the System Diagnostics screen.
4. In the Commands section at the lower right, type 10 or -10 in the Move Relative field to re-align the
aperture plate. You may need to visually check that you are moving the aperture plate in the correct
direction relative to the base plate hole.
The pin should now drop down through the hole in the back left of the aperture
plate down through the alignment block and into the base plate hole with
minimum force. If the pin still does not align with the base plate hole, continue to
move the aperture plate ten steps at a time until alignment is achieved.
5. Press or click F2 to turn on X-rays.
6. Locate the (2) 10-32 x .37L Hex should bolt screws on the bottom sled bracket of the Filter Drum
Sled assembly, and loosen them by 1/2 turn.
7. Adjust the rear sled adjustment to move the whole assembly left or right in the beam (figure 119).
The goal is to achieve the maximum voltage on the left side of the survey screen.
8. Press or click F2 to turn off x-rays.
9. Tighten the (2) 10-32 x .37L Hex bottom sled screws.
10. Perform skew alignment (See Perform Skew Filter Drum Base Alignment).
11. Repeat the lateral and skew process until beam is flat and at maximum amplitude.
12. Remove the pin and block when done.
2. Ensure that all components can freely adjust by turning the sled Skew adjustment screw. Loosen
filter drum base screws slightly if necessary (Figure 368).
Note Deflect the arm slightly to the left and right. The voltage should decrease when
the outside the beam path.
5. Adjust the Filter Drum Assembly Skew (Figure 370) until the amplitude displayed on the Device
Control Screen is flat.
Figure 312 Voltage Amplitude on Device Control Screen
If a detector at the end of the charts (e.g., Detector 44, Figure 371) has on an amplitude less than 6.0V and
less than the detector on the opposite side (e.g., Detector 171, Figure 314), the detector must be adjusted
front to back, so that both detectors display the same voltage.
Figure 313 Details Detector 44
8. Tighten screws.
Caution Make sure the pin and block have been removed before performing Aperture
Calibration.
1. From the Hologic main window, click Utilities > Service Utilities > System Diagnostics.
2. Click the C-Arm tab.
3. In the upper right corner of the screen, click Calibrate Aperture (Figure 375) to move the aperture to
its right and left physical limits.
Figure 317 Select Calibrate Aperture
The x-rays turn on as the aperture plate moves left to right. As the aperture moves through the beam path,
the signal is captured and displayed. The chart will identify the signal levels for each aperture specific
encoder positions.
The vertical lines on the final output should correspond to the center of the aperture openings. The data
contributes to software positioning and aperture identification.
Slope Value
4. Click Continue.
As the potentiometer moves in both directions, a linear change should be displayed in the Aperture
Calibration Plot (Figure 377).
5. Click Continue to open the Aperture Calibration Peaks screen (Figure 378)
Figure 320 Aperture Calibration Peaks
• The green lines should be centered at the peak of the signal for each aperture if the aperture is
moving correctly. This screen also gives an indication of whether the aperture plate is moving
erratically or binding.
• The red lines are typically displayed in pairs to reflect the pairing of the apertures. In between the
pairs, the beam should be off (maximum dark).
6. Click Continue to open the Aperture Calibration Data screen (Figure 379) and review the offset
numbers.
Figure 321 Aperture Calibration Data
Caution Be sure to restart the software in Service Mode after calibrating the aperture, so
the software can access the new data file. If the system is not rebooted, the
software will not be able to find a selected aperture correctly.
Caution Any adjustment of this potentiometer affects the QC highs and lows. You must
run detector flattening and recalibrate the machine after adjusting the A/D
gain. DO NOT ADJUST THE A/D GAIN UNLESS ABSOLUTELY
NECESSARY AS DEFINED in Step 9 in the section that follows (Verify and
Adjust the A/D Gain).
6.2.1 Principles
Radiation leakage from the HVPS/S results from a defect in the lead liner of the tank or its cover. If there is
a hole in the lead liner, the resulting leakage will produce a pencil beam of X-rays. If there is a crack in the
lead liner, the resulting leakage will produce a fan beam. In either case, the initial survey must be taken
close to and almost touching the tank with the survey meter probe. If a leak is found, the meter indicates a
spike in the reading. If this occurs, you must then move the probe one-meter (approximately 40 in.) along
the beam from that spot to determine if the leakage is within acceptable levels.
2. Start the Utility by selecting Utilities > Service Utilities > X-Ray Survey.
3. Change X-Ray Mode to 3 and Aperture to 7 to lock in the changes.
Warning: X-rays are generated during this procedure. Keep hands, head and other body
parts out of beam.
1. Position the survey meter one meter (approx. 40 in.) from the tank and move it all around the
outside of the instrument.
2. The reading must be 1.0 mrad/h (10 µGy/h)or less. Record the highest reading in the service report.
3. Press the <F2> key to turn off X-rays, remove the tool, and replace the tank cover.
Note Enter this information into the Radiation Measurement Report (CSD-0042-
F07).
This procedure flattens the X-ray beam for each scan mode and uses the metal edges of the table to
calculate the center table position in the x-axis.
Note All covers, table mat, etc., normally in the X-ray path, MUST be on the Scanner
before running Beam Flattening.
1. Restart the Horizon software in service mode (if not already). Press the Center Table button and
turn on the laser.
2. Using a DXA Quality Control Phantom (Part Number ASY-01564), place the phantom on end with
the flattening target facing up. Orient the phantom so the metal edge is to the right (head end) and
tissue to the left (foot end) as viewed from the front of the table. Align the phantom flattening target
to the laser crosshair.
Figure 324 Correct Flattening Position
3. Start the X-Ray Survey Utility by selecting Utilities > Service Utilities > X-Ray Survey.
4. Set Pulse to 1, High gain to 2, Low gain to 1, X-ray Mode to 3 and Aperture to 7 (for A, C, and W
models).
5. For Display, click Graph.
6. Click X-Rays (F2) to turn X-Rays on.
7. Check that the phantom covers the entire beam. This is critical. If phantom does not cover the entire
beam, move it until it does. Keep the phantom as straight as possible.
8. Turn off X-rays with the <F2> key and press <Esc> or click Close to exit X-Ray Survey.
9. Reboot the PC, logon as Field Service, and restart the software in Service Mode to return to Horizon
Main Menu screen.
10. Select Utilities > Service Utilities > Detector Flattening.
11. Click Flattening –Basic First.
Figure 325 Flattening –Basic First
This procedure will take approximately 15 minutes on a Ci and Wi, 25 minutes on a C and W, and
50 minutes on an A model system.
13. After this procedure, the X-Ray Source Assembly will be hot. You must wait for 30 minutes before
performing a Field Service Calibration of the machine.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam
system or if the customer has multiple systems, then a cross calibration
must first be performed to ensure that any longitudinal studies begun on
the QDR being removed can safely be continued on the new Horizon. Refer
to the cross-calibration procedure in Section 3.10 Fan Beam Cross-
Calibration Gold Standard System on Page 164 before uninstalling the
existing QDR. The Cross-Calibration procedure applies to fan beam
systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
3. Compare values displayed with actual values printed on the phantom. If there is a discrepancy,
check to make sure you have the phantom that shipped with the unit. Check with Hologic Technical
Support if needed.
6.3.3 Perform BCA Calibration Scans and Table Top Radiographic Uniformity (A, W
and Wi)
This procedure verifies proper X-Ray beam alignment and the uniformity of the X-Ray transparency of
entire tabletop in a manual process (for Horizon A, W and Wi models). BCA Calibration Scans must be
performed on the system prior to verifying the uniformity test scan.
Note On Models A, W, and Wi, after Flattening is performed and 30 minute cool
down is expired, perform a Daily QC scan. Automatically the system will
perform BCA calibration scans on the QC Phantom. After the calibration scans
are complete the system will inform you to remove the QC Phantom and clear
the table to perform the Table Top Uniformity Scan test. This test will report the
results automatically and must pass with the Hi and Low Air SD of less than
2.0.
If the system is passing normally, skip to next step.
If the system is not passing with less than SD 2.0, verify that table alignments
and beam alignments are correct, adjust as necessary and run Flattening again.
1. If necessary, restart the software in Service Mode.
2. Type WBAIRQC in the Patient Biography. Enter the serial number of the unit in the Patient ID field.
Make sure that only the pad is on the table. Clear the table of anything else.
3. Perform one Whole Body air scan (a scan with nothing on the table). Do NOT interrupt the scan for
any reason. When the scan completes, select Exit Exam. Do not analyze the scan that was just
performed.
4. From the Main Menu screen, select Utilities > Service Utilities > Table Top Radiographic
Uniformity.
5. Select the Whole Body scan and click the Next>> button.
6. Table Top Radiographic Uniformity results tabs are now displayed. Select the Low Air Tab.
7. If the Global Stats S.D. (2) is less than 2.0, the instrument is properly aligned. If the Global Stats S.D.
is greater than or equal to 2.0, then the machine is not aligned properly. Check the following:
• If the image appears “streaky” or “banded”, check for loose wires or other debris between the table
and the tank.
• Check the Aperture Calibration.
• Check AC Line voltage.
• Check +/- 15 V Power Supply.
• Check for a dedicated line.
• Check Detector Flattening.
• Re-run daily QC (detailed in MAN-03644 Horizon QDR Series User Guide).
Print a copy of the test results and include it with the other paperwork being returned to Hologic.
After realignment, repeat the Table Top Radiographic Uniformity test. If the global S.D. is less than 2.0,
you are done. If the global S.D. if equal or greater than 2.0, print a copy of the test results and include it
with the other paperwork being returned to Hologic.
If the global is equal to or greater than 2.0, the non-uniformity may be attributed to the table itself. If so,
continue with this procedure. Make sure that the machine has body composition loaded. If it does, then
follow the appropriate procedure below in Section 0.9 Instruments using Body Composition Analysis (BCA)
on Page 13.
3. Center the table, place the phantom on the table and align with the laser. The sides of the phantom
should be parallel to the table. Use a ruler if necessary.
4. Click Perform Exam from the application main screen.
5. At the Select a Patient for this Exam window, select the phantom, and click OK.
6. At the Patient Confirmation screen, click OK.
7. At Select Scan Type screen, select AP/Lateral, uncheck Use Default Scan Mode, and click Next>>.
8. At the Select AP Lumbar Spine Scan Mode for AP/Lateral Exam screen, select Array (a), and click
Next>>.
9. At the Select Lateral Scan Mode for AP/Lateral Exam screen, select Fast Array (f), and click Next.
10. Recheck phantom alignment with laser.
Note Do not move the arm or table, otherwise the test will have to be started over.
11. Place two sheets of paper against the left and right edges of the phantom (Figure 416) When done,
click Continue.
12. On the Scan Parameters screen, set the scan length to 6.0 in., click Enter and click Start Scan.
13. When scan has finished, select Spine as the analysis method, and click Next>>.
14. Analyze the scan and click Start Position.
15. Refer to Figure 439 Verify that the angle displayed under the image (near the bottom of the screen)
is less than ± 2 degrees.
16. Using the sheets of paper as a guide, carefully rotate the phantom on its back edge towards the C-
Arm so that the bone segments are facing the rear of the table as shown. Except for the rotation, do
not move the phantom in any other direction.
17. Click Close. When the Continuously Press Enable message appears, press and hold the Lateral
Enable switch until the C-Arm and table move to the Lateral position.
18. On the Scan Parameters screen, click Start Scan.
19. When the “Continuously Press Enable” message appears, press and hold the Lateral Enable switch
until the C-Arm and table move back to the AP position.
20. From the Exit Exam screen, select Analyze Scan.
21. Analyze the scan (refer to user guide for Lateral analysis) and close the screen.
22. Click Exit to close the Exit Analysis screen.
23. Restart APEX software in service mode (if not already). From the APEX Main window, select
Utilities > Service Utilities > Field Service Calibration.
24. Select Highs/Lows for Array Spine Phantom.
25. Select the AP scan just performed and click Next.
26. Click Continue on the Calculated highs and lows page.
27. Select Highs/Lows for Array Lateral Phantom.
28. Click Continue on the Calculated highs and lows page.
29. Click Temp Write to write temporary results and click Continue to exit the write dialog box.
Note For Horizon system with version APEX 5.6 and higher, only 10 reproducibility
scans for calibrations are required. Earlier versions required 20 scans. Acquire 5
scans with a period of 15 between groups.
14. Analyze the first scan manually selecting Lumbar Spine for the analysis method.
15. Note the scan #.
16. Click the Close button to exit analysis.
17. From the Main Menu screen, select Utilities > Service Utilities > Auto Analyze.
18. Click unanalyzed Scans and select the 10 scans, click Compare radio button and then click Next.
19. Now, select the QC Scan performed in step Section 3.11.2 Perform QC Scan on Spine Phantom, (C, Ci)
With BCA Calibration Scans and Table Top Radiographic Uniformity (A, W and Wi) on Page 199.
20. Click Next.
21. The 10 scans will now be compared with the auto-analyzed QC Scan.
6.3.7 Calibration (Area, BMD, and BMC) with ACF and BCF Correction Factors
Calibration for Area, BMD, and BMC is accomplished in three stages:
• Array Scan Thickness Measurement & Calibration
• Calibration of Area and BMC for Array Scan Modes
• Adding Array AP scans to the QC database
Follow the procedure, in order, and exactly as shown, for each stage of the calibration.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam system
or if the customer has multiple systems, then a cross-calibration must be
performed to ensure that any longitudinal studies begun on the QDR being
removed can safely be continued on the new Horizon. Refer to the cross-
calibration procedure in Section 3.10 Fan Beam Cross-Calibration Gold
Standard System on Page 164 before de-installing the existing QDR. The Cross
Calibration procedure applies to fan beam systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
1. From the Main Menu screen, select Utilities > Service Utilities > Field Service Calibration.
2. Click the ACF/BCF button.
3. Select the 11 reproducibility scans performed in previous step.
4. Click Next
5. Click Continue at the Successfully Calculated dialog box.
6. At the Field Service Calibration dialog box, click the Write button to write the values to arrc.txt file.
7. Click Continue to return to the Main Menu.
3. Set the following parameters in the Device Control Screen (Figure 359 and Figure 360) to the values
shown:
• High Gain = 2
• Low Gain = 1
• X-Ray Mode = 3
• Aperture = 7
4. Verify that the screen shows Pulse equal to 1.
5. Select Filter Wheel to ON to ensure the filter drum is spinning.
Note To turn X-rays Off and On you can press the F2 Function key on the keyboard
or by using the mouse toggle the F2 button on the screen.
7. Insert the X-ray beam alignment tool, TLS-05376, into the slot on the C-arm (Figure 362).
Note The handle of the alignment tool protrudes from the C-arm to prevent
inadvertently leaving the tool in the C-arm after performing an alignment.
8. Press the keyboard F2 function key or with the mouse toggle the F2 button on the screen to turn X-
rays On.
9. Check the voltage output on the Device Control Screen (Figure 363).
Note Regardless of the initial voltage readings, the goal of the adjustment is to center
the filter drum and aperture assembly in the middle of the X-rays beam to
achieve maximum amplitude with a uniform flat signal across the top of the
graph.
If the arm is deflected in either direction, when the filter drum assembly is
properly aligned, the voltage drop should be the same in both directions.
On Horizon A systems use Emergency Motion to raise the table (Figure 6) to
access the area underneath the table top.
Caution Do not drive the table to its maximum raised position. Raising the table to its
maximum height at abnormal speed can damage hardware.
2. From the Hologic main menu, select Utilities > Service Utilities >System Diagnostics > Detectors
tab and select aperture 7 on Horizon C, W, and A models only.
Note Do not use the arrow up/down function to select Aperture 7. Instead, use the
mouse to highlight the displayed aperture number, type 7 and press Enter.
3. Use the alignment block TLS-02698, and alignment pin 099-0111 to align the hole in the filter drum
base plate to the hole in the aperture plate and to the aperture base plate assembly.
The pin should drop through easily. If the pin does not drop in, select the C-Arm
tab on the System Diagnostics screen.
4. In the Commands section at the lower right, type 10 or -10 in the Move Relative field to re-align the
aperture plate. You may need to visually check that you are moving the aperture plate in the correct
direction relative to the base plate hole.
The pin should now drop down through the hole in the back left of the aperture
plate down through the alignment block and into the base plate hole with
minimum force. If the pin still does not align with the base plate hole, continue to
move the aperture plate ten steps at a time until alignment is achieved.
5. Press or click F2 to turn on X-rays.
6. Locate the (2) 10-32 x .37L Hex should bolt screws on the bottom sled bracket of the Filter Drum
Sled assembly, and loosen them by 1/2 turn.
7. Adjust the rear sled adjustment to move the whole assembly left or right in the beam (figure 119).
The goal is to achieve the maximum voltage on the left side of the survey screen.
8. Press or click F2 to turn off x-rays.
9. Tighten the (2) 10-32 x .37L Hex bottom sled screws.
10. Perform skew alignment (See Perform Skew Filter Drum Base Alignment).
11. Repeat the lateral and skew process until beam is flat and at maximum amplitude.
12. Remove the pin and block when done.
15. Restart the scan and verify that the leading edge of L4 spine is displayed between lines 30 – 35 in the
acquiring image.
16. Click the Restart Scan button and let the arm reposition.
17. Click Cancel scan.
18. Press center table on the control panel.
19. Move the phantom back to Laser Cross hairs at center table position.
20. Click Perform Exam and select the test patient biography just used.
21. Select AP Lumbar as the scan type.
22. Deselect Use Default Scan Mode.
23. Click Next.
24. Select Array mode.
25. Change length of scan to 6.0 inch. Click Start Scan.
26. Verify that L4 is displayed in the acquiring image between lines 30 – 35 and the spine is centered in
the image.
Caution Any adjustment of this potentiometer affects the QC highs and lows. You must
run detector flattening and recalibrate the machine after adjusting the A/D
gain. DO NOT ADJUST THE A/D GAIN UNLESS ABSOLUTELY
NECESSARY AS DEFINED in Step 9 in the section that follows (Verify and
Adjust the A/D Gain).
This procedure flattens the X-ray beam for each scan mode and uses the metal edges of the table to
calculate the center table position in the x-axis.
Note All covers, table mat, etc., normally in the X-ray path, MUST be on the Scanner
before running Beam Flattening.
1. Restart the Horizon software in service mode (if not already). Press the Center Table button and
turn on the laser.
2. Using a DXA Quality Control Phantom (Part Number ASY-01564), place the phantom on end with
the flattening target facing up. Orient the phantom so the metal edge is to the right (head end) and
tissue to the left (foot end) as viewed from the front of the table. Align the phantom flattening target
to the laser crosshair.
3. Start the X-Ray Survey Utility by selecting Utilities > Service Utilities > X-Ray Survey.
4. Set Pulse to 1, High gain to 2, Low gain to 1, X-ray Mode to 3 and Aperture to 7 (for A, C, and W
models).
5. For Display, click Graph.
6. Click X-Rays (F2) to turn X-Rays on.
7. Check that the phantom covers the entire beam. This is critical. If phantom does not cover the entire
beam, move it until it does. Keep the phantom as straight as possible.
