OmniScan SX MX2 HydroForm – Setup Overview
Back wall 1
Back wall 1
Back wall 1
Back wall 2
OmniScan SX MX2 HydroForm – TCG Creation cont.
Enter the position of back wall 2 (17mm below) and increase the gain so that back
wall 2 is >100% amplitude.
The amplitude of back wall 2 should increase without affecting the amplitude of back
wall 1 or the interface signal for the I gate.
Back wall 1
Back wall 2
OmniScan SX MX2 HydroForm – Detection Gate Configuration
Set gate A start as close to the interface signal as possible without touching and
ensure sufficient width to cover the first back wall. This may require modification in
analysis mode or when the scanner starts moving.
Set gate B over the second back wall as pictured below.
OmniScan SX MX2 HydroForm – Detection Gate Configuration
Change the peak selection of gate A and gate B to first peak edge mode as below.
OmniScan SX MX2 HydroForm – Thickness Reading Configuration
Select >Gate>Thickness>Source and change the thickness reading source to A/-I/.
How the thickness reading is calculated is determined based on this parameter.
T = Thickness for current A-scan and data point.
OmniScan SX MX2 HydroForm – Thickness Reading Configuration
Enter 2mm for the minimum thickness. This may be adjusted prior to inspection or in
in analysis mode based on the interface signal and surface condition.
Enter the nominal pipe thickness for maximum. (10mm below)
This range will limit the T thickness reading so invalid readings are not displayed or
evaluated.
OmniScan SX MX2 HydroForm – A-S-C Layout
Select the A-S-C layout from the title bar to display the position C-scan.
The thickness C-scan color pixel is determined by the thickness reading for each A-
scan based on the gate logic selected in previous step.
Readings are covered in detail in the analysis section of this training.
Thickness C-scan
OmniScan SX MX2 HydroForm – Inspection Configuration One Line Scan
Select >Scan>Inspection>Scan>Encoder 1.
This will change the inspection from a time based inspection to an encoded
inspection and the clock icon in the OmniScan notification area will change from a
clock to a mechanical wheel.
Thickness C-scan
OmniScan SX MX2 HydroForm – Inspection Configuration One Line Scan
Select >Scan>Encoder>Resolution and enter the resolution of the encoder.
If the encoder resolution is not known it can be verified or calibrated using the
calibration wizard in >Wizard>Calibration>Type:Encoder>Start.
Thickness C-scan
OmniScan SX MX2 HydroForm – Inspection Configuration One Line Scan
Select >Scan>Area and enter the scan start and end. This will define the length of
the C-scan for the one line scan.
The software will not allow a C-scan length to be created that will result in exceeding
the 300 Megabyte file size limit under current configuration.
Thickness C-scan
OmniScan SX MX2 HydroForm – Inspection Configuration One Line Scan
The inspection resolution is the interval at which all A-scans are recorded and is
normally defined in the corrosion procedure or work instruction.
Use 1mm or increase to 2mm if a larger C-scan is needed. This parameter is directly
related to scanner speed and file size.
Thickness C-scan
OmniScan SX MX2 HydroForm – Inspection Configuration One Line Scan
The OmniScan setup for the HydroForm inspection is now complete. Save the Ops
file and continue to the acquisition section.