Professional Documents
Culture Documents
Higaki Et Al 2018 Introduction To The Technical Aspects of Computed Diffusion Weighted Imaging For Radiologists
Higaki Et Al 2018 Introduction To The Technical Aspects of Computed Diffusion Weighted Imaging For Radiologists
org
1131
Imaging Physics
Introduction to the Technical
Aspects of Computed Diffusion-
weighted Imaging for Radiologists
Toru Higaki, PhD
Yuko Nakamura, MD, PhD Diffusion-weighted (DW) imaging is a magnetic resonance (MR)
Fuminari Tatsugami, MD, PhD imaging method. It is an indispensable sequence for the diagnosis
Yoko Kaichi, MD, PhD of acute cerebral infarction and is recognized as a standard tool in
Motonori Akagi, MD oncologic imaging. Computed DW imaging refers to the synthesiz-
Yuij Akiyama, RT ing of arbitrary b-value DW images from a set of measured b-value
Yasutaka Baba, MD, PhD images by voxelwise fitting. Computed DW imaging is advanta-
Makoto Iida, MD, PhD geous because it generates DW images with a higher diffusion effect
Kazuo Awai, MD, PhD than that achievable by using the MR imaging units in use today.
Additionally, computed DW imaging can reduce imaging time
Abbreviations: ADC = apparent diffusion co- while producing images characterized by a higher signal-to-noise
efficient, CNR = contrast-to-noise ratio, DW =
diffusion-weighted, FDG = fluorodeoxyglucose, ratio than what the acquired DW images would display at the cor-
MPG = motion probing gradient, SNR = signal- responding b values. By fitting input images acquired at a lower b
to-noise ratio, TE = echo time
value and correspondingly a shorter echo time, the signal intensity
RadioGraphics 2018; 38:1131–1144 of the resulting computed DW image is closer to the ideal case.
https://doi.org/10.1148/rg.2018170115 Computed DW images are generated by employing mathematical
Content Codes: models that use mono-, bi-, or triexponential equations. To generate
accurate computed DW images, the appropriate model must be se-
From the Department of Diagnostic Radiology,
Hiroshima University, 1-2-3 Kasumi, Minami- lected, and the image parameters for the input data must be chosen
ku, Hiroshima 734-8551, Japan (T.H.,Y.N., F.T., accordingly. In addition, to reduce artifacts on computed DW im-
Y.K, M.A., Y.B., M.I., K.A.); and Department of
Clinical Radiology, Hiroshima University Hospi- ages, the misalignment of input data must be corrected with the aid
tal, Hiroshima, Japan (Y.A.). Presented as an ed- of image registration techniques.
ucation exhibit at the 2016 RSNA Annual Meet-
ing. Received April 28, 2017; revision requested ©
RSNA, 2018 • radiographics.rsna.org
September 8 and received October 6; accepted
October 17. For this journal-based SA-CME ac-
tivity, the author K.A. has provided disclosures
(see end of article); all other authors, the editor,
and the reviewers have disclosed no relevant re-
lationships. Address correspondence to T.H.
Introduction
(e-mail: higaki@hiroshima-u.ac.jp). Diffusion-weighted (DW) imaging is an established magnetic reso-
©
RSNA, 2018
nance (MR) imaging method for whole-body imaging and can be
used to diagnose a wide spectrum of diseases (1–6). For example,
DW imaging is a standard tool for diagnosing ischemic stroke, as dif-
SA-CME Learning Objectives
fusion changes can be observed on DW images and apparent diffu-
After completing this journal-based SA-CME sion coefficient (ADC) maps earlier than with other imaging modali-
activity, participants will be able to:
ties. Analogously, as malignant tumors tend to be characterized by
■■Define the principles of DW imaging.
increased cellularity and typically demonstrate impeded diffusion at
■■Describe the basic theory and advan-
tages of computed DW imaging.
imaging, the signal intensity of malignant lesions on DW images is
■■Identify the clinical uses and applica-
higher than that of the normal parenchyma (1).
tions of computed DW imaging. A crucial parameter used in DW imaging is the b value (7),
See rsna.org/learning-center-rg. which refers to the strength of the motion probing gradient (MPG)
pulse; the contrast ratio on DW images increases as the b value
increases. However, the absolute signal intensity of a DW image is
lower with higher b values, resulting in a decrease in the signal-to-
noise ratio (SNR). Consequently, it is generally difficult to obtain
DW images with ultra-high b values at lower magnetic fields (1).
