Professional Documents
Culture Documents
Cobas C 311 Analyzer: Operator's Manual Software Version 01-04
Cobas C 311 Analyzer: Operator's Manual Software Version 01-04
Operator’s Manual
Software Version 01-04
cobas c 311 analyzer
Document information
Edition notice Operator’s Manual for users of the cobas c 311 analyzer.
Every effort has been made to ensure that all the information contained in this
document is correct at the time of issue. However, Roche Diagnostics GmbH reserves
the right to make any changes necessary without notice as part of ongoing product
development.
Any customer modification to the instrument will render the warranty or service
agreement null and void.
Software updates may only be carried out by Roche service representatives.
Intended use The cobas c 311 analyzer is an automated, discrete clinical chemistry analyzer
intended for the in vitro quantitative/qualitative determination of analytes in body
fluids. It is important that the operator reads this manual thoroughly before using the
analyzer.
Roche Diagnostics
2 Operator’s Manual · Version 2.0
cobas c 311 analyzer
Instrument approvals The cobas c 311 analyzer meets the protection requirements laid down in IVD
Directive 98/79/EC. Furthermore, our instruments are manufactured and tested
according to the following international standards:
o IEC 61010-1: 2001
o IEC 61010-2-010: 2003
o IEC 61010-2-081: 2001
o IEC 61010-2-101: 2002
o UL 61010-1: 2001
o CAN/CSA C22.2 No. 61010-1-04
o EN 61326-2-6:2006
The Operator’s manual meets the requirements of European Standard EN 591.
Compliance is demonstrated by the following marks:
Contact addresses
Roche Diagnostics
Operator’s Manual · Version 2.0 3
cobas c 311 analyzer
Roche Diagnostics
4 Operator’s Manual · Version 2.0
cobas c 311 analyzer
Table of contents
Roche Diagnostics
Operator’s Manual · Version 2.0 5
cobas c 311 analyzer
Maintenance Part C
13 General maintenance
Overview C–7
Maintenance items C–8
Maintenance pipes C–10
Maintenance types C–19
Maintenance report C–23
List of maintenance items C–24
List of maintenance checks C–28
Troubleshooting Part D
15 Data alarms
Introduction D–7
Data alarm list D–8
Data alarms (ISE unit) D–10
Data alarms (photometric unit) D–19
Alarms for calibrations D–28
Data alarms for controls D–39
Data problems without alarm D–43
Rerun list D–47
16 Troubleshooting
General troubleshooting D–51
General instrument troubleshooting D–56
ISE troubleshooting D–59
Troubleshooting of the photometric unit D–66
Glossary Part E
Glossary E–3
Index Part F
Index F–3
Roche Diagnostics
6 Operator’s Manual · Version 2.0
cobas c 311 analyzer
Preface
Observe the instructions of the Operator’s Manual for safe operation of the system
o If the system is used in a manner not specified in this Operator’s Manual, the
protection provided by the system may be impaired.
o Keep this manual in a safe place to ensure that it is not damaged and remains available
for use.
o This Operator’s Manual must be easily accessible at all times.
The following documents are provided to assist you in finding desired information
quickly:
Operator’s Manual Contains information about safety, hardware components and operating the analyzer
as well as maintenance and troubleshooting. A table of contents at the beginning of
the manual as well as at the beginning of each chapter, and an index at the end of this
manual help you to find information quickly.
Online Help Contains a detailed description of the software of the cobas c 311 analyzer. In
addition to the software description, the whole Operator’s Manual is included in the
Online Help. This makes it possible to retrieve information from both Online Help
and Operator’s Manual using the search functions available for electronically stored
documents.
You cannot run the COBI CD on the cobas c 311 analyzer because you need Adobe
Acrobat Reader to view the COBI CD. Adobe Acrobat Reader is not installed on the control
unit and must not be installed.
Roche Diagnostics
Operator’s Manual · Version 2.0 7
cobas c 311 analyzer
The software of the cobas c 311 analyzer has a context-sensitive online help feature to
aid you in operating the instrument. “Context-sensitive” means that wherever you
are located within the cobas c 311 software, choosing the help feature displays help
text relating to that area of the software. The online help system offers a quick and
convenient way to find information, such as explanations of screens and dialog boxes
and how to perform particular processes.
F1 Help There are two ways to enter the online help system: Via the help icon in the bottom
left of the screen or by pressing F1 on the keyboard. The context-sensitive entry
displays information relating to your current location in the software.
Customer information
Customer training Contact your local Roche service representative for any questions or information
regarding cobas c 311 analyzer training.
Contact customer service Contact your local Roche service representative for further information regarding the
cobas c 311 analyzer service agreement.
Ordering information Replacement parts, consumable materials, reagents, calibrators and controls should
be ordered from Roche Diagnostics. When ordering, please use the Roche
Diagnostics catalog number and reference name for each item. Contact your local
Roche service representative for the detailed ordering list.
Roche Diagnostics
8 Operator’s Manual · Version 2.0
cobas c 311 analyzer
Visual cues are used to help you locate and interpret information in this manual
quickly. This section explains the formatting conventions used in this manual.
Tip
a Start of procedure
End of procedure
o List item
e Cross-reference
f Path to the active window or screen
Electrical and electronic equipment marked with this symbol are
covered by the European directive WEEE.
The symbol denotes that the equipment must not be disposed of in
the municipal waste system.
Abbreviation Definition
A
ANSI American National Standards Institute
ASCII American Standard Code for Information Interchange
AU Analytical Unit. In the analyzer software this abbreviation is
sometimes also used to distinguish the photometric unit (AU) from
the ISE unit (ISE).
C
Cfas Calibrator for automated systems
CLAS 2 Clinical Laboratory Automation System 2
COBI CD compendium of background information
CSA Canadian Standards Association
CSF cerebrospinal fluid
CV coefficient of variation (%)
D
DAT Drugs of Abuse Testing
dB(A) decibel weighted against the A-frequency response curve. This curve
approximates the audible range of the human ear.
DIL diluent
E
EC European Community
EMF electromotive force
EN European Standard
Roche Diagnostics
Operator’s Manual · Version 2.0 9
cobas c 311 analyzer
Abbreviation Definition
H
HbA1c Hemoglobin A1c, glycosylated hemoglobin
I
ICVC initial cassette volume check
IEC International Electrotechnical Commission
INC. incubator bath
IS internal standard for the ISE unit
ISE ion selective electrode (ISE unit)
IVD in vitro diagnostic
IVDD In vitro diagnostics directive
K
KCL potassium chloride, reference solution for ISE unit
kVA kilovolt-Ampere. Unit for expressing rating of AC electrical
machinery.
L
LIS laboratory information system
LLD liquid level detection
M
MSDS material safety data sheet
N
n/a not applicable
P
P photometric unit of the analyzer
Q
QC quality control
R
RCM reaction calculation mode
REF reference solution for ISE unit
S
SD standard deviation
SIP ISE sipper syringe
STAT short turn around testing
Std standard, commonly used for calibrator
T
TDM Therapeutic Drugs Monitoring
U
UDF Universal Disk Format
UL Underwriters Laboratories Inc.
Roche Diagnostics
10 Operator’s Manual · Version 2.0
System description A
Before you start working with the cobas c 311 analyzer, it is essential that the
warnings, cautions, and safety requirements contained in this manual, as well as the
explanations of the safety labels to be found on the analyzer are read and understood
by the user.
Roche Diagnostics
A-4 Operator’s Manual · Version 2.0
cobas c 311 analyzer 1 General safety information
Safety classifications
Safety classifications
The safety alert symbol by itself (without a signal word) is used to promote awareness to
hazards which are generic or to direct the reader to safety information provided elsewhere
in the document.
These symbols and signal words are used for specific hazards:
WARNING
Indicates a hazardous situation which, if not avoided, could result in death or serious
WARNING injury.
CAUTION
Indicates a hazardous situation which, if not avoided, could result in minor or moderate
CAUTION injury.
NOTICE NOTICE
Indicates a hazardous situation which, if not avoided, may result in property damage.
Tip
Indicates additional information on correct use of the analyzer or useful tips.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-5
1 General safety information cobas c 311 analyzer
Safety precautions
Safety precautions
Particular attention must be paid to the following safety precautions. If these safety
precautions are ignored, the operator may suffer serious or fatal injury. Each precaution is
important.
Operator qualification
Personal protective equipment o Be sure to wear appropriate protective equipment, including, but not limited to,
safety glasses with side shields, fluid resistant lab coat, and approved disposable
gloves.
o Wear a face shield if there is a chance of splash or splatter.
Accuracy/precision of measured An incorrect result may lead to an error in diagnosis, therefore posing danger to the
results patient.
o For proper use of the instrument, measure control samples and monitor the
instrument during operation.
o Do not use reagents, calibrators, or quality controls that have exceeded their
expiration dates. Observe the specified storage conditions. Otherwise, inaccurate
results may be obtained.
o For diagnostic purposes always assess the results in conjunction with the patient’s
medical history, clinical examination and other findings.
Roche Diagnostics
A-6 Operator’s Manual · Version 2.0
cobas c 311 analyzer 1 General safety information
Safety precautions
Operating conditions o Operation outside of the specified ranges may lead to incorrect results or
malfunction of the instrument (see Specifications on page A-69).
o Use the instrument indoors only and avoid heat and humidity.
o Make sure that the analyzer's ventilation openings remain unobstructed at all
times.
o Perform maintenance according to the specified intervals to maintain the
operating conditions of the instrument.
o Keep the Operator’s Manual in a safe place to ensure that it is not damaged and
remains available for use. This manual must be easily accessible at all times.
Approved parts Use of non-approved parts or devices may result in malfunction of the instrument
and may render the warranty null and void. Only use parts and devices approved by
Roche.
Third-party software Installation of any third-party software that is not approved by Roche may result in
incorrect behavior of the analyzer. Do not install any non-approved software.
Power interruption A power failure or momentary drop in voltage may damage the analyzer or lead to
data loss. Perform regularly backups of measurement results. Operate only with an
uninterruptible power supply. Do not switch off power while the PC accesses the
hard disk or diskette.
Electromagnetic fields Devices that emit electromagnetic waves may cause the instrument to malfunction.
Do not operate the following devices in the same room where the analyzer is
installed:
o Mobile phone
o Transceiver
o Cordless phone
o Other electrical devices that generate electromagnetic fields
Instrument unused for an If the instrument is not used for an extended period of time, the main circuit breaker
extended period of time switch must be set to OFF. Remove and refrigerate any remaining reagents. For
further information, call technical support.
Relocation and transportation Do not attempt to relocate or transport the analyzer. Leave relocation and
transportation to personnel trained or authorized by Roche.
e For information about disposal of the analyzer, see:
Disposal of the analyzer on page A-20
Roche Diagnostics
Operator’s Manual · Version 2.0 A-7
1 General safety information cobas c 311 analyzer
Safety summary
Safety summary
This safety summary contains the most important and general warning and caution
messages. Additionally, you will find specific safety information at the beginning of
Part Operation and Part Maintenance.
Warning messages
Electrical safety
Biohazardous materials
Roche Diagnostics
A-8 Operator’s Manual · Version 2.0
cobas c 311 analyzer 1 General safety information
Safety summary
Waste
Waste must be treated in accordance with the relevant laws and regulations. Any
substances contained in reagents, calibrators, and quality controls, which are legally
regulated for environmental protection, must be disposed of according to the
relevant water discharge facility regulations. For the legal regulations on water
discharge, please contact the reagent supplier.
Two kinds of liquid waste are discharged by the analyzer:
o Concentrated waste solution that contains highly concentrated reaction solution.
This waste must be treated as infectious waste as specified by the relevant
regulations.
o Dilute waste: A non-concentrated waste solution diluted with rinsing water from
cell wash or water from the incubation bath. When using NaOH-D for washing
the reaction cells, alkaline concentration is 0.1 to 1.0 mmol/L in terms of sodium
hydroxide equivalence.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-9
1 General safety information cobas c 311 analyzer
Safety summary
Caution messages
Mechanical safety
Roche Diagnostics
A-10 Operator’s Manual · Version 2.0
cobas c 311 analyzer 1 General safety information
Safety summary
Spillage
Roche Diagnostics
Operator’s Manual · Version 2.0 A-11
1 General safety information cobas c 311 analyzer
Safety summary
Data security
Unauthorized access and data loss due to malicious software and hacker attacks
Portable storage media can be infected with and transmit computer malware, which may
CAUTION be used to gain unauthorized access to data or cause unwanted changes to software.
The cobas c 311 analyzer is not protected against malicious software and hacker attacks.
The customers are responsible for IT security of their IT infrastructure and for protecting it
against malicious software and hacker attacks. Failure to do so may result in data loss or
render the cobas c 311 analyzer unusable.
Roche recommends the following precautions:
o Allow connection to authorized external devices only.
o Ensure that all external devices are protected by appropriate security software.
o Ensure that access to all external devices is protected by appropriate security
equipment. Roche strongly recommends the use of a cobas IT Firewall.
o Do not copy or install any software on the cobas c 311 analyzer unless it is part of the
system software or you are instructed to do so by a Roche service representative.
o If additional software is required, contact your Technical support to ensure validation of
the software in question.
o Do not use the USB ports to connect other storage devices unless you are instructed to
do so by official user documentation or a Roche service representative.
o Exercise utmost care when using external storage devices such as USB flash drives,
CDs, or DVDs. Do not use them on public or home computers while connecting to the
cobas c 311 analyzer.
o Keep all external storage devices in a secure place and ensure that they can be
accessed by authorized persons only.
Roche Diagnostics
A-12 Operator’s Manual · Version 2.0
cobas c 311 analyzer 1 General safety information
Safety summary
Notices
Moving parts
Roche Diagnostics
Operator’s Manual · Version 2.0 A-13
1 General safety information cobas c 311 analyzer
Safety labels of the system
Warning labels have been placed on the analyzer to draw your attention to areas of
potential hazard. The labels and their definitions are listed below according to their
location on the instrument.
The safety labels on the analyzer comply with the following standards: ANSI Z535,
IEC 61010, IEC 60417, or ISO 7000.
If the labels are damaged, they must be replaced by Roche service personnel. For
replacement labels, contact your local Roche representative.
Spillage warning
This label indicates that the instrument may be damaged if a spillage
occurs within the vicinity of this label.
Do not place liquids in this area.
Infection warning
This label indicates that touching the instrument mechanism can cause
infection or injury to your fingers or hands.
Do not open the top cover while the instrument is in operation. Stop every
mechanism, then open the top cover.
Protective equipment warning
This label indicates that there is a danger of coming into contact with
corrosive material within the vicinity of this label.
Wear appropriate protective equipment (such as goggles and gloves).
Warning
This label indicates that there is a danger of hazardous situations arising
within the vicinity of this label, which may result in death or serious injury.
Refer to the Operator's Manual for instructions on safe operation.
Biohazard warning
This label indicates that there are potential biohazards within the vicinity
of this label.
The relevant laboratory procedures on safe use must be observed.
Electrical warning
This label indicates that there is a danger of coming into contact with
electrical components when gaining access to parts of the analyzer marked
with this label.
Refer to the Operator's Manual for instructions on safe operation.
Hot surface warning
This label indicates that the area within the vicinity of this label may be
hot.
Do not touch this area as you may be burned.
Barcode label direction
This label indicates that the barcode label has to be on the right side when
inserting the cobas c pack or other cassette into the reagent compartment.
Do not turn the cobas c pack the other way around as severe malfunctions
may occur.
Roche Diagnostics
A-14 Operator’s Manual · Version 2.0
cobas c 311 analyzer 1 General safety information
Safety labels of the system
The following sections briefly describe the meaning of the safety labels in the context
of where they occur on the instrument.
e For detailed information about the safety labels on the instrument, see:
Front view on page A-16
Side view on page A-17
Top view on page A-18
Rear view on page A-19
In addition to safety labels on the instrument there are safety notes in the
corresponding parts of the Operator’s Manual and Online Help.
e For more information, see:
Part Operation on page B-1
Part Maintenance on page C-1
These safety notes give more detailed information about potentially hazardous
situations that may arise during daily operation or when carrying out maintenance
procedures.
When working with the analyzer, be sure to observe both the safety labels on the
instrument and the safety notes in the Operator’s Manual and Online Help.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-15
1 General safety information cobas c 311 analyzer
Safety labels of the system
Front view
F-1
F-1
F-2
F-2
F-3
F-3
F-4
F-4
F-5
F-5 F-6
F-4 F-7
F-5
Roche Diagnostics
A-16 Operator’s Manual · Version 2.0
cobas c 311 analyzer 1 General safety information
Safety labels of the system
Side view
S-1
S-1
Roche Diagnostics
Operator’s Manual · Version 2.0 A-17
1 General safety information cobas c 311 analyzer
Safety labels of the system
Top view
Roche Diagnostics
A-18 Operator’s Manual · Version 2.0
cobas c 311 analyzer 1 General safety information
Safety labels of the system
Rear view
R-1
R-1 R-2
R-2 R-3
R-3
Roche Diagnostics
Operator’s Manual · Version 2.0 A-19
1 General safety information cobas c 311 analyzer
Disposal of the analyzer
Roche Diagnostics
A-20 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Table of contents
This chapter provides a detailed description of the control unit, the sampling unit, the
components of the reaction disk and the reagent area as well as the ISE unit. The
sample containers and other system components are also described in this chapter.
Roche Diagnostics
A-22 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Overview
Overview
Main components The main components of the cobas c 311 analyzer are:
o Analytical unit: performs the measurements.
o Control unit: controls and monitors the complete analytical process, saves and
outputs measurement results and supports maintenance functions.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-23
2 Description of the analyzer cobas c 311 analyzer
Overview
The analyzer comprises a photometric unit and an ISE unit (for ion-selective
electrode [ISE] determinations).
A
E
F
C
Photometric unit
The photometric unit provides the analyzer with a flexible photometric method of
assaying up to 300 in vitro tests per hour on a wide range of analytes. The following
are the main components of the cobas c 311 analyzer:
o Sampling system
o Reagent system
o Reaction disk system
Sampling system The sampling system is composed of a sample disk, a sample pipetter (consisting of a
pipetter arm and the sample probe), a sample syringe, and a rinse station for internal
and external rinsing of the sample probe.
e See Sampling area components on page A-37.
Roche Diagnostics
A-24 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Overview
Reagent system The reagent system is composed of a refrigerated reagent compartment consisting of
two storage rings for cobas c packs, and a reagent pipetting system. The reagent
pipetting system consists of a reagent pipetter (consisting of a pipetter arm and the
reagent probe), cap piercing device, a reagent syringe, and a rinse station for internal
and external rinsing of the reagent probe.
e See Reagent area components on page A-44.
Reaction disk system The reaction disk system is composed of a reaction disk, immersed in a reaction bath,
an ultrasonic mixing unit, a photometric measuring system, and a cell rinse unit for
cleaning the reaction cells once test measurement is complete.
e See Reaction disk area components on page A-49.
ISE unit
Moreover, the cobas c 311 analyzer has an integrated ISE unit, which provides the
system with a potentiometric method for assaying sodium, potassium and chloride
samples. The ISE unit can process up to 150 samples per hour. The following are the
main components of the ISE unit:
o ISE measuring compartment with measuring cartridges for Cl-, K+, Na+ and
Reference cartridge
o ISE sipper mechanism
o ISE reagent compartment
e See ISE area components on page A-59.
Other Components
Other components of the analyzer are the following:
o Water supply
o Vacuum system
o System interface port
o Waste solution container
Roche Diagnostics
Operator’s Manual · Version 2.0 A-25
2 Description of the analyzer cobas c 311 analyzer
Overview
100 applications:
o 86 photometric applications
o + 3 ISE applications + 3 serum indices + 8 calculated tests
Table A-1 Basic specifications of the cobas c 311 analyzer
Roche Diagnostics
A-26 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Power switches
Power switches
There are two types of power switches: One power switch each for the analytical unit
and the control unit, and in addition the superordinate power switch (main circuit
breaker).
A B C D
A Power switch of the analytical unit D Main circuit breaker of the analytical unit
B Maintenance switch
C Power switch of the control unit
Roche Diagnostics
Operator’s Manual · Version 2.0 A-27
2 Description of the analyzer cobas c 311 analyzer
Control unit
Correct order for switching on Please note that there is a prescribed order for switching on and off. The following
and off order must be observed:
o Warning: Do not switch on the analyzer unless you have been trained by Roche
and unless you are familiar with the safety precautions.
o For switching on, the analytical unit has to be switched on before the control unit.
o For switching off (shutdown), the control unit has to be switched off first by
choosing Shutdown from the software. After the computer power supply has
turned off, switch off power for the analytical unit, monitor and printer.
Control unit
The control unit uses a graphic user interface to control all instrument functions. The
following are components of the control unit:
Components of the control unit are subject to change without prior notice.
Roche Diagnostics
A-28 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Control unit
Computer
The computer monitors the functions and operational modes of the analyzer.
A hard disk drive is used to store system operating parameters as well as data from
patient samples, controls, and calibrators.
A 3 1/2 inch floppy disk drive and a DVD-RW rewrite device are available for reading
and writing parameters and other information for backup purposes. The DVD drive
is also used for installation of software.
Only Roche technical support may install software onto the analyzer.
Touchscreen Monitor
The system is equipped with a 15" color monitor with touchscreen adapter. The
monitor is used:
o To display information
o To navigate through the software
o To initiate instrument functions
Use your finger to touch what you want to request or change directly on the screen.
Most of the items within the software can be accessed using the touchscreen. Touch
the item desired (for example, menu bar, list box, text box, button, etc.) to complete
your task. For example, to display the Data Review screen in the Workplace menu,
touch Workplace, then the Data Review tab.
When touching the screen, be sure to “tap”, not “press”. The tap must be of short duration.
Selecting Items To select a consecutive range of items in a list, press the <Shift> key and touch the
first item in the range. While continuing to press the <Shift> key, touch the last item
in the range. All items, including the first and last items touched in the range, are
highlighted. You may also touch the first item in the list and drag your finger to the
last item in the list.
To select multiple, nonconsecutive items, press the <Ctrl> key, then touch the desired
items.
Height adjusting lever To adjust the height of the monitor base, lift the lever.
Keyboard
A 101-key enhanced keyboard is used to navigate through the software and to enter
information.
Most items that can be accessed by means of the touchscreen can also be accessed by
means of the keyboard.
e For more information, see Shortcut keys on page B-32
Roche Diagnostics
Operator’s Manual · Version 2.0 A-29
2 Description of the analyzer cobas c 311 analyzer
Control unit
Mouse
Printer
Only Roche technical support may install and connect the printer.
The following components take part in the data communication within the analyzer
and its network environment:
o Connected analyzer
o cobas link
o Host computer
There are two devices to accomplish the data communication: A serial port for
bidirectional communication with the Host computer and a hub for network
connections. The hub is located on the PC stand.
e See System interface on page A-77.
Roche Diagnostics
A-30 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
cobas link
cobas link
Overview
cobas link platform The cobas link platform is the gateway for retrieving and distributing information -
such as instructions for use, value sheets, important notes, and test- and lot-specific
analyzer settings - from Roche TeleService-Net to cobas analyzers in the laboratory.
cobas link is an integral and mandatory part of the cobas modular platform
analyzers.
TeleService-Net (TSN) TeleService-Net is the technical infrastructure to provide cobas analyzers and
operators with important product information from Roche. TeleService-Net offers
several applications to manage and display data and information of remotely
connected instruments.
TeleService-Net
cobas link
data station
cobas analyzers
cobas link data station The cobas link data station is a dedicated desktop computer with keyboard, mouse,
monitor, and printer.
cobas e-library The cobas e-library (e-library) is the user interface for working with cobas link on the
cobas link data station.
The main application on cobas link for the operator is the cobas e-library, consisting
of e-package inserts and e-barcodes. This application is used for searching, reviewing,
and printing of e-package inserts.
There is a separate Operator’s Manual for the cobas e-library available.
e For more information on working with the e-library, refer to the cobas e-library
Operator’s Manual.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-31
2 Description of the analyzer cobas c 311 analyzer
cobas link
cobas link update process cobas link receives data from TSN about the applications, calibrators, and controls
used on cobas analyzers. This ensures that the most up-to-date product information
is always available in the laboratory. All information relevant for the analyzer can be
downloaded from cobas link.
cobas link is updated daily via an automatic download if connected to the internet or
a phone line. The cobas link data station connects to TSN, typically overnight,
depending on the customer's requirements (configured by Roche service engineers
during installation). While connected, all newly available data are downloaded from
the TSN to your cobas link data station.
If your laboratory does not have an online connection to TSN, the TSN data is
provided on a CD, distributed by your local Roche organization.
e For more information on working with CD, refer to the cobas e-library Operator’s
Manual.
Main functions
From the operator’s point of view the main functions of cobas link are:
Download
e-barcodes
Provide
e-package inserts
e-package inserts o The cobas e-library stores many documents (e-package inserts) such as the
instructions for use for the applications, information on controls, calibrators, and
announcements.
e See Using e-package inserts on page A-34
Backup function o Providing the possibility of important data backup for the cobas c 311 analyzer.
e See Using the backup function on page A-34
Roche Diagnostics
A-32 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
cobas link
This section explains the working principles of the main cobas link functions.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-33
2 Description of the analyzer cobas c 311 analyzer
cobas link
Roche Diagnostics
A-34 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
cobas link
Overview of the user interface The screen in Figure A-13 is for illustrative purposes only. It shows the New Entries
for an analyzer with clinical chemistry and immunological modules.
D
B
A Navigation bar has a tab for the e-library application. Links to other applications
can also appear in this bar. The e-library application opens by
default.
B Navigation menu provides links to the e-library application screens.
C Online Help accesses the context-sensitive online help. A quick and
convenient way to find information, such as explanations of
screens and dialog boxes and how to perform particular
processes.
D Screen header displays the path of your active screen. The active operator name
is displayed on the right hand side of the Screen header.
E Message List displays the list of bulletins, letters, package inserts, value sheets
and product announcements.
F Footer explains the icons used in the document and version columns.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-35
2 Description of the analyzer cobas c 311 analyzer
cobas link
o Check the New Entries screen on a daily basis to make sure that you receive all
important information from Roche.
o Whenever you load a new application, make sure that you always add it to your
Preferences. Only the items you have selected in Preferences will be displayed on
the New Entries screen.
o If the cobas link data station is not connected to TeleService-Net, you have to update
regularly via CD (Update e-library screen).
e For more information on working with the e-library, refer to the cobas e-library
Operator’s Manual.
Roche Diagnostics
A-36 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Sampling area components
F
A
D
G
Roche Diagnostics
Operator’s Manual · Version 2.0 A-37
2 Description of the analyzer cobas c 311 analyzer
Sampling area components
Sample disk
The sample disk is located at the front on the left side. The rotor consists of two
concentrical rings with a total number of 108 positions for samples.
It is possible to reserve some positions for STAT samples. In addition, the sample
disk can hold different sample types, QCs and calibrators. When working without
barcode the positions for auto QC and calibrators must be defined by means of the
software. The position W1 is reserved for tubes with ISE cleaning solution/SysClean
and the position W2 for ISE Activator.
For easier adding and removing of samples, the sample disk could be taken out of the
analyzer. Make sure that the shield pipe is lifted up first.
Access Sample Disk lamp The green Access Sample Disk lamp displays the following statuses:
STAT positions
Use the STAT positions to start analysis of any samples directly, by-passing all
samples in the sample disk. The number of reserved positions for STAT samples is
definable (on the inner ring). A sample loaded in a STAT position will be processed
with higher priority than other samples on the sample disk.
Sample detector The sample detectors detect if there is a sample present in a sample position.
Roche Diagnostics
A-38 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Sampling area components
Barcode reader The barcode readers scan the barcode on each sample.
B
C
A Barcode reader (long slot) for outer ring C Barcode reader and sample detector for
B Sample detector (short slot) for outer ring inner ring
Figure A-15 Barcode readers and sample detectors in sample disk area
Barcode scan range The barcode label must be positioned so that it can be scanned through the opening
in the sample disk (see Figure A-16 on page A-39).
A
max. 88 mm
Figure A-16 Sample tubes on the outer ring with their barcode facing outwards
Ensure that the sample barcodes are facing the right way so the two barcode readers can
scan them. Samples placed in the inner ring must have their barcodes facing inwards,
samples placed in the outer ring must have their barcodes facing outwards.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-39
2 Description of the analyzer cobas c 311 analyzer
Sampling area components
The sample pipetting system is composed of a sample pipetter and a sample syringe.
After each pipetting, the sample probe is rinsed by the sample probe rinse station.
Sample pipetter
The sample pipetter consists of the pipetter arm and the sample probe.
When a sample tube is in the sampling position, the barcode is read three times for
safety reasons. Then the pipetter transports sample liquid from a sample tube to a
reaction cell. When aspirating, liquid level detection (LLD) is accomplished by a
highly sensitive capacitance measurement, as well as a clot detection by means of
pressure measurements.
To protect the probe against electrostatic noise, which would interfere with the
capacitance measurement, a shield pipe is mounted over the sampling position.
After sample has been aspirated, the probe is raised from the sample and is moved to
the reaction disk. The sample probe arm lowers the probe into the reaction cell at the
sample dispense position. Sample is dispensed while the beveled sample probe tip is
in contact with the bottom of the reaction cell. This ensures that a precise volume of
sample is deposited into the bottom of the cell even when using a low dispense
volume. The sample probe is spring-mounted on the arm to avoid damage to the
probe or reaction cell.
Roche Diagnostics
A-40 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Sampling area components
The rinse station is the home position for the sample probe when the pipetter is in
Standby.
In addition to the rinsing between pipetting, the sample probe is washed with Sample
Cleaner 1 and/or 2 from the sample probe detergent bottles:
o The daily and weekly maintenance actions include the washing of the sample
probe.
o Special washes can be programmed to prevent carryover.
e For more information, see:
Auxiliary reagents and cleaning solutions on page A-65
Daily maintenance on page C-46
Weekly maintenance on page C-58
Carryover evasion on page B-137
Sample syringe
The sample pipetter is connected by tubing to the sample syringe, which controls the
pipetting action.
A C
B
Figure A-18 Position of the sample syringe (inside the left front door)
The sample syringe, which is located behind the left front door of the analyzer, is
filled with degassed and deionized water. The syringe uses positive displacement to
aspirate and dispense samples by means of the sample probes.
The syringe motor retracts the plunger within the chamber of the syringe, and sample
is aspirated into the tip of the sample probe. The pipetter arm moves the sample
probe to the reaction disk. The sample probe lowers into the reaction cell and the
syringe motor reverses to dispense the sample. The pipetter arm lifts the sample
probe from the reaction cell and moves it to the sample probe rinse station.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-41
2 Description of the analyzer cobas c 311 analyzer
Sampling area components
Categories of samples
The cobas c 311 analyzer distinguishes the following categories of samples:
o Patient samples
O Routine samples
O STAT samples
O Rerun samples
o Calibrators
o Quality controls (QCs)
Sample containers
The following kinds of containers can be loaded onto the sample disk:
o Sample cups
O Hitachi standard cups
O Hitachi micro cups(a)
o Sample tubes (13 mm or 16 mm in diameter and 75 mm or 100 mm in length)
o In addition, non-standard tubes and false-bottom tubes can also be used.
G
H
A
B I
C J
D
E
F
Roche Diagnostics
A-42 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Sampling area components
Sample cups can be inserted into 16-millimeter sample tubes (cup on tube) or they
can be used without tubes.
In this document, "container" is used as a generic term for all kinds of cups and tubes.
e For sample container specifications, see
Sample cups and tubes on page A-77
Sample identification When operating in barcode mode, each sample’s barcode label is scanned by the
barcode reader of the sample disk. The sample barcode label provides the sample ID,
which is used for sample identification and test selection purposes.
e For specifications of barcodes and barcode labels, see
Barcode specifications on page A-78
Specifications of barcode labels on page A-79
The sample ID number provides tracking for the system. To track samples by means
of the software, select Sample Tracking from the System Overview screen.
When working with barcoded samples, only the STAT positions must be defined in
the software.
When working with Patient IDs, positions have to be assigned in the software for
routine and STAT samples for all sample types.
Non-barcode mode When operating in non-barcode mode, samples are identified by a sequence number
and their position in the sample disk. This assignment has to be done on the
Workplace > Test Selection screen.
When working with non-barcoded samples, define the position of each sample
category (controls, calibrators, and STAT positions) in the software.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-43
2 Description of the analyzer cobas c 311 analyzer
Reagent area components
D
A
B
C
F
A Rinse station and drying cylinder D Cutouts for the reagent probe
B Piercer compartment E Reagent pipetter
C Reagent compartment (with cutout) F Gate to the reagent loading station
G Barcode reader
Reagent compartment
Roche Diagnostics
A-44 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Reagent area components
The reagent pipetting system is composed of a reagent pipetter and a reagent syringe.
After each pipetting, the reagent probe is rinsed by the reagent probe rinse station.
B C D
Reagent pipetter
The reagent pipetter transports the reagents from the reagent compartment to the
reaction disk. The pipetter is mounted on an independent x-y-motion mechanism.
Before each pipetting, the reagent probe is externally and internally rinsed with
deionized water and dried. After reagent has been aspirated, the probe moves from
the reagent compartment to the reaction disk. There, the reagent volume is dispensed
into a reaction cell containing the sample. The mechanical cycle of the cobas c 311
analyzer allows for three different reagent timings: R1, R2, and R3.
The reagent probe is equipped with a pressure sensor and a liquid level detector
(LLD). When a new cobas c pack is loaded, the pressure sensor performs an initial
volume check—also referred to as initial cassette volume check (ICVC)—for each of
the bottles in a cobas c pack, if this option has been selected. The liquid level detector
is used for measuring the volume of the ISE reagents (ISE Dil. and ISE IS). Moreover,
the reagent pipetter picks up the piercer to open the cobas c pack by piercing the
tops.
e See Piercer on page A-48.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-45
2 Description of the analyzer cobas c 311 analyzer
Reagent area components
Reagent syringe
A C
B
Figure A-22 Position of the reagent syringe (inside the left front door)
The reagent syringe is located behind the left front door of the analyzer. It is filled
with degassed and deionized water, using positive displacement to aspirate and
dispense reagents.
Roche Diagnostics
A-46 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Reagent area components
Reagents for all Roche applications are provided in cobas c packs (COBAS
INTEGRA cassettes can also be used). These cassettes contain from one to three
specially designed reagent bottles and have barcode labels with detailed reagent and
test-related information.
The following components belong to the reagent management system:
o Reagent disk (inside the reagent compartment)
o Reagent loading station
o Barcode reader
o Piercer
Roche Diagnostics
Operator’s Manual · Version 2.0 A-47
2 Description of the analyzer cobas c 311 analyzer
Reagent area components
Barcode reader
The barcode reader scans data from the barcode label on the cobas c pack and checks
its integrity.
e For more information about the encoded data on the barcode label, see
Barcode label on page B-134.
After opening the gate of the reagent loading station, the operator scans the barcode
label and places the cobas c pack into the free position.
If the barcode of a cobas c pack or other cassette is unreadable, this cassette is not
accepted. The operator is requested to close the gate.
If the cobas c pack currently being scanned has not been on the analyzer before, the
analyzer registers the initial capacity, pierces the caps of the bottles and performs an
initial cassette volume check (ICVC)—if this option has been selected.
e For more information about the registration of cobas c packs, see
Reagent registration on page B-136.
Piercer
When a new cobas c pack is loaded into the reagent compartment, the reagent
pipetter picks up the piercer, which is kept near the rinse station. The piercer lowers
and pierces the caps of the reagent bottles in the cassette through the cutouts in the
compartment cover.
Roche Diagnostics
A-48 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Reaction disk area components
E
A
F
B
G
C
Reaction disk
The reaction disk of the cobas c 311 analyzer carries 66 semi-disposable plastic
reaction cells (cuvettes). These reaction cells are grouped together in six segments
with 11 cells each. All reaction cells are seated in a controlled-temperature bath. This
reaction bath maintains the cells at the required temperature of 37 0.1C. Therefore
it is also referred to as the incubator bath.
Cell covers Two cell covers protect the contents of the reaction cells from light. Some reactions
are light sensitive and intrusion of light could lead to faulty measurements. The cell
covers also serve to prevent contamination by reagents.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-49
2 Description of the analyzer cobas c 311 analyzer
Reaction disk area components
We recommend replacing the reaction cells once a month as they gradually deteriorate
over a prolonged period of use. Always carry out wash reaction parts and cell blank
measurement after replacing the reaction cells.
e See M8: Replacing reaction cells and cleaning incubator bath and drain filter on page C-68
C
F
Ultrasonic mixer
The ultrasonic mixing unit mixes the reagents within each of the reaction cells to
ensure a homogeneous distribution of reactants.
To avoid spillage, the water level of the incubator bath is checked before mixing by
calculating the volume. If the liquid level is too low or too high, an alarm (Mix.E) is
issued and mixing is not performed.
Contamination of the polished surface of the ultrasonic mixing unit will cause
inadequate mixing. It must be cleaned at least once every three months. The
ultrasonic output intensity is continually monitored. If the intensity falls below a
certain limit, an alarm (<Mix) is issued, and replacement of the ultrasonic mixing
unit is required. Contact your service representative for the replacement.
e See M15: Cleaning the ultrasonic mixer on page C-93.
Roche Diagnostics
A-50 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Reaction disk area components
Reaction bath
The circular reaction bath, positioned beneath the reaction disk, maintains the
reaction mixtures in the reaction cells at a temperature of 37C. Water in the reaction
bath is circulated by a pump through a refrigeration unit where it is cooled and then
onto the heater where it is heated, as necessary, to maintain the temperature
( 0.1C).
Two glass windows (inner and outer) are positioned in opposite walls of the reaction
bath. These windows permit light from the photometer lamp to pass through the
reaction bath water and through the reaction cells in the bath. The light beam
emerges from the outer window of the reaction bath and enters the instrument
photometer.
A liquid level sensor detects the water level of the bath. Deionized water is
automatically added to the reaction bath, as determined by the liquid level sensor, to
compensate for evaporation. This occurs even in Standby mode.
Top cover
The reaction disk area (as well as most of the sampling area and the reagent area) is
covered by the top cover.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-51
2 Description of the analyzer cobas c 311 analyzer
Reaction disk area components
Photometer
The cobas c 311 analyzer is equipped with a photometer to measure the absorbencies
of the reaction mixtures in the reaction cells.
The photometer lamp is located against the inner ring of the incubator bath beneath
the reaction disk. The detector is outside the incubator bath ring, near the cell rinse
unit.
e To locate the photometer detector, see Figure A-24 on page A-49.
Measurements are taken of all the 66 reaction cells, as the reaction disk is turning.
A B C D E F G H I J K L M
Photometer lamp The photometer lamp beneath the reaction disk, is encased in a constant-temperature
water jacket, which helps to maintain a constant energy output from the lamp, and
also extends the lamp life. Replace the photometer lamp if the photometer check
value at 340 nm exceeds 14000, or if lamp has been used for 750 hours or more.
e See M21: Replacing the photometer lamp on page C-113.
Light path The light from the photometer lamp passes through the following main components:
o Inner reaction bath window
o Reaction bath water
o Reaction cell and its contents
o Reaction bath water
o Outer reaction bath window
o ...and into the photometer
When the light beam enters the photometer, it strikes a diffraction grating. This
separates the light into its constituent wavelengths and reflects them onto a fixed
array of 12 photodiodes. Each photodiode is permanently positioned to detect light at
a different wavelength.
Roche Diagnostics
A-52 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Reaction disk area components
The computer uses the available assay parameter information to select the
wavelengths and the times at which a reaction mixture’s absorbance is read and
results calculated.
The instrument computer keeps track of which test is being performed in each
reaction cell. It also knows when each reaction cell passes through the photometer
light path. The computer uses this tracking ability and the programmed read
instructions to obtain test results.
The cell rinse unit is located to the left of the reaction disk. It cleans, rinses, and dries
the reaction cells once the chemical reaction of the reaction mixture has been
measured. To ensure cell integrity (optical characteristics), a photometric reading of
a cell containing water is performed during the cleaning process (cell blank) and
compared with the stored value from the weekly cell blank measurement.
e For information about the used rinsing reagents, see
Auxiliary reagents and cleaning solutions on page A-65.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-53
2 Description of the analyzer cobas c 311 analyzer
Reaction disk area components
Cell rinse sequence The following sequence reflects the order in which cleaning, rinsing, and cell
blanking of the reaction cells takes place.
Roche Diagnostics
A-54 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Reaction disk area components
The following figures and descriptions help to understand the flow of photometric
analysis:
o Figure A-29 on page A-55 shows the main functions and their positions on the
reaction disk area.
o Figure A-30 on page A-56 shows the main steps of photometric analysis flow.
o Table A-4 on page A-57 explains the operating principle and the photometric
analysis flow.
Figure A-29 Main functions and their positions on the reaction disk area
Roche Diagnostics
Operator’s Manual · Version 2.0 A-55
2 Description of the analyzer cobas c 311 analyzer
Reaction disk area components
1 Start
2 Resetting
5 Aspiration of water
6 Sampling
4 19
6 28
7 33
8 43
9 52
10 57
12 Automatic stop
Roche Diagnostics
A-56 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Reaction disk area components
Step Action
1–2 Upon start, the instrument resets the mechanisms, and then starts rinsing the
reaction cell.
In a single cycle (12 s), the reaction disk rotates by 65 reaction cells, divided into six
steps with temporary stops. The rotation steps are: +17 +24 +17 -12
+2 +17 reaction cells (+: forward, -: backward rotation, : temporary stop).
3 o Using the cell rinse nozzles A/B/C in the rinsing mechanism, reaction solution
is aspirated and the reaction cell is rinsed with deionized water.
o In the next 2 cycles, the reaction cell is washed with cell detergent 1 dispensed
from the cell rinse nozzle D and aspirated from nozzle E.
o In the next 2 cycles, the reaction cell is washed with cell detergent 2 dispensed
from the cell rinse nozzle F and aspirated from nozzle G.
o Deionized water is dispensed with nozzle H, it is aspirated with nozzle I.
o Using the nozzles J/K, the reaction cell is rinsed once again with deionized
water.
Note: The numbers of the cell rinse nozzles refer to Figure A-28 on page A-54.
4 Then a cell blank measurement is performed 3 times (M) after dispensing deionized
water with cell rinse nozzle L.
If the cell blank value differs by 0.1Abs or more from the stored value from the
weekly cell blank measurement, that cell will not be used for analysis.
5–6 When the reaction cell after rinsing and cell blank measurement advances to the
sampling position, sampling begins.
Sampling is carried out in the order starting from the sample that will take the
longest reaction time in order to shorten the time needed for completion of data
output.
7–9 Reagents R1, R2, and R3 are usable, and will be added at the determined time points
(0, 1.5 and 5 minutes).
After one of the reagents R1, R2, and R3 has been added, the liquid in the reaction
cell will be mixed at the mixing position by means of the ultrasonic mixer.
10 Sampling is carried out every 12 s (1 cycle), and measurement is performed once in
one full turn, that is 57 times in 10 minutes. On completion of measurement, the
concentration is calculated by use of the absorbance at the specified photometric
point.
11 The instrument dispenses the reaction solution with the rinsing mechanism and
conducts washing with cell detergents and rinsing with water.
12 Then the instrument goes into Standby.
Table A-4 Schedule of the photometric analysis
Roche Diagnostics
Operator’s Manual · Version 2.0 A-57
2 Description of the analyzer cobas c 311 analyzer
Reaction disk area components
Reaction monitoring
The reaction disk rotates one turn in 12 seconds and during this time, the absorbance
is measured and stored for all of the reaction cells that cross the optical path of the
photometer. These absorbance values are recorded in the reaction monitoring
system. For each reaction cell, cell blank (absorbance 0) is measured, and then
photometry is performed 57 times in 10 minutes. Therefore, photometry is less
affected by time lapse. Also, the concentration can be calculated by using desired
photometric points.
Absorbance
G
E
24 57
BC
7
A
1
H
F
D
0 5 10
Time / min.
Roche Diagnostics
A-58 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
ISE area components
The ISE unit provides an electronic method for measuring sodium, potassium, and
chloride ion activities in samples. It is designed to process up to 150 samples per hour
(450 tests/h).
The following components belong to the ISE unit:
o ISE sipper mechanism—for aspirating ISE IS and diluted sample from reaction
cells into the measuring system
o ISE measuring compartment
o Two ISE reagent compartments —containing bottles for ISE IS, ISE Dil., and
ISE Ref. (KCL)
The ISE analysis uses the reagent pipetter for pipetting ISE IS and ISE Dil. into
reaction cells.
A
D
E
B
Roche Diagnostics
Operator’s Manual · Version 2.0 A-59
2 Description of the analyzer cobas c 311 analyzer
ISE area components
The ISE sipper mechanism consists of a sipper nozzle and a sipper syringe.
D
A
Figure A-33 ISE sipper and ISE measuring system (covers removed)
Sipper nozzle The sipper nozzle aspirates either ISE IS solution or sample solution diluted with
diluent from a reaction cell into the ISE measuring flow path. The ISE IS solution and
the sample with the diluent are heated up to the measuring temperature (37C) in the
reaction bath.
The sipper nozzle is cleaned during daily maintenance.
Sipper syringe The sipper syringe is positioned behind the left front door of the cobas c 311
analyzer. It is the rightmost syringe (to the right of the sample syringe). The syringe
not only provides the negative pressure for the sipper, it also aspirates ISE Ref. from
its container and into the reference cartridge flow path when the pinch valve is
closed.
Roche Diagnostics
A-60 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
ISE area components
Measurement cartridges Three cartridges, each containing an ion-selective electrode, are directly connected to
form a flow path for the diluted sample and the ISE IS solutions.