8. Turn off X-rays with the <F2> key and press <Esc> or click Close to exit X-Ray Survey.
9. Reboot the PC, logon as Field Service, and restart the software in Service Mode to return to Horizon
Main Menu screen.
10. Select Utilities > Service Utilities > Detector Flattening.
11. Click Flattening –Basic First.
Figure 339 Flattening –Basic First
Note 1. If the Horizon being installed is to replace an existing QDR fan beam
system or if the customer has multiple systems, then a cross calibration
must first be performed to ensure that any longitudinal studies begun on
the QDR being removed can safely be continued on the new Horizon. Refer
to the cross-calibration procedure in Section 3.10 Fan Beam Cross-
Calibration Gold Standard System on Page 164 before uninstalling the
existing QDR. The Cross-Calibration procedure applies to fan beam
systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
3. Compare values displayed with actual values printed on the phantom. If there is a discrepancy,
check to make sure you have the phantom that shipped with the unit. Check with Hologic Technical
Support if needed.
7.4.3 Perform BCA Calibration Scans and Table Top Radiographic Uniformity (A, W
and Wi)
This procedure verifies proper X-Ray beam alignment and the uniformity of the X-Ray transparency of
entire tabletop in a manual process (for Horizon A, W and Wi models). BCA Calibration Scans must be
performed on the system prior to verifying the uniformity test scan.
Note On Models A, W, and Wi, after Flattening is performed and 30 minute cool
down is expired, perform a Daily QC scan. Automatically the system will
perform BCA calibration scans on the QC Phantom. After the calibration scans
are complete the system will inform you to remove the QC Phantom and clear
the table to perform the Table Top Uniformity Scan test. This test will report the
results automatically and must pass with the Hi and Low Air SD of less than
2.0.
If the system is passing normally, skip to next step.
If the system is not passing with less than SD 2.0, verify that table alignments
and beam alignments are correct, adjust as necessary and run Flattening again.
1. If necessary, restart the software in Service Mode.
2. Type WBAIRQC in the Patient Biography. Enter the serial number of the unit in the Patient ID field.
Make sure that only the pad is on the table. Clear the table of anything else.
3. Perform one Whole Body air scan (a scan with nothing on the table). Do NOT interrupt the scan for
any reason. When the scan completes, select Exit Exam. Do not analyze the scan that was just
performed.
4. From the Main Menu screen, select Utilities > Service Utilities > Table Top Radiographic
Uniformity.
5. Select the Whole Body scan and click the Next>> button.
6. Table Top Radiographic Uniformity results tabs are now displayed. Select the Low Air Tab.
7. If the Global Stats S.D. (2) is less than 2.0, the instrument is properly aligned. If the Global Stats S.D.
is greater than or equal to 2.0, then the machine is not aligned properly. Check the following:
• If the image appears “streaky” or “banded”, check for loose wires or other debris between the table
and the tank.
• Check the Aperture Calibration.
• Check AC Line voltage.
• Check +/- 15 V Power Supply.
• Check for a dedicated line.
• Check Detector Flattening.
• Re-run daily QC (detailed in MAN-03644 Horizon QDR Series User Guide).
Print a copy of the test results and include it with the other paperwork being returned to Hologic.
After realignment, repeat the Table Top Radiographic Uniformity test. If the global S.D. is less than 2.0,
you are done. If the global S.D. if equal or greater than 2.0, print a copy of the test results and include it
with the other paperwork being returned to Hologic.
If the global is equal to or greater than 2.0, the non-uniformity may be attributed to the table itself. If so,
continue with this procedure. Make sure that the machine has body composition loaded. If it does, then
follow the appropriate procedure below in Section 0.9 Instruments using Body Composition Analysis (BCA)
on Page 13.
3. Center the table, place the phantom on the table and align with the laser. The sides of the phantom
should be parallel to the table. Use a ruler if necessary.
4. Click Perform Exam from the application main screen.
5. At the Select a Patient for this Exam window, select the phantom, and click OK.
6. At the Patient Confirmation screen, click OK.
7. At Select Scan Type screen, select AP/Lateral, uncheck Use Default Scan Mode, and click Next>>.
8. At the Select AP Lumbar Spine Scan Mode for AP/Lateral Exam screen, select Array (a), and click
Next>>.
9. At the Select Lateral Scan Mode for AP/Lateral Exam screen, select Fast Array (f), and click Next.
10. Recheck phantom alignment with laser.
Note Do not move the arm or table, otherwise the test will have to be started over.
11. Place two sheets of paper against the left and right edges of the phantom (Figure 416) When done,
click Continue.
12. On the Scan Parameters screen, set the scan length to 6.0 in., click Enter and click Start Scan.
13. When scan has finished, select Spine as the analysis method, and click Next>>.
14. Analyze the scan and click Start Position.
15. Refer to Figure 439 Verify that the angle displayed under the image (near the bottom of the screen)
is less than ± 2 degrees.
16. Using the sheets of paper as a guide, carefully rotate the phantom on its back edge towards the C-
Arm so that the bone segments are facing the rear of the table as shown. Except for the rotation, do
not move the phantom in any other direction.
17. Click Close. When the Continuously Press Enable message appears, press and hold the Lateral
Enable switch until the C-Arm and table move to the Lateral position.
18. On the Scan Parameters screen, click Start Scan.
19. When the “Continuously Press Enable” message appears, press and hold the Lateral Enable switch
until the C-Arm and table move back to the AP position.
20. From the Exit Exam screen, select Analyze Scan.
21. Analyze the scan (refer to user guide for Lateral analysis) and close the screen.
22. Click Exit to close the Exit Analysis screen.
23. Restart APEX software in service mode (if not already). From the APEX Main window, select
Utilities > Service Utilities > Field Service Calibration.
24. Select Highs/Lows for Array Spine Phantom.
25. Select the AP scan just performed and click Next.
26. Click Continue on the Calculated highs and lows page.
27. Select Highs/Lows for Array Lateral Phantom.
28. Click Continue on the Calculated highs and lows page.
29. Click Temp Write to write temporary results and click Continue to exit the write dialog box.
Note For Horizon system with version APEX 5.6 and higher, only 10 reproducibility
scans for calibrations are required. Earlier versions required 20 scans. Acquire 5
scans with a period of 15 between groups.
14. Analyze the first scan manually selecting Lumbar Spine for the analysis method.
15. Note the scan #.
16. Click the Close button to exit analysis.
17. From the Main Menu screen, select Utilities > Service Utilities > Auto Analyze.
18. Click unanalyzed Scans and select the 10 scans, click Compare radio button and then click Next.
19. Now, select the QC Scan performed in step Section 3.11.2 Perform QC Scan on Spine Phantom, (C, Ci)
With BCA Calibration Scans and Table Top Radiographic Uniformity (A, W and Wi) on Page 199.
20. Click Next.
21. The 10 scans will now be compared with the auto-analyzed QC Scan.
7.4.7 Calibration (Area, BMD, and BMC) with ACF and BCF Correction Factors
Calibration for Area, BMD, and BMC is accomplished in three stages:
• Array Scan Thickness Measurement & Calibration
• Calibration of Area and BMC for Array Scan Modes
• Adding Array AP scans to the QC database
Follow the procedure, in order, and exactly as shown, for each stage of the calibration.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam system
or if the customer has multiple systems, then a cross-calibration must be
performed to ensure that any longitudinal studies begun on the QDR being
removed can safely be continued on the new Horizon. Refer to the cross-
calibration procedure in Section 3.10 Fan Beam Cross-Calibration Gold
Standard System on Page 164 before de-installing the existing QDR. The Cross
Calibration procedure applies to fan beam systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
1. From the Main Menu screen, select Utilities > Service Utilities > Field Service Calibration.
2. Click the ACF/BCF button.
3. Select the 11 reproducibility scans performed in previous step.
4. Click Next
5. Click Continue at the Successfully Calculated dialog box.
6. At the Field Service Calibration dialog box, click the Write button to write the values to arrc.txt file.
7. Click Continue to return to the Main Menu.
Using a test patient, perform at least one scan in each of the scan modes to verify that machine is fully
functional in all modes of operation. Verify that there are no electrical or mechanical problems during each
scan mode.
[ ] Perform an AP Spine Array, Fast Array, and Express scan.
[ ] Perform a Left Hip Array, Fast Array, and Express scan.
[ ] Perform a Whole Body scan (A, W, and Wi only).
[ ] Perform a Forearm scan.
[ ] Perform the IVA Scan QC Procedure described below.
17. From the Scan Parameters screen set the scan length to 10.0 in.
18. Click the Start Scan button to run a single-energy AP scan.
19. From the Hologic Analyze screen, perform the next 3 steps.
Note Intensity may be adjusted to visualize scan.
20. Count the 12 steps associated with the Step Wedge Phantom.
The number of steps visible on the Step Phantom should be 11 and 1 more for the table.
21. Skip this step on models Ci and Wi. By clicking the + sign to increase the magnification and using
the scroll bar to examine the x-ray test pattern image on the screen, you should be able to see three
lines for the 0.8 test pattern image.
22. When completed, click Cancel.
A radiation dose measurement must be performed at installation time, and done routinely whenever the
machine is serviced. This test requires the use of a Victoreen Model 450P, Unfors, or equivalent measuring
device set to measure dose over time (mR/Hr).
Note The dose measurements must be done with the machine fully assembled and the
table mat in place.
1. If using a Victoreen, place the meter in the center of the table approximately ½ in. to the right of the
laser crosshair. If using an Unfors, place the detector on the crosshair and raised 2.0 in. above the
table surface.
2. Click the Perform Exam button, create a test patient biography, and select the Array spine scan.
3. Set scan length for 9.5 in. and click the Start scan button.
4. The meter is now scanned.
5. Observe the scan to verify that the entire chamber of the meter is included in the scan.
6. Record the result from the meter.
Note The arrayExplorer spine scan dose should be less than 20mrad (200µGy).
A radiation scatter measurement must be performed at the time of installation. You must use a survey-
type radiation meter (Victoreen 450P or equivalent).
Perform the following:
1. Center the table and C-arm.
2. Place the spine phantom in the center of the table and align using the laser.
3. Start an AP Spine scan.
4. Using the <Esc> key, stop the scan when it is between the two middle vertebrae.
5. Start the X-ray Survey Utility by selecting Utilities > Service Utilities > X-ray Survey.
6. Using the survey meter, measure the scatter radiation at a distance of 2 meters from the phantom on
a horizontal plane all around the unit. Move the meter slowly.
Note If you cannot measure the scatter radiation at a distance of 2 meters because of
space restrictions, you can measure at 1 meter and divide the result by 4. This
reading must be less than 1.0 mrad/h (10 µGy/h) at 2 meters (approx. 80 in.) in
array AP scan mode.
7. Record the highest reading in the service report and then turn off X-rays by clicking on X-ray (F2).
Note Enter this information into the Radiation Measurement Report (Section 0.1.10
Radiation Measurement Report on Page 5).
7.10 Perform QC
Once the machine has been fully assembled and calibrated, at least one QC scan should be performed.
• Perform the daily QC procedure and enter the scan into the QC database.
Caution Make sure the pin and block have been removed before performing Aperture
Calibration.
1. From the Hologic main window, click Utilities > Service Utilities > System Diagnostics.
2. Click the C-Arm tab.
3. In the upper right corner of the screen, click Calibrate Aperture (Figure 375) to move the aperture to
its right and left physical limits.
Figure 342 Select Calibrate Aperture
The x-rays turn on as the aperture plate moves left to right. As the aperture moves through the beam path,
the signal is captured and displayed. The chart will identify the signal levels for each aperture specific
encoder positions.
The vertical lines on the final output should correspond to the center of the aperture openings. The data
contributes to software positioning and aperture identification.
Slope Value
4. Click Continue.
As the potentiometer moves in both directions, a linear change should be displayed in the Aperture
Calibration Plot (Figure 377).
5. Click Continue to open the Aperture Calibration Peaks screen (Figure 378)
Figure 345 Aperture Calibration Peaks
• The green lines should be centered at the peak of the signal for each aperture if the aperture is
moving correctly. This screen also gives an indication of whether the aperture plate is moving
erratically or binding.
• The red lines are typically displayed in pairs to reflect the pairing of the apertures. In between the
pairs, the beam should be off (maximum dark).
6. Click Continue to open the Aperture Calibration Data screen (Figure 379) and review the offset
numbers.
Figure 346 Aperture Calibration Data
Caution Be sure to restart the software in Service Mode after calibrating the aperture, so
the software can access the new data file. If the system is not rebooted, the
software will not be able to find a selected aperture correctly.
This procedure flattens the X-ray beam for each scan mode and uses the metal edges of the table to
calculate the center table position in the x-axis.
Note All covers, table mat, etc., normally in the X-ray path, MUST be on the Scanner
before running Beam Flattening.
1. Restart the Horizon software in service mode (if not already). Press the Center Table button and
turn on the laser.
2. Using a DXA Quality Control Phantom (Part Number ASY-01564), place the phantom on end with
the flattening target facing up. Orient the phantom so the metal edge is to the right (head end) and
tissue to the left (foot end) as viewed from the front of the table. Align the phantom flattening target
to the laser crosshair.
Figure 347 Correct Flattening Position
3. Start the X-Ray Survey Utility by selecting Utilities > Service Utilities > X-Ray Survey.
4. Set Pulse to 1, High gain to 2, Low gain to 1, X-ray Mode to 3 and Aperture to 7 (for A, C, and W
models).
5. For Display, click Graph.
6. Click X-Rays (F2) to turn X-Rays on.
7. Check that the phantom covers the entire beam. This is critical. If phantom does not cover the entire
beam, move it until it does. Keep the phantom as straight as possible.
8. Turn off X-rays with the <F2> key and press <Esc> or click Close to exit X-Ray Survey.
9. Reboot the PC, logon as Field Service, and restart the software in Service Mode to return to Horizon
Main Menu screen.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam
system or if the customer has multiple systems, then a cross calibration
must first be performed to ensure that any longitudinal studies begun on
the QDR being removed can safely be continued on the new Horizon. Refer
to the cross-calibration procedure in Section 3.10 Fan Beam Cross-
Calibration Gold Standard System on Page 164 before uninstalling the
existing QDR. The Cross-Calibration procedure applies to fan beam
systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
3. Compare values displayed with actual values printed on the phantom. If there is a discrepancy,
check to make sure you have the phantom that shipped with the unit. Check with Hologic Technical
Support if needed.
8.3.3 Perform BCA Calibration Scans and Table Top Radiographic Uniformity (A, W
and Wi)
This procedure verifies proper X-Ray beam alignment and the uniformity of the X-Ray transparency of
entire tabletop in a manual process (for Horizon A, W and Wi models). BCA Calibration Scans must be
performed on the system prior to verifying the uniformity test scan.
Note On Models A, W, and Wi, after Flattening is performed and 30 minute cool
down is expired, perform a Daily QC scan. Automatically the system will
perform BCA calibration scans on the QC Phantom. After the calibration scans
are complete the system will inform you to remove the QC Phantom and clear
the table to perform the Table Top Uniformity Scan test. This test will report the
results automatically and must pass with the Hi and Low Air SD of less than
2.0.
If the system is passing normally, skip to next step.
If the system is not passing with less than SD 2.0, verify that table alignments
and beam alignments are correct, adjust as necessary and run Flattening again.
1. If necessary, restart the software in Service Mode.
2. Type WBAIRQC in the Patient Biography. Enter the serial number of the unit in the Patient ID field.
Make sure that only the pad is on the table. Clear the table of anything else.
3. Perform one Whole Body air scan (a scan with nothing on the table). Do NOT interrupt the scan for
any reason. When the scan completes, select Exit Exam. Do not analyze the scan that was just
performed.
4. From the Main Menu screen, select Utilities > Service Utilities > Table Top Radiographic
Uniformity.
5. Select the Whole Body scan and click the Next>> button.
6. Table Top Radiographic Uniformity results tabs are now displayed. Select the Low Air Tab.
7. If the Global Stats S.D. (2) is less than 2.0, the instrument is properly aligned. If the Global Stats S.D.
is greater than or equal to 2.0, then the machine is not aligned properly. Check the following:
• If the image appears “streaky” or “banded”, check for loose wires or other debris between the table
and the tank.
• Check the Aperture Calibration.
• Check AC Line voltage.
• Check +/- 15 V Power Supply.
• Check for a dedicated line.
• Check Detector Flattening.
• Re-run daily QC (detailed in MAN-03644 Horizon QDR Series User Guide).
Print a copy of the test results and include it with the other paperwork being returned to Hologic.
After realignment, repeat the Table Top Radiographic Uniformity test. If the global S.D. is less than 2.0,
you are done. If the global S.D. if equal or greater than 2.0, print a copy of the test results and include it
with the other paperwork being returned to Hologic.
If the global is equal to or greater than 2.0, the non-uniformity may be attributed to the table itself. If so,
continue with this procedure. Make sure that the machine has body composition loaded. If it does, then
follow the appropriate procedure below in Section 0.9 Instruments using Body Composition Analysis (BCA)
on Page 13.
3. Center the table, place the phantom on the table and align with the laser. The sides of the phantom
should be parallel to the table. Use a ruler if necessary.
4. Click Perform Exam from the application main screen.
5. At the Select a Patient for this Exam window, select the phantom, and click OK.
6. At the Patient Confirmation screen, click OK.
7. At Select Scan Type screen, select AP/Lateral, uncheck Use Default Scan Mode, and click Next>>.
8. At the Select AP Lumbar Spine Scan Mode for AP/Lateral Exam screen, select Array (a), and click
Next>>.
9. At the Select Lateral Scan Mode for AP/Lateral Exam screen, select Fast Array (f), and click Next.
10. Recheck phantom alignment with laser.
Note Do not move the arm or table, otherwise the test will have to be started over.
11. Place two sheets of paper against the left and right edges of the phantom (Figure 416) When done,
click Continue.
12. On the Scan Parameters screen, set the scan length to 6.0 in., click Enter and click Start Scan.
13. When scan has finished, select Spine as the analysis method, and click Next>>.
14. Analyze the scan and click Start Position.
15. Refer to Figure 439 Verify that the angle displayed under the image (near the bottom of the screen)
is less than ± 2 degrees.
16. Using the sheets of paper as a guide, carefully rotate the phantom on its back edge towards the C-
Arm so that the bone segments are facing the rear of the table as shown. Except for the rotation, do
not move the phantom in any other direction.
17. Click Close. When the Continuously Press Enable message appears, press and hold the Lateral
Enable switch until the C-Arm and table move to the Lateral position.
18. On the Scan Parameters screen, click Start Scan.
19. When the “Continuously Press Enable” message appears, press and hold the Lateral Enable switch
until the C-Arm and table move back to the AP position.
20. From the Exit Exam screen, select Analyze Scan.
21. Analyze the scan (refer to user guide for Lateral analysis) and close the screen.
22. Click Exit to close the Exit Analysis screen.
23. Restart APEX software in service mode (if not already). From the APEX Main window, select
Utilities > Service Utilities > Field Service Calibration.