1132 July-August 2018 radiographics.rsna.org
Figure 3. Flow chart shows the generation process of an axial computed DW image. In Step 1 (teal line), the ADC values are calcu-
lated from signal intensities measured at two b values using Equation 2. In Step 2 (purple line), the signal intensity of the computed
DW image at the arbitrary target b value (bx) is calculated using Equation 1, using the measured signal intensity of b0 and the ADC
value calculated in step 1.
Sbi bi , (4)
the influence of noise when estimating the slope pixel value of Sbi or Sb0 approaches 0 asymptoti-
by increasing the interval between the b values. cally, the ADC value asymptotically approaches
negative infinity or positive infinity. When the
Target b Value.—A high b value enhances the ADC value is ± infinity, artifacts can be reduced
contrast of the lesion, whereas the SNR of the by setting the output pixel value to 0.
image decreases. The optimal b value is dependent
on the organ being imaged (8,9). In many cases, a Clinical Advantages of Computed
value that is slightly higher than that used during DW Imaging
the routine examination is selected (11,13,16,20). One of the advantages of using computed DW
imaging is that the b value can be changed arbi-
Misalignment between Input Images trarily. In particular, even at low magnetic fields,
When generating computed DW images, it is at which it is difficult to acquire high b-value
necessary to correct morphologic misalignment DW images, such images can be generated by
between the input images. Because computed computation.
DW images are calculated voxelwise from the sig- In general, computed DW imaging allows for
nal intensity of matching voxel positions in each the calculation of arbitrary b-value images by
input image, each voxel should be acquired from postprocessing after completion of the imag-
the same location in the patient’s body. ing examination (8,9). This benefits diagnostic
Among the several causes of misalignment dur- ability, as a high b value results in high lesion
ing imaging, one is gross motion during the exam- contrast. Additionally, increasing the b value
ination. Even if the patient’s body is stationary, the through computation can reduce the T2 shine-
position of the organ may shift during imaging ow- through effect. However, it is important to note
ing to the patient’s heartbeat or breathing (35–37). that the b value cannot be increased infinitely. As
The influence of cardiac pulsation is apparent on shown in Equation 1, signal intensity decreases
the arterial vasculature, and it has a nonnegligible as the b value increases; if the b value is set too
effect, especially on high b-value images, which are high, the SNR drops, resulting in considerable
sensitive to microscopic motion. image quality degradation.
Another cause of image misalignment is the Another benefit of computed DW imaging is
distortion associated with magnetic field inho- the use of a shorter TE. As described previously,
mogeneities, which can be caused by factors signal intensity decreases as the TE increases,
such as the patient’s magnetic susceptibility and causing the SNR on DW images to drop. In
the eddy current derived from the MPG pulses addition, extending TE elicits T2 shine-through
(37,38). The MPG pulse generates eddy currents effects, potentially causing false-positive high-in-
in various metallic components of the MR imag- tensity lesions to be diagnosed (16). TE elonga-
ing unit, resulting in unpredictable magnetic field tion is associated with the long duration of the
inhomogeneities that can distort DW images. MPG pulse required to obtain a high b-value
Image registration algorithms provide an effi- image. With the computed DW imaging tech-
cient tool to correct misalignments (39–43). Image nique, images can be generated at a short TE
registration is the process of transforming different with the desired high b value, using a short TE
images into the same coordinate system. Available and low b-value images as the input data.
image registration methods are rigid, affine, and
deformable registration. Such algorithms are clas- Improving Contrast-to-Noise Ratio
sified on the basis of the difference in the method There are two ways to improve the contrast-to-
applied for transforming the image (44,45). The noise ratio (CNR) on DW images, as described
registration algorithms can also be classified on the previously. The first is to shorten the TE to
basis of the method used to calculate the degree reduce image noise by inputting images with low
of alignment between images, including the sum b values. The second way to improve CNR is to
(mean) of squared difference (46), mutual infor- increase the b value of the generated computed
mation (42,43), and the correlation coefficient DW image to increase the contrast between
(40,41). Figure 9 shows computed DW images normal tissue and the lesion (48).
with and without misalignment correction. Figures 10–12 show cases in which the CNR
is improved by using computed DW imaging.