Reference cartridge The reference cartridge contains the reference electrode. ISE Ref. is aspirated through
a separate flow path by the sipper syringe. A reference electrode potential is
registered.
Colors of the cartridges The electrode potentials are measured in the color-coded cartridges as follows:
The difference between the potentials at the reference electrode and the ion-selective
electrode equals the electromotive force (EMF). For every test, the EMF of both
ISE IS and diluted sample solution are measured for each sort of ions (Cl-, K+, and
Na+). From these EMF values the results are calculated.
The two ISE reagent compartments contain three reagent bottles: Internal standard
(ISE IS), diluent (ISE Dil.) and reference solution (ISE Ref.).
e To locate the ISE reagent bottles, see ISE area on page A-59 (E and F).
The front reagent compartment contains a ISE Ref. (KCl) bottle. The sipper syringe
aspirates ISE Ref. to reference electrode when the pinch valve of the ISE measuring
system is closed.
The rear reagent compartment for ISE IS and ISE Dil. is equipped with level sensors
for each reagent bottle.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-61
2 Description of the analyzer cobas c 311 analyzer
ISE area components
It usually takes three cycles to complete one ISE measurement, since internal
standard (ISE IS) is measured before and after sample. However, if only ISE is
continuously measured, two cycles are sufficient for one measurement, since the
ISE IS measured after sample can be used in common with the ISE IS to be measured
before the following sample. The final ISE measurement always requires three cycles.
Meas. 1
Sample and diluent Sample and diluent
ISE IS ISE IS
Meas. 2
Meas. 2
Sample and diluent
ISE IS
Table A-5 ISE measurement sequence
For every measurement, the analyzer measures three electromotive force values
(EMF); for chloride, potassium, and sodium, where EMF denotes the difference in
potential between the respective ion-selective electrode and reference electrode.
Roche Diagnostics
A-62 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
ISE area components
Finally, the results are calculated from the electromotive forces of ISE IS and diluted
sample. Table A-6 on page A-63 summarizes the flow of an ISE analysis:
Roche Diagnostics
Operator’s Manual · Version 2.0 A-63
2 Description of the analyzer cobas c 311 analyzer
Behind the front doors
The following figure provides a front view of the cobas c 311 analyzer. The legend
lists the main components behind the front doors of the analyzer.
F
A
B
C
G
D
H
E
Vacuum system
The vacuum system is located at the front of the analyzer. It consists of a vacuum
pump, a vacuum tank, vacuum sensors, and connecting tubing. The vacuum system
aspirates reaction mixture waste from the reaction cells to the reaction waste
container and removes reaction cell rinse water from the analyzer through the main
drain line.
Water supply
The deionized water supply system consists of the water tank, located behind the left
front door, connecting tubing, and a series of electronic valves. Water is
automatically added to the water tank when necessary. Water from this source is
supplied directly to the cell rinse unit, to the rinse stations, and to the incubator bath.
Water for the cobas c 311 analyzer is to be deionized to a conductivity of 1 S/cm or
less. Deionized water will be consumed in about 12 L/h on average. The supply of
deionized water must have a capacity of 40 L/h or more.
Roche Diagnostics
A-64 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Behind the front doors
The sound volume adjustment knob allows adjustment of the sound volume level of
the alarm buzzer.
o Two bottles with detergents for the cell rinse process are located behind the left
front door of the analyzer:
O Cell Wash Solution I / NaOH-D (CellCln I)
O Cell Wash Solution II / Acid Wash (CellCln II)
e For more information, see:
To replace a cell detergent bottle on page B-141
To locate the bottles, see Figure A-34 on page A-64 (D and E).
o Two small detergent bottles (70 mL) are located close to the sample pipetter. They
are used for cleaning the sample probe.
O Position 1: Sample Cleaner 1
O Position 2: Sample Cleaner 2
e For more information, see:
To replace a sample probe detergent bottle on page B-142
To locate the bottles, see Figure A-14 on page A-37 (C).
o Three detergents are supplied in cobas c packs:
O NaOH-D/Basic Wash for reagent probe cleaning
O SMS/Acid Wash for reagent probe cleaning
O SCCS detergent (Special Cell Cleaning Solution)
o Hitergent is supplied in a cobas c pack. Hitergent is an additive to the reaction
bath to reduce surface tension and algal proliferation.
e For more information about those reagents, see:
Auxiliary reagents and detergents on page B-132
cobas c packs on page B-133.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-65
2 Description of the analyzer cobas c 311 analyzer
Rear view
Rear view
The Figure A-35 provides a rear view of the cobas c 311 analyzer. The legend lists the
main components on the back of the analyzer.
F
A
G
B
H
D I
Power supply (US only) There are two outlets (100 V AC, 5 A) inside the cover. One is for the PC (control
unit), and the other is for the monitor only. Use these only if the power cord is too
short to reach an external power source.
The connection of PC or monitor to outlets must be done by Roche technical support only.
Ethernet connector The Ethernet connector is used for communication between the analytical unit and
the control unit.
The supplied Ethernet cable must be used. Only trained Roche technical support personnel
(or similarly qualified personnel supervised by authorized service agents of Roche
Diagnostics GmbH) are qualified to connect the Ethernet cable.
Main circuit breaker The main circuit breaker is usually turned on for keeping the reagent cooling unit
active.
e See Power switches on page A-27.
Do not place any obstacles in front of the main circuit breaker. You must have
unobstructed access at all times so that you can immediately turn off the power in case of
emergency.
Roche Diagnostics
A-66 Operator’s Manual · Version 2.0
cobas c 311 analyzer 2 Description of the analyzer
Rear view
ISE drain port and waste The ISE drain port is used for discharging highly concentrated waste solution from
solution container the ISE unit into the waste solution container.
The waste solution container also contains the highly concentrated waste from the
reaction cells and is equipped with a liquid level sensor for issuing an alarm when it
becomes full.
Dilute waste line This line is used for discharging dilute waste solution.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-67
2 Description of the analyzer cobas c 311 analyzer
Rear view
Roche Diagnostics
A-68 Operator’s Manual · Version 2.0
cobas c 311 analyzer 3 Specifications
Table of contents
Specifications
This chapter gives an overview of the specifications of the cobas c 311 analyzer.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-69
3 Specifications cobas c 311 analyzer
Table of contents
Roche Diagnostics
A-70 Operator’s Manual · Version 2.0
cobas c 311 analyzer 3 Specifications
System specifications
System specifications
Dimensions
Operating conditions
The power distribution panel and the water supply and drainage facilities must be
available within 5 m of the analyzer.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-71
3 Specifications cobas c 311 analyzer
System specifications
Environmental conditions
Roche Diagnostics
A-72 Operator’s Manual · Version 2.0
cobas c 311 analyzer 3 Specifications
Control unit
Control unit
The following specifications apply to the control unit of the cobas c 311 analyzer.
Data storage Sample data (routine/STAT) 10 000 samples (routine, STAT and quality control
samples)
Reaction process data 10 000 tests (routine, STAT and quality control samples)
Quality control data Individual: 2500 samples, 100 tests/sample
Cumulative: 100 tests, 500 points/test
Roche Diagnostics
Operator’s Manual · Version 2.0 A-73
3 Specifications cobas c 311 analyzer
Analytical unit
Analytical unit
The following specifications apply to the analytical unit of the cobas c 311 analyzer.
Reaction system
Processing capacity varies depending on measuring conditions, test selection and system
configuration.
Sampling system
Roche Diagnostics
A-74 Operator’s Manual · Version 2.0
cobas c 311 analyzer 3 Specifications
Analytical unit
Reagent system
Photometric unit
Roche Diagnostics
Operator’s Manual · Version 2.0 A-75
3 Specifications cobas c 311 analyzer
Analytical unit
ISE unit
Roche Diagnostics
A-76 Operator’s Manual · Version 2.0
cobas c 311 analyzer 3 Specifications
Other specifications
Other specifications
In addition to specifications of the control unit and the analytical unit themselves,
this section provides an overview of specifications of the system interface, barcodes,
and sample containers.
System interface
e For information about the serial interface, refer to the Host Interface Manual.
All data processing equipment connected to the system must comply with the relevant
standards (IEC 950, UL 1950, CSA 22.2 No. 950).
Sample cups (both standard and micro cups) can be inserted into sample tubes with
16 mm diameter (cup on tube) or they can be used without tubes.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-77
3 Specifications cobas c 311 analyzer
Other specifications
Barcode specifications
Barcodes used with the cobas c 311 analyzer must be in compliance with one of the
following standards:
o NW7 (Codabar)
o Code 39
o ITF
o Code 128
Sample barcodes The following specifications apply to the various barcode types:
Roche Diagnostics
A-78 Operator’s Manual · Version 2.0
cobas c 311 analyzer 3 Specifications
Other specifications
Scan range for sample barcodes Stick the label in exact alignment with the centerline of the sample tube in order to
prevent scanning errors.
100 mm 75 mm
sample tube sample tube
Scan range
88 mm
Specifications of barcode labels Secure a margin of 5 mm or more on each edge of barcode label.
Roche Diagnostics
Operator’s Manual · Version 2.0 A-79
3 Specifications cobas c 311 analyzer
Other specifications
Roche Diagnostics
A-80 Operator’s Manual · Version 2.0
Operation B
Before starting to work with the analyzer, read the following safety messages
carefully. If you ignore these safety messages, you may suffer serious or fatal injury.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-3
4 Safety information for operation cobas c 311 analyzer
Roche Diagnostics
B-4 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
Table of contents
Software basics
Roche Diagnostics
Operator’s Manual · Version 2.0 B-5
5 Software basics cobas c 311 analyzer
Table of contents
Roche Diagnostics
B-6 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
General description of the user interface
The screen representations shown in this chapter and throughout this manual are for
illustrative purposes only. The screens do not necessarily show valid results.
The screen of the analyzer software is divided into four different areas. Following is
an example of a screen showing the different areas.
Fixed areas The three outer areas are fixed and accessible from every screen:
o The status line provides information about the current status.
o The Help button and the guidance prompt provide support that is specific to the
current menu and the location of the cursor.
o The global buttons provide access to specific functions.
e See Fixed areas on page B-8
See The help system of the instrument on page B-24
Menu area The menu area in the center changes according to the currently active menu or
function. Standard software elements like buttons, menu tabs, windows, etc. are used
in the menus. If you are not familiar with these elements, we recommend that you
read the section Software elements first.
e See Main menus on page B-11
See Software elements on page B-17
Roche Diagnostics
Operator’s Manual · Version 2.0 B-7
5 Software basics cobas c 311 analyzer
Fixed areas
Fixed areas
This section discusses the three areas that appear at the top, the lower left and the
right side of each screen.
Status line
A BC D E F
The status line displays the current status of the units, operational mode, operator ID,
date and time, and the time remaining when performing maintenance.
Status icons The three status icons on the left display the current statuses of the units:
o Host: external system. This icon is only visible if the host communication is
enabled in the Start Conditions window.
o Core unit
o Analytical unit (AU)
Roche Diagnostics
B-8 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
Fixed areas
Status colors The status colors represent the operational mode and additional information of each
unit or component. The following describes the status colors from the highest to the
lowest priority:
Color Meaning
Red Indicates an instrument alarm at the Stop, S.Stop or E.Stop level
or that a reagent is empty and there is no second cobas c pack
placed on this analyzer.
Yellow Indicates an instrument alarm at a caution level or that the
number of remaining tests is below the yellow alarm level
threshold.(a)
Purple Indicates that the number of tests remaining for a diluent, test
reagent or detergent volume is less than the daily requirement(a) -
only applicable if the Preventive Action check box has been
selected.
Light blue Indicates a status other than Standby (e.g., Operation or
Maintenance).
Light green Indicates Standby mode.
Table B-1 Color scheme of the status icons for each unit
(a) Specified on the Utility > System (Page 2/5) > Reagent Level Check window.
Screen accessibility
Some screens are not accessible with the operator level logon, or can be viewed with a
supervisor level logon but not edited. Some fields are viewed only at the administrator level
logon.
The Help button and the guidance prompt are shown in the lower left of each screen.
A B
Guidance prompt In this area, the software prompts you as to what kind of information to type and
what format to use when typing that information. The guidance prompt may also tell
what action to take, such as touching a button or pressing a key. For example, “Touch
screen or press <Enter>”. Each prompt is specific to the location of the cursor.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-9
5 Software basics cobas c 311 analyzer
Fixed areas
Global buttons
Use the global buttons to display screens that are used for specific functions. The
global buttons are accessible from every screen and appear on the right side of the
screen.
The Stop, S. Stop and Start buttons are also the system control buttons. The global
buttons can be selected by touchscreen, mouse or keyboard.
e For information about operation with the keyboard, see
Shortcut keys on page B-32.
Roche Diagnostics
B-10 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
Main menus
Main menus
The graphical user interface used by the control unit consists of the System Overview
screen and 5 main menus, Workplace, Reagent, Calibration, QC and Utility. Each
menu consists of two or more screens. Through these screens all instrument
functions are controlled.
Access to some screens, especially those of the Utility menu, is restricted by the
access level of the user. For this reason, the screen may not appear exactly as shown
here.
Standard software elements like buttons, menu tabs, windows, etc. are used in the
menus. If you are not familiar with these elements, we recommend that you read the
section Software elements first.
e See Software elements on page B-17
System Overview
The System Overview screen has a central role within the cobas c 311 analyzer
software. This screen provides an overview of the whole system at any given time. It
can be used to prepare the analyzer for daily routine operation.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-11
5 Software basics cobas c 311 analyzer
Main menus
Workplace
The Workplace menu consists of 2 screens: Test Selection and Data Review.
The Test Selection screen is used to perform the following:
o To make test selections
o To specify patient demographics
o To assign patient ID numbers and disk positions to samples
The Data Review screen is used to perform the following:
o To review data
o To backup data
o To edit data
o To delete data
o To send data to the Host
Roche Diagnostics
B-12 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
Main menus
Reagent
The Reagent menu consists of 2 screens: Setting and Status. These screens are used
to view detailed reagent information and to load and unload reagents.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-13
5 Software basics cobas c 311 analyzer
Main menus
Calibration
The Calibration menu consists of 3 screens: Status, Calibrator and Install. These
screens are used to request calibrations, define calibrators and view the reaction
curves for calibrators installed on the system. The calibration mode for the next start
up calibration or for an intermediate calibration can be determined from here. If an
intermediate calibration is necessary, it can be requested by test and calibration type.
Roche Diagnostics
B-14 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
Main menus
QC
Roche Diagnostics
Operator’s Manual · Version 2.0 B-15
5 Software basics cobas c 311 analyzer
Main menus
Utility
A Page button
Page buttons The Utility > System screen displays page buttons that can be used to display
additional buttons to display new screens. These page buttons are numbered (X/5), to
indicate which page you are on.
The page buttons only change the displayed buttons and not the entire appearance of the
screen.
Roche Diagnostics
B-16 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
Software elements
Software elements
These elements are used to navigate through the software, to select a choice or
function, and to type in information.
A
B
C
A A tab
B Menu tabs
C Screen tabs (for submenus)
D A screen (the Utility > System screen)
Menus The menu tabs are used to select screens and windows that are grouped together and
perform related tasks. The menu tabs are accessible from every screen, except the
global screens. This makes the software navigation within the menus and between
menus very easy.
The menu tabs are only for the 5 menus: Workplace, Reagent, Calibration, QC and
Utility. The menu tabs can be selected by touchscreen, mouse or keyboard. The
selected menu is colored blue.
Screens Each menu contains of two or more screens. Screens are frequently also called
submenus. The example asks you to select the Utility menu and therein the System
screen.
Example: Choose the Utility > System screen.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-17
5 Software basics cobas c 311 analyzer
Software elements
Tabs Tab is the common name for a special kind of button, with which you can switch
between menus.
Menus
Screens
A Tabs(a)
Figure B-12 Example of a tab and the hierarchical structure of menus, screens and tabs
(a) Tabs appear only in the Utility > Application screen, in the Workplace > Test selection screen, in a few
windows and a few global menus.
In this document, the term tab is mostly used as a generic term for subscreens or
subwindows. Subscreens appear in two screens (see Figure B-12) and a few windows
and global menus use subwindows (see Figure B-13, C).
Example: Choose the Calib. tab on the Utility > Application screen.
This is equal to: Choose the Utility > Application > Calib. screen.
Roche Diagnostics
B-18 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
Software elements
Buttons
Buttons are used for execution of functions, confirmation of entries and selections,
and displaying pop-up windows. The available buttons depend on the menu or
screen that is active.
B D
Black triangle The black triangle in the upper right corner of a button indicates that touching this
button displays another window (see Figure B-13, B).
The black triangle in the lower left corner of a button indicates that touching this
button performs the specified function and closes the window (see Figure B-13, D).
Text color on buttons If the text on the button is black, the button is available in the current mode. If the
text on the button is gray, the button is not available in the current mode.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-19
5 Software basics cobas c 311 analyzer
Software elements
Standard buttons
The following buttons are used throughout the software. These buttons are explained
below and will not be explained again with each individual screen.
OK Touch this button to accept the changes and entries made on a window and to close
the window.
Cancel Touch this button to close the window without saving the changes and entries made
on a window.
Yes Touch this button to accept the changes made on a window, execute the action of the
window (e.g., Delete) and close the window.
No Touch this button to close the window without saving the changes made on a window
or without executing the function of the window (e.g., Delete).
Save Touch this button to save any changes made to the screen/window currently
displayed. When changes have been made, this button turns yellow.
Update Touch this button to update any changes made to the screen/window currently
displayed. When changes have been made, this button turns yellow.
Execute Touch this button to accept the changes made on a window, to execute the action of
the window and to close the window.
Roche Diagnostics
B-20 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
Software elements
List boxes
List boxes display a list of choices. If there are more choices than can fit in the box, a
scroll bar is displayed. List box is a generic term for any type of dialog box option
containing a list of items the user can select.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-21
5 Software basics cobas c 311 analyzer
Software elements
Text boxes
Text boxes are used to type information. When you move to an empty text box, the
box is highlighted.
Check boxes
Check boxes are used to select a particular function. Sometimes when an option is
selected, additional fields become available. Multiple check boxes can be selected in
the same area of the screen.
es
Roche Diagnostics
B-22 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
Software elements
Options
An option is selected when a black circle is displayed within the white circle for the
option. Only one option can be selected at a time.
1. Touchscreen: Move to the option by touching it on the screen. The option
displays a black circle within the white circle when you touch the option.
2. Keyboard: Move to the option by pressing the <Tab> key. If multiple options are
available, the <Up Arrow> and <Down Arrow> keys move to the different
options.
Windows
Confirmation Window Many functions require confirmation prior to their execution (e.g., Delete).
Confirmation windows are used to confirm these functions. Touch Yes to confirm
the function and to close the confirmation window, or touch No to close the
confirmation window without carrying out the function.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-23
5 Software basics cobas c 311 analyzer
The help system of the instrument
The software of the cobas c 311 instrument has a context-sensitive online help system
to support you in operating the instrument.
Context-sensitive means that the help system always displays information related to
the current window or software element. Therefore, the help system offers a quick
and convenient way to find information.
A
A F1 Help button
About F1 Help
Roche Diagnostics
B-24 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
The help system of the instrument
Figure B-20 Area of the F1 Help window (example screenshot from the English version)
Button bar
The buttons at the top of the help window provide the following functions:
Hide / Show Allows you to show or hide the navigation pane. This feature
provides more space in the topic area, which may be desired for
displaying large figures.
Previous / Next Use these buttons to open the topic above or below the current
topic. If there is no topic to view above, the Previous button is
unavailable.
Back / Forward Use these buttons to backtrack through the topics you have
previously reviewed in the order in which you viewed them. The
record of topics viewed is deleted each time you exit the
F1 Help.
Refresh This is a standard function of Internet browsers for refreshing
the screen. You do not need this button to display the F1 Help.
Print Prints the topic currently displayed in the topic area.
Help Directs you to information about the help system (help on help).
Roche Diagnostics
Operator’s Manual · Version 2.0 B-25
5 Software basics cobas c 311 analyzer
The help system of the instrument
Navigation pane
The navigation pane appears to the left of the help window. There are four tabs
available at the top of the pane: Contents, Index, Search, and Favorites.
o The Contents tab uses book and page symbols to display the contents of the
F1 Help and remains synchronized with the topic displayed in the topic area.
o The Index tab provides access to the help index.
o The Search tab provides access to a full-text search feature.
o The Favorites tab allows organizing important topics with your personal
bookmarks.
Figure B-21 Areas of the F1 Help window (example screenshot from the English version)
e For information on the Contents, Index, Search, and Favorites tabs, see:
Searching for information in the help system on page B-27
Using favorites on page B-30.
Topic area
Help topics appear in the area to the right of the navigation pane. The topic area
comprises the description related to the topic selected in the navigation pane.
If the topic extends beyond the window, scrollbars allow you to view the rest of the
topic.
Related topics All subtopics to the current topic are listed at the end of each topic.
Alternatively, you can navigate to subtopics using the Next button or using the
Contents tab.
Links Cross-references to associated topics are indicated by blue text. When you hover over
a link, the mouse pointer changes to a hand icon.
Roche Diagnostics
B-26 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
The help system of the instrument
There are various methods of finding the desired information in the F1 Help. Select
your preferred method.
o The Contents tab displays the structure and contents of the cobas c 311 help
system allowing you to navigate to the desired location.
o The Index tab displays a list of predefined keywords based on the topic titles and
words or phrases in the help system.
o The Search tab allows you to perform a full-text search in the help system.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-27
5 Software basics cobas c 311 analyzer
The help system of the instrument
Figure B-22 Search tab (example screenshot from the English version)
Roche Diagnostics
B-28 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
The help system of the instrument
Recommendations for better The number of search results is displayed above the result list (Found: xx).
search results
Problem Recommendation
Too many search results 1. Select the Search titles only check box.
2. Deselect the Match similar words check box.
3. Select the Search previous results check box and enter a
new search term(s) that is associated with your desired
information.
4. Connect two or more search terms with AND, OR,
NEAR, or NOT (see arrow button on the right of the text
box).
The desired information is 1. Select the Match similar words check box.
not contained in the search 2. Deselect the Search previous results and the Search
results titles only check boxes.
3. Maybe you are using terms that are different from the
terms used in the help system. Use the Index tab for your
search.
Table B-3 Recommendations
Search options The following describes the search options for the F1 Help:
AND Use this option to join words or phrases when both (or all) terms
must appear in the search results.
OR Use this option to join synonymous or related terms when either
(or both) of the terms must appear in the search results.
NEAR Use this option to find words within 10 words of each other.
NOT Use this option to exclude a particular word or combination of
words from your search results.
Search previous results Select this check box to limit the search results and search
through your results list from previous search.
Deselect this check box if you want to search through all of the
files in the online help.
Match similar words Select this check box to find similar words and minor
grammatical variations for the phrase you search.
Search titles only Select this check box to search titles only.
Deselect this check box for a full text search.
Table B-4 Search options
Roche Diagnostics
Operator’s Manual · Version 2.0 B-29
5 Software basics cobas c 311 analyzer
The help system of the instrument
Using favorites
You can add a favorite, also called bookmark, to online help topics that you want to
visit again. The favorites work similar to the favorites in your web browser.
a To add a favorite
1 In the F1 Help, search for or navigate to the desired topic.
Figure B-23 Favorites tab (example screenshot from the English version)
a To rename a favorite
1 In the Favorites tab, select the desired favorite.
2 Right-click the favorite name and select Rename.
3 Type the new name and press <Enter>.
Roche Diagnostics
B-30 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
The help system of the instrument
You may find it useful to print a topic from the cobas c 311 online help.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-31
5 Software basics cobas c 311 analyzer
Shortcut keys
Shortcut keys
All functions can be initiated via the keyboard or the touchscreen for selection of
screen items.
All special keys and their functions are described below.
Shortcut Description
<F1> Use this key to display the online help.
<F2> Use this key to display the Start Conditions screen.
<F3> Use this key to display the Sample Stop screen.
<F4> Use this key to display the Stop screen.
<F5> Use this key to display the Workplace menu. This key does not
work if you are in a global screen.
<F6> Use this key to display the Reagent menu. This key does not
work if you are in a global screen.
<F7> Use this key to display the Calibration menu. This key does
not work if you are in a global screen.
<F8> Use this key to display the QC menu. This key does not work if
you are in a global screen.
<F9> Use this key to display the Utility menu. This key does not
work if you are in a global screen.
<F10> Use this key to display the Sample Pause/Scan screen.
<F11> Use this key to display the Alarm screen.
<F12> Use this key to display the System Overview screen.
<Print/Print Screen> Use this key to display the Print screen.
Press <Shift>+<Print Screen> simultaneously to print the
current screen.
<Scroll Lock> Use this key to display the Cancel Print window.
<Pause/Break> Use this key to display the Shutdown screen with the options
Logoff, Sleep and Shutdown.
<Esc> Use this key to exit a window or global screen.
<Tab> Use this key to move to the next field in a window or screen. To
go from a field to the previous one, press <Shift>+<Tab>
simultaneously.
<Enter> Use this key to confirm an entry.
<Shift> Simultaneously pressing <Shift> and a character key generates
a capital letter or a special character, for example on an English
keyboard <Shift>+<,> generates the character “<”.
<Backspace> Use this key to delete a character to the left of the cursor.
<Space> Use this key to generate a space.
<Delete> Use this key to delete a character to the right of the cursor.
<Home> Use this key to place the cursor at the beginning of a list or text
box.
<End> Use this key to place the cursor at the end of a list.
<Page Up> Use this key to scroll upward in a list, one page at a time.
<Page Down> Use this key to scroll downward in a list, one page at a time.
Table B-5 Shortcut keys (Sheet 1 of 2)
Roche Diagnostics
B-32 Operator’s Manual · Version 2.0
cobas c 311 analyzer 5 Software basics
Shortcut keys
Shortcut Description
<Arrow> Use these keys to move the cursor to the right, left, upwards or
downwards within a text box.
<Caps Lock> Use this key to lock the letter keys into the upper case mode.
<Num Lock> Use this key to activate the numerical keypad for entering
numbers and mathematical operators such as +, -, /, *.
Table B-5 Shortcut keys (Sheet 2 of 2)
Roche Diagnostics
Operator’s Manual · Version 2.0 B-33
5 Software basics cobas c 311 analyzer
Shortcut keys
Roche Diagnostics
B-34 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Table of contents
Daily operation
This chapter provides a description of the everyday tasks which are required for
running the cobas c 311 analyzer. Common procedures that are performed as part of
the daily workflow are also described here.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-35
6 Daily operation cobas c 311 analyzer
Table of contents
Roche Diagnostics
B-36 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Overview
Overview
Daily operation Daily operation encompasses the routine tasks that are required to prepare and
maintain the analyzer, and to analyze samples. The layout of the System Overview
screen provides the operator with an intuitive guide to the tasks required for routine
operation.
The following diagram gives an overview of daily operation.
Starting inspection
Analyzer start-up
(Initialization and standby)
• Check Alarm button
Preroutine operation
• Preventive action
• Calibration and QC
Routine operation
• Routine or STAT sampling Rerun assigned
------------------------------------------------------
Results
Maintenance
Shutdown
Certain maintenance items must be performed every day before the start of analysis.
We recommend performing the Power ON pipe every day during the Power ON procedure.
e For detailed information see:
Daily maintenance on page C-46
Power ON pipe on page C-17
Roche Diagnostics
Operator’s Manual · Version 2.0 B-37
6 Daily operation cobas c 311 analyzer
Starting the analyzer
Before you can start routine operation, you must first prepare the analyzer for
operation.
Start-up inspection
Before start-up, it is important to check that all of the following conditions are met. If
a condition is not met, take the necessary remedial action.
Before performing the following actions, observe the following safety precautions:
o Infection by samples and associated materials on page A-8
o Infection by waste solution on page A-9
o Personal injury due to contact with moving parts on page A-10
o Skin inflammation or injury caused by reagents and other working solutions on
page A-9
During operation, always check for any abnormal sound, water leakages or other abnormal
condition. If a trouble occurs, take suitable safety measures according to the condition and
contact your Roche service representative.
Roche Diagnostics
B-38 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Starting the analyzer
A B
2 Switch on the computer of the control unit, the monitor, and the printer.
After initialization, the Logon screen is displayed.
3 Enter your operator ID and password to log on.
4 Choose OK to gain access to the software and begin operation.
When initialization has been completed, the system goes into Standby.
Sleep mode
Instead of switching off power after the working day, the analyzer can also be put into
sleep mode. Sleep mode provides the possibility of an automatic start-up on the next day.
Note that when the analyzer wakes up from sleep mode, the control unit performs a
shutdown to reset the PC operation software (Windows XP).
e For information about sleep mode and shutting down the analyzer, see:
Analyzer shutdown and sleep mode on page B-81.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-39
6 Daily operation cobas c 311 analyzer
Starting the analyzer
Instrument statuses
This section explains the most important operational statuses (modes). The status is
displayed in the upper left of the monitor.
Check the upper left of the monitor before starting a new analysis or a maintenance
item:
o Stand By: If Stand By is displayed with a white background, the analyzer is ready
to start a new analysis or a system-controlled maintenance item. The top cover
must be closed.
o Maintenance mode: If the status line is displayed with a yellow background, the
analyzer is ready for maintenance actions that are performed with the top cover
open.
e For information about the required status for each maintenance item, see:
Overview of maintenance schedules on page C-34
Status Explanation
Power Up Status while the instrument is loading programs, performing
self-checks etc.
Initialize Status while the instruments performs any kind of initialization
– after power up or before being operative.
Standby Status when the instrument does not perform any measurement
activities or automatic maintenance procedures – but only a
limited initialization or preparation is required.
Preparation Status while the instrument prepares any kind of reagent or
other liquid.
Operation Status while the instrument is pipetting, measuring or
calculating.
Sample Reception Mode After all test requests are completed, the instrument waits for
additional samples. New samples will be processed immediately
after pressing the Start button.
Table B-7 Instrument statuses (modes) (Sheet 1 of 2)
Roche Diagnostics
B-40 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Starting the analyzer
Status Explanation
Sampling Pause After pressing Pause on the Sample Pause / Scan window
during operation, the status changes to Sampling Pause.
Additional samples can be loaded onto the sample disk. The
analyzer continues operation after pressing the Start button.
S.Stop After pressing the S.Stop button, the status changes to Sampling
Stop and then to Standby. Already pipetted samples will be
completed before sample processing is stopped.
E.Stop Status indicating that the system has performed an emergency
stop due to hardware failure or because any of the safety devices
requested an emergency stop. The system requires either a
complete power off or at least a reset (Utility > Maintenance) to
resume normal operation.
Standby with yellow Maintenance mode: Status for all actions that are performed
background with the top cover open. The low voltage power to the pipetters
and the reaction disk is switched off by means of the
maintenance switch without having to go into complete
shutdown.
e See: Maintenance mode on page B-42
Shutdown status The analyzer and control unit are disconnected and the
operation power switch is switched off. However, power for
keeping the reagents cool is still supplied.
Table B-7 Instrument statuses (modes) (Sheet 2 of 2)
Roche Diagnostics
Operator’s Manual · Version 2.0 B-41
6 Daily operation cobas c 311 analyzer
Starting the analyzer
Maintenance mode
In Maintenance mode, the motors driving the pipetter probes, the sample disk and
the reaction disk are turned off, while the analyzer remains powered on.
Any actions which require the top cover to be open (such as replacing ISE reagents or
maintenance actions, or checks) may only be performed in Maintenance mode (or
shutdown).
A Maintenance switch
1 Turn the maintenance switch (located on the right side of the analytical unit) to
Maintenance Mode.
The status line turns bright yellow. The yellow color indicates that the analyzer is
in Maintenance mode, even if Stand By is still displayed.
Roche Diagnostics
B-42 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Starting the analyzer
If an alarm was issued, the Alarm global button flashes. The color of this button
stands for the alarm level. Yellow indicates the caution level and red the stop level.
When the global button Alarm flashes, it is necessary to open the Alarm screen to
view the alarm. The Alarm screen identifies all system alarm conditions.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-43
6 Daily operation cobas c 311 analyzer
Starting the analyzer
The System Overview screen plays a central role within the cobas c 311 analyzer
software. This screen provides the operator with an overview of the whole system at
any given time and the analyzer can be prepared for daily routine operation (via the
Work Flow Guide).
It is possible to call up the System Overview screen from any point of the
software—Workplace, Reagent, Calibration, QC, or Utility— by choosing this icon
in the top left corner of the screen or by pressing <F12>.
e For a list of all shortcut keys, see:
Shortcut keys on page B-32.
The Work Flow Guide area at the top of the System Overview screen guides you
through the preroutine operation.
The Overview area provides a view of the current status of the analyzer and
components of the cobas c 311 analyzer.
The lower section of the Overview area displays information relevant to the
component currently selected such as incubator temperature (INC.).
Roche Diagnostics
B-44 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Starting the analyzer
Alarm This button displays the color corresponding to the highest priority alarm of the
currently selected component. Choose this button to display the global Alarm screen.
The alarm(s) corresponds to the component currently selected.
e See Check system alarms on page B-43.
e For a detailed description of the Alarm screen, refer to the Online Help.
Reagent Overview Choose this button to display an overview of the reagents currently loaded. This
button is only visible if you have selected AU (analyzer unit).
e See Reagent Overview button on page B-143.
Figure B-31 Work Flow Guide area of the System Overview screen
Sample Data Red The database is full (10 000 records). No additional samples can
Clear be processed until the hard disk is cleared.
Yellow The database is nearly full (more than 9400 records).
Reagent Red Reagent for at least one test or detergent required for the test is
Preparing not on board (mandatory) or is empty.
Yellow The number of tests remaining for a reagent, diluent, or wash
solution is below the yellow alarm threshold.
Purple The number of tests remaining for a diluent, test reagent or
detergent volume is less than the daily requirement - only
applicable if the Preventive Action check box has been selected.
Calibration and Yellow The system has recommended a calibration or quality control.
QC Select
Parameter Red A reagent, calibrator, or QC was loaded onto the analyzer, which
Download has not been installed before. New information for applications,
controls, and calibrators must be downloaded from cobas link.
Sample Blue This button is used to track samples, it does not change color.
Tracking
Roche Diagnostics
Operator’s Manual · Version 2.0 B-45
6 Daily operation cobas c 311 analyzer
Starting the analyzer
Yellow The number of tests remaining for a reagent is below the yellow
alarm level threshold.(a)
Purple The number of tests remaining for a reagent is below the purple
alarm level.
Red Indicates an instrument alarm at the Stop, S.Stop or E.Stop level
Alarm
Roche Diagnostics
B-46 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Starting the analyzer
The Work Flow Guide area consists of 6 buttons: Daily Maintenance, Sample Data
Clear, Reagent Preparing, Calibration and QC Select, Parameter Download, and
Sample Tracking. The first five buttons are placed in the suggested sequence of
preventive action. If a button is not highlighted, that action is not necessary.
The Sample Tracking button is used to search for samples on the analyzer. This is
possible at any time during operation as long as no samples have been removed from
the sample disk and no new run has been started.
e For more information about the meaning of the colors, see:
Color scheme Work Flow Guide area, Table B-8 on page B-45.
Preventive action
Preventive action is a look ahead at what might be required during daily routine
operation. When activated, preventive action triggers the following items:
o Reagent purple alarm
o Calibrate Now
At logon, the Preventive Action check box on the System Overview screen is
selected and therefore active.
For routine operation we recommend deselecting the Preventive Action check box
because it is not necessary to be informed about the triggered alarms.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-47
6 Daily operation cobas c 311 analyzer
Preroutine operation
Preroutine operation
Before you can start routine operation you must first prepare the system for
operation. Preroutine operation involves the following tasks:
o Performing maintenance actions
o Clearing and backing up data, if necessary
o Preparation of reagents, detergents and diluents
o Calibration
o Measurement of quality controls
o Download of parameters, if necessary
The Work Flow Guide area at the top of the System Overview screen guides you
through the preroutine operation.
Requirements: Define The Daily Maintenance button indicates when maintenance intervals are about to
maintenance types expire. However this function is available only if maintenance intervals for
maintenance items are defined.
e For more information, see
Color scheme Work Flow Guide area, Table B-8 on page B-45.
Maintenance types—scheduling and tracking maintenance items on page C-8.
Choosing Daily Maintenance on the Work Flow Guide area displays the
Maintenance screen. Use the Maintenance screen to execute maintenance actions or
maintenance pipe functions.
e See Performing maintenance items on page C-9.
Define a Power Up Pipe For a proper use of the analyzer Roche recommends performing some maintenance
items regularly. We recommend making these maintenance items automatic by
means of maintenance pipes.
e See Recommended maintenance pipes on page C-17.
o Daily maintenance items may be programmed as a Power Up Pipe and will then be
performed automatically when powering on or waking up the system.
o To backup the data of your analyzer to cobas link we recommend including the
maintenance item cobas link Essential Information Upload into the Power Up Pipe.
o The date of execution is updated each time a maintenance item is executed. It will be
updated even if operation is interrupted due to an alarm, etc. In this case, execute the
relevant maintenance item again and terminate it normally.
e For more information, see:
Maintenance pipes on page C-10
Power Up Pipe function on page C-14
Start Up Pipe function on page C-15
Roche Diagnostics
B-48 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Preroutine operation
2 While the analyzer executes the Power Up Pipe, perform the visual checks as
described (with top cover closed).
3 When the Power Up Pipe is finished, perform the visual checks as described.
4 Choose the System Overview screen.
o Choose the Daily Maintenance button if it is yellow or red. Perform the
indicated maintenance items.
o Before starting analysis, check that the temperature of the incubator bath is
within 37C 0.1C.
5 Continue with the following preroutine steps.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-49
6 Daily operation cobas c 311 analyzer
Preroutine operation
Use the Sample Data Clear button to delete all measurement results of routine and
Stat samples. Less sample data on the hard disk allows faster access to the data.
Backing up on a periodic basis is recommended.
o Once the hard disk is full, no more sample data can be saved. In the System
Overview screen the Sample Data Clear button turns to red. To save further data,
the hard disk must be cleared.
o It is strongly recommended that a sample clear is performed daily to minimize the risk
of data loss.
e For more information, see Color scheme Work Flow Guide
Table B-8 on page B-45
If a sample is measured with the same sample ID or same number as an already measured
sample, the results will be displayed with the original sample number or ID.
Roche Diagnostics
B-50 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Preroutine operation
Requirements: Define reagent The Reagent Preparing button indicates that insufficient amounts of reagent remain.
warning levels and mandatory However, this function is available only if the test counts (or volumes) for the reagent
tests level check and the mandatory tests were defined.
Define reagent warning levels The Reagent Preparing button changes its color to yellow or purple if the
corresponding alarm has been selected for one or more tests. Define reagent warning
levels for the yellow and purple alarm under Utility > System (2/5) > Reagent Level
Check for each reagent and detergent.
For the purple alarm, we recommend setting the volume of reagents to be used in one
day.
The purple alarm however, will only appear if the Preventive Action check box in the
System Overview screen has been selected.
Define mandatory tests The Reagent Preparing button changes its color to red if a reagent or detergent is
empty or if a reagent has been defined as mandatory and is not on board. Mandatory
tests are defined under Utility > Module Set > Test Assignment. Detergents and
diluents can not be defined as mandatory.
e For more information, see Color scheme Work Flow Guide:
Color scheme Work Flow Guide area on page B-45
Define reagent warning levels on page B-51.
Print Reagent Load/Unload List The Reagent Load/Unload List displays the cobas c packs and auxiliary reagents that
need to be replenished on the system.
Now, replace any required reagent, diluent, detergent, or wash solution according to
the Reagent Load/Unload List. Ensure that reagents have not exceeded their
expiration dates.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-51
6 Daily operation cobas c 311 analyzer
Preroutine operation
ISE reagents
The ISE auxiliary reagents (ISE IS, ISE Dil., and ISE Ref.) are stored in reagent bottles
in corresponding ISE reagent compartments.
Replace all required reagents as indicated on the Reagent Load List. Make sure to
place them according to their correct positions.
The analyzer must be in Maintenance mode (bright yellow status line) or shutdown
during replacement.
After replacement, the following steps must be performed:
1. Reagent level reset:
The volume must be reset under Reagent > Setting > Reagent Level Reset. For
ISE IS and ISE Dil., the remaining volume is then reset after the first reagent level
detection. This is automatically performed by the reagent probe before starting a
measurement.
2. Reagent prime:
After the replacement of ISE IS or ISE Ref., a Reagent Prime is required to fill the
flow path with the new liquid.
3. ISE calibration:
A changeover calibration will be requested in the Calibration Status screen.
Before performing the following actions, observe the following safety precautions:
o Skin inflammation or injury caused by reagents and other working solutions on
page A-9
o Infection and injury due to contact with instrument mechanism on page B-3
o Incorrect results due to incorrect reagent volume on page A-10
o Incorrect results due to expired reagents or mixing reagents on page B-4
Roche Diagnostics
B-52 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Preroutine operation
6 Perform a reset in Utility > Maintenance to return all mechanical parts to their
home positions.
7 Select the Reagent > Setting screen and select the reagent bottle that has been
replaced.
8 Choose Reagent > Setting > Reagent Level Reset and choose Yes.
A blank space is provided in the column Available Tests. The remaining volume
will be displayed after the first reagent level detection.
9 After replacing the ISE IS bottle, perform a Reagent Prime and an ISE calibration.
After replacing the ISE Dil. bottle, perform an ISE calibration.
e See To perform a Reagent Prime on page B-54.
4 Replace the ISE Ref. bottle, and place the tubing into the new bottle.
Make sure that the end of the tube touches the bottom of the bottle. Otherwise,
reagent may not be dispensed properly.
Incorrect results due to incorrect insertion of the ISE Ref. aspiration tube
If the aspiration tube is not inserted correctly, the reagent may not be dispensed
CAUTION properly. This may lead to incorrect results.
o Insert the aspiration tube so that the end of the tube touches the bottom of the
bottle.
o Do not bend the aspiration tube.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-53
6 Daily operation cobas c 311 analyzer
Preroutine operation
If you have replaced a reagent bottle, calibrate the ISE unit before you resume routine
operation.
e For details on the ISE calibration, see:
Calibration And QC Select button on page B-58
ISE calibration on page B-173.
Roche Diagnostics
B-54 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Preroutine operation
Loading cobas c packs It is only possible to load a cobas c pack when the analyzer is in Standby mode and
the applications have been downloaded from cobas link. Reagent registration takes a
few minutes as it includes piercing the cap and checking reagent volume.
The cobas c packs containing diluent for pre-dilution of samples, Hitergent for the
incubator water bath, and detergents for the reagent probe (NaOH-D and SMS) are
registered in the same way as other cobas c packs. For these solutions all relevant data
must be downloaded first from cobas link via Utility > Application > Download.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-55
6 Daily operation cobas c 311 analyzer
Preroutine operation
5 Slide a cobas c pack with the barcode label facing to the right side past the barcode
reader.
6 If reading is successful, you will be asked to confirm the cassette information. If
reading is not successful, an alarm occurs.
7 Place the cobas c pack in the empty position on the reagent disk.
8 Close the gate covering the reagent loading station.
If the reagent disk cover is left open for more than ten minutes, an alarm will occur.
Replacing auxiliary reagents e For instructions on how to replace auxiliary reagents, see:
To replace a cell detergent bottle on page B-141
To replace a sample probe detergent bottle on page B-142
Roche Diagnostics
B-56 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Preroutine operation
Roche Diagnostics
Operator’s Manual · Version 2.0 B-57
6 Daily operation cobas c 311 analyzer
Preroutine operation
It is necessary to calibrate all applications and measure quality control (QC) samples
regularly to verify the stability of reagents and the entire system. Make sure that
calibration has been completed successfully and that all QC results are OK before you
start routine operation.
Calibration and QC intervals vary from application to application and therefore each
application has its own configuration. In line with this configuration, the system
automatically recommends calibrations and QC measurements for all registered
applications.
Every time the system recommends a calibration this is indicated by the Calibration
And QC Select button turning yellow.
The following sections describe how to carry out a recommended calibration. The
instructions are based on the assumption that all calibration and QC configurations
have already been set.
e For general information on calibration, see:
Calibration concept on page B-151
Triggers for automatic calibration requests on page B-152
e For information on calibration settings, see:
Description of application parameters - Calib. tab on page B-224
To select tests for start-up group calibration on page B-158
e For information on QC settings, see:
To perform controls for active reagents on page B-186
To perform controls for standby reagents on page B-186
To perform QC after calibration on page B-187.
Roche Diagnostics
B-58 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Preroutine operation
Calibration or QC failure
If you have to manually assign a calibrator or QC container or other sample container—for
CAUTION example, due to an unreadable barcode—do not place any containers with barcodes in
manually assigned positions.
It is possible to do a calibration with a mixture of barcoded and non-barcoded calibrators.
However, you must assign the position of the non-barcoded calibrator and no position
should be assigned for the barcoded calibrator. If you put a barcoded container on a
position that was assigned manually, the calibration is not measured.
o Note that Hitachi micro cups are not applicable to calibrators and QCs.
o Before placing calibrators and controls in the sample disk, check that no bubbles or
foam are visible on the surface of the liquid.
o If you perform QC after calibration, QC samples do not have to follow directly after
calibrator samples. Other samples can be placed in between them.
Use the following procedure to request the recommended calibrations and controls.
If you need to select additional calibrations manually, select these from the
Calibration > Status screen.
e See Requesting and cancelling calibrations manually on page B-157.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-59
6 Daily operation cobas c 311 analyzer
Preroutine operation
2 Choose the buttons that are yellow in both the Calibration area and the QC area
to select the recommended calibrations and controls for measurement.
