Download as pdf or txt
Download as pdf or txt
You are on page 1of 3

Veterinary Clinical Pathology ISSN 0275-6382

B R I E F C O M M U N I C AT I O N

Evaluation of a portable glucose meter for use in cattle and sheep


Panagiotis D. Katsoulos1, Anastasios Minas2, Maria A. Karatzia3, Konstantinos Pourliotis3, Georgios
Christodoulopoulos1
1
Clinic of Medicine, School of Veterinary Medicine, University of Thessaly, Karditsa, Greece; 2Laboratory of Microbiology, Department of Medical
Laboratories, Technological Educational Institution of Larissa, Larissa, Greece; and 3Clinic of Farm Animals, School of Veterinary Medicine, Aristotle
University of Thessaloniki, Thessaloniki, Greece

Key Words Background: In farm animal practice, determination of blood glucose


Buiatrics, glucometer, hypoglycemia, ketosis, concentration under field conditions is often necessary.
point-of-care testing, ruminants Objective: As there is no portable glucose meter device developed for use
in farm animals, the analytical accuracy of a portable glucometer designed
Correspondence
for people was evaluated for its use in cattle and sheep.
Panagiotis D. Katsoulos, Kallipoleos 20,
57200, Lagadas, Thessaloniki, Greece
Methods: Blood samples from 90 cattle and 101 sheep were used in the
E-mail: pkatsoul@vet.uth.gr study. Glucose concentration was determined in whole blood immediately
after blood collection from the jugular vein with the One Touch Vita por-
DOI:10.1111/j.1939-165X.2011.00309.x table glucometer and in serum with an enzymatic colorimetric method.
The agreement between methods was assessed by Passing and Bablok re-
gression analysis. The precision and the accuracy of the measurements
were determined using the concordance correlation coefficient.
Results: There was a strong linear relationship between the glucose values
obtained using the portable glucometer and those obtained by the bench
method in both cattle and sheep. Precision was 95% for cattle and 88% for
sheep, whereas accuracy was 92% and 99%, respectively. The mean glu-
cose values obtained using the portable glucometer were significantly
lower by 8.3% in cattle and 3.2% in sheep than those determined by the
bench method.
Conclusion: The One Touch Vita portable glucometer can be used in
clinical practice to determine blood glucose concentrations in cattle
and sheep, but reference intervals (RI) must be corrected to allow for neg-
ative bias. Based on these equations the RI for blood glucose in cattle
and sheep using the portable glucometer were corrected to 1.84–4.17 and
2.41–4.35 mmol/L, respectively.

In farm animal practice it is often necessary to determine mals evaluated during 2009 at the Clinic of Medicine
blood glucose concentrations under field conditions. of the School of Veterinary Medicine of the University
Owing to the lower renal threshold for glucose in rumi- of Thessaly, were used in the present study following
nants, determination of blood glucose levels is regarded institutional guidelines for the care and use of animals.
as necessary before intravenous injection of a glucose Samples were collected directly into vacuum glass
solution.1 Furthermore, blood glucose determination can tubes (BD, Franklin Lakes, NJ, USA) without antico-
be used to detect neonatal hypoglycemia or for diagno- agulant by jugular vein puncture with an 18 G needle.
sing and differentiating types I and II ketosis.1 Use of a Glucose concentrations were determined in whole
portable glucometer would be beneficial in these cases. blood immediately after venipuncture using a One
As there is no portable glucose meter developed for use in Touch Vita portable glucometer (Lifescan, High
farm animals, we evaluated the analytical accuracy of a Wycombe, UK) and were expressed in mmol/L. Ac-
portable glucometer designed for measuring glucose in cording to the manufacturer, the glucometer uses a
people for use in cattle and sheep. glucose oxidase biosensor and is designed to determine
Blood samples from 90 cattle and 101 sheep, ob- glucose in human fresh capillary whole blood with
tained for various diagnostic investigations from ani- hematocrits that range from 30% to 55%.2 All samples

