Do’s & Don'ts while using Euro Lab 200 / 240

DO’s DON’Ts
• Do a daily & weekly maintenance procedure as per the instruction • Do not use the reagents immediately after taking it from
provided . refrigerator
• Load the reagents & wash solution in proper position • Do not overfill / top up the reagents
• Label the bottles mentioning the name of the parameter with R1 and • Do not use calibrator / control in the frozen state, thaw it
R2 correctly on the respective bottles and its caps. properly and then use it
• Make sure that the reagent bottle caps are removed and properly • Do not use control/ calibrator/ reagent beyond mentioned time
placed for performing the test. limit.
• Load sufficient reagents as per daily need. • Don’t use clotted/ hemolyzed / lipemic samples
• Avoid loading samples in primary tubes if not fitting in the
• Properly thaw the calibrator and control before using.
sample rack.
• Do calibration of parameters at the required intervals as per the • Don’t open the lid while machine is in use
instruction provided.
• Don’t load reagents during operation using pipettes
• Perform Quality Controls (QC) on a daily basis.
• Don’t put hand inside during operations or probe movement
• Centrifuge the samples properly • Don’t take out the rack in which sample is loaded
• Use sample in FRESH sample cups. • Don’t reuse cal/control once it is used
• Use fresh tips and sample vials for all purpose. • Don’t install any antivirus software in the computer connected
• Store reagents at 2-8°C, calibrators and controls at deep freezer into with the instrument.
sample vial. • Don’t use associated computer for any other purpose other than
• Make sure that the distilled water can is filled with enough water. instrument operation
• Ensure good internet connectivity is there for team viewer support • Don’t forget to switch off the instrument when not in use
• Make sure that the instrument is not going to sleep mode or screen • Don’t switch off the cooling system of the instrument when
saver mode. reagents are kept on the instrument .