In case there are reagents on board with less than ten tests remaining, a confirmation
window opens and it is possible to cancel the calibration/QC request for these
reagents.
3 Select the list you want to print from the Load List area:
o QC comprises all requested QCs including QC after calibration.
o Calibration comprises all requested calibrations.
4 Choose Print to print the selected Load Lists. The Calibrator and QC Load Lists
display all calibrators and controls that are necessary for performing the
requested measurements.
5 Load the calibrators and controls onto the analyzer as directed by the Calibration
load list and the QC Load List, see: To load required calibrators and controls on
page B-61.
Roche Diagnostics
B-60 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Preroutine operation
e For more information about the different information on printouts, refer to the Online
Help of the particular report.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-61
6 Daily operation cobas c 311 analyzer
Preroutine operation
If running barcoded calibrator or QC samples and the lot read is not installed, the
Parameter Download button on the System Overview screen turns red and the
samples are not run. Choose Parameter Download and download the necessary lot
number of the calibrators or QCs.
If automatic printout of calibration and QC is selected on Utility > System (Page 4/5) >
Automatic Printout, the Calibration Monitor and Control Result Monitor reports are
automatically printed when the measurements are completed.
Roche Diagnostics
B-62 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Preroutine operation
If calibration or QC failed… In case a calibration fails or a QC result falls outside the expected range, check for QC
data alarms on the Workplace > Data Review screen and calibration alarms on the
printouts.
Data alarms are issued when measurement values or results are in any way unusual or
unexpected. Data alarms appear on screen and on reports as short strings also
referred to as data flags. Look up the meaning of each data flag and possible remedies
in the troubleshooting part of this manual.
e See Chapter 15 Data alarms.
After taking the necessary measures, repeat the calibration or QC before you begin
routine analysis.
If a realtime QC rule is violated, the system issues a realtime QC alarm. Make sure that you
check the results of the QC to see if the values are within the range.
Quality control results The results of QC measurements are saved in the database and displayed on the Data
Review screen as well as on the QC > Individual screen. QC can also be checked in
the Run Status screen instead of in the Individual QC screen.
It is important to regularly consolidate these results (by pressing the Accumulate
button) in order to generate long term quality control data (QC > Cumulative).
e For information on accumulation of control data, see:
Accumulation of QC measurement data on page B-181.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-63
6 Daily operation cobas c 311 analyzer
Preroutine operation
When the Parameter Download button is red, new parameters for applications,
calibrators, and controls must be downloaded from cobas link. Choose Parameter
Download in the System Overview screen to open the To Do List window.
e For an overview of downloading parameters from cobas link, see:
Loading or updating applications on page B-214.
The Sample Tracking window provides an overview of the samples in the sample
disk and allows the operator to search for samples which have already been registered
by the system.
Sample tracking is part of the routine workflow. Therefore, the Sample Tracking
window is described in the next section.
e See: Sample tracking on page B-72
Roche Diagnostics
B-64 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Routine Operation
Routine Operation
After completing preroutine operation, you are ready to start routine operation.
Before continuing with routine operation, deselect the Preventive Action check box on
the System Overview screen. Otherwise, the system will issue purple and other irrelevant
alarms in the course of routine operation.
Before performing the following actions, observe the following safety precautions:
o Infection by samples and associated materials on page A-8
o Infection and injury due to contact with instrument mechanism on page B-3
o Incorrect results due to incorrect reagent volume on page A-10
o Incorrect results and interruption of analysis due to contaminated samples on
page A-11
o Incorrect results due to evaporation of samples or reagents on page A-11
o Incorrect results due to carryover on page A-11
o Incorrect results due to position mismatch on page B-4
o Incorrect results due to missing covers of the ISE unit on page B-4
o Incorrect results due to aspiration of air on page B-4
Roche Diagnostics
Operator’s Manual · Version 2.0 B-65
6 Daily operation cobas c 311 analyzer
Routine Operation
In this section, we describe the workflow for an analyzer that operates in connection
with a Host and is set to process routine samples in barcode mode. Processing
samples always includes the following steps:
1. Test selection
2. Loading samples
3. Starting the measurement
4. Checking the results
This includes to check for open test requests.
Test selection
Test selections for routine samples are usually downloaded from the Host. However,
these test selections may be changed manually—regardless of the mode (Standby,
Stop, Operation, or Sample Stop) the analyzer is in. When working without a Host,
the test selection must be entered manually.
e For information on manual test selections, see Requesting a test manually on page B-91.
Roche Diagnostics
B-66 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Routine Operation
A
max. 88 mm
Figure B-39 Sample tubes on the outer ring with their barcode facing outwards
3 Ensure that the sample barcodes are facing the right way so the two barcode
readers can scan them. Samples placed in the inner ring must have their barcodes
facing inwards, samples placed in the outer ring must have their barcodes facing
outwards (see Figure B-39 on page B-67).
e For specifications of the correct position of the barcode label, see:
Scan range for sample barcodes on page A-79
Roche Diagnostics
Operator’s Manual · Version 2.0 B-67
6 Daily operation cobas c 311 analyzer
Routine Operation
a To start processing
1 Choose Start (global button).
Roche Diagnostics
B-68 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Routine Operation
Interruption of measurements
If you press the Stop button or a serious red alarm stops the analyzer, then no results
will be calculated for already pipetted samples. Sample status will change to Masked
or Ordered. Eliminate the cause and restart the measurement.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-69
6 Daily operation cobas c 311 analyzer
Routine Operation
In this section, we describe the workflow for an analyzer that operates in connection
with a Host and is set to process STAT samples in barcode mode.
Test selection
Test selection for STAT samples is the same as for routine samples. That is, test
selections for samples are downloaded from the Host in realtime but may still be
changed from the analyzer’s control unit.
e For information on changing test selections, see Requesting a test manually on page B-91.
You can check the Sample Tracking window for reserved STAT positions in both
Patient ID mode and barcode mode (Sample ID mode).
Roche Diagnostics
B-70 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Routine Operation
As results are generated on the instrument, they are saved in the database located on
the internal hard disk of the control unit. When all the test results requested for a
particular sample are available, the analyzer sends them to the Host where they can be
validated (automatic Host communication is activated by default).
Status of results To view or edit results on the analyzer, use the Workplace > Data Review screen; all
test results in the database are displayed here.
The St. column on the left side displays the status of each sample:
e For a detailed description of the Data Review screen, refer to the Online Help.
e For more information, see:
Data Review screen on page B-105
Archiving patient data on page B-106
Editing or deleting sample data on page B-107
Tracking samples on the analyzer on page B-111
When checking results, it may become necessary to locate a certain sample for further
inspection. If a sample is still on the analyzer, use System Overview > Sample
Tracking to locate it.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-71
6 Daily operation cobas c 311 analyzer
Routine Operation
Sample tracking
In the Sample Tracking window, the analytical status of each sample in the sample
disk is displayed.
The Sample Tracking window provides an overview of all samples in the sample
disk. The Track Sample area on the right side of the screen displays detailed
information about each sample in the sample disk.
Position status The following statuses are displayed in the Sample Tracking window:
Cup Present A cup is present in this position but no request has been made or
no process has been started.
Processing The sample is being processed.
Roche Diagnostics
B-72 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Routine Operation
With SBS mode If the SBS mode function is activated, the sample database is scanned for open test
requests after a run (SBS = scan before sample stop). This allows the analyzer to
process open test requests automatically in most cases. In some exceptional cases,
however, intervention by the operator may still be required.
e For detailed information, see: Open test requests (with SBS mode) on page B-113.
Without SBS mode If the SBS function is not activated, open test requests will require operator
intervention. You must check each open test request and eliminate the cause before
the open test request can be processed. Therefore we strongly recommend using the
SBS mode.
e For detailed information, see: Open test requests (without SBS mode) on page B-114.
Additional tests requested by the host are not indicated in the analyzer software.
Information about additional tests to be measured must be provided by the host.
7 In non-barcode mode, note the sequence number (S. No.) of the affected samples.
Do not change the disk position of these samples.
8 After performing the remedy, perform the corresponding procedure described on
the following pages.
e Processing open test requests (with SBS mode) on page B-75
Processing open test requests (without SBS mode) on page B-76.
Roche Diagnostics
B-74 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Routine Operation
Roche Diagnostics
Operator’s Manual · Version 2.0 B-75
6 Daily operation cobas c 311 analyzer
Routine Operation
Barcode mode For barcoded samples, you have the following options:
o Move the sample to another position.
e See To restart measurement (barcode mode) on page B-76.
o Perform a manual rerun.
e See Requesting manual reruns (barcode or non-barcode mode) on page B-100.
Notes
o If the analyzer is in Standby, open or additional requests are performed
automatically when you restart measurement. The position of the affected samples
may be changed (or you can leave them in their original position). All samples with
test requests will be processed, not only those samples that were not measured
during the previous run.
o If the analyzer is in Sampling Stop (Sample Reception) or Sampling Pause
status, the position of the affected samples must be changed before you press
Start. Repositioned samples will be measured after all other samples are processed.
Non-barcode mode Depending on the status, you have the following options:
o If the analyzer is in Sampling Stop (Sample Reception) or Sampling Pause
status, perform a manual rerun.
e See Requesting manual reruns (barcode or non-barcode mode) on page B-100.
o If the analyzer is in Standby, the sequence number must be changed before you
press Start.
e See Restart measurement from Standby (non-barcode mode) (5) on page 120.
Roche Diagnostics
B-76 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Routine Operation
To check whether the automatic Host communication is activated or not, choose the
Result Upload Setting tab on the Utility > System (Page 1/5) > Host
Communication Setting window (available in administrator level only).
System parameters Backup after to a floppy disk via Utility > Maintenance > Parameter Read/Write
(re)programming
Reports Backup after printing to a floppy disk via Print > View > Backup Disk Write
or DVD
Figure B-45 Overview of data types and possible storage routines
(a) The data types to be stored on cobas link data station are configured by Roche service personnel during installation (e.g., calibrator and control
data).
Roche Diagnostics
Operator’s Manual · Version 2.0 B-77
6 Daily operation cobas c 311 analyzer
Routine Operation
3 Make a backup copy if you have changed any system parameters (e.g., application
parameters, see Table B-37 on page B-266).
e For more information, see:
Saving system parameters on page B-265
Roche Diagnostics
B-78 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Routine Operation
For STAT samples, the results will be printed automatically if automatic printout of
Emergency Samples is selected on Utility > System (Page 4/5) > Automatic Printout.
e For more information about the different information on printouts, refer to the Online
Help of the particular report.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-79
6 Daily operation cobas c 311 analyzer
Routine Operation
Sampling Stop
After all test requests are completed, sampling stops. After the last result is calculated,
the system remains in Sample Reception mode for a set period of time. This time
period is set in Utility > System > Sample Reception area. Thereafter, the entire
system goes into Standby.
The period of time that is to elapse before the analyzer switches from operation to
Standby mode is defined under Utility > System. A time must be entered in the
combo box to specify the interval between starting operation and Standby. If the
Sample Reception Mode is disabled in the Start Conditions screen, the system goes
into Standby immediately after the last result is calculated.
Roche Diagnostics
B-80 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Shutting down the analyzer
This section discusses the tasks to be performed at the end of analysis. It gives a
detailed description of the shutdown including both complete shutdown and sleep
mode.
After routine operation is finished and all required maintenance has been performed,
the analyzer can be shut down. There are two options for analyzer shutdown:
o Complete shutdown
o Entering sleep mode
Shutdown In Shutdown status the analytical unit and the control unit are powered off. Only the
power for keeping reagents cooled is supplied.
Sleep mode During sleep mode most of the parts are powered off. Only the control unit, cooling
compartment and control board are powered on. The monitor is automatically
powered off to reduce power consumption when it is not used for a fixed period of
time. Move the mouse to wake up the monitor.
Sleep mode provides the possibility of an automatic start-up on the next day. This
way, the system can perform initialization and any programmed start up
maintenance items in the absence of an operator, before the beginning of a working
session. When the operator arrives, the analyzer is ready for operation.
Note that when the analyzer wakes up from sleep mode, the control unit performs a
shutdown to reset the PC operation software (Windows XP).
e For more information about checks after waking-up from sleep mode, see:
Start-up inspection on page B-38.
Perform one of the following procedures to shut down the analyzer after all required
daily maintenance has been performed.
e For information about the daily maintenance items that must be performed before
shutdown, see:
Daily manual maintenance on page C-38.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-81
6 Daily operation cobas c 311 analyzer
Shutting down the analyzer
2 If the maintenance item ISE Wash is included in the sleep pipe, make sure that
Sample Cleaner 1 and the cups with Activator and SysClean are on board.
3 Choose Utility > Maintenance and execute the pipe (e.g., daily sleep pipe).
The analyzer goes to sleep mode as soon as the pipe is finished.
4 Perform the regular checks after shutdown.
Roche Diagnostics
B-82 Operator’s Manual · Version 2.0
cobas c 311 analyzer 6 Daily operation
Shutting down the analyzer
Wake up after sleep mode Wake up can be programmed to a set time using the Power Up Pipe function. Of
course you can also wake up the analyzer manually at any time.
In case a Wake Up Time is set, a maintenance pipe can be performed automatically
after wake up.
e See Power Up Pipe function on page C-14.
If power to the analyzer is switched off prior to the complete shutdown of the
computer, the instrument may not start up properly when power is supplied again.
5 Switch off the power switch on the right side of the analytical unit.
6 Turn off the water supply, if necessary.
After shutting down the analyzer, check individual parts of the instrument according
to the given maintenance recommendations.
e See Checks after shutdown on page B-84.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-83
6 Daily operation cobas c 311 analyzer
Shutting down the analyzer
The checks and maintenance actions required after shutdown are the same as those
required before start-up.
e See Start-up inspection on page B-38.
If the analyzer has been set for automatic start-up, it is essential to perform all
required maintenance actions either at the beginning or the end of each working
session working session.
e See Overview of daily maintenance schedules on page C-34.
Below is the list of conditions that are to be checked before start-up. Check that:
ISE unit o All electrode cables and tubings are correctly connected. Tubing is not leaking.
If a trouble occurs, take suitable measures according to the condition and contact
your Roche service representative.
Roche Diagnostics
B-84 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Table of contents
This chapter provides descriptions of special tasks that are not usually part of the
daily workflow. It is meant to complement the chapter Daily operation, where
everyday tasks and common procedures for running the cobas c 311 analyzer are
described.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-85
7 Orders and results cobas c 311 analyzer
Table of contents
Roche Diagnostics
B-86 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Overview
Overview
The following sections describe tasks and data that can be accessed by the screens
under the Workplace menu. The Workplace menu comprises two screens: The
Workplace > Test Selection and the Workplace > Data Review screen.
Workplace > Test Selection Use this screen to enter test selections and demographics for a sample. The available
screen fields for sample identification vary, depending whether STAT or routine test
selections are being entered and whether the analyzer is operating in barcode mode
or not.
e For information on special operation procedures, see:
Requesting a test manually on page B-91
Entering unreadable sample barcodes on page B-103
Workplace > Data Review Use this screen to perform tasks related to reviewing and editing routine and STAT
screen results. Control results can also be viewed here. Other tasks that can be performed
from this screen include editing demographic information, sending data to the Host,
deleting data individually or in batches, backing up data and editing data.
e For information on special operation procedures, see:
Archiving patient data on page B-106
Editing or deleting sample data on page B-107
Displaying archived patient data on page B-110
Tracking samples on the analyzer on page B-111
e For a complete description of all fields and commands under the Workplace menu, refer
to the Online Help.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-87
7 Orders and results cobas c 311 analyzer
Test Selection screen
Use this screen to enter test selections and demographics for a sample. The available
fields for sample identification vary, depending whether STAT or routine test
selections are being entered and whether the analyzer is operating in barcode mode
or non-barcode mode.
If test selections are downloaded from a Host, this screen is commonly not used.
e For more information about the different sample information fields and the buttons on
this screen, refer to the Online Help of the particular field or button.
e For information about the test selection matrix, see Test selection matrix on page B-89
e For information on special operation procedures, see:
Requesting a test manually on page B-91
Entering unreadable sample barcodes on page B-103
A maximum of 160 test keys can be programmed on the analyzer. The test matrix has
32 keys in each group and up to 5 groups of test keys can be programmed on
Utility > System (Page 1/5) > Key Setting. One group at a time is displayed on the
Test Selection screen. Use the tabs above the test matrix to switch between the
groups. Group names are defined under Utility > System (Page 1/5) > Key Setting.
e For more information about Key Setting, refer to the Online Help.
Colors
The test keys display different colors to indicate the status of the test for the sample
ID displayed in the Sample ID text box. The following colors are available:
This test has been performed, a result is available, and can be ordered again.
A rerun has been performed for this test and the result is available.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-89
7 Orders and results cobas c 311 analyzer
Test Selection screen
Markings
One test or profile can be assigned to each key. The test keys display special markings
depending on the test's status or programming. The different markings and their
meanings are outlined below:
Normal test key for a test or profile without any masking indicators. The
reagent is on board of the analyzer and registered.
Mask (test channel or patient samples are masked)
o Test manually masked on the Masking window (Start (global button) >
Masking > T-Mask)
o Measurement of patient samples is manually masked (Start (global
button) > Masking > P-Mask). However, calibrations and controls can
be performed.
o The test can be requested. However, the test is performed only after the
test is manually unmasked.
o The test is not assigned to the analyzer in Utility > Module Set.
Reagent mask (no reagent available)
o Test is automatically masked by the analyzer. The volume of reagent, or
detergent, that is required for the respective test is either insufficient or
the reagent is not present on the analyzer. The test can be requested;
however, the test is performed only if a new reagent or detergent is placed
on the analyzer (e.g., Sample Cleaner 1, when a special wash is
programmed, see Special Wash on page B-255).
Calibration mask (calibration failed)
Test is automatically masked by the system if no valid calibration is available
and the test requires a calibration.
The purple bar will not be displayed on the test key if:
o Calibration auto masking setting for the individual test is not selected on
Utility > Application > Calib.
o Calibration auto masking setting for the analyzer is not selected on
Utility > System (Page 2/5) > Calib Mask Setting.
Increased sample volume
o Test will run with an increased sample volume.
Decreased sample volume
o Test will run with a decreased sample volume.
Dilution
o Test will run with a dilution. The number after the triangle indicates the
ratio of the dilution (for example, 3 is 1:3).
Profile
o A brown bar on the key indicates the key has been assigned to a profile.
When selected, all tests assigned to that profile are highlighted.
o If the reagent for a test is not on board, the corresponding test key is marked with
reagent mask and calib. mask.
o If a diluent for a test is insufficient, this test will not be performed. However, the
corresponding test key is NOT marked with a red bar. The test will only be marked with
a "M" in the Status column of the Test Review window (M status flag).
Roche Diagnostics
B-90 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Test Selection screen
Tests can be manually selected for STAT and routine samples when the analyzer is in
either barcode or non-barcode mode.
This manual procedure is necessary if the test selection is not received from an
external system (host).
Samples which have been manually prediluted can be measured, but the
corresponding dilution factors are NOT taken into consideration when the results are
calculated by the analyzer. It is the operator's responsibility to calculate the final
results.
7 Select the sample container type and press <Enter>. The cursor moves to the
Sample Volume / Dilution box.
8 Select the necessary dilution, if any, for the sample and press <Enter>. The cursor
moves to the test key.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-91
7 Orders and results cobas c 311 analyzer
Test Selection screen
9 Select the test, combination of tests, or test profiles for the sample. Selected tests
and profile keys appear white.
10 If necessary, type in demographic information about the sample and the patient,
such as the draw date, the sex and the age, and a comment in the Demographics
window.
11 Choose Save to save the test selection.
Completing the test selection for all samples, continue with To check and correct
registered tests and to start analysis on page B-92.
Roche Diagnostics
B-92 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Test Selection screen
Samples which have been manually prediluted can be measured, but the
corresponding dilution factors are NOT taken into consideration when the results are
calculated by the analyzer. It is the operator's responsibility to calculate the final
results.
9 Select the sample container type and press <Enter>. The cursor moves to the
Sample Volume / Dilution box.
10 Select the necessary dilution, if any, for the sample and press <Enter>. The cursor
moves to the test key.
11 If necessary, type in demographic information about the sample and the patient,
such as the draw date, the sex and the age, and a comment.
12 Select the test, combination of tests, or test profiles for the sample. Selected tests
and profile keys appear white.
13 Choose Save to save the test selection.
Completing the test selection for all samples, continue with To check and correct
registered tests and to start analysis on page B-94.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-93
7 Orders and results cobas c 311 analyzer
Test Selection screen
3 Type in the number of samples for repeating the test selection in the Repeat
Count text box.
4 Choose OK.
In the Repeat operation, demographic data are not copied. When the demographic data
are necessary, they have to be entered with each sample ID after collective entry.
Roche Diagnostics
B-94 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Test Selection screen
Processing reruns
Automatic reruns
If a test result is flagged with a data alarm (e.g., the result is outside the technical
limits, >Test alarm) and the automatic rerun function is activated, this test will be
measured again.
Automatic reruns can be performed for most of the data alarms, except for some
alarms according to the rerun list.
e For information on data alarms that can cause an automatic rerun, see:
Rerun list on page D-47.
The following table shows in which cases an automatic rerun will be performed for a
specific test if the result of this test is flagged with a data alarm:
System-wide setting The system-wide setting is displayed in the Automatic Rerun area on the Start
Conditions window.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-95
7 Orders and results cobas c 311 analyzer
Test Selection screen
3 Select the Routine check box or STAT check box (or both) to process reruns
without operator intervention. Routine and STAT reruns can be selected
separately.
4 Choose OK to save the rerun setting.
5 Choose Cancel to close the Start Conditions screen.
Application-specific setting Automatic reruns are performed depending on whether the Automatic Rerun check
box on Utility > Application > Range is selected or not for each test. If it is, the test is
automatically requested for rerun and remeasured whenever a result is flagged with a
specific data alarm. This applies to data alarms which indicate that a result is outside
the technical or repeat limit (>Test, <Test, >Rept, <Rept data alarms). Selection of the
Automatic Rerun check box is an application-specific setting; that is, it has to be set
for each individual application.
e For information on using the application-specific rerun setting, see:
Automatic Rerun on page B-229
Repeat limit on page B-230.
Manual reruns
If Automatic Rerun is set to NO in the Start Conditions window, all reruns are to be
processed manually (regardless of application-specific settings). This is an advantage
when you aim for an optimized throughput.
e For information about performing manual reruns, see:
Performing manual reruns on page B-98.
Roche Diagnostics
B-96 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Test Selection screen
o Not all displayed dilution ratios are available for every test.
o If you choose a dilution ratio in the S. Vol./D.Ratio field, this selection will overwrite
the pre-programmed dilution ratio indicated in the Utility > Application > Analyze
screen.
If for example the test has a pre-programmed dilution ratio of 1:5 and you select a
dilution ratio of 1:2 for S. Vol./D.Ratio, the test is processed with a dilution of 1:2.
o If the sample volume of a test is higher than 20 L, a manual dilution request is not
accepted and a pop-up window is displayed.
Make sure that sufficient diluent is loaded for the system to dilute samples.
Automatic dilutions Individual tests can be configured with a dilution for the first run and/or rerun.
Dilutions defined in the system are automatically performed and calculated by the
system. Administrator access is required to define these dilution parameters
(Utility > Application > Analyze).
e For more information, see:
Sample Volume area on page B-223
Manually requested dilutions Dilutions can be requested by the operator. Choose the Sample Volume / Dilution
box on Workplace > Test Selection and select Decrease, Increase, or a dilution ratio
(1:3 - 1:50).
e For more information, see Requesting a test manually on page B-91.
Request from Host Dilutions can also be requested by a Host computer. In this case, the samples are in
the sample disk and processed again.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-97
7 Orders and results cobas c 311 analyzer
Test Selection screen
Prediluted samples
Manually prediluted samples are samples which have been prediluted before they
were put on the analyzer. (These are not to be confused with manually requested
dilutions.)
Samples which have been manually prediluted can be measured, but the corresponding
dilution factors are NOT taken into consideration when the results are calculated by the
system. It is the operator's responsibility to calculate the final results.
In case you predilute a sample before you reload it onto the analyzer, make sure to
select the Pre-dilution check box on Workplace > Test Selection. The results will
then be flagged with a P (either on the printout or when sending the result to the
host). This indicates that the sample was prediluted and the customer must calculate
the final result.
Samples can be manually selected for rerun if a single test result is flagged with a data
alarm. Even if a result is not flagged with a data alarm, it may sometimes be desirable
to verify a result by a rerun.
This manual procedure is necessary if the Automatic Rerun setting in the Start
Conditions screen is disabled.
e For general information about reruns, see:
Automatic reruns on page B-95.
Roche Diagnostics
B-98 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Test Selection screen
Roche Diagnostics
Operator’s Manual · Version 2.0 B-99
7 Orders and results cobas c 311 analyzer
Test Selection screen
6 Choose Add in the Rerun Assignment window to add this sample to the Manual
Rerun list in the bottom of this window.
7 Choose OK to close the Rerun Assignment window.
8 On the Test Selection screen, select the tests that require a manual rerun.
The color of a selected test key changes from yellow to white (manual rerun) or
from gray to white (3rd result). Tests may also be added which have not been
measured before.
9 Choose Save on the Test Selection screen.
10 Repeat the previous procedure up to the last sample that requires a rerun.
11 Choose Start (global button).
12 Choose Start to initiate the rerun.
When pipetting of a rerun sample begins, this sample will be deleted from the
Manual Rerun List.
If a sample listed in the Manual Rerun List is not placed on the sample disk, this
request will be deleted.
Roche Diagnostics
B-100 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Test Selection screen
This section describes optional routines that may be used during a run:
o Additional tests
o Additional samples
If the 6th and final SBS database check has already been performed before additional
tests are ordered, restart the measurement to perform the tests during the next run.
e See: Additional samples or tests on page B-113.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-101
7 Orders and results cobas c 311 analyzer
Test Selection screen
o If a routine sample is currently being pipetted, it takes about 3-4 pipetting cycles
before the instrument enters Sampling Pause status.
o If a STAT sample is currently being pipetted, pipetting of all tests of the STAT
sample will be completed before entering Sampling Pause status.
o Pipetting of calibrator or QC samples cannot be paused. Press the Pause button
again when pipetting of all requested calibrators and QCs is completed.
If non-barcoded samples have been exchanged, all test requests must be deleted.
Choose Exchanged and the Start Conditions screen appears again. Now, choose
S.Stop (global button) to put the analyzer into Standby and delete the test requests.
After deletion, all test selections must be re-entered or re-downloaded from the host.
Roche Diagnostics
B-102 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Test Selection screen
The barcode label must be attached in such a way that it is visible through the
openings in the sample disk.
Figure B-52 Samples on the outer ring with their barcode facing outwards
Roche Diagnostics
Operator’s Manual · Version 2.0 B-103
7 Orders and results cobas c 311 analyzer
Test Selection screen
5 Type the position in the Disk Pos. text box and press <Enter>.
6 Type the sample ID of the sample in the Sample ID text box and press <Enter>.
7 Choose Add. The sample type, position, and sample ID are displayed.
8 Repeat steps 3 to 7 for any further samples with a barcode read error.
9 Choose OK.
Disk position assignment is deleted after the sample has been measured
If you want to manually rerun a sample with unreadable barcode, you have to reassign the
position of this sample in the Barcode Read Error window. Then this sample can be put
in the Manual Rerun list in the Rerun Assignment window.
e See: Requesting manual reruns (barcode or non-barcode mode) on page B-100
Roche Diagnostics
B-104 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Data Review screen
Use this screen to perform tasks related to reviewing and editing routine and STAT
results. Control results can also be viewed here.
Other tasks that can be performed from this screen include editing demographic
information, sending data to the Host, editing, and deleting data individually or in
batches, backing up data.
Status column The St. column displays the sample status codes.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-105
7 Orders and results cobas c 311 analyzer
Data Review screen
Disk column This column displays the disk position number (001-108) of the current sample. It
also shows how the sample was run, using the following code:
Code Sample
N Routine (normal sample)
E STAT (emergency sample)
C QC (control sample)
Table B-19 Disk position codes in the Data Review screen
e For more information about the different sample information fields and the buttons on
this screen, refer to the Online Help of the particular field or button.
e For information on special operation procedures, see:
Archiving patient data on page B-106
Editing or deleting sample data on page B-107
Displaying archived patient data on page B-110
Tracking samples on the analyzer on page B-111
Use this function to make backup copies of test results on a DVD (when operating
without a Host or for a backup additionally to the Host).
4 Select the medium to which the data will be written, for example Floppy Disk,
from the Media area.
o If only one medium is available, for example, floppy disk, the Media area is not
active.
o Mass Storage denotes the DVD disk.
Roche Diagnostics
B-106 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Data Review screen
7 Select the format in which the data is to be written, ASCII or Binary, and select
the Base Data or All Data option in the File Format area.
o ASCII files can be read by other PC systems, but cannot be reloaded into the
cobas c 311 analyzer.
o Binary files can be reloaded into the cobas c 311 analyzer (Workplace > Data
Review > choose Backup Media from the data list box), but cannot be read by
other PC systems.
o Base Data will only save the measurement results and the data alarms.
o All Data will additionally save the units of the measurement results and the sample
volume/dilution ratio.
8 Choose OK to back up the selected data after confirmation. The Save Data
window is displayed. This closes after the data has been backed up.
If a confirmation window with the message Mass storage media is full or Floppy disk is
full is displayed during backup, no data is stored on the media.
o Insert an empty medium and repeat the backup process
e For applicable DVD media types, see: Control unit on page A-73.
Sample data displayed on the Data Review screen can be edited or deleted if required.
Use the following procedures to edit or delete sample data.
QC View is also available as an option in the list box in the Data area, but this view
displays only QC data.
e For information on using a Backup Disk, see Displaying archived patient data on
page B-110.
3 Select a sample from the list on the left of the screen. The results, 1st and Rerun, of
the tests performed on this sample are displayed in the list on the right of the
screen.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-107
7 Orders and results cobas c 311 analyzer
Data Review screen
4 Choose Test Review to display the Test Review window. This window displays
more information on the test results.
5 Select the test result in the Data column. The result is activated (blue background)
and highlighted white.
Roche Diagnostics
B-108 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Data Review screen
a To delete a sample
1 Choose Workplace > Data Review.
2 Select an individual sample, or a number of samples, to delete from the list on the
left of the screen.
3 Choose Delete Record to delete the sample(s) selected after confirmation.
When a QC sample is deleted from Routine View, this will be transferred to the QC View
Database. To delete this you have to delete the QC sample from the QC View Database.
The Sample Count is unchanged after a record has been deleted. To delete all records,
use Delete All. This will reset the sample count to 0.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-109
7 Orders and results cobas c 311 analyzer
Data Review screen
Only data saved in binary format can be reloaded onto the control unit.
e See Archiving patient data on page B-106
Roche Diagnostics
B-110 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Data Review screen
o Only STAT and routine samples can be searched for in the Sample Tracking window.
o The location of a sample is only indicated on the Sample Tracking window once a
sample has been scanned.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-111
7 Orders and results cobas c 311 analyzer
Open test requests
4 Choose the Stat or Routine tab, depending on what kind of sample you are
searching for.
5 Select one of the following search options (e.g., to search by sample type):
o Sample Type
o Sample ID (barcode mode only)
o Sequence No. (non-barcode mode only)
o Disk position (non-barcode mode and STAT only)
o Main comment
6 Choose OK.
The samples that match the search criteria will be marked in the graphics
representing the sample disk.
After a run has been completed, you may find that some of the tests you requested
could not be performed. These open test requests (or open orders) may occur, for
example, if you request additional tests for a sample, or if a data alarm occurs while
automatic rerun is inactive.
With SBS mode The SBS mode function allows the analyzer to process open test requests
automatically without operator intervention in most cases. In some exceptional cases,
however, intervention by the operator may still be required.
e See Open test requests (with SBS mode) on page B-113.
Without SBS mode If the SBS function is not activated, open test requests will require operator
intervention. You must check each open test request and eliminate the cause before
the open test request can be processed. Therefore we strongly recommend using the
SBS mode.
e See Open test requests (without SBS mode) on page B-114.
Roche Diagnostics
B-112 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Open test requests
If SBS mode is activated, the analyzer scans the sample database after a run to check
for open test requests (SBS = scan before sample stop). The analyzer processes the
following categories of open test requests without operator intervention:
o Test orders without results (exceptions see below)
o Additional test orders for first run or rerun
o Tests which you have previously unmasked during operation (P-Mask, T-Mask,
Calib. Mask)
Pause/Scan and S.Stop buttons Both buttons can be used during SBS mode, as in normal operation.
Additional samples or tests You may add additional samples when the Pause/Scan button is pressed (Sampling
Pause). The additional samples are measured after restarting from Pause.
Additional tests can be requested during normal operation if the sample is already
completed, or during Sampling Pause.
The SBS database can be checked several times during a run (up to six times). So even
if the first database check has already been performed, additional tests or samples will
be measured in the next cycle. (After the sixth database check during a run, the
analyzer goes into Sample Reception mode or Standby. If you have added samples,
calibrators or controls several times during a run, there may still be some open orders
remaining after six database checks. In this case you only need to initiate a restart.)
e For more information, see:
To order additional tests in SBS mode on page B-101
To process additional samples during operation on page B-102.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-113
7 Orders and results cobas c 311 analyzer
Open test requests
Pipetting priority If samples are added during Sampling Pause in SBS mode, they are processed with a
lower priority than samples already identified during the SBS database check.
e For more information on priority levels, see:
General pipetting priority on page B-121
After a run has been completed, you may find that some requested tests could not be
performed. These leftover tests or open test requests (or open orders) fall into the
following categories:
o Open test requests with a Confirmation window
o Additional test requests:
a) Requested by user (with a Confirmation window)
b) Requested by host (cannot be detected by the analyzer software)
Open test requests with a Although a test request exists, it is not carried out. This may occur under certain
Confirmation window circumstances (see Table B-20 on page 117). This category of open test requests is
identified by the following Confirmation window.
Roche Diagnostics
B-114 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Open test requests
Additional test requests The cobas c 311 analyzer only scans the sample disk at the beginning of a run.
Therefore, the analyzer cannot automatically process any additional tests requested
during the run (by user or host).
Additional test request is used as a generic term for the following cases:
o Manual reruns
o 3rd results
o Additional test selections
Two subcategories of additional test requests must be distinguished:
a) In case the user requested additional tests via the analyzer software, the same
confirmation window (Figure B-60) appears after the run.
b) Additional tests requested by the host are not indicated in the analyzer
software. Information about additional tests to be measured must be sent
manually from host to the analyzer.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-115
7 Orders and results cobas c 311 analyzer
Open test requests
Additional tests
Open test requests
with a Confirmation window requested requested
by user by host
Working in barcode or
3. Yes No
non-barcode mode? Barcode mode?
or:
Page B-98
Roche Diagnostics
B-116 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Open test requests
5 After performing the remedy, follow the procedures described in Processing open
test requests (without SBS mode) on page 119.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-117
7 Orders and results cobas c 311 analyzer
Open test requests
Additional tests requested by the host are not indicated in the analyzer software.
Information about additional tests to be measured must be provided by the host.
4 Follow the procedures described under Processing open test requests (without SBS
mode) on page 119.
Roche Diagnostics
B-118 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Open test requests
The procedure for restarting the measurement depends on whether the analyzer is in
barcode or non-barcode mode.
a To restart measurement
1 Make sure the cause of the open test request is eliminated.
2 Remove the affected sample from the sample disk and check the sample container
visually for sample clot or sample short.
3 Place the sample into a new position on the disk.
4 Choose Start (global button) > Start to restart the measurement.
All samples with test requests are processed. No further action is required.
Notes
o If the analyzer is in Standby, open or additional requests are performed
automatically when you restart measurement. The position of the affected samples
may be changed (or you can leave them in their original position). All samples with
test requests will be processed, not only those samples that were not measured
during the previous run.
o If the analyzer is in Sampling Stop (Sample Reception) or Sampling Pause
status, the position of the affected samples must be changed before you press
Start. Repositioned samples will be measured after all other samples are processed.
Barcode mode For barcoded samples, you have the following options:
o Move the sample to another position.
e See To restart measurement on page B-119.
o Perform a manual rerun.
e See Requesting manual reruns (barcode or non-barcode mode) in the Operator’s
Manual.
Non-barcode mode Open test requests can be processed either from Standby, Sampling Stop, or Sampling
Pause status.
o If the analyzer is in Sampling Stop (Sample Reception) or Sampling Pause
status, perform a manual rerun using the Rerun Assignment window.
e See Requesting manual reruns (barcode or non-barcode mode) in the Operator’s
Manual.
o If the analyzer is in Standby, the sequence number must be changed before you
press Start.
e See Restart measurement from Standby (non-barcode mode) (5) on page 120.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-119
7 Orders and results cobas c 311 analyzer
Open test requests
a To restart measurement
1 Make sure the cause of the open test request is eliminated.
2 On the Data Review screen, note the lowest sequence number (S. No.) of the
affected samples.
3 Change the test selection if required. Make sure to save the test selection.
4 Choose Start (global button).
5 In the Start Sample No. text box, enter the lowest sequence number.
Roche Diagnostics
B-120 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
General pipetting priority
This overview shows in which priority samples are processed, regardless of whether
SBS mode is activated or not.
Basic pipetting priority: During normal operation as well as in SBS mode, samples are pipetted with the
following priority:
1. STAT sample(a) with rerun orders
2. STAT sample with 1st run orders
3. Routine sample with rerun orders
4. Routine sample with 1st run orders
Calibration and QC requests: Calibration and QC can be requested during Sampling Pause. This affects the
pipetting priority.
1. Calibration
2. QC after calibration
3. STAT sample with rerun orders
4. STAT sample with 1st run orders
5. Routine QC / auto QC
6. Routine sample with rerun orders
7. Routine sample with 1st run orders
Additional samples If samples are added during Sampling Pause in SBS mode, they are processed with a
lower priority than samples already identified during the SBS database check.
1. STAT sample with rerun orders
2. STAT sample with 1st run orders
3. STAT sample added (during Sampling Pause)
4. Routine sample with rerun orders
5. Routine sample with 1st run orders
6. Routine sample added (during Sampling Pause)
(a) Regarding to SBS mode, STAT sample here always means a STAT sample already identified during the
last SBS database check.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-121
7 Orders and results cobas c 311 analyzer
Preparing a DVD disk
The cobas c 311 control unit computer is optionally equipped with a optical disk
drive. The optical disk drive can be used for installing software updates and backing
up and restoring data.
e See Archiving patient data on page B-106
Supported disk types The following disk types can be used for backing up data.
Unsupported disk types All disks that are not listed in Supported disk types cannot be used for backing up
data, e.g. the following disk types:
o CD-R
o CD-RW
o DVD-R DL
o DVD+R DL
o DVD-RW DL
o DVD+RW DL
o DVD-RAM
o All types of Blu-ray disks
Formatting is required All disks must be formatted (initialized) on the cobas c 311 control unit computer,
before they can be used for backing up data on them.
e See Formatting a disk on page B-123
Compatibility to other Most disks must be made compatible so they can be read by other computers than the
computers cobas c 311 control unit computer.
e See Making a disk compatible on page B-125
Roche Diagnostics
B-122 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Preparing a DVD disk
Formatting a disk
All disks must be formatted (initialized) on the cobas c 311 control unit computer,
before you can backing up data on it.
Formatting several times Recordable disks can be formatted once only. Rewritable disks can be formatted
several times. The formatting function can be used to delete all data from rewritable
disk.
e For Information about the disk types, see Supported disk types on page B-122
Formatting on other computers Only format disks on the cobas c 311 control unit computer. The cobas c 311 control
unit computer may not be able to read disks formatted on any other computer.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-123
7 Orders and results cobas c 311 analyzer
Preparing a DVD disk
Different types of formatting The cobas c 311 control unit computer can perform either a Quick (erase) or Full
format of the discs. The following table describes the fundamental differences
between the two.
a To format a disk
o The procedure for formatting a disk may differ, depending on the configuration of your
system.
o Once the formatting operation is started, it cannot be cancelled.
o Formatting a disk deletes all data on the disk.
4 Type a name for the disk in the Volume label text box if desired.
5 Choose Quick (erase) or Full from the Format type area.
Depending on the type of disk only one option is available.
e For more information about format types, see Different types of formatting on
page B-124
Roche Diagnostics
B-124 Operator’s Manual · Version 2.0
cobas c 311 analyzer 7 Orders and results
Preparing a DVD disk
Most disks must be made compatible so they can be read by other computers than the
cobas c 311 control unit computer.
Availability for disk types The make compatible function is not for all disk types available. The following table
describes the availability for the disk types.
Disk type Make compatible available? Can more data be added after using
make compatible?
DVD-R Yes No
DVD+R Yes No
DVD-RW Yes Yes
DVD+RW No n/a
Table B-23 Make compatible: Availability for disk types
Adding data after making After a disk has been made compatible usually it is not possible to add data on it.
compatible However, for some disk types it is possible to add data after the disk has been made
compatible. After adding data the disk must be made compatible again so it can be
read by another computer than the cobas c 311 core unit computer.
e For information about the disk types, see Availability for disk types on page B-125
o The procedure for making a disk compatible may differ, depending on the
configuration of your system.
o Once the making compatible operation is started, it cannot be cancelled.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-125
7 Orders and results cobas c 311 analyzer
Preparing a DVD disk
Roche Diagnostics
B-126 Operator’s Manual · Version 2.0
cobas c 311 analyzer 8 Reagents
Table of contents
Reagents
This chapter describes all types of reagents used on the cobas c 311 analyzer. It
provides information about the system’s reagent management as well as detailed
information on how the operator can monitor the status of loaded reagents
(operator-related reagent management).
Roche Diagnostics
Operator’s Manual · Version 2.0 B-127
8 Reagents cobas c 311 analyzer
Table of contents
Roche Diagnostics
B-128 Operator’s Manual · Version 2.0
cobas c 311 analyzer 8 Reagents
Reagent concept
Reagent concept
Reagents This section provides information about reagents used on the cobas c 311 analyzer,
how they are used for ISE applications and photometric applications. It describes the
reagent containers and explains how the analyzer registers new reagents and how it
monitors reagent consumption.
This section describes all reagents necessary to run the cobas c 311 analyzer’s ISE unit
and the reagents that are specific for each available ISE application. The following ISE
applications are available:
Reference information for ISE The following table compiles reference information for all ISE reagents, such as
reagents auxiliary reagents, calibrators, and controls.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-129
8 Reagents cobas c 311 analyzer
Reagent concept
Reagents for measurements Each ISE measurement uses the following reagents:
o Internal standard (ISE IS), used for one-point calibrations performed before and
after each sample determination
o Reference solution (ISE Ref.), used for reference electrode measurements
o Diluent (ISE Dil.), used for 1:31 dilution of sample
ISE auxiliary reagents (ISE IS, ISE Ref., ISE Dil.) are supplied in reagent bottles and
do not use a barcode for registration. ISE auxiliary reagents are to be replaced only
when the instrument is in Maintenance mode or in Shutdown status.
After the replacement of an ISE reagent, the following steps are required:
1. Reset the initial volume.
2. Perform a Reagent Prime to fill the flow path with the new liquid (not required
after replacement of ISE Dil.).
3. Calibrate the ISE unit.
e For more information about replacement of ISE reagents, see:
ISE reagents on page B-52.
Calibrators The following ISE calibrators are used depending on the calibration method:
o Std (1) or S1: ISE Low, a water-based solution
o Std (2) or S2: ISE High, a water-based solution
o Std (3) or S3:
o For global use: ISE Comp., a serum-based solution, used for blank
calibrations, full calibrations as well as for conditioning the electrodes
o For use in US only: ISE High (compensated) with compensated set points is used
for full calibrations.
The calibration interval for all ISE applications is 24 hours.
Quality controls The following products are used for quality controls:
o PNU or PNU Plus
o PPU or PPU Plus
Quality controls can be used either with barcode or without. The operator can decide
whether to use barcodes, change the settings in the software if necessary (Barcode
Setting area on Utility > System), but then has to stick to the respective mode for QC
measurements.
Do not put barcoded sample tubes to assigned positions for calibrators or quality controls.
Be consistent with the software settings (Position Assignment under Calibration >
Calibrator and under QC > Control).
Roche Diagnostics
B-130 Operator’s Manual · Version 2.0
cobas c 311 analyzer 8 Reagents
Reagent concept
This section describes all reagents necessary to run the photometric unit. It provides
an overview of diluents, auxiliary reagents (such as basic and acidic detergents), as
well as calibrators and controls.
Diluents
NaCl For photometric applications, either water or a NaCl solution is used as diluent. The
water is deionized water from the analyzer’s water tank. NaCl solution is supplied in
cobas c packs NaCl 9%.
Whenever used as diluent, the 9% solution from the cobas c pack is diluted with
water to a concentration of 0.9% by the analyzer.
Make sure you have sufficient diluent on the analyzer—especially if a higher demand of
diluent can be expected.
