Professional Documents
Culture Documents
Omnia RM en 15.0
Omnia RM en 15.0
Omnia RM en 15.0
Reference Manual
COSMED Srl
This manual has been written to the best of COSMED’s knowledge and in accordance
with the state of the art.
However, this does not exclude the possibility of errors or inaccuracies.
The images shown in this manual are for illustration purposes only and may not be an
exact representation of the product or its parts.
COSMED assumes no liability for the end user interpretation of this manual or for
casual or consequential damages from the provision, representation or use of this
documentation.
No parts of this manual may be reproduced or transmitted in any form without the
expressed written permission of COSMED.
This manual has been written in accordance with IEC 60601-1 and IEC 82079 standards.
All trademarks, registered trademarks and logos mentioned in this reference manual
are property of their respective owners.
www.cosmed.com
Table of Contents
1 - Get Started
This reference manual describes OMNIA with the most complete configuration with
all available modules. Detailed descriptions on how to run tests are demanded to
the device user manual.
Note: This reference manual can be used in conjunction with the user manual of
the device intended to communicate with the software.
Note: OMNIA users are assumed to be competent with the functions and risks
associated with the connected device.
1.2.2 Warnings
OMNIA is used in conjunction with one or multiple devices. Please refer to the
device user manual for specific Warnings.
1.2.3 Precautions
Make sure you follow the procedures and recommendations reported in
section 11 - Maintenance to ensure your data are protected and secured
and your system includes the latest available updates for highest
performance and data reliability.
Carefully read sections 1.2.4, 1.2.5 and 1.2.6 to understand your
responsibility on sensitive data protection.
OMNIA is used in conjunction with one or multiple devices. Please refer to
the device user manual for specific Precautions.
As part of your security concept, you are responsible for establishing controls to
protect your network and any networked devices, including the PC where the
OMNIA SW is installed, against cyber security threats.
Network Cyber Security Measures
Only connect COSMED devices operated using OMNIA to networks which are
explicitly intended for that purpose.
If you connect the device to a network, it is highly recommended that you deploy
measures to protect the network against cyber security threats. This includes the
deployment of firewalls to limit remote access and antivirus protection of standard
computer systems.
Network Security Requirements
For managing risks in clinical network deployments, COSMED recommends that you
apply a formal process such as the IEC 80001 series of standards to address safety,
effectiveness, and data and system security.
OMNIA is used in conjunction with many COSMED devices, from spirometers to lung
functions systems, from metabolic carts to body composition equipment.
COSMED devices can communicate with OMNIA via USB, RS-232 or Bluetooth®
connections.
OMNIA software allows the user to:
Calibrate the device.
Perform test.
Manage subject data.
Review and edit test results.
The table below lists all COSMED devices compatible with OMNIA; the software
communication type and available software features are listed for each device:
Note: From this point on “Quark CPET” refers to instructions which relate to
Quark CPET O2 Paramagnetic and Quark CPET O2 GFC.
The table below lists all OMNIA modules indicating the compatible devices:
Compatible
Pony FX MIP/MEP
Spiropalm 6MWT
Device
Quark RMR
Quark i2m
Quark PFT
Q-NRG+
Pony FX
Q-NRG
Module
Q-Box
K4b2
Code
K5
A-670-100-001 x x x x x x x x x x ● ● ● x x x
A-670-100-002 ● ○ ○ ● x ● x x x x x x x x x x
A-670-100-003 ○ x x x x ● x x x x x x x x x x
A-670-100-004 ○ x x x x ○ x x x x x x x x x x
A-670-100-005 ○ ● ○ x x x x ● x x x x x x x x
A-670-100-006 x x ○ x x x x x x ○ x x x x x x
A-670-100-007 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
A-670-100-008 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
A-670-100-009 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
A-670-100-010 ○ ○ ○ x x x x ○ x x x x x x x x
A-670-100-011 ○ x x ○ ● ○ x x x x x x x x x x
A-670-100-012 ○ x x ○ x ○ x x x x x x x x x x
A-670-100-014 ○ x x x x ● x x x x x x x x x x
A-670-100-015 ○ x x x x x x x x x x x x x x x
A-670-100-016 x x ● x x x x x ○ ○ x x x x x x
A-670-100-022 ○ ● ● x x x x ● ○ ○ x x x x x x
A-670-100-023 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
A-670-100-025 x x x x x x x x x x x x x x ● ●
A-670-100-026 x x x x x x x x x x x x x x ○ ○
A-670-100-027 x x x x x x x x x x x x x x ○ ○
A-670-100-028 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
A-670-100-029 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
C04550-02-11 ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○ ○
2 - Installation
2.1 PC Requirements
Note: For OMNIA Network additional requirements apply. Please refer to OMNIA
Network Reference Manual (see section 1.1.5).
Note: The software is copyright protected and should only be installed from the
original CD/USB device.
Depending on the device you want to connect to OMNIA, the software and
hardware activation codes reside on a USB dongle or in the device itself.
Pony FX MIP/MEP *
Device
Spiropalm 6MWT *
BOD POD GS
Quark Spiro
Quark CPET
Quark RMR
Quark i2m
Quark PFT
Pony FX *
License
Q-NRG+
Q-NRG
Q-Box
Type K4b2
K5
USB
● ● ● ● ● ● ● ● ●
dongle
Onboard ● ● ● ● ● ● ●
* USB dongle for devices which have been upgraded to OMNIA software from
previous Suite software.
Note: All non-COSMED devices require a USB dongle if not used as auxiliary
devices.
Note: Each dongle is associated with a specific device SN (printed on the dongle
label) and only works with that device.
Power on the device and connect it to the PC via the available wired
connection (see section 1.3). The necessary drivers for communication will
be installed automatically by Windows®.
Run OMNIA and login (see section 2.3.1)
Authorization lasts 30 days and needs to be renewed if an authorized device has not
been connected during that period. OMNIA can be re-authorized only via a wired
communication.
Note: Carefully store the password in a safe place. See section 11.6.1 to reset the
password.
At this stage you are logged in as OMNIA Administrator and you can create and
configure users and user roles, see section 9.6.
First User login (Domain: Omnia)
Users must leave the password field blank to be redirected to the page where to set
their own password.
First User login (Active Directory Domain)
Users must use their usual domain password.
3 - Operating OMNIA
3.1 Overview
OMNIA user interface is based on a hierarchical structure. The software panels are
structured as a series of encapsulated boxes and, as a general rule, the functions
located inside the larger box have effects on the smaller ones.
For each panel, tiles and buttons are displayed according to the devices connected
to the software and the existence of any additional modules. This reference manual
describes OMNIA with the most complete configuration.
The user interface for each specific software section is carefully described in the
following chapters.
When navigating through the software the screen is divided in three main areas:
The main features of each area are described below. Certain features may only be
available in certain software sections or configurations.
Minimize
Resize
Close
When a tile is clicked, the corresponding software function opens and a new panel
populates the central area. All software panels are carefully described within this
reference manual.
Some tile may lead to other tiles in accordance with the hierarchical structure of the
software.
Home panel
The central area of the Home panel is populated as follows:
Database
Testing
Calibration
HOME
Quality Control
Utility
Resource Center
Functionalities and features of each tile of the Home panel are carefully described in
the following chapters of this reference manual.
OMNIA version
User logout
4 - Settings
4.1 Overview
Environment
Calibration
QC
Predicteds
Tests
Settings
ECG
Admin
Export
Licenses
Appearances
About
4.2 International
Settings Description
Language Select OMNIA language from the dropdown
Use HL7®-compatible Genders Check to enable the use of HL7® compatible genders.
HL7® genders include: Female, Male, Transgender,
Hermaphrodite/Undetermined, Other
Unit of Measurement Customize the unit of measurement for each
parameter
Note: Changes related to languages and unit of measurement are applied only
after restarting the software.
4.3 Environment
Environment
Settings Description
Temperature Enter the environmental temperature
Pressure Enter the barometric pressure
RH Enter the environmental relative humidity
Note: Device integrated sensors will overwrite the above default values at test
start and adjust the parameters above throughout the test.
Flowmeter
Settings Description
Enter the temperature at the flowmeter (default
Temperature
34°C)
Enter the relative humidity at the flowmeter (default
RH
100%)
BTPS
Settings Description
Apply Inspiratory BTPS Check to apply the BTPS correction factor during
Apply Expiratory BTPS Expiration / Inspiration
4.4 Calibration
4.4.1 Flowmeter
Syringe
Settings Description
Volume Enter the certified value of the calibration syringe
(default 3000 mL)
N2 Reference Gas
Settings Description
O2, CO2, O2 Enter the certified concentration of N2WO gas
cylinders (O2: default 16.0% , CO2: default 5.0%, O2
100%: default 100.0%)
Scale
Settings Description
Calibration Weight Enter the certified calibration weight for Scale
calibration procedure
Cylinders
Settings Description
Calibration Volume Enter the certified calibration Volume for Bod Pod
calibration (set by technician on installation)
Pediatric Calibration Volume Enter the certified Pediatric calibration Volume for
Bod Pod calibration (set by technician on installation)
Volume Phantom Value of an external Calibration Volume used to
calibrate Bod Pod (set by technician on installation).
Calibration Volume which was used for a previous
Bod Pod (GS, EX, etc.)
4.5 QC
4.5.1 Flowmeter
Verification
Settings Description
Accepted Range Enter the acceptable range for the three flow ranges
(Low, Medium and High) used during Flowmeter
Verification
Max Error Enter the maximum acceptable error for Flowmeter
Verification
Simulated
Settings Description
Simulator Volume Enter the net Volume of the Q-Box simulator
System Resistance Enter the simulator Resistance
System Dead Space Enter the simulator Dead Space
Subject inside QBox during test Check when the simulated test is performed with a
subject inside the Q-Box (e.g. Erlenmeyer Flask
simulated test where the bulb is squeezed by a
Subject)
4.6 Predicteds
All predicted sets are grouped by test type and users can select the sets from the
dropdown for each test type. A link to the reference literature is provided by the
software for each predicted set (see section 13.2). The applicable area (age, gender,
height,…) of predicted sets is defined according to the corresponding reference.
Options
Settings Description
Show z score Check to display the z-score for each parameter
(number of SDs from predicted value) when available
in the predicted set
Show Post BD results as % of Check to display Post BD results as a % of predicted
predicted values rather than % of baseline results
Use Extrapolated Predicted Check to extend the predicted set if the subject falls
Values outside the reference range of the predicted set
Note: Not every predicted set can be extrapolated. When predicted data cannot
be extrapolated, the corresponding predicted values will remain blank if the
values are outside of the ranges of the predicted set.
Calculation Method
Settings Description
LLN/ULN Calc Select the guidelines for the calculation of Lower
Limit of Normal and Upper Limit of Normal:
80%-120% of predicted
ATS/ERS 2005 [1] (Pred ±1.645 SD)
Height from ulna length formula Select the formula for the calculation of height using
ulna length from the dropdown menu
Hemoglobin Reference
Settings Description
Hb Ref. (Male) Enter the reference Hb level for Male population
(used in DLCO Hb correction, default 14.6 g/dL)
Hb Ref. (Female and Children) Enter the reference Hb level for Female and Children
population (used in DLCO Hb correction, default 13.4
g/dL)
4.7 Tests
4.7.1 Spirometry
General
Settings Description
Show QC Messages Show quality control messages according to ATS/ERS
2005 [2] or ATS/ERS 2019 [3] guidelines
Use breathing valve during Check when using the breathing valve during
spirometry spirometry to apply the correct calibration factors
Show calibration warning Check to show warning related to the necessity of daily
message calibration/verification
Start test immediately (Avoid Check to start the test immediately after selection from
pressing Start button) the Testing menu (removing the necessity to press the
Start button)
Charts
Settings Description
Autoscale F/V and V/t for small Check to auto scale the Flow/Volume and Volume/Time
curves curves: both axes are halved if curve points and related
predicteds are within the half of the standard range
Show entire FVC maneuver on Check to show the spirogram on V/t chart
V/t chart
Extend F/V loop scale up to 16 Check to extend the Flow axis up to 16 L/s
L/s
Show Normal Range on F/V Show standard deviation bars in correspondence of the
chart main predicted parameters (when available) on FVC
charts
4.7.2 DLCO
General
Settings Description
Show QC Messages Check to show quality control messages
Chart
Settings Description
Show mouth pressure signal Check to show mouth pressure signal during DLCO
(DLCO SB) Single Breath (useful as quality control feedback during
breath-hold; aids the detection of unwanted Muller or
Valsalva maneuvers)
Auto-Refresh CO graph during Check to refresh the CO graph during editing, according
editing to the Zero Wet correction
Show CO/CH4 conc. vs volume Check to show CO/CH4 concentration curve against
volume (ATS/ERS 2017 [4])
Correction Factors
Settings Description
COHb/CO Back Pressure Select COHb/CO Back Pressure correction:
None
Zero Wet (use a correction on Hb for
“anaemia effect” in addition to the Zero Wet
for CO back pressure).
Formula (use a single formula based on
COHb% for both “anaemia effect” and CO
back pressure)
ATS 2017 [4] (use a formula based on FaCO
for “anaemia effect” and Zero Wet for CO
back pressure)
Breath-hold
Settings Description
Time Enter the duration of Breath-Hold
Method Select method to calculate the Breath-Hold:
Jones and Meade (from beginning of the
inspiration to the beginning of alveolar
sampling collection)
OGILVIE (from 30% of the inspiratory time to
the middle of alveolar sampling collection)
ESP (from 50% of inspiratory volume to the
beginning of alveolar sampling collection).
Washout Volume
Settings Description
Method Select method to identify Washout volume (and related
Sampling volume):
Automatic: from the end of Breath-Hold to
Sampling (Sampling 500mL / 1000mL);
software automatically detects the flattening
of tracer gas curve and identify the start of
Sampling volume.
ATS/ERS 2005 [5]: 1000mL from the end of
Breath-Hold (Sampling 1000mL)
ATS/ERS 2017 [4]: end of Breath-Hold to
Sampling (Sampling 200mL); software
automatically detects the flattening of tracer
gas curve and identify the start of Sampling
volume.