Removing Outlier Pixel Values Figure 10 shows a patient with acute isch-
During the computed DW imaging generation emic stroke; Figure 11 shows brain images of
process, outlier pixel values may appear. Such a healthy subject; and Figure 12 shows images
pixels arise from image noise or artifacts in the obtained in a patient with prostate cancer.
input images and result in artifacts on computed In the case of a patient with acute ischemic
DW images. For example, in Equation 2, if the stroke (Fig 10), it would have been desirable to
1138 July-August 2018 radiographics.rsna.org
Figure 11. Brain images of a healthy 34-year-old male subject. (a) Axial computed DW image shows the
brain, obtained with a target b value of 4000 sec/mm2 and a TE of 98 msec. This image was generated from
input images with b values of 250, 750, 1250, and 2000 sec/mm2, using the biexponential model described
in Equation 9. (b) Axial DW image shows the brain, obtained with a b value of 4000 sec/mm2 and a TE of
122 msec. The DW image appears noisy, but this effect is reduced on the computed DW image.
1140 July-August 2018 radiographics.rsna.org
Figure 12. Lesion depiction in a 77-year-old man with prostate cancer. (a) Axial T2-weighted MR image (T2WI) of the
prostate shows a lesion (arrowheads). (b) Axial DW image (DWI) obtained with a b value of 2000 sec/mm2 shows the
lesion (arrowheads). Dashed square = area of interest magnified in c. (c) Axial computed DW images (cDWI) with pro-
gressively increasing b value application show the lesion. The signal intensity of the lesion increases with the increasing b
values, resulting in clear lesion depiction (arrowhead) at 2000 sec/mm2. The images were generated from input images
with b values of 0 and 1000 sec/mm2, using the monoexponential model. The target b values for the computed DW im-
ages ranged from 200 to 2000 sec/mm2.
(Fig 11a) compared to that of the measured DW false-positive high-intensity lesion. Figures 13–15
image (Fig 11b) because the effect of the T2 demonstrate to what extent the T2 shine-through
relaxation is decreased by using a shorter TE. effect is reduced by using computed DW imaging.
In Figure 12, the signal intensity of the lesion in- Figure 13 shows images obtained in a patient with
creases with the application of progressively higher multiple hepatic cysts; Figure 14 shows images
b values, resulting in clearer lesion depiction. The obtained in a patient with hepatic metastasis from
diagnostic ability of computed DW imaging when colon cancer; and Figure 15 shows images ob-
applied to prostate cancer has been reported to tained in a patient with brain metastasis.
improve by increasing the b value (11,13,20). In Figure 13b, the hepatic cysts appear as
high-intensity lesions. However, the ADC map
Eliminating T2 Shine-Through Effect (Fig 13c) does not display diffusion restric-
The computed DW imaging technique makes it tion for these lesions, which is thought to be
possible to generate computed DW images with attributable to the T2 shine-through effect. By
a short TE and a high b value by relying on input generating a computed DW image with a b value
images meeting these requirements. A shorter of 1500 sec/mm2, the lesions can be correctly
TE and/or high b value eliminates the T2 shine- evaluated without referring to the ADC map.
through effect that may result in the depiction of a Nakamura et al (16) reported the usefulness of
RG • Volume 38 Number 4 Higaki et al 1141
Figure 13. Hepatic cysts in a 74-year-old woman. (a) Axial computed DW image (cDWI) shows the hepatic cysts (arrowheads),
imaged with a target b value of 1500 sec/mm2, generated from input DW images with b values of 600 and 150 sec/mm2, using the
monoexponential model. ms = msec. (b) Axial DW image (DWI) shows the hepatic cysts (arrowheads) as high-intensity lesions, imaged
with a b value of 1000 sec/mm2. (c) Axial ADC map shows the hepatic cysts. By generating a computed DW image with a b value of
1500 sec/mm2, the lesions can be correctly evaluated without referring to the ADC map.
computed DW imaging for imaging noncompli- Although the theory underlying computed DW
cated hepatic cysts. imaging is simple, there are several important key
In Figure 14b, the tumor appears as a uniform aspects that should be observed. Most notably,
lesion on the measured DW image with a b value when computed DW images with ultra-high b
of 600 sec/mm2. On the computed DW image, values are generated, accurate images can be ob-
only the edges of the lesion display high signal tained only by applying the triexponential model.
intensity, as the signal intensity in the necrotic Even for the generation of lower b-value com-
region decreases with reduced T2 shine-though puted DW images, the appropriate computation
effect. This matches the pattern of FDG PET models must be combined with carefully chosen
uptake (Fig 14d). b values for the input DW images. Computed
In Figure 15, on the measured DW image, the DW imaging falls in the category of synthetic
edematous region exhibits isointense signal com- MR imaging, which has had growing visibility
pared with that of the surrounding region; this is in recent years. For the widespread application
thought to be due to the T2 shine-through effect. of computed DW imaging, the framework of
Since the TE is short on the computed DW im- synthetic MR imaging may be helpful for the ef-
age, the T2 shine-through effect is reduced, and ficient acquisition of input images.
the signal intensity of the edema decreases.