A1CD2 The diluent for HbA1c whole blood tests is also supplied in a cobas c pack.
e For information about the HbA1c whole blood application, see:
Loading the HbA1c Whole blood application on page B-217.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-131
8 Reagents cobas c 311 analyzer
Reagent concept
Hitergent Hitergent Surfactant for incubator bath to prevent algae 1 cobas c pack of Reagent disk
proliferation and foaming 50 mL
NaOHD NaOH-D Detergent 1 for special washes (carryover evasion, 1 cobas c pack of Reagent disk
reagent probe and cell wash, if necessary) and cell 50 mL
wash (Maintenance item (6) Wash Reaction Parts)
SMS SMS Detergent 2 for special washes (carryover evasion, 1 cobas c pack of
reagent probe and cell wash, if necessary) 50 mL
Table B-27 Photometric auxiliary reagents and detergents
(a) used in this documentation
Remaining volume control for The remaining volumes of the auxiliary reagents are monitored by countdown,
reagent bottles starting from a given initial volume. The remaining volumes are displayed on the
Reagent > Status screen. Whenever a reagent bottle is replaced, the operator has to
reset the initial volume. Reagent registration is not performed automatically
whenever a new bottle is placed on the analyzer.
e For more information on monitoring the reagent status, see:
Reagent Status screen on page B-140
Replacing auxiliary reagents on page B-140
Roche Diagnostics
B-132 Operator’s Manual · Version 2.0
cobas c 311 analyzer 8 Reagents
cobas c packs
Most of the calibrators and controls have to be reconstituted and pipetted into cups.
cobas c packs
Apart from a number of reagent bottles for auxiliary reagents, the principal reagent
container is the cobas c pack (COBAS INTEGRA cassettes can also be used).
All information about cobas c packs in this manual also applies to COBAS INTEGRA
cassettes.
In addition to the cobas c packs and COBAS INTEGRA cassette types, there is the
cobas c pack MULTI, which is an empty barcoded cobas c pack. These kits are used
for development channel applications and lyophilized reagents, which have to be
mixed with solvent just prior to use. To prepare a cobas c pack MULTI for use on the
analyzer follow the instructions given in the instructions for use in cobas link.
A cobas c pack can contain up to three reagent bottles. The center position is
position A. With the barcode facing to the rear, the left position is position B and the
right position is position C. Pipettings at different timings (R1, R2, or R3) are not
restricted to any position.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-133
8 Reagents cobas c 311 analyzer
cobas c packs
Barcode label
Each cobas c pack is equipped with a barcode label. The analyzer scans this barcode
label each time a cassette is loaded onto the analyzer at the reagent loading station.
If the barcode of a cobas c pack or other cassette is unreadable, this cassette is not
accepted. The operator is requested to close the gate.
Roche Diagnostics
B-134 Operator’s Manual · Version 2.0
cobas c 311 analyzer 8 Reagents
Handling of reagents
Handling of reagents
This section describes how the analyzer registers reagents and how it keeps track of
the remaining volumes of reagents in the course of their consumption. The
unloading and reloading of reagents is explained. The last sub-section covers
carryover evasion.
ISE IS and ISE Dil. are stored in reagent bottles in one of the ISE reagent
compartments. At analyzer startup, the remaining volume is checked for those ISE
reagents. The reagent pipetter probe is equipped with a level detector (capacitance
method), which monitors the liquid level in the reagent bottles.
The ISE Ref. reagent bottle is not equipped with a level detector. The remaining
volume of ISE Ref. is counted down with each test.
When you replace a reagent bottle, you have to reset the remaining volume manually.
After replacement, the liquid level is not detected. Therefore, use only new bottles for
replacement. The Reagent Overview screen displays the present remaining volume
of each reagent (System Overview > Reagent Overview screen).
If the ISE IS or ISE Dil. bottle is not on board at analyzer startup or if the liquid level
falls below the limit where the bottle needs to be replaced, a yellow alarm (Reagent
Short) is issued on the Alarm screen. The AU Unit and the Reagent Overview
button turn red.
When the number of available tests is “0”, all ISE tests will be masked and thus no
subsequent measurement is possible.
e For instructions on how to replace ISE reagents, see ISE reagents on page B-52.
o Make sure bubbles do not form in the ISE IS and ISE Dil. bottle. If bubbles form,
reagent level may not be detected correctly by liquid level detection.
o ISE reagents are to be replaced only when the instrument is in Maintenance mode
or in Shutdown status.
o Make sure that you place the reagents in their correct positions (refer to labels on
the analyzer).
o For ISE Ref., make sure that the end of the tube touches the bottom of the bottle.
Otherwise, reagent may not be pipetted properly.
o After replacing an ISE IS or ISE Ref. bottle, reagent priming must be carried out
to fill the flow path with the new liquid.
o Whenever replacing any ISE reagent, perform an ISE calibration before you
resume routine operation.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-135
8 Reagents cobas c 311 analyzer
Handling of reagents
Reagent registration
In order to register a new cobas c pack (or other cassette) the analyzer must read the
information stored on the barcode label. The operator clicks the Loading button in
the Reagent Setting screen and the analyzer unlocks the gate. The operator opens the
gate and scans the barcode label. After the cobas c pack is accepted, the operator
places the cassette in the reagent loading station.
e For information about the reagent loading station, see
Reagent loading station on page A-47.
If the cobas c pack currently being scanned has not been on the analyzer before, the
analyzer proceeds with the following actions:
o By reading the cassette barcode, the system checks the availability of the
corresponding test application.
o The piercer pierces the caps of the bottles in the cobas c packs.
o For standard cobas c packs, the analyzer registers the initial capacity encoded in
the barcode of the cobas c pack.
o For cobas c packs MULTI and COBAS INTEGRA cassettes, the reagent pipetter
probe performs an initial volume check—also referred to as initial cassette
volume check (ICVC)—for each of the bottles in the cobas c pack, if this ICVC
option has been selected.
After a cobas c pack has been registered the analyzer counts every pipetting out of
this cassette. Thus, the cassette’s initial number of available test is counted down.
When the number of remaining tests is lower than a defined limit, an alarm is issued
(yellow or purple alarm, according to configuration under Utility > System
(Page 2/5) > Reagent Level Check). When the number of remaining tests becomes
zero, an alarm is issued (red alarm) and the test is masked.
e For more information on monitoring the reagent status, see:
Reagent Status screen on page B-140
Roche Diagnostics
B-136 Operator’s Manual · Version 2.0
cobas c 311 analyzer 8 Reagents
Handling of reagents
The analyzer counts down each cassette’s initial number of available tests with every
pipetting out of that cassette. The thus obtained number of available tests is stored by
the analyzer for up to 3000 cobas c packs. If you unload a cobas c pack and reload it
later, the analyzer recognizes the cassette and continues the test count at the point
where it was unloaded.
When reloading a cobas c pack onto the same analyzer, the analyzer does not
perform a second volume check—like the initial cassette volume check (ICVC) when
the cassette was new. It is assumed that the cassette’s reagent volumes remain
unchanged while the cobas c pack is not on the analyzer.
e For further instructions, see Unloading cobas c packs on page B-57
Carryover evasion
Even though all pipetter probes and reaction cells are washed and rinsed thoroughly
after each use, it could be possible that traces of sample liquid or reagent remain on
contacted parts (probes or reaction cells) and are carried over at the next pipetting.
By means of the special wash function, the cobas c 311 analyzer prevents traces of
sample liquid or reagent of one test from being carried over to another test and bias
the result of the latter.
e A current carryover evasion list is available on cobas link. Refer to the important note for
NaOHD / SMS / SmpCln 1.
e For more information on programming special washes, see:
Programming a reagent probe wash on page B-255
Programming a cell wash on page B-257
Programming a sample probe wash on page B-258
Roche Diagnostics
Operator’s Manual · Version 2.0 B-137
8 Reagents cobas c 311 analyzer
Reagent screen overview
Reagents The following sections describe tasks and data that can be accessed by the screens
under the Reagent menu. The Reagent menu comprises two screens: The Reagent >
Setting screen and the Reagent > Status screen. In addition, the Reagent Overview
button on the System Overview screen as well as windows accessed via this button
are explained.
Reagent > Setting screen The Reagent > Setting screen provides detailed information about each test reagent
and position. The compiled data include the numbers of available and remaining
tests, reagent codes, lot numbers, sequence numbers, expiration dates (shelf-life) as
well as on board stability of reagents (in days). The following tasks can be performed
from this screen:
o Loading and unloading cobas c packs
o Reagent level reset of ISE reagents
o Priming ISE reagents
o Registration of Development Channel reagents
Reagent > Status screen The Reagent > Status screen provides all important data about test reagents to
monitor the reagent status on the analyzer. The data are sorted according to the
names of the applications. Moreover, there is an additional list, which provides
information about auxiliary reagents. The following task can be performed from this
screen:
o Resetting the remaining volume control for auxiliary reagents
e For a complete description of all fields and commands under the Reagent menu, refer to
the Online Help.
e For more information, see:
Reagent Setting screen on page B-139
Reagent Status screen on page B-140
System Overview > Reagent By changing its color, the Reagent Overview button indicates potentially insufficient
Overview button amounts of reagent. Choose this button to view details about the status of loaded
reagents.
e See Reagent Overview button on page B-143.
Roche Diagnostics
B-138 Operator’s Manual · Version 2.0
cobas c 311 analyzer 8 Reagents
Reagent Setting screen
Reagent Level Reset Choose this button to reset the initial volume of ISE reagents.
e For more information, see:
To replace ISE IS or ISE Dil. on page B-52
To replace ISE reference solution (ISE Ref.) on page B-53
Reagent Prime Choose this button to prime reagents of the ISE unit. Both ISE IS and ISE Ref can be
primed.
e For more information on replacing and priming ISE reagents, see:
ISE reagents on page B-52
To perform a Reagent Prime on page B-54.
Development Channel Choose this button to reserve a Development Channel application to be assigned to a
cobas c pack MULTI.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-139
8 Reagents cobas c 311 analyzer
Reagent Status screen
A Test list
B Reagent Details list
C Auxiliary Reagent list (for photometric assays only)
The Reagent > Status screen displays reagent data for all tests on the analyzer. The
screen comprises a test list, a Reagent Details list (on the top right), and an Auxiliary
Reagent list (on the bottom right).
e For more information about the different fields and buttons on this screen, refer to the
Online Help of the particular field or button.
When replacing auxiliary reagent bottles (such as cell detergent or sample probe
detergent bottles), you have to manually reset the software’s reagent volume count.
o A detergent whose expiration date has expired must not be used. Otherwise,
measurement may fail to achieve a correct result.
o Avoid adding a new detergent to the old one. Otherwise, measurement may fail to
achieve a correct result.
o Always replace empty detergent bottles with new and full ones. Otherwise the volume
cannot be correctly determined.
Follow the procedures below to replace auxiliary reagent bottles and reset their
volume count:
Roche Diagnostics
B-140 Operator’s Manual · Version 2.0
cobas c 311 analyzer 8 Reagents
Reagent Status screen
4
3
3 Remove the tubing from the detergent bottle(s) and unscrew the filter from the
tube end.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-141
8 Reagents cobas c 311 analyzer
Reagent Status screen
4 Replace required reagent bottle. Two kinds of 70 mL bottles can be used. Sample
Cleaner 1 and Sample Cleaner 2 bottles must be placed on detergent positions 1
and 2 respectively.
5 Choose Reagent > Status > Reagent Volume Reset to reset the volume count.
6 Choose OK to confirm the reset.
Resetting initial volumes of The remaining volumes of cell detergents and other auxiliary reagents are monitored
auxiliary reagents by counting down from the initial volume of a full reagent bottle. When you place a
new reagent bottle, you always have to manually reset this volume countdown.
Always replace empty detergent bottles with new and full ones. Otherwise the volume
cannot be correctly determined.
Roche Diagnostics
B-142 Operator’s Manual · Version 2.0
cobas c 311 analyzer 8 Reagents
Reagent Overview button
By changing its color, the Reagent Overview button in the System overview screen
indicates potentially insufficient amounts of reagent:
Yellow The number of tests remaining for a reagent is below the yellow
alarm threshold (caution level).
For ISE reagents, the remaining volume is below the yellow alarm
threshold (caution level).
Purple The number of tests remaining for a reagent is below the purple
alarm level.
Table B-28 Colors of Reagent Overview button in Overview area of System Overview screen
Choose the Reagent Overview button to display the Reagent Overview window.
This window displays an overview of the consumables currently loaded on the
analyzer.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-143
8 Reagents cobas c 311 analyzer
Reagent Overview button
Choose System Overview > Reagent Overview to display this screen. It consists of
three areas, the Reagent area, Inventory area, and ISE area.
A
B
C
A reagent mandatory message is displayed on the Reagent Overview screen when the
analytical unit is shown red on the System Overview screen. This may occur if a
mandatory reagent is not on board.
This function is available only, if the test of the missing reagent has been defined as
mandatory under the Utility > Module Set > Test Assignment.
Reagent area Use this area to view the status of the cobas c packs loaded on the analyzer. This area
comprises a graphic representing the reagent disk and a legend.
The graphic is divided into 2 large rings. The 14 segments of the inner large ring and
the 28 segments of the outer large ring represent the 42 positions for cobas c packs.
Each of the segments is subdivided into 3 subrings. The colors of each segment
correspond to the following statuses:
Roche Diagnostics
B-144 Operator’s Manual · Version 2.0
cobas c 311 analyzer 8 Reagents
Reagent Overview button
Legend of the Reagent area The legend at the bottom of the Reagent area explains the colors represented in the
sections of the reagent disk graphic.
1. Inner subring: Cas (Cassette)
Warning Remaining tests in this cobas c pack below yellow alarm level
(Utility > System (Page 2/5) > Reagent Level Check).
Zero tests, Used up The cobas c pack has been used up and is empty. There is still
reagent available in another cassette on this analyzer.
Caution There is no more reagent available for this assay on this
analyzer.
Stand-by This is a standby cassette.(a)
(a) Standby reagents are reagent cassettes already on board but not in use at present. Opposite of active
reagents.
(a) In case QC samples have never been measured, the outer subring is shown as N.A.
Inventory area The Inventory area displays the amount of auxiliary reagents on the analyzer
previously selected on the System Overview screen.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-145
8 Reagents cobas c 311 analyzer
Reagent Overview button
ISE area
A
C
B
D
A Bar-chart for the remaining volume C Color-coded bar for the remaining volume of
B Color-coded bar for the calibration status ISE reagents
D Color-coded bar for the QC status
The ISE area displays the amount of reagents by way of bar-charts. The first bar
below each bar-chart indicates the remaining volume of each ISE reagent. The second
and the third bar below the bar-charts provide additional, color-coded information
on the statuses of calibration and QC.
1. Remaining volume of ISE reagent:
(a) In case QC samples have never been measured, this is shown as N.A.
Roche Diagnostics
B-146 Operator’s Manual · Version 2.0
cobas c 311 analyzer 8 Reagents
Reagent Overview button
Detail Choose the Detail button to display the Detail window. This window displays more
detailed information about the selected cassette, namely: Reagent Type, Test Name,
Remaining Tests, Calibration Date, Calibration Type, Reagent Lot No., Reagent
Sequence Number, First Registration Date, and Expiration Date.
Date format
All dates are displayed according to the selected Set Date/Time setting on the Utility >
System screen.
Month and Year of expiration are indicated by 2 digits each. The format can be mm/yy or
yy/mm depending on the Set Date/Time setting on the Utility > System screen.
Example: For the settings dd/mm/yy and mm/dd/yy, the Expiration Date format is mm/yy.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-147
8 Reagents cobas c 311 analyzer
Reagent Overview button
Roche Diagnostics
B-148 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Table of contents
Calibration
This chapter provides descriptions of special tasks that are not usually part of the
daily workflow. It is meant to complement the chapter Daily operation, where
everyday tasks and common procedures for running the cobas c 311 analyzer are
described.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-149
9 Calibration cobas c 311 analyzer
Table of contents
Roche Diagnostics
B-150 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration concept
Calibration concept
Calibration The following section gives an overview of the calibration concept applied by the
cobas c 311 analyzer; it provides you with a basic understanding of the various
calibration functionalities implemented on the analyzer and its software.
e For instructions on how to perform calibrations, see Calibration And QC Select button on
page B-58.
What is calibration? Calibration is the process that establishes a relation between measurement values
(such as absorbance values) and corresponding results (concentration of an analyte).
This process may comprise the complete setting up of a new calibration curve or it is
only the update of one or two parameters of an existing calibration curve.—In either
case the term calibration is appropriate.
Automatic calibration The relation between measurement values and results is subject to various
environment and reagent conditions and may drift in the course of time. Therefore it
is necessary to repeat calibrations regularly. To keep the resulting calibration
management simple and efficient the analyzer automatically recommends
calibrations.
To get an overview of the functions provided by the cobas c 311 analyzer for
automating calibration refer to the following section.
e See Triggers for automatic calibration requests on page B-152.
Manually requesting calibration The operator can manually order a calibration. It is possible to request a single test
for calibration or a group of calibration tests.
If a predefined set of calibration tests are to be carried out at the start of analysis run,
it is advisable to define a start-up setting calibration. The tests for Start up calibration
are defined under Calibration > Status > Start Up Setting.
e See To select tests for start-up group calibration on page B-158.
The following calibration settings are defined under the Utility > Application >
Calib. screen.
e See Description of application parameters - Calib. tab on page B-224.
Validation of calibration After a calibration is performed, it needs to be validated. The analyzer performs this
task by means of automatic calibration checks. Whenever it detects an irregular
condition or result, the calibration is classified as failed, a calibration alarm is issued,
and the calibration is automatically recommended to be repeated.
Calibration masking If a calibration fails one or more quality criteria applied in the validation process,
tests are no longer performed with the affected reagent. This is achieved
automatically by means of the auto masking function.
e For more information, see Calibration masking on page B-155.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-151
9 Calibration cobas c 311 analyzer
Calibration concept
Failed If a calibration fails one or more quality criteria, this calibration gets the status Failed.
Failed calibration data are not available for the corresponding cobas c pack.
Changeover calibration Calibrations have to be performed if certain physical events occur. These include:
o A change in reagent lot (of cobas c packs)
o A change of a cobas c pack (regardless of the lot; if requested by application)
Timeout Calibrations are performed at regular intervals to compensate changes over time in
reagents and in the measurement systems. Timeout calibrations can be performed for
lot timeout and for cassette timeout.
QC-triggered calibration Calibration will be initiated if the QC results are outside the confidence limit. Three
(QC Violation) different controls can be defined.
For each application you have to decide whether calibration should be performed
time-triggered (Timeout) or QC-triggered (QC violation).
Calib Now This automatic calibration function is active only if Preventive Action is activated in
the System Overview screen. If a calibration runs out of time during the period
specified in the Remaining Time box on the Calibration > Status screen, the
analyzer recommends a calibration indicating Calib Now as a calibration trigger.
e For information on the necessary QC violation settings, see Auto Calibration on
page B-225.
Roche Diagnostics
B-152 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration concept
Calibration concept
Types of calibration curves There are six different types of calibration curves possible on a cobas c 311 analyzer.
That is, there are six different types of mathematical functions possible to describe
the relation between measurement value and result. In this document and in the user
interface (UI), these types of calibration curves are called calibration types. The
names of the calibration types are:
Calibration curve parameters A specific calibration curve is defined by its calibration type (mathematical function)
and its parameters. The names of these parameters in the user interface are:
o S1Abs., K, A, B, C, (L, H, I only for serum indices)
A Linear calibration curve, for example, is defined by two parameters (S1Abs. and K),
an RCM calibration curve is defined by four, and a Spline can require up to six
parameters. When the analyzer performs a calibration, it redetermines these
parameters to fit the calibration curve to new measurement values.
For the Spline calibration curve the calculation are neither printed out nor displayed on the
Calibration Result screen. Only S1Abs. is displayed.
Calibration methods Calibrations are performed with varying numbers of calibrators. Up to six calibrators
are used for a full calibration of certain photometric tests. However, not all calibrators
available for a test need to be used in every calibration. To define which calibrators
are used there is a choice of up to four different calibration methods.
The availability of calibration methods depends on the kind of test that has to be
calibrated. The following table displays all calibration methods and the
corresponding calibrators.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-153
9 Calibration cobas c 311 analyzer
Calibration concept
ISE calibrators can be used with or without barcodes. The labels for the ISE Standards 1
and 2 (ISE Low and ISE High) can be found on the back of the boxes. The label for the
ISE Comp. can be found inside the box.
o Calibrations are carried out in triplicate.
Full calibration Full calibration for Na+, K+, and Cl- requires the following 3 calibrator solutions:
o ISE Standard 1 (ISE Low)
o ISE Standard 2 (ISE High)
o ISE Standard 3 (ISE Comp.)
o ISE Standard 3 (ISE High) for US only
The slope of the calibration curve is calculated from ISE Standards 1 and 2. ISE Low
and ISE High are aqueous standards. ISE Comp. / ISE High (compensated) is
designed to reduce matrix effects. A full calibration is required every 24 hours.
One-point calibration The ISE internal standard (ISE IS) is measured before and after each routine sample
(only one measurement for sample analyses in succession). These measurements are
used to correct for system-related drifts (junction potential differences, differences in
electrode conditions, etc.).
The ISE IS is also measured during calibration as well.
Sample type specific ISE It is possible to set up and use two independent sets of calibration curves (ISE-A and
calibration curves ISE-B) for different sample types for Na+, K+, and Cl-.
If only one calibration curve is used for the different sample types, only one
calibration is performed.
e For information on configuration and assignment of calibrations, see
ISE calibration on page B-173
Roche Diagnostics
B-154 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration concept
Lot calibration Lot calibration data are reagent lot specific for the test. Lot calibration data are
transferred to other cobas c packs belonging to the same reagent lot. The lot
calibration data are applied to result calculations of all patient samples and controls.
A lot calibration can only be generated from a fresh cobas c pack, which means that
the calibration must be performed within 24 hours after the cassette has been loaded
on the analyzer.
Cassette calibration Cassette calibration data are cassette-specific. A cassette calibration is generated if a
cobas c pack is calibrated which is loaded on the analyzer more than 24 hours ago
and all calibration quality criteria are met.
Most of the photometric tests need to be calibrated only if the reagent lot is changed
(lot calibration). Some of the tests, however, need to be calibrated whenever a new
cobas c pack is inserted (cassette calibration).
Newest calibration Newest calibration is a function that permits to replace empty cobas c packs without
performing a calibration.
For a reagent cassette for which no applicable calibration data exists, the newest
calibration result is transferred at the time of reagent registration. This function is
used to avoid the following situation: A cassette of a new lot, that was loaded on the
system, does not have any calibration data. Always the newest calibration data (the
data of the last valid lot calibration) are transferred.
Calibration masking
Failed calibrations If a calibration is successful, the calibration data are available for the measurement of
patient samples and controls. If a calibration fails one or more quality criteria, this
calibration gets the status Failed. Failed calibration data are not available for the
corresponding cobas c pack.
If automatic calibration masking is activated, the corresponding test with the Failed
calibration data is masked. If automatic calibration masking is not activated, the test
is not masked but data alarms are attached to the results.
The Failed calibration can be rejected by the operator (Reject button on
Calibration > Status). In this case, the last successful calibration will be used.
However, the original calibration recommendation still remains for the cobas c pack.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-155
9 Calibration cobas c 311 analyzer
Calibration Screen Overview
Calibration The Calibration menu contains three screens: Calibration > Status, Calibration >
Calibrator, and Calibration > Install. This chapter describes important commands
accessed via these screens as well as various information displayed on them.However,
this chapter does not cover all possible commands. For a complete description of all
fields of the software’s user interface, refer to the Online Help.
e For more information about the calibration menu, refer to the Online Help.
The following sections explain certain tasks associated with the Calibration > Status
screen. Not all tasks are described but the most important ones.
e For a complete description of all elements of the user interface, refer to the Online Help.
e For information about particular tasks associated with this screen, see:
Requesting and cancelling calibrations manually on page B-157
Reviewing calibration data on page B-159
Selecting tests for start-up calibrations on page B-158
Calibration Factors on page B-165.
Exclamation Point
This column header turns yellow if the displayed status list should be updated. The
displayed information may be out-of-date, e.g., if a calibration has been performed in the
meantime. Choose the Exclamation Point to update the status list.
Roche Diagnostics
B-156 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration Status screen
e For more information about the different fields and buttons on this screen, refer to the
Online Help of the particular field or button.
According to the intended use, calibrations will be automatically requested from the
analyzer and from the operator via the System Overview screen. Regardless of these
requests from the analyzer calibrations can be manually selected and deselected as
well.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-157
9 Calibration cobas c 311 analyzer
Calibration Status screen
The start-up calibration function allows you to select a number of tests and treat
them as a group. Tests selected for start-up calibration will be requested for
calibration when the Start Up button is chosen either from the Calibration and QC
Select window, or directly from Calibration > Status.
The set of tests that are selected in a start-up calibration group, including the
respective calibration methods, is defined under the Calibration > Status > Start Up
Setting window.
The selected test is highlighted in blue and the calibration method is highlighted in
green. For ISE assay, specify Full.
7 Choose Update.
8 Repeat steps 2-7 for each group.
9 After completing settings for all of the groups, choose OK.
10 Select Start Up Group from the drop-down box and choose Start Up on the
Calibration > Status screen to activate the selected calibrations or from the
Calibration and QC Select button on the System Overview screen.
11 Choose Save to confirm the start-up calibrations.
Roche Diagnostics
B-158 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration Status screen
Roche Diagnostics
Operator’s Manual · Version 2.0 B-159
9 Calibration cobas c 311 analyzer
Calibration Status screen
S1Abs., K, A, B, and C = Calibration curve parameters (S1Abs. and K for linear calibration
curves)
This window gives information about the most recent calibrations for the
displayed photometric tests.
o To display the calibration curve of a selected test choose Working
Information.
o To display the calibration factors for the cassette calibration, lot calibration,
and newest calibration of a selected test choose Calibration Factor.
e For more details about calibration factors, see
Calibration curve parameters on page B-153
Calibration Factors on page B-165.
Roche Diagnostics
B-160 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration Status screen
4 Use the drop-down list above the graph to select a specific measurement. The
extensions 1st and 2nd in the list correspond to the first and second measurement
of each calibrator, respectively.
e For a complete description of all fields and buttons, refer to the Online Help.
e For background information about photometric principles, refer to the COBI CD.
The measuring points of the reaction monitor of a calibration can be printed out: First,
select a test from the Calibration > Status screen, then, choose Print (global button),
select Reaction Monitor from the list on the left, and choose Print.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-161
9 Calibration cobas c 311 analyzer
Calibration Status screen
This window displays a graph showing calibration results for both the Std (1)
calibrator and the calibrator with the maximum concentration on one plot.
The left and right y-axes are scaled independently:
o The left axis refers to the Std (1) calibrator represented by .
In case of a rate assay, it indicates the primary wavelength absorbance at
photometric measuring point 1 of Std (1).
o The right axis refers to the maximum calibrator represented by .
e For a complete description of all fields and buttons, refer to the Online Help.
The calibration trace list can be printed out. First, select a test from the Calibration >
Status screen, then, choose Print (global button), select Calibration Trace from the list
on the left, enter date range, and choose Print.
Roche Diagnostics
B-162 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration Status screen
The entries ISE-A and ISE-B refer to two different calibration curves that can be
assigned to different sample types. For example, ISE-A is assigned to serum/plasma
and ISE-B to urine samples. This assignment is set on Utility > System (Page 3/5) >
ISE Calib Setting.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-163
9 Calibration cobas c 311 analyzer
Calibration Status screen
This window displays information about the most recent successful ISE
calibration: Electromotive force (EMF), slope, and target concentration values.
The three Compensated Value text boxes display the difference between the
target value of the ISE compensator [ISE Std (3)] and its measured values for Na+,
K+, and Cl-. This difference is added to all measured routine samples and
controls. This is done to compensate for the fact that the internal standard
(ISE IS) and the calibrators Std (1) and Std (2) are aqueous solutions whereas the
patient samples being measured are serum-based solutions.
The slope of the calibration curve is calculated from ISE Standards 1 and 2.
ISE Comp. is designed to reduce matrix effects. It only affects the intercept and
not the slope.
For US only, ISE High (compensated) is used instead of ISE Comp.
Roche Diagnostics
B-164 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration Status screen
Calibration Factors
The term calibration factors refers to coefficients (S1Abs, K, A, B, C) used for setting
up calibration curves for photometric tests. The analyzer stores calibration factors for
each calibration curve of all registered cobas c packs.
The following kinds of calibration factors are available: Cassette, Lot, and Newest.
Lot calibration factors Lot calibration factors are reagent lot specific for the test. Lot calibration factors are
transferred to other cassettes belonging to the same reagent lot. The lot calibration
data are applied to result calculations of all patient samples and controls.
A lot calibration can only be generated from a fresh cobas c pack, which means that
the calibration must be performed within 24 hours after the cassette has been loaded
on the analyzer.
Cassette calibration factors Cassette calibration factors are cassette-specific. A cassette calibration is generated
when a cobas c pack is calibrated which is older than 24 hours after being loaded on
the analyzer for the first time and all calibration quality criteria are met.
Most of the photometric tests need to be calibrated only when the reagent lot is
changed (lot calibration). Some tests, however, need to be calibrated whenever a new
cobas c pack is inserted (cassette calibration).
Newest calibration factors Newest calibration is a function that allows you to replace empty cobas c packs
without performing a calibration (if no changeover calibration is recommended for
this application).
For a reagent cassette for which no applicable calibration data exists, the newest
calibration is transferred at the time of reagent registration. The newest calibration is
always the last generated lot calibration of this test. This is used to prevent that a
cobas c pack of a new lot loaded on the analyzer does not have any calibration data.
Always the newest calibration data are transferred.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-165
9 Calibration cobas c 311 analyzer
Calibration Install screen
The following sections contain certain tasks associated with the Calibration > Install
screen. Not all tasks are described but the most important ones.
To display this screen choose Calibration > Install.
The Calibration > Install screen is used to review information about currently
registered calibrators, to update calibrator data, and to install new calibrators.
e For more information about the different fields and buttons on this screen, refer to the
Online Help of the particular field or button.
Roche Diagnostics
B-166 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration Install screen
In case calibrators are used without barcodes, assign the calibrator positions manually.
3 Select one of the available search options (Required Item, Calibrator Name, Lot
Number, or Release Date).
o Lot Number is the recommended search condition.
o The Lot Number must always be entered with 8 digits. The last two digits are
zeros (XXXXXX00).
o The option Release Date refers to the date when the calibrator was released by
Roche Diagnostics.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-167
9 Calibration cobas c 311 analyzer
Calibration Install screen
4 Choose Search to start searching for the selected criteria. The search results are
displayed.
In the column Note you can find additional information about the calibrator such
as reason for change.
To change the order of the list select the Sort By option.
5 Select the Selection check box of each list item to be downloaded.
6 Choose Download and confirm with OK.
Roche Diagnostics
B-168 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration Install screen
In case calibrators are used without barcodes, assign the calibrator positions manually.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-169
9 Calibration cobas c 311 analyzer
Calibration Install screen
The concentration value of a calibrator can be edited on the Calibration > Install
screen. You can use this function to define the number of decimal places for Std (1)
used for reportable data.
4 Select a test (analyte) from the list whose calibrator concentration value is to be
entered or edited.
5 Select a box in the Std Concentration area and type the appropriate
concentration value. Observe the respective unit of measure displayed below the
Std Concentration area.
Std (1) is used for the blank calibrator. Std (2) through Std (6) are used for all
other calibrators.
The number of decimal places used for Std (1) defines the number of decimal places
used for reportable data.
It is recommended to select the number of decimal places at the time of installing the
application.
Roche Diagnostics
B-170 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration Install screen
1 Backup the database and delete patient data (System Overview > Sample Data
Clear > Backup and Clear).
If Sample data clear is executed, all the records of the samples are deleted and QC
data are moved to the QC View.
e For details see
Archiving patient data on page B-106
Sample Data Clear button on page B-50.
2 Unload all the cobas c packs for this test (Reagent > Setting > Unloading).
e For details see Unloading cobas c packs on page B-57.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-171
9 Calibration cobas c 311 analyzer
Calibration Install screen
Any cobas c packs for this test that were already in use on the analyzer cannot be
reloaded onto the analyzer after the test has been deleted.
6 Install the calibrator values for this test under Calibration > Install > Download.
e For details see Loading calibrator data on page B-167.
7 Change the decimal places of calibrator 1 (Std (1) (Calibration > Install > Edit).
8 Install the control values for this test under QC > Install > Download.
e For details see Loading control data on page B-203.
If concentration values for a particular calibrator have been reassigned, you are
informed by means of the cobas e-library (Value Sheets or Important Notes). In this
case you have to reassign the calibrator values on the cobas c 311 analyzer.
2 Perform a new lot calibration for the current reagent lot and for all previously
registered reagent lots that will be reused.
e See Requesting and cancelling calibrations manually on page B-157
3 In addition perform a cassette calibration for the currently used cobas c pack and
for all temporary unloaded cobas c packs.
e See Requesting and cancelling calibrations manually on page B-157
Roche Diagnostics
B-172 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration Install screen
ISE calibration
ISE calibrators low and high can be used with or without barcode labels. Without a
barcode you have to register the calibrators, assign a calibrator position to them, and
enter the relevant concentration values for each calibrator as described above.
When the analyzer configuration is completed, the three calibrators —ISE Low,
ISE High, and ISE Comp.—used in an ISE calibration must always be placed in their
assigned positions.
For US only, ISE High (compensated) is used instead of ISE Comp.
e For details see:
Requesting and cancelling calibrations manually on page B-157
Editing concentration values on page B-170
Assigning calibrator positions on page B-175.
Two ISE calibration curves For ISE tests it is possible to apply one of two independent calibration curves. After
installation and assignment, the respective calibrator values are displayed under
ISE-A and ISE-B on Utility > Application > Other.
The assignment of one or the other calibration curve to a certain sample type is set on
Utility > System (Page 3/5) > ISE Calib Setting.
Figure B-85 Utility > System (Page 3/5) > ISE Calib Setting window
Roche Diagnostics
Operator’s Manual · Version 2.0 B-173
9 Calibration cobas c 311 analyzer
Calibration Calibrator screen
The following sections contain certain tasks associated with the Calibration >
Calibrator screen. Not all tasks are described, but the most important ones.
To display this screen choose Calibration > Calibrator.
The Calibration > Calibrator screen is used to review information about currently
registered calibrators and to manually assign calibrators to specific positions. The
position assignment is necessary for calibrators that are used without barcode
identification.
e For information about particular tasks associated with this screen, see:
Assigning calibrator positions on page B-175.
e For more information about the different fields and buttons on this screen, refer to the
Online Help of the particular field or button.
Roche Diagnostics
B-174 Operator’s Manual · Version 2.0
cobas c 311 analyzer 9 Calibration
Calibration Calibrator screen
Use the following procedure to assign a calibrator to a position if you are using cups
without a barcode, or if the analyzer cannot read the barcode on the calibrator.
2 Select the calibrator to be assigned to a position from the list on the left.
3 Select a position from the list on the right. The position selected for the
assignment has to be an empty line in the list.
4 Choose Assign to assign the selected calibrator to the selected position.
To undo the assignment choose Remove or choose Cancel to leave the Calibrator
Position Assignment window without saving any changes.
5 To assign further calibrators repeat steps 2 to 4.
6 After all necessary assignments are completed, choose OK to save the changes.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-175
9 Calibration cobas c 311 analyzer
System Overview screen
The following section explains calibration tasks associated with the System Overview
screen.
Choose the graphical icon representing the configuration in the status line (top left of
the screen), or press F12 to display the System Overview screen.
e For a detailed description of the calibration process at the beginning of routine operation,
see:
Calibration And QC Select button on page B-58
Requesting calibration and QC and printing load lists on page B-59
Measuring calibrators and controls on page B-62.
Roche Diagnostics
B-176 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
Table of contents
QC
This chapter explains certain tasks associated with the QC menu. Not all tasks are
described but the most important ones.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-177
10 QC cobas c 311 analyzer
Table of contents
Roche Diagnostics
B-178 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC concept
QC concept
QC methods
The following QC methods are available on a cobas c 311 analyzer: Individual (intra-
day) QC, cumulative (long-term) QC, and realtime QC.
Individual and cumulative QC All results of QC measurements can be viewed on the QC > Run Status screen and
the QC > Individual screen. It is recommended to transfer QC data at the end of a
day from the individual QC to the cumulative QC. Under cumulative QC you find all
the long-term quality control data stored on the analyzer.
e For more information see:
QC Run Status screen on page B-188
QC Individual screen on page B-190
QC Cumulative screen on page B-199
e For more details on the accumulation of QC data (from individual to cumulative QC), see:
Accumulation of QC measurement data on page B-181
Accumulation of QC results on page B-198.
Realtime QC Independent of individual and cumulative QC, the realtime QC function allows the
evaluation of QC measurement immediately after their results become available
(realtime) utilizing the Westgard algorithm.
Realtime QC for one test always uses two kinds of controls and compares QC results
against the controls’ known standard deviation (SD) and mean values.
e For more information on realtime QC, see Configuring and using realtime QC on
page B-193.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-179
10 QC cobas c 311 analyzer
QC concept
Routine QC Each test has one or more controls assigned to it. Moreover, a test must be not only
assigned to a control but also activated for that control to make a QC measurement
possible. Routine QC comprises all activated tests of all installed controls. You can
request a QC measurement for all of these tests—for example at the beginning of a
work shift—with one single command (System Overview > Calibration and QC
Select > Routine QC).
e For more information about routine QC, see:
Requesting QC measurements on page B-186
To perform controls for active reagents on page B-186
Standby reagent QC QC measurements can be requested individually for standby cobas c packs. Standby
cobas c packs are already on board but not in use at present.
e For more information about standby reagent QC, see:
Requesting QC measurements on page B-186
To perform controls for standby reagents on page B-186
Automatic QC It is possible to keep frequently used controls on the sample disk—ready to be used
any time before or during routine operation. Thus, in combination with Timeout
QC, the analyzer is able to perform QC measurements without any intervention by
the operator.
e See Programming auto QC measurements on page B-209.
QC after calibration For this kind of QC measurements no special settings are needed. QC measurements
are performed for all calibrated tests without explicit requests from the operator,
when the control is activated and on board.
e See To perform QC after calibration on page B-187.
Manual QC This function allows you to measure QCs of any test based on your judgment.
e See Requesting QC measurements on page B-186.
Definitions concerning QC
This section explains some QC expressions that are used in the following:
o Active cobas c packs are currently in use. Sometimes called current reagents. The
opposite are standby cobas c packs.
o Controls or control samples are used synonymously for quality controls (QCs).
Controls are special samples that are used for QC.
Roche Diagnostics
B-180 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC concept
Two accumulation methods There are two accumulation methods available: Mean-R and X-R. The choice for one
or the other method is made under Utility > System (Page 1/5) > QC Setting.
o X-R transfers only one QC result from individual QC to cumulative QC.
Therefore you must specify a measurement when accumulating QC data.
o Mean-R calculates cumulative QC data on the basis of all individual QC results
being accumulated.
Two calculation methods If Mean-R is set as accumulation method, two methods are available for the
calculation of mean values and SD values in cumulative QC: DayToDay and Overall.
o The Overall method uses a weighting factor for each accumulated set of data.
o In the DayToDay calculation method all accumulations have the same weight.
The calculation method is selected under Utility > System (Page 1/5) > QC Setting.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-181
10 QC cobas c 311 analyzer
QC concept
Handling of QC results
The following chart shows the handling of QC results starting with the measurement
of QC samples and ending with the deletion of QC results. It shows the different
screens that display QC results, and shows the particular action that moves (deletes,
accumulates) the results from one screen to another screen.
QC measurements
displayed in… displayed in…
performed
Delete All
transfer to…
transfer to…
o Deletes QC in QC View
o Deletes QC in QC Individual
Recommendation
Workflow for handling QC results in order to keep as little QC data as possible
Roche Diagnostics
B-182 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC concept
Summary QC View is linked with the QC individual database. If a QC result is deleted in the
QC View database, it is as well deleted in the QC individual database and vice versa.
Therefore it is recommended to first accumulate individual QC results before
deleting QC data on Data Review > QC View.
QC data of the QC cumulative database should only be deleted, if the particular data
is no longer required.
On the Utility > System screen under Barcode Setting, you can verify or change
whether the analyzer uses barcodes for controls or not.
When working with barcodes, controls are automatically identified. In case controls
are processed in non-barcode mode, it is necessary to assign a position for each
control.
The current position assignment is listed on the QC > Control screen. To provide
additional assignments or remove existing assignments choose Position
Assignment.
o For assigned controls (without barcode), be sure to use the specified position. A
misplacement would go undetected by the analyzer and yield invalid results.
o Do not use barcoded and assigned controls on the same QC position.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-183
10 QC cobas c 311 analyzer
QC Status screen
QC Status screen
The following sections explain certain tasks associated with the QC > Status screen
screen. Not all tasks are described but the most important ones.
To display this screen choose QC > Status.
The QC > Status screen displays all installed controls with the corresponding
assigned and activated tests.
The test display can be sorted by test name (alphabetical order), control name
(alphabetical order), QC test selection, or QC cause. Tests can be selected with the
Select/Deselect button. Selected tests are highlighted in green. A control test
selection is made for the subsequent control run. The control test selections of the
status list disappear after the control is measured.
Only the test selection of pipetted control tests disappear. If a test is requested but the
control sample is not on board, the QC request remains.
Routine QC Assign The Routine QC Assign button selects all Routine QC tests (active reagents only) for
a control measurement. Routine QC tests are all activated tests for all installed
controls.
Tests that are not to be measured can be deselected on the Status screen: Highlight
the tests on the status list and then use the Deselect button.
Routine QC To display a list of all currently activated Routine QC tests, choose Routine QC.
Stand By Bottle QC Use this button to select cobas c packs for stand by QC.
Roche Diagnostics
B-184 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Status screen
Stability Timer Reset Use this button to reset the On Board Stability Time (set in Utility > Application >
Range). In case the on board stability time of the control is expired (highlighted
yellow in QC > Status plus a yellow alarm message), the timer can be reset with this
button after the control material is exchanged. This function is only available in
Standby or Pause.
Direct access from System From the System Overview screen, select the Calibration and QC Select button and
Overview screen then the Routine QC button in the QC area on the right. This is essentially the same
as selecting the Routine QC Assign button on QC > Status.
e For more information about the different fields and buttons on this screen, refer to the
Online Help of the particular field or button.
Timeout QC
A control interval in hours can be defined for each application on the Range tab on
Utility > Application. Select the Control Interval Time check box and type the
appropriate time in the adjacent text box.
The analyzer checks the timeout every 30 minutes. When the specified time has
elapsed, a QC request is triggered. This QC request is indicated on the QC > Status
screen by the word Timeout in the Cause column.
Be sure to place an appropriate QC sample on the sample disk in advance if the
control is defined as auto QC.
The control is measured with the active (in use) cobas c pack.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-185
10 QC cobas c 311 analyzer
QC Status screen
Requesting QC measurements
2 Select the appropriate test and control. Multiple tests and controls can be
highlighted.
Roche Diagnostics
B-186 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Status screen
3 Choose Select to select all highlighted tests and controls. A green bar appears in
the Selection column. The Select button toggles to Deselect.
Choosing Deselect deselects a highlighted control and the Deselect button toggles
to Select.
You can also print a QC load list from System Overview > Calibration and QC
Select. Select the QC option in the Load List area.
Calibration or QC failure
If you have to manually assign a calibrator or QC container or other sample
CAUTION container—for example, due to an unreadable barcode—do not place any containers
with barcodes in manually assigned positions.
It is possible to do a QC with a mixture of barcoded and non-barcoded QCs. However,
you must assign the position of the non-barcoded QC and no position should be
assigned for the barcoded QC. If you put a barcoded container on a position that was
assigned manually, the QC is not measured.
If you perform QC after calibration, QC samples do not have to follow directly after
calibrator samples. Other samples can be placed in between them.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-187
10 QC cobas c 311 analyzer
QC Run Status screen
Use the Run Status screen to get an overview of the last 500 quality control (QC)
measurements performed on the analyzer or to view the most recent QC
measurements for one particular test specified.
Use the Test box to select all tests or one specific test to be displayed on the chart. Use
the scroll bar at the bottom of the chart to move the focus from one result to the next.
e For more information about the different fields and buttons on this screen, refer to the
Online Help of the particular field or button.
QC run status details The Run Status screen consists of a details area in the upper part of the screen and a
run status chart in the center of the screen. The details area displays information that
relates to the QC measurement selected on the run status chart. The chart in the
center of the Run Status screen displays up to 500 results of QC measurements.
Chart background A white background refers to today’s measurements. A blue background refers to
past days’ measurements.
Roche Diagnostics
B-188 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Run Status screen
Chart symbols Shape: Circles are used if QC violation is activated for a test in Utility >
Application > Calib; otherwise triangles are used.
e For more information on QC violation, see QC-triggered calibration (QC Violation) on
page B-152.
Color: Red denotes that a QC result falls outside the confidence limits, and yellow
denotes that it is outside 2 SD limits. Green symbols are used for results within the
relevant limits.
Special fields in the details area Status: This field displays a reagent’s status of precedence. Possible values are
Current, SB1, SB2, SB3…
The order of precedence of standby cobas c packs within the same lot is assessed by the
time of registration; the oldest standby cobas c pack is used first.
cl-low / cl-high: These fields display the confidence limits specified for the selected
test. If the selected test is not configured for QC violation but for timeout calibration,
the fields under cl-low and cl-high are blank.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-189
10 QC cobas c 311 analyzer
QC Individual screen
QC Individual screen
The following sections explain certain tasks associated with the QC > Individual
screen. Not all tasks are described but the most important ones.
To display this screen choose QC > Individual.
e For information about particular tasks associated with this screen, see:
To exclude an individual QC point and add a comment on page B-192
e For more information about the different fields and buttons on this screen, refer to the
Online Help of the particular field or button.
