Professional Documents
Culture Documents
Assessment of Interchangeable Applicability of Two
Assessment of Interchangeable Applicability of Two
Junzhe Wu
Als Dissertation genehmigt von der
Medizinischen Fakultät der Friedrich-Alexander-Universität
Erlangen-Nürnberg
i
Table of contents
Abstract .................................................................................................................... 4
Zusammenfassung ................................................................................................... 7
1 Introduction ...................................................................................................... 10
1.1 Importance of wear measurement of THA .................................................10
1.1.1 Indications of THA ..............................................................................10
1.1.2 Staging and treatment of hip osteoarthritis..........................................11
1.1.3 Prosthetic interface commonly used in total hip arthroplasty ...............12
1.1.4 Importance of different types of polyethylene......................................12
1.1.5 Common reasons for the failure of THA ..............................................13
1.1.6 Definition and features of wear ...........................................................13
1.1.7 Applications of wear measurement .....................................................14
1.2 Wear measurement techniques .................................................................14
1.2.1 Classification of wear measurement techniques .................................15
1.2.2 Common wear measurement techniques ...........................................15
1.2.3 Comparison between two dimensional (2D) techniques and three
dimensional (3D) techniques ...........................................................................18
1.2.4 Specific principles of wear measurement techniques in this experiment
18
1.3 Gaps in clinical routine ..............................................................................20
1.3.1 Applications of preoperative planning software programs ...................20
1.3.2 Causes of wear measurement using plain pelvic radiographs.............22
1.3.3 Influencing factors of artificial joint wear .............................................23
1.3.4 Correlation of pelvic positions and wear measurement .......................25
1.4 Aim, hypotheses, and innovation of this study ...........................................26
2 Material and Methods ...................................................................................... 27
2.1 Measurement setup...................................................................................28
2.2 Phantom model .........................................................................................30
2.3 Measurement protocol...............................................................................33
2.4 Data analysis.............................................................................................36
2.5 Statistics....................................................................................................37
3 Results ............................................................................................................ 38
3.1 3Dwear measurement result ........................................................................38
3.2 2Dwear measurement result ........................................................................42
ii
3.3 Comparison between 3Dwear and 2Dwear measurement results ...................44
3.3.1 Comparison between 3Dwear and 2Dwear measurement results in the
pelvic neutral position ......................................................................................44
3.3.2 Comparison between 3Dwear and 2Dwear measurement results in all
pelvic positions as a whole ..............................................................................45
3.3.3 Comparison between 3Dwear and 2Dwear measurement results in all
pelvic positions individually ..............................................................................48
4 Discussion ....................................................................................................... 50
4.1 Restating the findings of the main results ..................................................50
4.2 Comparison to other studies of wear measurement...................................50
4.2.1 Comparison to other individual 2D techniques by computer-assisted
software programs ...........................................................................................51
4.2.2 Comparison to other individual 3D techniques by RSA .......................51
4.2.3 Comparison to other combined 2D and 3D techniques .......................52
4.2.4 Comparison to other studies involving pelvic orientations ...................53
4.3 Strengths of the experiment ......................................................................54
4.3.1 2Dwear measurement systems can meet clinical requirements.............54
4.3.2 Evaluating accurate wear measurement ranges by setting different
wear groups .....................................................................................................55
4.3.3 Simulating the problem of pelvic misalignment during clinical filming by
setting different pelvic orientations ...................................................................55
4.4 Limitations of the experiment .....................................................................56
4.4.1 Factors affecting the precision and accuracy of wear measurement ...57
4.4.2 Limitations ..........................................................................................57
4.5 Future work ...............................................................................................58
4.6 Conclusion ................................................................................................59
5 Reference list................................................................................................... 60
6 List of abbreviations ......................................................................................... 67
7 List of publications ........................................................................................... 69
8 Appendix ......................................................................................................... 70
9 Acknowledgements.......................................................................................... 71
iii
Abstract
Abstract
Introduction / Background
Osteoarthritis (OA) constitutes the most common reason for total hip arthro-
plasty (THA).
However, the change of patient positions, for example, tilt or rotation of the
pelvis, may influence the accuracy of wear measurement in THA.