Disclosures of Conflicts of Interest.—K.A. Activities related
Conclusion to the present article: disclosed no relevant interests. Activities
not related to the present article: research grants from Toshiba
Computed DW imaging allows for the genera- Medical Systems, Hitachi HealthCare, Fujitsu, Eisai, Daiichi
tion of new contrast images at postprocessing. Sankyo, Bayer Yakuhin, and Fuji Yakuhin; member of the
1142 July-August 2018 radiographics.rsna.org
Figure 14. Hepatic metastasis (arrowhead) in a 79-year-old woman. (a) Axial computed DW image (cDWI) shows the hepatic metas-
tasis with high signal intensity around the edges (b = 1500 sec/mm2). The image was generated from input DW images with b values
of 600 and 0 sec/mm2, using the monoexponential model. ms = msec. (b) Axial DW image (DWI) shows a uniform lesion, imaged with
a b value of 600 sec/mm2. (c) Axial gadolinium ethoxybenzyl–enhanced hepatobiliary phase MR image shows the hepatic metastasis.
(d) Axial fluorine 18 fluorodeoxyglucose (FDG) positron emission tomography (PET) color-mapped fusion image shows the hepatic
metastasis. The signal intensity of the tumor on the computed DW image (a) matches the pattern of FDG PET uptake.
Figure 15. T2 shine-through effect in a 75-year-old woman with brain metastasis. (a) Axial computed DW image (cDWI) of the
brain shows edema with low signal intensity, obtained with a target b value of 4000 sec/mm2. The image was generated from input
images with b values of 250, 750, 1250, and 2000 sec/mm2, using the biexponential model (Eq 9). ms = msec. (b) Axial DW image
(DWI) (b = 4000 sec/mm2) shows the edema with iso–signal intensity compared with that of the surrounding region, thought to be
due to the T2 shine-through effect. (c) Axial ADC map of the brain shows the edema. Since the TE on the computed DW image (a)
is short, the T2 shine-through effect is reduced, and the signal intensity of the edema decreases.
RG • Volume 38 Number 4 Higaki et al 1143
Medical Advisory Board for GE Healthcare. Other activities: nance imaging for prostate cancer diagnosis. Eur Radiol
disclosed no relevant relationships. 2013;23(12):3509–3516.
21. Barral M, Taouli B, Guiu B, et al. Diffusion-weighted MR
imaging of the pancreas: current status and recommendations.
References Radiology 2015;274(1):45–63.
1. Koh DM, Collins DJ. Diffusion-weighted MRI in the body: 22. Kallehauge JF, Tanderup K, Haack S, et al. Apparent diffusion
applications and challenges in oncology. AJR Am J Roentgenol coefficient (ADC) as a quantitative parameter in diffusion
2007;188(6):1622–1635. weighted MR imaging in gynecologic cancer: dependence
2. Schaefer PW, Grant PE, Gonzalez RG. Diffusion-weighted on b-values used. Acta Oncol 2010;49(7):1017–1022.
MR imaging of the brain. Radiology 2000;217(2):331–345. 23. Karki K, Hugo GD, Ford JC, et al. Estimation of optimal
3. Hagmann P, Jonasson L, Maeder P, Thiran JP, Wedeen b-value sets for obtaining apparent diffusion coefficient free
VJ, Meuli R. Understanding diffusion MR imaging tech- from perfusion in non-small cell lung cancer. Phys Med Biol
niques: from scalar diffusion-weighted imaging to diffusion 2015;60(20):7877–7891.
tensor imaging and beyond. RadioGraphics 2006;26(suppl 24. Le Bihan D, Turner R, Moonen CT, Pekar J. Imaging of
1):S205–S223. diffusion and microcirculation with gradient sensitization:
4. González RG, Schaefer PW, Buonanno FS, et al. Diffusion- design, strategy, and significance. J Magn Reson Imaging
weighted MR imaging: diagnostic accuracy in patients 1991;1(1):7–28.
imaged within 6 hours of stroke symptom onset. Radiology 25. Dijkstra H, Baron P, Kappert P, Oudkerk M, Sijens PE. Ef-
1999;210(1):155–162. fects of microperfusion in hepatic diffusion weighted imaging.