Roche Diagnostics
B-190 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Individual screen
Individual QC Chart window On the QC > Individual screen, choose Chart to open the Individual QC Chart
window.
This window displays all non-accumulated QC results of the selected test in a Levey-
Jennings chart. A Levey-Jennings chart provides a visual method of monitoring
trends in QC results. The result values are scaled to the standard deviation (SD) of the
respective test.
Up to three controls can be selected (Select button) and displayed on one chart.
Comment Use the Comment button to comment, exclude, or comment and exclude a single
measurement. The chart uses three special symbols with the following meaning:
A comment is attached to this point (in QC >Individual QC > Chart > Comment).
The point is excluded and a comment is attached (in QC >Individual QC > Chart >
Comment).
o Excluded points are removed from statistics but still displayed on the chart. Excluded
points cannot be included later.
o Control values of standby cobas c packs are always excluded.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-191
10 QC cobas c 311 analyzer
QC Individual screen
Roche Diagnostics
B-192 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Individual screen
Quality control by the Shewhart The instrument measures two controls (X and Y) in pairs, whose mean value ( X ) and
method standard deviation (SD) are known for each test. Each QC result is processed
according to the judgment standard (multi-rule) and it is judged whether it can be
reported or not.—When the algorithm yields an error, an alarm is generated
indicating whether the error is a random error, a QC error, or a systematic error.
The kinds of realtime QC samples are 100 at a maximum. The following figure shows
an illustration of the multi-rule chart. The abbreviations used in this figure are
described in Table B-31 on page B-194.
No
1-2SD Inside control range
Yes
No
No No No No No No No No
1-2.5SD 1-3SD 2-2SA R-4SD 2-2SW 4-1SA 4-1SW 10XA 10XW
Configuration of realtime QC Configure realtime QC in Realtime QC window (QC > Individual > Realtime QC):
Choose Select to specify control (X) and control (Y) for each relevant test. Then,
choose Rules to select the rules.
The following table briefly explains all rules that are selectable for realtime QC and
lists the corresponding alarms.
e For more details, refer to the COBI CD.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-193
10 QC cobas c 311 analyzer
QC Individual screen
Table B-31 QC evaluation by Shewhart multi-rule method and corresponding alarms (Sheet
1 of 2)
Roche Diagnostics
B-194 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Individual screen
-1SD
8 The last four X or Y 4-1SW S4-1Sw
Y
data values are above
-1SD
1SD
+1SD or below -1SD.
1SD
X
-1SD
Table B-31 QC evaluation by Shewhart multi-rule method and corresponding alarms (Sheet
2 of 2)
Deselecting realtime QC In cases where quality control using one QC sample is conducted instead of realtime
QC, specify None for X and Y corresponding to each analytical item and enter 0 for
SD. If in QC > Install the SD value is 0, a realtime QC check is not performed.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-195
10 QC cobas c 311 analyzer
QC Individual screen
Configuring realtime QC
This window shows a diagram with the QC results for a selected test and a selected
pair of controls. The diagram is scaled to the target SD of control (X) and control (Y),
respectively. The newest QC result is shown in red.
Select Choose this button to assign two controls—control (X) and control (Y)— to be
evaluated by the multi-rule method.
Roche Diagnostics
B-196 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Individual screen
Rules Choose this button to select rules used to evaluate the QC results. A pair of controls is
compared against a known standard deviation (SD) and mean. If one or both of the
controls fail a rule, the analyzer continues applying the testing criteria for all selected
rules. When at least one rule violation is found, the corresponding data alarm is
issued.
Reject Choose this button to reject realtime QC results that have specific data alarms
attached. Affected result points are removed from the plot.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-197
10 QC cobas c 311 analyzer
QC Individual screen
Accumulation of QC results
Figure B-98 Utility > System (Page 1/5) > QC Setting window
Accumulation Method Select one of the two accumulation methods, Mean-R or X-R, from this box.
e See Accumulation of QC measurement data on page B-181.
QC Calculation If the accumulation method is Mean-R, select one of the two calculation methods,
DayToDay or Overall, from this box to specify how mean and SD values are
calculated in the cumulative QC.
This box is not available if X-R is selected in the Accumulation Method box.
Accumulating QC results
a To accumulate QC data
1 Select the data to be accumulated from the list on the QC > Individual screen.
2 Choose Accumulate.
3 If Mean-R is set as accumulation method, go to step 4.
If X-R is set as accumulation method, type the consecutive number of the QC
results to be used for the accumulation in the box or type 0 (zero) to transfer the
mean value; then choose OK.
4 In the confirmation window choose OK to accumulate the selected data.
After accumulation the corresponding data are deleted from the QC > Individual
screen and a new standard deviation (SD) and mean value is calculated and
displayed on QC > Cumulative.
Roche Diagnostics
B-198 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Cumulative screen
QC Cumulative screen
e For more information about the different fields and buttons on this screen, refer to the
Online Help of the particular field or button.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-199
10 QC cobas c 311 analyzer
QC Control screen
QC Control screen
The QC > Control screen provides an overview of all controls that are downloaded
and assigned to a position on the sample disk.
When working with barcodes, controls are automatically identified. In case controls
are processed in non-barcode mode, it is necessary to assign a disk position number
for each control. The current position assignment is listed on the QC > Control
screen.
Choose QC > Control to display this screen.
Roche Diagnostics
B-200 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Control screen
2 On the Position Assignment window select the control from the list on the left
and an unassigned disk position from the list on the right.
3 Choose Assign to register the new assignment.
4 To remove an existing assignment select the concerned disk position from the list
on the right and choose Remove.
5 Choose OK to save the changes and close the window.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-201
10 QC cobas c 311 analyzer
QC Install screen
QC Install screen
The following sections explain certain tasks associated with the QC > Install screen
screen. Not all tasks are described but the most important ones.
To display this screen choose QC > Install.
Use this screen to add and delete controls from the analyzer and to assign control
values.
The list on the left of the screen displays all controls that are installed on the analyzer.
For all installed controls the control name is highlighted in green. Up to a 100
controls can be installed on the analyzer.
e For information about particular tasks associated with this screen, see:
Loading control data on page B-203
Editing control values on page B-206
Activating control tests on page B-208
e For more information about the different fields and buttons on this screen, refer to the
Online Help of the particular field or button.
Roche Diagnostics
B-202 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Install screen
Control data for all available applications can be downloaded to the analyzer from
cobas link. An operator ID of supervisor level or higher is required for this operation.
This operation is a part of the installation process for a new application.
e For an overview of the complete installation process of a new application, see:
Loading or updating applications on page B-214
o In case controls are used without barcodes, assign the control positions manually.
o Controls may be added to the analyzer but the control tests will not run until they are
activated on the QC > Install screen.
2 Select one of the available search options (Required Item, Control Name, Lot
Number, or Release Date).
o Lot Number is the recommended search condition.
o The Lot Number must always be entered with 8 digits. The last two digits are
zeros (XXXXXX00).
o The option Release Date refers to the date when the calibrator was released by
Roche Diagnostics.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-203
10 QC cobas c 311 analyzer
QC Install screen
3 Choose Search to start searching for the selected criteria. The search results are
displayed.
In the column Note you can find additional information about the control such as
reason for change.
To change the order of the list select the Sort By option.
4 Select the Selection check box of each list item to be downloaded.
5 Choose Download and confirm with OK.
Roche Diagnostics
B-204 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Install screen
Roche Diagnostics
Operator’s Manual · Version 2.0 B-205
10 QC cobas c 311 analyzer
QC Install screen
Control values can be edited on the QC > Install screen. You can use this function,
for example, if an Important Note concerning reassigned target values for a particular
lot of controls is released. Instead of downloading the control values from cobas link
you may also edit them manually.
5 Choose Calculate to recalculate the confidence limits on the basis of the edited
target and SD values and the QC violation rule (set under Utility > Application >
Calib.).
In case target and SD values are edited or entered manually, the calculation of
confidence limits is achieved by choosing the Calculate button. However, the
confidence limits can also be typed directly into the text boxes under Confidence
Limit.
The Calculate button and the boxes High and Low are not available if the selected
test is not configured for QC Violation.
Roche Diagnostics
B-206 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
QC Install screen
If target values for a particular control have been reassigned, you are informed by
means of the cobas e-library (Value Sheets or Important Notes). In this case you have
to reassign the control values on the cobas c 311 analyzer.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-207
10 QC cobas c 311 analyzer
QC Install screen
To activate a test the test must have target and SD values entered for the
control.
e See Editing control values on page B-206.
Roche Diagnostics
B-208 Operator’s Manual · Version 2.0
cobas c 311 analyzer 10 QC
Programming auto QC measurements
When a control interval time is specified on Utility > Application > Range for an
application, it is possible to let the analyzer automatically perform necessary QC
measurements with controls that permanently reside in the sample disk.
At intervals of 30 minutes, the instrument checks whether a preset time is exceeded
or not. If a time period after the last QC measurement exceeds the preset time, the
instrument issues a recommendation of execution of QC measurement.
In case an automatic QC selection was made and quality control samples are on the
sample disk, the instrument will carry out automatic QC measurement by
intervening the current measurement of routine samples.
An on board stability time of QC samples on the instrument can be set for each test.
An alarm will be issued if the on board stability time of QC samples is exceeded.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-209
10 QC cobas c 311 analyzer
Programming auto QC measurements
Roche Diagnostics
B-210 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Table of contents
Configuration
This chapter describes the configuration of the cobas c 311 analyzer for customers.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-211
11 Configuration cobas c 311 analyzer
Table of contents
Roche Diagnostics
B-212 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Application
Application
When the system was installed by your Roche Service representatives all required
applications are installed. This section explains how to add new applications and
configure specific application parameters.
Overview
General information Up to 100 tests (86 photometric tests, 3 electrolyte tests, 8 calculated tests, and
3 serum indices) can be installed on the analyzer.
The following test numbers are fixed:
Loading a new application Before you can load a cobas c pack of a new application onto the analyzer, the
application parameters have to be downloaded from cobas link. Otherwise, the
cobas c pack will be rejected by the system.
e Refer to the following sections to get information about:
cobas link on page A-31
Loading application parameters on page B-215
Description of application parameters on page B-221
Roche Diagnostics
Operator’s Manual · Version 2.0 B-213
11 Configuration cobas c 311 analyzer
Application
The cobas c 311 analyzer requires application parameters, calibrator data, and
control data for each application. All data required for a new application can be
downloaded from cobas link. Existing applications can be updated via cobas link.
Installation Guide To be able to use a new application on the analyzer, you have to perform the
following steps:
1. Load the application parameters from cobas link. Customize the default
parameters if necessary.
e For more information, see:
Loading application parameters on page B-215
Description of application parameters on page B-221
2. Load the calibrator data from cobas link. Or: Check and/or edit the number of
decimal places for Std (1) manually.
e See Loading calibrator data on page B-167
3. Load and activate the control data for the new application.
e See Loading control data on page B-203.
4. Assign sample disk positions for calibrators and controls (only necessary when
working with non-barcoded calibrators or controls).
e For more information, see:
To assign a calibrator to a specific position on page B-175
To assign a calibrator to a specific position on page B-175
To assign a control to a specific position on page B-201
5. Add special washes if applicable.
e See Special Wash on page B-255
6. Load the cobas c pack(s) onto the analyzer.
e See Loading cobas c packs on page B-55
7. Define a key setting for the new application.
e See To assign a test or a profile to a key on page B-244.
8. Define reagent level check alarms if necessary. Choose Utility > System
(Page 2/5) > Reagent Level Check.
e For more information, please refer to the Online Help for that particular window.
Roche Diagnostics
B-214 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Application
2 Select one of the available search options (Required Item, Application Code,
Application Name, Cassette No., or Release Date).
o Application Code: Select the application code (ACN) from the list box.
The application code number (ACN) is available in the instructions for use of
the corresponding cobas c pack. Up to four applications (ACN) per
cobas c pack are possible.
The Application Code is the recommended search condition.
o Application Name: Select the application name from the list box.
o Cassette No.: Enter the Cassette number in the text box.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-215
11 Configuration cobas c 311 analyzer
Application
3 Choose Search to start searching for the selected criteria. The search results are
displayed.
In the column Note you can find additional information about the application
such as reason for change.
To change the order of the list select the Sort By option.
4 Select the Selection check box of each list item to be downloaded.
If the required item is not contained in the list, you can go to the next page by clicking
on the page number (e.g., 1/2) in the upper right corner of the list.
The short test name assigned to the application automatically displays in the Test
Name box. However, it is possible to assign a different test name if needed.
It is also possible to change the channel number. This may be needed to change
the pipetting sequence.
Roche Diagnostics
B-216 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Application
This section describes the necessary steps to utilize HbA1c whole blood application
on the system.
Information on the general installation procedure for a new application is given in
the previous section.
e See Loading or updating applications on page B-214
Overview
Necessary applications To obtain results for HbA1c (%), the following applications are necessary:
Two cassettes are necessary for the HbA1c test. One cobas c pack contains the
hemolyzing reagent (application code number 952). The application code numbers
871 and 881 refer to the second cobas c pack which contains reagent for both
applications (Hb-W3 and A1-W3).
Roche Diagnostics
Operator’s Manual · Version 2.0 B-217
11 Configuration cobas c 311 analyzer
Application
Reaction course The following figure shows the reaction course of the application:
70%
Cycle (time)
0 6 23 57
R3
S, R1
0 34 Cycle (time) 70
Result calculation The result for the Hemoglobin A1c analysis is calculated according to IFCC
standardization:
A1-W3 HbA1c
Equation B-1 RWI3 = ----------------- 100% (that means HbA1c % = ------------------ 100% )
HB-W3 Hb
This calculated test can be downloaded as application RWI3 (Application code 891)
Roche Diagnostics
B-218 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Application
5 Program a calculated test for the HbA1c values according to Protocol 1 (IFCC).
Use following settings:
o Sample Type: Suprnt
o Unit of Measure: mM/M
o Report Name: HbA1c Gen.3 IFCC
o Item: RWI3
o Formula: (A1-W3 / HB-W3)*1000
e For more information, see
Programming calculated tests on page B-250
Roche Diagnostics
Operator’s Manual · Version 2.0 B-219
11 Configuration cobas c 311 analyzer
Application
8 Assign the RWI3 test to a key on the Suprnt tab of the Key Setting window.
e For more information, see:
To assign a test or a profile to a key on page B-244
Result calculation according to Roche Diagnostics provides the application RWI3 (Application code 891) to calculate
DCCT/NGSP reporting range the Hemoglobin A1c% results according to DCCT/NGSP reporting range.
A1-W3
Equation B-2 RWD3 = ----------------- 91.5 % + 2.15
HB-W3
To report the results according to IFCC reporting range, you have to configure the
calculated test ACN 891 (Formula tab of the Calculated Test Formula window):
A1-W3
Equation B-3 RWD3 = ----------------- 1000
HB-W3
Downloading control data for Control data for HbA1c Control N (357) and HbA1c Control P (358), or
IFCC reporting range PreciControls Hba1c Norm/Path are downloaded with values for the DCCT/NGSP
reporting range. The control data must be manually updated to the IFCC reporting
range values. These values can be obtained from the control value sheets.
e See Editing control values on page B-206
Sample container, sample For processing the HBA1c whole blood application only standard tubes with the
pipetting following specifications can be used:
o 13 mm x 75 mm
o 13 mm x 100 mm
o 16 mm x 75 mm
o 16 mm x 100 mm
For HbA1c whole blood QC measurements, sample cups directly placed on a sample
disk can be used.
For the use of blood collection systems as sample tubes, please contact your technical
support.
Clot detection Clot detection is automatically deactivated when pipetting whole blood samples.
Roche Diagnostics
B-220 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Application
e For instructions on changing default application parameters on the Calib. tab, see:
To define calibration timeouts on page B-226
To define application auto masking on page B-225
e For instructions on changing default application parameters on the Range tab, see:
To change the unit of measure for an application after calibration on page B-228
To define application report name on page B-229
e To change the technical limit on page B-230
To change the repeat limit on page B-230
To change qualitative fields (photometric tests) on page B-231
To change expected values on page B-232
To change default settings on page B-233
Roche Diagnostics
Operator’s Manual · Version 2.0 B-221
11 Configuration cobas c 311 analyzer
Application
The following sections explain certain settings associated with the Analyze tab on
Utility > Application screen. Not all settings are described, but the most important
ones.
To display this screen, choose Utility > Application > Analyze.
Figure B-112 Analyze tab on Utility > Application screen (photometric tests)
Roche Diagnostics
B-222 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Application
Sample Volume area It is possible to perform automatic dilution on patient samples. Usually the first run is
performed with the normal sample volume settings.
The dilution settings are displayed as follows:
A B C
Photometric test
A Sample volume [L] (undiluted)
B Volume of the diluted sample [L]
C Diluent volume [L]
If a dilution is selected on the Workplace > Test Selection screen, it takes priority.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-223
11 Configuration cobas c 311 analyzer
Application
The following sections explain certain settings associated with the Calib. tab on
Utility > Application screen. Not all settings are described, but the most important
ones.
To display this screen, choose Utility > Application > Calib.
Figure B-114 Calib. tab on Utility > Application screen (Photometric test)
Figure B-115 Calib. tab on Utility > Application screen (ISE test)
e For instructions on changing default application parameters on the Calib. tab, see:
To define application auto masking on page B-225
To define calibration timeouts on page B-226
Roche Diagnostics
B-224 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Application
Auto masking
If Auto Masking is selected, the particular test requiring calibration due to a failed
calibration will be masked automatically.
To activate the auto masking function, select the Auto Masking check box on Utility >
System (Page 2/5) > Calib Mask Setting.
Auto Calibration
Calibration can be requested automatically in the following cases:
o Changeover (Cassette or lot change): If a new cobas c pack or a new reagent lot is
loaded on the analyzer.
o Timeout: If the calibration time period for a cobas c pack or the reagent lot has
expired.
o QC Violation: If a QC result falls outside the confidence limit.
When you load a new application onto the system, the changeover- and time-
triggered calibration is activated by default. If you intend to work with QC-triggered
calibration, you must configure the QC violation parameters and activate QC-
triggered calibration (QC Violation).
e For more information, see:
To activate QC violation on page B-225
To define calibration timeouts on page B-226
a To activate QC violation
1 On the Calib. tab of the Utility > Application screen select the test to be edited
from the Test list on the left.
2 Select QC Violation.
3 Select the calibration update method (Blank, Span, 2 Point, or Full) from the
Method list box.
e For more information about calibration methods, see:
Calibration concept on page B-151
Roche Diagnostics
Operator’s Manual · Version 2.0 B-225
11 Configuration cobas c 311 analyzer
Application
4 Select the rule for checking the QC measurement from the Rule list box.
Calibration will be recommended if a QC result falls out of the specified range
(expressed in multiples of one standard deviation).
5 Select the controls that should trigger a calibration.
Up to three controls can be assigned. If any of the QC results falls out of the
specified range, a new calibration is recommended.
For each of the assigned controls, an individual confidence interval must be
specified. The confidence interval can be calculated automatically using the
standard deviation selected in the Rule field or it can be entered manually.
e For more information, see Editing control values on page B-206.
Roche Diagnostics
B-226 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Application
The following sections explain certain settings associated with the Range tab on
Utility > Application screen. Not all settings are described, but the most important
ones.
To display this screen, choose Utility > Application > Range.
Figure B-116 Range tab on Utility > Application screen (Photometric test)
Figure B-117 Range tab on Utility > Application screen (ISE test)
e For instructions on changing default application parameters on the Range tab, see:
To change the unit of measure for an application after calibration on page B-228
To define application report name on page B-229
To define application automatic rerun on page B-229
To change the technical limit on page B-230
To change the repeat limit on page B-230
To change qualitative fields (photometric tests) on page B-231
To define serum indices on page B-231
To change expected values on page B-232
To change default settings on page B-233
Roche Diagnostics
Operator’s Manual · Version 2.0 B-227
11 Configuration cobas c 311 analyzer
Application
Unit of measure
The current unit of measure is displayed in this field. The unit has first been selected
when the application was downloaded from cobas link.
Changing the unit also recalculates the concentration based application data, for
example, repeat limit, expected value. The list box can offer up to four different units.
1 Backup the database and delete patient data (System Overview > Sample Data
Clear > Backup and Clear).
If Sample data clear is executed, all the records of the samples are deleted and QC
data are moved to the QC View.
e For details see
Archiving patient data on page B-106
Sample Data Clear button on page B-50.
2 Unload all the cobas c packs for that test (Reagent > Setting > Unloading).
e For details see Unloading cobas c packs on page B-57.
Any cobas c packs for this test that were already in use on the analyzer cannot be
reloaded onto the analyzer after the test has been deleted.
7 Install the calibrator values for this test under Calibration > Install > Download.
e For details see Loading calibrator data on page B-167.
8 Install the control values for this test under QC > Install > Download.
e For details see Loading control data on page B-203.
Roche Diagnostics
B-228 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Application
Report name
If specified, the report name will be used for printouts instead of the short test name.
For the example in Figure B-116 on page B-227, Test02 will be printed out for
SGLUC reports.
Automatic Rerun
o If automatic rerun is selected, the sample has to remain on the sample disk until the
results of the sample are available.
o To perform automatic reruns, the automatic rerun function must be selected (Start
(global button) > Automatic Rerun).
Roche Diagnostics
Operator’s Manual · Version 2.0 B-229
11 Configuration cobas c 311 analyzer
Application
Repeat limit
For each test a clinical relevant range can be typed in here. If the test result is outside
this limit but inside the Technical Limit of the application, the test is repeated using
the same sample volume and dilution as in the first run.
o The concentration range of the Repeat Limit must lie within the Technical Limit.
o A Repeat Limit Flag (>Rept / < Rept) is only added to a result if this is enabled in
Utility > System (Page 1/5) > Alarm.
5 Verify the settings of the Repeat Limit Flag and the With Automatic Rerun
check boxes. If both are selected, an automatic rerun will be performed if the
result is outside the repeat limit (>Rept, <Rept data alarms).
Roche Diagnostics
B-230 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Application
To use the serum index function, the serum index application must be installed and the
serum index requested on the Workplace > Test Selection screen.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-231
11 Configuration cobas c 311 analyzer
Application
Expected values
Expected values area is used to define the normal range for men and women in three
different age groups. If the result for this test falls outside the ranges entered here, the
system issues an alarm (H, L (only on printouts)). The last row of the field does not
allow an age limit to be defined. These expected values include patients older than the
upper limit of the second age group.
A C
B D
E
Roche Diagnostics
B-232 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Application
The following section explains how to set up two independent sets of calibration
curves for different sample types for each ISE test Na+, K+, and Cl-.
To display this screen select Utility > Application > Others.
Figure B-120 Others tab on Utility > Application screen (ISE test)
Use the Others tab to set up two independent sets of calibration curves for different
sample types for each ISE test Na+, K+, and Cl-. Choose an ISE test from the test list to
display calibrator codes, concentration values, and disk position numbers.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-233
11 Configuration cobas c 311 analyzer
Application
If only one calibration type is used—if you use the same calibration curve for all sample
types— then you must enter the same calibrator codes for ISE-A and ISE-B. It is not
possible to inactivate one of the calibrator curves.
If a field is grayed out, the field is read only for the password level logged onto the system.
At an administrator level, if the field is not grayed out, it can be edited.
Roche Diagnostics
B-234 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
System configuration
System configuration
This section explains how to perform certain tasks associated with the Utility >
System menu.
e For a complete description of all fields and commands under the System menu, refer to
the Online Help.
The entire system parameter settings should be set before the other settings (control,
calibrator settings).
e For information about particular tasks associated with this screen, see:
Programming system parameters on page B-236
Saving system parameters on page B-265
Restoring system parameters on page B-267
Roche Diagnostics
Operator’s Manual · Version 2.0 B-235
11 Configuration cobas c 311 analyzer
System configuration
Roche Diagnostics
B-236 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
System configuration
1/5
2/5
3/5
4/5
5/5
Roche Diagnostics
Operator’s Manual · Version 2.0 B-237
11 Configuration cobas c 311 analyzer
System configuration
By default, the analyzer uses Sample IDs. If there are several sample types of the same
patient, each sample has a different ID.
Using Patient IDs means to assign the same ID to all samples types (serum/plasma,
urine, CSF, and supernatant) of one patient. To distinguish the different sample
types, they have to be assigned to different position areas on the sample disk.
To assign positions, the Patient ID mode has to be activated first. The Change button in
the Position Assignment area will only be displayed in this mode. Therefore you have to
delete all data in the Data Review screen.
4 With regard to routine and STAT samples, assign each of the positions for each of
sample types.
5 Choose OK to save the changes made.
Roche Diagnostics
B-238 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
System configuration
The SBS mode function allows the analyzer to process open test requests
automatically without operator intervention in most cases. In some exceptional cases,
however, intervention by the operator may still be required.
e For detailed information about this function, see
Open test requests on page B-112.
Figure B-125 Utility > System (page 1/5) > Host Communication window
Roche Diagnostics
Operator’s Manual · Version 2.0 B-239
11 Configuration cobas c 311 analyzer
System configuration
6 Check the setting of the Ignore host orders for samples with already existing
1st results check box.
This check box determines which host orders are measured in SBS mode.
o Select this check box if host orders for tests which already have 1st results
should be ignored and not be measured again. In this case only additional
orders sent from host (manually via batch) will be measured, host requests for
rerun and 3rd results will be ignored.
o Deselect this check box if all orders from host should be measured in SBS
mode, even if not necessary. In this case the complete test selection sent from
host will be measured, resulting in unneccessary rerun orders.
It is always possible to generate 2 results for each test: 1st run and rerun. Select the
3rd Results Acceptance check box on Utility > System to expand the ability to
generate multiple results for a single test.
Description of 3rd Results When 3rd Results Acceptance is activated, it is possible to request a test more than
Acceptance two times after confirming a safety warning. Any third or subsequent result will
overwrite the previous rerun result.
The repeated rerun is only performed if the previous rerun result has been sent to the
Host. In case the previous result is not transmitted prior to the repetition request, the
sample is not pipetted.
Preconditions for 3rd Results Before activating 3rd Result Acceptance it is necessary to check the following
Acceptance conditions.
Roche Diagnostics
B-240 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
System configuration
Repeat Limit Flag / With When this feature is enabled, the system utilizes the repeat limit flag and repeat limit
Automatic Rerun values as defined under Utility > Application > Range. Whenever a sample result is
outside the specified repeat limit for a particular test, a data flag >Rept / <Rept is
attached to each result.
e For information on how to define the repeat limit, see: Repeat limit on page B-230.
When the Repeat Limit Flag check box is selected, the With Automatic Rerun check
box also is available. If it is selected as well, the particular test for this is sample is
repeated with the same sample volume as the in the first run.
To perform automatic reruns, the automatic rerun function must be selected (Start (global
button) > Automatic Rerun).
e For more information about the Alarm Setting window, refer to the Online Help for that
particular window.
The Host Communication Setting window is used to define all Host computer
interface parameters. The parameters are defined during installation by a Roche
Service engineer and should not be changed as this may cause data transferring
problems.
The Host Communication Setting window is divided into 3 tabs, which are
described in the following.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-241
11 Configuration cobas c 311 analyzer
System configuration
Use this tab to define the text settings used for Host communications.
The abbreviation TS stands for Test Selection.
e For a detailed description, refer to the Host Manual.
Figure B-128 Host Communication Setting window - Result Upload Setting tab
Use this tab to define the upload settings used for the transfer of result data from the
analyzer to the Host.
Option Meaning
by sample All results of one sample are collected before uploading
ISE/Chemistry ISE and all clin.-chem. results are uploaded separately
by Test (only for STAT samples) As soon as the result is available
Table B-35 Routine Sample Host upload setting
No Automatic Result Upload When this function is enabled, no result is uploaded in real time. In this case, the
results can be uploaded manually by selecting samples from the Workplace > Data
Review screen and choosing Send To Host.
Roche Diagnostics
B-242 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
System configuration
To display this window, choose Utility > System (Page 1/5) > Key Setting.
This window is used to assign tests or profiles to test keys. One test or profile can be
assigned to each test key. Tests can be assigned for up to 5 groups of 32 keys per
sample type. In total 160 test keys are available for each sample type.
The group name is not displayed on the Key Setting window. It is displayed in the
Test Selection screen.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-243
11 Configuration cobas c 311 analyzer
System configuration
o If you want to delete a test key assignment, select Undefined at the end of the list.
o Before you can select a profile, the profile has to be defined.
a To define a profile
Profiles are sample type specific, this means the sample type has to be selected.
1 Choose the tab of a sample type on the Key Setting window, for example, Ser/Pl.
2 Choose Profile Setting on the Key Setting window. The Profile Setting window
is displayed.
Roche Diagnostics
B-244 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
System configuration
Use this window to set the number of sample disk positions that are reserved for
STAT samples (0 to 53 positions). The assignment of the STAT positions is displayed
on the top left of the Utility > System screen.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-245
11 Configuration cobas c 311 analyzer
System configuration
Use this window to define the dimensions of false bottom tubes/cup on non standard
tubes. The following measures must be defined:
o Container height - the length of the tube: 73 mm - 102 mm
o Bottom Level (see Figure B-134): 0.1 mm - (Container height - 20) mm
If you use tubes with conical bottom, select Conical Bottom.
When using false bottom tubes or cup on non-standard tubes, the distance between
the upper edge of the tube and the bottom level must be more than 20 mm.
90 mm (for example)
Container Height
> 20 mm
Bottom Level
15 mm (for example)
o Note that only one false bottom tube or non-standard tube can be specified.
o The analyzer automatically detects whether a false bottom tube is set in a sample disk
position. You do not need to assign positions for false bottom tubes.
o For exact specifications of the sample containers, contact its manufacturer.
Roche Diagnostics
B-246 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Module Set
Module Set
This section explains how to perform certain tasks associated with entire modules
(photometric unit or ISE unit) rather than specific samples or tests. For the
cobas c 311 analyzer, the word module is used synonymously to unit.
The entire module set (module configuration, test assignment) should be performed before
the other settings (control, calibrator settings).
Roche Diagnostics
Operator’s Manual · Version 2.0 B-247
11 Configuration cobas c 311 analyzer
Module Set
Assigning a test
2 Select a test in the Test Name list box of the Test Assignment window.
3 Select the Assign check box.
4 Select the Mandatory check box if the reagent has to be on board all the time.
5 Select Na, K, Cl or Na, K.
o Na, K, Cl: This option allows sodium (Na+), potassium (K+) and chloride
(Cl-) ion testing on this ISE unit.
o Na, K: This option allows only sodium (Na) and potassium (K) testing on this
ISE unit.
6 Choose OK.
Roche Diagnostics
B-248 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Calculated Tests
Calculated Tests
This section explains how to program calculated tests and compensated tests.
To display this screen, choose Utility > Calc. Test.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-249
11 Configuration cobas c 311 analyzer
Calculated Tests
Calculated test results are not performed on the system but are derived by applying a
test formula to the results of clinical chemistry tests performed on the analyzer. Yet,
test formulas are not to be applied on control material.
If test data to be used in the calculation is unavailable, the program does not perform
the calculation. Calculated test results cannot be edited. Use the following procedures
to add, edit or delete a calculated test formula.
4 Select the type of sample (Ser/Pl, Urine, CSF, Suprnt) for which the calculated
test formula is being created from the Sample Type list.
5 Type the unit of measure in the Unit Of Measure box.
6 Type a name for the calculated test as it should appear on the report in the Report
Name box.
7 Type the short name of the calculated test in the Test box.
8 Use the Test list to select a test that is part of the formula. After highlighting the
test, choose Select to add the test to the formula. The formula appears to the right
of the Test text box, after the equal sign. Use the Calculator area in the window to
enter mathematical operators and numerals in the correct sequence along with
test names until the formula is complete.
9 Now, set the range parameters.
Roche Diagnostics
B-250 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Calculated Tests
2 Type the expected value ranges or the qualitative parameters in the Range area, if
needed. Select the Default setting, if needed.
e For information about editing the expected values, see:
Expected values on page B-232
3 Choose OK to save the new formula and close the Calculated Test Formula
window. The new formula is shown in the Calculated Test list box at the top of
the Calc. Test screen.
The number of digits displayed to the right of the decimal in the calculated test result is
determined by the number of digits entered to the right of the decimal for the first
Male Expected Value field on the Range tab of the Calculated Test Formula
window.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-251
11 Configuration cobas c 311 analyzer
Calculated Tests
Roche Diagnostics
B-252 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Calculated Tests
Clinical chemistry test results can be processed on the system and adjusted by
applying a compensated test formula. Use the following procedures to add, edit or
delete a compensated test formula.
4 Select the test and sample type to be compensated from the Test Name/Sample
Type list box in the upper left corner of the window.
5 Use the Test list to select any test that is part of the formula. Choose Select after
selecting the test to add it to the formula, which is shown at the top of the
window. Use the Calculator area to enter mathematical operators and numerals
in the correct sequence with test names.
6 Choose OK to save the new formula and close the Compensated Test Formula
window. The new formula is shown in the Compensated Test list box at the
bottom of the Calc. Test screen.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-253
11 Configuration cobas c 311 analyzer
Calculated Tests
Roche Diagnostics
B-254 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Special Wash
Special Wash
The cobas c 311 analyzer is a random access system. Therefore, there is potential for
sample probe, reagent probe and reaction cell carryover and thus interferences
between tests.
To avoid carryover and cross-contamination between tests use the special wash
function. This function allows you to preset combinations of reagents or samples
where carryover may occur. The analyzer then performs a special wash cycle
whenever this combination is encountered during analysis.
The following section explains how to set the carryover avoiding function.
The reagent probe dips into the reagents when aspirating. To avoid a reagent
carryover, program a special wash cycle for the reagent probe.
e For more information refer to the package insert NaOHD/SMS/SmpCln 1 and the
carryover list.
2 Select an empty line in the Reagent Probe list to add a reagent probe wash. To
edit an existing reagent probe wash, select the corresponding line.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-255
11 Configuration cobas c 311 analyzer
Special Wash
3 Choose Edit, on the right of the Reagent Probe list, to display the Edit Reagent
Probe Wash window.
4 In the From Reagent area, select the test name for the causal reagent in the Test
list box. Select the appropriate reagent type (R1, R2, or R3) in the Reagent Type
list box.
5 In the To Reagent area, select the test name for the affected test in the Test list
box. Select the appropriate reagent type (R1, R2, or R3) in the Reagent Type list
box.
6 Select the Detergent type (D1, D2, D3, or Water) from the Type list box for each
reagent probe selected.
o D1 = NaOH-D (application code 947)
o D2 = SMS (application code 948)
o D3 = SCCS (application code 949)
7 Then type the wash solution volume (from 20 to 180 L) in the Volume text box.
8 Choose OK to add the new special wash to the list on the Special Wash screen
and close the window.
9 Check the amount of the appropriate detergent cassette on the instrument on the
Reagent > Status screen.
Roche Diagnostics
B-256 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Special Wash
To avoid the carryover of traces of the test mixture from one test to the next test
measured in that same reaction cell, program a special wash cycle for the reaction
cell.
4 Select a test from the Test list box that requires cell washing.
5 Select the R1 check box to activate a cell wash for this test.
6 If more than 180 L detergent solution is needed, select the R2 check box as well.
7 Select the Detergent Type (D1, D2, D3, or Water) from the Type list box.
o D1 = NaOH-D (application code 947)
o D2 = SMS (application code 948)
o D3 = SCCS (application code 949)
8 Enter the detergent solution volume (from 20 to 180 L) in the Volume text box.
9 Choose OK to save the settings, add the programming to the Cell list box and
close the window.
10 Check the amount of the appropriate detergent cassette on the instrument on the
Reagent > Status screen.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-257
11 Configuration cobas c 311 analyzer
Special Wash
The sample probe dips into the sample when it aspirates sample liquid. To avoid a
carryover of sample liquid, program a special wash cycle for the sample probe.
The Whole Blood Detergent Type is only available when HbA1c calculation is
selected as the test. This will wash the sample probe after the whole blood sample is
pipetted.
Roche Diagnostics
B-258 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Special Wash
5 Select the detergent type from the Detergent Type list box.
This solution is used for washing the sample probe.
o 1 (Sample Cleaner 1)
o 2 (Sample Cleaner 2)
o 1+2 (D1+D2)
o water
The detergent aspiration volume is 45 L.
6 Choose OK to save the settings.
7 Check that the appropriate wash solution is placed on the instrument in the
correct position.
8 Check the amount of the appropriate detergent on the Reagent > Status screen.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-259
11 Configuration cobas c 311 analyzer
Special Wash
Roche Diagnostics
B-260 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Report Format
Report Format
This section explains how to customize the report format of the patient report.
The patient report (Print (global button)> Workplace> Data Print) can be printed
in two formats: Monitor format or Report format. Only the Report format can be
customized.
Customizing the report format of the patient report consists of the following
steps:
o Choose the page layout.
e To choose the page layout on page B-262
o Define the title, the positions of all other items, and the print order.
e To define the title and the positions of all other items on page B-262
o Choose Save to save the changes.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-261
11 Configuration cobas c 311 analyzer
Report Format
Roche Diagnostics
B-262 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Report Format
1 In the Print Order area, select the test from the list.
2 Type the line number on which the test is to print in the Line text box.
To check the report setup, print a report example from Print (global button) > Utility >
Report Example.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-263
11 Configuration cobas c 311 analyzer
Report Format
Roche Diagnostics
B-264 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Saving system parameters
You can format a floppy disk using the maintenance item (13) Floppy Disk Utility, if
required.
Do not eject the floppy disk when the access lamp of the floppy disk drive is lit.
8 When processing is complete, remove the floppy disk from the disk drive and set
aside for use later in this procedure.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-265
11 Configuration cobas c 311 analyzer
Saving system parameters
Roche Diagnostics
B-266 Operator’s Manual · Version 2.0
cobas c 311 analyzer 11 Configuration
Restoring system parameters
Use the following procedure to restore system parameters from a backup floppy disk.
Roche Diagnostics
Operator’s Manual · Version 2.0 B-267
11 Configuration cobas c 311 analyzer
Restoring system parameters
Roche Diagnostics
B-268 Operator’s Manual · Version 2.0
Maintenance C
Before performing any maintenance, read the following safety messages carefully. If
you ignore these safety messages, you may suffer serious or fatal injury.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-3
12 Safety information for maintenance cobas c 311 analyzer
e For information about water pressure, see Operating conditions on page A-71.
Roche Diagnostics
C-4 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
Table of contents
General maintenance
This chapter provides general information about maintenance of the cobas c 311
analyzer. Concepts, such as maintenance pipes, maintenance pipe functions and
maintenance types are described as well as instrument care. Lists of maintenance
items and checks are also provided.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-5
13 General maintenance cobas c 311 analyzer
Table of contents
Roche Diagnostics
C-6 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
Overview
Overview
Definitions
System-controlled maintenance Maintenance items that are performed without operator interaction (such as Air
Purge).
Operator-controlled Maintenance items that require operator interaction (such as Cleaning Sample
maintenance Probe).
User-definable maintenance Users may define their own maintenance items for the laboratory. User-definable
maintenance items are always maintenance actions that have to be performed
manually.
Maintenance pipe A combination of sequential maintenance items programmed into a fully automated
procedure, which can be performed by the analyzer without operator intervention.
Maintenance pipe function A function that automatically starts a maintenance pipe at a particular time (e.g., at
power-up).
Roche Diagnostics
Operator’s Manual · Version 2.0 C-7
13 General maintenance cobas c 311 analyzer
Maintenance items
Maintenance concept
Performing maintenance items Maintenance items can be performed manually on the Utility > Maintenance screen.
manually
Operator-controlled and user-definable maintenance items cannot be part of a
maintenance pipe. Thus they have to be performed manually.
e See Performing maintenance items on page C-9
Maintenance pipes—performing A maintenance pipe is a set of system-controlled maintenance items (batch set). You
maintenance items can use a maintenance pipe so that certain maintenance items occur in a certain
automatically sequence.
Before usage, maintenance pipes must be defined under Utility > System
(Page 2/5) > Pipe Setting. Up to 20 pipes can be defined and each pipe can have up
to ten maintenance items including parameters belonging to them.
Maintenance pipes can be executed in two different ways:
o Manually by the user
o Automatically by means of maintenance pipe functions at power up of the system
(Power Up Pipe function) or before start of analysis (Start Up Pipe function)
e For more information on maintenance pipes, see:
Maintenance pipes on page C-10
Maintenance types—scheduling Maintenance items and maintenance pipes can be grouped together in a maintenance
and tracking maintenance items type according to certain functions (such as daily or weekly maintenance).
Within a maintenance type you can assign maintenance intervals (Period) and
warning levels to each maintenance item and maintenance pipe. Thus, you can create
and customize a maintenance schedule for every periodic maintenance task.
If a maintenance item or a maintenance pipe has been scheduled, you can track its
status (date of last execution and warning level) on the Maintenance screen. The
status of maintenance types is also displayed on the System Overview screen.
Maintenance types are configured in Utility > System (Page 2/5) > Maintenance
Setting.
e For more information on defining and using maintenance types, see:
Defining and editing maintenance types on page C-19
Tracking maintenance on page C-22.
Maintenance report—recording The analyzer records the execution of maintenance items in the Maintenance
maintenance Report.
The Maintenance Report is requested through Print (global button) > Utility >
Maintenance Report.
e See Maintenance report on page C-23
Maintenance items
Roche Diagnostics
C-8 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
Maintenance items
The date of execution will also be updated even if operation is interrupted due to an
alarm, etc. In this case, perform the relevant maintenance item again and terminate it
normally.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-9
13 General maintenance cobas c 311 analyzer
Maintenance pipes
Maintenance pipes
Maintenance pipes Before usage, maintenance pipes must be defined under Utility > System
(Page 2/5) > Pipe Setting. The name of a pipe is freely definable, e.g. Power On pipe.
Roche recommends using maintenance pipes for automating instrument care, but
there are no pre-defined pipes on the system by default.
e For more information, see:
Defining and editing maintenance pipes on page C-11
Recommended maintenance pipes on page C-17
Maintenance pipes can be executed manually by the user. Alternatively, they can be
executed automatically by means of maintenance pipe functions.
Maintenance pipe functions A maintenance pipe function triggers the system to start a maintenance pipe at one of
those three particular times (events):
o At power-up or wake up (Power Up Pipe function)
o Before start of analysis (Start Up Pipe function)
o Before entering sleep mode after the last maintenance item is completed (Sleep
Pipe function)
e Using Maintenance pipe functions on page C-14
Example The Power On pipe contains multiple maintenance items (Figure C-2). It can be
executed manually or be assigned e.g., to the Power Up Pipe function. In the latter
case, the pipe function will automatically start the maintenance pipe at power-up or
wake up.
Maintenance pipe:
Note that:
o Power On is used as a name for a maintenance pipe. (This name is freely definable.)
o Power Up Pipe is a fixed name of a maintenance pipe function.
The following sections describe how maintenance pipes are defined, executed, and
deleted.
Roche Diagnostics
C-10 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
Maintenance pipes
Use the following procedure to program a series of maintenance items into a specific
maintenance pipe. Up to 20 maintenance pipes are definable and up to ten items can
be defined in each pipe. These items are performed one after the other. When a
maintenance item is performed as part of a pipe, an asterisk appears on the left side of
the date on the Maintenance Report.
e See Maintenance report on page C-23
2 Choose an existing pipe in the Pipe Name list to edit the pipe or choose the first
blank line to define a new pipe.
3 Choose Edit to display the Edit Pipe window.
4 Type an unused name for the new pipe in the Pipe Name text box or edit the
name of the existing pipe if you want to change its name.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-11
13 General maintenance cobas c 311 analyzer
Maintenance pipes
5 Add a maintenance item to the Pipe Item List by choosing the maintenance item
on the Maintenance Item List, then choose Add.
For automatic transition to sleep mode after completion of a maintenance pipe,
select the Sleep check box.
6 Choose Parameter to define the parameters such as the number of cycles to be
performed depending on the maintenance item highlighted.
e For more information about the different maintenance items, see:
List of maintenance items on page C-24.
e For more information about the parameters of the different maintenance items, refer
to the Online Help for that particular window.
7 Choose OK to save the settings and return to the Edit Pipe window.
8 Repeat steps 5 to 7 for all items that are to be included in the pipe.
9 Delete items from the pipe by choosing the appropriate item on the Pipe Item
List, then choose Remove.
Items must be listed in the sequence that they are to be performed by the system. If the
items are not in the desired order, choose the item to be moved to highlight it, then
choose the appropriate arrow to move the item up or down in the list.
10 Choose OK when the pipe settings are complete. The pipe then appears in the
Pipe Name list in the Maintenance Pipe Setting window and as an option on
Utility > System (Page 2/5) > Power Up Pipe.
Roche Diagnostics
C-12 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
Maintenance pipes
The pipe must have been defined as a maintenance type first (see Defining and editing
maintenance types on page C-19).
Roche Diagnostics
Operator’s Manual · Version 2.0 C-13
13 General maintenance cobas c 311 analyzer
Maintenance pipes
2 Select the pipe to be performed at power up from the drop-down list box for each
day of the week.
3 In case you want to set a wake-up time, do the following:
o Select the check box to the right of the weekday for which you want to set a
wake-up time. (If the check box is deselected, the specified maintenance pipe
is executed manually, e.g after power up.)
o Set the time in the two boxes to the right of the check box.
4 Choose OK.
The PC will be restarted and the analytical unit initialized 10 minutes before the
specified time. Thereafter the maintenance pipe will be performed.
During sleep mode the wake-up time is displayed in the Logon window.