4
Abstract
Results
From the perspective between different wear detection methods, the precision
and accuracy were 0.206 mm and 0.159 mm for 2Dwear and 0.043 mm and
0.017 mm for 3Dwear, respectively. With respect to different pelvic positions, for
2Dwear, p values of the four relative positions were all less than 0.05; for 3Dwear,
the p-value between pelvic lateral tilted left (obliquity left) 5-degree position
and the pelvic neutral position was less than 0.05. From the perspective be-
tween different wear groups, for 2Dwear, the medium wear group was not sig-
nificantly different among all wear positions (p = 0.068); for 3Dwear, the low wear
group was not significantly different among all wear positions (p = 0.235).
Discussion / Conclusion
For the two methods used in this experiment, i.e. 3Dwear and 2Dwear, precision
and accuracy using 3Dwear were approximately by factor 5 and 9 better than
for the 2Dwear approach despite the pelvic positions. The low wear group in
3Dwear and the medium wear group in 2Dwear were relatively unaffected by pel-
vic positions. Attention should be paid to the potential malposition of the pelvis
during the image acquisition because this affected the 2Dwear measurement
outcome significantly. If this is taken into account, an in vivo wear measure-
ment will be possible using the 2Dwear technology because 2Dwear is an excel-
lent alternative to 3Dwear.
5
Abstract
Partial of the present dissertation will be turned into an article for the
future publication to scientific peer review journal.
6
Zusammenfassung
Zusammenfassung
Einleitung / Hintergrund
Die Osteoarthrose (OA) ist der häufigste Grund für eine Hüfttotalendoprothese
(THA).
Ergebnisse
Bezogen auf die verschiedenen Verschleißerkennungsmethoden betrug die
Präzision und Genauigkeit 0,206 mm und 0,159 mm für 2Dwear bzw. 0,043 mm
und 0,017 mm für 3Dwear. Bezüglich der verschiedenen Beckenpositionen
waren bei 2Dwear die p-Werte der vier relativen Positionen alle kleiner als 0,05;
bei 3Dwear war der p-Wert zwischen der seitlich nach links geneigten
(Schräglage links) 5-Grad-Position des Beckens und der neutralen
Beckenposition kleiner als 0,05. Im Vergleich zwischen den verschiedenen
Verschleißgruppen unterschied sich bei 2Dwear die mittlere Verschleißgruppe
nicht signifikant zwischen allen Verschleißpositionen (p = 0,068); bei 3D wear
unterschied sich die niedrige Verschleißgruppe nicht signifikant zwischen allen
Verschleißpositionen (p = 0,235).
8
Zusammenfassung
Diskussion / Fazit
Für die beiden in diesem Experiment verwendeten Verschleißmessmethoden,
d.h. 3Dwear und 2Dwear, waren Präzision und Genauigkeit unter Verwendung
von 3Dwear trotz der unterschiedlichen Beckenpositionen um den Faktor 5 und
9 besser als bei der 2Dwear-Methode. Die ermittelten Abriebraten für die
Gruppe mit geringem Verschleiß in 3Dwear und die Gruppe mit mittlerem
Verschleiß in 2Dwear waren unbeeinflusst von den Beckenpositionen. Es sollte
allerdings bei der Bildaufnahme auf eine mögliche Fehllagerung des Beckens
geachtet werden, da diese das Messergebnis von 2Dwear erheblich beeinflusst
hat. Wenn dies berücksichtigt wird, ist eine in-vivo-Verschleißmessung mit der
Technologie von 2Dwear möglich, da 2Dwear eine ausgezeichnete Alternative zu
3Dwear ist.
9
Introduction
1 Introduction
In recent years, there has been an increasing number in total hip arthroplasty
(THA) because the technology of THA is becoming more mature and perfect,
which depends on the development of artificial joint prostheses. The lifespan
of artificial hip joints is still a major problem due to inevitable wear, so there are
several wear detection techniques. With the improvement of the wear proper-
ties of artificial joint materials, the wear rate is now decreasing, which places
higher demands on the accuracy of wear measurement methods. At present,
Roentgen Stereophotogrammetric Analysis (RSA) is a relatively accurate and
reliable method for in vivo wear measurement, presenting the current gold
standard (Uddin et al. 2016). However, the experimental equipment and soft-
ware available in the clinical practice are limited, and the radiographic imaging
process is generally affected by pelvic positions. Based on everyday clinical
practice, the experiment is designed to compare two software programs in
seven different pelvic positions.