CO Based: from end of Breath-Hold to
Sampling (Sampling 500mL); software
automatically detects the start of Sampling
Volume when the CO curve crosses the
regression line obtained considering the
points belonging to middle third of CO
concentration vs. volume
Calculation Method
Settings Description
Alveolar Volume Select method to calculate Alveolar Volume:
ATS/ERS 2005 [5]
Mass Balance (as per ATS/ERS 2017 [4])
Overall results Select method to calculate Overall results:
Average
ATS/ERS 2017 [4]
Anatomic dead space Select method to calculate Anatomic dead space:
ATS/ERS 2005 [5] (2,2mL·kg-1)
ATS/ERS 2017 [4] (2,2mL·kg-1 if BMI<30kg·m-2,
h2/189.4 if BMI≥30kg·m-2)
Fowler
TLC(DLCO) for Obstructed Check to apply the TLC(DLCO) correction for Obstructed
Subject Subject according to: Correction of Single-Breath
Helium Lung Volumes in Patients With Airflow
Obstruction: Naresh M. Punjabi, MD; David Shade, BA;
and Robert A. Wise, MD, FCCP. CHEST 1998; 114:907–
918
Calculation Method
Settings Description
sRaw Select method to calculate the sRaw:
Average
Median
FRC(Pleth) Select method to calculate the FRC(Pleth) if Custom
volumes calculation method is selected (see below):
Average
Median
Select volume calculation method for all tests in the archive from the following:
Settings Description
ATS Preferred FRC=average, ERV=average, RV=FRC-ERV, TLC=RV+VC
(ERV, FRC and VC are measured)
ATS Optional FRC=average, TLC=max, RV=TLC-VC, ERV=FRC-RV (TLC,
FRC and VC are measured)
Custom All settings related to volumes are available
A diagram teaches the operator about how to perform the maneuver (correct
sequence and way of breathing: quiet breathing followed by panting is only
required for TGV capture).
Multi-Breath
Settings Description
End Test Criteria (FetN2) Enter the FetN2 value for End of Test determination (at
least three consecutive breaths with this concentration
must be recorded)
Depth of breathing Check (and enter the range) to have visual feedback on
the depth of breathing (L) during test execution
Frequency Check (and enter the range) to have visual feedback on
the respiratory frequency (1/min) during test execution
FiN2 Enter the value of FiN2 (%) contained in the inspiratory
mixture (default 0%)
Single-Breath
Settings Description
Calculation Method Select the calculation method for Single Breath:
25%-75% (regression line between 25%-75%
of the expired volume)
V III (regression line related to the phase III
itself)
4.7.5 CPET
General
Settings Description
Start recording after 2 minutes Check to automatically begin the recording of the test
after 2 minutes (if Rec is not pressed)
Print Stored ECG Check to automatically print the ECG
Subject Type Select the default Subject Type in the pre-test window
interface:
Clinical
Healthy
Dyspnea/Leg Pain Select the scale to be used for Dyspnea/Leg Pain
evaluation:
Borg Scale (6-20)
Modified Borg Scale (0-10)
Calculate MVV Enter the calculation method of MVV based on FEV1
Show calibration warning Check to show a warning message to trigger air
message calibration before test start
Data Filtering
Select how to filter data from the following:
Settings Description
None No filtering is applied
Time Average Filter is based on time average. Enter time (default 30s)
Smoothing Filter is based on steps smoothing. Number of steps can
be selected from 3 to 10
Rolling Time Average Filter is based on rolling time average (one data point
every 10 seconds representing average data over the
previous 30 seconds)
Accepted Range
Check to configure automatic filtering in order to ignore non-physiological data.
Settings Description
Vt Tidal Volume > Min. Enter Min
VO2 Min < Oxygen Uptake < Max. Enter Min and Max
RF Min < Respiratory Frequency < Max. Enter Min and Max
RQ Min < Respiratory Quotient < Max. Enter Min and Max
Additional Analyses
Settings Description
Enable Training Zones Check to show the Training Zones tab in CPET Edit
panel
Enable FatMax Check to show the FatMax calculation in the Training
Zones tab in CPET Edit panel
Enable O2 Kinetics Check to show the O2 Kinetics tab in CPET Edit panel
4.7.6 REE
General
Settings Description
Show calibration warning Check to show a warning message to trigger air
message calibration before test start
Data Filtering
Select how to filter data from the following:
Settings Description
None No filtering is applied
Time Average Filter is based on time average. Enter time (default 30s)
Smoothing Filter is based on steps smoothing. Number of steps can
be selected from 3 to 10
Rolling Time Average Filter is based on rolling time average (one data point
every 10 seconds representing average data over the
previous 30 seconds)
Accepted Range
Check to configure automatic filtering to ignore non-physiological data.
Settings Description
VT Tidal Volume > Min. Enter Min
VO2 Min < Oxygen Uptake < Max. Enter Min and Max
RF Min < Respiratory Frequency < Max. Enter Min and Max
RQ Min < Respiratory Quotient < Max. Enter Min and Max
Canopy
Settings Description
Normal Range Enter the normal range for FeCO2 during test execution
(warning trigger)
Enable FiO2 reading Check (and enter) the reading of FiO2 during test
Temp, RH@Flowmeter Select method for calculating Temperature and RH at
the Flowmeter:
Automatic (sensor based)
Custom (enter the value)
Quark RMR blower adjustment Select the method for blower adjustment:
PC SW
Manual knob
P0.1
Settings Description
Vt Stability Threshold Select the Vt required stability before activation
Modality Check and select the modality for the test:
Environment
O2
CO2
Rocc
Occlusion
Settings Description
Insp/Exp Select the configuration for the occlusion
Flow Threshold / Peak Select the Flow Threshold (and enter threshold value)
or Peak for the occlusion
Type Select the occurrence of the occlusion:
> (whenever the flow is above the threshold)
Random (randomically)
Manual (manually activated)
Calculation Method
Settings Description
Average/Median Select method to calculate the results:
Average
Median
FOT
Settings Description
Calculation Method Select method to calculate the results:
Average
Median
Coherence Threshold Enter the coherence threshold for trail analysis
Signal Level Enter the signal level for FOT perturbation for Quark
i2m (default 40)
Broncho-Challenge
Settings Description
Default Protocol Select the default protocol for BC tests from the list of
available protocols (refer to section 9.4)
Default Reference Select the default reference for BC test:
Diluent (if present in the protocol)
Pre (if present in the same visit)
Dosimeter
WARNING: Read carefully before using the device. Users are assumed to be
competent with the associated functions and risks.
Settings Description
Nebulizer Driving Pressure Enter the nebulizing driving pressure (used to check the
effective pressure error when nebulizing)
Nebulizer Output Enter the correct Nebulizer output based on the
Nebulizer used; it is used for the calculations in
broncho-provocation protocols (dosimeter mode)
Reference values:
REF C05118-01-10 (RESPIRONICS
SIDESTREAM):
Nebulizer Output = 0.35 mL/min
REF A-862-010-001 (DEVILBISS):
Nebulizer Output = 0.9 mL/min
Review broncho-challenge protocols based on the
nebulizer used
Accepted Pressure Error (%) Enter the maximum acceptable error for driving
pressure
Start Delay (ms) Enter the delay in activation of nebulizer when
triggered
Triggering Flowmeter Select Flowmeter used for triggering the nebulizer
4.7.10 Advanced
Anti-Bacterial Filter
Settings Description
Dead Space (VDf) Enter the Anti-Bacterial Filter dead space
Flowmeter
Settings Description
X9 system Dead Space (VDs) Enter the X9 Flowmeter dead space
Turbine system Dead Space Enter the Turbine dead space
(VDs)
Valve Dead Space (VDv) Enter the Breathing Valve dead space
Flow-REE system dead Space Enter the Flow-REE dead space
4.8 ECG
This menu allows the configuration of the GE® Cardiosoft-CASE and Cardiolex® EC
Sense interface.
Settings Description
Data folder Select destination of GE® Cardiosoft-CASE/Cardiolex®
EC Sense data
PDF folder Select destination of GE® Cardiosoft-CASE/Cardiolex®
EC Sense PDF report
Executable Path Select the path to the executable file used in testing
mode 1 (refer to section 13.4.5) to automatically start
the program
Protocol Select Cardiosoft-CASE/Cardiolex® EC Sense
communication protocol with OMNIA from the
dropdown (see section 13.4.5)
4.9 Admin
Login
Settings Description
Login session timeout Enter the timeout for user’s session (OMNIA will
require password entry after timeout is expired)
Automatic login authentication using Check to use automatic login authentication using
Active Directory Active Directory
Password Expiration
Settings Description
Password does not expire Select to remove the password expiration
Password expires in ## Days Select (and enter) the duration of the password
verification
Password Complexity
Settings Description
Policy Select the password policy:
None
Custom (enter length and select
characteristics checkboxes)
4.10 Export
4.10.1 File
File Name Pattern
Settings Description
File Name Define a standardized filename format using the
available options
PDF
Settings Description
PDF path Select the location for PDF reports
Open PDF file after exporting Check for automatic PDF opening after exporting
Use Bullzip PDF printer Check to use Bullzip PDF to generate PDF (it can
be enabled only if Bullzip is installed on the PC)
Generate PDF/A Check to create the PDF in PDF/A standard
format
Geotag
Settings Description
Geotag Select the option for display of GPS data (K5)
XML
Settings Description
Encrypt Check to encrypt the XML file
Include test curves Check to include data related to the single curve
Deidentify subject data Check to deidentify subject data
Include parameters for each trial Check to include test parameters of each trial
(NIOSH) (according to NIOSH specs)
Parameters Select the parameters to be exported:
Results
Edit
Print
Any
4.10.2 GDT
General
Settings Description
Enable Check to enable the GDT communication
protocols
Parameters Select which parameters are exported according
to the customization (Results, Edit, Print, Any)
Import Settings
Settings Description
Data Path Select the location for the file containing the
requests to process
File Name Enter the file name related to the requests
Delete file after import Check to delete file after importing
Export Settings
Settings Description
Data Path Select the location for the file containing the
results
File Name Enter a file name related to the response
PDF Path Select the location for PDF export
4.11 Licenses
This menu contains the list of all installed and activated OMNIA modules on the
connected device(s). See section 1.4 for further details.
4.12 Appearance
Resolution
Settings Description
Scaling Slide to adjust the scaling factor for OMNIA
resolution
Startup Display
Settings Description
Open Omnia on Assign the display where OMNIA will open (when
multiple screens are available)
4.13 About
This menu contains a direct link to COSMED website and information about:
Software version.
Current user.
SQL Server.
Database Size.
5 - Database
5.1 Overview
The Database is accessible by clicking on the Database tile from the Home panel.
In the Database users can:
View subjects, visits and test information.
Create new subjects, visits and tests.
Delete subjects, visits and tests.
Edit subjects.
Access visits and test edit screens.
Print visits and tests report.
Access subjects trend analysis.
Query the database and print query.
Import subjects, visits and tests in xml proprietary format.
Download tests from portable COSMED devices.
Export subjects, visits and tests in proprietary xml format.
Note: Activities in the Database may be limited based on the roles assigned to the
current user.
5.2 Structure
The main features of each area are described below. Certain features may only be
available in certain database status or software configurations.
Buttons Description
Create new subject (see section 5.3)
Buttons Description
Open the edit subject panel.
Refer to section 5.3 for the complete list of fields
Create a previous visit (date < present date) for the selected
subject. Only an Offline Test can be created for a previous visit
(see section 0)
Delete the selected subject
Other action buttons related to the most common functions are available in the top
section of the subject card:
Buttons Description
Create new visit for the selected subject (see section 5.3.2)
Create new test for the selected subject (see section 6 - Testing)
Open the trend analysis for the selected subject (see section
5.3.5)
From the visits card users can access each test performed during the selected visit:
Icons Description
Click to show the list of available tests within a visit
When a test is selected, the corresponding line is highlighted and the following
action buttons appear at the end of the line:
Buttons Description
Open the selected test (see section 6.1.3).
The test can also be opened by double-clicking on the
corresponding line
Print the selected test.
Notes and Interpretation are not available for single test
printouts
Export the selected test
From the Database panel users can access the subpanels and windows described
below.
Buttons Description
Read data from the connected device and show all tests stored
on the device
Read data from the oximeter and show all tests stored on the
device, after device selection from the dropdown
Merge two or more test performed in the same day into a single
test
To the left side of the panel all tests stored on the connected device are listed,
including basic test information. The list can be populated by clicking on and can
be filtered by toggling the Select New Test Only button. The selected test can be
transferred into OMNIA database creating a new subject or b
eing linked to an existing one. Tests on device and subjects in OMNIA database can
be browsed using the two dedicated search bars.
5.3.5 Trend
The trend panel displays the progression of selected parameters across the selected
visits of a subject. The trend appearance can be selected from the Configuration
dropdown (see section 9.2.6) and the visits time interval can be set in the Start and
End fields.
6 - Testing
6.1 Overview
Test functions are accessible by clicking on the Testing tile from the Home panel of
OMNIA after having created a visit. Refer to the diagram below (with the complete
list of testing functions) to access the required test.
Spirometry
Body Plethysmography
Nitrogen Washout
Metabolic
Pulse Oximetry
ECG
Body Composition
Note: For appearance and algorithms applied to test results refer to section 4.7.
# Step Description
1 Required Items Collect the necessary items according to specific test as
indicated by your device user manual
2 Warm-up and Follow the recommended warm-up time as indicated
Calibrations by your device user manual
Perform periodic calibrations as needed
3 Data entry and Test Select existing subject from the database or create a
Selection new one.
Add a new visit for the current day, if not present (refer
to section 5 - Database ).