Vet Clin Pathol 40/2 (2011) 245–247 


c 2011 American Society for Veterinary Clinical Pathology 245
1939165x, 2011, 2, Downloaded from https://onlinelibrary.wiley.com/doi/10.1111/j.1939-165X.2011.00309.x by Ufrgs - Universidade Federal Do Rio Grande Do Sul, Wiley Online Library on [26/06/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
Portable glucometer for cattle and sheep Katsoulos et al

were determined to have hematocrits that fell within confidence intervals (CI) for cattle and sheep
this range. Before each measurement, a control solu-
tion (One Touch Vita control solution, Lifescan) was Cattle : y ¼0:965ð95%CI : 1:247  0:719Þ
tested. Serum was separated immediately after clotting þ 1:229ð95%CI : 1:147  1:320Þx
by low speed centrifugation at 1600 g for 10 minutes, Sheep : y ¼0:144 ð95%CI : 0:000  0:453Þ
transferred to plastic vials, and sent to the Laboratory
þ 0:947ð95%CI : 0:839  1:000Þx
in the Clinic of Medicine, School of Veterinary Medi-
cine, University of Thessaly for analysis of serum glu- Based on these equations the RI for blood glucose
cose concentrations. The same individual (P.D.K.) in cattle and sheep using the portable glucometer
performed all analyses using a commercial diagnostic were corrected to 1.84–4.17 and 2.41–4.35 mmol/L,
kit (Glucose Liquid-Fast, Zafiropoulos-Karavitis Ltd., respectively.
Athens, Greece) that utilizes the glucose oxidase- Analysis of Bland–Altman plots for cattle and
peroxidase (GOD-PAP) enzymatic colorimetric method3; sheep (Figure 1) indicated that the 2 methods were
the assay was performed within 30 minutes after blood linear and that the portable glucometer was adequate
collection. The kit was used according to the manufac- for determination of blood glucose in these species.
turer’s instructions: 0.02 mL of serum was mixed with Precision was 95% for cattle and 88% for sheep; such
2 mL of reagent and incubated for 5 minutes at 371C. high precision suggests that there is strong linear rela-
Absorbance of the mixture was measured at 510 nm tionship between the glucose values determined by the
using a spectrophotometer (UV-1601, Shimadzu, 2 methods. Accuracy represents a bias correction
Tokyo, Japan), and serum glucose concentration was
calculated in mmol/L based on absorbance of a stan-
dard solution (5.55 mmol/L). The reference intervals
(RI) for serum glucose established in our clinic using
this bench method were 2.28–4.18 mmol/L for cattle
and 2.39–4.44 mmol/L for sheep.
Analysis of the data were done using MEDCALC
9.2 software (MedCalc Software, Mariakerke,
Belgium). The t-test for paired samples was run
to evaluate the significance of differences between
glucose values determined by the reference method
and the portable glucometer for each animal species.
One-way analysis of variance was used to compare
the mean absolute relative deviations of the
glucose values determined by the glucometer among
3 groups: low (o RI for the bench method), middle
(within RI for the bench method), and high (4 RI for
the bench method) glucose concentrations in both
species. Post hoc multiple comparisons were made
using the Bonferroni test. A significance level of
P o.05 was used for all comparisons. The agreement
between methods was assessed using Passing and
Bablok regression analysis,4,5 and Bland–Altman plots
were created for both species. Precision and accuracy
of the measurements were tested using the concordance
correlation coefficient.6

Results and Discussion


Figure 1. Bland–Altman plots showing the difference between glucose
The regression analysis according to Passing and Ba-
concentrations obtained with the bench (reference) method and the One
blok revealed the following equations between the Touch Vita portable glucometer plotted against the mean of the 2 meth-
glucose values obtained from the portable glucometer ods. (A) Blood samples from cattle (n = 90). (B) Blood samples from sheep
(y) and those from the bench method (x) with 95% (n = 101).

246 Vet Clin Pathol 40/2 (2011) 245–247 


c 2011 American Society for Veterinary Clinical Pathology
1939165x, 2011, 2, Downloaded from https://onlinelibrary.wiley.com/doi/10.1111/j.1939-165X.2011.00309.x by Ufrgs - Universidade Federal Do Rio Grande Do Sul, Wiley Online Library on [26/06/2023]. See the Terms and Conditions (https://onlinelibrary.wiley.com/terms-and-conditions) on Wiley Online Library for rules of use; OA articles are governed by the applicable Creative Commons License
Katsoulos et al Portable glucometer for cattle and sheep