10
Introduction
2014 in the United States (Park et al. 2018). Osteoarthritis, rheumatoid arthritis,
femoral head necrosis, developmental dysplasia of the hip, and femoral neck
fracture are the most common indications for THA. Over 80% of all primary hip
surgery is due to osteoarthritis (OA) of the hip (Seidlitz and Kip 2018), which
is the most significant reason for requiring hip joint replacement. Total hip ar-
throplasty (THA) is a crucial technique for OA, which is increasingly perfect
over the past twenty years, increasing the quality of life.
11
Introduction
MoPE interface was previously the most widely applied and most classic type
of total hip prosthesis. For example, a survey study showed that MoPE was
the most commonly reported bearing, with a proportion of 51% between 2005
and 2006 in the United States (Bozic et al. 2009). However, American Joint
Replacement Register (AJRR) reported that the most widely used bearing was
a CoPE combination with 67.3% in 2018 (Surgeons 2019). In 78.4% of all THA
treatments in Germany, a CoPE bearing was used (Grimberg et al. 2019). In
clinical practice, CoC is at risk of breakage (Jeffers and Walter 2012), and MoM
is concerned about the release of metal ions (Clarke et al. 2003). On the other
hand, consider that when a CoC or MoM bearing is used, the visibility of the
head contour in the cup with x-ray or stereo x-rays may be unsatisfactory.
Therefore, this experiment was conducted on MoPE, and then the wear meas-
urements were simulated to calculate the linear wear of polyethylene.
12
Introduction
2012 to 14% in 2018, which was the case in the US (Surgeons 2019). Similarly,
the antioxidant PE ratio in Germany was 17.7% based on the annual report
2019 of Endoprostheses Register Germany (Grimberg et al. 2019).
13
Introduction
In vivo, wear is assessed as penetration of the femoral head into the liner. The
linear penetration measurement includes creep (plastic deformation) and wear
(material removal). Creep does not produce particles initially (Galvin et al.
2010). Still, creep occurs early after the total hip replacement to increase the
head-liner contact region and decrease contact pressure (Penmetsa et al.
2006). One study showed that most of the creep behavior occurred within half
a year after implantation, which was termed the bedding-in phase, and con-
cluded that the penetration behavior was determined by creep within half a
year and attributed to wear after one year (Glyn-Jones et al. 2008).
14
Introduction
dinates, optics, etc., whose tools consist of physics equipment and corre-
sponding software. However, there is no consensus in terms of the accuracy
and precision of these various methods.
2012). Livermore et al. (1990) used a transparent template with concentric cir-
cles, including various radii in increments of 1 mm, and a compass to deter-
mine the center of the femoral head, which was the most regularly applied in
radiographic methods then, despite a manual method as well (Grillini and Af-
fatato 2013). The approach of Dorr and Wan (1996) drew an acetabular face
reference line to calculate linear wear, which was another significant manual
wear technique (Barrack et al. 2001). There were several manual assessment
methods (Charnley and Halley 1975, Pollock et al. 2001, Kang et al. 2003).
16
Introduction
17
Introduction
For 2Dwear in the experiment, mediCAD is the software program for wear meas-
urement, which uses a digital edge-detection algorithm to determine the center
①
Homepage mediCAD HECTEC GmbH, mediCADclassic: https://www.medicad.eu/
[07.10.2020, 17:43]
②
Homepage AGFA, IMPAX: https://www.agfa.com/corporate/ [07.10.2020, 17:43]
18
Introduction
of the acetabular cup and the femoral head and calculates the relative linear
wear value. The concrete procedures in the mediCAD wear analysis are as
follows: (i) scaling images, (ii) drawing reference lines, (iii) defining the head
center by 3 points, and (iv) calculation of wear (Figure 2).
19
Introduction
Figure 2: Analytical environment of wear measurement using mediCAD® Classic. The tur-
quoise line is a reference line. The black and blue circles are respectively the centers of the
femoral head and the acetabular cup. The white sphere near the reference line is the cali-
bration sphere.
20
Introduction
https://www.endomap.de/
https://www.agfa.com/corporate/
https://www.medicad.eu/
https://www.traumacad.com/
https://www.lexi.co.jp/en/
21
Introduction
The 2Dwear linear analysis can detect 90.1% of the total linear wear detected
by the 3D analysis and the repeatability of 3D analysis is poor (Martell et al.
