DMC HAMBURG - SERVICE HELPDESK

Originator: Joerg Breitenfeld Date: 6-Sep-2013

Applies to: Juno DRF

Symptom : With dual console configuration the movement


is jerky

Cause : 1. address conflict, both consoles sending on


the same address
OR
2. Data loss

Remedy : 1. - power off the Juno DRF cabinet


- disconnect both consoles CAN cables X8 from
the CAN splitter board A21
- power on the Juno DRF cabinet, red window
with CAN error 1500 pops up
- select "work offline" on the console, enter
the code 2573
- press console 1 for the control room and
console 2 in the examination room
- power off the Juno DRF cabinet
- reconnect the CAN cables X8 to the correct
port
- check function

2. - press setup button and check which


console has Joy Version 0, (normal is Joy 1.02)
- power off the Juno DRF cabinet
- disconnect the related console CAN cable X8
from the CAN splitter board A21
- connect the provided CAN terminator plug
(120R Pin1-2) to the end of that CAN Cable
- power on the Juno DRF cabinet, red window
with CAN error 1500 pops up
- select "work offline" on the console, enter
the code 2573
- press console 1 for the control room or
console 2 for the examination room
- power off the Juno DRF cabinet
- remove terminator, reconnect the CAN cables
X8 to the correct port
- check function