Roche Diagnostics
C-14 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
Maintenance pipes
The Start Up Pipe Setting area displays the maintenance pipe currently selected
as Start Up pipe. If no maintenance pipe is selected, None is displayed.
2 Choose Change in the Start Up Pipe Setting area.
3 Select the desired start up pipe from the list box on the Pipe Function window.
4 Choose OK to accept the change.
5 Choose Start (global button) to perform the displayed pipe.
After completion of the start up pipe, analysis will be started automatically.
Once the start up pipe begins, the Start Up Pipe Setting changes back to None.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-15
13 General maintenance cobas c 311 analyzer
Maintenance pipes
4 Select the Sleep check box to activate the Sleep Pipe function.
The system automatically enters Sleep mode after the pipe is completed.
e See Recommended maintenance pipes on page C-17
5 Choose OK to save the settings and return to the Maintenance Pipe setting
window.
Roche Diagnostics
C-16 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
Maintenance pipes
Automating instrument care We recommend using the maintenance pipe functions to perform maintenance pipes
automatically. This ensures a maximum of instrument care with minimal operator
intervention.
e For information about maintenance pipes and maintenance pipe functions, see:
Maintenance pipes on page C-10
Defining and editing maintenance pipes on page C-11
Using Maintenance pipe functions on page C-14
Power ON pipe
We recommend performing this maintenance pipe by means of a maintenance pipe
function—either after initialization of the analyzer (Power Up Pipe function) or
before start of analysis (Start Up Pipe function).
Note that even though a maintenance pipe can be executed automatically some
maintenance items require visual checks by the operator.
Sleep pipe
By use of the Sleep Pipe function the analyzer will automatically turn to sleep mode
after the maintenance pipe is completed.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-17
13 General maintenance cobas c 311 analyzer
Maintenance pipes
Weekly pipe
We recommend performing this maintenance pipe every week before shutdown or
start up.
Before you shut down the analyzer, check the cell blank values in the Print View window.
The values will be lost after shutdown.
Roche Diagnostics
C-18 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
Maintenance types
Maintenance types
Predefined maintenance types Two Maintenance types are predefined: Maintenance and Check.
o Maintenance: Contains all available maintenance items except check routines
and service routines.
o Check: Contains all check routines.
Due time For each maintenance item or maintenance pipe, the control term (due time) is
definable. A maintenance item or pipe which has passed its scheduled term will be
indicated in yellow or red (depending on criticality) on screen:
o Indication in yellow: At warning level
o Indication in red: Due or overdue
Use the following procedure to group a series of maintenance items into a specific
maintenance type.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-19
13 General maintenance cobas c 311 analyzer
Maintenance types
3 Enter a name for the Maintenance Type in the text box at the top of the window.
7 For maintenance tracking, enter the frequency period in the text box and select
the time period (Hours, Days, Months) from the Period list box.
If a frequency period is not required, enter 0 in the field to cancel.
8 Select a timer function from the Timer list box:
o Realtime: The actual time
o Power On: Cumulative power-on time of the analyzer (excluding sleep time)
o Operation: Cumulative operation time
Power On time and Operation time are both listed on the Maintenance
Report (Print > Utility > Maintenance Report).
e See Maintenance report on page C-23.
9 Enter the warning level (%) in the Warning Level text box.
Example:
If you define a period of 10 hours in real time and a warning level of 90%, the
analyzer will remind you to perform the maintenance item after 9 hours.
If a warning level is not required, enter 0 in the field to cancel.
10 By selecting the Comment check box you can add a comment during execution
that will be printed out on the Maintenance Report.
11 Choose Update to add the item to the Maintenance Items list.
Roche Diagnostics
C-20 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
Maintenance types
12 Repeat steps 3 to 11 for all maintenance items for the maintenance type.
13 Choose OK to save the group and its settings.
14 Choose Close to close the Edit Maintenance Type window.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-21
13 General maintenance cobas c 311 analyzer
Maintenance types
Tracking maintenance
Roche Diagnostics
C-22 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
Maintenance report
Maintenance report
To print out a report for a certain maintenance item, choose Utility > Maintenance and
select the desired maintenance item. Then choose Print (global button).
e For more information about the Maintenance Report, refer to the Online Help for
that particular window.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-23
13 General maintenance cobas c 311 analyzer
List of maintenance items
The following section lists all maintenance items and a brief description of their
function. The time period in brackets denotes the approximate system time of the
maintenance item in (min:sec).
(ISE): Denotes that a function applies to the ISE unit only.
(P): Denotes that a function applies to the photometric unit only. In the maintenance
windows, the abbreviation AU is used to refer to the photometric unit (P).
(P+ISE): Denotes that a function applies to both units.
(1) Reset (P+ISE) (0:30): Mechanical parts are reset to their home positions.
(2) Photometer Check (P) (2:40): Photometer output is checked by measuring a water blank of cell no. 1.
Perform this maintenance item daily.
Check the photometer values on Print > Photometer Check. Measurement results
must be below 14000. If results are higher, check the light path; if necessary, replace
the photometer lamp, then perform a cell blank measurement.
Before performing this maintenance item, check that the incubator temperature is
within 37 0.1C (see System Overview screen on page B-44).
Depending on the ambient temperature it can take up to 30 minutes after switching
on the analyzer or after maintenance item (4) Incubation Water Exchange for the
incubator to reach the correct temperature.
e See To check the light intensity on page C-113.
(3) Cell Blank Measurement (P) (16:50): Water blanks of all cells are measured. If there is a difference greater than
0.1 absorbance units in the results for one cell compared to the cell No. 1, that cell
number is written to the Abnormal Cell List. To view this list choose Print (global
button) > View. Perform this maintenance item weekly.
e For information about the cell blank measurement, see:
To perform a cell blank measurement on page C-60
To check the Cell Blank Measurement report on page C-60.
(4) Incubation Water Exchange (P) (5:30): Incubator bath water is exchanged and Hitergent (4.3 mL) is added.
Perform this maintenance item daily. Incubator bath is also referred to as reaction
bath.
After this maintenance item has been performed, check that the incubator
temperature is within 37 0.1C (see System Overview screen on page B-44) before
performing maintenance item (2) Photometer Check. Depending on the ambient
temperature it can take up to 30 minutes for the incubator to reach the correct
temperature.
Roche Diagnostics
C-24 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
List of maintenance items
(5) Air Purge (P) (1:45): Any air is purged which may have become trapped in the sample and
reagent syringes. Water as well as any air are discharged at the rinse stations. Perform
this maintenance item daily.
With top cover closed, visually check that water is dispensed from the sample probe
and/or the reagent probe in a straight flow.
After air purge (in Standby mode), visually check the syringes and the tubing system
for leakage and air bubbles.
e See To perform an air purge (sample syringe or reagent syringe) on page C-106.
(6) Wash Reaction Parts (P) (29:30): All cells and probes are washed. The sample probe is washed with Sample
Cleaner 1 from the 70 mL detergent bottle (approx. 300 L). The reagent probe and
reaction cells are washed with detergent from a cobas c pack (NaOH-D, application
code 947, 26 mL).
Perform this maintenance item weekly or with new cells. If you are running DAT
assays, then perform this item daily. For more information please refer to the
instructions for use in cobas link of the assay.
e See To wash the reaction parts on page C-59.
(7) Reagent Prime (ISE): Reagents are primed according to the option chosen (IS+REF or REF).
Perform a REF reagent prime daily.
Whenever you replace either the ISE IS or ISE Ref. bottle, a reagent prime and an ISE
calibration must be carried out to fill the flow path with the new liquid and to
calibrate the ISE unit.
e For the REF option, see: To prime the ISE Ref. tubing on page C-75
For the IS+REF option, see: To prime lines on page C-90.
(8) Cell Detergent Prime (P) (8:55): Reagent lines for Cell Wash Solution I and II are primed; reagent lines are
purged of air and cells are filled up and emptied by vacuum.
You must select which detergents need to be primed (Detergent 1, Detergent 2, or
both). Perform when new bottles are loaded or when the filters are cleaned.
e See To perform a cell detergent prime on page C-79.
(9) Incubator Bath Cleaning (P): Performs transition to incubator bath cleaning mode: the photometer lamp is
switched off and the incubator bath water is automatically drained. Then the analyzer
will perform a complete system shutdown.
Switch off the analyzer after the control unit has powered off. After cleaning, power
on the analyzer as usual; the incubator bath is refilled with some water. A yellow
alarm is issued, indicating that the level of incubation water is below the lower limit.
Perform maintenance item (4) Incubation Water Exchange to exchange the
incubator bath water and to add Hitergent (4.3 mL).
Perform this maintenance item monthly to clean the incubator bath and the drain
filter, and quarterly to clean the ultrasonic mixer.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-25
13 General maintenance cobas c 311 analyzer
List of maintenance items
e See To remove reaction cells and to clean the incubator bath on page C-69
To clean the surface of the ultrasonic mixer on page C-93.
(10) ISE Wash (P+ISE) (28:15): The sample probe and the flow path of the ISE unit are washed. The
sample probe is washed with Sample Cleaner 1 from the 70 mL detergent bottle. The
ISE flow path is washed with SysClean from position W1 on sample disk. The ISE
electrodes are conditioned with Activator from position W2.
Perform this maintenance item daily if running ISEs. This item can be performed in a
sleep pipe.
e For more information, see:
Sleep pipe on page C-17
To wash ISE flow path, ISE electrodes and the sample probe on page C-50.
(11) Sample Probe Wash (P) (6:10 for 5 cycles): The sample probe is washed with Sample Cleaner 1 from the
70 mL detergent bottle (approx. 300 L). This item can be used in case of a clogged
probe.
e See To wash the sample probe on page C-123.
(12) Flow Path Wash (P) (20:30): The flow path, the drain port and the drainage tube of concentrated waste
solution are cleaned with CellCln II (detergent consumption: approximately 30 mL).
Perform this maintenance item monthly.
e See To wash the reaction system on page C-86.
(13) Floppy Disk Utility A floppy disk inserted into the floppy disk drive will be formatted after confirmation.
(14) Parameter Read/Write Parameter settings are read from or written to a floppy disk.
e See To save system parameters on a floppy disk on page B-265
See To restore system parameters to the system on page B-267.
Roche Diagnostics
C-26 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
List of maintenance items
cobas link Essential Information Important data such as calibrator and control data is uploaded and stored on the
Upload cobas link data station. The last 5 backups will be stored, the oldest backup will be
overwritten with the newest backup. Hence the last 5 backups are available to restore
to the analyzer if required.
It is recommended performing this maintenance item every day. This maintenance
item can only be performed in a maintenance pipe. Please contact either your Roche
service representative or technical support to include cobas link Essential
Information Upload in a maintenance pipe.
e For more information, see:
Defining and editing maintenance pipes on page C-11
Recommended maintenance pipes on page C-17
Roche Diagnostics
Operator’s Manual · Version 2.0 C-27
13 General maintenance cobas c 311 analyzer
List of maintenance checks
The following section lists all maintenance checks and gives a brief description of
their function.
(1) Disk Check Files are checked on hard disk or floppy disk and a directory listing is printed out. If a
malfunction occurs, the system issues an alarm. The hard disk check can also be used
to determine the software version.
If the Backup Datafile option is selected, the system collects and stores system data
for maintenance purposes.
e For using the Backup Datafile option, see Backup of the system data files (getlog
procedure) on page D-53
(2) ISE Check (ISE) (18:00 for 10 cycles): The electrode output of ISE cartridges (Cl, K, Na and REF
cartridges) is measured with ISE IS and a listing of EMF values is printed out.
Perform this check twice after replacing the ISE cartridges and ISE tubing. Wait
10 minutes between each check.
e See To perform an ISE Check and calibrate the ISE unit on page C-90.
(3) Mechanisms Check (P+ISE) (9:30 for 30 cycles): Performs a functional check of mechanical parts of the
entire sample line of the analyzer (according to the number of cycles).
If the check box With Photo Interrupter Check is selected, the system checks
automatically all relevant photocoupler sensors during this maintenance function.
Only the defective sensors are shown in the Print View window.
Choose Stop (global button) to end this maintenance check. Perform this check as
needed after replacing the sample or reagent probe and after replacing the nozzle tips
on the cell rinse nozzles.
e See To perform a Mechanisms Check on page C-130.
(4) Barcode Reader Check (P) The barcode on a sample tube or cobas c pack is scanned. After pressing Stop, the
data are printed out for verification.
To check the function of the barcode reader:
o Select Sample Barcode Reader or Reagent Barcode Reader.
o In the former case, place a barcoded tube onto the sample disk and start the
Barcode Reader Check.
In the latter case, start the Barcode Reader Check and scan a cobas c pack.
If the check box With Photo Interrupter Check is selected, the system checks
automatically all relevant photocoupler sensors during this maintenance function.
Only the defective sensors are shown in the Print View window.
The reading results are shown in the Print View window.
(5) Probe Check (P+ISE) Puts the analyzer in probe adjust mode. In this mode, a probe—selected in
the Probe Check window—can be moved to designated positions by means of the
S.Stop button. As this function must be performed with closed cover, it is not
possible to sight check the alignment of the probe.
Select Stop to stop the Probe Check.
(6) Cuvette Mixing (P+ISE) (5:05 for 10 cycles): Performs a functional check on the ultrasonic mixer.
Roche Diagnostics
C-28 Operator’s Manual · Version 2.0
cobas c 311 analyzer 13 General maintenance
List of maintenance checks
The reagent probe pipettes deionized water into the corresponding cuvette. Then the
cuvette is moved to the mixing position and ultra sonic mixing will be executed.
Perform this check after the quarterly cleaning of the ultrasonic mixer.
e See To check the intensity of the ultrasonic output on page C-95.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-29
13 General maintenance cobas c 311 analyzer
List of maintenance checks
Roche Diagnostics
C-30 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Table of contents
This chapter describes the maintenance actions required for correct and efficient
running of the cobas c 311 analyzer. The maintenance of both the ISE unit and the
photometric unit is discussed. The schedule of required periodic maintenance actions
(daily, weekly, monthly…) is provided as well as maintenance actions that are
performed as needed.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-31
14 Maintenance of the analyzer cobas c 311 analyzer
Table of contents
Roche Diagnostics
C-32 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Maintenance schedules
Maintenance schedules
This section provides you with an overview of the maintenance actions that are
required for proper instrument care. The information is divided into a tabular
overview and a brief description of the maintenance actions.
The information is sorted according to the frequency with which the actions must be
performed.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-33
14 Maintenance of the analyzer cobas c 311 analyzer
Maintenance schedules
The intervals for maintenance and checkup recommended in this document are
independent of the system operation time unless otherwise noted (e g. maintenance
actions which are based on test count or time).
Roche Diagnostics
C-34 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Maintenance schedules
Roche Diagnostics
Operator’s Manual · Version 2.0 C-35
14 Maintenance of the analyzer cobas c 311 analyzer
Maintenance schedules
The following section provides you with a brief description of the maintenance items
and maintenance actions that are required for proper instrument care.
Roche Diagnostics
C-36 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Maintenance schedules
Checks before start-up of the There are various conditions that need to be checked before an analyzer start-up.
analyzer Therefore it is important to inspect the analyzer before starting up.
e For more information, see Start-up inspection on page B-38.
Maintenance items that can be Below is a list of maintenance items that are suitable for use in a maintenance pipe. In
performed in a Power Up pipe connection with the Power Up Pipe function this maintenance pipe will be executed
automatically at power or wake up on of the analyzer.
e For information about maintenance pipe functions, see:
Using Maintenance pipe functions on page C-14
e For information about power on and sleep pipes, see:
Recommended maintenance pipes on page C-17
Roche Diagnostics
Operator’s Manual · Version 2.0 C-37
14 Maintenance of the analyzer cobas c 311 analyzer
Maintenance schedules
Roche Diagnostics
C-38 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Maintenance schedules
Weekly maintenance
The following table provides you with an overview of the maintenance actions that
must be performed on a weekly basis.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-39
14 Maintenance of the analyzer cobas c 311 analyzer
Maintenance schedules
Monthly maintenance
The following table provides you with an overview of the maintenance actions that
must be performed on a monthly basis.
Roche Diagnostics
C-40 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Maintenance schedules
Quarterly maintenance
The following table provides you with an overview of the maintenance actions that
must be performed on a quarterly basis.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-41
14 Maintenance of the analyzer cobas c 311 analyzer
Maintenance schedules
Roche Diagnostics
C-42 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Maintenance schedules
As needed maintenance
The following table provides you with an overview of the maintenance actions that
must be performed as needed.
Rinsing the sample When the sample probe gets clogged, this is o Perform maintenance item (11) Sample
probe – elimination automatically detected by the probe. Sampling Probe Wash.
of clogging stops and an alarm is issued. o If the clogging is not eliminated, clean the C-122
sample probe manually.
M23: Replacing Sample aspiration error occurs frequently. o Shut down the analyzer. C-122
sample and reagent An alarm for clogging of the sample probe occurs o Remove the probe, clean it, and eliminate
probes—elimination and is not reset after cleaning (maintenance item clogging.
of clogging (11) Sample Probe Wash). o If a probe is bent or damaged, replace it.
o Perform maintenance item (5) Air Purge.
A probe is bent or otherwise damaged.
With top cover closed, visually check that
Upon detection of abnormal sample aspiration water is dispensed from the tip of the probe
caused by a clot, the sample probe is washed in a straight flow.
automatically by the analyzer. o Perform maintenance check
(3) Mechanisms Check (30 cycles).
M21: Replacing the The lamp has been used for more than o Perform maintenance item (2) Photometer C-113
photometer lamp 750 hours(a). Check.
A photometer check value exceeds 14000. o If any of the current data exceeds 14000,
check the light path for contamination or
scratches. Check that cell no. 1 is at least
half filled with water.
o Shut down the analyzer. Wait 30 minutes
for the photometer lamp to cool down.
o Replace the photometer lamp.
o Wait 30 minutes for the photometer lamp
to stabilize.
o Perform maintenance item (3) Cell Blank
Measurement.
Table C-16 Instrument care - as needed (Sheet 1 of 2)
(a) This is Power On Time, i.e. the time the analyzer is ON. You can check the Power on time of the analyzer by looking at the printed Maintenance
Report (Print > Utility > Maintenance Report).
Roche Diagnostics
Operator’s Manual · Version 2.0 C-43
14 Maintenance of the analyzer cobas c 311 analyzer
Maintenance schedules
Roche Diagnostics
C-44 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Periodic replacement of parts
Some parts require periodic replacement for preventive maintenance. Replace these
parts regularly and in accordance with your workload and the recommendations
given in this document. For purchase, inform your local service representative of the
part number.
Spare parts
For proper instrument care use only original spare parts provided by Roche.
o The necessary spare parts are listed in the description of the corresponding
maintenance action.
o To order spare parts, please contact your Roche representative.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-45
14 Maintenance of the analyzer cobas c 311 analyzer
Daily maintenance
Daily maintenance
In this section you will find all the maintenance actions that must be performed every
day.
The maintenance items listed in the Power ON pipe and the Sleep Pipe must also be
performed every day.
e For more information, see:
Power ON pipe on page C-17
Sleep pipe on page C-17.
Roche Diagnostics
C-46 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Daily maintenance
At the end of analysis each day, the ISE flow path and the sample probe have to be
washed. This is a combined maintenance procedure for both ISE and photometric
unit.
The inside of the sample probe is cleaned as well as the ISE flow path and the
electrodes; after cleaning, the electrodes are reconditioned with Activator.
This maintenance action comprises the following procedures:
1. To set the number of washing cycles
2. To check the detergents and ISE reagents for the amounts remaining
3. To wash ISE flow path, ISE electrodes and the sample probe (maintenance item
(10) ISE Wash)
Operator time: approximately 1 minute
System time: approximately 29 minutes
o Roche recommends performing the cleaning of the ISE flow path and the sample probe
and the conditioning of the electrodes within a sleep pipe. Then you do not have to
perform this maintenance step manually. Make sure that there are sufficient quantities
of reagent remaining.
o Even if you are not using the ISE unit, you must nevertheless perform a wash after daily
operation. You can use either ISE Wash or Sample Probe Wash to do this.
e See Sleep Pipe function on page C-16.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-47
14 Maintenance of the analyzer cobas c 311 analyzer
Daily maintenance
a To check the detergents and ISE reagents for the amounts remaining
1 Make sure that the analyzer is in Standby.
2 Choose System Overview > Reagent Overview button
3 Check the detergents and ISE reagents for the amounts remaining in the Reagent
Overview screen.
o If inadequate, replace detergent or reagent bottles.
e See To replace ISE IS or ISE Dil. on page B-52.
e See To replace ISE reference solution (ISE Ref.) on page B-53.
e See To replace a sample probe detergent bottle on page B-142.
o Set detergents onto sample disk. Pipette suitable amounts of SysClean and
Activator into sample cups as shown in Figure C-12.
Roche Diagnostics
C-48 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Daily maintenance
A
F
Roche Diagnostics
Operator’s Manual · Version 2.0 C-49
14 Maintenance of the analyzer cobas c 311 analyzer
Daily maintenance
a To wash ISE flow path, ISE electrodes and the sample probe
1 Make sure that the analyzer is in Standby.
2 Check the detergents and ISE reagents for the amounts remaining in the Reagent
Overview screen.
e See To check the detergents and ISE reagents for the amounts remaining on page C-48.
5 Select Maintenance (1) from the Maintenance Type list on the left.
6 Select (10) ISE Wash from the Maintenance Items list on the right.
7 Choose Select to open the ISE Wash window.
8 Choose Execute.
The inside of the sample probe is cleaned as well as the ISE flow path and the
electrodes.
Make sure that you calibrate the ISE unit before you resume routine analysis.
e For details on how to calibrate the ISE unit, see:
Requesting and cancelling calibrations manually on page B-157.
If ISE wash is interrupted for some reason, perform maintenance item (6) Wash
Reaction Parts. Afterwards perform maintenance item (7) Reagent Prime and select
IS+REF.
e For information on how to perform these maintenance items, see:
M5: Washing the reaction parts on page C-59
To prime lines on page C-90.
9 If DAT (Drugs of Abuse Testing) assays are measured on the analyzer, perform
M5: Washing the reaction parts on page C-59 additionally.
Roche Diagnostics
C-50 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Daily maintenance
At the end of analysis each day, clean the outside of the pipetter probes (sample
probe, reagent probe) and the ISE sipper nozzle to remove residual solution and
precipitation. This is a combined maintenance procedure for both ISE and
photometric unit.
Replace pipetter probes when they are bent or otherwise damaged.
When sample aspiration error occurs frequently or if an alarm for clogging of the
sample probe occurs and is not reset after cleaning (maintenance item (11) Sample
Probe Wash, page C-123), you must remove the sample probe and eliminate
clogging.
e For instructions on how to eliminate clogging, see:
M23: Replacing sample and reagent probes—elimination of clogging on page C-122
Roche Diagnostics
Operator’s Manual · Version 2.0 C-51
14 Maintenance of the analyzer cobas c 311 analyzer
Daily maintenance
Figure C-14 Cleaning the sample probe and the shield pipe
Use a new lint-free gauze pad for each probe to prevent cross-contamination. Be
careful not to soak the gauze pad so that liquid does not drip down into the reaction
cells.
3 Wipe (from top to bottom) the outside of the sample probe with gauze pads
moistened with alcohol.
4 Lift up the shield pipe and moving the sample probe by hand to a position above
the sample disk.
5 Also wipe the shield pipe with a gauze pad moistened with alcohol.
Roche Diagnostics
C-52 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Daily maintenance
Figure C-15 Cleaning the ISE sipper nozzle and the reagent probe
Roche Diagnostics
Operator’s Manual · Version 2.0 C-53
14 Maintenance of the analyzer cobas c 311 analyzer
Daily maintenance
10 Close the top cover of the analyzer and lock it (or: continue with the next
maintenance action using Maintenance mode).
11 Turn the maintenance switch back to Operation mode (or power up the analyzer
if it was in Shutdown status). The analyzer returns to Standby.
12 Perform a Reset in Utility > Maintenance to return all mechanical parts to their
home positions.
Roche Diagnostics
C-54 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Daily maintenance
At the end of analysis each day, clean the cell rinse nozzles. Regular cleaning prevents
contamination, crystal formation, and clogging.
Replace the nozzle tip if it has worn down. The replacement cycle is typically 1 to
2 years, depending on the conditions of use.
e For instructions on how to replace nozzle tip, see:
M24: Replacing nozzle tips on cell rinse nozzles on page C-131
Roche Diagnostics
Operator’s Manual · Version 2.0 C-55
14 Maintenance of the analyzer cobas c 311 analyzer
Daily maintenance
Figure C-16 Dismounting the cell rinse unit and cleaning the nozzles
3 Loosen the retaining screw of the cell rinse unit and lift off the entire unit.
4 Moisten a lint-free gauze pad with deionized water and gently wipe all tips of the
cell rinse nozzles in a downward motion.
5 If a nozzle is clogged, insert the probe cleaning wire (stainless steel wire, 0.5 mm
diameter) into the tip of the nozzle and eliminate the clogging.
6 Attach the cell rinse unit.
7 Close the top cover of the analyzer and lock it (or: continue with the next
maintenance action using Maintenance mode).
8 Turn the maintenance switch back to Operation mode (or power up the analyzer
if it was in Shutdown status). The analyzer returns to Standby.
9 Perform a Reset in Utility > Maintenance to return all mechanical parts to their
home positions.
Roche Diagnostics
C-56 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Daily maintenance
At the end of analysis each day, clean the outlet of the ISE drain port. Regularly
perform this maintenance to prevent the accumulation of crystals and clogging of the
drain port.
Operator time: approximately 3 minutes
Roche Diagnostics
Operator’s Manual · Version 2.0 C-57
14 Maintenance of the analyzer cobas c 311 analyzer
Weekly maintenance
Weekly maintenance
In this section you will find all the maintenance actions that must be performed at
least once a week.
The maintenance items listed in the weekly pipe must also be performed.
e For more information, see:
Weekly pipe on page C-18.
Roche Diagnostics
C-58 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Weekly maintenance
Contamination of the reaction cells or the waste solution flow path will result in
incorrect measurement results. Wash the reagent probe and reaction cells at least
once a week. Perform this maintenance action daily, if DAT (Drugs of Abuse
Testing) assays are performed on the analyzer. For more information please refer to
the instructions for use in cobas link of the assay.
After washing the reaction system, perform a cell blank measurement to check the
condition of the reaction cells.
Operator time: approximately 1 minute
System time: approximately 47 minutes (30 minutes for (6) Wash Reaction Parts and
17 minutes for (3) Cell Blank Measurement)
This maintenance action comprises the following procedures:
1. To wash the reaction parts (maintenance item (6) Wash Reaction Parts)
2. To perform a cell blank measurement (maintenance item (3) Cell Blank
Measurement)
3. To check the Cell Blank Measurement report
Roche Diagnostics
Operator’s Manual · Version 2.0 C-59
14 Maintenance of the analyzer cobas c 311 analyzer
Weekly maintenance
7 Choose Execute.
The procedure is finished when the analyzer returns to Standby.
8 Perform a cell blank measurement.
e See To perform a cell blank measurement on page C-60
Roche Diagnostics
C-60 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Weekly maintenance
CELL NO. 340 376 415 450 480 505 546 570 600 660 700 800
001 12953 11414 11152 10756 10519 10333 10057 9918 9705 9442 9283 8938
002 12972 11444 11170 10771 10534 10350 10072 9932 9717 9459 9299 8952
003 12973 11446 11172 10776 10538 10354 10075 9934 9719 9459 9297 8949
004 12976 11444 11172 10775 10538 10354 10075 9933 9716 9455 9296 8949
005 12991 11468 11191 10793 10555 10369 10091 9949 9734 9475 9314 8963
006 12985 11460 11184 10790 10553 10367 10088 9947 9730 9469 9309 8959
007 12975 11450 11177 10780 10544 10359 10080 9938 9722 9462 9304 8954
008 12949 11413 11148 10752 10513 10327 10049 9909 9695 9436 9276 8930
009 12970 11451 11175 10779 10540 10356 10079 9937 9723 9465 9307 8956
010 12964 11433 11165 10771 10532 10346 10068 9927 9712 9449 9290 8940
… … … … … … … … … … … … …
065 12950 11423 11157 10763 10526 10341 10063 9922 9709 9446 9288 8943
066 12965 11440 11169 10775 10538 10354 10077 9937 9723 9462 9303 8956
Table C-21 Example of a Cell Blank Measurement report with Abnormal Cell List
Roche Diagnostics
Operator’s Manual · Version 2.0 C-61
14 Maintenance of the analyzer cobas c 311 analyzer
Weekly maintenance
Two cell covers protect the contents of the reaction cells from light. Some reactions
are light sensitive and intrusion of light could lead to faulty measurements. The cell
covers also serve to prevent contamination by reagents. If reagent adheres to the front
or rear face of a cell cover, the analytical accuracy may decrease.
At the end of analysis each day, check the cell covers for contamination. Clean the cell
covers at least once a week.
Operator time: approximately 3 minutes
System time: approximately 2 minutes
Roche Diagnostics
C-62 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Weekly maintenance
6/7
A U-cover
4 Wipe the front and rear faces of the cell covers using a gauze pad moistened with
alcohol.
5 Wipe the openings of cell covers using a cotton swab moistened with alcohol.
6 Pinch the sides of the U-cover (A) and pull it off the cell cover.
7 Wash the U-cover with deionized water. In case dust and dirt do not come off,
wipe the U-cover using a cotton swab moistened with alcohol.
In case you are also performing monthly maintenance, we recommend
continuing with the following action:
e See M8: Replacing reaction cells and cleaning incubator bath and drain filter on
page C-68.
8 Attach the dried U-cover and put the cell covers back in place.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-63
14 Maintenance of the analyzer cobas c 311 analyzer
Weekly maintenance
9 Close the top cover of the analyzer and lock it (or: continue with the next
maintenance action using Maintenance mode).
10 Turn the maintenance switch back to Operation mode (or power up the analyzer
if it was in Shutdown status). The analyzer returns to Standby.
11 Perform a reset in Utility > Maintenance to return all mechanical parts to their
home positions.
Roche Diagnostics
C-64 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Weekly maintenance
Clean the rinse stations of the sample probe and the reagent probe at least once a
week to prevent bacterial growth or precipitation that may clog the rinse stations.
Operator time: approximately 5 minutes
System time: approximately 2 minutes
Roche Diagnostics
Operator’s Manual · Version 2.0 C-65
14 Maintenance of the analyzer cobas c 311 analyzer
Weekly maintenance
5
3 6
B C
3 Using a cotton swab moistened with 2% Hitergent solution, clean the inside of the
sample probe rinse station.
Be careful not to splash liquid (water or Hitergent) into the drying cylinder.
4 Inject about 10 mL of 2% Hitergent solution into the sample probe rinse station.
5 Inject about 100 mL of deionized water into the rinse station to rinse it.
6 Wipe the drying cylinder (used for vacuum suction in the sample probe rinse
stations) with a cotton swab moistened with alcohol.
Roche Diagnostics
C-66 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Weekly maintenance
3
5
6
A B
7 Clean the rinse station and the drying cylinder for the reagent probe in the same
way as you clean the sample probe rinse station.
8 Close the top cover of the analyzer and lock it.
9 Turn the maintenance switch back to Operation mode (or power up the analyzer
if it was in Shutdown status). The analyzer returns to Standby.
10 Perform a reset in Utility > Maintenance to return all mechanical parts to their
home positions.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-67
14 Maintenance of the analyzer cobas c 311 analyzer
Monthly maintenance
Monthly maintenance
In this section you will find all the maintenance actions that must be performed at
least once a month.
M8: Replacing reaction cells and cleaning incubator bath and drain filter
Replace the reaction cells once a month because they gradually deteriorate over a
prolonged period of use. We recommend cleaning the incubator bath and incubator
bath filter at the same time.
Contamination inside the incubator bath (reaction bath) or on the photometric
window will reduce the reproducibility of measurement results. Clean the incubator
bath and photometric window at least once a month.
This maintenance action comprises the following procedures:
1. To remove reaction cells and to clean the incubator bath
2. To clean the drain filter of the incubator bath
3. To reinstall reaction parts and to terminate the incubator bath cleaning mode
Operator time: approximately 10 minutes
System time: approximately 47 minutes (30 minutes for (6) Wash Reaction Parts and
17 minutes for (3) Cell Blank Measurement)
Materials required m Hitergent cassette (surfactant for the incubator, consumption 4.3 mL)
m Lint-free gauze pads
m Deionized water or 2% Hitergent solution
Roche Diagnostics
C-68 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Monthly maintenance
Figure C-22 Removing cell rinse unit and segments of reaction cells
7 Remove the cell rinse unit, the sipper cover and the cell covers.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-69
14 Maintenance of the analyzer cobas c 311 analyzer
Monthly maintenance
8 Remove each segment of the reaction cells from the reaction disk. Be careful not
to touch the surfaces of the reaction cells.
11 10
Roche Diagnostics
C-70 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Monthly maintenance
1 Remove the incubator bath drain filter. Grasp the filter by its handle and pull the
filter out.
2 Wash the filter with tap water, rinse it with deionized water and return it to its
place.
In case you are only performing monthly maintenance, continue with the next
procedure: To reinstall reaction parts and to terminate the incubator bath
cleaning mode.
Otherwise we recommend continuing with the following items:
e For quarterly maintenance, see
M15: Cleaning the ultrasonic mixer on page C-93
e For every six months maintenance, see
M21: Replacing the photometer lamp on page C-113
Roche Diagnostics
Operator’s Manual · Version 2.0 C-71
14 Maintenance of the analyzer cobas c 311 analyzer
Monthly maintenance
Roche Diagnostics
C-72 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Monthly maintenance
Inspect the ISE Ref. aspiration filter, which is attached to the tube end in the ISE Ref.
bottle. Clean the filter each time you replace the ISE Ref. bottle but at least once a
month. Clogging of the filter will reduce the accuracy of ISE Ref. aspiration and data
reliability.
This maintenance action comprises the following procedures:
1. To clean the ISE Ref. aspiration filter
2. To reset the remaining volume of ISE Ref.
3. To prime the ISE Ref. tubing
Operator time: approximately 5 minutes
System time: approximately 2 minutes
Roche Diagnostics
Operator’s Manual · Version 2.0 C-73
14 Maintenance of the analyzer cobas c 311 analyzer
Monthly maintenance
3 Remove the tubing from the ISE Ref. bottle and unscrew the filter from the tube
end.
4 Wash the filter with tap water.
5 Rinse the filter with deionized water.
6 Replace the ISE Ref. bottle if necessary.
7 Finally, screw the filter on the tube end and place it back into the bottle.
8 Close the top cover of the analyzer and lock it.
9 Turn the maintenance switch back to Operation mode (or power up the analyzer
if it was in Shutdown status).The analyzer returns to Standby.
10 Perform a reset in Utility > Maintenance to return all mechanical parts to their
home positions.
11 If you have replaced the ISE Ref. bottle, reset the remaining volume.
e See: To reset the remaining volume of ISE Ref. on page C-75
Roche Diagnostics
C-74 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Monthly maintenance
Roche Diagnostics
Operator’s Manual · Version 2.0 C-75
14 Maintenance of the analyzer cobas c 311 analyzer
Monthly maintenance
Inspect the detergent aspiration filters, which are attached to the tube end in the cell
detergent bottles. The cell detergent bottles (CellCln I and CellCln II) are located
behind the left front door. Clean the filter each time you replace a bottle but at least
once a month. Clogging of the filter will reduce the accuracy of detergent aspiration
and will lead to insufficient cell cleaning.
This maintenance action comprises the following procedures:
1. To clean the detergent aspiration filters
2. To reset the remaining volume
3. To perform a cell detergent prime
Operator time: approximately 5 minutes
Roche Diagnostics
C-76 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Monthly maintenance
4
3
3 Remove the tubing from the detergent bottle(s) and unscrew the filter from the
tube end.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-77
14 Maintenance of the analyzer cobas c 311 analyzer
Monthly maintenance
Roche Diagnostics
C-78 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Monthly maintenance
Roche Diagnostics
Operator’s Manual · Version 2.0 C-79
14 Maintenance of the analyzer cobas c 311 analyzer
Monthly maintenance
Clean the radiator filter of the cooling unit at least once a month to prevent dust or
dirt accumulation. Clogging of the filter may cause the analyzer to overheat.
Operator time: approximately 5 minutes
3
5
Roche Diagnostics
C-80 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Monthly maintenance
Contamination inside the tank will result in contamination of the entire flow path
and adversely affect all measurements. Check the water tank at least once a month
and clean the tank if necessary.
Operator time: approximately 10 minutes
System time: approximately 2 minutes
Roche Diagnostics
Operator’s Manual · Version 2.0 C-81
14 Maintenance of the analyzer cobas c 311 analyzer
Monthly maintenance
A Closed tap
Roche Diagnostics
C-82 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Monthly maintenance
7 Remove the float assembly from the tank while pulling out the tank and place the
float on a paper towel.
8 Visually check that the water is clear. If not, perform thorough cleaning.
e See: To thoroughly clean the water tank on page C-84
9 If the water is clear, empty the water from the tank and rinse it with deionized
water three times.
10 Fill the tank at least 1/3 full of deionized water.
11 Reconnect the water tank.
e See: To reconnect the water tank on page C-84
Roche Diagnostics
Operator’s Manual · Version 2.0 C-83
14 Maintenance of the analyzer cobas c 311 analyzer
Monthly maintenance
3 Reconnect the water hose to the water tank and open the tap (Figure C-29).
4 Close left front door.
5 Turn on the external water supply.
6 Ensure the top cover is locked, before you switch on the analyzer again.
Roche Diagnostics
C-84 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Monthly maintenance
Roche Diagnostics
Operator’s Manual · Version 2.0 C-85
14 Maintenance of the analyzer cobas c 311 analyzer
Monthly maintenance
Roche Diagnostics
C-86 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Every two months maintenance
In this section you will find all the maintenance actions that must be performed at
least once every two months.
The electrical response level and the slope value (sensitivity) of each measuring
cartridge slightly decrease with time and use. Replace an ISE measuring cartridge if
one of the following criteria is met:
o The cartridge has been in service more than 2 months.
o The test count has reached 9000 tests.
o The slope value of the cartridge falls outside of the normal range.
In the latter case, an alarm is issued as described in the figure below.
+ 40 + 45 + 50 + 55 + 60 + 65 + 68 mV
- 68 - 65 - 60 - 45 - 40 - 35 - 30 mV
Figure C-32 ISE data alarms and corresponding slope values (EMF values)
Roche Diagnostics
Operator’s Manual · Version 2.0 C-87
14 Maintenance of the analyzer cobas c 311 analyzer
Every two months maintenance
System time: approximately 56 minutes (10 minutes for (7) Reagent Prime (IS+REF),
2*18 minutes for (2) ISE Check, plus 10 minutes in between)
3 Remove the sipper cover and the cover of the ISE measuring compartment.
4 Disconnect the three electrode wires (Cl, K and Na).
Roche Diagnostics
C-88 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Every two months maintenance
5 Pull the release lever toward the RELEASE position to loosen the cartridges from
the mounting block. The lever is equipped with a spring—take care of your
fingers.
6 Use large tweezers to remove the cartridges from the mounting block.
7 During the replacement:
o Thoroughly wipe up any spilled liquid or liquid adhering to connecting parts.
Remaining liquid on ion selective electrodes may lower measurement
precision.
o If an O-Ring from the cartridges remains inside the ISE measuring
compartment, use tweezers to remove it.
8 Insert the new cartridges in the right color-coded position into the ISE measuring
compartment while holding the lever at RELEASE.
Verify that the connecting part and the new cartridges are provided with O-rings, but
be sure to remove the black rubber protectors that are located over the O-ring
receptacles on the cartridges.
9 Set the lever at LOCK to fasten the cartridges, then reconnect the four color-
coded electrode cords to their respective cartridges.
10 Replace the sipper cover and the cover of the ISE measuring compartment.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-89
14 Maintenance of the analyzer cobas c 311 analyzer
Every two months maintenance
a To prime lines
e For information about reagent consumption, see:
(7) Reagent Prime on page C-25
Roche Diagnostics
C-90 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Every two months maintenance
6 After waiting ten minutes, execute another ISE check (10 cycles).
Make sure the printed EMF values of the internal standard solution are normal
and the difference in successive values for the same electrode are stable within
0.2 mV. Otherwise retry ISE check.
Normal IS EMF values lie between NA: -10 to -90 mV, K -10 to -90 mV and Cl 80
to 160 mV, Ref -7 to 7 mV. The maximum deviation for Ref EMF over all values
should be no more than 2 mV.
If an EMF value is abnormal (level error or noise error for example), the
corresponding alarm name will be printed out.
e See ISE data alarms and corresponding slope values (EMF values) on page C-87.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-91
14 Maintenance of the analyzer cobas c 311 analyzer
Quarterly maintenance
Quarterly maintenance
In this section you will find all the maintenance actions that must be performed at
least once every three months.
Roche Diagnostics
C-92 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Quarterly maintenance
Clean the ultrasonic mixer at least once every three months. Contamination and
precipitation on the surface of the ultrasonic mixer may cause inadequate mixing and
thus lead to inaccurate results.
This maintenance action can be combined with the monthly maintenance action M8:
Replacing reaction cells and cleaning incubator bath and drain filter on page C-68.
Do not open the top cover of the analyzer until the analyzer is in shutdown.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-93
14 Maintenance of the analyzer cobas c 311 analyzer
Quarterly maintenance
6/7
A
A Polished surfaces of the ultrasonic mixer
Figure C-36 Cleaning the surfaces of the ultrasonic mixer with cotton swabs
Roche Diagnostics
C-94 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Quarterly maintenance
In case you are also performing monthly maintenance action M8: Replacing reaction
cells and cleaning incubator bath and drain filter, continue with Step 6 of To reinstall
reaction parts and to terminate the incubator bath cleaning mode on page C-72.
In case you also want to perform every six months maintenance action M21:
Replacing the photometer lamp, continue with M21: Replacing the photometer lamp
on page C-113.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-95
14 Maintenance of the analyzer cobas c 311 analyzer
Quarterly maintenance
After long use, the tubing will gradually wear out and the accuracy of sample
aspiration will decrease. Replace the ISE pinch valve tubing once every three months.
Operator time: approximately 6 minutes
System time: approximately 46 minutes (2*18 minutes for (2) ISE Check, plus
10 minutes in between)
A Pinch valve
3 Remove the sipper cover and the cover of the ISE measuring compartment.
Roche Diagnostics
C-96 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Quarterly maintenance
4 Gently remove both ends of the ISE pinch valve tubing from their tube connectors
and pull the tubing through the pinch valve (A). Discard the old tubing.
5 Insert the new tubing through the pinch valve and attach both ends of the tubing
to the connectors. Make sure that there is no slack in the tubing.
In case you also want to replace the ISE sipper tubing, continue with Step 4 of
M17: Replacing the ISE sipper tubing on page C-98. Perform the ISE Check
afterwards.
6 Replace the sipper cover and the cover of the ISE measuring compartment.
7 Close the top cover of the analyzer and lock it.
8 Turn the maintenance switch back to Operation mode (or power up the analyzer
if it was in Shutdown status). The analyzer returns to Standby.
9 Perform a reset in Utility > Maintenance to return all mechanical parts to their
home positions.
10 Perform maintenance check (2) ISE Check (10 cycles) to condition the electrodes.
11 After waiting 10 minutes, perform maintenance check (2) ISE Check (10 cycles)
again. Check the printed EMF values. Then calibrate the ISE unit before you
resume routine analysis.
e See To perform an ISE Check and calibrate the ISE unit on page C-90.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-97
14 Maintenance of the analyzer cobas c 311 analyzer
Quarterly maintenance
After long use, the tubing will gradually wear out and the accuracy of sample
aspiration will decrease. Replace the ISE sipper tubing once every three months.
Operator time: approximately 6 minutes
System time: approximately 46 minutes (2*18 minutes for (2) ISE Check, plus
10 minutes in between)
Roche Diagnostics
C-98 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Quarterly maintenance
3 Remove the sipper cover, the arm cover of the sipper nozzle, and the cover of the
ISE measuring compartment.
4 Gently remove both ends of the ISE sipper tubing from the tube connectors.
Discard the old tubing.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-99
14 Maintenance of the analyzer cobas c 311 analyzer
Quarterly maintenance
6 Replace the arm cover of the sipper nozzle, the cover of the ISE measuring
compartment, and the sipper cover.
7 Close the top cover of the analyzer and lock it.
8 Turn the maintenance switch back to Operation mode (or power up the analyzer
if it was in Shutdown status). The analyzer returns to Standby.
9 Perform a reset in Utility > Maintenance to return all mechanical parts to their
home positions.
10 Perform maintenance check (2) ISE Check (10 cycles) to condition the
electrodes.
11 After waiting 10 minutes, perform maintenance check (2) ISE Check (10 cycles)
again. Check the printed EMF values. Then calibrate the ISE unit before you
resume routine analysis.
e See To perform an ISE Check and calibrate the ISE unit on page C-90.
Roche Diagnostics
C-100 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Quarterly maintenance
Replace the syringe seals every three months or after 112,500 samples for the reagent
and sample syringe or after 56,250 samples for the ISE sipper syringe. Worn-out
syringe seals can cause leakage and inaccurate pipetting.
There are three syringes behind the left front door of the cobas c 311 analyzer
(Figure C-39). The procedure for replacing the syringe seals is the same for all
syringes. Note, however, that there are different spare parts for each of the three
syringe types (sample, reagent, and ISE sipper syringe).
o The reagent syringe is used as an example in the rest of this section. All illustrations
show the reagent syringe.
o We recommend completing the procedures for the first syringe before starting with the
next. Do not dismount more than one syringe at the same time because the internal
parts of the syringes differ and may be mistaken easily with each other.