2003a). On the other hand, wear along the x-axis and y-axis directions (medial
and proximal) is widely regarded as where the majority of wear occurs
22
Introduction
(McCalden et al. 2005). A study also indicated that femoral head penetration
of 95% of patients in the research occurred in the AP plane (Sychterz et al.
1999). In two studies, one used cemented cups (Derbyshire and Barkatali
2017) and the other used cementless cups (Stilling et al. 2010). Their 2Dwear
measurements with AP were both compared with RSA, and the accuracy was
similar. Another study showed the line wear in the AP plane could afford a
predictable estimation of the total wear volume (Ilchmann et al. 2012). Further-
more, compared to direct measurements from retrieved components, AP im-
ages were also applicable to detect 2-dimensional in vivo wear (Parran et al.
2018).
Not many researchers have addressed the problem of comparing 2D wear de-
tection with 3Dwear detection. And little is known about the interchangeability
between the 3D RSA method and computerized software methods using plain
radiographs. For example, a study compared three methods (the simple and
noncomputerized Scheier-Sandel, Charnley-Duo methods, and the computer-
ized EBRA method) with RSA and found that the EBRA method had the best
accuracy (Ilchmann et al. 1995). In addition, a study comparing two RSA tech-
niques (3D) and three standard radiographic techniques (2D) for measurement
of wear demonstrated that the 2D techniques were less accurate (Schewelov
et al. 2004). Another comparative study with 5 years of follow-up showed
greater wear outcomes than RSA at each time point measured with Martell Hip
Analysis Suite software, but the difference between the two methods was re-
duced after 2 years (Bragdon et al. 2006a). The interchangeability between
the 3Dwear method and 2Dwear detection is still unresolved.
In most instances, the two factors, young and active patients (Tikhilov et al.
2018), are liable to increase wear. For head size, the femoral head of larger
size will lead to more wear theoretically. However, a 10 to 14 year follow-up
23
Introduction
showed no relation between femoral head size and linear wear rate, and larger
femoral heads were correlative to more volumetric wear (Lachiewicz et al.
2016). For fixation mode, according to knowledge, cemented fixation mode
generates less wear. Nevertheless, when HXLPE was employed, the PE wear
rate was similar between cemented sockets and uncemented liners (Morita et
al. 2017). Prosthesis type, liner thickness, head type are all design factors. A
study showed that design factors might be pronounced regarding the risk of
revision (Johanson et al. 2017). For the femoral offset, the restored offset can
cause less wear. A clinical study involving implants on both sides that were
similar except for implant offset showed that the more closely restored femoral
component, which was corresponding to preoperative hip biomechanics, de-
creased polyethylene wear (Sakalkale et al. 2001). Gender is a vague influ-
encing factor. A study identified a gender-dependent discrepancy in wear ac-
cording to head size, with 32 mm heads being linked with increased wear in
females (Stambough et al. 2016). A randomized controlled trial demonstrated
that there was no significant difference in the conventional polyethylene group,
and no significant difference in the cross-linked polyethylene group, regarding
gender factor (McCalden et al. 2009). Weight is correlated to artificial joint wear,
and obesity seems to be a risk influencing factor. A multiple factor analysis
indicated that weight was one of the four most dominant variables (Floerke-
meier et al. 2017).
There are some other influencing factors, for example, pelvic position and ac-
etabular cup position with the parameters: inclination and anteversion. A ret-
rospective study compared the relationship between the pelvic tilt angle, cup
anteversion, cup inclination, and polyethylene wear in the conventional poly-
ethylene liners group and the highly cross-linked polyethylene liners group in
the supine and standing positions. This study concluded that the pelvic tilt an-
gle correlated with the annual rate of polyethylene linear wear in the conven-
tional polyethylene liners group, but the mechanism was unknown (Tezuka et
al. 2015). Recently, a ten to sixteen-year follow-up indicated that the cup ab-
duction angle and anteversion were correlative to the linear wear rate in THA
(Cheung et al. 2019).
24
Introduction
25
Introduction
The hypothesis was that the 3Dwear detection method would show superior pre-
cision and accuracy compared to the 2Dwear detection method (Hypothesis 1).