24. Select Highs/Lows for Array Spine Phantom.
25. Select the AP scan just performed and click Next.
26. Click Continue on the Calculated highs and lows page.
27. Select Highs/Lows for Array Lateral Phantom.
28. Click Continue on the Calculated highs and lows page.
29. Click Temp Write to write temporary results and click Continue to exit the write dialog box.
Note For Horizon system with version APEX 5.6 and higher, only 10 reproducibility
scans for calibrations are required. Earlier versions required 20 scans. Acquire 5
scans with a period of 15 between groups.
14. Analyze the first scan manually selecting Lumbar Spine for the analysis method.
15. Note the scan #.
16. Click the Close button to exit analysis.
17. From the Main Menu screen, select Utilities > Service Utilities > Auto Analyze.
18. Click unanalyzed Scans and select the 10 scans, click Compare radio button and then click Next.
19. Now, select the QC Scan performed in step Section 3.11.2 Perform QC Scan on Spine Phantom, (C, Ci)
With BCA Calibration Scans and Table Top Radiographic Uniformity (A, W and Wi) on Page 199.
20. Click Next.
21. The 10 scans will now be compared with the auto-analyzed QC Scan.
8.3.7 Calibration (Area, BMD, and BMC) with ACF and BCF Correction Factors
Calibration for Area, BMD, and BMC is accomplished in three stages:
• Array Scan Thickness Measurement & Calibration
• Calibration of Area and BMC for Array Scan Modes
• Adding Array AP scans to the QC database
Follow the procedure, in order, and exactly as shown, for each stage of the calibration.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam system
or if the customer has multiple systems, then a cross-calibration must be
performed to ensure that any longitudinal studies begun on the QDR being
removed can safely be continued on the new Horizon. Refer to the cross-
calibration procedure in Section 3.10 Fan Beam Cross-Calibration Gold
Standard System on Page 164 before de-installing the existing QDR. The Cross
Calibration procedure applies to fan beam systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
1. From the Main Menu screen, select Utilities > Service Utilities > Field Service Calibration.
2. Click the ACF/BCF button.
3. Select the 11 reproducibility scans performed in previous step.
4. Click Next
5. Click Continue at the Successfully Calculated dialog box.
6. At the Field Service Calibration dialog box, click the Write button to write the values to arrc.txt file.
7. Click Continue to return to the Main Menu.
Using a test patient, perform at least one scan in each of the scan modes to verify that machine is fully
functional in all modes of operation. Verify that there are no electrical or mechanical problems during each
scan mode.
[ ] Perform an AP Spine Array, Fast Array, and Express scan.
[ ] Perform a Left Hip Array, Fast Array, and Express scan.
[ ] Perform a Whole Body scan (A, W, and Wi only).
[ ] Perform a Forearm scan.
[ ] Perform the IVA Scan QC Procedure described below.
17. From the Scan Parameters screen set the scan length to 10.0 in.
18. Click the Start Scan button to run a single-energy AP scan.
19. From the Hologic Analyze screen, perform the next 3 steps.
Note Intensity may be adjusted to visualize scan.
20. Count the 12 steps associated with the Step Wedge Phantom.
The number of steps visible on the Step Phantom should be 11 and 1 more for the table.
21. Skip this step on models Ci and Wi. By clicking the + sign to increase the magnification and using
the scroll bar to examine the x-ray test pattern image on the screen, you should be able to see three
lines for the 0.8 test pattern image.
22. When completed, click Cancel.
A radiation dose measurement must be performed at installation time, and done routinely whenever the
machine is serviced. This test requires the use of a Victoreen Model 450P, Unfors, or equivalent measuring
device set to measure dose over time (mR/Hr).
Note The dose measurements must be done with the machine fully assembled and the
table mat in place.
1. If using a Victoreen, place the meter in the center of the table approximately ½ in. to the right of the
laser crosshair. If using an Unfors, place the detector on the crosshair and raised 2.0 in. above the
table surface.
2. Click the Perform Exam button, create a test patient biography, and select the Array spine scan.
3. Set scan length for 9.5 in. and click the Start scan button.
4. The meter is now scanned.
5. Observe the scan to verify that the entire chamber of the meter is included in the scan.
6. Record the result from the meter.
Note The arrayExplorer spine scan dose should be less than 20mrad (200µGy).
A radiation scatter measurement must be performed at the time of installation. You must use a survey-
type radiation meter (Victoreen 450P or equivalent).
Perform the following:
1. Center the table and C-arm.
2. Place the spine phantom in the center of the table and align using the laser.
3. Start an AP Spine scan.
4. Using the <Esc> key, stop the scan when it is between the two middle vertebrae.
5. Start the X-ray Survey Utility by selecting Utilities > Service Utilities > X-ray Survey.
6. Using the survey meter, measure the scatter radiation at a distance of 2 meters from the phantom on
a horizontal plane all around the unit. Move the meter slowly.
Note If you cannot measure the scatter radiation at a distance of 2 meters because of
space restrictions, you can measure at 1 meter and divide the result by 4. This
reading must be less than 1.0 mrad/h (10 µGy/h) at 2 meters (approx. 80 in.) in
array AP scan mode.
7. Record the highest reading in the service report and then turn off X-rays by clicking on X-ray (F2).
Note Enter this information into the Radiation Measurement Report (Section 0.1.10
Radiation Measurement Report on Page 5).
8.9 Perform QC
Once the machine has been fully assembled and calibrated, at least one QC scan should be performed.
• Perform the daily QC procedure and enter the scan into the QC database.
Note If installed, the aperture assembly must be removed in order to access the filter
drum. (Required Tool TLS-01886.)
1. The DXA system must be powered on. Exit APEX without shut down, and start APEX in Service
Mode. On the filter drum controller board, press SW1 to reset the board.
2. Remove the aperture assembly to access the filter drums.
a. Remove two screws for motor bracket.
b. Remove four screws (2 front, 2 rear) for filter drum.
c. Remove aperture assembly.
d. Unplug two cables on the board that go to aperture.
e. Remove strain relief.
3. On the new filter drum controller board, move the Run/Calibrate DIP switch (SW2-4) to the ON
position. ON position is towards connector J3. ON is to the right in the orientation shown in
Figure 351. OFF is to the left (Figure 351).
4. Momentarily press the CAL push-button switch (SW3), (Figure 351). Both drums will rotate slowly
for three revolutions, and then power is removed from the drum motors so the drums can be
rotated freely.
5. When the drums have stopped rotating, manually rotate the Inner and Outer drums to the
calibration position, where the slots in the hub of each drum lines up with the slot in the end plate
at the rear drum motor end plate (Figure 107).
Note The Home position is not the same as the calibration position. The leading edge
of outer drum aluminum is at top center of rotation with leading edge of the
brass (inner drum) just slightly behind. It is important that this visual check be
made following calibration. If not, repeat the alignment procedure - steps 3
through 10.
Figure 352 Drum Home Position
Note The calibration key should be perpendicular to and flush with the end plate. The
end of the key with the flat surface should be towards the motor and the stepped
side facing down, so that it engages the slot in the end plate and slots in both
drums
Figure 353 Alignment Tool Placement
1. Insert the calibration key (part number TLS-01886 (Figure 354) into the slots in the hubs of both
drums and the rear end plate, forcing the slots in the drums to align with the slot in the rear end
plate (Figure 355).
Figure 354 Alignment Key
Note The calibration key should be perpendicular to and flush with the end plate.
The end of the key with the flat surface should be towards the motor and the
stepped side facing down so that it engages the slot in the end plate and slots in
both drums (Figure 355).
Note he Home position is not the same as the calibration position. The leading edge of
outer drum aluminum is at top center of rotation with leading edge of the Brass
(inner drum) just slightly behind. It is important that this visual check be made
following calibration. If not, repeat the alignment procedure - steps 3 through
11.
3. Set the following parameters in the Device Control Screen (Figure 359 and Figure 360) to the values
shown:
• High Gain = 2
• Low Gain = 1
• X-Ray Mode = 3
• Aperture = 7
4. Verify that the screen shows Pulse equal to 1.
5. Select Filter Wheel to ON to ensure the filter drum is spinning.
Note To turn X-rays Off and On you can press the F2 Function key on the keyboard
or by using the mouse toggle the F2 button on the screen.
7. Insert the X-ray beam alignment tool, TLS-05376, into the slot on the C-arm (Figure 362).
Note The handle of the alignment tool protrudes from the C-arm to prevent
inadvertently leaving the tool in the C-arm after performing an alignment.
8. Press the keyboard F2 function key or with the mouse toggle the F2 button on the screen to turn X-
rays On.
9. Check the voltage output on the Device Control Screen (Figure 363).
Note Regardless of the initial voltage readings, the goal of the adjustment is to center
the filter drum and aperture assembly in the middle of the X-rays beam to
achieve maximum amplitude with a uniform flat signal across the top of the
graph.
If the arm is deflected in either direction, when the filter drum assembly is
properly aligned, the voltage drop should be the same in both directions.
On Horizon A systems use Emergency Motion to raise the table (Figure 6) to
access the area underneath the table top.
Caution Do not drive the table to its maximum raised position. Raising the table to its
maximum height at abnormal speed can damage hardware.
2. From the Hologic main menu, select Utilities > Service Utilities >System Diagnostics > Detectors
tab and select aperture 7 on Horizon C, W, and A models only.
Note Do not use the arrow up/down function to select Aperture 7. Instead, use the
mouse to highlight the displayed aperture number, type 7 and press Enter.
3. Use the alignment block TLS-02698, and alignment pin 099-0111 to align the hole in the filter drum
base plate to the hole in the aperture plate and to the aperture base plate assembly.
The pin should drop through easily. If the pin does not drop in, select the C-Arm
tab on the System Diagnostics screen.
4. In the Commands section at the lower right, type 10 or -10 in the Move Relative field to re-align the
aperture plate. You may need to visually check that you are moving the aperture plate in the correct
direction relative to the base plate hole.
The pin should now drop down through the hole in the back left of the aperture
plate down through the alignment block and into the base plate hole with
minimum force. If the pin still does not align with the base plate hole, continue to
move the aperture plate ten steps at a time until alignment is achieved.
5. Press or click F2 to turn on X-rays.
6. Locate the (2) 10-32 x .37L Hex should bolt screws on the bottom sled bracket of the Filter Drum
Sled assembly, and loosen them by 1/2 turn.
7. Adjust the rear sled adjustment to move the whole assembly left or right in the beam (figure 119).
The goal is to achieve the maximum voltage on the left side of the survey screen.
8. Press or click F2 to turn off x-rays.
9. Tighten the (2) 10-32 x .37L Hex bottom sled screws.
10. Perform skew alignment (See Perform Skew Filter Drum Base Alignment).
11. Repeat the lateral and skew process until beam is flat and at maximum amplitude.
12. Remove the pin and block when done.
2. Ensure that all components can freely adjust by turning the sled Skew adjustment screw. Loosen
filter drum base screws slightly if necessary (Figure 368).
Note Deflect the arm slightly to the left and right. The voltage should decrease when
the outside the beam path.
5. Adjust the Filter Drum Assembly Skew (Figure 370) until the amplitude displayed on the Device
Control Screen is flat.
Figure 370 Voltage Amplitude on Device Control Screen
If a detector at the end of the charts (e.g., Detector 44, Figure 371) has on an amplitude less than 6.0V and
less than the detector on the opposite side (e.g., Detector 171, Figure 314), the detector must be adjusted
front to back, so that both detectors display the same voltage.
Figure 371 Details Detector 44
8. Tighten screws.
Caution Make sure the pin and block have been removed before performing Aperture
Calibration.
1. From the Hologic main window, click Utilities > Service Utilities > System Diagnostics.
2. Click the C-Arm tab.
3. In the upper right corner of the screen, click Calibrate Aperture (Figure 375) to move the aperture to
its right and left physical limits.
Figure 375 Select Calibrate Aperture
The x-rays turn on as the aperture plate moves left to right. As the aperture moves through the beam path,
the signal is captured and displayed. The chart will identify the signal levels for each aperture specific
encoder positions.
The vertical lines on the final output should correspond to the center of the aperture openings. The data
contributes to software positioning and aperture identification.
Slope Value
4. Click Continue.
As the potentiometer moves in both directions, a linear change should be displayed in the Aperture
Calibration Plot (Figure 377).
5. Click Continue to open the Aperture Calibration Peaks screen (Figure 378)
Figure 378 Aperture Calibration Peaks
• The green lines should be centered at the peak of the signal for each aperture if the aperture is
moving correctly. This screen also gives an indication of whether the aperture plate is moving
erratically or binding.
• The red lines are typically displayed in pairs to reflect the pairing of the apertures. In between the
pairs, the beam should be off (maximum dark).
6. Click Continue to open the Aperture Calibration Data screen (Figure 379) and review the offset
numbers.
Figure 379 Aperture Calibration Data
Caution Be sure to restart the software in Service Mode after calibrating the aperture, so
the software can access the new data file. If the system is not rebooted, the
software will not be able to find a selected aperture correctly.
Caution Any adjustment of this potentiometer affects the QC highs and lows. You must
run detector flattening and recalibrate the machine after adjusting the A/D
gain. DO NOT ADJUST THE A/D GAIN UNLESS ABSOLUTELY
NECESSARY AS DEFINED in Step 9 in the section that follows (Verify and
Adjust the A/D Gain).
8.15.1 Principles
Radiation leakage from the HVPS/S results from a defect in the lead liner of the tank or its cover. If there is
a hole in the lead liner, the resulting leakage will produce a pencil beam of X-rays. If there is a crack in the
lead liner, the resulting leakage will produce a fan beam. In either case, the initial survey must be taken
close to and almost touching the tank with the survey meter probe. If a leak is found, the meter indicates a
spike in the reading. If this occurs, you must then move the probe one-meter (approximately 40 in.) along
the beam from that spot to determine if the leakage is within acceptable levels.
2. Start the Utility by selecting Utilities > Service Utilities > X-Ray Survey.
3. Change X-Ray Mode to 3 and Aperture to 7 to lock in the changes.
Warning: X-rays are generated during this procedure. Keep hands, head and other body
parts out of beam.
1. Position the survey meter one meter (approx. 40 in.) from the tank and move it all around the
outside of the instrument.
2. The reading must be 1.0 mrad/h (10 µGy/h)or less. Record the highest reading in the service report.
3. Press the <F2> key to turn off X-rays, remove the tool, and replace the tank cover.
Note Enter this information into the Radiation Measurement Report (CSD-0042-
F07).
This procedure flattens the X-ray beam for each scan mode and uses the metal edges of the table to
calculate the center table position in the x-axis.
Note All covers, table mat, etc., normally in the X-ray path, MUST be on the Scanner
before running Beam Flattening.
1. Restart the Horizon software in service mode (if not already). Press the Center Table button and
turn on the laser.
2. Using a DXA Quality Control Phantom (Part Number ASY-01564), place the phantom on end with
the flattening target facing up. Orient the phantom so the metal edge is to the right (head end) and
tissue to the left (foot end) as viewed from the front of the table. Align the phantom flattening target
to the laser crosshair.
Figure 382 Correct Flattening Position
3. Start the X-Ray Survey Utility by selecting Utilities > Service Utilities > X-Ray Survey.
4. Set Pulse to 1, High gain to 2, Low gain to 1, X-ray Mode to 3 and Aperture to 7 (for A, C, and W
models).
5. For Display, click Graph.
6. Click X-Rays (F2) to turn X-Rays on.
7. Check that the phantom covers the entire beam. This is critical. If phantom does not cover the entire
beam, move it until it does. Keep the phantom as straight as possible.
8. Turn off X-rays with the <F2> key and press <Esc> or click Close to exit X-Ray Survey.
9. Reboot the PC, logon as Field Service, and restart the software in Service Mode to return to Horizon
Main Menu screen.
10. Select Utilities > Service Utilities > Detector Flattening.
11. Click Flattening –Basic First.
Figure 383 Flattening –Basic First
This procedure will take approximately 15 minutes on a Ci and Wi, 25 minutes on a C and W, and
50 minutes on an A model system.
13. After this procedure, the X-Ray Source Assembly will be hot. You must wait for 30 minutes before
performing a Field Service Calibration of the machine.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam
system or if the customer has multiple systems, then a cross calibration
must first be performed to ensure that any longitudinal studies begun on
the QDR being removed can safely be continued on the new Horizon. Refer
to the cross-calibration procedure in Section 3.10 Fan Beam Cross-
Calibration Gold Standard System on Page 164 before uninstalling the
existing QDR. The Cross-Calibration procedure applies to fan beam
systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
3. Compare values displayed with actual values printed on the phantom. If there is a discrepancy,
check to make sure you have the phantom that shipped with the unit. Check with Hologic Technical
Support if needed.
8.16.3 Perform BCA Calibration Scans and Table Top Radiographic Uniformity (A, W
and Wi)
This procedure verifies proper X-Ray beam alignment and the uniformity of the X-Ray transparency of
entire tabletop in a manual process (for Horizon A, W and Wi models). BCA Calibration Scans must be
performed on the system prior to verifying the uniformity test scan.
Note On Models A, W, and Wi, after Flattening is performed and 30 minute cool
down is expired, perform a Daily QC scan. Automatically the system will
perform BCA calibration scans on the QC Phantom. After the calibration scans
are complete the system will inform you to remove the QC Phantom and clear
the table to perform the Table Top Uniformity Scan test. This test will report the
results automatically and must pass with the Hi and Low Air SD of less than
2.0.
If the system is passing normally, skip to next step.
If the system is not passing with less than SD 2.0, verify that table alignments
and beam alignments are correct, adjust as necessary and run Flattening again.
1. If necessary, restart the software in Service Mode.
2. Type WBAIRQC in the Patient Biography. Enter the serial number of the unit in the Patient ID field.
Make sure that only the pad is on the table. Clear the table of anything else.
3. Perform one Whole Body air scan (a scan with nothing on the table). Do NOT interrupt the scan for
any reason. When the scan completes, select Exit Exam. Do not analyze the scan that was just
performed.
4. From the Main Menu screen, select Utilities > Service Utilities > Table Top Radiographic
Uniformity.
5. Select the Whole Body scan and click the Next>> button.
6. Table Top Radiographic Uniformity results tabs are now displayed. Select the Low Air Tab.
7. If the Global Stats S.D. (2) is less than 2.0, the instrument is properly aligned. If the Global Stats S.D.
is greater than or equal to 2.0, then the machine is not aligned properly. Check the following:
• If the image appears “streaky” or “banded”, check for loose wires or other debris between the table
and the tank.
• Check the Aperture Calibration.
• Check AC Line voltage.
• Check +/- 15 V Power Supply.
• Check for a dedicated line.
• Check Detector Flattening.
• Re-run daily QC (detailed in MAN-03644 Horizon QDR Series User Guide).
Print a copy of the test results and include it with the other paperwork being returned to Hologic.
After realignment, repeat the Table Top Radiographic Uniformity test. If the global S.D. is less than 2.0,
you are done. If the global S.D. if equal or greater than 2.0, print a copy of the test results and include it
with the other paperwork being returned to Hologic.