Select the Test from Testing menu
4 Test Set-up Set-up device and accessories according to your device
user manual
When ready, press Start to begin the test
5 Real-time execution Instruct the subject on how to perform the test.
Perform multiple trials, if necessary, then stop the test
or wait for automatic end
6 Edit trials (if necessary) Edit Trials: performed trials can be accepted/discarded
7 Visit Review Review test results and edit visit information (if
necessary)
8 Dispose and Cleaning Dispose of non-reusable items
Clean and disinfect reusable items according to the
instructions in the Cleaning and Disinfection user
manual
Button Function
Start the test/activate gas delivery/start recording as
per defined in your device user manual
The bottom bar indicates if BTPS has been disabled from the Settings (refer to
section 4.3) for some pulmonary function tests. For test-specific functions please
refer to the sections below.
Maximize graph
Tabular data may also be seen throughout the navigation and during test execution.
Refer to section 9.2 for parameter selection.
Results which are displayed in Visit Review can be customized in Utility (refer to
section 9.2 for further details).
The following functions are available:
Button Function
Perform a new test in the current visit
The compact PFT report is organized as follows: the header fields are in white
(complete or partial fields are shown depending on the report page, see section
9.2.1). Below the header, all PFT tests performed within the visit are shown in a
compact fashion: each test is preceded by a grey bar with some test-specific fields
(including test type and details) and numerical results are shown for the Visit
Review Overview panel parameters. Charts are shown for each test except for MVV,
DLCO IB, MIP/MEP and Rocc. The interpretation is also shown according to the
selection. Below the compact test section, all tests selected in the print options are
shown in detail.
Additional details on test-specific visit review features are available for each test
described in this chapter.
Visit Review Overview
The Visit Review Overview panel contains information for all tests performed within
the selected visit, including the visit status. The Visit Review Overview panel
comprehends:
Numerical results and their normal ranges, predicted values and z-scores.
Interpretation and Notes.
Interpretation and Notes are editable by clicking on the corresponding tabs to the
right. Five visit statuses are available:
To be processed.
In progress.
Preliminary.
Final.
Corrected.
Edit Trials
Print test
6.2 Spirometry
Pediatric Incentive
Pediatric incentive option is available, to help children perform acceptable FVC
and FVC post BD tests using animations and graphics. It is on by default if the
subject is ≤ 7 years. In any case, the incentive can be enabled/disabled for all
subjects by clicking on the Incentive button. OMNIA offers different animation
and configuration of both flow and volume targets, complying with ATS/ERS
2007 [6]. The following functions are available:
Button/slider Function
Enable/disable the pediatric incentive option and select the
desired animation by clicking on the corresponding button
%PEF Pred Select flow target as a % of predicted PEF for goal achievement
(from 30% to 150%)
%FVC Pred Select volume target as a % of predicted FVC for goal
achievement (from 30% to 150%)
WARNING: If FEV1 decreases more than 20% when compared to baseline FVC, as
the Trial Edit panel opens a message prompts indicating the target has
been reached. Please end the test as continuing can be dangerous and
is NOT recommended.
Non-dosimeter protocols
After protocol selection, the below interface appears if the selected protocol is
not a Dosimeter Protocol (as defined in section 9.4). Users can modify the
protocol-specific feature (e.g. the agent dose) for the current protocol step. The
Test button shall be pressed after the provocation step has been performed to
execute the FVC maneuver, or alternatively Cancel is available to abort.
Button/Field Function
Skip to the next step
Dosimeter protocols
If the selected protocol is a Dosimeter Protocol (as defined in section 9.4) the
dosimeter protocol interface below is shown. Users can modify the agent for
the current protocol step and navigate through the steps. The Stop button shall
be pressed after the provocation step has been performed to execute the FVC
maneuver, or alternatively Cancel is available to abort.
Button/Field Function
Start the provocation step (it may include one or more
breaths, as per defined in the protocol)
System quality control for FEV1 and FVC results is provided by OMNIA in
agreement with [3]. The software also displays the operator quality control and
subject position during the test, both of which are entered by the operator.
The following predicted values are shown on FVC graphs, if available:
MEF25%, MEF50%, MEF75%, PEF (horizontal line) and FVC on
Flow-Volume graph.
FEV1, FVC (horizontal line) and 6s line, as defined by ATS/ERS.
In addition to FVC Pre results, the dose of the administered BD agent is shown
in FVC Post BD panel. Graphs and numerical results show FVC Pre and Post
results.
FVC post BC
In FVC post BC panel, Flow-Volume and Volume-Time graphs and numerical
results at each protocol step show the best FVC inspiratory and expiratory
phases of performed trials within each step. A graph showing FEV1 fall at each
step is displayed and by clicking on users can either select a Fall % or a %
Baseline visualization. Numerical results include PD20, PD15 and PD10 and are
shown for each protocol step. Users can visualize either Dose or Cumulative
dose results by selecting the corresponding representation from the dropdown.
Note: If first dose (X) caused >20% decrease in FEV1, the dosage is listed as
<X. For all other steps, the dosage is calculated as an interpolation
between two steps, in accordance with ATS guidelines.
SVC
SVC panel allows switching between all SVC tests performed (SVC, SVC post BD) by
clicking on the corresponding tab to the right.
In SVC Pre and Post BD panels, graphical and numerical results show the best SVC
maneuver among the trials performed, independently of the performed maneuver
(IVC or EVC). VC, IRV, VT and ERV are graphically displayed on SVC graph.
In FVC post BD Edit panel, the user can modify the BD agent dose administered
during the test.
FVC Post BC
From FVC Post BC Visit Review panel, users can independently edit trials for each
step of the FVC Challenge, by clicking on the corresponding button. Users can
also modify the Operator QC and Position fields. The following functions are
available in addition to those in section 6.1.4.
Button Function
Perform a new trial for the current step
Accept only the three best curves according to ATS/ERS criteria for
the current step
Accept only the three best SVC trials according to ATS/ERS criteria
Single SVC trials can also be edited after clicking on the corresponding icon.
EELV (black dashed horizontal line) and EILV (red dashed horizontal line) for the
selected curve, which detect VT, can be adjusted graphically. Pressing OK changes
are confirmed (Cancel disregards them). VT and IC values beneath the graph are
automatically updated as soon as EELV and EILV values are adjusted.
In SVC post BD Edit panel, the administered BD agent dose can be modified.
MVV
In MVV Edit panel user can accept/discard trials, choose the best MVV among the
performed trials and show/hide trials. Users can modify the Operator QC and
Position fields. The following functions are available in addition to those in section
6.1.4.
Button Function
Perform a new MVV trial
Accept only the three best MVV trials according to ATS/ERS criteria
DLCO test consists in a diffusion gas maximal inspiration from RV to TLC, followed by
a breath-hold period and an exhalation (Single-Breath technique). Inspired and
exhaled CO and tracer gas are analysed to calculate the transfer factor.
As an alternative for subjects who lack the ability to hold their breath, the DLCO
Intra-Breath technique allows a slow exhalation to replace the breath-hold period.
OMNIA also allows to evaluate 𝐷𝑀 and 𝑉𝐶 for both techniques if at least two trials
are performed with different oxygen fractions in the inspired gas (Membrane
Diffusion test).
DLCO IB
DLCO IB Visit Review panel shows the results of the Diffusing Lung Capacity (Intra-
Breath) test. In the DLCO IB panel, the volume chart and numerical results show the
best DLCO from the trials performed.
The graph plots CO and CH4 % concentrations, the flow and volume from the
beginning of the maneuver and the recommended expiratory flow range (between
250 and 750 mL/s). The washout volume and sampling volume are displayed on the
graph where a horizontal line indicates the volume corresponding to 90% of best VC
performed during the visit.
Results are reported as a percentage of the predicted values and as z-scores.
Predicted values and normal ranges for each parameter are also reported, according
to the selected predicted set. Gauges which graphically display z-scores of the main
DLCO parameters are also shown for easy interpretation of the results.
The software displays the system quality control, the operator quality control and
subject position during the test, both of which are entered by the operator. The
following functions are available in addition to those described in section 6.1.3.
Button Function
Review gas calibration factors
Single DLCO trials can also be edited after clicking on the corresponding icon.
The sampling volume and the washout volume can be adjusted by moving the two
red dashed lines directly on the graphs or by moving the sliders below the graphs.
These volumes can also be adjusted according to standard methods which users can
select from the Washout Volume dropdown to the right. If the Mass Balance
calculation method is selected from OMNIA Settings (refer to section 4.7.2) an
additional slider allow to adjust the end of test. Users can also adjust the correction
factors by selecting the method for COHb and back pressure adjustment from the
dropdown and enable the TLC(DLCO) calculation for obstructed subjects by clicking
on the corresponding checkbox. Software also allows to enable membrane diffusion
which requires entering fraction of inspired oxygen %. To choose the breath-hold
method from the dropdown and to zoom in on the Volumes graph, toggle the
Magnifier button. For further details on DLCO options please refer to section 4.7.2.
Pressing OK, changes are confirmed (Cancel disregards them).
DLCO IB
In DLCO IB Edit panel user can accept/discard trials, choose the best DLCO
maneuver from the performed trials and show/hide trials. Users can also modify the
Operator QC and Position fields and enter subject’s hemoglobin (Hb) and
carboxyhemoglobin (COHb) concentration values at test start (see section 4.7.2 for
further details). The following functions are also available in addition to those
described in section 6.1.4.
Button Function
Perform a new DLCO trial
Single DLCO trials can be edited by clicking on the corresponding icon. The
sampling and washout volumes can be adjusted by moving the two red dashed lines
directly on the graph or by moving the sliders below it. These volumes can also be
adjusted according to ATS/ERS [4]. Users can adjust the correction factors by
selecting the method for COHb and back pressure adjustment from the dropdown
and enable the TLC(DLCO) calculation for obstructed subjects by selecting the
corresponding checkbox. Software allows to enable membrane diffusion by entering
the fraction of inspired oxygen %. To zoom in on the graph toggle the Magnifier
button. Pressing OK, changes are confirmed (Cancel disregards them).
Button Function
Choose washout volume and sampling volume according to
ATS/ERS criteria
Raw Challenge
Body Plethysmography test measures the Thoracic Gas Volume (TGV) and Airways
Resistance (Raw). TGV measurements by body plethysmography results highly
accurate particularly in subjects with lung diseases associated with gas trapping and
obstructed subjects, for whom dilution techniques may underestimate lung
volumes.
OMNIA provides three different testing modes: Thoracic Gas Volume/Raw, Thoracic
Gas Volume Post Dilator and Raw challenge (after BC). The TGV/Raw Simulated test
is available for verification purposes.
To the right, there are three vertical bars representing real-time respiratory
frequency, tidal volume and box pressure, which help to check for stability. During
test execution the user interacts with the device using the additional commands:
Button Function
Perform Q-Box test with door open, no values will be displayed
Open vent valve (to be used if pressure is not stable inside the box, i.e.
no thermal equilibrium is reached)
Toggle the panting filter on/off: when toggled on, subject should
breath at a respiratory frequency between 0.5 Hz and 1 Hz
Raw challenge
During test execution the Thoracic Gas Volume/Raw real-time interface is shown
(TGV Captures and SVC/IC Captures graphs are disabled).
After protocol selection, the protocol interface is as described in Forced Vital
Capacity Challenge. Refer to sections 6.2.2 and 6.2.3.
TGV/Raw Simulated
During test execution the Thoracic Gas Volume/Raw real-time interface is shown.
In addition to TGV/Raw pre results, the dose of the administered BD agent is shown
in the TGV/Raw Post BD panel. Pre and Post absolute numerical results and %
changes are both displayed.
TGV/Raw (post BC)
In the TGV/Raw post BC panel, the sRaw graph shows all maneuvers (different
protocol steps are indicated with different colors) and numerical results for each
protocol step show the average results (or median, refer to section 4.7.3) for all
accepted trials within each protocol step. Two graphs showing the % decrease in
sGaw and the % increase in sRaw against the protocol steps are also displayed.
Numerical results include PD45, PD40 and PD35 referring to sGaw % decrease and
PD100 referring to sRaw % increase. Users can visualize either Dose or Cumulative
dose results by selecting the item from the dropdown. The following functions are
available in addition to those described in section 6.1.3.
Button Function
Show/hide TGV/Raw Pre graphical and numerical results
Single captures can be edited by clicking on the corresponding icon and can be
cycled through using the left/right arrows. Captures can be accepted or discarded
by checking the Accepted checkbox in the lower left. Users can also modify the
Operator QC and Position fields. The following functions are also available in
addition to those described in section 6.1.4.
Button Function
Export capture raw data in xls format
N2 Multi-Breath Washout
Nitrogen Washout
N2 Single-Breath Washout
Inert gas washout tests, performed using the single- or multiple-breath washout
techniques, measure ventilation distribution inhomogeneity at differing lung
volumes and offer complementary information to standard lung function tests, such
as spirometry.
Multi-Breath Nitrogen Washout (N2 MBW) aims at measuring FRC and Single-Breath
Nitrogen Washout (N2 SBW) aims at measuring Closing Volume (CV), and as a
consequence Closing Capacity (CC) if RV is known. CV reflects airway closure
occurring preferentially in dependent lung regions and peripheral airway
obstruction.
To the right, FetN2, FeN2, CFRC, CEV and LCI real-time values are shown and two
vertical bars show real-time respiratory frequency and tidal volume, which help to
identify a normal breathing pattern. Tabular data is at the bottom.
N2 Single-Breath Washout
During test execution multiple graphs show real-time values. At test start a volume-
time graph is shown to aid in determining when to activate oxygen delivery. After
the initiation of oxygen delivery, the graph to the left will display expired N2
fraction vs. volume. To the right, flow is plotted against time and the target flow
range, as defined by ATS/ERS [7], is indicated by two dashed lines.
N2 SBW
N2 SBW Visit Review panel shows the results of N2 Single-Breath Washout test. In
N2 SBW panel, charts (CV graph and flow graph) and numerical results show the
best N2 single-breath washout among the trials performed.
The CV graph shows expired % nitrogen against volume. The flow graph shows the
expiration flow against time. On this graph the recommended flow range (between
400 and 500 mL/min) is indicated by two horizontal dashed lines.
Single N2 MBW trials can also be edited after clicking on the corresponding
icon.