A
B
Roche Diagnostics
Operator’s Manual · Version 2.0 C-101
14 Maintenance of the analyzer cobas c 311 analyzer
Quarterly maintenance
Roche Diagnostics
C-102 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Quarterly maintenance
3 4 5
A
B
C
D
B
3 To disconnect the tubing at the top and bottom of the syringe, loosen and remove
the upper and lower tube connectors (A). Use a dry gauze pad to absorb any liquid
leaking from the tubing or syringe.
4 Loosen the syringe holder (B) and lift it up about 1 cm (lift it up over the
thread (D) of the syringe holder).
5 Remove the syringe (C) by carefully pulling it toward you.
6 Place the removed syringe on a dry gauze pad.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-103
14 Maintenance of the analyzer cobas c 311 analyzer
Quarterly maintenance
D
H
1 Loosen the retaining screw (G) with the spanner wrench (A). Be careful not to
bend the plunger.
2 Carefully remove the plunger (H) together with the retaining screw (G) and the
syringe seals (C, E) from the syringe barrel.
Roche Diagnostics
C-104 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Quarterly maintenance
I
E
G
I
3 Remove the upper seal piece (C), spacer (D), and lower seal piece (E) from the
plunger (H).
4 If the plunger, syringe or spacer are contaminated, wipe them with a gauze pad or
cotton swab moistened with deionized water.
5 Mount the retaining screw, spring, new lower seal piece, spacer, and new upper
seal piece onto the plunger as shown in Figure C-42.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-105
14 Maintenance of the analyzer cobas c 311 analyzer
Quarterly maintenance
6 Place the plunger (H) in the syringe barrel (B). Check that the spring (F) is
centered on the lower seal piece (E). If the spring is not centered, it could be
damaged during tightening.
7 Tighten the retaining screw with the spanner wrench.
6 Choose the syringe to be purged of air in the Syringe area and choose Execute.
The respective pipetter will operate. At the rinse station, water as well as any air
are discharged from the probe.
Roche Diagnostics
C-106 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Quarterly maintenance
1 Ensure the top cover is locked, before you switch on the analyzer again.
2 Choose Utility > Maintenance.
3 Select Maintenance (1) from the Maintenance Type list on the left.
4 Select (7) Reagent Prime from the Maintenance Items list on the right.
5 Choose Select to open the Reagent Prime window.
6 Choose Parameter to open the Reagent Parameter Setting window.
7 Choose IS+REF in the ISE area and OK.
8 Back in the Reagent Prime window, choose Execute.
9 After reagent prime, check connections following the instructions below.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-107
14 Maintenance of the analyzer cobas c 311 analyzer
Quarterly maintenance
Roche Diagnostics
C-108 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Quarterly maintenance
3 Inspect the syringe to ensure that no air remains in the syringe and no leaks are
visible at any of the fittings.
o If leakage is found at the syringe holder or tube connector, try to tighten.
o If leakage is found at the bottom of the syringe, try to attach again.
o If there are any air bubbles visible in the syringe, remove them by lightly
tapping your finger against the syringe while the liquid is flowing.
If bubbles cannot be removed, disconnect the syringe again (see To remove the syringe
on page C-103). Then wipe the plunger with gauze moistened with deionized water
(see To replace the syringe seals on page C-104).
Roche Diagnostics
Operator’s Manual · Version 2.0 C-109
14 Maintenance of the analyzer cobas c 311 analyzer
Every six months maintenance
In this section you will find all the maintenance actions that must be performed at
least once every six months.
Clean the inlet water filter at least once every six months to prevent clogging of the
water system.
Operator time: approximately 5 minutes
Roche Diagnostics
C-110 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Every six months maintenance
After cleaning, tighten the filter cap securely and make sure there is no water leakage.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-111
14 Maintenance of the analyzer cobas c 311 analyzer
Every six months maintenance
Clean the cooling fan at the rear of the analyzer to remove dust and dirt.
Operator time: approximately 5 minutes
Roche Diagnostics
C-112 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Every six months maintenance
(a) This is Power On Time, i.e. the time the analyzer is ON. You can check the Power on time of the
analyzer by looking at the printed Maintenance Report (Print > Utility > Maintenance Report).
Roche Diagnostics
Operator’s Manual · Version 2.0 C-113
14 Maintenance of the analyzer cobas c 311 analyzer
Every six months maintenance
If the current readings exceed 14000 at any wavelength, check the following points
and then replace the photometer lamp:
o Verify that the reaction cells, the incubator bath, and the photometric windows
are free of contamination or bubbles.
o Verify that the reaction cells are not scratched or cracked.
o Verify that reaction cell no. 1 is at least half filled with water.
If the current data value is quite different from the previous one, check if one of the
above applies.
Roche Diagnostics
C-114 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Every six months maintenance
If the reaction disk is detached with the reaction cells left in place, water drops
adhering to the outside of the reaction cells may drip onto the space surrounding the
photometer lamp. This may cause an alarm to be issued.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-115
14 Maintenance of the analyzer cobas c 311 analyzer
Every six months maintenance
6 Loosen the three retaining screws and remove the reaction disk from the system.
A Connector cover
7 Rotate the connector cover (A) counterclockwise and disconnect the connector of
the lamp wire.
8 Loosen the two lamp retaining screws and take out the photometer lamp.
When the screws become loose and turn easily, the lamp can be detached.
9 Carefully remove the retaining screws from the lamp base. These are needed to
install the new lamp.
Roche Diagnostics
C-116 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Every six months maintenance
2/3
Be careful not to touch the glass part of the new photometer lamp. If you do touch the
glass, wipe it off with a gauze pad.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-117
14 Maintenance of the analyzer cobas c 311 analyzer
Every six months maintenance
7 Close the top cover of the analyzer and lock it (or: continue with the next
maintenance action in Shutdown status).
8 Switch on the analyzer again.
9 Wait 30 minutes for the photometer lamp to stabilize and perform a cell blank
measurement before you resume routine operation. This is necessary to
compensate for a potential change in light intensity.
e For instructions, see To perform a cell blank measurement on page C-60.
In case you are also performing the monthly maintenance action M8: Replacing
reaction cells and cleaning incubator bath and drain filter, continue with Step 6 of
To reinstall reaction parts and to terminate the incubator bath cleaning mode on
page C-72. This includes a cell blank measurement.
Roche Diagnostics
C-118 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Every six months maintenance
Like any ISE cartridge, the ISE reference cartridge slowly deteriorates with use over
time. Eventually, the electromotive force (EMF) values of all cartridges may become
unstable. Therefore, replace the reference cartridge every six months.
This maintenance action comprises the following procedures:
1. To replace the ISE reference cartridge
2. To prime lines
3. To perform an ISE Check and calibrate the ISE unit
Operator time: approximately 5 minutes
System time: approximately 48 minutes (2 minutes for (7) Reagent Prime (REF),
2*18 minutes for (2) ISE Check, plus 10 minutes in between)
Roche Diagnostics
Operator’s Manual · Version 2.0 C-119
14 Maintenance of the analyzer cobas c 311 analyzer
Every six months maintenance
4
3
Figure C-51 Removing ISE covers and disconnecting the reference electrode wire
3 Remove the sipper cover and the cover of the ISE measuring compartment.
4 Disconnect the electrode wire of the ISE reference cartridge.
Roche Diagnostics
C-120 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
Every six months maintenance
6
7
Figure C-52 Removing the ISE reference cartridge and wiping up spilled liquid
5 Pull the release lever toward the RELEASE position to loosen the cartridges from
the mounting block.
6 Use tweezers to remove the reference cartridge from the mounting block.
7 During the replacement:
o Thoroughly wipe up any spilled liquid or liquid adhering to connecting parts.
Remaining liquid on ion selective electrodes may lower measurement
precision.
o If an O-Ring from the cartridges is remaining inside the ISE measuring
compartment, use tweezers to remove it.
8 Insert the new ISE reference cartridge into the ISE measuring compartment while
holding the lever at RELEASE.
Verify that the connecting part and the new cartridge is provided with O-rings, but be
sure to remove the black rubber protectors that are located over the O-ring receptacles
on the cartridge.
9 Set the lever at LOCK to fasten the cartridge, then reconnect the wire to the
cartridge.
10 Replace the sipper cover and the cover of the ISE measuring compartment.
11 Go through the following procedures before you resume routine operation. Select
the REF option for the reagent prime.
e See: To prime lines on page C-90
To perform an ISE Check and calibrate the ISE unit on page C-90
Roche Diagnostics
Operator’s Manual · Version 2.0 C-121
14 Maintenance of the analyzer cobas c 311 analyzer
As needed maintenance
As needed maintenance
In this section you will find all the maintenance actions that must be performed as
needed and are not subject to a regular time schedule.
e This section discusses the following maintenance actions:
M23: Replacing sample and reagent probes—elimination of clogging on page C-122
M24: Replacing nozzle tips on cell rinse nozzles on page C-131
M25: Draining the vacuum tank on page C-133
It is necessary to remove a probe to clean its inside, eliminate clogging, as well as for
replacement. Replace the sample, ISE or reagent probes when they are bent or
otherwise damaged.
This maintenance action comprises the following procedures:
1. To wash the sample probe
2. To remove the sample probe
3. To remove the reagent probe
4. To eliminate clogging
5. To put back the sample probe
6. To put back the reagent probe
7. To check the tubing system for leakage
8. To perform an air purge and check the operation of the probe
9. To perform a Mechanisms Check
Operator time: approximately 7 minutes
System time: approximately 12 minutes
Materials required m Sample Cleaner 1 (SmpCln 1): approximately 300 L (in case the sample probe is
to be washed)
m Sample or reagent probe (in case replacement is required)
m Sample probe seal (in case it has been pulled-off and the probe is not replaced)
m Cleaning wire (stainless steel) of 0.2 and 0.5 mm diameter
m Deionized water
Roche Diagnostics
C-122 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
As needed maintenance
Roche Diagnostics
Operator’s Manual · Version 2.0 C-123
14 Maintenance of the analyzer cobas c 311 analyzer
As needed maintenance
5
4
A Probe seal
3 Remove the pipetter arm cover by pressing the arm cover releases—located on
either side of the fixed bottom part of the pipetter arm—and gently lift up the
cover. This exposes the wires and the tubing.
4 Disconnect the wire of the liquid level sensor.
5 Loosen the tubing from the probe by unscrewing it, taking care not to drop and
lose the probe seal (A). Check that the probe seal stays either on the end of the
probe tubing or in the retaining nut.
6 Lift the probe from the probe arm.
Roche Diagnostics
C-124 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
As needed maintenance
Roche Diagnostics
Operator’s Manual · Version 2.0 C-125
14 Maintenance of the analyzer cobas c 311 analyzer
As needed maintenance
a To eliminate clogging
1 To eliminate clogging use probe cleaning wire (stainless steel) with the following
diameters:
o For the sample probe: 0.2 mm diameter
o For the reagent probe: 0.5 mm diameter
2 Insert the probe cleaning wire into the probe tip and run it through the probe.
Roche Diagnostics
C-126 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
As needed maintenance
A B C
A Foam cushion with holding slit for tubing C Tab for aligning the pipetter in the
B Probe seal pipetter arm
Roche Diagnostics
Operator’s Manual · Version 2.0 C-127
14 Maintenance of the analyzer cobas c 311 analyzer
As needed maintenance
Roche Diagnostics
C-128 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
As needed maintenance
6 In the Syringe area select the appropriate option (Sample Syringe, Reagent
Syringe or All).
7 Choose Execute to initiate the air purge.
The respective pipetter will operate (System time: 1:45). At the rinse station, water
as well as any air which may have become trapped are discharged.
8 With top cover closed, visually check that water is dispensed from the tip of the
probe in a straight flow.
If water is not dispensed in a straight flow, eliminate clogging of the probe.
e For more information, see:
M23: Replacing sample and reagent probes—elimination of clogging on page C-122.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-129
14 Maintenance of the analyzer cobas c 311 analyzer
As needed maintenance
Roche Diagnostics
C-130 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
As needed maintenance
Replace the nozzle tips on the cell rinse nozzles if they are worn. The replacement
cycle is typically one to two years depending on conditions of use.
Replace a nozzle tip if its corner or bottom is worn so that water remains in the
reaction cell.
Operator time: approximately 6 minutes
System time: approximately 10 minutes
Figure C-59 Dismounting the cell rinse unit and the nozzle tips
3 Loosen the retaining screw of the cell rinse unit and lift off the entire unit.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-131
14 Maintenance of the analyzer cobas c 311 analyzer
As needed maintenance
4 Grip the nozzle tip with radio pliers and pull it off.
7
A
A Guide bar
5 Attach a new nozzle tip by hand and align it with the guide bars (A).
6 Attach the cell rinse unit.
7 Make sure that the new nozzle tip is correctly aligned with the reaction cell.
8 Close the top cover of the analyzer and lock it.
9 Turn the maintenance switch back to Operation mode (or power up the analyzer
if it was in Shutdown status). The analyzer returns to Standby.
10 Perform a reset in Utility > Maintenance to return all mechanical parts to their
home positions.
11 Perform maintenance check (3) Mechanisms Check (30 cycles).
e See To perform a Mechanisms Check on page C-130.
Roche Diagnostics
C-132 Operator’s Manual · Version 2.0
cobas c 311 analyzer 14 Maintenance of the analyzer
As needed maintenance
At the end of analysis each day, we recommend checking the vacuum tank. An alarm
will be issued (Liquid in vacuum tank) when there is water or waste solution in the
vacuum tank. If this alarm appears, the vacuum tank must be drained. If this happens
frequently, contact your local service representative.
Operator time: approximately 5 minutes
3 Remove the cap holding the drain tube of the vacuum tank.
4 Drain the waste solution into a beaker.
5 Reattach the cap to secure the drain tube.
Roche Diagnostics
Operator’s Manual · Version 2.0 C-133
14 Maintenance of the analyzer cobas c 311 analyzer
As needed maintenance
Spills on the instrument surface could be biohazardous and may damage the surface.
Clean up all spills immediately. Use this procedure to ensure that all surfaces are
clean.
Roche Diagnostics
C-134 Operator’s Manual · Version 2.0
Troubleshooting D
Data alarms
This chapter contains the data alarms generated by the cobas c 311 analyzer. In
addition to a description of the cause of each alarm and remedy, the analytical unit of
origin and automatic rerun conditions are described here.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-3
15 Data alarms cobas c 311 analyzer
Table of contents
Roche Diagnostics
D-4 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Table of contents
Roche Diagnostics
Operator’s Manual · Version 2.0 D-5
15 Data alarms cobas c 311 analyzer
Table of contents
Roche Diagnostics
D-6 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Introduction
Introduction
If any abnormality occurs while the analyzer is switched on, the analyzer notifies the
operator of the potential problem by issuing an alarm. These alarms comprise data
alarms (referring to irregular measurement results or conditions) and instrument
alarms (referring to irregular system conditions). Alarms are classified into five levels:
Alarm levels Data alarm This alarm is attached to the results of calibrations or to measurement
results of QC or patient samples. If a data alarm occurs at calibration or
QC sample measurement and if it will affect subsequent measurements,
an instrument alarm of the warning level occurs at the same time. The
analyzer does not stop operation.
Warning This alarm is generated by data alarms or troubles on the instrument. If
this alarm occurs during operation, the analyzer does not stop operation.
The operator must judge whether to continue or interrupt measurement.
Sampling stop This alarm concerns troubles on the instrument. For the pipetted sample,
measurement is continued. The analyzer stops pipetting new samples.
Stop This alarm concerns troubles on the instrument. The analyzer stops
operation within one cycle (12 seconds). For the sample under
measurement, no result data are obtained and the measurement must be
repeated.
Emergency stop This alarm concerns troubles on the instrument. The analyzer
immediately stops operation. For the samples under measurement, no
result data are obtained and the measurement must be repeated.
Alarm indication Data alarms are indicated on the Workplace > Data Review screen, on the Test
Review window (Workplace > Data Review > Test Review), on printed reports, and
are sent to Host. If a data alarm occurs, a symbol (also referred to as flag) is attached
to the measurement result. These flags are strings of three to six characters, which are
all explained in this chapter.
Instrument alarms are indicated by the Alarm button (global button) on the System
Overview screen as well as by an audible alarm. If an alarm occurs, the Alarm button
flashes. Its color indicates the alarm level: Yellow indicates warning level, red
indicates sampling stop, stop, or emergency stop.
In case of an alarm choose the Alarm button to open the Alarm window. This
window provides an alarm list and descriptions and remedies for each listed alarm.
Automatic rerun If a data alarm occurs on a test for which automatic rerun is selected (check box on
Utility > Application > Range), a rerun is automatically requested. For execution of
the automatic rerun during measurement, however, automatic rerun needs to be
activated on the Start Conditions screen as well.
For QC and patient samples, a data alarm consists of an alarm message and a result
message. The result messages are specific to the cobas c 311 analyzer tests.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-7
15 Data alarms cobas c 311 analyzer
Data alarm list
The following table displays the data alarms applicable to the ISE unit, the
photometric unit, and their presentation (flag) on screens and reports.
Roche Diagnostics
D-8 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarm list
Roche Diagnostics
Operator’s Manual · Version 2.0 D-9
15 Data alarms cobas c 311 analyzer
Data alarms (ISE unit)
ADC.E
Remedy 1 If other instrument alarm exist, correct those alarms and resume operation.
2 Choose Utility > Maintenance and perform maintenance item (1) Reset. If the
alarm recurs call technical support.
Calc.?
Roche Diagnostics
D-10 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarms (ISE unit)
Remedy 1 Check the test that is flagged with an error message in the calculation. Dilute the
sample and rerun the sample.
2 Check the calibration type list box on Utility > Application > Calib.
3 Resume operation. If alarm recurs, call technical support.
Cal.E appears on each QC and patient sample for the affected test until the problem is
resolved.
Cause Any alarm, for example, Std.E, ISE.E, ISE.N, …, occurred during calibration.
Remedy 1 Correct the condition causing the alarm that occurred during the latest
calibration.
2 Recalibrate.
ClcT.E
Cause A data alarm has occurred for a test needed in the calculation. This is not valid for the
following errors and alarms:
o Calculation not possible (Calc.?)
o Unable to calculate compensated test (CmpT.?)
Remedy 1 Correct the data alarm of the test to be used on the calculation.
2 Recalibrate.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-11
15 Data alarms cobas c 311 analyzer
Data alarms (ISE unit)
CmpT.?
Description During test-to-test compensation calculation, the denominator became zero. Blank
space is left in the report.
Cause o The test used for test-to-test compensation has not been measured yet.
o Any test used for test-to-test compensation has data alarm calculation not
possible (Calc.?) or data error in compensated test (CmpT.E).
o Any test used in the compensation formula has a data alarm that leaves the result
blank (for example, Samp.S, Reag.S).
Remedy 1 Correct the data alarm of the test to be used for compensation.
2 Rerun the sample.
CmpT.E
Description 1 In test-to-test compensation calculation, a data alarm other than those shown
below is indicated for the compensation test data.
2 Calculation not possible, test-to-test compensation disabled, overflow.
Remedy Correct the data alarm on the test to be used for compensation. Rerun the sample.
Edited
Description An edited first result or replaced rerun result is marked with an Edited alarm on the
Data Review screen. This also prints on the patient report.
Roche Diagnostics
D-12 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarms (ISE unit)
H (only on printouts)
Description For patient samples, the calculated concentration is greater than the upper limit of
the expected value range.
For control samples, a concentration exceeded the 2 SD values specified on QC >
Install.
>Index
Description In one or more tests, the L, H, I value is larger than the specified reference value
(Utility > Application > Range).
The serum index check is performed after all test results are generated. If a data alarm
is already attached to the result, no serum index check will be performed.
ISE.E
Description During measurement of internal reference, the EMF was not within the following
ranges (Internal standard solution):
Roche Diagnostics
Operator’s Manual · Version 2.0 D-13
15 Data alarms cobas c 311 analyzer
Data alarms (ISE unit)
Remedy 1 Set the reagent and perform maintenance item (7) Reagent Prime for the ISE unit
(IS+REF).
2 Replace the electrode and perform maintenance item (7) Reagent Prime for the
ISE unit (IS+REF).
3 Make sure that the O-ring is attached to the electrode joint and then perform
maintenance item (7) Reagent Prime for the ISE unit (IS+REF).
4 Set the internal standard solution (ISE IS) properly and perform maintenance
item (7) Reagent Prime for the ISE unit (IS+REF).
5 Connect the cord correctly.
ISE.N
Description In ISE measurement, the fluctuation in electromotive force exceeds the following
value:
Na+ 0.7 mV
K+ 1.0 mV
Cl- 0.8 mV
Remedy 1 Check for sufficient reagent volume, make sure the tubing sits correctly in the
ISE Ref. bottle, and perform maintenance item (7) Reagent Prime for the ISE unit
(IS + REF).
2 Make sure that the O-ring is attached to the electrode joint and perform
maintenance item (7) Reagent Prime for the ISE unit (IS + REF).
3 Replace the sipper tube and perform maintenance item (7) Reagent Prime for the
ISE unit (IS + REF).
4 Eliminate deposits.
5 Carry out cleaning, checkup, and maintenance of the sipper syringe.
6 Make sure that the ISE compartment cover and the cover of the ISE sipper are
reinstalled after maintenance.
Roche Diagnostics
D-14 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarms (ISE unit)
L (only on printouts)
Description For patient samples, the calculated concentration is less than the lower limit of the
expected value range. For control samples, a concentration was less than the 2 SD
values entered on QC > Install.
<Mix
Mix.E
Over.E
Alarm Overflow
Description Display is not possible because the output figure exceeds 6 digits
Cause o The data is obtained in more than six digits including a negative sign and decimal
point.
Remedy o Minimize the difference in concentration between the sample and calibrator.
o Obtain the result data with six characters including negative sign and decimal
point.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-15
15 Data alarms cobas c 311 analyzer
Data alarms (ISE unit)
Reag.S
Description There is insufficient ISE reagent volume when alarm is associated with sodium,
potassium and chloride values.
Remedy 1 Verify adequate reagent volumes. Replace low reagent, as necessary. Update the
new reagent volume manually. Prime new reagent and recalibrate.
2 If adequate reagent volumes are present, verify volumes on the Reagent screen.
Update the values manually, if necessary. Rerun the sample.
3 Resume operation. If alarm recurs, call technical support.
>Rept / <Rept
Description The result falls outside the repeat limit range programmed on Utility >
Application > Range.
Cause The sample concentration is higher (>Rept) or lower (<Rept) than the set value.
Remedy This alarm can be activated on Utility > System (Page 1/5) > Alarm. If activated, the
analyzer can also be programmed to automatically repeat this test with normal
sample volume.
Samp.S
Roche Diagnostics
D-16 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarms (ISE unit)
Samp.C
Remedy 1 Perform maintenance procedure To wash the sample probe on page C-123.
2 If alarm recurs, remove the sample probe, clean it manually, and eliminate clog. If
a probe is bent or damaged, replace it with a new one.
e See M23: Replacing sample and reagent probes—elimination of clogging on page C-122
>Test
Description If the data is out of following ranges, ISE range over is issued.
4 Resume operation.
If alarm recurs, call technical support.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-17
15 Data alarms cobas c 311 analyzer
Data alarms (ISE unit)
>Test / <Test
Description The measured value is outside the technical limit range entered on Utility >
Application > Range.
o Over the technical limit: Value is greater than the upper limit (>Test).
o Under the technical limit: Value is less than the lower limit (<Test).
Cause 1 The sample concentration is higher than the set value (>Test).
2 The sample concentration is lower than the set value (<Test).
3 An improper technical limit range is specified.
Remedy 1 Manually dilute and rerun the diluted sample until the measured concentration is
within the specified range. Be sure to calculate the original concentration from the
measured value, using the correct dilution factor.
In case of urine sample s with a <Test alarm, a manual rerun can be performed
with decreased volume (6.5 L instead of normal volume).
2 Perform maintenance item (11) Sample Probe Wash.
3 Clean the sample probe manually.
4 Call technical support.
Roche Diagnostics
D-18 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarms (photometric unit)
>Abs
Description The absorbance value to be used for calculation after cell blank correction exceeded
3.3.
Remedy o If only one sample is affected: Check whether the sample is grossly lipemic or has
an extremely high value. Follow your laboratory protocol for this situation.
o If only one application is affected: Check reagent preparation for that test.
o If all samples are affected:
O Remove any obstructions in the optical path of the photometer. Make sure the
lamp is on.
O Choose Utility > Maintenance and perform maintenance item
(4) Incubation Water Exchange
O Clean the incubator bath if it is contaminated.
e For further instructions, see:
M8: Replacing reaction cells and cleaning incubator bath and drain filter on
page C-68.
O Choose Utility > Maintenance and perform maintenance item
(2) Photometer Check.
O Check the Photometer Check results on the printout (Abs < 14000).
O Replace the photometer lamp, if necessary.
O Resume operation. If alarm recurs, call technical support.
o If all samples are affected sporadically:
O Check if some reaction cells are scratched. Replace all cells if necessary.
e For further instructions, see:
M8: Replacing reaction cells and cleaning incubator bath and drain filter on
page C-68
Roche Diagnostics
Operator’s Manual · Version 2.0 D-19
15 Data alarms cobas c 311 analyzer
Data alarms (photometric unit)
ADC.E
Remedy 1 If other instrument alarms exist, correct those alarms and resume operation.
2 Choose Utility > Maintenance and perform maintenance item (1) Reset.
3 If the alarm recurs, remove the reaction disk and check for water droplets or dust
on sensors. Clean if necessary.
4 If the alarm recurs call technical support.
Calc.?
Cal.E appears on each QC and patient sample for the affected test until the problem is
resolved.
Remedy 1 Correct the condition causing the alarm that occurred during the latest
calibration.
2 Recalibrate.
Roche Diagnostics
D-20 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarms (photometric unit)
ClcT.E
e Cause and remedy are identical to the data alarm of the ISE unit.
See ClcT.E on page D-11.
CmpT.?
Description During test-to-test compensation calculation, the denominator became zero. Blank
space is left in the report.
e Cause and remedy are identical to the data alarm of the ISE unit.
See CmpT.? on page D-12.
CmpT.E
Description 1 In test-to-test compensation calculation, a data alarm for other than those shown
below is indicated for the compensation test data.
2 Calculation not possible (Calc.?), data error in compensated test (CmpT.E),
overflow (Over.E), random error (R4SD), systematic error (S2-2Sa, S2-2Sw,
S4-1Sa, S4-1Sw, S10Xa, S10Xw), QC error (Q2.5SD, Q3SD), and outside of
expected value (H/L only on the printout).
e Cause and remedy are identical to the data alarm of the ISE unit.
See CmpT.E on page D-12.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-21
15 Data alarms cobas c 311 analyzer
Data alarms (photometric unit)
>Cuvet
Description For a diluted test the cell blank value of the second reaction cell exceeds the reference
by more than 0.1Abs. The reference is the stored value from the (3) Cell Blank
Measurement.
If a reaction cell fails the cell blank measurement, the system excludes it from any
measurements.
For diluted tests however, the system uses two reaction cells:
o 1st cell: Contains sample and diluent.
o 2nd cell: Used for discharging diluted sample from the 1st cell, adding reagent,
and performing measurement.
The >Cuvet alarm indicates a failed cell blank measurement for the 2nd cell.
e For more information about cell blank measurement, see
Cell rinse unit on page A-53
Flow of photometric analysis on page A-55
Cell Blank Measurement window on page E-284
Roche Diagnostics
D-22 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarms (photometric unit)
Det.S
Description If a special wash for the reagent probe is defined, the reagent probe plunges into the
cobas c pack with the specified detergent. This alarm occurs, if the test is carried out
even though the detergent volume is insufficient.
Edited
Description An edited first result or replaced rerun result is marked with an Edited alarm on the
Data Review screen. This also prints on the patient report.
H (only on printouts)
>Index
Description In one or more tests, the L, H, I value is larger than the specified reference value
(Utility > Application > Range).
The serum index check is performed after all test results are generated. If a data alarm
is already attached to the result, no serum index check will be performed.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-23
15 Data alarms cobas c 311 analyzer
Data alarms (photometric unit)
>Kin
Description The prozone check value exceeds the specified limit value. (Reaction rate method)
Remedy 1 Dilute and rerun the sample or rerun the sample with a decreased sample volume.
2 To avoid check, set [0] [0] [0] [0] [0] [0] [Inside] [0] [0] for Prozone Limit (on
Utility > Application > Analyze).
>Lin
Description In a rate assay, the reaction linearity exceeds the specified limit value.
L (only on printouts)
Roche Diagnostics
D-24 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarms (photometric unit)
Mix.E
<Mix
Description The ultrasonic monitor value is lower than the reference value.
Over.E
Alarm Overflow
Description Display is not possible because the output figure exceeds 6 digits
Cause o The K factor has more than six digits due to the use of a wrong calibrator.
o The data is obtained in more than six digits including a negative sign and decimal
point.
Remedy o Minimize the difference in concentration between the sample and calibrator.
o Obtain the result data with six characters including negative sign and decimal
point.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-25
15 Data alarms cobas c 311 analyzer
Data alarms (photometric unit)
>Proz
Description The prozone check value exceeds the specified limit value. (Antigen readdition
method)
>React
Description In a rate assay, main wavelength absorbance exceeds the specified limit value.
Remedy 1 Verify the setting in the Increase/Decrease box on Utility > Application >
Analyze.
2 Dilute and rerun the sample or rerun the sample with a decreased sample volume.
3 Prepare the reagent newly.
ReagEx
Description The alarm indicates that an expired reagent was used; the test result is not
guaranteed. The alarm can be inactivated under Utility > Application > Alarm.
Remedy 1 Expired reagents are listed on the reagent unload list with the ReagEx alarm.
2 Exchange expired reagents.
Roche Diagnostics
D-26 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarms (photometric unit)
>Rept / <Rept
Description The result falls outside the repeat limit range programmed on Utility >
Application > Range.
e Cause and remedy are identical to the data alarm of the ISE unit.
See >Rept / <Rept on page D-16.
Samp.?
Description The absorbance of a sample is found equal or greater than the theoretical maximum
absorbance (for infinite analyte concentration). The result field will be left blank on
the report and Data Review screen. This blank result is transmitted, together with the
alarm code “>” to the Host.
Remedy Dilute the sample, if required, and rerun. If automatic rerun is programmed, the
sample will be rerun with a decreased sample volume.
Samp.C
Samp.S
>Test / <Test
Description The measured value is outside the technical limit range entered on Utility >
Application > Range.
e Cause and remedy are identical to the data alarm of the ISE unit.
See >Test / <Test on page D-18.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-27
15 Data alarms cobas c 311 analyzer
Alarms for calibrations
The following table displays all calibration data alarms applicable to ISE tests,
photometric tests, and their presentation (flag) on screens and reports.
Description The current ISE calibrator concentration value or slope value differs from the
previous one by more than the specified Compensated Limit. The Compensated
Limit is a limit for the difference | previous value – current value | expressed as
percentage of the average (previous value + current value)/2.
The Cal.E alarm is a warning only, not necessarily indicative of a problem with the
calibration. Check the test’s control recovery before accepting the new calibration result.
Cause o The ISE calibrators or ISE reagents are not placed in proper position.
o The ISE reagents have deteriorated or the ISE calibrators have become
concentrated due to evaporation.
Roche Diagnostics
D-28 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Alarms for calibrations
Cond.E
Description The slope value is 68.1 mV or greater for Na+ or K+ electrodes, or it is -68.1 mV or
less for the Cl- electrode. The conditioning is inadequate.
+ 40 + 45 + 50 + 55 + 60 + 65 + 68 mV
- 68 - 65 - 60 - 45 - 40 - 35 - 30 mV
Figure D-1 ISE data alarms and corresponding slope values (EMF values)
Roche Diagnostics
Operator’s Manual · Version 2.0 D-29
15 Data alarms cobas c 311 analyzer
Alarms for calibrations
Dup.E
Description The difference between the first and second measurement (absorbance) of a
calibrator is outside the specified range.
The following steps describe how a decision is made to flag a calibration for violating
the duplication limit.
1 The absorbance for a calibrator (N) is measured twice.
2 The % of error and absorbance error are calculated.
3 Is the absorbance error < Duplicate Limit Abs. that appears on Utility >
Application?
If no, go to step 5.
If yes, go to step 4.
4 Continue with result calculations. No Dup.E alarm is issued.
5 Is the % error < the % Duplicate Limit?
If no, go to step 6.
If yes, go to step 4.
6 Dup.E alarm is issued for this result.
Cause The difference between the first and second measurement (absorbance) of a
calibrator is greater than Duplicate Limit Abs and greater than the % Duplicate Error.
Remedy 1 Recalibrate.
2 Check reagent preparation and expiration date. Prepare new reagent if necessary
and recalibrate.
3 If alarm recurs, call technical support.
If this alarm occurs, a Std.E alarm is issued. The Std.E alarm prevents updating of
calibration for the affected test.
Roche Diagnostics
D-30 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Alarms for calibrations
IStd.E
Description The concentration of the Internal Standard solution (ISE IS) was not within the
following ranges:
Remedy 1 If the EMF of ISE IS is abnormal on the calibration report, check the ISE IS
reagent volume and preparation. If necessary, prepare fresh reagent, prime and
recalibrate.
2 If the EMF of ISE IS is normal on the calibration report, check ISE reagent
syringe. The EMF of ISE IS lies ideally midway between the low and the high
standard. The ideal ISE IS concentrations values are: Na+: 140 mmol/L; K+:
5 mmol/L; Cl-: 100 mmol/L.
3 Resume operation. If alarm recurs, call technical support.
4 Perform maintenance item (10) ISE Wash to wash the flow path of the ISE unit.
e See M1: Washing ISE flow path and sample probe on page C-47
Roche Diagnostics
Operator’s Manual · Version 2.0 D-31
15 Data alarms cobas c 311 analyzer
Alarms for calibrations
Prep.E
Description The slope value is within the following range: 45.0 to 49.9 mV for Na+ or K+
electrodes, -39.9 to -35.0 mV for the Cl- electrodes.
+ 40 + 45 + 50 + 55 + 60 + 65 + 68 mV
- 68 - 65 - 60 - 45 - 40 - 35 - 30 mV
Figure D-2 ISE data alarms and corresponding slope values (EMF values)
If the alarm “ISE slope marginal (Prep.E)” appears, you may continue the day’s
analyses but the cartridges must be replaced the next day. Check the QC to verify that
the slope is acceptable
Remedy 1 Perform maintenance item (10) ISE Wash to wash the flow path of the ISE unit.
e See M1: Washing ISE flow path and sample probe on page C-47
Roche Diagnostics
D-32 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Alarms for calibrations
Rsp1.E
Description The A Factor (response compensation factor) is outside the following limits:
Rsp2.E
Roche Diagnostics
Operator’s Manual · Version 2.0 D-33
15 Data alarms cobas c 311 analyzer
Alarms for calibrations
S1A.E
Description During calibration, expected absorbance is outside the S1 Abs Limit. S1 is read
bichromatically for endpoint assays, monochromatically for rate assays.
SD.E
Description During nonlinear or multipoint linear calibration, the SD value was larger than the
SD limit programmed on Utility > Application.
Roche Diagnostics
D-34 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Alarms for calibrations
Sens.E
Description Sensitivity is checked for both linear (2 to 6 points) and nonlinear calibrations. This
alarm is indicated if the sensitivity value obtained in a calibration falls out of the
sensitivity limits specified on Utility > Application > Calib.
A sensitivity value is calculated from the measured absorbance values ( Abs ) and
given concentration values ( Conc ) of the blank calibrator ( S1 ) and the calibrator
SN :
S N = Std 2 for 2 Point calibrations and span calibrator for multipoint calibrations
For span calibration, the previous S1 Abs (linear) or previous mean absorbance (nonlinear)
of calibrator (1) is used for the sensitivity check.
Remedy 1 Check preparation and expiration dates of calibrators and reagents. Recalibrate
the affected test.
2 Check sample pipetter for leaks and recalibrate the affected test.
3 Choose Utility > Application > Calib. to check the sensitivity limit and
recalibrate the affected test.
4 Resume operation. If alarm recurs, call technical support.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-35
15 Data alarms cobas c 311 analyzer
Alarms for calibrations
Slop.E
Description The slope value is less than 45.0 mV for Na+ or K+ electrodes, or
greater than -35 mV for the Cl- electrode.
+ 40 + 45 + 50 + 55 + 60 + 65 + 68 mV
- 68 - 65 - 60 - 45 - 40 - 35 - 30 mV
Figure D-3 ISE data alarms and corresponding slope values (EMF values)
Remedy 1 Replace the appropriate ISE cartridge before continuing routine analysis.
e See M14: Replacing ISE measuring cartridges (Cl, K, Na) on page C-87.
Roche Diagnostics
D-36 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Alarms for calibrations
Std.E (ISE)
Description 1 During ISE calibration, any one of the following alarms was encountered:
The calibration is not updated if this alarm is issued; i.e. the prior calibration remains in
effect.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-37
15 Data alarms cobas c 311 analyzer
Alarms for calibrations
Std.E (P)
Description 1 During photometric calibration, any one of the following alarms was
encountered:
The calibration is not updated if this alarm is issued; i.e. the prior calibration remains in
effect.
Cause o The reagent has not been prepared properly or is empty and has to be replaced.
o The concentration of calibrator is improper or the calibrator is not placed in the
proper position.
o An improper check value is specified.
Roche Diagnostics
D-38 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarms for controls
Q3SD
Description In Realtime QC the control X or control Y data value is above 3 SD or below -3 SD.
This check is performed only when RULE 1-3 SD is selected (in QC > Individual >
Realtime QC > Rules).
Remedy 1 Check that controls and reagents are properly prepared, positioned, and stored.
2 Check correct lot number and expiration dates of controls and reagents.
3 Check that the control values (mean value, standard deviation) are entered
correctly on QC > Install.
4 Prepare a new control and analyze it.
5 Prepare a new calibrator and perform a calibration.
6 Load a new cobas c pack or replace with a new cobas c pack.
7 Check the pipetter (Maintenance Check).
e See M23: Replacing sample and reagent probes—elimination of clogging on page C-122
Roche Diagnostics
Operator’s Manual · Version 2.0 D-39
15 Data alarms cobas c 311 analyzer
Data alarms for controls
Q2.5SD
Description In Realtime QC the control X value or control Y data value is above 2.5 SD or below
-2.5 SD.
This check is performed only when RULE 1-2.5SD is selected (in QC > Individual >
Realtime QC > Rules).
R4SD
Description In Realtime QC one of X and Y data values is above 2 SD and the other is below
-2 SD.
This check is performed only when RULE R-4SD is selected. N = run size entered on
Select Rules window (in QC > Individual > Realtime QC > Rules).
S2-2Sa
Description In Realtime QC the control X and Y values are above 2 SD or below -2 SD.
This check is performed only when RULE 2-2SD is selected (in QC > Individual >
Realtime QC > Rules).
Roche Diagnostics
D-40 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data alarms for controls
S2-2Sw
Description In Realtime QC the last two control X or last two control Y values were above 2 SD or
below -2 SD.
Cause o Control values (mean value, standard deviation) are not correctly specified.
o The reagents are not properly prepared.
o Reagent is deteriorated (e.g., linearity of working curve degraded).
o Poor precision due to leakage of the pipetter joint.
This check is performed only when RULE 2-2SD is selected (in QC > Individual >
Realtime QC > Rules).
Remedy 1 Check that the control values (mean value, standard deviation) are entered
correctly on QC > Install.
2 Check that reagents are properly prepared, positioned, and stored.
3 Check correct lot number and expiration dates of reagents.
4 Load a new cobas c pack or replace with a new cobas c pack.
5 Prepare a new calibrator and perform a calibration.
6 Prepare a new control and analyze it.
7 Check the pipetter (Maintenance Check).
e See M23: Replacing sample and reagent probes—elimination of clogging on page C-122
S4-1Sa
Description In Realtime QC the last two X and Y data values are above +1 SD or below -1 SD.
This check is performed only when RULE 4-1SD is selected (in QC > Individual >
Realtime QC > Rules).
Roche Diagnostics
Operator’s Manual · Version 2.0 D-41
15 Data alarms cobas c 311 analyzer
Data alarms for controls
S4-1Sw
Description In Realtime QC the last four control X or last four control Y values are above 1 SD or
below -1 SD.
This check is performed only when RULE 4-1SD is selected (in QC > Individual >
Realtime QC > Rules).
S10Xa
Description In Realtime QC the last five X and Y data values fall on the + or - side of the mean
value.
This check is performed only when RULE 10X is selected (in QC > Individual >
Realtime QC > Rules).
S10Xw
Description In Realtime QC the last 10 control X or last 10 control Y values are positive-above the
mean value or negative-below the mean value.
This check is performed only when RULE 10X is selected (in QC > Individual >
Realtime QC > Rules).
Roche Diagnostics
D-42 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data problems without alarm
Remedy 1 Avoid leaving the sample in the sample cup for a long time.
2 Perform maintenance item (10) ISE Wash to wash the flow path of the ISE unit.
e See M1: Washing ISE flow path and sample probe on page C-47
Erroneous operation
Remedy 1 Carry out preliminary and/or periodical check according to the specified
procedure.
2 Change the channel, use the wash program, or take any other measure after
consulting with the reagent manufacturer.
3 Eliminate fibrin or dust. Be sure to check the sample and reagent before setting
them.
4 Perform maintenance item (11) Sample Probe Wash.
5 Use the recommended sample container.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-43
15 Data alarms cobas c 311 analyzer
Data problems without alarm
Poor reproducibility
Remedy 1 Carry out daily checks and periodical maintenance according to the specified
maintenance procedure.
2 Replace with a new cobas c pack.
3 Replace with new ISE reagents. Do not add or mix old and new reagent.
4 The water quality must be 1 S/cm (microsiemens per cm) or less.
5 Use the recommended reagent handling.
6 Check the completeness of the special wash list. If necessary, install the special
wash list according to the recommendation of the manufacturer.
7 Check for air bubbles and liquid leakage at the syringes.
o For sample syringe or reagent syringe perform an air purge (see To perform an
air purge (sample syringe or reagent syringe) on page C-106).
o For ISE sipper syringe perform a reagent prime (see To perform an air purge
(sample syringe or reagent syringe) on page C-106).
8 Perform maintenance item (10) ISE Wash to clean the ISE flow path or replace
the ISE measuring cartridges with new ones.
e See:
M14: Replacing ISE measuring cartridges (Cl, K, Na) on page C-87
M22: Replacing the ISE reference cartridge on page C-119
Roche Diagnostics
D-44 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Data problems without alarm
Remedy 1 Avoid leaving the sample or control in the sample cup for a long time.
2 Avoid leaving the calibrator in the sample cup open for a long time.
3 Replace with new ISE reagents. Do not add or mix old and new reagent.
4 Use the recommended reagent, control and calibrator handling.
Remedy 1 Use the calibrator immediately after opening the cap of the vial.
2 Prepare the reagent again by the correct method.
3 Set the concentration of calibrator, measuring wavelength, etc. properly.
4 Replace with new ISE reagents. Do not add or mix old and new reagent.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-45
15 Data alarms cobas c 311 analyzer
Data problems without alarm
Roche Diagnostics
D-46 Operator’s Manual · Version 2.0
cobas c 311 analyzer 15 Data alarms
Rerun list
Rerun list
Roche Diagnostics
Operator’s Manual · Version 2.0 D-47
15 Data alarms cobas c 311 analyzer
Rerun list
Roche Diagnostics
D-48 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
Table of contents
Troubleshooting
Roche Diagnostics
Operator’s Manual · Version 2.0 D-49
16 Troubleshooting cobas c 311 analyzer
Table of contents
Roche Diagnostics
D-50 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
General troubleshooting
General troubleshooting
Troubleshooting procedures
To identify and isolate problems effectively, you must understand the theory of
operation, operating procedures, emergency procedures and test reaction
descriptions covered in this manual. The following areas are covered:
The basic operator is not responsible for troubleshooting electrical problems except as
covered in this part of the Operator’s Manual.
When troubleshooting, observe and record the alarms and isolate the problem to the
area denoted by the alarms. Follow instructions of the alarm message and follow the
remedies herein. In many cases, you may be able to find the problem, correct it and
then resume processing. The remainder of this chapter provides instructions and
guidelines to aid in isolating problems.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-51
16 Troubleshooting cobas c 311 analyzer
General troubleshooting
Chemistry problem
o test(s) affected and other tests on board
o special wash programming
o catalog number, lot numbers and expiration dates of reagents
o catalog number, lot numbers and expiration dates of calibrators and controls
o lot number of ISE cartridges
o calibration monitor printout from the last few calibrations performed and/or
calibration trace printout
o control results from the last few controls performed
o patient results (with correlation results, if relevant)
o Reaction Monitor report for affected test(s) printed out via Print > Workplace >
Reaction Monitor
Instrument problem
o account number/customer ID number
o instrument serial number and software version
o description of the problem including relevant alarm(s) and alarm code numbers
o other instrument or maintenance related information.
Roche Diagnostics
D-52 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
General troubleshooting
Roche Diagnostics
Operator’s Manual · Version 2.0 D-53
16 Troubleshooting cobas c 311 analyzer
General troubleshooting
5 Press <Windows> + <e> on the control unit keyboard to access the Windows
Explorer.
6 Browse to C:\7600\GetLog and open the directory named after the respective
date.
7 Right-click on the file MERGE.TXT and select Rename. Change the file name to
MERGE.BAT and press <Enter>.