If the global is equal to or greater than 2.0, the non-uniformity may be attributed to the table itself. If so,
continue with this procedure. Make sure that the machine has body composition loaded. If it does, then
follow the appropriate procedure below in Section 0.9 Instruments using Body Composition Analysis (BCA)
on Page 13.
3. Center the table, place the phantom on the table and align with the laser. The sides of the phantom
should be parallel to the table. Use a ruler if necessary.
4. Click Perform Exam from the application main screen.
5. At the Select a Patient for this Exam window, select the phantom, and click OK.
6. At the Patient Confirmation screen, click OK.
7. At Select Scan Type screen, select AP/Lateral, uncheck Use Default Scan Mode, and click Next>>.
8. At the Select AP Lumbar Spine Scan Mode for AP/Lateral Exam screen, select Array (a), and click
Next>>.
9. At the Select Lateral Scan Mode for AP/Lateral Exam screen, select Fast Array (f), and click Next.
10. Recheck phantom alignment with laser.
Note Do not move the arm or table, otherwise the test will have to be started over.
11. Place two sheets of paper against the left and right edges of the phantom (Figure 416) When done,
click Continue.
12. On the Scan Parameters screen, set the scan length to 6.0 in., click Enter and click Start Scan.
13. When scan has finished, select Spine as the analysis method, and click Next>>.
14. Analyze the scan and click Start Position.
15. Refer to Figure 439 Verify that the angle displayed under the image (near the bottom of the screen)
is less than ± 2 degrees.
16. Using the sheets of paper as a guide, carefully rotate the phantom on its back edge towards the C-
Arm so that the bone segments are facing the rear of the table as shown. Except for the rotation, do
not move the phantom in any other direction.
17. Click Close. When the Continuously Press Enable message appears, press and hold the Lateral
Enable switch until the C-Arm and table move to the Lateral position.
18. On the Scan Parameters screen, click Start Scan.
19. When the “Continuously Press Enable” message appears, press and hold the Lateral Enable switch
until the C-Arm and table move back to the AP position.
20. From the Exit Exam screen, select Analyze Scan.
21. Analyze the scan (refer to user guide for Lateral analysis) and close the screen.
22. Click Exit to close the Exit Analysis screen.
23. Restart APEX software in service mode (if not already). From the APEX Main window, select
Utilities > Service Utilities > Field Service Calibration.
24. Select Highs/Lows for Array Spine Phantom.
25. Select the AP scan just performed and click Next.
26. Click Continue on the Calculated highs and lows page.
27. Select Highs/Lows for Array Lateral Phantom.
28. Click Continue on the Calculated highs and lows page.
29. Click Temp Write to write temporary results and click Continue to exit the write dialog box.
Note For Horizon system with version APEX 5.6 and higher, only 10 reproducibility
scans for calibrations are required. Earlier versions required 20 scans. Acquire 5
scans with a period of 15 between groups.
14. Analyze the first scan manually selecting Lumbar Spine for the analysis method.
15. Note the scan #.
16. Click the Close button to exit analysis.
17. From the Main Menu screen, select Utilities > Service Utilities > Auto Analyze.
18. Click unanalyzed Scans and select the 10 scans, click Compare radio button and then click Next.
19. Now, select the QC Scan performed in step Section 3.11.2 Perform QC Scan on Spine Phantom, (C, Ci)
With BCA Calibration Scans and Table Top Radiographic Uniformity (A, W and Wi) on Page 199.
20. Click Next.
21. The 10 scans will now be compared with the auto-analyzed QC Scan.
8.16.7 Calibration (Area, BMD, and BMC) with ACF and BCF Correction Factors
Calibration for Area, BMD, and BMC is accomplished in three stages:
• Array Scan Thickness Measurement & Calibration
• Calibration of Area and BMC for Array Scan Modes
• Adding Array AP scans to the QC database
Follow the procedure, in order, and exactly as shown, for each stage of the calibration.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam system
or if the customer has multiple systems, then a cross-calibration must be
performed to ensure that any longitudinal studies begun on the QDR being
removed can safely be continued on the new Horizon. Refer to the cross-
calibration procedure in Section 3.10 Fan Beam Cross-Calibration Gold
Standard System on Page 164 before de-installing the existing QDR. The Cross
Calibration procedure applies to fan beam systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
1. From the Main Menu screen, select Utilities > Service Utilities > Field Service Calibration.
2. Click the ACF/BCF button.
3. Select the 11 reproducibility scans performed in previous step.
4. Click Next
5. Click Continue at the Successfully Calculated dialog box.
6. At the Field Service Calibration dialog box, click the Write button to write the values to arrc.txt file.
7. Click Continue to return to the Main Menu.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam
system or if the customer has multiple systems, then a cross calibration
must first be performed to ensure that any longitudinal studies begun on
the QDR being removed can safely be continued on the new Horizon. Refer
to the cross-calibration procedure in Section 3.10 Fan Beam Cross-
Calibration Gold Standard System on Page 164 before uninstalling the
existing QDR. The Cross-Calibration procedure applies to fan beam
systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
3. Compare values displayed with actual values printed on the phantom. If there is a discrepancy,
check to make sure you have the phantom that shipped with the unit. Check with Hologic Technical
Support if needed.
8.17.3 Perform BCA Calibration Scans and Table Top Radiographic Uniformity (A, W
and Wi)
This procedure verifies proper X-Ray beam alignment and the uniformity of the X-Ray transparency of
entire tabletop in a manual process (for Horizon A, W and Wi models). BCA Calibration Scans must be
performed on the system prior to verifying the uniformity test scan.
Note On Models A, W, and Wi, after Flattening is performed and 30 minute cool
down is expired, perform a Daily QC scan. Automatically the system will
perform BCA calibration scans on the QC Phantom. After the calibration scans
are complete the system will inform you to remove the QC Phantom and clear
the table to perform the Table Top Uniformity Scan test. This test will report the
results automatically and must pass with the Hi and Low Air SD of less than
2.0.
If the system is passing normally, skip to next step.
If the system is not passing with less than SD 2.0, verify that table alignments
and beam alignments are correct, adjust as necessary and run Flattening again.
1. If necessary, restart the software in Service Mode.
2. Type WBAIRQC in the Patient Biography. Enter the serial number of the unit in the Patient ID field.
Make sure that only the pad is on the table. Clear the table of anything else.
3. Perform one Whole Body air scan (a scan with nothing on the table). Do NOT interrupt the scan for
any reason. When the scan completes, select Exit Exam. Do not analyze the scan that was just
performed.
4. From the Main Menu screen, select Utilities > Service Utilities > Table Top Radiographic
Uniformity.
5. Select the Whole Body scan and click the Next>> button.
6. Table Top Radiographic Uniformity results tabs are now displayed. Select the Low Air Tab.
7. If the Global Stats S.D. (2) is less than 2.0, the instrument is properly aligned. If the Global Stats S.D.
is greater than or equal to 2.0, then the machine is not aligned properly. Check the following:
• If the image appears “streaky” or “banded”, check for loose wires or other debris between the table
and the tank.
• Check the Aperture Calibration.
• Check AC Line voltage.
• Check +/- 15 V Power Supply.
• Check for a dedicated line.
• Check Detector Flattening.
• Re-run daily QC (detailed in MAN-03644 Horizon QDR Series User Guide).
Print a copy of the test results and include it with the other paperwork being returned to Hologic.
After realignment, repeat the Table Top Radiographic Uniformity test. If the global S.D. is less than 2.0,
you are done. If the global S.D. if equal or greater than 2.0, print a copy of the test results and include it
with the other paperwork being returned to Hologic.
If the global is equal to or greater than 2.0, the non-uniformity may be attributed to the table itself. If so,
continue with this procedure. Make sure that the machine has body composition loaded. If it does, then
follow the appropriate procedure below in Section 0.9 Instruments using Body Composition Analysis (BCA)
on Page 13.
3. Center the table, place the phantom on the table and align with the laser. The sides of the phantom
should be parallel to the table. Use a ruler if necessary.
4. Click Perform Exam from the application main screen.
5. At the Select a Patient for this Exam window, select the phantom, and click OK.
6. At the Patient Confirmation screen, click OK.
7. At Select Scan Type screen, select AP/Lateral, uncheck Use Default Scan Mode, and click Next>>.
8. At the Select AP Lumbar Spine Scan Mode for AP/Lateral Exam screen, select Array (a), and click
Next>>.
9. At the Select Lateral Scan Mode for AP/Lateral Exam screen, select Fast Array (f), and click Next.
10. Recheck phantom alignment with laser.
Note Do not move the arm or table, otherwise the test will have to be started over.
11. Place two sheets of paper against the left and right edges of the phantom (Figure 416) When done,
click Continue.
12. On the Scan Parameters screen, set the scan length to 6.0 in., click Enter and click Start Scan.
13. When scan has finished, select Spine as the analysis method, and click Next>>.
14. Analyze the scan and click Start Position.
15. Refer to Figure 439 Verify that the angle displayed under the image (near the bottom of the screen)
is less than ± 2 degrees.
16. Using the sheets of paper as a guide, carefully rotate the phantom on its back edge towards the C-
Arm so that the bone segments are facing the rear of the table as shown. Except for the rotation, do
not move the phantom in any other direction.
17. Click Close. When the Continuously Press Enable message appears, press and hold the Lateral
Enable switch until the C-Arm and table move to the Lateral position.
18. On the Scan Parameters screen, click Start Scan.
19. When the “Continuously Press Enable” message appears, press and hold the Lateral Enable switch
until the C-Arm and table move back to the AP position.
20. From the Exit Exam screen, select Analyze Scan.
21. Analyze the scan (refer to user guide for Lateral analysis) and close the screen.
22. Click Exit to close the Exit Analysis screen.
23. Restart APEX software in service mode (if not already). From the APEX Main window, select
Utilities > Service Utilities > Field Service Calibration.
24. Select Highs/Lows for Array Spine Phantom.
25. Select the AP scan just performed and click Next.
26. Click Continue on the Calculated highs and lows page.
27. Select Highs/Lows for Array Lateral Phantom.
28. Click Continue on the Calculated highs and lows page.
29. Click Temp Write to write temporary results and click Continue to exit the write dialog box.
Note For Horizon system with version APEX 5.6 and higher, only 10 reproducibility
scans for calibrations are required. Earlier versions required 20 scans. Acquire 5
scans with a period of 15 between groups.
14. Analyze the first scan manually selecting Lumbar Spine for the analysis method.
15. Note the scan #.
16. Click the Close button to exit analysis.
17. From the Main Menu screen, select Utilities > Service Utilities > Auto Analyze.
18. Click unanalyzed Scans and select the 10 scans, click Compare radio button and then click Next.
19. Now, select the QC Scan performed in step Section 3.11.2 Perform QC Scan on Spine Phantom, (C, Ci)
With BCA Calibration Scans and Table Top Radiographic Uniformity (A, W and Wi) on Page 199.
20. Click Next.
21. The 10 scans will now be compared with the auto-analyzed QC Scan.
8.17.7 Calibration (Area, BMD, and BMC) with ACF and BCF Correction Factors
Calibration for Area, BMD, and BMC is accomplished in three stages:
• Array Scan Thickness Measurement & Calibration
• Calibration of Area and BMC for Array Scan Modes
• Adding Array AP scans to the QC database
Follow the procedure, in order, and exactly as shown, for each stage of the calibration.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam system
or if the customer has multiple systems, then a cross-calibration must be
performed to ensure that any longitudinal studies begun on the QDR being
removed can safely be continued on the new Horizon. Refer to the cross-
calibration procedure in Section 3.10 Fan Beam Cross-Calibration Gold
Standard System on Page 164 before de-installing the existing QDR. The Cross
Calibration procedure applies to fan beam systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
1. From the Main Menu screen, select Utilities > Service Utilities > Field Service Calibration.
2. Click the ACF/BCF button.
3. Select the 11 reproducibility scans performed in previous step.
4. Click Next
5. Click Continue at the Successfully Calculated dialog box.
6. At the Field Service Calibration dialog box, click the Write button to write the values to arrc.txt file.
7. Click Continue to return to the Main Menu.
Note Proper calibration of the motors depends upon other motor calibrations. You
must calibrate motors in the following order: TZ, AY, TY, TX, AR, and
Aperture. Proper motor calibration depends on all motors always being
calibrated in the order shown.
The motor controller board control and drive the table and C-arm movements. Calibrate motors in the
following order: AY, TY, TX.
Use Table 86 to determine the calibration procedures you need to perform for each Horizon model.
Perform the calibration procedures identified for each motor (√) in order from left to right.
Table 86 Calibration Procedures
Model TZ AY TY TX AR
A √ √ √ √ √
W NA1 √ √ √ NA1
Wi NA1 √ √ √ NA1
C NA1 √ NA1 √ NA1
Ci NA1 √ NA1 √ NA1
1. Not applicable to the system.
Note Perform the motor calibration in the order shown below. In the Tests section
click Calibrate. See the instructions for each motor:
• 1.3.2 Tz Motor Calibration on Page 4
• 1.3.4 Calibrate MOTOR$TY (Horizon Models A, W, Wi) on Page 10
• 1.3.5 Calibrate MOTOR$TX on Page 12
• Section Calibrate MOTOR$AR (Horizon A) on Page 390
5. Reboot the system after all motor calibrations are completed.
Note Some motor calibration changes are not applied until the system is re-booted.
Calibrate MOTOR$TZ
Check Calibration Values
1. From the Hologic Main Window, click Utilities.
2. Click Service Utilities.
3. Select System Diagnostics. Select the Tz Motor tab.
4. Place SW6 (Figure 1) on the Tz motor driver board in Service mode.
Use the up/down service switches to move the pedestals to the lowest position.
Caution Drive both pedestals at the same time. If the pedestals are not synchronized, the
board will stop functioning.
Caution The expected travel of the motors is 22.9 cm. Enter the actual measure travel
distance on the calibration screen.
Use a metric ruler to carefully measure the travel distance of the table. This is a
critical measure for proper system function.
The program moves the C-arm and draws an encoder versus distance calibration grid. The positive
and negative linear fit parameters are displayed at the top left and top right of the plot in the form Y
= Intercept + Slope * X.
5. If the two slopes are within 0.3 of each other (Figure 6), click Continue.
The Table (AY Motor) Calibration dialog box displays the calibration parameter values.
6. Check that the parameters meet specifications. For more information, see MOTOR$AY Calibration
Parameters.
The last value on the calibrate_position line in the dialog box must be 3750 ± 5 ticks. To adjust this
value, refer to Section MOTOR$AY Calibration Parameters on Page 385.
7. Click Save.
Note Because the encoder is underneath the X-ray Source Assembly, adjusting it can
be counter-intuitive.
1. For example, if a value of 3770 appears (20 ticks above the correct calibration
position at the left physical stop), move the C-arm all the way to the right.
2. If the value on the right is 500, adjust it to 480 to reduce the corresponding
value on the left stop.
3. Hit Calibrate to move the C-arm to the left stop. The value should be
reduced by 20 ticks.
MOTOR$AY Calibration Parameters
Values are separated by commas within calibration fields.motor_direction determines whether the
positive step direction is the reverse of the direction of increasing encoder values. The value is 0 if the
direction is not reversed and 1 if the direction is reversed. This value is set from the corresponding
parameter in the MOTOR_XX.PRO file and is a constant for each motor. The direction of increasing
encoder values must agree with the system coordinates. The positive step direction is set by the wiring
harness for the stepper motor.
calibrate_position has eight values:
(1)Tolerance. The position tolerance for absolute moves in encoder ticks. The calibration program
sets this field to the value found in the corresponding calibrate_position field in the
MOTOR_XX.PRO file.
(2,3)NumberOfSteps,StepDistance. The number of motors steps corresponding to the step
distance in µm. The ratio of these two numbers determines the step size (41187/2288 18 µm). The
calibration program sets these fields to the values found in the corresponding calibrate_position
fields in the MOTOR_XX.PRO file. These values are a property of the mechanical design of the
system and should never change.
(4,5)NumberOfTicks,TickDistance. The number of encoder ticks corresponding to the tick distance
in µm. The ratio of these two numbers (61339/209 293 µm) determines the encoder calibration.
These numbers can change from calibration to calibration, however, the ratio should remain
approximately the same.
(6)EncoderOffset. The value subtracted from the encoder readback to set the origin of motion. The
origin of AY, TY, and TX motion is the extreme right mechanical stop. This value should be the same
as the NegLimit value.
(7,8)NegLimit,PosLimit. The encoder readings for the negative (right) and positive (left)
mechanical stops.
pos_limit_position is the limit of motion, in µm, in the positive direction.
neg_limit_position is the limit of motion, in µm, in the negative direction.
Note Because the encoder is underneath the X-ray Source Assembly, adjustment can
seem counter-intuitive. When we perform calibration, pressing Esc as the
C-arm travels left should stop it at 3750 ± 5.
If, for example, the C-arm stops at 3770, 20 stops greater than it should be, we
need to send the C-arm all the way to the right to determine where it stops.
If the C-arm stops at 500 on the right, we need to reduce the travel by 20 to 480.
We then must recalibrate, press Esc as the C-arm travels left, and then check the
left stop once again to see if the C-arm has stopped at 3750 ± 5.
Figure 394 Click Cancel when TY Motor Reaches Left Mechanical Stop
3. When TY reaches the left mechanical stop (Figure 396), click OK.
Figure 396 Click OK when TY Motor Reaches Left Mechanical Stop
4. When TY reaches the right mechanical stop (Figure 397), click OK.
Figure 397 Right Mechanical Stop
The program moves the C-arm and draws an encoder versus distance calibration grid. The positive and
negative linear fit parameters are displayed at the top left and top right of the plot in the form
Y = Intercept + Slope * X.
5. If the two slopes are within 0.3 of each other (Figure 398), click Continue.
Figure 398 TY Calibration Plot
Calibrate MOTOR$TX
The TX motor centers the table and C-arm.
1. Select the TX Motor tab.
2. Click Calibrate.
3. A dialog box prompts moving the C-arm to 0° degrees (Figure 13).
Figure 399 Motor$TX Calibration
Caution The arm must be at 0°.
Note This message only applies to Horizon A systems. Use a digital level to set the C-
arm to 0° ± 0° for the A systems.
Ignore this message for other models.
4. Click OK.
5. When the TX motor reaches the outer mechanical limit, click Cancel (or press Esc) and verify that
the starting position is correct (1000 ±5).
6. Start Calibrate again and when TX reaches the outer mechanical limit (Figure 13), click OK.
Figure 400 Outer Mechanical Limit
The program moves the table and draws an encoder versus distance calibration grid. The positive
and negative linear fit parameters are displayed at the top left and top right of the plot in the form
Y = Intercept + Slope * X.
7. If the two slopes are within 0.3 of each other (Figure 14), click Continue.
Caution Confirm that the table is level front-to-back before calibrating MOTOR$AR.
Unlevel tables cause positioning tolerance errors on whole body, lateral, and
IVA scans.
1. Remove the table pad.
2. Place the digital level on the table at the foot end. The digital level
3. value must be 0.0°.
4. Move the level to the head of the table.
5. Confirm that the level is still at 0.0°.
If the level result is not 0.0°, adjust the leveling ft until the result is 0.0°.