There were three innovation points in this study. The first point was that com-
paring the common clinically used method (preoperative planning software)
with the gold standard method (RSA). On the other hand, in terms of experi-
mental design, the pelvic orientation factor was concerned, which was more in
line with the actual clinical practice. The third point was that different wear
value ranges were arranged in the experimental protocol.
26
Material and Methods
27
Material and Methods
Figure 3: Experimental flowchart (In the narrative that follows, 3D wear is used to represent
the 3D technique of RSA in this experiment, and 2Dwear is used to represent the 2D tech-
nique in this experiment.)
This setup was able to capture RSA as well as conventional x-ray images par-
allelly (Figure 4C). For each position of simulated wear, radiographs of each
28
Material and Methods
group would be obtained in the same phantom setting when the phantom was
not moved from the radiology table. The radiological images of RSA and plain
AP were obtained sequentially by shifting the position of the radiology table in
the horizontal direction and adjusting the height of the ceiling-fixed roentgen
tube.
A) B)
C)
Figure 4: (A) uniplanar RSA measurement setup (patients in supine position) using a
ceiling-fixed and mobile roentgen tube (i.e. RSAleft) (Mobilett Plus, Siemens, Berlin,
Germany). (B) classical standard x-ray setup for AP x-ray in the supine position using
one ceiling-fixed roentgen tube (Multix RD 82477-01 Vertix ACS, Siemens, Berlin, Ger-
many). (C) The conversion process of two systems: When x-raya.p. moves to the loca-
tion of RSAright, system B is converted to system A.
29
Material and Methods
30
Material and Methods
Figure 5: A phantom model with foam sawbones, total hip prostheses, and dial microme-
ters. The phantom model consists of an uncemented, 58-mm titan cup (Peter Brehm, Wei-
sendorf, Germany) and an uncemented stem (Peter Brehm GmbH, Weisendorf, Germany)
with a 32-mm cobalt-chromium head (Peter Brehm GmbH, Weisendorf, Germany) im-
planted into a foam hip sawbone. The pelvis is in a neutral position, with the anterior superior
iliac spine and symphysis pubis in the same vertical plane. The femoral stem is attached to
a group of three dial micrometers with a resolution of 0.01 mm (Mitutoyo Deutschland
GmbH, Neuss, Germany).
Figure 6: A 58-mm titan cup (Peter Brehm, Weisendorf, Germany), and an uncemented
stem (VECTOR-TITAN, Peter Brehm, Weisendorf, Germany), and a 32-mm cobalt-chro-
mium head (Peter Brehm, Weisendorf, Germany).
31
Material and Methods
Figure 7: Cup position with the inclination 44.5 degrees and the anteversion 14.4 degrees
32
Material and Methods
A) B)
Figure 9: Global coordinate system with three planes (mediolateral (m.l.), craniocaudal (c.c.),
anteroposterior (a.p.)). (A) RSA: Origin = “Center” of migrating model at reference time point
T0, Direction = Aligned with calibration box. (B) The femoral head moved in superior, medial,
and posterior directions during the experiment.
Seven experiments were performed to simulate the change of the pelvic posi-
tions of patients during the shoot of x-ray separately, which corresponded to
seven different pelvic positions, including neutral, tilted ±5 deg, rotated ±5 deg,
lateral tilted (obliquity) ±5 deg (Figure 12). In addition to adjusting the direc-
tions of the pelvis in the phantom model, the other procedures were identical.
33
Material and Methods
Each experiment was repeated five times as well. Below was an overview of
the experimental process (Figure 13).
Figure 10: Wear protocol illustrated by the Global coordinate system. Blue line represents
distance moved on the x-axis in the wear protocol. Green line represents distance moved
on the y-axis in the wear protocol. Red line represents distance moved on the z-axis in the
wear protocol. Yellow line represents the total distance moved by the x, y, and z axes in the
wear protocol, expressed as a vector.
34
Material and Methods
35
Material and Methods
Wear analysis using EGS-RSA (3Dwear) utilized algorithms to fit the head and
cup models to the regions of interest marked on the peripheries of the head
and cup in the stereo radiographs. The process of 3Dwear analysis included to
generate a new RSA scene, to perform image calibration, to detect contours
in the images, to detect contours in the images, to achieve the pose estimation,
and to calculate wear between several scenes (Figure 1).
36
Material and Methods
to draw reference lines, to define the head center by 3 points, to define cup
center by 3 points, and calculation of wear (Figure 2).