Main washout parameters are reported for each breath and the end of the phase
for the FRC calculation is indicated by a vertical dashed line on the FRC graph. Such
limit can be adjusted by moving the dashed line directly on the graph or by moving
the slider below the graph. Such editing tool is particularly useful for subjects who
accidentally inhale air during test execution with consequent increases in % FetN2:
the following breaths can be discarded by adjusting the end-of-phase line and the
FRC can be calculated upon the breaths preceding the air inhalation if the estimate
is not too rough and without the need to repeat the test.
The end of the phase for FRC calculation is restored to the test end if Restore is
pressed.
N2 SBW
In N2 SBW Edit panel user can accept/discard trials, choose the best maneuver
among the performed trials and show/hide trials. Users can also modify the
Operator QC and Position fields and decide the CV calculation method from the
dropdown. For further details on calculation methods please refer to section 0. The
following functions are also available in addition to those described in section 6.1.4.
Button Function
Execute a new N2 SBW trial
Single N2 MBW trials can also be edited after clicking on the corresponding
icon.
Three vertical dashed lines indicate the limits between the phases which can either
be dragged on the graph or the corresponding sliders below the graph can be
moved for adjustment. The phase III slope is shown on the graph with a dashed line,
whose regression equation is reported at the top of the graph. The calculation
method for the selected trial can be selected from the dropdown to the right (refer
to section 0 for further information).
Button Function
Autodetect starting and ending points of phase III
Maximum Pressure
Forced Oscillation
MIP and MEP are indices of the respiratory muscle strength at the mouth, which is
independent of the lung condition. MIP is an index of diaphragm strength, while
MEP measures the strength of abdominal and intercostal muscles [8]. Rocc test
measures airways resistance using the interrupter technique. Respiratory Drive
(P0.1) is the mouth occlusion pressure developed 0.1 s after the onset of
inspiration. The Forced Oscillation Technique (FOT) measures respiratory system
impedance at different frequencies.
Button Function
Activate shutter closure for MIP/MEP capture
In addition to FOT pre results, the dose of the administered BD agent is shown in
FOT Post BD panel. Pre and Post results and absolute and % changes are shown.
Single Maximum Pressure trials can also be edited after clicking on the
corresponding icon. The start and the end (red dashed vertical lines) of the MIP
(MEP) plateau can be changed to calculate MIP (MEP) properly. Volume-time curve
is plotted on the same graph to help locate the plateau correctly. A dashed
horizontal line moves as the plateau duration is adjusted, and it indicates the
maximum pressure of the trial. Pressing OK confirms the changes (Cancel disregards
them). MIP (MEP) and plateau Duration values beneath the graph are automatically
updated as the plateau interval is adjusted.
Single airways resistance trials can also be edited after clicking on the corresponding
icon. A pressure-time graph shows the overlap of all captures (occlusions)
performed within the trial. Single captures can be shown/hidden and
accepted/discarded by checking/unchecking the related checkboxes. Pressing Save
confirms the changes (Cancel disregards them).
Respiratory Drive
In Respiratory Drive Edit panel users can accept/discard trials and accept/discard
trial steps. Users can also modify the Operator QC and Position fields. The following
functions are also available in addition to those described in section 6.1.4.
Button Function
Accept all the trials
Single Respiratory Drive trials can also be edited after clicking on the corresponding
icon. Single captures can be accepted/discarded by checking/unchecking the
corresponding checkbox. Captures can be discarded by deselecting them from the
graphs. Pressing Save confirms the changes (Cancel disregards them).
Forced Oscillation (pre/post BD)
In the Forced Oscillation Edit panel user can accept/discard trials and select one trial
to see additional details. Beneath the resistance and reactance graph, which shows
all the accepted trials and mean values, two graphs display additional information
for the selected trial: by clicking on the corresponding graph, users can choose
to display either flow and pressure vs. time or smoothed flow vs. time to the left
graph and either frequency-specific tabular data or coherence-frequency plot to the
right graph.
Users can also modify the coherence threshold, the Operator QC and Position fields.
In FOT post BD Edit panel, the user can modify the BD agent dose administered
during the test.
6.7 6MWT
Note: The test cannot be performed using OMNIA, OMNIA only allows data
review and edit.
To download the test from the Spiropalm 6MWT, please refer to the section 5.3.3.
Field Description
6MWD Enter distance covered during the 6MWT
Baseline dyspnea Select baseline dyspnea value from the dropdown
Final dyspnea Select final dyspnea value from the dropdown
Baseline fatigue Select baseline fatigue value from the dropdown
Final fatigue Select final fatigue value from the dropdown
Other symptoms Select symptoms from the dropdown
Suppl. O2 Select if supplemental oxygen was provided
Sup. O2 flow Enter the flow of supplemental oxygen
P Syst Enter systolic pressure
P Diast Enter diastolic pressure
Baseline FEV1 Enter baseline FEV1
Cardiopulmonary Exercise Testing (CPET) allows evaluation of the heart and lungs
under stress conditions that may be induced via several different protocols, by
means of the measurement of Oxygen consumption and Carbon Dioxide production
rates. The two common methods for analysing exhaled gases use either a mixing
chamber or breath-by-breath measurements, and both methods can be used for
performing a CPET test on Quark PFT. Although the mixing chamber mode does not
provide real-time values, at high respiratory frequencies (>60 breaths per minute) it
is more accurate than the Breath-by-Breath method.
Ergometer*
Field Description
Select ergometer from the dropdown (refer to section 13.6 for
Ergometer
further information)
Select protocol from the list of available protocols in the dropdown
Protocol
(as defined in section 9.3.1)
Port Enter ergometer communication port
*this field becomes inoperable if GE® Cardiosoft-CASE or Cardiolex® EC Sense are selected as
HR source (see below).
Spirometry
Enter the following spirometry information to estimate peak minute ventilation and
calculate the breathing reserve. These fields are pre-populated if spirometry has
been performed in the same visit (refer to section 4.7.5 for MVV multiplying factor
adjustment).
Field Description
FEV1 Enter subject’s FEV1
IC Enter subject’s IC
FVC Enter subject’s FVC
VC Enter subject’s VC
Integrated Devices
Field Description
Select the HR source and/or ECG source from the dropdown:
None: no HR monitor
Metabolic Cart: HR retrieved from a HR monitor, TTL ECG
or oximeter connected to the metabolic cart (refer to
sections 13.4.1 and 13.4.3 for further information)
PC Software ECG: ECG parameters retrieved from
integrated PC ECG (refer to section 13.4.2 for further
information)
Ergometer
HR Source
Mortara XScribe: ECG parameters retrieved from
Mortara® ECG (refer to section 13.4.4 for further
information)
GE Cardiosoft-CASE: ECG parameters retrieved from GE®
Cardiosoft-CASE ECG (refer to section 13.4.5 for further
information)
Cardiolex EC Sense: ECG parameters retrieved from
Cardiolex® EC Sense ECG (refer to section 13.4.6 for
further information)
Blood Pressure Select the blood pressure source from the dropdown:
None
Manual: manual blood pressure monitoring
Bike
Tango®: systolic and diastolic blood pressure retrieved
from Tango® (refer to section 13.4.6 for further
information)
Port Enter blood pressure communication port
Emogas Analyzer Select emogas analyzer source from the dropdown:
None
Radiometer (MonLink 2.0): ABG parameters retrieved
from Radiometer® TCM5 (refer to section 13.4.8 for
further information)
SenTec DMS: ABG parameters retrieved from SenTec®
DMS (refer to section 13.4.7 for further information)
Port Enter emogas analyzer communication port
During test execution the user can choose the real-time dashboard from the
available dashboards, as listed in section 9.2, by clicking on the corresponding tab
below the tabular data. Charts and tabular data can be customized as defined in
section 9.2.
User can also interact with the device via the following commands:
Button Function
Begin recording data (refer to section 4.7.5 for automatic start)
Rest
Warm-up
Exercise
Recovery
Insert a text marker by either typing the marker label or selecting an
existing marker label from the dropdown (refer to section 10 -
Resource Center). The marker is inserted as soon as OK is pressed
Enter blood pressure values (automatic or manual entry, as defined at
test start)
Select the RPE value for dyspnea according to the scale selected in
OMNIA Settings (refer to section 4.7.5)
Select the RPE value for leg pain according to the scale selected in
OMNIA Settings (refer to section 4.7.5)
Two widgets can be used for real-time protocol adjustment and can be switched
through the left/right arrows on the protocols widget.
Widget Function
The flow-volume graph and the volume-time graph are shown and dyspnea and leg
pain RPE values can be entered by clicking on the corresponding button. A powerful
interface allows the offline review and editing of all EFVL captures (refer to section
6.8.4).
Results
Graphical and numerical results are shown in the Results panel. The chart can be
customized according to section 6.1.2 and 9.2. Protocol information (including if an
ECG was used as HR source) are reported below the graph and numerical results are
shown to the right, divided by type. From the Summary dropdown users can choose
the results to display, as per defined in section 9.2. Gauges relative to normal ranges
are also shown for easy interpretation of the results. The following functions are
available in addition to those described in section 6.1.3.
Button Function
Edit test
Dashboards
From the Dashboard panel users can choose the results dashboard from the
available dashboards, as defined in section 9.2, by clicking on the corresponding tab
below the tabular data. Charts and tabular data can be customized as defined in
section 9.2 and each chart can be printed by clicking .
ECG
The ECG panel shows the ECG report in PDF format. The following function is
available in addition to those described in section 6.1.3.
Button Function
Edit ECG (enabled according to the ECG configuration). ECG editing is
described in section 6.8.4
Analysis
The Analysis panel allows users to analyse data using regression tools. To the left a
chart is shown. To the right the fitting parameters for the selected graph are divided
by regression type.
Button Function
Add new fitting set for the selected graph
A fitting window opens and the selected chart is shown with vertical lines indicating
the regression interval. The time interval can be adjusted by dragging the lines
directly on the graph or by moving the slider to the right. Users can also choose the
regression type from the dropdown and enter the number of iterations (Max
Attempts). The fitting can then be calculated using the following:
Button/slider Function
Calculate parameters for the selected regression type
Restore the default fitting parameters. Always press this button before
Calculate if the interval has changed
Pressing OK saves the fitting and lists it in the Analysis panel. Users can edit each
fitting via the following functions:
Button Function
Edit selected fitting
Note: Fittings are not saved in the Analysis panel after exiting the visit.
ABG
The ABG panel shows entered ABG values for all steps with ABG markers (both
received/entered during the test and entered during test editing).
EFVL
The EFVL panel shows the flow-volume loops analysis, if edited. To the left all EFVL
events are shown. To the right the summary of all accepted events and VE and
power plot against time are shown. Charts can be customized according to section
6.1.2.
O2 Kinetics
The O2 Kinetics panel shows the O2 kinetics analysis, if edited. To the left a chart
plots O2 and power against time and can be customized according to section 6.1.2.
Edited O2 kinetics fittings are listed to the right and are graphically superimposed
on the O2 curve with different colors.
Steady State
The Steady State panel shows the steady state analysis, if edited. To the left a chart
plots the steady state stages and can be customized according to section 6.1.2. To
the right tabular data summarize the parameters values at the different stages.
Training Zones
The Training Zones panel shows the training zones analysis, if edited. To the left a
chart highlights the training zones according to the color code set for each zone (see
section 9.5). To the right tabular data summarize the parameters values for the
different zones.
CPET Trial Edit panel allows switching between the tabs described above. As the
edits are performed, the corresponding tabs receive a green check mark.
Immediately after a test has ended, only the VO2max is automatically calculated
and shows up on the VO2 Max tab.
Overview
CPET Edit Overview panel allows users to review and edit graphs, numerical data
and test information.
Field/slider Description
Subject type Select subject type from the dropdown:
Clinical
Healthy
Test type Select test type from the dropdown:
Maximal
Submax
Other
FVC Post Ex (L) Enter FVC post exercise
FEV1 Post Ex (L) Enter FEV1 post exercise
Reason for test Select reason for test from the dropdown. Default values are listed.
Refer to section 10 - Resource Center for additional values
Reason for Select reason for stopping the test from dropdown. Default values
stopping test are listed. Refer to section 10 - Resource Center for additional values
Test Purpose Select test purpose from the dropdown. Default values are listed.
Refer to section 10 - Resource Center for additional values
Test Position Select test position from the dropdown:
Undefined
Standing
Sitting
Supine
ECG Response Select ECG response from dropdown:
None
Normal
Borderline
Abnormal
Anti Bacterial Filter Check/uncheck box if anti-bacterial filter was used during the test
Set the test as best test (best test data are used in the Visit Review
Overview tab and in printouts)
Charts can be customized according to section 6.1.2. The following functions are
available:
Button Function
Edit and filter tabular data (description below)
Single steps can be ignored by selecting the Ignore checkbox and ABG markers
can be added by selecting the ABG checkbox. All fields can be modified. The
following functions are also available:
Button Function
Delete selected steps
Edit ECG
The function is available only if a PDF is generated in the ECG software.
The ECG Edit function automatically opens the ECG software where all test edits
can be performed. According to the ECG configuration, OMNIA software may
show the steps required for editing a test, the software where to perform the
actions and the related buttons.
Note: After test editing a new PDF should be generated in PC ECG software.
Button Function
Autodetect the time intervals for VT1/LT1 and VT2/LT2
Button Function
Autodetect the exercise phase interval
By pressing OK the fitting is saved and listed in the O2 Kinetics panel. Users can edit
each fitting via the following functions:
Button Function
Edit selected fitting
EFVL
The EFVL panel shows all EFVL events to the right. The FVC pre loop is reported in
grey on all EFVL events, if performed within the same visit. The basal loop is shown
is red and accepted loops for each event are colored in blue.
To the left the summary of all accepted events is shown and VE and power are
plotted vs. time, with markers indicating when EFVL were performed. The summary
graph shows EELV, EILV, leg pain and dyspnea values for all events and a grey
dashed line indicates Basal IC - 400 mL. The distance between the EILV and EELV,
graphically represented by the grey area, is the VT. Charts can be customized
according to section 6.1.2. Users can also select the Dynamic Hyperinflation
Confirmed checkbox when the possibility of dynamic hyperinflation is detected (i.e.