Continue the Calibration Procedure
The program moves the TY and AY motors to center positions, then moves
the TZ motor to topmost position.
6. Remove the C-arm X-ray Source Assembly cover.
7. Level the X-ray Source Assembly to 0.0° ± 0.1°.
a. Place the digital level on the top of the X-ray Source Assembly (Figure 402).
b. b. Use the Table IN/OUT switch on the operator panel to move the C-arm until the level is 0.0° ±
0.1°
8. Click OK.
9. When the AR motor reaches the AP mechanical limit (Figure 403), click Cancel.
Figure 403 AR Motor AP Mechanical Limit
5. Observe the detector counts on the diagnostics display. Adjust the right string encoder until the
value is 325 ± 2.
6. Set SW6 on the TZ motor driver board to the Normal position.
7. Calibrate to see whether the unit is within limits (325 ± 2).
8. If unit exceed limits, on the Tz PCB, disconnect the right string encoder connector (J-9).
9. Move the left string encoder connector from J-7 to J-9.
10. Set SW1 on the TZ motor driver board to the Normal position.
11. Observe the detector counts on the diagnostics display.
12. Adjust the left string encoder until the value is 325 ± 2 of the set value the right string encoder was
set to in Step .
13. Move the left string encoder connector back to J-7.
14. Reconnect the right string encoder to J-9.
15. Use the up/down service switches (SW2 and SW3) to move the pedestal to a position at least one (1)
inch above the lowest position.
Caution The expected travel of the motors is 22.9 cm. Enter the actual measure travel
distance on the calibration screen.
Carefully measure the travel distance of the table. This is a critical measure for
proper system function.
20. At the conclusion, the program will prompt for details similar to the text below:
set_table_calibration=325,2664,325,2668
calibrate_position=10,1000,1000,819,50000,500,500,3494
pos_limit_position=182784
neg_limit_position=0
The four values for the set table calibration are, respectively:
• the left pedestal lower and upper encoder limits, and
• the right pedestal lower and upper encoder limits
the left lower and right lower values need to be 325 ± 2 counts and within 2 counts of each other.
The range of motion for the two pedestals (i.e., the upper encoder limit-lower encoder limit)
should be less than or equal to 40 counts of each other. These values should be within 20 counts
of each other.
21. If the value is not within specifications, adjust the appropriate pedestal encoder.
22. Shut down the computer.
23. Turn off the circuit breaker at the back of the system.
24. Turn system on.
25. Log in as QDR.
26. Run the following scans to test, all of which should pass:
a. QC Scan
b. BCA calibration scans
c. Table Top Radiographic Uniformity (TTU).
7. In the Patient window (Figure 410), enter Flat/Lat Verification in Last Name.
8. Click OK.
Figure 410 Patient Window
9. Enter a height and weight for the patient in the Patient Confirmation Window (Figure 411).
10. Click OK.
11. At the Select Scan Type screen, select AP/Lateral (Figure 412).
12. Deselect Use Default Scan Mode.
Figure 412 Select AP/Lateral
17. At the Select Lateral Scan Mode for AP/Lateral Exam screen, select Fast Array (f).
18. Click Next.
Figure 414 Select Fast Array (f)
19. Follow the instructions (Figure 415) regarding patient (spine phantom) alignment with the laser.
20. Click Continue.
Figure 415 Alignment Warning
21. Place two sheets of paper against the left and right edges of the phantom (Figure 416).
Figure 416 Paper against Right Edges of Phantom
25. After the scan has been completed, on the Analyze Setup screen (Figure 418), select Choose
Analysis Method.
26. Select Lumbar Spine from the pull-down.
27. Click Next.
Note If Lumbar Spine is not available from the drop down menu, you must load base
analysis protocols from the application disk first.
32. If image adjustment is necessary to visualize the bone, select the SUN symbol to
open the Image Toolbox (Figure 421).
33. When done, click the Sun symbol to switch back to the Global ROI Toolbox.
Figure 421 Global ROI Toolbox
35. Select the top line of the Global ROI and move it to the vertebral space that separates L1 and L2
(Figure 422)
Figure 422 Move Top Line of Global ROI to Vertebral Space
39. Verify that the angle displayed under the image (near the bottom of the screen) is less than ± 2
degrees (Figure 427).
Figure 427 Bone Map Start Position
40. Using the sheets of paper as a guide, carefully rotate the phantom on its back edge towards the C-
arm so that the bone segments are facing the rear of the table as shown (Figure 428).
Except for the rotation, do not move the phantom in any other direction.
45. From the Exit Exam screen, select Analyze Scan (Figure 430).
46. From the Analysis Setup screen, select Choose Analysis Method.
47. Select Supine Lateral from the pull-down menu (Figure 431).
Figure 431 Supine Lateral Analysis Method
50. Select Line Mode. Select the bottom line of the Global ROI and make the sides vertical (Figure 433).
Figure 433 Line Mode
51. Verify the left edge of the ROI is at the left edge of the image.
52. Select Whole Mode (Figure 434).
53. Using the right arrow key, count the number of steps required to align the left edge of the Vertebral
Boundaries ROI (the middle vertical line in Figure 434) to the left edge of spine.
54. Record the number of steps taken here: ______________.
55. The steps will be less than 6 when properly calibrated.
56. If the number of steps taken was 6 or less, click Cancel on the scan analysis screen, click Yes and Exit
to confirm analysis exit and proceed to Step 66.
57. If the number of steps exceeds 6 steps, you must adjust the table height.
58. Click Cancel on the scan analysis screen,
59. Click Yes and
60. Exit to confirm analysis exit and proceed to Step 61.
61. Click Utilities > Service Utilities > Table Height.
62. Refer to Figure 435.
a. Subtract the actual number of steps recorded in Step 53 from the number 10.
b. Enter the difference into the text box as a positive number to lower the table (Example for this
image: +4).
c. Record value entered into the text box ______________. Click Next when done.
63. From the APEX or QDR main screen, select Exit, Exit QDR without shut down and click OK.
64. Re-boot the system and login as user Field Service. When the APEX or QDR main screen displays,
select Exit, Exit QDR without shut down and click OK.
65. Start APEX or QDR software in Service Mode.
66. Detector Flattening (Big Flat) must now be run since the table height has been changed. Click
Utilities > Service Utilities > Detector Flattening and follow the on-screen instructions.
Note The Table Height Utility parameter limits are ±5 to raise or lower the table;
however, actual limits may be less for any given system. If the limits are
exceeded, the Lateral Position setup cannot function and a Table Height Out
of Limits or similar error message will display. If this is the case, enter
countering numbers until values are within limits (indicated by no error
message).
For Example: you entered +5 and the Table Height Out of Limits or similar
error message displayed. To counter, enter -2. This will result in you having
input the value +3. Any time you run the Table Height utility, you must always
re-run Detector Flattening (Big Flat).
67. Select Perform Exam and select patient Flat/Lat Verification in the last name field and
68. Click OK.
69. Click OK.
70. Repeat Steps 14 through 54.
This procedure flattens the X-ray beam for each scan mode and uses the metal edges of the table to
calculate the center table position in the x-axis.
Note All covers, table mat, etc., normally in the X-ray path, MUST be on the Scanner
before running Beam Flattening.
1. Restart the Horizon software in service mode (if not already). Press the Center Table button and
turn on the laser.
2. Using a DXA Quality Control Phantom (Part Number ASY-01564), place the phantom on end with
the flattening target facing up. Orient the phantom so the metal edge is to the right (head end) and
tissue to the left (foot end) as viewed from the front of the table. Align the phantom flattening target
to the laser crosshair.
Figure 436 Correct Flattening Position
3. Start the X-Ray Survey Utility by selecting Utilities > Service Utilities > X-Ray Survey.
4. Set Pulse to 1, High gain to 2, Low gain to 1, X-ray Mode to 3 and Aperture to 7 (for A, C, and W
models).
5. For Display, click Graph.
6. Click X-Rays (F2) to turn X-Rays on.
7. Check that the phantom covers the entire beam. This is critical. If phantom does not cover the entire
beam, move it until it does. Keep the phantom as straight as possible.
8. Turn off X-rays with the <F2> key and press <Esc> or click Close to exit X-Ray Survey.
9. Reboot the PC, logon as Field Service, and restart the software in Service Mode to return to Horizon
Main Menu screen.
10. Select Utilities > Service Utilities > Detector Flattening.
11. Click Flattening –Basic First.
Figure 437 Flattening –Basic First
This procedure will take approximately 15 minutes on a Ci and Wi, 25 minutes on a C and W, and
50 minutes on an A model system.
13. After this procedure, the X-Ray Source Assembly will be hot. You must wait for 30 minutes before
performing a Field Service Calibration of the machine.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam
system or if the customer has multiple systems, then a cross calibration
must first be performed to ensure that any longitudinal studies begun on
the QDR being removed can safely be continued on the new Horizon. Refer
to the cross-calibration procedure in Section 3.10 Fan Beam Cross-
Calibration Gold Standard System on Page 164 before uninstalling the
existing QDR. The Cross-Calibration procedure applies to fan beam
systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
3. Compare values displayed with actual values printed on the phantom. If there is a discrepancy,
check to make sure you have the phantom that shipped with the unit. Check with Hologic Technical
Support if needed.
10.4.3 Perform BCA Calibration Scans and Table Top Radiographic Uniformity (A, W
and Wi)
This procedure verifies proper X-Ray beam alignment and the uniformity of the X-Ray transparency of
entire tabletop in a manual process (for Horizon A, W and Wi models). BCA Calibration Scans must be
performed on the system prior to verifying the uniformity test scan.
Note On Models A, W, and Wi, after Flattening is performed and 30 minute cool
down is expired, perform a Daily QC scan. Automatically the system will
perform BCA calibration scans on the QC Phantom. After the calibration scans
are complete the system will inform you to remove the QC Phantom and clear
the table to perform the Table Top Uniformity Scan test. This test will report the
results automatically and must pass with the Hi and Low Air SD of less than
2.0.
If the system is passing normally, skip to next step.
If the system is not passing with less than SD 2.0, verify that table alignments
and beam alignments are correct, adjust as necessary and run Flattening again.
1. If necessary, restart the software in Service Mode.
2. Type WBAIRQC in the Patient Biography. Enter the serial number of the unit in the Patient ID field.
Make sure that only the pad is on the table. Clear the table of anything else.
3. Perform one Whole Body air scan (a scan with nothing on the table). Do NOT interrupt the scan for
any reason. When the scan completes, select Exit Exam. Do not analyze the scan that was just
performed.
4. From the Main Menu screen, select Utilities > Service Utilities > Table Top Radiographic
Uniformity.
5. Select the Whole Body scan and click the Next>> button.
6. Table Top Radiographic Uniformity results tabs are now displayed. Select the Low Air Tab.
7. If the Global Stats S.D. (2) is less than 2.0, the instrument is properly aligned. If the Global Stats S.D.
is greater than or equal to 2.0, then the machine is not aligned properly. Check the following:
• If the image appears “streaky” or “banded”, check for loose wires or other debris between the table
and the tank.
• Check the Aperture Calibration.
• Check AC Line voltage.
• Check +/- 15 V Power Supply.
• Check for a dedicated line.
• Check Detector Flattening.
• Re-run daily QC (detailed in MAN-03644 Horizon QDR Series User Guide).
Print a copy of the test results and include it with the other paperwork being returned to Hologic.
After realignment, repeat the Table Top Radiographic Uniformity test. If the global S.D. is less than 2.0,
you are done. If the global S.D. if equal or greater than 2.0, print a copy of the test results and include it
with the other paperwork being returned to Hologic.
If the global is equal to or greater than 2.0, the non-uniformity may be attributed to the table itself. If so,
continue with this procedure. Make sure that the machine has body composition loaded. If it does, then
follow the appropriate procedure below in Section 0.9 Instruments using Body Composition Analysis (BCA)
on Page 13.
3. Center the table, place the phantom on the table and align with the laser. The sides of the phantom
should be parallel to the table. Use a ruler if necessary.
4. Click Perform Exam from the application main screen.
5. At the Select a Patient for this Exam window, select the phantom, and click OK.
6. At the Patient Confirmation screen, click OK.
7. At Select Scan Type screen, select AP/Lateral, uncheck Use Default Scan Mode, and click Next>>.
8. At the Select AP Lumbar Spine Scan Mode for AP/Lateral Exam screen, select Array (a), and click
Next>>.
9. At the Select Lateral Scan Mode for AP/Lateral Exam screen, select Fast Array (f), and click Next.
10. Recheck phantom alignment with laser.
Note Do not move the arm or table, otherwise the test will have to be started over.
11. Place two sheets of paper against the left and right edges of the phantom (Figure 416) When done,
click Continue.
12. On the Scan Parameters screen, set the scan length to 6.0 in., click Enter and click Start Scan.
13. When scan has finished, select Spine as the analysis method, and click Next>>.
14. Analyze the scan and click Start Position.
15. Refer to Figure 439 Verify that the angle displayed under the image (near the bottom of the screen)
is less than ± 2 degrees.
16. Using the sheets of paper as a guide, carefully rotate the phantom on its back edge towards the C-
Arm so that the bone segments are facing the rear of the table as shown. Except for the rotation, do
not move the phantom in any other direction.
17. Click Close. When the Continuously Press Enable message appears, press and hold the Lateral
Enable switch until the C-Arm and table move to the Lateral position.
18. On the Scan Parameters screen, click Start Scan.
19. When the “Continuously Press Enable” message appears, press and hold the Lateral Enable switch
until the C-Arm and table move back to the AP position.
20. From the Exit Exam screen, select Analyze Scan.
21. Analyze the scan (refer to user guide for Lateral analysis) and close the screen.
22. Click Exit to close the Exit Analysis screen.
23. Restart APEX software in service mode (if not already). From the APEX Main window, select
Utilities > Service Utilities > Field Service Calibration.
24. Select Highs/Lows for Array Spine Phantom.
25. Select the AP scan just performed and click Next.
26. Click Continue on the Calculated highs and lows page.
27. Select Highs/Lows for Array Lateral Phantom.
28. Click Continue on the Calculated highs and lows page.
29. Click Temp Write to write temporary results and click Continue to exit the write dialog box.
Note For Horizon system with version APEX 5.6 and higher, only 10 reproducibility
scans for calibrations are required. Earlier versions required 20 scans. Acquire 5
scans with a period of 15 between groups.
14. Analyze the first scan manually selecting Lumbar Spine for the analysis method.
15. Note the scan #.
16. Click the Close button to exit analysis.
17. From the Main Menu screen, select Utilities > Service Utilities > Auto Analyze.
18. Click unanalyzed Scans and select the 10 scans, click Compare radio button and then click Next.
19. Now, select the QC Scan performed in step Section 3.11.2 Perform QC Scan on Spine Phantom, (C, Ci)
With BCA Calibration Scans and Table Top Radiographic Uniformity (A, W and Wi) on Page 199.
20. Click Next.
21. The 10 scans will now be compared with the auto-analyzed QC Scan.
10.4.7 Calibration (Area, BMD, and BMC) with ACF and BCF Correction Factors
Calibration for Area, BMD, and BMC is accomplished in three stages:
• Array Scan Thickness Measurement & Calibration
• Calibration of Area and BMC for Array Scan Modes
• Adding Array AP scans to the QC database
Follow the procedure, in order, and exactly as shown, for each stage of the calibration.
Note 1. If the Horizon being installed is to replace an existing QDR fan beam system
or if the customer has multiple systems, then a cross-calibration must be
performed to ensure that any longitudinal studies begun on the QDR being
removed can safely be continued on the new Horizon. Refer to the cross-
calibration procedure in Section 3.10 Fan Beam Cross-Calibration Gold
Standard System on Page 164 before de-installing the existing QDR. The Cross
Calibration procedure applies to fan beam systems only.
2. To ensure stability, the instrument must be completely powered up for a
minimum of 30 minutes prior to running any of the following tests. Also,
because the tabletop pad will have a slight affect on the test results, all scans
must be performed with the tabletop pad in place.
Warning: X-rays are produced during most of these tests. Keep hands, head and other
body parts out of the X-ray beam path. The tester should also be wearing an
approved radiation dosimetry badge.
1. From the Main Menu screen, select Utilities > Service Utilities > Field Service Calibration.
2. Click the ACF/BCF button.
3. Select the 11 reproducibility scans performed in previous step.
4. Click Next
5. Click Continue at the Successfully Calculated dialog box.
6. At the Field Service Calibration dialog box, click the Write button to write the values to arrc.txt file.
7. Click Continue to return to the Main Menu.
Note Some of the screens may change in appearance over time as Microsoft updates
and patches the Word 2013 program.
Procedure:
1. Power up computer.
2. Select Admin then enter the Password as password.
3. From the APEX software select Exit then Exit Without Shutdown to shut down the APEX software.
Network Setup
1. Plug the internet connection cable into the back of the PC.
2. If available connect to site's available network in room. Go to next section.
3. Alternatively: Use a MiFi wireless cellular adaptor for internet connectivity. If functional, use with a
USB direct cable connection to the MIFI or on data enabled I-phone.
a. Or, obtain a CMP-00813 Wireless USB Adapter.
b. Turn on the MiFi hotspot.
c. Insert the Netgear Resource CD into the DVD drive run setup.
d. If the Netgear setup program starts automatically, go to step h.
e. If the Netgear setup program does not start automatically, click Start > Run > Browse.
f. Double-click autorun.exe.
g. Follow prompts, insert USB Netgear when instructed to.
h. When the Netgear installation is complete it will display a list of available wireless networks.
i. Select the one that corresponds to your MiFi hotspot.
j. Select manually configure from the next window.
k. Enter the password for the MiFi hotspot.
l. Verify you have internet access.
m. Reboot the PC.
n. Select Admin then enter the Password as password.
o. From the APEX software select Exit then Exit Without Shutdown to shut down the APEX
software.
Note For APEX Versions 4.x, install the APEX Reporting Engine V2013 only if you
have purchased the Physician’s Viewer™ Report Writer option.
For APEX 5.5 and higher, the installation of Physicians Viewer™ and the
DXReport option is not required on the DXA scanner. The option is now built
into the APEX application.
The APEX Reporting Engine V2013 can be loaded on the scanner PC or on another PC that can be used as
a remote viewing station.
• Load APEX Reporting Engine V2013 on the scanner PC for Hologic Applications Training
purposes.
Caution Do not activate the software on the scanner, if the software is being installed on
the Docs computer.
Prerequisites
Your system must have:
• a DVD drive
• a product key
• the APEX Reporting Engine V2013 Recovery DVD
Note APEX Reporting Engine V2013 Recovery must be loaded prior to the APEX
software, or if required Physician’s Viewer and DXReporting options software.
Note In bottom right corner of the screen, ensure that internet access is active.
If no network connection is available, use the radio button for phone activation.
16. When the MS Word 2013 installation is complete, uninstall the Netgear software through Control
Panel > Programs and Features by clicking on the Netgear software and selecting Uninstall. Follow
the on screen prompts.
17. Reboot the PC.
18. Log in as QDR.
5. Select Control Panel > Hardware and Sound under Devices and Printers.
6. Click on Device Manager (from Category view) or Control Panel > Device Manger (from Icon
view).