In the last procedure, the wear measurements of two methods between 3Dwear
and 2Dwear were recorded, respectively.
2.5 Statistics
The statistics were processed by SPSS (Version 23, SPSS Inc., Illinois, USA).
According to different statistical purposes, the statistical methods included:
paired t-test, one Way ANOVA, Wilcoxon matched-pair signed-rank test, and
Kruskal-Wallis test. The paired t-test was applied between each paired wear
group to assess the interchangeability of the methods above. Paired t-test or
Wilcoxon matched-pair signed-rank test was applied between each paired
wear group or position, according to whether the differences between the two
groups or positions matched the normal distribution. One Way ANOVA or
Kruskal-Wallis test was applied among each wear group from all wear posi-
tions, according to whether the differences matched the normal distribution.
Paired t-test and Wilcoxon matched-pair signed-rank test were used to answer
the hypotheses: 1) The 3Dwear detection method would show superior precision
and accuracy compared to the 2Dwear detection method; 2) 3Dwear detection
methods were not sensitive to pelvic misalignment. One Way ANOVA and
Kruskal-Wallis test were used to answer the hypothesis 2).
For precision and accuracy, precision was expressed as the standard devia-
tion (SD) of the repeated measures, and accuracy was expressed as the bias
of measured values of wear measurement, whereas the bias was given a def-
inition with the average difference between the measured and true values.
Box plots were applied to observe possible outliers and extreme values within
experiment analysis. Special scatter plots, with limits of agreements (LOA),
were calculated to assess the interchangeable applicability of 2Dwear and
3Dwear detection methods (Bland and Altman 1986). (LOA) = mean difference
± 1.96 × SD
37
Results
3 Results
This study aimed to assess the interchangeable applicability of wear detection
between 2Dwear and 3Dwear methods and the effect of pelvic malpositioning on
the results of wear measurement. Using the methods described above, differ-
ent wear measurements were obtained.
The results were organized into three sections: 3Dwear measurement result,
2Dwear measurement result, and comparison between 3Dwear and 2Dwear meas-
urement results. Different aspects of each section would be illustrated.
For the X axis (medio-lateral axis), the p values of two relative positions were
both less than 0.05 (pelvic tilted back 5-degree position vs. pelvic neutral po-
sition, pelvic lateral tilted left 5-degree position vs. pelvic neutral position) (Fig-
ure 14).
For the Y axis (superior-inferior axis), the p values of three relative positions
were all less than 0.05 (pelvic rotation right 5-degree position vs. pelvic neutral
position, pelvic lateral tilted left 5-degree position vs. pelvic neutral position,
pelvic lateral tilted right 5-degree position vs. pelvic neutral position) (Figure
15).
For the Z axis (posterior-anterior axis), the p values of two relative positions
were both less than 0.05 (pelvic lateral tilted left 5-degree position vs. pelvic
neutral position, pelvic lateral tilted right 5-degree position vs. pelvic neutral
position) (Figure 16).
For 3Dwear as a whole, the p-value between pelvic lateral tilted left 5-degree
position and the pelvic neutral position was less than 0.05 (Figure 17). To
38
Results
indicate each wear group in each pelvic position further, the differences among
the low wear groups were significantly less than the other two groups (Figure
18). In further statistical analysis, the low wear group was not significantly dif-
ferent among all wear positions (p = 0.235). Nevertheless, medium and high
wear groups were significantly different (both p < 0.05).
39
Results
Figure 14: Box plot of 3Dwear for the X axis (The x-axis shows different wear positions, and
the y-axis shows the difference between measurement value and true value. Stars (*): p <
0.05).
Figure 15: Box plot of 3Dwear for the Y axis (The x-axis shows different wear positions, and
the y-axis shows the difference between measurement value and true value. Stars (*): p <
0.05).
40
Results
Figure 16: Box plot of 3Dwear for the Z axis (The x-axis shows different wear positions, and
the y-axis shows the difference between measurement value and true value. Stars (*): p <
0.05).
Figure 17: Box plot of 3Dwear (The x-axis shows different wear positions, and the y-axis
shows the difference between measurement value and true value. Stars (*): p < 0.05).