EELV graph is above the Basal IC - 400 mL line).
Single EFVL events can be customized by clicking on the corresponding icon.
By selecting the Accepted checkbox the event is accepted and included in the
summary. Users can confirm flow limitation by selecting the corresponding
checkbox. Dyspnea and leg pain RPE values can be entered from the dropdowns.
The breath to associate with the EFVL event is highlighted in orange on both the
flow-volume and volume-time graphs and can be selected navigating through the
left/right arrows.
Steady State
The Steady State panel allows users to analyse the trend of parameters during a
test, assuming that a steady state protocol has been used. To the left a
customizable chart (see section 6.1.2) shows Power, Speed, Grade and VO2/Kg vs.
time. To the right all the steady state parameters are listed for each stage. Steady
state stages can be added and edited by means of:
Button Function
Add a stage before the selected one
Users can define the time range of selected stages by moving the corresponding
time slider or dragging the dashed vertical lines on the graph.
Zones
The Zones panel is visible if the corresponding setting has been enabled (see section
4.7.5). The Zones panel allows to edit the training zones and the Maximum Fat
Oxidation, if the corresponding setting has been enabled (see section 4.7.5). It
shows two editable graphs (Fat vs. time and VO2/Kg vs. HR by default). On the
Maximum Fat Oxidation chart users can select the FatMax (FM) point by adjusting
the corresponding slider or dragging the vertical line. The selected FM is then
highlighted on the Training Zones chart by means of a marker.
To the right FM tabular data are shown according to the selection and training
zones tabular data are shown after selecting the Zones protocol from the dropdown
(see section 9.5 for training zones protocols edit).
REE Canopy
REE Ventilator
Resting Energy
REE Ventilator (Q-NRG)
Expenditure
REE Mask
Oxygen consumption (VO2) and Carbon Dioxide Production (VCO2) are measured to
calculate the Resting Energy Expenditure (REE), or Resting Metabolic Rate (RMR),
and the Respiratory Quotient (RQ), representing the substrates of metabolism.
REE measurements act as a precise tool for determining a subject’s basal nutritional
requirements and provide information which allow for adequate nutritional intake.
Button Function
Start recording data
Average data over the specified interval (this can be edited after the
test)
Results
Graphical and numerical results are shown in the Results panel. The chart can be
customized according to section 6.1.2. Protocol information are reported below the
graph and numerical results are shown to the right, divided by type. Gauges relative
to normal ranges are also shown for easy interpretation of results. The following
functions are available in addition to those described in section 6.1.3.
Button Function
Edit test
Dashboards
From the Dashboard panel users can choose the results dashboard from the
available dashboards, as listed in section 9.2.4, by clicking on the corresponding tab
below the tabular data. Charts and tabular data can be customized as defined in
section 6.1.2 and each chart can be printed by clicking on .
Analysis
The Analysis panel allows users to analyse data using regression tools. To the left a
chart is shown. To the right the fitting parameters for the selected graph are divided
by regression type.
Button Function
Add new fitting set for the selected graph
A fitting window opens and the selected chart is shown with vertical lines bordering
the regression interval (highlighted). The time interval can be adjusted by dragging
the lines directly on the graph or by moving the slider to the right. Users can also
choose the regression type from the dropdown. The fitting can then be calculated
using the following:
Button/slider Function
Calculate parameters for the selected regression type
Pressing OK saves the fitting and lists it in the Analysis panel. Users can edit each
fitting via the following functions:
Button Function
Edit selected fitting
Note: Fittings are not saved in the Analysis panel after exiting the visit.
N/A
Low
Medium
High
Sedation Select sedation level from dropdown:
N/A
Low
Medium
High
Severity (APACHE (REE Vent only)
III) Select severity from dropdown according to APACHE III
classification
Test Purpose Select test purpose from the dropdown. Default values are listed.
Refer to section 10 - Resource Center for additional field values
Set the test as best test (best test data are used in the Visit Review
Overview tab and in printouts)
Charts can be customized according to section 6.1.2. The following functions are
available:
Button Function
Edit and filter tabular data (description below)
Single steps can be ignored by selecting the Ignore checkbox and all fields can
be modified. The following functions are also available:
Button Function
Delete selected steps
Button Function
Add additional averaging interval after the selected interval
Note: When using Bluetooth®, data loss may occur due to radio interferences or
exceeded transmission range; the entire test can be downloaded to PC after
test is completed.
Protocol Options
Field Description
Name Enter protocol name
Baseline Enter baseline phase duration
Exercise Enter exercise phase duration
Recovery Enter recovery phase duration
Walk Path Enter the walk path distance
Select the lap increment from the dropdown (1 or 2 walk
Lap Increment
paths)
Pause Check to set options for exercise timer and pause timer
Select exercise timer options from the following:
Exercise Timer Reset: exercise timer restarts from 0 after each pause
Cumulative: exercise timer cumulates all exercise phases
Real-time execution
During test execution a graph to the left displays real-time HR and SpO2% values,
and to the right real-time values are shown for HR, SpO2 %, walking distance, PI %,
exercise duration, recovery duration and test duration. The plethysmographic
waveform is also shown on the bottom right corner. Tabular data is at the bottom.
During test execution the icons below should be checked to ensure reliable
measurements.
Icons Indication
Poor detecting signal. Data are not transmitted
Heart rate
During test execution, user interacts with the device via the following commands:
Button Function
Enter Recovery phase and automatically open subject and test info
box
Select the RPE value for dyspnoea according to the scale selected in
OMNIA settings
Select the RPE value for leg pain according to the scale selected in
OMNIA settings
Protocol phases are highlighted on the graph according to the following colour
code.
Phase Color
Rest Dark yellow
Exercise Green
Pause Dashed
Recovery Yellow
Print test
Export test
POx Edit panel allows to edit tabular data and to enter additional information for
the selected test.
Field Description
Laps Completed # Enter the number of complete laps
Final Partial Lap Enter the distance of the final partial lap
Total Pauses # Enter the number of pauses during test
Total Pauses Enter the total pause duration
Duration (mm:ss)
mMRC Dyspnea Select the mMRC value from the dropdown
Scale
FEV1 Enter FEV1
Walking Speed Enter the walking speed
PaO2 (mmHg) Enter PaO2 at the end of the test
Reason for Select reason for stopping the test from dropdown. Default values
Stopping Test are listed. Refer to section 10 - Resource Center for additional values
Medication before Enter medication before test
test
Supplemental Select supplemental oxygen type from the dropdown:
Oxygen Type Pulse dose
Continuous flow
Delivery Device Select supplemental oxygen delivery device from the dropdown:
Nasal cannula
Oxygen pendant
Mode of Transport Select supplemental oxygen transport mode from the dropdown:
Carried shoulder unit
Pushed/pulled cylinder
Walking Aid Select walking aid from the dropdown
Walking Enter walking limitations
Limitations
Comments Enter comments
The graph and tabular data are shown on the left and a vertical black line highlights
the selected step on the graph. To the right, test data are shown, and additional
information can be entered or selected from the dropdown.
The following functions are also available.
Button/slider Function
Edit the exercise phase interval. See below for further
information
Set the test as best test (best test data are used in the Visit
Review Overview tab and in printouts)
section 6.10.2. The protocol phases are shown in different colours on the graph and
users can adjust the exercise phase by moving the corresponding sliders or dragging
the start and end vertical lines on the graph.
Advanced edit
By clicking on a window appears which allows to delete steps or edit
parameters. A parameter can be selected from the dropdown and the following
functions are available:
Value: replace all values with entered value
Correction%: apply % correction factor to all values
Offset: add offset to values
The deletion or the parameter edit are applied to the adjusted time range, if
selected, and to the adjusted parameter range (defined by parameter, condition
and value), if selected.
6.11 ECG
ECG Stress
ECG
ECG Resting
The overall goal of performing an ECG is to obtain information about the electrical
function of the heart. ECG Resting records and measures short ECG tests on subjects
in resting position. ECG Stress records and measures ECG tests on subjects under
stress conditions using a pre-defined test protocol.
During test execution user interacts with the device using the following commands:
Button Function
Save test in OMNIA (requires PDF generation in PC ECG Stress
software)
Note: After test editing a new PDF must be generated in PC ECG software.
In ADP Visit Review panel, results show the best test performed during the visit and
includes a distribution curve relative to the population selected for %Fat and
graphical trends for %FFM, Fat Mass, Fat Free Mass, and Body Mass. Numerical
results are shown to the right.
The following functions are available in addition to those described in section 6.1.3.
Button Function
Export test in Excel format
7 - Calibration
7.1 Overview
Air Calibration
Calibrate Scale
N2 Washout - Air
X9 Linearization Table
Calibration
Gas Calibration N2 Washout
Calibration
Polytropic Factor
Metabolic
Calibration Calibration-ERGO
Mixing Chamber Flowmeter
Calibration
Metabolic
Calibration-ICU
I2m calibration
Q-i2m
Characterization
Q-i2m Calibration
Q-i2m Calibration
Calibration History
Q-i2m Filter
Characterization
Note: At calibration start, user must confirm whether the flowmeter is connected
to the breathing valve. Calibrations with the breathing valve only compute
the calibration parameters to apply to tests using the breathing valve.
Scale calibration is required for the BOD POD device to ensure accuracy of
measurements. To perform scale calibration, refer to your device user manual. On-
screen instructions, which ask to place and remove the reference calibration
weights from the scale repeatedly, are provided during calibration.
The following functions are available during calibration execution in addition to
those described in section 7.1.1.
Button Function
End calibration and save results
At the end of calibration, calibration details show measured numerical results along
with the acceptable ranges and a Pass/Fail result is shown to the right.
Note: At calibration start, user must confirm whether the X9 is connected to the
breathing valve. Calibrations with the breathing valve only compute the
calibration parameters to apply to tests using the breathing valve (two
separate linearization tables are created for the X9 with/without the
breathing valve).
Gas calibration procedures consist of multiple points calibrations: room air and
known reference calibration gas mixtures from certified gas cylinders. Air calibration
is the only single-point calibration. To calibrate Gas Analyzers refer to your device
user manual. Gas cylinder concentration values and O2 sensor span can either be
edited at calibration start or in Settings/Calibration (refer to section 4.4.2).
During calibration procedures actual concentrations are plotted together with the
expected values, as shown in the example below.
At calibration end, details of the calibration (including measured parameters and
normal ranges) can be visualized by clicking on the Show Details checkbox.
I2m calibration is required by the Quark i2m device to ensure accurate impedance
measurements. A reference impedance is used through which the system sends a
PRN (Pseudo-Random Noise) signal. Flow and pressure are shown in the real-time
interface during i2m calibration. Impedance readings are performed and correction
factors are then calculated to match the reference impedance value.
The loudspeaker signal level and impedance value can be edited in
Settings/Calibration (refer to section 4.4). To perform i2m calibration refer to your
device user manual.
OMNIA maintains a record of all calibrations performed using the software. This can
be from a single or multiple devices. Calibrations can also be manually added to the
record by clicking on : a window opens and asks to select device, serial number,
calibration type, flowmeter and measured calibration parameters.
Manually inserted calibrations are particularly useful for certified calibration data
entry, whenever it is not possible to perform the calibration directly from OMNIA or
to automatically transfer the calibration information to OMNIA.
All calibrations are listed along with the device information and the overall
calibration result (Passed, Failed and Accepted, Manual). If no calibration result is
shown default settings are applied to the calibration.
The following functions are displayed in the Calibration History panel:
Button/Dropdown Function
Filter On Filter calibrations by date
Select Device Filter calibrations by device
Serial Number Filter calibrations by serial number
Calibrations Show/hide record of calibrations
QC Show/hide record of quality controls
Export calibration
8 - Quality Control
8.1 Overview
Quality Control is used to validate that the device is within calibration limits and as
an advanced tool to assess a good performance of the device.
Quality Control functions are accessible by clicking on the Quality Control tile from
the Home panel of OMNIA. Refer to the diagram below to access the required
function.
All quality control procedures performed with OMNIA, except for those executed in
the Control Panel, are listed in the Calibration History.
Practice Subject
Control Panel Measurements
Export QC
QC History
Accept calibration
Leakage Check is an operation which is performed to check seal of the Q-Box and
controlled leaks. It calculates peak pressure and leakage time (i.e., depressurization
time delay from peak pressure) after Q-Box pressurization.
To perform a Leakage Check, refer to section 4.5.2 and to your device user manual.
The check is automatic after Start is pressed.
Q-i2m test load is used to check Q-i2m impedance readings. A reference impedance
(FOT test load) is used through which the system sends a PRN (Pseudo-Random
Noise) signal. Flow and pressure are shown in the real-time interface during Q-i2m
test load.
To perform Q-i2m test load refer to your device user manual.
On screen-instructions are provided for all BOD Pod quality controls. The following
functions are displayed during BOD POD quality control execution:
Button Function
Proceed to the following step
Go to previous step
At scale check end the measured weight is shown along with a Pass/Fail result to
the right.
8.5.2 Autorun
The Autorun is used to assess environmental and BOD POD stability and consists of
six sequential automatic measurements. To perform the Autorun refer to your
device user manual. During the Autorun individual measurements are plotted and
linearly fitted. At autorun end the measurements average and the fitting standard
deviation and slope are shown to the right along with a Pass/Fail result.
8.5.4 Autorun-6X
The Autorun-6X activity takes one-hour, but can be stopped early, and consists of
six sequential Autorun sequences with a five-minutes break between. This is used
to warm-up the BOD POD prior the routine QC activities. To perform the autorun-6X
refer to your device user manual. Intermediate and final results are displayed. The
following functions are displayed:
Button Function
End Autorun-6X and save results
During the Autorun-6X individual autorun measurements are plotted and linearly
fitted. At Autorun-6X end all the autoruns performed are plotted and listed along
with averages, standard deviations and slopes. The average of the autorun averages
and the standard deviation and slope of the autorun averages fitting are shown to
the right along with a Pass/Fail result.