7. Right-click on Other devices > Other PCI Bridge and select Properties.
8. Select the Drive tab.
9. Click Update Driver.
10. Select Browse My Computer and click Browse.
11. Navigate to the DVD drive to the APEX System Software > Device Driver folder.
12. Click OK.
13. Click Next.
14. Select Install this software anyway on the Windows Security popup screen.
15. After the driver installs, the Hologic CC Card interface card is recognized by Windows.
16. Click Close.
17. Turn off the PC.
Note The installation program begins. The progress of the SQL Server 2012 Express
installation will be displayed in the command window. A message will be
displayed when it has completed. This may take several minutes, so please be
patient.
4. Press any key to continue.
5. Turn off the PC.
The installation program begins and setup information is displayed. Then the InstallShield Wizard
and then the Welcome screen is displayed.
Figure 443 InstallShield Wizard
18. Click Next to accept the C:\QDR folder as the destination location. A progress bar displays while
the program files are being installed.
19. The InstallShield Wizard Complete (Figure 449) screen appears.
Yes, I want to restart my computer now option is selected.
Figure 449 InstallShield Wizard Complete
20. Remove the APEX applications DVD from the DVD drive.
21. Click Finish.
The system reboots.
22. Log in as QDR User.
23. Navigate to the APEX Task Bar and click Help > About APEX.
The About Hologic APEX Software screen (Figure 450) appears.
Figure 450 About APEX
The correct installed Software Version should be displayed in the Version box.
24. Click OK.
Restore the Scan Files from The Removable Drive and System Recovery
1. Copy all of the .pxx and .rxx files from the removable drive to the C:\QDR\data folder.
2. Recover the system backup performed just prior to exchanging the PC's.
3. After reconcile completes, Exit APEX with Shutdown.
4. Restart the PC and login as Admin.
5. Go to Utilities > Report tab > DXReport tab.
6. Recover the report backup .zip file from the removable drive.
Note Physicians Viewer V7.3 and DX Report V7.3 are provided to install as upgrade
to remote viewing station used at site.
Neither are required to be installed with APEX 5.6.0.3 systems.
The functionality for VFA analysis and DX Reporting is built into the APEX
application.
Note When Hologic Connect is started and connection is made to the system by the
Help Desk, the APEX application will turn off the table communications. After
connection is terminated the Apex application must be restarted to use the table.
6. Install Hologic Connect.
Note Physician’s Viewer and Central DICOM Server are two separate applications
that exist on the same installation disk. This section applies only to the
installation of Physician’s Viewer.
Note Physician’s Viewer cannot be installed on a computer that has Central DICOM
Server installed. Prior to installing Physician’s Viewer, ensure that Central
DICOM Server is not installed.
Note If you want to sync Physician’s Viewer to a Central DICOM Server, both
Physician’s Viewer and Central DICOM Server must be properly configured.
1. If this computer has QDR™ system software installed, log in as QDR User and then exit QDR
without shutdown.
4. At this screen you will need your System Administrator (normally an MIS person) to type in the
password.
5. When the password is provided, click OK. Figure 452 displays.
Note If you don’t know, or aren’t sure, what the Local DICOM Port and Local
DICOM AE Title should be, accept the defaults. These can be changed later
during configuration.
Please note that DICOM Server is a service associated with Physician’s Viewer.
DICOM Server is not the same as Central DICOM Server, which is a separate
application.
Figure 455 Physician’s Viewer Configuration
Note:
13. Click Install. A progress bar (Figure 457) displays while the program files are being installed.
Figure 457 Installing Hologic Physician’s Viewer
2. In the Control Panel (Figure ), click on Network and Internet to open the Network ands Sharing
Center (Figure ).
Control Panel
In this window,
1. select Use the following DNS server addresses and
2. enter the IP Address for the preferred DNS server and the Alternate DNS Server.
For a list of Free and Public DNS Servers, see Table 88 and Table 89.
3. Click OK.
4. Restart the computer and login as QDR.
Note The IP Address, Subnet mask, and Default gateway may be configured to a
static IP address if assigned by the facility.
Note The application must be restarted after the remote session is ended.
2. Enable Remote Desktop connection by checking Allow connection from computer running any
version of Remote Desktop (less secure).
3. Click OK.
4. Close the System Properties window and the Exit box to close the control panel.
Note The configuration of the Remote Desktop connection option in Windows may
not be permitted by the IT department of the facility.
puntCAT12 109.69.8.51
1. The free DNS servers listed above as Level3 will automatically route to the nearest DNS server operated by
Level3 Communications, the company that provides most of the ISPs in the US their access to the Internet
backbone.
2. Google also offers IPv6 public DNS servers: 2001:4860:4860::8888 and 2001:4860:4860::8844
3. DNS.WATCH also has IPv6 DNS servers at 2001:1608:10:25::1c04:b12f and 2001:1608:10:25::9249:d69b. In an
uncommon but much appreciated move, DNS.WATCH publishes live statistics for both of their free DNS
servers. Both servers are located in Germany which could impact performance if used from the US or other
remote locations.
4. OpenDNS also offers DNS servers that block adult content, calledOpenDNS FamilyShield. Those DNS
servers are 208.67.222.123 and 208.67.220.123. A premium DNS offering is also available, calledOpenDNS
Home VIP.
5. The Norton ConnectSafe free DNS servers listed above block sites hosting malware, phishing schemes, and
scams, and is called Policy 1. Use Policy 2 (199.85.126.20 and 199.85.127.20) to block those sites plus those
with pornographic content. Use Policy 3 (199.85.126.30 and 199.85.127.30) to block all previously men-
tioned site categories plus those Norton deems "non-family friendly." Be sure to check out the list of things
blocked in Policy 3 - there are several controversial topics in there that you may find perfectly acceptable.
6. GreenTeamDNS "blocks 18 categories which include malware, botnets, dangerous websites, adult related
content, aggressive/violent sites as well as advertisements and drug-related websites" according to their
FAQ page. Premium accounts have more control.
7. Register here with SafeDNS for content filtering options in several areas.
8. The DNS servers listed here for OpenNIC are just two of many in the US and across the globe. Instead of
using the OpenNIC DNS servers listed above, see their complete list of public DNS servers here and use
two that are close to you or, better yet, let them tell you that automatically here. OpenNIC also offers some
IPv6 public DNS servers.
9. FreeDNS says that they "never log DNS queries." Their free DNS servers are located in Austria.
10.The censurfridns.dk DNS servers are uncensored, operated by a privately funded individual, and are
physically located in Denmark. You can read more about them here. IPv6 DNS servers are also available at
2002:d596:2a92:1:71:53:: and 2001:67c:28a4::.
11.Hurricane Electric also has an IPv6 public DNS server available: 2001:470:20::2.
12.puntCAT is physically located near Barcelona, Spain. The IPv6 version of their free DNS server is
2a00:1508:0:4::9.
Note Primary DNS servers are sometimes called preferred DNS servers and
secondary DNS servers are sometimes called alternate DNS servers. Primary
and secondary DNS servers can be mixed and matched to provide another
layer of redundancy.
Logs: No
Other: Be nice.
logs kept.
Other: fvz-rec-us-ga-01
Logs: Logging completely disabled.
Note Central DICOM Server can only be installed on a computer running the
Windows XP Pro or Windows 7 operating system.
Central DICOM Server cannot be installed on a computer that has Physician’s
Viewer installed. Prior to installing Central DICOM Server, ensure that
Physician’s Viewer is not installed.
Central DICOM Server is typically installed on a separate, server computer
that runs 24 hours a day, seven days a week. If the server computer has QDR
system software installed, log in as QDR User and then exit QDR without
shutdown, before installing Central DICOM Server.
4. At this screen you will need your System Administrator (normally an MIS person) to type in the
password. When the password is provided, click OK. Figure 460 displays.
Note If you don’t know, or aren’t sure, what the Local DICOM Port and Local
DICOM AE Title should be, accept the defaults. These can be changed later
during configuration
14. Click Next. Figure 465 appears.
Figure 465 Hologic DICOM Server Ready to Install the Program
15. Click Install. A progress bar displays (Figure 466) while the program files are being installed.
Note For Physician’s Viewer, the CentralDICOMSever parameters are set up during
installation and called DICOM Server.
If the CDS parameters need to be changed after Physician’s Viewer is installed, the same procedure is used.
Note Contact Technical Support if the default Physician’s Viewer configuration does
not suit your installation requirements:
For example:
• DICOM Central Server with multiple remote PV installs
• IT at the site needs to change ports to reduce data transfer
11.5.1 Procedure
1. In Windows Explorer, browse to C:\Program Files (x86)\Hologic\Physician's Viewer (Figure 468).
Figure 468 Browse to Physician’s Viewer
2. Double-click DicomServerConfig.exe.
3. In the Hologic DICOM Server Configuration window (Figure 469), configure the parameters as
necessary.
4. When done, click OK.
11.6.1 Purpose
These instructions are provided to guide in the installation and configuration of Hologic Connect.
Features of Hologic Connect on APEX V4.6/V5.6 include the following
Location Site
Primary Connect.hologic.com 209.202.167.57
US - East Coast: ghsom1.axeda.com 209.202.157.179
Location Site
US - West Coast: ghsj1.axeda.com 152.8.82.253
UK: ghuk1.axeda.com 89.234.8.217
Japan: gas-bo6.axeda.com 209.202.186.30
Australia: gas-bo7.axeda.com 216.34.120.236
Note If you need to uninstall the APEX software, the Hologic Connect client will
have to uninstalled first.
Note Hologic Connect will only install with APEX software versions 4.6, 5/6, or 6.0.
If a warning appears (Figure ) indicating that the correct version of the software
has not been installed, the APEX software must be upgraded before Hologic
Connect will install correctly.
Note After you click on setup.exe there will be a delay after the preparing to install
screen in the installation popup display. This is normal. Be patient and wait
approximately 2 min.
6. Click Yes to restart the system when the Hologic Connect Installer Information Screen (Figure )
prompts you to restart.
Upon restart, the Hologic Connect Configuration Icon (Figure ) will appear on the desktop.
Hologic Connect Configuration Icon
4. Scroll down the list of active services and verify that Hologic Connect:
a. appears in the list of services, and
b. its status is Stopped.
Note If the site does not have a DNS Server you may have to use a public DNS server.
See list at bottom of document.
Contact Information
Control Description
Common Data Block
Online Click the Online button to toggle the state of the HolxQdr driver between
Online and Offline
Emergency Stop When ON, indicates the Emergency Stop button is pressed
Interlock Inhibit When ON, indicates an interlock inhibit error condition is raised in the
HolxQdr driver
Collision Imminent When YES, the hardware has moved to a position that the HolxQdr driver
has flagged as dangerous
Control Panel Click the Control Panel button to enable, or disable, the control panel
ComCon Version Displays the version number for the Communications Controller board
firmware
Number of ComCon Displays the number of Communications Controller interrupts received
Interrupts
Missed ComCon Interrupts Displays the number of Communications Controller interrupts that have
been lost
Number of DOS Interrupts Displays the number of DOS interrupts received
Tests Block
Link Addresses (Check to When run, the Communications Controller (C/C) card is queried at each of
select the test) the first eight addresses on the serial bus and results are displayed in the
Test Results window.
A successful query result is displayed as E_OK and a failed query is
displayed as E_TIMEOUT_I. Queries can be expected to fail for devices that
are not physically present on the system (e.g. the TZ motor on Horizon C
and W systems).
If the test as a whole fails (an error other than a query failure), the error
message is displayed.
Emergency Watchdog When run, this test verifies that A/C line frequency interrupts are received
(Check to select the test) from the C/C. If so, a number of C-arm queries are sent to the C/C, and then
stopped. Results are displayed in the Test Results window.
If the test completes successfully, PASSED is displayed along with the time it
takes for an emergency stop condition to be asserted following the last C-
arm query.
If the test fails, the error code identifier string for the error condition is
displayed with no timing data.
Run Selected Tests Click to run the selected tests
Control Description
C-arm Data Block
Device State Displays the device state code for the C-arm as an error code identifier
string. Possible values include:
E_OK - the C-arm is functioning normally.
E_EMULATE_I - the C-arm is being emulated.
E_NOT_READY_I - the C-arm has not yet recovered from a reset
E_UNEXPECTED_RESPONSE_I - the last query to read the C-arm status
failed
Xray Signal Level Displays the current X-ray energy level. Possible values include:
XRAY_OFF, XRAY_LO and XRAY_HI represent instantaneous states while
the X-rays are pulsing
XRAY_NONE indicates X-rays are off
Filter Wheel Click the Filter Wheel button to turn the filter wheel ON (spinning) or OFF
Filter Drum Position Displays the current position of the filter wheel as a single string.
“Hi” or “Lo” indicates HI_FILTER (brass section of the inner drum in the X-
ray beam) or LO_FILTER (air section of the inner drum in the X-ray beam)
“Bone”, “Tiss” or “Air”, indicates Bone (aluminum), Tissue (plastic) or Air
segment of the outer drum is in the X-ray beam
“H0” or “H1” indicates which half of the filter drum is nearest the X-ray
source
“P0”, “P1”, “P2” or “P3”, indicates which of the four X-ray pulses per filter
drum phase is currently taking place
Xrays Click the Xrays button to turn X-rays ON or OFF
Control Description
Hi Voltage Click the Hi Voltage button to turn high voltage ON or OFF
Warm Up Click the Warm Up button to turn the X-ray warm-up current ON or OFF
Laser Click the Laser button to turn the positioning laser ON or OFF. Note: Only
works when Enable Laser is DISABLED.
Enable Laser Click the Enable Laser button to turn keypad control of the laser ON or OFF
Laser Duration To change the length of time the laser stays on, enter a new number in the
edit box and press Enter (only applies when the laser is turned on from the
keypad, i.e., when Enable Laser is ENABLED).
Filter Abort When the Filter Abort button is ENABLED a filter error will cause an abort.
When the Filter Abort button is DISABLED a filter error will not cause an
abort.
HiLo Abort When the HiLo Abort button is ENABLED a High-Low error will cause an
abort. When the HiLo Abort button is DISABLED a High-Low error will not
cause an abort.
Filter Errors Displays a count of how many times the filter drum was out of phase with
the X-ray generator
HiLo Errors Displays a count of how many times the X-ray signal level was wrong for
the phase
Frequency Displays the line current frequency
Cycle Interval Displays the time interval between system interrupts
Warm Up Delay To change the warm-up delay, enter a new number in the edit box and press
Enter
Hi Voltage Delay To change the X-rays On high voltage delay, enter a new number in the edit
box and press Enter
Driver version Displays the version number for the driver
Device version Displays the version number for the C-arm microprocessor
Xray Error Displays a status code for the X-ray generator as an error code identifier
string. E_OK indicates the X-ray generator is functioning normally.
Pulse Per Seg Displays the number of Xray pulses in each segment
Xray On Phase: Click one button in each sub-control to identify the filter wheel position
(filter, segment, pulse and drum half) at which X-rays will be turned on:
Filter Click HI or LO for desired filter
Segment Click BONE, AIR or TISSUE for desired segment
Pulse Click 0, 1, 2 or 3 for desired pulse
DrumHalf Click 0 or 1 for desired drumhalf
Control Description
Xray Order Indicates the voltage level of the corresponding X-ray pulse during one filter
wheel segment
AC Line High Displays the instantaneous state of the AC line, which may be high (TRUE)
or low (FALSE).
Filter Drum Locked Displays TRUE when the filter drum is locked (preventing it from spinning),
displays FALSE when the filter drum is not locked
Brass Segment Displays TRUE when the filter drum brass filter is in the X-ray beam,
displays FALSE when the brass filter is not in the X-ray beam
Aperture Position Displays the position of the aperture plate in mils
Attenuator Position Displays the simulated position of the attenuator in mils
Xray Mode The edit box displays a numeric code value indicating X-ray voltage, power
and timing. Click the up or down arrows to increase or decrease the value
displayed in the edit control and change the X-ray voltage, power and
timing accordingly.
mA kVp Monitor Clicking the mA kVp Monitor button turns monitoring on very briefly
(fraction of a second) and the mA and kVp values are updated.
If X-rays are turned on, the monitor turns on (for almost a second) then off,
and the mA and kVp values are continuously updated during that time.
When X-rays are turned off, the monitor turns on (for almost a second) then
off, and the mA and kVp values are continuously updated during that time.
An error is generated if X-rays fail to turn on, or turn off.
mA Displays the X-ray current in milliamps (as reported by the C-arm
microprocessor)
kVp Displays the X-ray voltage in kilovolts (as reported by the C-arm
microprocessor)
Tests Block
Emergency Relay When run, this test first verifies that an emergency stop condition does not
exist. Then it sends an emergency stop command to the C-arm and verifies
that an emergency stop condition is asserted.
If the test completes successfully, PASSED is displayed. If the test fails, the
error code identifier string for the error condition is displayed.
Laser When run, this test turns on the laser positioning light and prompts: “Did
the positioning light in the C-arm turn on?”
Clicking Yes causes the test to proceed to the next step.
Clicking No ends the test with a failed status and error result.
The test then turns off the laser positioning light, and prompts: “Did the
positioning light in the C-arm turn off?”
Clicking Yes causes the test to be successfully completed.
Clicking No ends the test with a failed status and error result.
Results are displayed in the Test Results window.
Control Description
High Voltage When run, this test turns on the C-arm high voltage relay and prompts: “Did
the high voltage contactor (relay) turn on?”
Clicking Yes causes the test to proceed to the next step.
Clicking No ends the test with a failed status and error result.
The test then turns off the C-arm high voltage relay, and prompts: “Did the
high voltage contactor (relay) turn off?”
Clicking Yes causes the test to be successfully completed.
Clicking No ends the test with a failed status and error result.
Results are displayed in the Test Results window.
Drum Synch When run, this test stops the filter drum, then starts the filter drum and
measures the time it takes the drum to synchronize to the A/C line
frequency.
Results are displayed in the Test Results window.
Aperture When run, this test moves the aperture plate in both directions, checks the
encoder readings, and computes the distance per step. Then it moves the
aperture plate to the position of each aperture slot and records the encoder
reading.
Results are displayed in the Test Results window.
Run Selected Tests Click to run the selected tests
Calibrate Aperture Click to run the aperture calibration.
The aperture calibration procedure begins by moving the aperture plate
over its entire range and collecting X-ray levels at each point. A graph of X-
ray intensity versus plate position is displayed (Aperture Calibration Scan)
during data collection. When data collection is complete the Continue
button is enabled.
When Continue is pressed the peaks and minima in the intensity graph are
computed. Offsets in mils, for each peak, are labeled as “Steps” and are
graphed against the measured encoder values. This plot is displayed
(Aperture Calibration Plot) and the Continue button is enabled. When
Continue is pressed the peaks and minima in the intensity graph are
determined and displayed (Aperture Calibration Peaks). The Continue
button is enabled.
When Continue is pressed a table of the positions of peaks and minima
intensity is displayed (Aperture Calibration Data). The peaks represent the
positions of the aperture slots. The operator is queried as to whether the
results should be saved or discarded.