41
Results
Figure 18: Box plot of 3Dwear among each wear group (The x-axis shows different wear
positions, and the y-axis shows measurement value minus true value. The green, orange,
blue boxes represent low, medium, and high wear groups, respectively).
To indicate each wear group in each pelvic position, the differences among the
medium groups were significantly less than the other two groups (Figure 20).
The medium wear group was not significantly different among all wear posi-
tions (p = 0.068). Nevertheless, low and high wear groups were significantly
different. (both p < 0.05)
42
Results
Figure 19: Box plot of 2Dwear (The x-axis shows different wear positions, and the y-axis
shows the difference between measurement value and true value. Stars (*): p < 0.05).
Figure 20: Box plot of 2Dwear among each wear group (The x-axis shows different wear
positions, and the y-axis shows measurement value minus true value. The green, orange,
blue boxes represent low, medium, and high wear groups, respectively).
43
Results
In this section, the wear measurement results were presented with three dif-
ferent aspects: the data of the pelvic neutral position, the data of all pelvic
positions as a whole, the data of all pelvic positions individually.
P-value in the medium wear group was 0.239, which showed that the medium
wear group was not significantly different between 2Dwear and 3Dwear. Never-
theless, low and high wear groups were significantly different (Both p = 0.00).
44
Results
As observed, conformity between 3Dwear and 2Dwear was visible in some cases,
especially in medium wear group (Figure 21). There was an outlier value from
the high wear group, which might happen due to the loosening of the experi-
mental device during the experiment process.
The box plot and statistical analysis were adopted to assess the interchange-
ability of the two methods. The differences between measurement value and
45
Results
true value were highlighted (Figure 22). In statistical analysis, the paired t-test
or Wilcoxon matched-pair signed-rank test was applied between each paired
wear group from seven pelvic positions according to whether the difference
between the two methods matched the normal distribution. P values in all three
wear groups were all less than 0.05, which showed that all three wear groups
were significantly different between 2Dwear and 3Dwear.
To further evaluate precision and accuracy overall, all the data from seven
pelvic positions was regarded as a whole. There was a significant difference
in precision and accuracy between the two methods (Table 7, Table 8). Nev-
ertheless, with regard to the accuracy, the medium wear group in 2Dwear was
relatively accurate compared to 3Dwear.
46
Results
Figure 21: Bland-Altman plot between 2Dwear and 3Dwear (The x-axis shows the average of
the measurements of two methods, and the y-axis shows the difference between the meas-
urements of two methods. The dashed black lines denote the 95% limits of agreement, the
blue line represents the bias from 0, and the dashed red line, at y = 0, represents the line
of perfect average agreement. The green, orange, and blue dots represent low, medium,
and high wear groups, respectively).
Figure 22: Box plot between 2Dwear and 3Dwear (The x-axis shows different wear groups,
and the y-axis shows measurement value minus true value. The blue and green boxes rep-
resent 2Dwear and 3Dwear, respectively. Stars (*): p < 0.05).
47
Results
Table 7 Precision of 2Dwear and 3Dwear from seven pelvic positions. (Unit: mm)
Table 8 Accuracy of 2Dwear and 3Dwear from seven pelvic positions. (Unit: mm)
48
Results
Figure 23: Box plot between 2Dwear and 3Dwear (The x-axis shows different wear positions,
and the y-axis shows measurement value minus true value. The blue and green boxes rep-
resent 2Dwear and 3Dwear, respectively. Stars (*): p < 0.05).
49
Discussion
4 Discussion
THA is a satisfactory surgery, which has dramatically improved the quality of
life of patients. However, wear remains an inevitable factor and a major unre-
solved question, although there are many available wear measurement meth-
ods. RSA is not a standard wear measurement tool because it requires special
measurement equipment (2 x-ray sources, calibration box) and leads to addi-
tional radiation exposure for patients. This limits its further clinical application.
In clinical practice, standard a. p. radiographs of the pelvis are often used for
follow-up, which can be widely applied in any follow-up time points instead of
RSA. However, standard a. p. radiographs are not always consistently stand-
ard because the pelvic positions of patients are not acceptable invariably. The
study also focused on the impact of pelvic orientation on wear measurements.