8.5.6 Export QC
Export QC allows to export BOD POD QC results to a comma delimited file (required
for troubleshooting). To export QC click Save, select folder and click Finish.
Control Panel is an advanced diagnostic tool for users with an understanding of the
device measurement technology. It allows to monitor several different parameters.
Name, real-time value, unit of measurement and acceptability range (Min, Max) are
shown for each parameter. By checking the Enabled checkbox, the corresponding
parameter can be enabled/disabled for graphical visualization, which can be
configured by selecting the Y-axis from the Show dropdown (time is represented on
the X-axis). The real-time graph allows to plot up to two parameters from the list.
The enabled parameters can be recorded and saved in a text file by clicking
below the parameters list and choosing location and file name. Users can also adjust
the parameters update interval by moving the slider below the graph.
Two different control panels are displayed depending on the device used.
The first control panel below shows the list of parameters to the left and the real-
time graph to the right.
Button Function
Begin data display
Edit chart: choose time, range of acceptability (Min, Max), step, auto
scale and to show the Y-axis for Y1 and Y2
Clear graph
The following control panel shows the list of parameters to the left, the real-time
graph and a window with additional commands to the right.
Button/Command Function
Begin data display
Clear graph
8.7 QC History
OMNIA maintains a record of all quality controls performed using the software. This
can be from a single device or multiple devices. In the Network version of OMNIA,
this record will contain all quality controls performed on the network.
All quality controls are listed along with the device information and the overall
quality control results.
The following functions are displayed in the QC History panel:
Button/Dropdown Function
Filter On Filter quality controls by date
Select Device Filter quality controls by device
Serial Number Filter quality controls by serial number
Calibrations Show/hide record of calibrations
QC Show/hide record of quality controls
For each quality control the following functions are also available:
Button Function
Open quality control
9 - Utility
9.1 Overview
Utility functions are accessible by clicking on the Utility tile from the Home panel of
OMNIA. Refer to the diagram below to access the required function.
Customize
Metabolic Protocols
Broncho-Challenge Protocols
Restore Database
Event Log
Audit Log
9.2 Customize
The customize panel allows to customize test reports and dashboards. It contains six
tabs: Header, PFT, CPET, REE, ADP, Trends.
Parameters Selection
All PFT tests are listed in the Test type column and the complete set of test-specific
parameters is shown for the selected test. By checking the Print, Results, Edit and
Overview checkboxes the corresponding parameter is shown/hidden in test
printouts, Visit Review test panel, Test Edit panel and Visit Review Overview panel,
respectively.
Measured values are always shown and users can choose to show/hide normal
ranges, predicted values, % of predicted values and z-scores by checking the
corresponding checkboxes. Users can also drag and drop the rows and the columns
to modify the arrangement.
Print Options
For each test users can customize the following print options by selecting the
corresponding checkboxes:
Option Description
Select charts appearance on report printout:
None: hide charts from printout
Charts Default: select default charts appearance
Occupational Health: select occupational health charts
appearance
Show/hide best FVC and additional 2 trials on report printout (FVC
Best FVC + 2 trials
only)
Show/hide volume chart on report printout (for DLCO SB, DLCO IB,
Volume chart
N2 MBW)
Gas vs Volume Show/hide gas-volume chart on report printout (DLCO SB only)
Chart
6MWT step detail Show/hide steps detail on report printout (6MWT only)
Show/hide gauges (graphical bars with predicted range and
Gauges
measured values), if available, on report printout
QC Show/hide quality control on report printout
Trial Results Show/hide results of performed trials on report printout
Tabular Data Show/hide tabular data on report printout
Z-score Format Set z-score format on report printout as:
Both: z-score values and bars on report printout
Value only: z-score values on report printout
Bar only: z-score bars on report printout
Details
Option Description
Device Show/hide device type and S/N on report printout
Test Show/hide test information on report printout
Calibration Show/hide calibration data on report printout
Environment Show/hide environmental data on report printout
Note: A complete set of functions is shown. Some PFT tests may have a subset of
such customising options.
The resulting report printout is organized as follows: the header fields are in white
(complete or partial fields are shown depending on the report page, see section
9.2.1), below details are shown in grey. Below the details, charts and gauges are
shown according to test type and further below numerical data, QC and trial results
are shown.
Parameters Selection
The complete set of CPET parameters, grouped by type, is shown.
By checking the Results and Overview checkboxes the corresponding parameter
is shown or hidden in the Summary section of the Results tab of the Visit
Review CPET panel and in the Visit Review Overview panel, respectively.
Measured values, first threshold (VT1/LT1) and maximum values are always
shown and users can choose to show/hide values during rest and warm-up
phases, FatMax, second threshold (VT2/LT2), predicted values, % of predicted
All dashboards (standard and customized) are listed to the left and the selected one
is shown in the panel. Dashboards can be ordered by drag and drop. Dashboards
may contain charts (number varies according to the dashboard template), tabular
data at the bottom and main parameters to the right (only available in real-time).
The following functions are available:
Button Function
Create a new dashboard
Dashboard Options
Dashboard options in the lower-left corner allow to select the dashboard type
for the selected dashboard and to configure it as Real-time, Results (shown in
the Dashboard tab of the Visit Review CPET panel) or Any (configured both as
Real-time and Results).
Charts
Each chart can be customized through the following functions:
Button Function
Select parameters to view on the chart. Standard graphs can be
selected or alternatively Edit Chart is available to customize graph
Maximize graph
Tabular data
Tabular data are listed below the charts and contain measured values of
selected parameters for the entire duration of the test.
To customize tabular data:
Reports
CPET report printouts are customisable by clicking on Customize CPET Reports tab.
Reports (standard and customized) are listed to the left and the selected one is
shown in the panel. Users can select:
the results to show on report choosing among the list of Summaries
defined in the Results tab.
up to two dashboards to show on report choosing among the one defined
in the Dashboard tab.
the tabular data to show on report choosing among the one defined in the
Dashboard tab.
The following functions are available:
Button Function
Create a new report template
Print Options
Users can customize the following print options by selecting the corresponding
checkboxes:
Option Description
Show/hide gauges (graphical bars with predicted range and
Gauges
measured values), if available, on report printout
Page Set vertical or horizontal page orientation
Orientation
ABG Show arterial blood gases (ABG) values on report printout
F/V loops Show/hide flow-volume loops on report printout
O2 Kinetics Show/hide O2 kinetics on report printout
Steady State Show/hide Steady State analysis on report printout
Training Zones
Option Description
Zones Show/hide training zones on report printout
Description Show/hide training zones description on report printout
Details
Option Description
Device Show/hide device type and S/N on report printout
Test Show/hide test information on report printout
Calibration Show/hide calibration data on report printout
Environment Show/hide environmental data on report printout
Protocol Show/hide information about the protocol used
The resulting report printout is organized as follows: the header fields are in white
(complete or partial fields are shown depending on the report page, see section
9.2.1), below details are shown in grey. Below the details, the dashboard and the
results summaries are shown according to the selected report and further below
gauges, F/V loops data and charts and the O2 kinetics are shown. Tabular data are
shown at the end of the report according to the selected report.
Parameters Selection
The complete set of REE parameters, grouped by type, is shown in the Results
tab. By checking the Results and Overview checkboxes the corresponding
parameter is shown/hidden in the Visit Review REE panel and in the Visit
Review Overview panel, respectively. Measured values and % of CV values are
always reported by default and users can choose to show/hide predicted
values, % of predicted values and class by checking the corresponding
checkboxes. Users can also drag and drop the rows and the columns to modify
the arrangement.
Dashboards
Real-time and Results dashboards are customisable by clicking on REE Dashboards
tab.
All dashboards (standard and customized) are listed to the left and the selected one
is shown in the panel. Dashboards may contain charts (number varies according to
the dashboard template), tabular data at the bottom and main parameters to the
right (only available in real-time).
The following functions are available:
Button Function
Create a new dashboard
Dashboard Options
Dashboard options in the lower-left corner allow to select the dashboard type
for the selected dashboard and to configure it as Real-time, Results (shown in
the Dashboard tab of the Visit Review REE panel) or Any (configured both as
Real-time and Results).
Charts
Each chart can be customized through the following functions:
Button Function
Select parameters to view on the chart. Standard graphs can be
selected or Edit Chart is available to customize graph.
Maximize graph.
Field Description
Parameter Choose the parameter to plot
Choose if the graph scale must be automatically set
Autoscale
Widgets
Widgets allow an easy viewing of some real-time parameters. Each widget can
be customized by clicking on the icon . Six widgets are available for REE
testing and users can select a subset of them, up to a maximum of four, to be
displayed on the dashboard.
Widget Function
Tabular data
Tabular data are listed below the charts and contain measured values of
selected parameters for the entire duration of the test.
To customize tabular data:
Click on next to tabular data box on the dashboard panel.
Parameters are divided by type. Check the checkboxes to show the
corresponding parameters in the tabular data.
Click OK to save changes.
Main parameters
Mian parameters are shown to the right. The following function is available:
Button Function
Add new main parameter from the list of all available parameters.
Parameter and label color can be customised by clicking on the
button. The button can be scrolled to be resized
Reports
REE report printouts are customisable by clicking on Customize REE Reports tab.
Reports (standard and customized) are listed to the left and the selected one is
shown in the panel. Users can select:
the results to show on report choosing among the list of Summaries
defined in the Results tab
up to two dashboards to show on report choosing among the one defined
in the Dashboard tab
the tabular data to show on report choosing among the one defined in the
Dashboard tab
The following functions are available:
Button Function
Create a new report template
Print Options
Users can customize the following print options using the corresponding
checkboxes:
Option Description
Show/hide gauges (graphical bars with predicted range and
Gauges
measured values), if available, on report printout
Page Set vertical or horizontal page orientation
Orientation
Details
Device Show/hide device type and S/N on report printout
Test Show/hide test information on report printout
Calibration Show/hide calibration data on report printout
Environment Show/hide environmental data on report printout
Protocol Show/hide information about the protocol used
The resulting report printout is organized as follows: the header fields are in white
(complete or partial fields are shown depending on the report page, see section
9.2.1), below details are shown in grey. Below the details, the dashboard and the
results summaries are shown according to the selected report and further below
gauges and tabular data are shown.
Parameters Selection
The complete set of parameters for body composition is shown. By checking the
Print, Results, Edit and Overview checkboxes the corresponding parameter is
shown/hidden in test printouts, Visit Review test panel, Test Edit panel and Visit
Review Overview panel, respectively.
Measured values are always shown. Users can also drag and drop the rows and the
columns to modify the arrangement.
Print Options
Users can customize the following print options using the corresponding
checkboxes.
Details
Option Description
Device Show/hide device type and S/N on report printout
Test Show/hide test information on report printout
Calibration Show/hide calibration data on report printout
Environment Show/hide environmental data on report printout
Button Function
Open a preview of selected test report
The resulting report printout is organized as follows: the header fields are in white
(complete or partial fields are shown depending on the report page, see section
9.2.1), below details are shown in grey. Below the details, charts, numerical data
and text explanations are shown.
Charts
Each chart can be customized through the following functions:
Button Function
Edit Chart
Maximize graph
Tabular data
Tabular data are listed below the charts and contain measured values of selected
parameters. To customize tabular data:
Click on next to tabular data box on the trend panel.
Parameters are divided by type. Check the checkboxes to show/hide the
corresponding parameters from the tabular data.
Click OK to save changes.
Options
Users can customize the following options:
Option Description
Print Options
Users can customize the following print options by selecting the corresponding
checkboxes:
Option Description
Charts Show/hide charts on report printout
Tabular Data Show/hide tabular data on report printout
Metabolic Protocols panel allows to view, customize and create exercise and resting
protocols.
Manual protocols
Manual protocols allow to manually add new steps by clicking on and to enter
data for each protocol step.
Incremental protocols
Incremental protocols allow to customise the following fields:
Button/Field Description
Rest Set data for Rest phase (duration and load)
Warm up Set data for Warm up phase (duration and load)
Exercise Set data for Exercise phase (duration, load and step duration)
Recovery Set data for Recovery phase (duration and load)
Ramp protocols
Ramp protocols allow to customise the following fields:
Button/Field Description
Rest Set data for Rest phase (duration and load)
Warm up Set data for Warm up phase (duration and load)
Exercise Set data for Exercise phase (duration and load)
Recovery Set data for Recovery phase (duration and load)
After protocol configuration, a graph shows the trend of the load. Single steps are
listed below the graph and can be modified by double-clicking on the step. Steps
appear in different colors depending on the protocol phase, as shown below.
Phase Color
Rest Dark yellow
Warm up Light blue
Exercise Green
Recovery Yellow
For each step the user can set the software to automatically start some events (BP,
ABG, ECG, EFVL) by selecting the corresponding checkbox or to delete the selected
step.
After protocol configuration, the resting intervals are highlighted on a graph. Single
steps are listed below the graph and can be modified by double-clicking on the step.
Training Zones Protocols panel allows to view, customize and create training zones
protocols to be used for CPET tests. All available protocols are listed to the left. The
panel shows the steps and features of the selected protocol.
For each training zone the following customization options are available:
Color selection from the dropdown.
Zone name and description entry.
Selection of reference Start Parameter and End Parameter, defining the
zone limits, among the following: VO2 Max, VO2@FatMax, VO2@VT1/LT1,
VO2@VT2/LT2, VO2 Reserve, HR Max.
Zone start and end expressed as a % of the reference parameters (see
above) by means of Start% and End% entry.
9.6.1 Overview
The device manager allows to configure all connected devices.
Device Connection
To the left side of the Device Manager panel a list of all COSMED devices that can be
interfaced with OMNIA is shown. The following functions are available:
Button Function
Scan all connected and authorized devices
By default, only the enabled devices are shown. A device is enabled if the
corresponding Enable checkbox is selected. By activating the Show all devices slider,
all COSMED devices are shown.
The communication port for each device can be selected from the Port dropdown. If
communication is successful, by clicking on the Show button software shows the
serial number of the selected device.
Bluetooth® communication is also available for some COSMED devices and can be
enabled from Device Manager panel (see following section for further details).