Commands Block
Move Absolute To move the aperture plate to a specific position, enter a number in the
Move Absolute edit box and click Move Absolute. The number is
interpreted as a location of the aperture plate, in mils.
Move Relative To move the aperture plate a specified distance from the current position,
enter a number in the Move Relative edit box and click Move Relative. The
number is interpreted as a distance to move the aperture plate, in mils. The
distance may be positive or negative.
Control Description
Emulate Click the Emulate button to set the C-arm emulation mode to the value
selected in the list box. The list box values are:
NONE - no emulation
FULL - emulate all C-arm devices
DRUM - just emulate the filter wheel
Note: The box does not display the current emulation state (there is no way
to read it).
Return Code: Following the execution of a Move Absolute, Move Relative or Emulate
command, the Return Code field displays the value returned by the
command, either E_OK or as an error code identifier string.
1. For descriptions of controls on the Common Data Block see Table 91.
Control Description
TZ Motor Data Block
Control Description
Device State Displays the device state code for the motor. Possible values include:
E_OK - the motor is functioning normally
E_EMULATE_I - the motor is being emulated
E_INVALID_PARAM_I, E_TOO_LONG_I, E_NOT READY_I or
E_WAITING_I - the last query to read the motor status failed
Pedestal Mode Displays the TZ motor mode as either OPERATIONAL or
MANUAL. The TZ motor only accepts commands when the mode is
OPERATIONAL.
Table Status Displays the status of the MOTOR_TZ system. The possible values are:
ACTIVE, TIMEOUT, BADALIGN, OK, INACTIVE,
SERVICE, CALIBRATE and RESET.
Pos Motion Switch Click to start the motor moving in the positive direction
(ENABLED). The motor continues to move until the button is
pressed again. Click again to stop the motor (DISABLED).
Neg Motion Switch Click to start the motor moving in the negative direction
(ENABLED). The motor continues to move until the button is
pressed again. Click again to stop the motor (DISABLED).
Left Pedestal Displays the status of the left pedestal. The possible values are
RUNNING, STALLED, STOPPED and AT_TARGET.
Right Pedestal Displays the status of the right pedestal. The possible values are
RUNNING, STALLED, STOPPED and AT_TARGET.
Current Position Displays the table position in microns.
Pos Limit Position Displays the table’s upper limit position (greatest allowed position
reading) in microns.
To change the upper limit position, enter a new number in the edit
box and press Enter.
Neg Limit Position Displays the table’s lower limit position (smallest allowed position
reading) in microns.
To change the lower limit position, enter a new number in the edit
box and press Enter.
Encoder Position Displays the position reported by the encoder, in encoder ticks.
Encoder Neg Limit Displays the lower pedestal encoder limit, in encoder ticks.
Encoder Pos Limit Displays the upper pedestal encoder limit, in encoder ticks.
Position Average The edit box displays the number of encoder readings that are
averaged per position reading.
To change, enter a new number in the edit box and press Enter.
Driver Version Displays the version number for the driver
Device Version Displays the version number for the motor controller microprocessor
Current SQDRIVER.INI Displays the upper left and right, and the lower left and right, pedestal
Limits (ticks) encoder limits (in encoder ticks).
Control Description
Tests Block
Motor Encoder When run, this test moves the TZ motor in both directions, and
measures the encoder distance.
Results are displayed in the Test Results window.
Run Selected Tests Click to run the selected tests
Calibrate Click to run the TZ motor calibration.
The calibration routine moves the table to its lowest position, and prompts
the operator to mark the position before continuing. When complete,
click OK.
Next, the calibration routine moves the table to its highest position and
prompts the operator to measure and enter (in cm) the distance moved.
When complete, click OK.
The operator is then queried as to whether the entered value is correct. If
correct, click Yes.
Finally, a complete set of motor calibration values are displayed and the
operator is queried as to whether the results should be saved.
Click Save, or Cancel, when prompted.
Commands Block
Move Absolute To move the motor to a specific position, enter a number in the Move
Absolute edit box and click Move Absolute. The number is interpreted as a
location of the motor, in microns.
Move Relative To move the motor a specified distance from the current position, enter a
number in the Move Relative edit box and click Move Relative. The number
is interpreted as a distance to move the motor, in microns. The distance may
be positive or negative.
Emulate Change the value in the list box and click the Emulate button to set the
motor emulation mode to ON or OFF. Note: The box does not display the
current emulation state (there is no way to read it).
Return Code Following the execution of a Move Absolute, Move Relative or Emulate
command, the Return Code field displays the value returned by the
command, either E_OK or as an error code identifier string.
Control Description
AR (AY, TX, TY) Motor Data Block
Motion State Displays the current state of the motor. The values are:
NONE - stopped
ACCEL - coming up to speed
STEP - at speed
DECEL - stopping
Device State Displays the device state code for the motor. Possible values include:
E_OK - the motor is functioning normally
E_EMULATE_I - the motor is being emulated
E_EMERGENCY_STOP_I - the Emergency Stop button is pressed
E_INTERLOCK_INHIBIT_I - no A/C line interrupts have been
received for too long a period, and the system has been shut down
E_NOT_READY_I - the motor has not yet recovered from a reset
E_INVALID_PARAM_I or E_TIMEOUT_I - the last query to read the
motor status failed
Control Description
Motion Error Displays the error code value for the motor. Possible values include:
E_OK - the motor is functioning normally
E_NO_MOTION_I - the motor has failed to move in response to a
command
E_MOTION_REVERSED_I - the motor has moved in the wrong
direction
E_TOLERANCE_I - the motor stopped too far from the target position
Pos Motion Switch Click to start the motor moving in the positive direction (ENABLED).
The motor continues to move until the button is pressed again. Click
again to stop the motor (DISABLED).
Neg Motion Switch Click to start the motor moving in the negative direction (ENABLED).
The motor continues to move until the button is pressed again. Click
again to stop the motor (DISABLED).
Pos Limit Switch Displays the current state of the positive limit switch. The values are:
NONE - not at a limit
SOFT - at a position limit
HARD - at a switch limit
BOTH - at both kinds of limit
UNKNOWN
Neg Limit Switch Displays the current state of the negative limit switch. The values are:
NONE - not at a limit
SOFT - at a position limit
HARD - at a switch limit
BOTH - at both kinds of limit
UNKNOWN
Position Displays the current motor position in microns (µm) (for the AR
Motor, the position is displayed in arc seconds) and in encoder ticks.
Move State Displays a string indicating the current status of most recent move
command. The possible values are:
DONE - last move completed successfully
START - move started by background thread
ACTIVE - move acknowledged by foreground thread
STOP - move stopped by background thread)
ABORT - move stopped by foreground thread
FAILED - last move failed
First Rate The current stepping rate for the motor at start-up, in microns (µm)
(for the AR Motor, arc sec) per second, is displayed in the edit box and
in steps per second in the static text field.
To change the initial stepping rate, enter a new number in the edit box
and press Enter.
Control Description
Final Rate The current stepping rate for the motor at full speed, in microns (µm)
(for the AR Motor, arc sec) per second, is displayed in the edit box and
in steps per second in the static text field.
To change the final stepping rate, enter a new number in the edit box
and press Enter.
Hold Power Displays the “resting” power level for the motor. Possible values are
OFF, LOW, MEDIUM, HIGH and MAX.
To change the resting power level, enter the first letter of a power level
in the edit box or click the spin control
Step Power Displays the active power level for the motor. Possible values are OFF,
LOW, MEDIUM, HIGH and MAX.
To change the active power level, enter the first letter of a power level
in the edit box or click the spin control
Accel Distance Displays the distance in microns (µm) (for the AR Motor, arc sec)
allowed for the motor to accelerate to full speed.
To change the acceleration distance, enter the new number in the edit
box and press Enter.
Accel Type Click the Accel Type button to set the acceleration profile to:
PARA - parabolic
EXP - exponential
LIN - linear
Min Motion The edit box displays the minimum time delay to wait to see if the
motor has moved, in power line cycles.
To change the minimum time delay, enter a new number in the edit
box and press Enter.
Motion Detect Click Motion Detect to toggle the state of motion detection between
ENABLED and DISABLED
Pos Limit Position The greatest allowed position reading, in microns (µm) (for the AR
Motor, arc sec), is displayed in the edit box and in encoder ticks in the
static text field.
To change the greatest allowed position value, enter a new number in
the edit box and press Enter.
Neg Limit Position The smallest allowed position reading, in microns (µm) (for the AR
Motor, arc sec), is displayed in the edit box and in encoder ticks in the
static text field.
To change the smallest allowed position value, enter a new number in
the edit box and press Enter.
Driver Version Displays the version number for the driver
Device Version Displays the version number for the motor controller
Position Average The edit box displays the number of encoder readings that are
averaged to get a position reading.
To change, enter a new number in the edit box and press Enter.
Control Description
Rate Scale Factor Displays the scaling factor used in computing the actual stepping rate
for the motor
Step Status Displays a string indicating the current stepping status. The possible
values are:
STOPPED
ACCEL - accelerating
SLEW - stepping at a constant rate
DECEL - decelerating
BUSY - the motor controller is re-calculating the acceleration profile
RESET
Direction Displays a string indicating the direction the motor is moving in, or, if
it is stopped, the direction it would move in. The possible values are
POS (positive) and NEG (negative).
Input Switches Displays a hexadecimal number encoding the values of the general
purpose input switches (read back from the motor controller
microprocessor).
Accel Steps Displays the number of steps used for acceleration from FirstRate to
FinalRate.
Number of Steps Displays the total number of steps to take in the current move
(including AccelSteps).
Tests Block
Motor Encoder When run, this test moves the motor in both directions, and measures
the encoder distance.
Results are displayed in the Test Results window.
Step Size When run, this test moves the motor in both directions, and measures
both the distance moved and the number of steps moved. It computes
the step size in each direction from these measurements.
Results are displayed in the Test Results window.
Step Rate When run, this test moves the motor in both directions, and measures
both the time required to move and the number of steps moved. It
computes the step rate in each direction from these measurements.
Results are displayed in the Test Results window.
Run Selected Tests Click to run the selected tests
Control Description
Calibrate Click to run the motor calibration.
For all motors except AR, the calibration routine begins by moving the
motor to its maximum limit, and prompting the operator to indicate
when the limit has been reached. AR motor calibration moves to
specified intermediate positions before moving to the limits, but the
operator is prompted in the same manner for each position. Click OK,
or Cancel (to exit calibration) when prompted.
Next, for all motors except the TX motor, the calibration routine moves
the motor to its minimum limit, and prompts the operator to indicate
when the limit has been reached. TX motor calibration skips this step.
Click OK, or Cancel when prompted.
Next, the calibration routine moves the motor from its minimum limit
to its maximum limit and back, in small increments. During this
operation a graph of encoder readings versus position is displayed.
When the graph is complete, the calibration routine calculates and
displays regression lines for the graphed data. Click Continue, or
Cancel when prompted.
Finally, a complete set of motor calibration values are displayed and
the operator is queried as to whether the results should be saved. Click
Save, or Cancel when prompted.
Commands Block
Move Absolute To move the motor to a specific position, enter a number in the Move
Absolute edit box and click Move Absolute. The number is interpreted
as a location of the motor, in microns (µm) (arc seconds for the AR
Motor).
Move Relative To move the motor a specified distance from the current position, enter
a number in the Move Relative edit box and click Move Relative. The
number is interpreted as a distance to move the motor, in microns
(µm) (arc seconds for the AR Motor). The distance may be positive or
negative.
Emulate Change the value in the list box and click the Emulate button to set the
motor emulation mode to ON or OFF.
Note: The box does not display the current emulation state (there is no
way to read it).
Return Code: Following the execution of a Move Absolute, Move Relative or
Emulate command, the Return Code field displays the value returned
by the command, either E_OK or as an error code identifier string.
1. For descriptions of controls on the Common Data Block see Table 91.
AW-00335_001-0902
When the Graph button is active and the system is producing X-rays, the data is presented in a bar graph
(Figure 476). Scale markings on the bottom of the graph indicate detector numbers. Scale markings aling
the sides indicate the numeric value of the column height. A label alone the left half of the graph indicates
which numeric value for the column is being displayed (for field engineering purposes this will always be
Volts). A label above the right half of the graph indicates which drum segment is being displayed (for feild
engineering purpose this will always be Air Segment). The bar graph shaded green is the hi voltage, the
bar graph shaded red is lo voltage.
AW-00335_001-0903
When the Table button is active and the system is producing X-rays, numeric data is displayed
(Figure 476). A label above the left half of the graph indicates which value is being displayed (for field
engineering purpose this will always be Volts). The display has a vertical scroll bar on the right because
only a sixth of the data for all of the detectors is showing at one time.
AW-00335_001-0904
Warning: While the X-ray button is ON the system is producing X-rays, take proper
precautions.
Note The next set of controls have the label Filter Wheel Off Phase which indicates the
state of the Filter Wheel. These controls (HI/LO, BONE/AIR/TISSUE, 0/1/2/3
and 0/1) indicate where the Filter Wheel will stop when it is not moving.
HI/LO buttons
The default for these buttons is HI. The Field Engineer does not use the
HI/LO buttons and they should always remain the default.
Bone/Air/Tissue Buttons
The default for these buttons is BONE. The Field Engineer will not use the BONE/AIR/ TISSUE buttons
and they should always remain the default.
0/1/2/3 buttons
The default for these buttons is 0. The Field Engineer will not use the 0/1/2/3 buttons and they should
always remain the default.
0/1 buttons
The default for these buttons is 0. The Field Engineer will not use the 0/1 buttons and they should always
remain the default.
AW-00335_001-0906
Pulse 0/1/2/3
The default setting of this button is 1. This indicates what data is displayed and not what the scanner is
doing. The Field Engineer will not use the Pulse 0/1/2/3 buttons and they should always remain the
default.
Hi Gain
This is a spinner control with a range of 0 to 11. Changing Hi Gain changes the displayed detector values
for the Hi portions of the current segment. The associated actual gain values are shown in the following
table.
Lo Gain
This is a spinner control with a range of 0 to 11. Changing Lo Gain changes the displayed detector values
for the Lo portions of the current segment. The associated actual gain values are shown in Table 95.
This is a spinner control with a range of 0 to 8. The default mode is 3. This control determines the voltage,
current and duty cycle for the X-ray generator. Most scans use X-ray Mode 3, Horizon A Whole Body uses
1, Horizon W Whole Body uses 3 and IVA (single energy) uses 6.
The following table shows the full set of defined X-ray modes.
Aperture
This is a spinner control with a range of 0 to 13. This controls the slit through which the X-rays pass to limit
the dimensions of the X-ray beam. This control allows the selection of a specific aperture in the system and
can be used to test aperture movement. The number next to the spinner is the physical position of the aperture.
Note: This control is slow. If you touch it (any control) before it settles, you will get an error.
Average Voltage
This displays the average voltage across all detectors for the Hi Air segment
AW-00335_001-0907
12.5.8 Display
This has two buttons: GRAPH and TABLE. This controls what is seen on the Main Display. In GRAPH the
Main Display shows as a bar graph (Figure 476) with scale markings on the left side according to how
Scale is set (see below). It shows the lo (red tinted) and the hi (green tinted) voltages. In TABLE the Main
Display shows the data from the detectors as numeric data.
Note “lo” = X-rays generated with low kV, “hi” = X-rays generated with high kV.
12.5.9 Scale
This has three buttons: RAW, LOG and VOLTS. The Default is VOLTS. The Field Engineer will not use
RAW or LOG and the button should always remain the default (VOLTS). This control sets up the scale
markings on the Main Display when the Display is set to GRAPH and changes the units for the numeric
value when the TABLE is displayed.
Segment
This has three buttons: BONE, TISSUE and AIR. The default is AIR. This controls the data segment that are
displayed in the GRAPH. The Field Engineer will not use either BONE or TISSUE and the button should
always remain the default (AIR).
AW-00335_001-0908
Status Bits
Detector values come back from the scanner as sixteen bit numbers, traveling through some parts of the
hardware on sixteen separate wires. The Status Bits field displays the most recent value for each bit, and
color-codes whether that bit has ever changed. Red bits have not been changed, green bits have. The Reset
button sets all the bits to red. If a particular bit never turns green under any circumstances, the
corresponding wire may be broken.
Reset
The Reset button re-starts the process of computing the means and standard deviations for each filter-
segment. It resets the number of samples to zero. It also resets the bits in the Status Bits display.
Select Detector
This is an edit box that displays specific data about the selected detector. Select Detector has a default of
108 (the middle of the bar graph as shown in Figure 476). This displays both the currently selected detector
(number) and allows a new detector to be selected by inserting a number in the edit box. A new detector
can also be selected by dragging the vertical blue line on the Main Display bar graph (Figure 482). The
number displayed in the edit box will be automatically updated to reflect the line position.
AW-00335_001-0909
Show Details
Show Details defaults to off. When clicked, a small window pops up containing detailed information
about one detector, in tabular form (Figure 483). This information includes the current (most recent) value,
mean value, standard deviation and number of samples (values used to calculate the mean and standard
deviation) for each of the six filter segments. The pop-up window can be dragged wherever the user wants
and remains until Show Details is actively released.
Control Description
D.A.S. Data Block
Device State Displays the device state code for the DAS as an error code identifier string.
Possible values include:
E_OK - the DAS is functioning normally.
E_EMULATE_I - the DAS is being emulated.
E_EMERGENCY_STOP_I - the Emergency Stop button is pressed
E_INTERLOCK_INHIBIT_I - no A/C line interrupts have been received for
too long a period, and the system has been shut down
E_NOT_READY_I - the DAS has not yet recovered from a reset
E_TOO_LONG_I, E_TIME_OUT_I, E_BAD_BLOCK_I,
E_INVALID_ADDR_I, or E_CANCELLED_I - the last query to read the
DAS status failed
Total Readings Displays the number of detectors being read. Note: this value is only
updated in the drivers when the detectors are read. They are read when the
Detectors property page is being displayed, but not when the DAS property
page is being displayed. The result is that this field displays 0 as a value
until the operator views the Detectors property page.
First Channel Displays the array index of the first detector being read.
To change the first detector, enter a new number in the edit box and press
Enter.
Control Description
Last Channel Displays the array index of the last detector being read.
To change the last detector, enter a new number in the edit box and press
Enter.
Hi Gain Displays the current gain value for high voltage X-rays, as a code value.
A: displays the A/D gain multiplier
M: displays the MUX gain multiplier
To change the high voltage gain, enter a new number in the edit box or click
the up/down arrows to raise/lower the selected high voltage gain (code
value).
Lo Gain Displays the current gain value for low voltage X-rays, as a code value.
A: displays the A/D gain multiplier
M: displays the MUX gain multiplier
To change the low voltage gain, enter a new number in the edit box or click
the up/down arrows to raise/lower the selected low voltage gain (code
value).
Driver version Displays the version number for the driver
Device version Displays the version number for the DAS microprocessor
Sample Mode The current DAS sample mode is indicated by the active (pressed) button:
SINGLE or DUAL.
To set the DAS sample mode, press the button for the mode desired.
