Saracom TMC Manual
Saracom TMC Manual
<INDEX>
AA. TMC UNIT Drawing.
BB. Installing the Fluxgate sensor.
CC. Calibration/Deviation procedure.
DD. TMC Calibration/Deviation table.
EE. Dip switch setting.
ADJUSTING THE FLUXGATE(TM-42) (#2)
1. ADJUSTING THE SF-30(FLUXGATE) 2. DEVIATION ADJSUTMENT.
1 2 3
PRESS “DEV/CAL” BUTTON THE “CALu” WILL BE DISPLAYED. THE “CALd “WILL BE DISPLAYED.
CONTINUOUSLY MORE THAN PRESS SHORTLY 1TIME AGAIN PRESS SHORTLY “DEV/CAL”
3 SECOND(LONG TIME) FOR “CALd” BUTTON
4 5 6
AA) PRESS “DEV/CAL” LONG TIME. THE SELECT “CALd” by “DIM +” BUTTON. THE PRESS SHORT
TIME TO ENTER A000 MODE THEN PRESS “DIM+ or “DIM-” button to go “A200”(200~210DEGREE)
BB) THEN PRESS “DEV/CAL” BUTTON SHORT TIME.
CC) THEN PRESS “DIM+” SHORT TIME AND ADD THE VALUE. “0.3”
DD) THEN PRESS LONG TIME TO EXIT THE MENU.
NOTICE : IF THE VALUE IS NOT ADDED, PLEASE ALSO TRY “A190” OR “A210”.
DEPARTMENT OF SERVICE(A/S)
ERROR MESSAGE/TROUBLESHOOTING
A. SYMPTOM :
i) “ErrF” - Installation stage : Possibly wrong connection.
- After delivery : Loose connection/defective SF-30(Fluxgate sensor).