Test Default Configuration
To the right side of the Device Manager panel the default devices and default
flowmeters can be selected from the dropdowns for each test type. Selecting the
default devices and default flowmeters in the Device Manager allows to skip the
device and flowmeter selection at test start.
By creating a new user or editing users’ information a window appears which allows
to enter Domain (if the Active Directory is available) and to enter the fields
described below, according to the selected domain.
Note: Active Directory users can access OMNIA with the same username and
password to log in the Active Directory.
Note: OMNIA Network allows to load groups of users from the Active Directory.
Note: By default administrators are assigned with all rights except servicing rights.
Roles’ names can be changed by double clicking on the current name and entering
the new one.
Note: Technical Support roles are not allowed to access or modify the database,
except for the DEMO subject, and are only assigned with servicing rights
(e.g. perform calibrations, access the event log, etc.).
Note: Some rights are encapsulated and require other rights to be enabled.
Note: COSMED assumes no liability for damage or loss of data in the event of
hard drive, PC, external storage damage should occur. To mitigate this risk,
multiple backups should be created and both backup files should be stored
in a safe place (e.g. external hard disk).
Note: For OMNIA network version the database backup is not managed by
OMNIA but the IT department shall perform it on server.
Restore Database function is available for OMNIA standalone version only and
allows to restore the database from a previously backed-up file.
To restore a database:
Select a backup file from the list of files in the backup folder (refer to
section 9.7). allows to review the backup folder in the filesystem.
Choose password options (it is possible to keep the current usernames and
passwords or to restore the one of the corresponding selected backup. It is
also possible to reset the Administrator password to the default one).
Press Restore. A progress bar will show the status of the operation.
Note: For OMNIA network version restoring the database is not managed by
OMNIA but the IT department shall restore it from server.
Data import from CSV is only available with the optional OMNIA Competitive Data
Import module (refer to section 1.4). It allows to import data from third
manufacturers’ equipment by means of a predefined *.csv (comma separated
values) file which is provided with OMNIA.
Note: COSMED does not guarantee the accuracy of the data in the CSV file nor
that the data is compatible with OMNIA database, user must verify the
data contained in the CSV file.
Note: If an error occurs during import in progress no data will be imported in the
OMNIA database.
The Event Log panel shows a record of all events logged by the software which can
be accessed for servicing purposes. Events can be sorted by date, severity (Warning,
QOS, Info, Error), username or machine. Events can be filtered by date by selecting
the day on the calendar and by severity by selecting or deselecting the
corresponding checkboxes. Additional information for each event are shown to the
right side after selection from the Log.
The Audit Log panel shows a record of all audits logged by the software. Audits can
be sorted by date, OS logged user, username, machine or audit information. Audits
can be filtered by date by selecting the day on the calendar and by information type
(Logoff, Logon, Information, Success audit) by selecting or deselecting the
corresponding checkboxes. Additional information for each audit are shown to the
right side after selection from the Log.
10 - Resource Center
10.1 Overview
Resource Center is a database of custom text strings which the user can directly
access during software navigation to automatically populate certain fields.
This software functionality avoids overtyping, reduces operating time and increases
consistency of stored information.
The following fields can be populated using the resource center database:
Software section Fields
Visit Review Interpretation, Notes
Visit Card Physician, Operator, Grouping
Metabolic Testing Reason for stopping test, Reason for test, Test purpose, Marker
Body Composition Density Model
For each field the user can create new strings by clicking on ; when a string is
selected, it can be modified or deleted by clicking respectively on or .
Interpretation and Notes
When creating a string for Interpretation and Notes, it is also possible to assign a
Tag. When the user wants to use a string from the resource center database to
populate Interpretation and/or Notes within the Visit Review panel, it must click on
(see section 6.1.3). The list of available strings is shown and the Tag can be used
to quickly search for the desired string. Default Notes strings are in agreement with
ATS/ERS guidelines. Placeholders referred to the main test parameters are also
available when creating custom interpretation comments and can be added to the
Description from the list to the right (parameters are listed per test type). When the
interpretation comment is selected in the interpretation field of the Visit Review
panel, the placeholder automatically fills with the numerical value of the parameter,
when available.
Density Model
When creating a Density Model you must enter Name, Fat Density and Fat Free
Mass Density values; the corresponding %fat formula is shown.
The user can choose any custom Density Model when starting a Body Composition
test. The configured Density Model will affect final test results.
11.1 Overview
New software versions availability is communicated to OMNIA end users via RSS
Feed. Information on latest software version are always available at
www.cosmed.com/update.
You can request your software upgrade by filling the form linked in the RSS Feed or
in the above COSMED website page. Your request will be processed and you will
promptly receive an e-mail from COSMED Support with upgrade eligibility
information and instructions.
For device Firmware Upgrade please refer also to your device user manual. To
install the latest firmware version, make sure OMNIA software version is updated.
To upgrade firmware:
Check www.cosmed.com/update for firmware updates and submit your
upgrade request if necessary.
Download the upgrade file received from COSMED Support and follow the
received instructions.
Connect the device to the PC via a USB cable or a serial connection and
make sure it is powered ON.
Select Device Manager, click on the device you would upgrade and specify
the communication port (USB or COM port).
Click on Upload Firmware and confirm. A progress bar will prompt to
indicate the update status.
When a message of successful upload prompts, click Ok to complete
firmware installation.
Note: The firmware update takes several minutes, DO NOT interact with or power
off the device during the process.
Firmware upgrade for K5 and Q-NRG devices are managed directly on the unit,
without using OMNIA functionality; refer to K5 and Q-NRG user manuals.
Refer also to Device Manager section (Device Connection) for details about how to
update Firmware on the device.
Both the OMNIA software and the COSMED device need to be authorized to
operate properly.
Each device may authorize differently, depending on the model and serial number
of the unit. Some devices use license dongles, others communicate directly with the
OMNIA software.
Note: Dongles are NOT interchangeable, they are associated with specific device
(S/N). The dongle also contains the software and firmware version
authorizations. Do not attempt to update software and/or firmware
without a valid license, otherwise the software and/or the device will
become inoperable.
If a dongle is removed from a computer and inserted into another, close and restart
OMNIA to obtain licenses.
Note: The license stored on the device contains the software and firmware version
range for which it is authorized. Do not attempt to update software and/or
firmware without having a valid license. If an attempt is made, the software
and/or device may become inoperable.
Device Management
Tab Functionality
Device Selection Select the device and the related communication port
Device Information Once a device is selected and powered on, selecting Read
Information: the license information related to the device
are shown; Clear cache* command removes temporary
files on HDD
Upload Firmware Update firmware on the selected device
Upgrade License Upgrade license on the device (HW and/or SW Module);
see below for further details
Note: *Cached files contain information about authorized devices which have
been connected to the computer in the past 30 days. Clearing the cached
files, the 30-day period for this device will end (i.e. an authorized device is
required to open the OMNIA).
Note: The following procedures can only be performed by the same Windows®
Administrator who installed OMNIA software and should only be performed
in case of necessity.
Note: Use only installation package for the same OMNIA version as the
currently installed one. DO NOT UNINSTALL the software to execute
this task.
Note: Use only installation package for the same OMNIA version as the
currently installed one. DO NOT UNINSTALL the software to execute
this task.
12 - Troubleshooting
Recurrent errors not supported by specific messages are reported below together
with the relevant corrective actions:
Problem Cause Corrective Action
After running COSMED DB The selected database has Remove the
Converter, the database is been previously converted arcconverter.log file from
not converted the Suite software
installation folder and try
again
After software installation, Setup installed only the SQL Start again the software
OMNIA icons are not engine setup
created on PC Desktop and
Antivirus blocked the Disable the antivirus and
the Start menu
program installation start the setup again
The SQL (COSMED) service The database is corrupted Reinitialize the Omnia
return errors during the Database and restore a
manual start backup (if any). Refer to
section 11.6.2
The SQL service engine is N/A
corrupted
While running OMNIA, error messages and notifications may be prompt (usually on
a black vertical bar on the right side of the screen) to report technical failures or
software navigation information. The main and most common messages are
reported in this section.
Note: To solve potential issues, always refer to the instruction included in the
following table (if any). If the problem persists, export the error log (refer to
section 9.10) and contact COSMED Customer Support (see section 13.7.1).
118 Body Box pressure is too Pbox doesn't reach Check that the Body
low. Calibration impossible the 10% of target Box door is closed.
within 6 s Check that the T/RH
probe is properly
inserted in the Body
Box.
Check that the gas
cylinder is not
empty, set to the
correct output
pressure and
opened.
Repeat calibration
123 Motherboard error. Board issues N/A
Contact COSMED Customer
Support
124 Body Box pressure is too Pbox >0.8cmH2O N/A
high
125 Body Box pressure is too Pbox <0.8cmH2O N/A
low
126 Body Box pressure is too N/A
high. Contact COSMED
Customer Support
127 CO2 scrubber is absent or CO2 scrubber not Insert or replace the
expired detected CO2 scrubber
304 Error while loading test. Software issues Try again
Select the test again
312 Error while exporting data Software issues Try again
313 Error while importing data Software issues Try again
353 Error while upgrading the Firmware issues N/A
firmware. Switch the device
OFF/ON and try again
804 License is not found. OMNIA USB dongle Check if the dongle
Connect an authorized license is not is correctly plugged
device or a OMNIA detected in the PC USB port.
hardware key Double check trying
a different USB port.
OMNIA onboard Check if the device is
license is not correctly plugged in
detected the PC USB port.
Double check trying
a different USB port.
Check if the
COSMED Device is
present in the
Windows® Device
Manager and if
there are no errors
reinstall the Omnia
Drivers
The serial number of Check serial number
the device is not on the information
detected (devices menu of the device
with onboard license and report it to
only) COSMED Customer
Support.
806 Licence is not valid. Please Software license not N/A
contact Customer Support valid
808 Network License Services Communication Check connection
are unreachable issues with the license
server and try again
816 Server database has a This client version is Upgrade this client
newer version. Update older than server version
Client version or contact version
the Administrator
817 Client has a newer version. This client version is Upgrade server
Update server database older than server version
version or contact the version
Administrator
820 Unable to load OMNIA, Client and server Align client and
upgrade the Database to versions mismatch server OMNIA
continue versions
822 Database cannot be Existing database Upgrade to an
upgraded to this version. version to old or intermediate
Uninstall OMNIA and install incompatible with the previous version and
a previous version or new one then to the new one
contact COSMED Customer
Support
835 Server Database If “Error: Cannot start Manually start the
unreachable. Error: service SQL (COSMED)
MSSQL$COSMED on service in the
computer”, the SQL Windows® services
(COSMED) service list, if error persists
cannot start. contact COSMED
Customer Support.
If “Error: Login failed Reinitialize the
for user Omnia Database
‘pgmSystem’”, the
database is not
initialized.
1027 Error while updating the Concurrent operation Try again
database. Try again ongoing
1280 Flow out of range Perform syringe
strokes at the
indicated flow
range.
Check flowmeter for
visual damages
1282 Error while processing the Incorrect parameters Try again
data. Restart during operation
1285 Error while uploading the PNT X9 linearization Try again
data. Press YES to retry table issues
1585 Error while saving file. Issues while trying to Look for possible
Check 'Event Log' for save customization resolution
details settings information in Event
Log (see section
9.10)
1766 Invalid data. Restart Incorrect parameters Try again
during MIP/MEP test
1782 Communication error. Incorrect parameters Try again
Restart during DLCO test
BTPS correction
1792 The device in use is not User selected a Change protocol or
compatible with the Dosimeter mode select an
selected protocol protocol and the appropriate device
connected device
does not have an
integrated dosimeter
2049 Error while creating report. Issues while creating Look for possible
Refer to 'Event Log' for report resolution
details information in Event
Log (see section
9.10)
2306 Sentence ID not found GDT error Check input file
2307 Sentence ID 8100 not found GDT error Check input file
or not well-formed or
incorrect file size
2308 Sentence ID {0} not GDT error Check input file
handled
2309 Incorrect sentence length, GDT error Check input file
line {0}
2310 Incorrect sentence format, GDT error Check input file
line {0}
2312 No matching Subject GDT error
2313 No matching Visit GDT error
2320 No matching Test GDT error
List of notifications
Error # Message Note
25 Device is not ready yet. Warm-up is in Allow for proper device
progress warm up according to
87 CO2 calibration is already in progress the device user manual.
Note: The device can be paired (not simultaneously) with multiple PCs. Moving
the same BTA from one PC to another requires re-pairing process.
If the pairing process between the device and BTA or wireless link fails, please
follow the actions summarized below.
Repeat 4-5 for all Generic USB Hub and USB Root Hub.
Run OMNIA and try the pairing process.
PC where the BTA is installed or a nearby device. Turn off nearby Wi-Fi stations and
verify again the device wireless performance.
Note: You may use a 5GHz Wi-Fi network to resolve possible Wi-Fi interferences
13 - Appendix
Complete lists of parameters are shown below per test type, regardless of the
device used. Subsets of such parameters may be visible depending on the device.
13.1.1 Spirometry
Parameter Description Parameter Description
Best FEV1 Best Forced Exp FIV3 Forced Insp. Volume in
Volume in 1 s 3s
Best FVC Best Forced Vital FIV6 Forced Insp. Volume in
Capacity 6s
Best PEF Best Peak Expiratory FIVC Forced Insp. Vital
Flow Capacity
BEV Back Extrapolation FVC Forced Vital Capacity
Volume
BEV/FVC% BEV as % of FVC FVC_QC Quality control grade
referred to FVC
BEV_ins Extrapolated Volume Hesitation Time Hesitation Time
(Insp.)