5. Srinivasan A, Goyal M, Al Azri F, Lum C. State-of-the-art Eur Radiol 2012;22(4):891–899.
imaging of acute stroke. RadioGraphics 2006;26(suppl 26. Chandarana H, Lee VS, Hecht E, Taouli B, Sigmund EE.
1):S75–S95. Comparison of biexponential and monoexponential model
6. Rajeshkannan R, Moorthy S, Sreekumar KP, Rupa R, Prabhu of diffusion weighted imaging in evaluation of renal lesions:
NK. Clinical applications of diffusion weighted MR imaging: preliminary experience. Invest Radiol 2011;46(5):285–291.
a review. Indian J Radiol Imaging 2006;16(4):705–710. 27. Lin M, Yu X, Chen Y, et al. Contribution of mono-expo-
7. Le Bihan D, Breton E, Lallemand D, Grenier P, Cabanis nential, bi-exponential and stretched exponential model-
E, Laval-Jeantet M. MR imaging of intravoxel incoherent based diffusion-weighted MR imaging in the diagnosis and
motions: application to diffusion and perfusion in neurologic differentiation of uterine cervical carcinoma. Eur Radiol
disorders. Radiology 1986;161(2):401–407. 2017;27(6):2400–2410.
8. Blackledge MD, Leach MO, Collins DJ, Koh DM. Com- 28. Hayashi T, Miyati T, Takahashi J, et al. Diffusion analysis
puted diffusion-weighted MR imaging may improve tumor with triexponential function in liver cirrhosis. J Magn Reson
detection. Radiology 2011;261(2):573–581. Imaging 2013;38(1):148–153.
9. Blackledge MD, Wilton B, Messiou C, Koh DM, Leach MO, 29. Hayashi T, Miyati T, Takahashi J, et al. Diffusion analysis
Collins DJ. Computed diffusion weighted imaging (cDWI) with triexponential function in hepatic steatosis. Radiol Phys
for improving imaging contrast (abstr). In: Proceedings of the Technol 2014;7(1):89–94.
Seventeenth Meeting of the International Society for Magnetic 30. Ueda Y, Takahashi S, Ohno N, et al. Triexponential function
Resonance in Medicine. Berkeley, Calif: International Society analysis of diffusion-weighted MRI for diagnosing prostate
for Magnetic Resonance in Medicine, 2009; 4005. cancer. J Magn Reson Imaging 2016;43(1):138–148.
10. Kawahara S, Isoda H, Fujimoto K, et al. Additional benefit of 31. Cercueil JP, Petit JM, Nougaret S, et al. Intravoxel incoherent
computed diffusion-weighted imaging for detection of hepatic motion diffusion-weighted imaging in the liver: comparison
metastases at 1.5T. Clin Imaging 2016;40(3):481–485. of mono-, bi- and tri-exponential modelling at 3.0-T. Eur
11. Maas MC, Fütterer JJ, Scheenen TW. Quantitative Radiol 2015;25(6):1541–1550.
evaluation of computed high b value diffusion-weighted 32. Hu YC, Yan LF, Sun Q, et al. Comparison between ultra-
magnetic resonance imaging of the prostate. Invest Radiol high and conventional mono b-value DWI for preoperative
2013;48(11):779–786. glioma grading. Oncotarget 2017;8(23):37884–37895.
12. Gatidis S, Schmidt H, Martirosian P, Nikolaou K, Schwenzer 33. Xueying L, Zhongping Z, Zhoushe Z, et al. Investigation
NF. Apparent diffusion coefficient-dependent voxelwise com- of apparent diffusion coefficient from ultra-high b-values in
puted diffusion-weighted imaging: an approach for improving Parkinson’s disease. Eur Radiol 2015;25(9):2593–2600.
SNR and reducing T2 shine-through effects. J Magn Reson 34. Ogura A, Koyama D, Hayashi N, Hatano I, Osakabe K,
Imaging 2016;43(4):824–832. Yamaguchi N. Optimal b values for generation of com-
13. Yoshida R, Yoshizako T, Katsube T, Tamaki Y, Ishikawa puted high-b-value DW images. AJR Am J Roentgenol
N, Kitagaki H. Computed diffusion-weighted imaging us- 2016;206(4):713–718.
ing 1.5-T magnetic resonance imaging for prostate cancer 35. Pang Y, Yu B, Zhang X. Hepatic fat assessment using ad-
diagnosis. Clin Imaging 2017;41:78–82. vanced magnetic resonance imaging. Quant Imaging Med
14. Bammer R. Basic principles of diffusion-weighted imaging. Surg 2012;2(3):213–218.
Eur J Radiol 2003;45(3):169–184. 36. Kandpal H, Sharma R, Madhusudhan KS, Kapoor KS.
