ZN Nhom 6

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

Method: Zn 19.09.

2023 Page 1 Date: 9/19/2023 2:01:13 PM

====================================================================================================
Sequence No.: 46 Autosampler Location:
Sample ID: blank Date Collected: 9/19/2023 1:58:57 PM
Analyst: Data Type: Original

----------------------------------------------------------------------------------------------------
Replicate Data: blank Analyte: Zn 213.86
Repl SampleConc StndConc BlnkCorr Time Signal
# mg/L mg/L Signal Stored
1 [0.00] 0.905 1:58:58 PM No
2 [0.00] 0.906 1:59:03 PM No
3 [0.00] 0.901 1:59:07 PM No
Mean: [0.00] 0.904
SD: 0.0000 0.0027
%RSD: 0.00% 0.30
Auto-zero performed.

====================================================================================================
Sequence No.: 47 Autosampler Location:
Sample ID: chuan 1 Date Collected: 9/19/2023 1:59:16 PM
Analyst: Data Type: Original

----------------------------------------------------------------------------------------------------
Replicate Data: chuan 1 Analyte: Zn 213.86
Repl SampleConc StndConc BlnkCorr Time Signal
# mg/L mg/L Signal Stored
1 [0.2] 0.111 1:59:16 PM No
2 [0.2] 0.117 1:59:20 PM No
3 [0.2] 0.112 1:59:25 PM No
Mean: [0.2] 0.113
SD: 0.000 0.0034
%RSD: 0.00% 2.97
Standard number 1 applied. [0.2]
Correlation Coef.: Slope: 0.89834 Intercept: 0.00000
Standard absorbance and concentration values are not in the same order.
S-shaped calibration curve detected. Two-coefficient equation used.

====================================================================================================
Sequence No.: 48 Autosampler Location:
Sample ID: chuan 2 Date Collected: 9/19/2023 1:59:35 PM
Analyst: Data Type: Original

----------------------------------------------------------------------------------------------------
Replicate Data: chuan 2 Analyte: Zn 213.86
Repl SampleConc StndConc BlnkCorr Time Signal
# mg/L mg/L Signal Stored
1 [0.4] 0.225 1:59:37 PM No
2 [0.4] 0.220 1:59:41 PM No
3 [0.4] 0.213 1:59:46 PM No
Mean: [0.4] 0.219
SD: 0.000 0.0062
%RSD: 0.00% 2.82
Standard number 2 applied. [0.4]
Correlation Coef.: Slope: 1.25751 Intercept: 0.00000
Standard absorbance and concentration values are not in the same order.
S-shaped calibration curve detected. Two-coefficient equation used.
Method: Zn 19.09.2023 Page 2 Date: 9/19/2023 2:01:13 PM

====================================================================================================
Sequence No.: 49 Autosampler Location:
Sample ID: chuan 3 Date Collected: 9/19/2023 1:59:55 PM
Analyst: Data Type: Original

----------------------------------------------------------------------------------------------------
Replicate Data: chuan 3 Analyte: Zn 213.86
Repl SampleConc StndConc BlnkCorr Time Signal
# mg/L mg/L Signal Stored
1 [0.6] 0.317 1:59:55 PM No
2 [0.6] 0.316 1:59:59 PM No
3 [0.6] 0.323 2:00:03 PM No
Mean: [0.6] 0.319
SD: 0.000 0.0035
%RSD: 0.00% 1.11
Standard number 3 applied. [0.6]
Correlation Coef.: Slope: 0.79675 Intercept: 0.00000
Standard absorbance and concentration values are not in the same order.
S-shaped calibration curve detected. Two-coefficient equation used.

====================================================================================================
Sequence No.: 50 Autosampler Location:
Sample ID: chuan 4 Date Collected: 9/19/2023 2:00:11 PM
Analyst: Data Type: Original

----------------------------------------------------------------------------------------------------
Replicate Data: chuan 4 Analyte: Zn 213.86
Repl SampleConc StndConc BlnkCorr Time Signal
# mg/L mg/L Signal Stored
1 [0.8] 0.394 2:00:13 PM No
2 [0.8] 0.405 2:00:17 PM No
3 [0.8] 0.409 2:00:21 PM No
Mean: [0.8] 0.403
SD: 0.000 0.0079
%RSD: 0.00% 1.95
Standard number 4 applied. [0.8]
Correlation Coef.: 0.999979 Slope: 0.57869 Intercept: 0.00000

----------------------------------------------------------------------------------------------------
Calibration data for Zn 213.86 Equation: Nonlinear Through Zero
Entered Calculated
Mean Signal Conc. Conc. Standard
ID (Abs) mg/L mg/L Deviation %RSD
blank 0.000 0 0.000 0.00 0.30
chuan 1 0.113 0.2 0.200 0.00 2.97
chuan 2 0.219 0.4 0.398 0.01 2.82
chuan 3 0.319 0.6 0.602 0.00 1.11
chuan 4 0.403 0.8 0.799 0.01 1.95
Correlation Coef.: 0.999979 Slope: 0.57869 Intercept: 0.00000

====================================================================================================
Sequence No.: 51 Autosampler Location:
Sample ID: nhom 6 Date Collected: 9/19/2023 2:00:38 PM
Analyst: Data Type: Original

----------------------------------------------------------------------------------------------------
Replicate Data: nhom 6 Analyte: Zn 213.86
Repl SampleConc StndConc BlnkCorr Time Signal
Method: Zn 19.09.2023 Page 3 Date: 9/19/2023 2:01:13 PM

# mg/L mg/L Signal Stored


1 0.050 0.050 0.029 2:00:38 PM No
2 0.095 0.095 0.054 2:00:42 PM No
3 0.087 0.087 0.050 2:00:47 PM No
Mean: 0.077 0.077 0.044
SD: 0.0237 0.0237 0.0135
%RSD: 30.68% 30.68% 30.49