ChgFEV1 Absolute change in IC Inspiratory Capacity
FEV1 after BD test
ChgFEV1% Relative change in IRV Inspiratory Reserve
FEV1 after BD test Volume
ChgFVC Absolute change in IRV/FVC IRV/FVC
FVC after BD test
ChgFVC% Relative change in FVC IVC Inspiratory Vital
after BD test Capacity
Computer_QC Computer Overall LungAge Lung Age
Quality control grade
Dosimeter Dosimeter Pressure MEF25% Max Exp. Flow @ 75%
Pressure FVC
Effort_QC Quality control grade MEF50% Max Exp. Flow @ 50%
referred to subject s FVC
effort
13.1.6 6MWT
Parameter Description Parameter Description
∆ IC Delta IC IV Max Maximum Inspiratory
Volume
6MWD Distance walked MVV Maximum Voluntary
during Six Minute Ventilation
Walk Test
6MWW Six Minute Walk Work P Diast Diastolic Blood
Pressure
BR Breathing Reserve P Syst Systolic Blood
Pressure
Dyspnea Ratings of Perceived Rf Respiratory Frequency
Exertion
FEV1 Forced Exp Volume in SpO2 Oxygen saturation as
1s measured by pulse
oximetry
HR Heart Rate Sup. O2 flow Supplemental oxygen
(flow)
IC Inspiratory Capacity Suppl. O2 Supplemental oxygen
IV Inspiratory Volume VE Minute Ventilation
13.1.7 Metabolic
CPET
Parameter Description Parameter Description
% Fat % Fat PaO2 Arterial oxygen
tension
Barom_Altitude Barometric Altitude PAO2 Alveolar oxygen
tension
REE
Parameter Description Parameter Description
% Fat % Fat REE Resting Energy
Expenditure
Bias Flow Bias Flow Rf Respiratory Frequency
PRO% Protein %
13.1.9 ECG
Parameter Description Parameter Description
DP Double product S V4 Slope of S-T segment
in Lead V4
HR Heart Rate S V5 Slope of S-T segment
in Lead V5
HRR Heart Rate Reserve S V6 Slope of S-T segment
in Lead V6
HRR_1_minute Heart rate recovery ST aVF Level of S-T segment
after a minute in Lead aVF
P Diast Diastolic Blood ST aVL Level of S-T segment
Pressure in Lead aVL
P Syst Systolic Blood Pressure ST aVR Level of S-T segment
in Lead aVR
RR_ECG R-R interval current ST I Level of S-T segment
heart beat in Lead I
S aVF Slope of S-T segment in ST II Level of S-T segment
Lead aVF in Lead II
S aVL Slope of S-T segment in ST III Level of S-T segment
Lead aVL in Lead III
S aVR Slope of S-T segment in ST V1 Level of S-T segment
Lead aVR in Lead V1
SI Slope of S-T segment in ST V2 Level of S-T segment
Lead I in Lead V2
13.2.1 Spirometry
Predicted Name Reference
GLI Multi-ethnic reference values for spirometry for the 3–95-yr
age range: the global lung function 2012 equations, Philip H.
Quanjer, Sanja Stanojevic, European Respiratory Journal 2012
40: 1324-1343.
ERS93 Standardized Lung Function Testing: Lung Volumes and Forced
ventilatory Flows.Official Statement of the European
Respiratory Society, The European Respiratory Journal Volume
6, Supplement 16, March 1993, 5-40
Zapletal 69 Compilation of reference values for lung function
measurements in children: Ph. H. Quanjer, J. Stocks, G.Polgar,
M. Wise, J. Karlberg, G. Borsboom; ERJ 1989, 2, Supp.4,184s-
261s.
MDB Spirometric reference values from a Mediterranean population:
J, Roca, Navajas, Bull, Eur, Physiopathol, Respir, 1986, 22, 217-
224
Falaschetti Prediction equations for normal and low lung function from the
Health Survey for England, Falaschetti, Eur Respir J. 2004
Mar;23(3):456-63.
Kuster Reference equations for lung function screening of healthy
never-smoking adults aged 18-80 years, Kuster, Eur Respir J.
2008 Apr;31(4):860-8
Forche 27. Verordnung des Bundesministers für Arbeit und Soziales
über die GesundheitsüberwaschungVerordnung des
Bundesministers für Arbeit und Soziales über die
Gesundheitsüberwaschung
Langhammer Forced spirometry reference values for Norwegian adults: the
Bronchial Obstruction in Nord-Trøndelag Study, Langhammer,
Eur Respir J. 2001 Nov; 18(5):770-9.
Solymar Nitrogen single breath test, flow-volume curves and spirometry
in healthy children, 7-18 years of age, Solymar, Eur J Respir Dis.
1980 Oct;61(5):275-86.
13.2.2 DLCO
Predicted Name Reference
ECCS Standardization of the measurement of transfer factor
(diffusing capacity). Report Working Party Standardization of
Lung Function Tests, European Community for Steel and Coal.
Official Statement of the European Respiratory Society, Cotes,
Eur Respir J Suppl. 1993 Mar;16:41-52.
Paoletti Reference equations for the single-breath diffusing capacity. A
cross-sectional analysis and effect of body size and age,
Paoletti, Am Rev Respir Dis. 1985 Oct;132(4):806-13
13.2.8 6MWT
Predicted Name Reference
Enright & Sherril Reference equations for the six-minute walk in healthy adults,
enright, Am J Respir Crit Care Med. 1998 Nov;158(5 Pt 1):1384-
7.
Troosters Six minute walking distance in healthy elderly subjects,
Troosters, Eur Respir J. 1999 Aug;14(2):270-4
Gibbons Reference values for a multiple repetition 6-minute walk test in
healthy adults older than 20 years, Gibbons, J Cardiopulm
Rehabil. 2001 Mar-Apr;21(2):87-93.
Camarri Six minute walk distance in healthy subjects aged 55-75 years,
Camarri, Respir Med. 2006 Apr;100(4):658-65. Epub 2005 Oct
17.
Geiger Six-minute walk test in children and adolescents, Geiger, J
Pediatr. 2007 Apr;150(4):395-9, 399.e1-2.
Chetta Reference values for the 6-min walk test in healthy subjects 20-
50 years old, Chetta, Respir Med. 2006 Sep;100(9):1573-8. Epub
2006 Feb 7.
This section describes the integrated devices and how to configure them on OMNIA
software to execute the tests. Always refer to the device user manual for
information on device operation, connections and parts.
13.4.1 Oximeter
Oximeters measure the hemoglobin saturation and retrieve real-time SpO2 and HR
to OMNIA.
Integrated Oximeters
Integrated oximeters can be used during CPET and RMR tests. To integrate the
parameters measured by the oximeter during a CPET test, Metabolic Cart must be
selected from the HR Source dropdown in the CPET start window.
Note: If more than one HR source is connected to the Metabolic Cart, OMNIA
records the HR value in this order of priority: HR monitors, TTL ECG,
oximeter.
Note: If you are using an adapter and the PC already has an integrated
Bluetooth® radio, be sure that the internal PC Bluetooth® radio is
disabled.
13.4.2 PC ECG
COSMED PC ECGs can be used either as integrated or as standalone devices.
Standalone Configuration
The standalone configuration is described in section 6.10.
Integrated Configuration
The integrated configuration is supported by CPET tests. To execute a CPET test with
the Stress PC ECG software:
Select PC Software ECG from the HR Source dropdown in the CPET start
window. The Stress PC ECG software opens automatically and allows to
configure the ECG settings by pressing on button. The control of
equipment is relegated to OMNIA and the protocol phases, load values and
blood pressure are imported and displayed on PC Software ECG (remote
protocol configuration disables PC ECG software protocol buttons).
Press in OMNIA to start the test.
Run the CPET test as described in section 6.8 and use the Stress PC ECG
software to store events.
Press in OMNIA to stop the test and to store the recorded ECG in
OMNIA in PDF format.
Press in OMNIA to edit the ECG on the Stress PC ECG software and
generate PDF report.
During the editing OMNIA shows a message on the side bar. Press Ok in the
side bar to save the edited ECG in OMNIA.
TTL ECGs can be used as HR sources during CPET tests to integrate the HR values
only. Metabolic Cart must be selected from the HR Source dropdown in the CPET
start window to configure a TTL ECG.
Note: If more than one HR source is connected to the Metabolic Cart, OMNIA
records the HR value in this order of priority: HR monitors, TTL ECG,
oximeter.
13.4.4 Mortara®
Mortara® ECG can be used as an integrated device during CPET tests. The
integration requires the auxiliary Mortara® X-Scribe CP software and Mortara®
Stress, GDT Cardiopulmonary and Connectivity licenses. However, refer to Mortara®
for all up-to-date information.
To execute a CPET test with the integrated Mortara® ECG:
Select Mortara® XScribe from the HR Source dropdown in the CPET start
window. X-Scribe is automatically launched in Cardiopulmonary mode (X-
Scribe CP) and works as a slave application to OMNIA. The control of
equipment is relegated to OMNIA and the protocol phases, load values and
blood pressure are imported and displayed on Mortara® X-Scribe CP.
X-Scribe CP enters the Observe phase. Wait for the automatic detection of
the average complex on X-Scribe CP and press in OMNIA to start the
test.
Run the CPET test as described in section 6.8.
Press in OMNIA to stop the test.
X-Scribe CP enters a Monitoring phase. During this phase the interpretation
can be written.
Press Save on X-Scribe CP to export the ECG report (Mortara® PDF must be
selected to export the report as a PDF) and then press Exit to save report in
OMNIA database.
GE® Cardiosoft, starting from version 7.0, can be used to integrate GE® ECGs during
CPET tests. The integration can be implemented in three different modes,
depending on the physical link used to connect the software:
GE® Cardiosoft and OMNIA running on the same PC.
GE® Cardiosoft and OMNIA running different PCs, connected by a crossover
LAN cable or by using a switch or an HUB connected with patch cable.
GE® Cardiosoft and OMNIA running different PCs, connected by RS-232 null
modem cable.
In all modes GE® Cardiosoft works as a Master while OMNIA works as a Slave,
therefore OMNIA toolbars become inoperable during testing and protocol creation
and ergometers control are relegated to GE® Cardiosoft.
Note: Ensure that before the exercise phase at least two phases are created,
which can either be left blank or set to rest and warm-up (multiple warm-up
phases accepted).
Exchanged data (protocol information, HR, P Diast and P Syst) do not differ between
modes but procedures to run a test are different. OMNIA shows the required steps
for each mode when running the CPET-ECG integrated test.
Mode 1
Mode 3
To execute a CPET test:
Select GE Cardiosoft-CASE from the HR Source dropdown in the CPET start
window.
Open GE® Cardiosoft.
Enter patient data in GE® Cardiosoft.
Select protocol and start the test in GE® Cardiosoft.
Control test phases, including ergometer loads and BP values, from GE®
Cardiosoft.
Press Stop in Cardiosoft and close Cardiosoft.
Stop the CPET test in OMNIA.
Open GE® Cardiosoft to view the ECG PDF report or edit the ECG test.
13.4.7 Tango®
Suntech® Tango® M2 (REF A-661-200-018) is an automated blood pressure monitor
which measures and displays blood pressure synchronized with the subject's ECG.
Tango® retrieves a subject’s systolic and diastolic blood pressure during a CPET test.
It must be powered on before COSMED equipment and requires the integration
with an ECG. Tango® must be selected from the Blood Pressure dropdown in the
CPET start window to configure a Tango®.
Note: Tango® requires the AUX Devices Integration module (REF A-670-100-023).
Note: N-SenTec® DMS requires the AUX Devices Integration module (REF A-670-
100-023).
Note: Radiometer® TCM5 requires the AUX Devices Integration module (REF A-
670-100-023).
13.4.10 PhysioFlow
PhysioFlow devices integrate with OMNIA retrieving real-time Cardiac Output,
Stroke Volume, Cardiac Index and HR (refer to section 13.1.7) during CPET tests. In
the CPET start window the PhysioFlow communication port must be selected from
the Physioflow dropdown to configure it.
Note: PhysioFlow devices require the AUX Devices Integration module (REF A-670-
100-023).
Note: The database conversion is only available for Suite software versions >10.0.
For previous versions please contact COSMED Support.
Note: After database conversion, do not use the Suite software on the same PC as
OMNIA. To avoid errors, remove the Suite software installation folder from
the PC where OMNIA is installed.
Treadmills
Brand/Model Name on OMNIA Cable Note
COSMED COSMED Treadmill / REF C01714-01-12 Several models.
Treadmill HP COSMOS Mph and Kmh
COSMED -- Several models.
Treadmill v4 Mph and Kmh
HP Cosmos® -- Several models.
Mph and Kmh
HP Cosmos®v4 -- Several models.
Mph and Kmh
Trackmaster® Trackmaster REF C01936-01-12 Mph and Kmh
Trackmaster® Trackmaster_v1.2 -- --
Trackmaster® 425 Trackmaster 425 -- Mph and Kmh
Trackmaster® 428 Trackmaster TM428 REF C01936-01-12 Mph and Kmh
Others
Type Note
Arm Ergometer Ergometer not controlled. Can be selected as an additional
test information
Field Testing Can be selected as an additional test information (e.g. when
using a wearable device)
Bike not interfaced Bike is not controlled however you can set a protocol
Treadmill not interfaced Treadmill is not controlled however you can set a protocol
Treadmills
Brand/Model Name on OMNIA Cable Note
Archimed Archimed -- Mph
C-Safe C-Safe --
Technogym®Excit Excite-Treadmill REF C02964-01-12 Mph
e
Imbramed Imbramed Treadmill- --
ATL / Imbramed
Treadmill Master
Lode® Lode Treadmills REF C02567-01-12 Mph and Kmh
Lode® v2 Lode Treadmills v2 REF C02567-01-12 Mph and Kmh
MarquetteT2000 MarquetteT2000 -- Mph and Kmh
MarquetteT2100 MarquetteT2000 -- Mph and Kmh
Power Jog Power Jog
Ram 770 RAM 770 REF C01936-01-12
Technogym®RunR TechnogymRunRace --
ace
Techmachine Tecmachine 1800 --
Track Emul Track Emul --
Woodway® Woodway REF C01714-01-12 Mph and Kmh
Desmo / Front
Line
Woodway® Pro Woodway Pro REF C01714-01-12 Kmh
13.8 References
13.9 Notes
info@cosmed.com | cosmed.com