15. Holdsworth SJ, Yeom KW, Antonucci MU, et al. Diffusion- Respiratory-triggered versus breath-hold diffusion-weighted
weighted imaging with dual-echo echo-planar imaging for MRI of liver lesions: comparison of image quality and ap-
better sensitivity to acute stroke. AJNR Am J Neuroradiol parent diffusion coefficient values. AJR Am J Roentgenol
2014;35(7):1293–1302. 2009;192(4):915–922.
16. Nakamura Y, Higaki T, Akiyama Y, et al. Diffusion-weighted 37. Chilla GS, Tan CH, Xu C, Poh CL. Diffusion weighted
MR imaging of non-complicated hepatic cysts: value of 3T magnetic resonance imaging and its recent trend: a survey.
computed diffusion-weighted imaging. Eur J Radiol Open Quant Imaging Med Surg 2015;5(3):407–422.
2016;3:138–144. 38. Andre JB, Bammer R. Advanced diffusion-weighted magnetic
17. Feuerlein S, Pauls S, Juchems MS, et al. Pitfalls in abdomi- resonance imaging techniques of the human spinal cord. Top
nal diffusion-weighted imaging: how predictive is restricted Magn Reson Imaging 2010;21(6):367–378.
water diffusion for malignancy. AJR Am J Roentgenol 39. Arlinghaus LR, Welch EB, Chakravarthy AB, et al. Motion
2009;193(4):1070–1076. correction in diffusion-weighted MRI of the breast at 3T. J
18. Koh DM, Collins DJ, Orton MR. Intravoxel incoherent mo- Magn Reson Imaging 2011;33(5):1063–1070.
tion in body diffusion-weighted MRI: reality and challenges. 40. deSouza NM, Orton M, Downey K, et al. Distortion cor-
AJR Am J Roentgenol 2011;196(6):1351–1361. rection of echo-planar diffusion-weighted images of uterine
19. Le Bihan D, Breton E, Lallemand D, Aubin ML, Vignaud cervix. J Magn Reson Imaging 2016;43(5):1218–1223.
J, Laval-Jeantet M. Separation of diffusion and perfusion 41. Hancu I, Lee SK, Hulsey K, et al. Distortion correction
in intravoxel incoherent motion MR imaging. Radiology in diffusion-weighted imaging of the breast: performance
1988;168(2):497–505. assessment of prospective, retrospective, and combined
20. Ueno Y, Takahashi S, Kitajima K, et al. Computed (prospective + retrospective) approaches. Magn Reson Med
diffusion-weighted imaging using 3-T magnetic reso- 2017;78(1):247–253.
1144 July-August 2018 radiographics.rsna.org
42. Mazaheri Y, Do RK, Shukla-Dave A, Deasy JO, Lu Y, Akin 46. Higaki T, Kaneda K, Tamaki T, Date N, Azemoto S. Non-
O. Motion correction of multi-b-value diffusion-weighted rigid image registration for medical diagnosis using free-form
imaging in the liver. Acad Radiol 2012;19(12):1573–1580. deformation with multiple grids. J Inst Image Electron Eng
43. Liu W, Liu X, Yang G, et al. Improving the correction of Japan 2007;37(3):286–292.
eddy current-induced distortion in diffusion-weighted images 47. Kapur T, Pieper S, Fedorov A, et al. Increasing the impact
by excluding signals from the cerebral spinal fluid. Comput of medical image computing using community-based open-
Med Imaging Graph 2012;36(7):542–551. access hackathons: the NA-MIC and 3D Slicer experience.
44. Maintz JB, Viergever MA. A survey of medical image registra- Med Image Anal 2016;33:176–180.
tion. Med Image Anal 1998;2(1):1–36. 48. Blackledge MD, Collins DJ, Koh DM, Leach MO. Signal
45. Zitová B, Flusser J. Image registration methods: a survey. to noise ratio of high b-value diffusion weighted images is
Image Vis Comput 2003;21(11):977–1000. improved using computed diffusion weighted imaging [abstr].
In: Proceedings of the Eighteenth Meeting of the International
Society for Magnetic Resonance in Medicine. Berkeley, Calif:
International Society for Magnetic Resonance in Medicine,
2010; 4707.
TM
This journal-based SA-CME activity has been approved for AMA PRA Category 1 Credit . See rsna.org/learning-center-rg.