8 Choose Yes to confirm the confirmation window about changing the file name
extension.
The file name changes to MERGE.BAT.
Roche Diagnostics
D-54 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
General troubleshooting
Roche Diagnostics
Operator’s Manual · Version 2.0 D-55
16 Troubleshooting cobas c 311 analyzer
General instrument troubleshooting
This chapter provides information about troubleshooting procedures for the analyzer
as a whole. Procedures for troubleshooting general instrument problems as well as
problems during power up are described.
Troubleshooting at Power Up
Conditions that can affect instrument power up are presented in Table D-6.
To troubleshoot a problem, determine the category below that best describes the
problem, and follow the recommended remedy. If all remedies are unsuccessful, call
technical support.
Description: The instrument does not power up when the power switch of the analytical unit is
switched on.
e For the power switch of the analytical unit, see Power switches on page A-27
Roche Diagnostics
D-56 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
General instrument troubleshooting
The control unit controls and monitors all mechanical functions of the instrument.
When a mechanical problem arises within the instrument, it is immediately
recognized. The alarm indicator on the global Alarm button flashes, alerting you to
the problem. Touch Alarm (global button) to display the Alarm screen with the
specific alarm code, date and time the alarm occurred and a description of the alarm.
Touch a specific alarm to display the alarm details and appropriate remedy.
For certain problems affecting the instrument's performance, the analyzer terminates
the operation mode and enters the sampling stop or stop mode. In the sampling stop
mode, the analyzer allows completion of the samples in process unaffected by the
failure. In the stop mode the analyzer stops immediately and terminates the
operation mode.
e For the power switches and circuit breaker, see Power switches on page A-27
Roche Diagnostics
Operator’s Manual · Version 2.0 D-57
16 Troubleshooting cobas c 311 analyzer
General instrument troubleshooting
Description: The touchscreen does not come on when the power switch of the control unit is
switched on.
Bubbles in syringe
Roche Diagnostics
D-58 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
ISE troubleshooting
ISE troubleshooting
Roche Diagnostics
Operator’s Manual · Version 2.0 D-59
16 Troubleshooting cobas c 311 analyzer
ISE troubleshooting
Roche Diagnostics
D-60 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
ISE troubleshooting
Roche Diagnostics
Operator’s Manual · Version 2.0 D-61
16 Troubleshooting cobas c 311 analyzer
ISE troubleshooting
Roche Diagnostics
D-62 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
ISE troubleshooting
Roche Diagnostics
Operator’s Manual · Version 2.0 D-63
16 Troubleshooting cobas c 311 analyzer
ISE troubleshooting
Roche Diagnostics
D-64 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
ISE troubleshooting
Roche Diagnostics
Operator’s Manual · Version 2.0 D-65
16 Troubleshooting cobas c 311 analyzer
Troubleshooting of the photometric unit
For problems with high test results on the photometric unit, troubleshoot according
to the steps below:
1 Incubation bath temperature is incorrect.
o If the bath temperature does not read 37 0.1C, call technical support.
2 Poor calibration results.
o Check calibrator preparation.
o Check proper calibration programming and calibration results. Repeat
calibration if necessary.
3 Calibrators were not properly prepared.
o Check calibrator preparation and calibration results. Repeat calibration.
4 Evaporation of sample, calibrator or control.
o Repeat analysis with fresh sample, calibrator and/or control.
If calibrators and controls have been loaded on the sample disk for more than
2 hours, evaporation of calibrator may lead to lower results for patient
samples.
5 Reagents are expired.
o Check the expiration date of the reagent.
6 Information is not correct Calibration > Install.
o Verify the calibration points on Calibration > Install and compare the
displayed data with the documentation for a specific test.
o Verify the calibration sample volume in the application parameters.
7 Incorrect sampling or dilution of sample.
o Check correct assembly of sample probe and pipetter parts.
o Check all fittings for leaks.
o Replace O-rings and seals.
8 Insufficient reagent volume.
o Check reagent pipetting system for leaks.
o Replace cobas c pack and repeat analysis.
9 Check that the ultrasonic mixer operates properly.
o Perform maintenance check (6) Cuvette Mixing.
e See To check the intensity of the ultrasonic output on page C-95.
Roche Diagnostics
D-66 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
Troubleshooting of the photometric unit
For problems with low test results on the photometric unit, troubleshoot according to
the steps below:
1 Reagents are expired.
o Prepare new reagents (see instructions for use in cobas link for stability of the
prepared reagent).
2 Reagents were not properly stored.
o Prepare new reagents (see instructions for use in cobas link for proper
storage).
3 Reagents were not properly prepared.
o Prepare new reagents (see instructions for use in cobas link for proper
preparation instructions).
4 Incubation bath temperature is incorrect.
o If the bath temperature does not read 37 0.1C, call technical support.
5 Calibrators were not properly prepared.
o Check calibrator preparation and repeat the calibration with fresh calibrators.
6 Information is not correct on Calibration > Install.
o Check Calibration > Install and compare the displayed data with the
documentation for a specific test.
o Verify the calibration sample volume in the application parameters.
7 Ensure there is sufficient sample in the container. Check instrument
specifications for minimum sample volumes.
8 Check sample for fibrin clotting.
9 Check sample pipetting systems for leaks and air bubbles.
10 Check sample probe for contaminants and obstructions.
11 Repeat analysis with appropriate sample volume.
12 Check sample probe for barbs and proper alignment.
13 Check the cell rinse nozzles are not dripping.
14 Check that the ultrasonic mixer operates properly.
o Perform maintenance check (6) Cuvette Mixing.
e See To check the intensity of the ultrasonic output on page C-95.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-67
16 Troubleshooting cobas c 311 analyzer
Troubleshooting of the photometric unit
For problems with erratic test results on the photometric unit, troubleshoot
according to the steps below:
1 Fibrin clot in 1 sample container or in sample probe (if low values printed for
several samples).
o Check sample for fibrin clot; remove fibrin and repeat analysis.
o Check sample probe for fibrin clot; clean probe (perform maintenance item
(11) Sample Probe Wash) and perform an air purge.
e See also M2: Cleaning pipetter probes and ISE sipper nozzle on page C-51.
o Replace sample probe and sample probe seal.
2 Sample probe does not reach the bottom of the reaction cell when dispensing
sample.
o Perform mechanisms check and verify that the probe reaches the bottom of
the cell.
o Check the spring mechanism to make sure the probe moves up and down
freely.
o The sample probe tip may be damaged. Replace the sample probe.
3 Maintenance was not performed properly or at the recommended frequency on
sample or reagent pipetter or probe.
o Check the Maintenance screen and perform any overdue maintenance
functions.
o If maintenance was recently performed on the sample probe, reagent probe,
rinse nozzles, or any pipetters:
O Was an air purge performed after maintenance?
O Were all parts correctly assembled?
O Have all tubings and seals been checked for air leaks?
O Were sample and reagent probe seals replaced?
4 Insufficient sample volume.
o Repeat analysis with sufficient sample.
5 Contaminated incubation bath.
o Check for particles in the incubation bath. If particles exist, perform the
incubation cleaning procedure
e See: M8: Replacing reaction cells and cleaning incubator bath and drain filter on
page C-68
o Check for foaming, perform incubation bath exchange.
o Check for sufficient Hitergent on the analyzer. Perform incubation water
exchange.
6 Check for sufficient volumes of cell cleaning detergents.
7 Observe cell wash for proper operation:
o Are wash/rinse solutions being properly added and aspirated?
o Are cells being dried?
Roche Diagnostics
D-68 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
Troubleshooting of the photometric unit
8 Check proper probe wash and cell wash programs are in place.
9 Check that the ultrasonic mixer operates properly.
o Perform maintenance check (6) Cuvette Mixing.
e See To check the intensity of the ultrasonic output on page C-95.
For problems with erratic or biased test results on the photometric unit, troubleshoot
according to the steps below:
1 Verify that the deionized water supply is free from contamination.
2 Check calibrators used on all units.
o Calibrators were not properly prepared. Repeat calibration with fresh
calibrator.
o Check calibrator preparation.
3 Information is not correct on Calibration > Install.
o Verify the calibration points on Calibration > Install and compare the
displayed data with the documentation for a specific test.
o Check that the calibration sample volume in the application parameters is
correct.
4 Check sample for fibrin clotting.
5 Ensure there is sufficient sample in the container. Check instrument
specifications for minimum sample volumes.
o Repeat analysis with appropriate sample volume.
6 If you are still experiencing problems, call technical support.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-69
16 Troubleshooting cobas c 311 analyzer
Troubleshooting of the photometric unit
For problems with a single sample or control on the photometric unit, troubleshoot
according to the steps below:
1 Verify samples and controls are placed in the proper positions. If necessary,
correct the sample or control placement and rerun the sample.
2 Verify the control value ranges and lot numbers entered on QC > Install are
correct. If necessary, correct the control value range or lot number on QC >
Install.
3 Verify the sample and/or control volume is sufficient. Verify the selected sample
cup on Workplace > Test Selection.
4 Verify the sample integrity is acceptable (fibrin, lipemia, hemolysis, icterus). If
necessary, collect fresh sample and rerun.
5 Verify the appropriate sample type is selected (serum/plasma, CSF, urine,
supernatant, other) and the specimen was collected appropriately. If necessary,
correct the sample type; check the instructions for use in cobas link for acceptable
specimen types. Check the specimen collection; check the instructions for use in
cobas link for acceptable specimen collection methods.
6 Verify the collection time and date of the sample are correct. If necessary, collect
fresh sample.
7 Verify correct test selections were made on Workplace > Test Selection. If
necessary, correct any selections and rerun the sample.
8 If you are still experiencing problems, call technical support.
For repeated or consistent problems with a single test on the photometric unit,
troubleshoot according to the steps below:
1 Verify that reagents have not expired, or are not contaminated or discolored. If
necessary, insert a new cobas c pack.
2 Verify the correct calibrator code and setpoints are used. If necessary, correct the
calibrator code and setpoints and repeat the calibration.
3 Verify the special wash programming, if applicable.
4 If you are still experiencing problems, call technical support.
Roche Diagnostics
D-70 Operator’s Manual · Version 2.0
cobas c 311 analyzer 16 Troubleshooting
Troubleshooting of the photometric unit
For problems with tests with more than 1 calibration point on the photometric unit,
troubleshoot according to the steps below:
1 Verify calibrators were properly prepared and stored. If necessary, prepare new
calibrators and recalibrate.
2 Verify the assigned calibrators are in the correct positions. If necessary, place
calibrator(s) in correct position(s), recalibrate and rerun samples. Verify that
required diluents are on board.
3 If you are still experiencing problems, call technical support.
For problems with multiple tests on the photometric unit, follow the steps below:
1 Verify there are sufficient volumes of special wash solutions and detergents. If
necessary, replace needed special wash solutions, detergent and rerun samples.
2 Check the reagent probe for barbs, obstructions or leaks.
3 Verify the reagent system is not leaking. Perform an air purge. If the system is
leaking, check connections in the reagent pipetter and check connections in the
reagent syringe.
4 Verify the incubation bath is free of debris and foam. If necessary, perform
incubation bath maintenance.
5 Perform a photometer check. Verify the Photometer Check report is within
acceptable limits (< 14000). If it is not, replace the photometer lamp. Perform a
cell blank. Calibrate all photometric tests.
6 If you are still experiencing problems, call technical support.
Roche Diagnostics
Operator’s Manual · Version 2.0 D-71
16 Troubleshooting cobas c 311 analyzer
Troubleshooting of the photometric unit
For problems with all photometric tests on the photometric unit, troubleshoot
according to the steps below:
1 Verify the sample probe is not blocked or does not have barbs at the tip. If
necessary, clean/replace the probe. Perform an air purge. Check proper dispense.
2 Verify the sample system is not leaking. If necessary, check tubings and
connections. Perform an air purge and check if there are air bubbles in the
syringe.
3 Verify controls/calibrators were properly prepared and stored. If necessary,
prepare new controls/calibrators.
4 Check that the ultrasonic mixer operates properly.
5 Check photometer lamp.
6 If you are still experiencing problems, call technical support.
For problems with biased enzymes on the photometric unit, troubleshoot according
to the steps below:
1 Verify the incubation bath level is above the photometer lens.
2 Verify the incubation bath temperature displayed on the System Overview screen
is 37 0.1C. Verify there is no Incubation Bath Temperature alarm present.
If the temperature is out of range, perform an incubation water exchange, allow
the temperature to stabilize and recheck the bath temperature. If the temperature
is still unacceptable, call technical support.
3 Verify the sample and reagent syringe seals are in good condition. If necessary,
change the syringe seals.
4 Verify the syringe fittings are not loose. If necessary, correct any loose fittings.
5 Verify controls were prepared using volumetric pipettes. If not, prepare new
controls using a volumetric pipette.
6 Perform a full calibration or blank update.
7 Perform a Photometer Check. Verify the Photometer Check report is within
acceptable limit (< 14000). If not, replace the photometer lamp. Perform a cell
blank. Calibrate all photometric tests.
8 If you are still experiencing problems, call technical support.
Roche Diagnostics
D-72 Operator’s Manual · Version 2.0
Glossary E
cobas c 311 analyzer Glossary
absorbance (Abs) - calibration quality criteria
Glossary
This glossary is a compendium in which to look up the bichromatic measurement Photometric technique
meaning of technical terms used in conjunction with the measuring absorbance values at two different
cobas c 311 analyzer. wavelengths (primary and secondary wavelength).
A C
absorbance (Abs) The instrument displays an calculated result See calculated test.
absorbance by ten-thousandfold value.
calculated test A test result calculated from different
assay 1. A specific test. individual tests with a given formula such as ratio A/B.
2. A procedure where the concentration of an analyte in a
sample is determined. calibration The process that establishes the relation
between measured signals (e.g. from a photometer,
assigned value (Roche-defined) Roche-defined photomultiplier, or ion-selective electrodes) and
concentration for calibration material that is encoded on corresponding concentration values of a calibrator.
the calibrator barcode card or transfer sheet.
calibration curve A plot of measured signal values
Automated Download (ADL) A service that provides (determined during calibration) versus known
information necessary for analysis (such as analytical concentration values of calibrators.
parameters) from the data center. ADL is a cobas
TeleService application. See also cobas TeleService. calibration factor The reciprocal value of the slope
(K) of a calibration curve (only applicable to linear
automatic calibration 1. Automatic timeout calibrations). One of the factors used to create a
calibration. A calibration of a parameter performed if a calibration curve (S1Abs, K, A, B, C).
specified time interval expires. The calibration can be
defined for each method separately. calibration frequency A specified interval at which
2. Automatic calibration after a cassette or a cassette of a an assay should be calibrated. Also referred to as
new lot is registered. The calibration can be defined for calibration stability. It is implemented in the application
each method separately. parameters which are downloaded via cobas link.
3. Automatic calibration on QC failure. A calibration
request is generated by the system if a QC value is outside calibration function The type of calibration (for
a predefined range. example, Rodbard function, linear function, or cutoff
function). A mathematical model that describes the
automatic QC A quality-control function that relationship between a signal and a concentration in the
automatically samples QC material, test-specifically calibration curve. See calibration curve.
according to defined time intervals.
calibration masking A function that masks a reagent
automatic rerun The repetition, without operator when no valid calibration is available.
intervention, of tests that have results with data alarms.
calibration monitor A function that prints the
auxiliary reagent A non-test-specific reagent that is measured absorbance of the standard solution and the
needed to wash and rinse reaction cells, sample probes calibration factors, at the time of calibration, for every
and reagent probes. measurement item.
barcode type Typical sample barcode types used in calibration quality criteria Calibration checks
the IVD industry are Code39, NW7 (Codabar), ITF, and applied to the auto-validation of every calibration on the
Code 128. analyzer.
Roche Diagnostics
Operator’s Manual · Version 2.0 E-3
Glossary cobas c 311 analyzer
calibration trace - compensated test
calibration trace A graph used to review the check digit A verification number used in barcodes
measurements of the 50 most recent calibrations for a and software.
specific test.
check sum The result of a mathematical procedure to
calibration type 1. One of the following: Linear, RCM, validate the integrity of a set of data.
RCM2T1, RCM2T2, Spline, Line Graph. Each calibration
type corresponds to one specific type of mathematical circuit breaker See main circuit breaker.
function. See calibration function.
2. Lot calibration or cassette calibration. clot detection 1. A device built into the pipetting
system to detect clots and to avoid false pipetting.
calibration validation Analysis, performed by 2. The procedure of detecting a clot.
software, to check a calibration data set against specific
criteria. Calibration validation results are: successful or cobas The "umbrella brand name" for a broad range
failed. of products and services for use in professional IVD
laboratories and physician's office settings.
calibrator A material of known composition and
properties that can be presented to the analytical cobas c pack The name of the reagent cassette used
instrument for calibration purposes. on cobas c systems and COBAS INTEGRA systems.
COBAS INTEGRA cassettes can also be used on the
calibrator code The identification number of the cobas c 311 analyzer.
standard solution in calibration measurement.
cobas c pack MULTI The name given to an empty,
capacitance An electrical property that provides the but assembled and barcoded kit cassette that can be used
basis for liquid level detection in sample probes. The for Roche or non-Roche reagents. The word MULTI in
probes carry a high-frequency, low-voltage electrical this name stands for multi-purpose. The cobas c pack
charge. Frequency and electrical charge characteristics MULTI has to be assigned to a development channel
are altered and sensed when the probe touches liquid. application before loading it onto the analyzer.
carryover Contamination of a reaction mixture of one cobas link The infrastructure of network connections
assay by reagent(s) or sample material of another that enables cobas TeleService to exchange information
(previous) assay. between the Roche service network and a customer's
laboratory.
cassette The integrated reagent carrier consisting of
one or more reagent bottles and barcoded labels. The cobas link data station A dedicated desktop
term cassette, on cobas systems, is used generically to computer, located in the laboratory, that has been
refer to cobas c packs, cobas c pack MULTI, or COBAS configured to act as a gateway between Roche systems
INTEGRA cassettes. and the Internet. As well as providing a communication
link, the data station also stores data and documentation
cassette calibration Calibration data are specific to a for assay processing and can provide a data archive.
single cobas c pack. Most of the photometric tests need
to be calibrated only if the reagent lot is changed (lot cobas TeleService The set of software applications
calibration). See lot calibration. that use cobas link to exchange service information
between Roche service network and a customer's
cell blank The process of measuring the absorbance of laboratory. cobas TeleService provides remote
all reaction cells, filled with water, at all wavelengths. The monitoring and diagnosis, hotline support, and software
cell blank values are stored on the hard disk. By periodic and documentation updates.
cell blank measurement, reaction cells can be screened
for contamination or damage. Code39 A barcode type for sample tubes that can be
read by the barcode reader.
cell rinse unit A device for cleaning the reaction cells
with detergent and water and for dispensing and coefficient of variation A statistical measure used to
aspirating cell blank water. describe imprecision. Often abbreviated to CV.
cerebrospinal fluid (CSF) The fluid in the ventricles compensated test A test that has the result modified
of the brain, between the arachnoid and pia mater, and by a formula that takes account of known or defined
surrounding the spinal cord. interference factors.
Roche Diagnostics
E-4 Operator’s Manual · Version 2.0
cobas c 311 analyzer Glossary
Complete - dynamic range
control ID The abbreviated name for a control data flag See data alarm.
material, for example PNU and PPU. Control IDs are
used on software screens and windows where limited dead volume See minimum sample volume.
space prevents the use of longer names.
default profile A predefined set of tests that the
control material A material of known composition analyzer automatically applies to a sample unless the
used to assess the performance of an analytical procedure operator specifies a different set of tests.
or part of an analytical procedure. Also called the control
sample. demographics Patient-related data such as name, date
of birth, and gender.
control name The name of a control material, for
example Precinorm Universal. development channel An open application where all
parameters must be set manually. Any non-standard
control SD value The acceptable variation SD value of reagents can be used but must be filled into the
a quality control sample. cobas c pack MULTI. A development channel
application has to be downloaded and installed via
control unit A computer by which an analytical system cobas link.
is controlled. The control unit also serves as the user
interface. deviation of duplicate measurements See
duplicate limit.
correction item A function that corrects the
measurement result of one item by using figures or the dilute waste solution A waste solution resulting after
measurement result of other tests. rinsing with water.
cumulative QC The accumulated data and associated dilution factor A software preset or manually assigned
statistics of individual QC data. dilution ratio that is used by the analyzer to perform a
requested dilution.
cup-on-tube The placement of a smaller secondary
sample container (for example a Hitachi Cup) on top of a disk position A dedicated position on the reagent or
primary sample tube. sample disk.
cycle The instrument time interval during which down time The period of non-operation between an
pipetting or measurement can be carried out. instrument failure and the resumption of operation.
cyclic QC Controls run at fixed intervals. dual value method A mode of expression of the
control chart in real-time quality control. For X-axis and
Y-axis, measure simultaneously the average and the
standard deviation of control of a low value and a high
value, and display them by X and Y coordinates,
respectively.
Roche Diagnostics
Operator’s Manual · Version 2.0 E-5
Glossary cobas c 311 analyzer
electromotive force (EMF) - ISE check
E H
electromotive force (EMF) The physical principle Hitergent 1. A detergent, with antibacterial properties,
that provides the basis for electrolyte measurement. The that can be added to the reaction bath where it acts as a
EMF is measured in millivolts surfactant, reducing the formation of foam.
2. A surfactant diluted for use in some cleaning
emergency stop An instrument alarm level that procedures.
immediately stops all instrument functions. See also E-
stopped. homogeneous immunoassay (HIA) An analytical
technique employing antigens and antibodies. An HIA
endpoint assay An analytical technique taking uses assay protocols similar to clinical chemistry without
measurements after a reaction is completed or has been a bound-free separation (for example, latex assays).
halted. See also rate assay.
Host communication Data exchange with a clinical
error handling A process during which the analyzer laboratory information system (LIS).
attempts to recover from an error condition. If the
analyzer cannot successfully recover from error, an alarm Host computer 1. A computer used for overall
is issued and the instrument is halted. management and control of the computer network.
2. A clinical laboratory computer that stores and
E-stopped A status indicating that the system has processes patient requests and results. A Host is able to
performed an emergency stop (E. Stop). This could be communicate with analytical instruments.
due to hardware failure or because any of the safety
devices have requested an emergency stop. The system Host interface protocol A technical description that
requires either complete power off, or at least defines data transfer between a Host computer and an
initialization, to resume normal operation. analytical system.
extended dynamic range The measuring range for Initialization The operational mode of an analyzer that
an assay at its highest dilution. occurs immediately after power on and during which the
instrument prepares itself for operation.
F instrument alarm A displayed alarm that indicates an
first registration The date and time when a reagent or unusual instrument condition such as an abnormal
sample was successfully recognized by the barcode reader reaction bath temperature or a mechanical malfunction.
for the very first time.
Instrument Manager Typically, PC-based software
functional sensitivity The concentration of an analyte that controls or supervises one or more analytical
at which a predefined level of imprecision is obtained. instruments.
Roche Diagnostics
E-6 Operator’s Manual · Version 2.0
cobas c 311 analyzer Glossary
ISE drain port - Online Help
ISE drain port Port that discharges reaction waste manual rerun A retest function. Although a list of
from the ISE unit and collects it in a container. samples required for a rerun is created by following data
alarms, rerun is not carried out automatically. After
ISE Ref. See reference solution (ISE Ref.). organizing the rerun sample list, the operator directs the
rerun.
K
masking A function that temporarily suspends the
K factor A factor used in conversion of absorbance measurement or calibration of a specific test, depending
values to concentration values or activities. on the condition of the instrument or reagent.
log file A set of data, typically stored in the control unit, Microcup A secondary sample cup made by Hitachi
that traces instrument-related or operator-related with a small dead (residual) volume.
activities such as maintenance.
minimum sample volume The sample volume that
lot calibration A mandatory calibration when a new must be present in a container to ensure faultless sample
lot of reagents is introduced to an analytical instrument. aspiration. Otherwise, a Sample Short alarm will occur.
The minimum sample volume depends on the size and
M shape of the sample container. Also called dead volume.
main circuit breaker Main power switch at the rear of Sample Cleaner 1 1 N NaOH solution for cleaning
the instrument. It controls the power supply of the entire the sample probe.
analyzer including the cooling unit (usually not switched
off).
N
maintenance item A maintenance procedure non-barcode mode A mode of instrument operation
performed by the system or the operator. in which the instrument identifies samples by sequence
numbers corresponding to their positions on the sample
maintenance key A button for position movement, disk.
used for a probe position check.
normal range See expected range.
maintenance pipe A combination of sequential
maintenance items programmed into a fully automated
procedure, which can be performed by the analyzer O
without operator intervention. on board 1. A technical device or function that is part
of the analytical instrument and can be used by the
maintenance procedure A procedure that must be instrument at any time.
performed on a regular basis (for example daily, weekly, 2. The availability of reagents and consumables on an
monthly, or every three months) to secure reliable analytical instrument for use at any time.
operation of the analyzer.
one-way serial processing Sample flow and
manual dilution An off-system, preanalytical step processing along a single, serial process lane that allows
performed by laboratory staff to reduce the analyte no bypass function and no rerun.
concentration in a sample.
Online Help On-screen documentation that a user can
request in a context-sensitive manner and search for any
given term.
Roche Diagnostics
Operator’s Manual · Version 2.0 E-7
Glossary cobas c 311 analyzer
online support - prozone
online support A service that supports the preparation post-analytical The sample management process,
for analysis and maintenance management by typically storage and archiving, after results have been
exchanging information over networks. See also cobas reported.
TeleService.
potentiometric assay An assay in which analytes (for
open request An ordered test that has not yet been example Na, K, or Cl) are measured in millivolts by ion-
performed or completed. Results for a sample may be selective electrodes.
partially available while some tests are not yet completed.
Power Up The system status while it is loading
Operate The operational mode during which the programs, performing self-checks, and so on.
instrument processes samples.
pre-analytical The sample management process
operator ID An alphanumeric ID that a system uses to before the analytical phase. Preanalytical processing
identify a particular operator. There are several levels typically involves actions such as sorting and aliquoting.
including operator, supervisor, and administrator.
precision The closeness of agreement between
order Also called a request. A test selected for a specific independent test results obtained under prescribed
sample or control. conditions.
order date/time A field used to maintain the arrival pre-dilution A dilution step performed before samples
date and time of an order in the laboratory. The are analytically processed on the analyzer.
date/time data may be entered manually or transmitted
by LIS protocols. preventive action A series of actions, suggested by the
system, that the operator should perform before daily
order ID The sample order identification refers to a operation to ensure sufficient inventory during the day
number of sample tubes (one or more specimen types) of (for example, the replenishment of reagents and
a given patient collected for a panel of different tests. consumables).
Typically, the sample order identification is printed on
order sheets. primary tube The original tube containing the sample
that has been drawn from the patient.
P
profile A user-defined set of test requests.
patient ID A set of alphanumeric data that
unmistakably identifies a particular patient. For example, protocol 1. A convention or standard that controls or
a social security number and a sample number. enables the connection, communication, and data
transfer between two computing end points. Protocols
photometric assay An assay in which analytes are can be implemented by hardware, software, or a
measured by a photometer. combination of the two.
2. A set of rules that guides how an activity must be
photometric window A window that enables light to performed.
pass from the light source lamp and into the reaction
bath. prozone The antigen/antibody complex formation is
predictable as long as an excess of reagent (antibody)
piercer A device that pierces the caps of a cobas c pack. exists. However, in patient samples with very high levels
of antigen, the reaction may begin to reverse
pinch valve A valve that pinches the suction tube and (deagglutination) because of the effect of the excess
switches the flow path. antigen. This is called a prozone effect, and without
checking for this phenomenon, abnormally high samples
pipe See maintenance pipe. may give incorrect or even false normal results. There are
two prozone check methods available: Antigen
pipetter A device used for pipetting (aspirating and readdition method and reaction rate method.
dispensing) a fixed amount of sample or reagent from a
sample or reagent container to reaction vessel.
Roche Diagnostics
E-8 Operator’s Manual · Version 2.0
cobas c 311 analyzer Glossary
QC error - rinse bath
qualitative assay An assay that does not allow the reagent probe rinse station An area located between
determination of the concentration of an analyte, but the reagent disks and reaction disk where reagent probes
yields a classification of the sample, such as are rinsed both internally and externally with water.
reactive/nonreactive or positive/negative, with regard to
a certain analyte. realtime QC Real-time quality control. A method in
which two quality control samples of low and high values
quantitative assay An assay that allows the are measured, the quantitative values are judged in real
determination of the quantity (concentration or activity) time, and an alarm is generated if necessary.
of an analyte.
recalibration The repetition of a calibration.
query download A communication process between
instrument PC and LIS by which a predefined data set is reference electrode Electrode through which the
transmitted upon request of the analytical instrument. reference solution flows to provide a reference potential
for ISE measurements (also called reference cartridge).
R
reference solution (ISE Ref.) A KCl solution
random access The ability of an analytical instrument aspirated through the ISE reference cartridge. Also
to process requests from a patient sample in any order. known as ISE Ref. or reference electrode solution. See
reference electrode.
rate assay A determination in which measurements
are taken as the reaction proceeds. The rate of the reflex testing A request for additional testing based on
reaction is proportional to the sample component being customer-defined algorithms or rules and previous test
analyzed. Also known as a kinetic assay. results.
raw data The unprocessed values obtained during the repeat limit A user-definable limit at which a test is
analytical process on an instrument (for example mVolt run again under unchanged conditions.
or absorbance).
reportable range The range of results that can be
reaction bath A bath for keeping reaction cells at a reported for the assay. It stretches from the lower
defined temperature (37C). Also called incubation bath detection limit to the high end of the calibration curve.
or incubator bath.
request See order.
reaction cell A plastic cuvette into which sample and
reagents are pipetted for the chemical or immunological rerun The performance of the same test on a sample
reactions. again under changed conditions.
reaction disk A rotatable disk that holds the semi- rerun – concentrated The performance of the same
disposable reaction cells used for photometric test with a less diluted sample, either by decreasing the
measurement. diluent or by increasing the sample volume.
reagent compartment A temperature-controlled Reset Operational mode during which the analyzer sets
section on an analyzer that holds reagents and diluents. and aligns all mechanical parts to their home positions.
reagent disk The device in the reagent compartment Result Date/Time The instrument fills the result date
into which cobas c packs are placed. and time after the result calculation is finished. It may be
maintained by work area management systems for
reagent mask A function that automatically stops information purpose.
analysis of the current test when a required reagent is
empty or not present on the system. A red bar appears on rinse bath See rinse station.
the test key in the test selection screen.
Roche Diagnostics
Operator’s Manual · Version 2.0 E-9
Glossary cobas c 311 analyzer
rinse nozzle - STAT sample position
rinse nozzle A nozzle that supplies or drains detergent sampling stop An instrument alarm level that
or water used for rinsing reaction cells. indicates a problem with the sampling system. See also
S.Stop.
rinse station A technical device that cleans probes or
disposable tips with deionized water or cleaning secondary tube A sample container of variable size
solutions to avoid contamination and carryover. into which aliquots are transferred
Rodbard function A mathematical function used to sequence number When operating in non-barcode
convert measured signals into concentrations. It uses mode, samples are identified by a sequence number and
four parameters to define the shape and position of their position in the sample disk. This assignment has to
calibration curves. be done on the Workplace > Test Selection screen.
S1Abs The absorbance of standard solution 1. The Sleep mode Also called sleeping. A mechanical and
displayed value is 10000 times greater than the actual electrical status of an analytical instrument during which
measured absorbance. no immediate processing can be initiated by the
operator.
sample blank The absorbance of the sample plus
reagent 1 of a photometric test. The sample blank is SMS Selective Mode Solution, 0.2N HCl in a
subtracted from the actual absorbance reading to obtain cobas c pack for cleaning the reagent probe and in a
the absorbance value relevant for the result calculation. 70 mL bottle for cleaning the sample probe.
sample cleaner SmpCln 1 and SmpCln 2. See Sample Stand By An operational mode of the analyzer during
Cleaner 1 and SMS. which power is on but no sample analysis or
maintenance procedures are being performed.
sample cup A small container that is used for samples
and also for calibrator and control material. Sample cups standard Traceable reference material used to create a
can be inserted into 16-millimeter sample tubes (cup on (master) calibration curve. Also referred to as calibrator.
tube) or can be used without tubes. Compared to a
sample tube, a sample cup allows the use of smaller liquid standard deviation A statistic used as a measure of
volumes and so reduces the dead volume. the dispersion or variation in a distribution of data.
sample ID A set of alphanumeric data that standby reagent QC The measurement of a quality
unmistakably identifies a particular sample. See also control sample from a reagent that is on board the
patient ID. analyzer but not presently in use for routine testing.
sample probe Pipetter probe used to transfer sample Start up An operational mode of an analyzer, following
material from sample containers to reaction cells. power-on, during which the instrument prepares itself
for operation.
sample tube A glass or plastic container for liquid
samples to be used with the system. It may or may not STAT application A special test application (for
have a barcode label, which may be used for positive example, reduced incubation time) for STAT or
sample identification. A sample tube contains sample of emergency samples to achieve faster result reporting.
one specific specimen (sample) type.
STAT sample Emergency sample. Results will be
sample type One of four types of sample that can be available within shortest possible time.
analyzed: serum/plasma, urine, cerebrospinal fluid
(CSF), and supernatant. The sample volume and normal STAT sample position Reserved positions in the
value are settable for each type. sample disk for STAT samples, which will be processed
with priority.
Roche Diagnostics
E-10 Operator’s Manual · Version 2.0
cobas c 311 analyzer Glossary
Supernatant - water supply tube
T
target range The allowed range of recovery for an
analyte in a control material.
Roche Diagnostics
Operator’s Manual · Version 2.0 E-11
Glossary cobas c 311 analyzer
water supply tube - water supply tube
Roche Diagnostics
E-12 Operator’s Manual · Version 2.0
Index F
cobas c 311 analyzer Index
Index
Symbols Application
– configuring, HbA1c, B–219
<Mix (data alarm), ISE unit, D–15 – installing HbA1c, B–219
<Mix (data alarm), photometric unit, D–25 – loading new application parameters, B–215
>Abs (data alarm), photometric unit, D–19 – troubleshooting, D–51
>Cuvet (data alarm), photometric unit, D–22 Application auto masking, B–225
>Index (data alarm), photometric unit, D–13, D–23 Application data, A–33
>Kin (data alarm), photometric unit, D–24 Application parameters
>Lin (data alarm), photometric unit, D–24 – description, B–221
>Proz (data alarm), photometric unit, D–26 – installation process, A–33
>React (data alarm), photometric unit, D–26 Application report name, B–229
>Rept / <Rept (data alarm) Application settings
– ISE unit, D–16 – autmatic rerun, B–229
– photometric unit, D–27 – auto masking, B–225
>Test (data alarm), ISE unit, D–17 – calibration timeouts, B–226
>Test / <Test (data alarm) – expected values, B–232
– ISE unit, D–18 – qualitative fields, B–231
– photometric unit, D–27 – repeat limit, B–230
– report name, B–229
Numerics – technical limit, B–230
Approvals, 3
3rd Results Acceptance, B–240 Approved parts, A–7
Archive
A – saving patient data, B–106
– saving system parameters, B–265
Abbreviations, 9 Areas of the analyzer, overview, A–24
Abnormal conditions, during operation, B–38 Assigning
Absorbance, reaction course, A–58 – barcoded samples to calibrator positions, B–130
Access – calibrator positions, B–175
– cannot access another software screen, D–57 – STAT positions, B–238
– no access due to operator level, B–9 – tests to units, B–248
– random access, E–9 Assignment
Access lamp for sample disk, A–37–A–38 – calibration curve, B–173
Accessory – Key setting window, B–244
– cobas c pack MULTI, B–133 – mandatory tests, B–51
– printer, A–30 – non-barcode mode, A–43
Accuracy/precision of measured results, A–6 – Test Assignment window, B–248
Activate AU, abbreviation, 9
– control, B–208 Auto masking, Utility > Application screen, B–225
– test, B–208 Automatic Printout, B–62, B–79
ADC.E (data alarm) Auxiliary reagents
– ISE unit, D–10 – for ISE applications, B–129
– photometric unit, D–20 – for photometric applications, B–132
Alarm – ISE applications, B–129
– check for, B–43 – overview, A–65
Alarm indication, D–7 – sample probe, A–37
Alarm levels, D–7
Alarm screen, reviewing, B–43
Analytical unit (analyzer), location, A–23
Roche Diagnostics
Operator’s Manual · Version 2.0 F-3
Index cobas c 311 analyzer
B Calibration
– Calibration > Calibrator screen, B–174
Backup – Calibration > Install screen, B–166
– Data files, D–53 – ISE unit, B–157
– overview, B–77 – load calibrator, B–176
– patient data, B–106 – request, B–60
– system parameters, B–265 – review data, B–159
– test results, B–77 – reviewing measurements, B–161
Backup function, A–32, A–34 – start-up calibration, B–158
Barcode – validating, B–63
– read error, B–103 Calibration > Calibrator screen, B–174
– scan range on sample containers, A–79 Calibration > Install screen, B–166
– specification of barcode types, A–78 Calibration and QC Load List
– specification of labels, A–79 – load calibrators and controls, B–61
Barcode reader – printing, B–60
– for cobas c packs, A–47, A–48 Calibration and QC Select, operation, B–58
– for samples, A–38 Calibration causes, B–152
Biased results Calibration error, ISE unit, D–28
– enzyme results, D–72 Calibration failure data
– troubleshooting, D–69 – ISE unit, D–11
Biohazardous materials, A–8 – photometric unit, D–20
Blood Calibration Load List report, B–61
– Whole > see HbA1c Calibration masking, B–155
Bubbles, in syringes, D–58 Calibration methods, B–153
Buttons Calibration timeouts, B–226
– global buttons, B–10 Calibration triggers, B–152
– software basics, B–19 Calibrator
– standard buttons, B–20 – assign position, B–175
– with black triangles and gray buttons, B–19 – for ISE applications, B–130
– load calibrator, B–61
C – measuring, B–62
– reassign calibrator values, B–172
Cal.E (Calib flag), ISE unit, D–28 Calibrator data, A–33
Cal.E (Sample flag) – loading, B–167
– photometric unit, D–20 Calling technical support, D–52
Cal.E (Sample flag), ISE unit, D–11 Carryover
Calc.? (data alarm) – auxiliary reagents, B–132
– ISE unit, D–10 – evasion, B–255
– photometric unit, D–20 – rinse station, A–40, A–46
Calculated test Cartridge
– program, B–250 – ISE measurement, A–61
– programming, B–249 – reference, A–61
Calculated test formula Caution, safety classification, A–5
– deleting, B–252 Cell blank, A–53
– entering, editing, B–250 Cell blank, abnormal (data alarm), D–22
Calculation Cell cleaners
– not possible (ISE unit), D–10 – See cell wash.
Cell covers, A–49
– cleaning, C–62
Cell detergents
– See cell wash.
Cell rinse unit, A–53
Roche Diagnostics
F-4 Operator’s Manual · Version 2.0
cobas c 311 analyzer Index
Roche Diagnostics
Operator’s Manual · Version 2.0 F-5
Index cobas c 311 analyzer
Roche Diagnostics
F-6 Operator’s Manual · Version 2.0
cobas c 311 analyzer Index
E Favorites tab
– F1 Help, B–30
E.Stop, B–41 Filters
e-barcodes, A–32 – detergent aspiration filters, C–76
– application data, A–33 – ISE Ref. aspiration filter, C–73
– calibrator data, A–33 – radiator filter, C–80
– control data, A–33 – water filter, C–110
Edited (data alarm) Finalizing a DVD disk, B–125
– ISE unit, D–12 Flow path
– photometric unit, D–23 – cleaning, C–47
Editing – ISE measuring system, A–61
– control values, B–206 Formatting a disk, B–123
– maintenance pipes, C–11 Fuses and circuit breakers, A–13
– maintenance types, C–19
Edition notice, 2 G
Electrical safety, A–8
Electrical supply, power not available, D–57 Gate, to the reagent loading station, A–44
Electromotive force, ISE analysis, A–62 GetLog file, D–53
e-library, A–31, A–35 Global buttons, B–10
– e-barcodes, A–32
– e-package inserts, A–32, A–34 H
– introduction, A–34
– user interface, A–35 H (data alarm)
Emergency stop, E–6 – ISE unit, D–13
Environmental conditions, A–72 – photometric unit, D–23
e-package inserts, A–32, A–34 Hard disk
Erratic result – control unit components, A–29
– ISE unit, D–60 – saving space on, B–50
– Photometric unit, D–68, D–69 HbA1c
Erroneous operation, D–43 – Calculation, B–220
Evaporation of samples, A–11 – Control data, B–220
Exclamation Point, Calibration Status screen, B–156 – Loading Application, B–217
Expected values Height adjusting lever, A–28, A–29
– changing, B–232 Help
Expected values, changing, B–232 – system, instrument, B–24
Expiration Date format, B–147 – See F1 Help
High result, Photometric unit, D–66
F High value
– ISE internal standard, D–63
F1 Help, B–24 – sodium, D–64
– areas in the help window, B–25 Hitachi cups, specifications, A–77
– buttons, B–25 Hitachi micro cup, A–42
– contents tab, B–27 Hitergent, A–51, B–132
– Favorites tab, B–30 Host interface, specifications, A–77
– index tab, B–27
– navigation, B–26 I
– printing topics, B–31
– search options, B–29 ICVC (initial cassette volume check)
– search recommendations, B–29 – registration of cobas c packs, B–136
– search tab, B–28 INC., abbreviation, 10
– searching for information, B–27 Increased sample volume, D–47
– topic area, B–26 Incubator bath, A–51
False bottom tube, A–42 Incubator bath, see also Reaction bath
– settings, B–246 Index tab, F1 Help, B–27
– specifications, A–77 Individual QC Chart window, B–191
Fatigue due to long hours of operation, A–11 Information, where to find, 7
Roche Diagnostics
Operator’s Manual · Version 2.0 F-7
Index cobas c 311 analyzer
Roche Diagnostics
F-8 Operator’s Manual · Version 2.0
cobas c 311 analyzer Index
Roche Diagnostics
Operator’s Manual · Version 2.0 F-9
Index cobas c 311 analyzer
Power ON QC results
– order for switching on the analyzer, order, B–39 – handling, B–182
– troubleshooting, D–56 Qualitative fields,
Power ON pipe, C–17 photometric application parameters, B–231
Power supply, specifications, A–71 Quality controls
Power switches, A–27 – for ISE applications, B–130
Power Up Pipe function, activating, C–14
Pre-diluted sample R
– requesting a test manually, B–91
Prediluted sample, B–98 R4SD (data alarm), D–40
– requesting a test manually, B–91 Radiator filter, A–64
Prep.E (data alarm), ISE unit, D–32 – cleaning, C–80
Preparing a disk, B–122 Random access, E–9
Preroutine operation, B–48 Random error, realtime QC, D–40
Preventive action Reaction bath, A–51, C–68
– check box, B–47 Reaction cell, A–49
– in daily routine, B–47 – rinse unit, A–53
Print Reaction course, of photometric analysis, A–58
– Automatic Printout, B–62, B–79 Reaction disk, A–49
– customizing report format, B–261 Reaction Monitor
– Data Print, B–79 – calibration measurements, B–161
– load list, B–59 Reaction system, specifications, A–74
– QC load list, B–187 Reag.S (data alarm), ISE unit, D–16
– reagent load/unload list, B–51 Reagent
– Report format, B–264 – barcode label of cobas c packs, B–134
– report name, B–229 – handling, B–135
– Requisition List, B–67 – ISE applications, B–129
Print order, B–261 – loading cobas c packs, A–47
Printer, A–30 – notes for replacement of ISE reagents, B–130
Printing – photometric applications, B–131
– help topic, B–31 – registration of cobas c packs, B–136
Printout – registration of ISE reagents, B–135
– Sample No. (S. No.), B–106 – replacement of ISE reagents, B–52
Priority, sample pipetting, B–121 Reagent area components, A–44
Processing Reagent cassettes
– reruns, B–95 – reagent compartment, A–44
– routine samples, B–66 Reagent compartment, A–44
– STAT samples, B–70 Reagent load/unload list
Programming – daily operation, B–51
– calculated tests, B–249, B–250 – printing, B–51
– compensated tests, B–253 Reagent loading station, A–47
– system parameters, B–236 Reagent management system, A–47
Reagent Overview window, B–143
Q Reagent pipetting system, A–45
Reagent Preparing button, B–51
Q2.5SD (data alarm), D–40 Reagent prime
Q3SD (data alarm), D–39 – for ISE reagents, B–54
QC Reagent probe, A–45
– request, B–60 – programming special washes, B–255
– validating, B–63 Reagent storage, location, A–44
QC error 1 (1-3S), D–39 Reagent syringe, A–46
QC error 2 (1-2.5S), D–40 – bubbles in, D–58
QC Load List report, B–61 Reagent system, specifications, A–75
QC points Reagent troubleshooting, photometric unit, D–70
– adding comments, B–192 ReagEx (data alarm), photometric unit, D–26
– excluding, individual points, B–192
Roche Diagnostics
F-10 Operator’s Manual · Version 2.0
cobas c 311 analyzer Index
Roche Diagnostics
Operator’s Manual · Version 2.0 F-11
Index cobas c 311 analyzer
Roche Diagnostics
F-12 Operator’s Manual · Version 2.0
cobas c 311 analyzer Index
Roche Diagnostics
Operator’s Manual · Version 2.0 F-13
Index cobas c 311 analyzer
Roche Diagnostics
F-14 Operator’s Manual · Version 2.0