Test Signal Press the Test Signal button to toggle the state of the test signal between ON
and OFF
Control Description
Tests Block
Test Signal When run, this test reads the detectors, turns on the ADC test signal and
reads the detectors again. It verifies that the detector values increased.
Results are displayed in the Test Results window.
Performance Test When run, this test measures detector output with X-rays off, then measures
detector output with X-rays on, and calculates a performance factor for each
detector.
Results are displayed in the Test Results window.
Upon completion, the operator is prompted to save the performance factors
and the bad detector data.
Run Selected Tests Click to run the selected tests
Control Description
Plot Results Click Plot Results to display a series of plots of the Performance Test results
(Performance Test must be run before the Plot Results button becomes
active). The plots are:
Plot 1: PERFORMANCE TEST - plots NOISE / (SIGNAL - OFFSET) per A/D
Channel
Plot 2: NOISE - plots mean values of Noise measurements (in volts)
Plot 3: NOISE - plots Standard Deviations of Noise measurements
Plot 4: SIGNAL - plots mean values of Signal measurements
Plot 5: SIGNAL - plots Standard Deviations of Signal measurements
Plot 6: OFFSET - plots mean values of Offset measurements
Plot 7: OFFSET - plots Standard Deviations of Offset measurements
Commands Block
Emulate Click the Emulate button to set the DAS emulation mode to the value
selected in the list box. The list box values are:
NONE - no emulation
COMCON - emulate the DAS on the ComCon card
SPECIAL - emulate the DAS on the PC, with diagnostic information in
channels 0-6
PC - emulate the DAS on the PC
Note: The box does not display the current emulation state (there is no way
to read it).
Return Code: Following the execution of an Emulate command the Return Code field
displays the value returned by the command, either E_OK or as an error
code identifier string.
1. For descriptions of controls on the Common Data Block see Table 91.
Control Description
Panel Data Block
Device State Displays the device state code for the Panel as an error code identifier string.
Possible values include:
E_OK - the Panel is functioning normally
E_EMULATE_I - the Panel is being emulated
E_INVALID_PARAM_I, E_TOO_LONG_I, E_NOT READY_I or
E_WAITING_I - the last query to read the panel status failed
Reset Displays the state code of the hardware reset flag, either TRUE or FALSE
Blink Rate To change the blink cycle time, enter a new number (milliseconds) in the edit
box and press Enter.
Tape Switch Displays TRUE if the collision detection tape has registered a collision, or
FALSE if it has not
Switches The Switches column displays the current state (either OFF or PRESSED) of
each of the following switches: C-arm Left, C-arm Right, Table In, Table Out,
Load/Unload, Home, Enable and Laser.
Control Description
Lights The Lights column displays the current command state (either ON, OFF or
BLINK) for each of the following lights: C-arm Left, C-arm Right, Table In,
Table Out, Load/Unload, Home and Enable.
Click the up or down arrow on the spin control for each light to change the
state of that light.
Driver version Displays the version number for the driver
Device version Displays the version number for the Panel microprocessor
Tests Block
Switches When run, this test prompts the operator to press the switches on the control
panel one switch at a time and observe that a light turns on or off in
response. It then prompts: “Did all the switches operate properly?” Click
Yes or No.
Results are displayed in the Test Results window.
Lights When run, this test sets all the lights on the control panel to the flashing
state. It then prompts: “Are the lamps on the control panel flashing?” Click
Yes or No.
Results are displayed in the Test Results window.
Emergency Stop When run, this test prompts the operator to press the Emergency Stop
switch.
Results are displayed in the Test Results window.
Emergency Strip When run, this test prompts the operator to press the emergency strip
switch located underneath or on the side of the table.
Results are displayed in the Test Results window.
Run Selected Tests Click to run the selected tests
Commands Block
Emulate Click the Emulate button to set the Control Panel emulation mode to either
ON or OFF. Note: The box does not display the current emulation state
(there is no way to read it).
Return Code: Following the execution of an Emulate command the Return Code field
displays the value returned by the command, either E_OK or as an error
code identifier string.
1. For descriptions of controls on the Common Data Block see Table 91.
Control Description
State Machine Data Block
Device State Displays the state machine status code. Possible values are:
NO_STATE - nothing loaded
PROGRAM - state program loaded
TASK - task table loaded
READY - state program and task table loaded
EXEC - running
BREAK - stopped by break
ABORT - stopped by abort
ERROR - stopped by error
Number of Tasks Displays the number of loaded tasks
Number of Buffers Displays the number of buffers allocated
Size of Buffers Displays the number of bytes per buffer
Number of Registers Displays the size of the array of registers
Size of Program Displays the number of bytes in the loaded program
Buffer Address Displays the address of the currently selected buffer (in standard memory
address format)
Control Description
Flags The Flags control consists of a row of sixteen buttons (one for each flag in
the Flags array). Click a button to toggle the corresponding flag bit between
T (set) and F (cleared).
Tasks The Tasks control consists of a column labeled “Task[x]:”, where x ranges
from 0 to 7, and adjoining columns labeled “Index and “Status.”
The Index column displays the index of the current instruction for the
corresponding task. The index is the offset of the instruction in the Program
array.The Status column displays the status of the corresponding task. The
status is an error code identifier string.
Program Address Displays the address of the loaded program (in standard memory address
format)
Driver version Displays the version number for the driver
Time Out Enable Displays whether or not the state machine tests for the time out condition.
Possible values are TRUE and FALSE.
Register Address Displays the address of the Registers array (in standard memory address
format)
Register[x] Displays “Register[x]:” where x is the index of the currently selected
register in the Registers array.
To change the value of the currently selected register, edit the value in the
edit box or select a new value from the drop-down list.
1. For descriptions of controls on the Common Data Block see Table 91 on Page 452.
13 Preventive Maintenance
This section lists the procedures that should be performed by trained service personnel at least once per
year, and preferably at six-month intervals
Use and complete form CSD-0042-F06 Bone Densitometer PM Checklist while performing the Preventive
Maintenance.
Note The Horizon system meets applicable FDA radiation performance standards
through its useful and expected life provided no components or parts are
removed from the system and no unauthorized adjustment or unauthorized
replacement of certified components is performed.
Caution Do Not Recalibrate The Machine. Call Technical Support if drift or a shift is
noted.
4. In the QDR > LogFiles folder, check the ERROR.LOG file for any errors.
Disk Cleanup
1. Exit QDR without shutdown.
2. Then run the Windows utility Disk Cleanup.
For each drive belt (C-arm Y, C-arm R, Table X, and Table Y) perform the following:
1. Loosen the 2 mounting bolts holding the tension block.
2. Tighten the tension nut so that the spring is compressed to 7/8 in. The bracket cutout can be used as
a measuring guide (the inside of the washer should be flush with the bracket cut). For the AR
tension spring, adjust to 7/8 in. from the inside of one washer to the inside of the other.
3. Tighten the two mounting bolts holding each drive belt’s tension block.
Note It is not necessary to perform the MOTOR calibration procedures after
tensioning the motor drive belts.
11. Test all tape switches. The tape switches are located along the length of the front and back of the
table, and mounted on the frame under the table. Verify that pressing a tape switch immediately
stops all C-arm motion, table motion, and X-ray production. X-ray production should be monitored
by a Victoreen 450P or equivalent.
Note The front (table) tape switch is disabled until the C-arm is at 0°, so this switch
should be checked with the C-arm at a position greater than 30°.
Measure kV
1. Connect oscilloscope 10X probe ground connection to SRC gnd.
2. Connect the oscilloscope 10X probe
3. channel 1 to kV mon.
4. Configure the oscilloscope:
5. set time to 5.00ms
6. voltage to 1 volt/division
7. DC Coupled
8. move the ground trace line to bottom of screen
9. Press F2 to turn on X-rays.
10. Observe the oscilloscope max kV pulse amplitude (Figure 489).
KVP: Divide the voltage displayed on the scope screen by 0.05625 to obtain the actual kV.
Measure mA
11. Move oscilloscope 10X probe channel 1 to mA mon.
12. Configure the oscilloscope: Change time to 10.0mS
13. Press F2 to turn on X-rays observe oscilloscope max 10mA pulse.
14. Press F2 to turn off x-rays.
15. Change set up to Hi Gain 2, Low Gain 1, Xray Mode 1, Aperture 7.
16. Press F2 to turn on X-rays observe oscilloscope max 3mA pulse (Figure 490).
mA: Divide the displayed voltage by 0.75 to obtain the actual mA.
Figure 490 Example: 2.3V /0.75 = 3.06 mA
Virus Check
1. Run a virus checker that is approved by Hologic for use on bone densitometer systems.
2. Refer to MAN-00508 for details about approved virus check software products.
Guide Bearing
Fitting Cap
Seal
Guide Rail
Consult Hologic Technical Support for more information on obtaining the proper tools, material, and
procedure.
1. Voltage indicators = green; motor drive status = red. If red led is on, there is a failure in the PCB. The CB
should be tripped.
Jumpers
LED1 PWR_OK
LEDs Voltage /
Signal Source Jumpers
LED10 On +5VDC Int. Voltage JW1 (NORMAL) In
Regulator
LED9 On +24VDC Ext. Signal Dist. JW3 (TEST) Out
LED8 Flash STATUS - - NORMAL/ SERVICE NORM
(4 LED group)
LED11 ON +3.3VDC Int. Voltage
Regulator
Table 105 64-Channel Detector Array Assembly (Horizon C, Ci, W, Wi) PCB-00889
D12 On LOCKED
X-ray
15 Troubleshooting
Contact the HelpDesk for assistance with any hardware and software issue. Do not order major parts until
you have performed a thorough diagnostic. Then contact the HelpDesk about the issue.
Provide the following information:
• serial number
• model of the unit
• software version
• installation site
• error message number
Note Before ordering any replacement parts, always verify that the system is properly
calibrated. The calibration process can help to identify mechanical problems.
Note The numbers on the left (in red) in Figure 493 are line numbers, not Index
numbers.
• A defective part: Pedestal, TZ Motor Controller, String Encoder problem, or a loose cable from the
Distribution board to the AR/TZ Motor Controller (if constant or very often).
• An improperly leveled table. Incorrect calibration of the AR Motor. Check the table for level at both
the head and foot end of the table from front to back of the table. The value you must have is 0.0
degrees front to back. If it is not, adjust it and then re-calibrate the AR Motor.
• Upgrade Required: This error code may also be an indication of the need for the TZ fix which
changes an EPROM on the TZ Controller board (P/N ASY-00549).
Error Code Index 318 during QC. X-ray Hi/lo Does Not Match Filter Drum Position.
This error message can occur in two scenarios: Software setup during an upgrade or Filter Drum problems
(loose belts, pulleys, etc.).
Scenario 1
When upgrading to Horizon or APEX, the HOLOGIC.INI is rewritten to read APER=False. The Horizon
Upgrade Tools Disk should change this line to True for a movable aperture. If it remains =False and you
will get either an Index 309 and/or an Index 318 error message. Check HOLOGIC.INI and modify the line if
necessary.
Scenario 2
The other time when you can get this error message is when the filter drum position encoders at the rear of
the filter drum are dirty and read the filter drum position inconsistently. Normally, the drum reads as
follows:
(Clean Detectors) Air-Lo, Air-High, Tissue-Lo, Tissue-Hi, Bone-Lo, Bone-Hi (Dirty Detectors) Air-Lo, Air-
Lo, Tissue-Lo, Tissue-Hi, Bone-Lo, Bone-Lo
It has also been reported that this error can also be caused by a sticky aperture assembly.
In the example above, the dirty detector caused you to miss the Air-Hi and Bone-Hi pulses. But, because of
the intermittent nature of the error, you will get Index 318 errors. If you had a completely dead detector(s),
you would be more likely to get X-rays do not match filter drum position.
To fix this problem, use compressed air to clean out the photo detectors. Index Error 323
This is usually caused by either a bad C-arm Interface board, P/N 140-0090 a defective Filter Drum, or X-ray
Source Assembly/XRC.
3. Go to Start > Run > MSCONFIG and go to the Startup tab (Figure 495).
4. Deselect any that are not on this list.
Index
A Boards
Distribution ....................................................21
A/D Gain Control Adjustment 141, 296, 313, 362 Motor Controller ...........................................21
Activation PCIe Communication Controller ................19
APEX Reporting Engine ............................420
Adjustments C
A/D Gain Control ...............141, 296, 313, 362
Cables, Upper C-arm ...........................................70
Belts ...............................................................106
Calibration
C-arm Parallelism .......................................106
Cross-Calibration ............................... 168, 173
Encoders .......................................................109
Calibration Errors ...............................................498
Filter Drum Assembly Left to Right 129, 288,
Calibrations .........................................................108
..........................................................354
after FRU Replacement or Repair .............281
Front to Back T-Rail and Table Edge/Rail Gap
Aperture ............................. 134, 293, 326, 359
Adjustment ....................................105
AR (Horizon A) ...........................................117
Motor Drive Belts ........................................481
AY (Horizon A, W) .....................................110
Sled ............................................... 129, 288, 354
Field Service 157, 300, 316, 331, 366, 373, 411
Table Edge to T-Rail ...................................104
HVPS Assembly Replacement ..................282
Alignments
Motors ...........................................................107
after FRU Replacement or Repair .............281
Requirements, after FRU Replacement ....281
Checks ...........................................................481
Sequence .......................................................107
Cross-Table X-ray Beam ............ 130, 289, 355
TX ..................................................................115
Requirements, after FRU Replacement ....281
Tz Motor .......................................................108
Skew Filter Drum Base .............. 128, 287, 353
C-arm
Table ..............................................................103
Hardware .......................................................70
X-ray Beam, Cross-Table ........... 130, 289, 355
Installation
Amplitude, Voltage .......................... 130, 289, 355
Horizon A ........................................ 66, 69
Aperture Assembly ..............................................16
Horizon C, Ci, W and Wi ......................68
Aperture Calibration ................134, 293, 326, 359
Interface Board ........................................... 8–9
Aperture Motor
Parallelism Adjustment (Horizon A) .......106
Assembly ..........................................................9
Subsystem ........................................................9
APEX 5.6 or later, installation ..........................208
Central DICOM Server Installation .................439
APEX Reporting Engine ....................................418
Checklists
APEX Reporting Engine Activation ................420
after Computer Replacement ....................201
AR Motor Cover, remove ..................................119
prior to Computer Replacement ...............199
B Replace DXA CPU before Software Upgrade
..........................................................199
Balance Detectors .............................. 132, 291, 357 Circuit Board Function, System .........................10
Base Frame Level .......................................... 71–72 Circuit Breaker ......................................................77
Beam Filter ..............................................................4 Clearances
Beam Filtration .....................................................16 Horizon A .......................................................53
Beam Quality .................................................16 Horizon C,Ci ..................................................54
Beam Quality ........................................................16 Horizon W,Wi ...............................................55
Belt Adjustment ..................................................106 Components
Belt Tension .........................................................106 C-arm Interface Board ............................... 8–9
Block Diagram C-arm Subsystem ............................................9
Distribution and Motor Control .................10 Detector Board ...............................................10
Distribution Board Distribution Board ................................... 7, 10
Interconnection ......................................19 Interconnection Diagram ......................21
Horizon C-arm Block Diagram .....................9 Interface .....................................................7
Horizon Operator Console ............................7 Routing ......................................................8
Horizon System ...............................................6 Motor Control Subsystem ..............................8
PCIe Communication Controller Board ....19 Pentium PC ......................................................7
BMC, BMD, AND AREA Gold Standard System Configuration
Averages ..............................................................169 Hologic Connect ................................. 446–447
Configurations E
CentralDICOMServer .................................444
DICOM .........................................................213 Emergency Motion .............................................479
IP Address and Subnet Mask ....................217 EMI Cables, replace ...........................................274
Network .............................................. 198, 206 Encoder, adjust ...................................................109
Connectivity Survey Form ................................203 Error Messages, QDR ........................................500
Control Panel Board ............................................10 Exam Mode Performance ....................................13
Corrective Service Procedures .........................198
Counterweights, C-arm .......................................67 F
Cross-Calibration ...................................... 168, 173 Field Replacement Units .......................... 275–280
Cross-Table X-ray Beam Alignment 130, 289, 355 Cables ............................................................279
D Computer and Isolation Module ..............278
Electronics Tray ...........................................275
Daily Calibration Internal Reference System .....4 Internal Computer Assemblies .................278
Destination Path Measurement ..........................58 Lower C-arm ................................................277
Detector Assembly Miscellaneous ..............................................279
Screw Locations ......................... 133, 292, 358 Mobile ...........................................................280
Detector Flattening ..142, 298, 314, 330, 364, 409 Special Tools ................................................280
Detector Signal Levels Check .......... 130, 289, 355 Table X ..........................................................276
Detectors, balance ............................. 132, 291, 357 Table Y ..........................................................276
Determining Alignment and Calibration Re- Upper C-arm ................................................278
quirements ..........................................................281 Field Service Calibrations 157, 300, 316, 331, 366,
Deviation from Indicated Values, Maximum ..16 ....................................................................... 373, 411
Diagram, Wiring Filter Drum Assembly ...............4–5, 8–9, 16, 241
Horizon C,Ci DIN Rail AC Input/Output .82 Aperture Controller ......................................11
Horizon W,Wi DIN Rail AC Input/Output ... Aperture replacement ................................242
82 Interface ..........................................................11
Diagrams, see Block Diagrams Left to Right Adjustment ..........129, 288, 354
DICOM Flattening, Detector ..142, 298, 314, 330, 364, 409
Configuration ..............................................213 Free and Public DNS Servers ................... 437–438
Host Configuration .....................................213 Front to Back T-Rail and Table Edge/Rail Gap
Server Configuration ..................................444 Adjustment ..........................................................105
Testing ..........................................................218 FRU, see Field Replacement Unit
Dimensions ...........................................................13
Horizon A ......................................................53 G
Horizon C,Ci ..................................................54 Gold Standard Phantom Label .........................170
Horizon W,Wi ...............................................55 Gold Standard Phantom Scan ..........................174
DIN Rail Guide Rail and Bearing Maintenance .............485
AC Input/Output Wiring Diagram ............82
Horizon ...........................................................74 H
Horizon C, Ci .................................................75
W, Wi ..............................................................75 Hardware
Distribution & Motor Control Block Diagram .10 C-arm ..............................................................70
Distribution Board ......................... 7–8, 10, 20–21 Filter Drum Assembly ....................................4
Interconnection Diagram .............................21 Hologic Connect
Interface ............................................................7 Service Verification .....................................448
Interface Connections ...................................20 Hologic Connect Configuration .............. 446–447
Power ..............................................................20 Hologic Connect Installation ................... 446–447
Routing .............................................................8 Horizon C-arm
DNS Servers ............................................... 437–439 Block Diagram .................................................9
Documentation, required ....................................48 Hardware .......................................................70
Doorway Size, minimum ....................................49 Installation
Drive and Monitor Issues .................................497 Horizon A ........................................ 66, 69
Duty Cycle .............................................................15 Horizon C, Ci, W and Wi ......................68
Interface Board ........................................... 8–9
Parallelism Adjustment (Horizon A) .......106
Subsystem ........................................................9
www.hologic.com | info@hologic.com