factor, which measures the deviation of the best-fit line and sheep in field conditions; however, results are un-
from the 451 line through the origin, and in the present derestimated. Thus, corrected RI should be used.
study was 92% for cattle and 99% for sheep, indicating
that the best-fit line almost fits the 451 line.
The mean glucose concentrations obtained from the Disclosure: The authors have indicated that they have
portable glucometer were significantly lower (P =.001 no affiliations or financial involvement with any orga-
for cattle, P =.008 for sheep) than those from the bench nization or entity with a financial interest in, or in
method in both cattle (mean  SEM: 2.98  0.10 financial competition with, the subject matter or
and 3.25  0.08 mmol/L for the portable glucometer materials discussed in this article.
and bench method, respectively, or 8.3%) and sheep
(mean  SEM: 3.06  0.07 and 3.16  0.08 mmol/L for References
the portable glucometer and bench method, respec-
tively, or 3.2%), indicating that, compared with the 1. Rollin F. Tools for a prompt cowside diagnosis: What
bench method, the portable glucometer underestimates can be implemented by the bovine practitioner?
blood glucose concentrations in cattle and sheep. As the Proceedings of the XXIV World Buiatrics Congress,
hematocrits of the animals used in the study (data not Nice, France, October 15–19, 2006; 322–336.
shown) were within the reference limits for hematocrit 2. OneTouch Vita Blood Glucose Monitoring System. User
suggested by the manufacturer,2 the exact cause for this guide 2009. Available at: http://www.lifescan.co.uk/sites/
underestimation is currently unknown and should be default/files/pdf/booklets/06607902A_Vita_User_Guide.
further investigated. Possible explanations include the pdf. Accessed March 21, 2011.
low precision of the portable glucometer at low blood 3. Barham D, Trider P. Improved colour reagent for
glucose concentrations.2 The glucometer was designed determination of blood glucose by oxidase system.
for use in people, who have higher blood glucose con- Analyst. 1972;97:142–145.
centrations than do cattle and sheep. This is confirmed 4. Passing A, Bablok W. A new biometrical procedure for
by the fact that the mean absolute relative deviation at testing the equality of measurements from two
low glucose concentrations was significantly greater different analytical methods. Application of linear
than at middle and high glucose concentrations in both regression procedures for method comparison studies
cattle (mean  SEM: 23  3%, 12  2%, and 8  2% for in clinical chemistry: part I. J Clin Chem Clin Biochem.
low, middle, and high glucose levels, respectively, 1983;21:709–720.
P =.001) and sheep (mean  SEM: 15  3%, 9  2%, 5. Passing A, Bablok W. Comparison of several regression
and 10  2% for low, middle, and high glucose levels, procedures for method comparison studies and
respectively, P =.008). determination of sample sizes. Application of linear
According to the American Diabetes Association,7 regression procedures for method comparison studies
the maximum bias (analytical and user error) com- in clinical chemistry: part II. J Clin Chem Clin Biochem.
1984;22:431–445.
pared with a reference method value should not ex-
ceed 10%; however, deviation of more than 10% is 6. Lin L-K. A concordance correlation coefficient to
observed in more than a third of the glucose values evaluate reproducibility. Biometrics. 1989;45:255–268.
when portable glucometers are used to measure glucose 7. American Diabetes Association.. Self-monitoring of
concentrations in human blood.8 Absolute relative de- blood glucose (consensus statement). Diabetes Care.
viations of more than 10% were observed in 46% of the 1996;19:62–66.
values for cattle and in 33% of the values for sheep. De- 8. Savoca R, Jaworek B, Huber AR. New ‘‘plasma
viations of more than 20% were seen in the 26% and referenced’’ POCT glucose monitoring systems – are
10% of obtained values for cattle and sheep, respec- they suitable for glucose monitoring and diagnosis of
tively. These biases are considered acceptable with the diabetes? Clin Chim Acta. 2006;372:199–201.
methodology used in the present study, given that glu- 9. Clinical and Laboratory Standards Institute. Evaluation
cose concentrations for each sample were determined of Precision Performance of Quantitative Measurement
only once with either the glucometer or the wet labora- Methods. CLSI document EP05-A2. Wayne, PA: CLSI; 2004.
tory method, despite CLSI recommendations.9,10 10. Clinical and Laboratory Standards Institute. User
Based on our results, it can be concluded that One Verification of Performance for Precision and Trueness;
Touch Vita portable glucometer is sufficiently accurate Approved Guideline. CLSI document EP15-A2. 2nd ed.
to determine blood glucose concentrations in cattle Wayne, PA, USA: CLSI; 2005.

Vet Clin Pathol 40/2 (2011) 245–247 


c 2011 American Society for Veterinary Clinical Pathology 247

You might also like