50
Discussion
In the study, the precision and accuracy of 2Dwear (mediCAD) in wear meas-
urement were presented. It was more convenient in the process of obtaining
images and analyzing data for 2Dwear. Although there was a gap between
2Dwear and 3Dwear in terms of precision and accuracy (Table 7, Table 8), 2Dwear
might be an alternative method to assess wear detection in some respects,
which was consistent with other 2D techniques (Crockarell Jr and Snearly
2012, Stilling et al. 2012, Langlois et al. 2015, Derbyshire and Barkatali 2017).
For the neutral pelvic position, precision in 2Dwear was 0.111 mm, and accuracy
in 2Dwear was 0.112 mm. This solution method could be applied without diffi-
culty to clinical situations. In a report on the wear rate (Teeter et al. 2017), the
average wear rate in crosslinked polyethylene was 0.04 mm/year and the av-
erage wear rate in conventional polyethylene was 0.08 mm/year, which were
within the range of experimental wear protocol.
51
Discussion
based RSA is a conventional and gold standard RSA method in need of the
insertion of tantalum markers. To overcome the inconveniences of the Marker-
Based RSA, Model-Based RSA for CAD and reverse-engineered shapes,
which is generated by a laser or optical scanning design model of the implant,
is developed. To further facilitate the CAD model-based RSA and RE model
RSA method, EGS-RSA is introduced by utilizing geometrical bodies to calcu-
late migration by rigid body kinematics.
EGS-RSA, which was an approach used in this study, was regarded to have
interchangeable applicability compared to the gold standard Marker-Based
RSA method in Author's laboratory (Jacobsen et al. 2018). On the other hand,
for the measurements in the frontal plane, the precision and accuracy were
similar between EGS-RSA and RE Model-Based RSA (Stilling et al. 2012).
Another research comparing four RSA methods in wear measurements
demonstrated that the marker-less RSA method was feasible in terms of cost,
time, and security (Börlin et al. 2006). On the other hand, similar to previous
research (Kärrholm et al. 2006, Stilling et al. 2012), the RSA wear measure-
ment along the anterior-posterior axis (z) had the worst precision. One reason
was that the z-axis from the sagittal plane consisted of the out-of-plane motion,
which was difficult to measure precisely. Another reason was that the femoral
head might touch the acetabular cup in the z-axis direction for high wear
groups.
For neutral pelvic position, precision in RSA was 0.046 mm, and accuracy in
RSA was 0.024 mm, which were both close to other RSA measurement stud-
ies (Ryd et al. 2000, Pineau et al. 2010, Li et al. 2014). Therefore, in experi-
mental phantom studies, EGS-RSA was sufficient to represent 3D RSA tech-
niques, avoiding adopting Marker-Based RSA or other Model-Based RSA.
52
Discussion
However, PolyWare was exclusively devised for wear measurement and not a
preoperative planning software program. Some similar situations occurred in
other combined 2D and 3D technique studies (Sychterz et al. 1999, Hui et al.
2003, Schewelov et al. 2004). So this was the first time that the research had
presented the problem of comparing preoperative planning software programs
with RSA in wear measurement to improve clinical convenience further.
53
Discussion
54
Discussion
2Dwear. In this experiment of using 2Dwear, the precision and accuracy of wear
measurements at seven different pelvic positions were relatively high (P:
0.111-0.400 mm, A: 0.112-0.196 mm). On the other hand, the actual clinical
wear condition is within the setting range of the 2Dwear measurement systems.
In summary, clinical requirements can be met for the simplicity, universality,
practicability, and low cost of 2Dwear measurement systems.
The strength of setting different wear groups was to evaluate accurate wear
measurement ranges for two wear detection methods. By calculating the pre-
cision and accuracy of low wear, medium wear, and high wear groups individ-
ually, this study could provide insight into a clear comparison of the correlation
about wear values in different ranges between the two methods above.
In the experiment, there was more significant conformity between the two wear
detection methods in the medium wear group than the other groups, namely,
the medium wear group in 2Dwear was relatively accurate compared to 3Dwear.
In clinical practice, considering the wear rate of the material and referring to
different wear ranges, a more cost-effective measurement method can be de-
termined. The range of medium wear group was from 0.1 mm to 0.5 mm, which
meant that 2Dwear was available in initial wear measurements.
55
Discussion
56
Discussion
4.4.2 Limitations
The basis of these two software data analyses is the visibility of the femoral
head contour in the cup, which limits the application of various materials in this
experiment. One of the limitations of this study was that only one material and
one size of cup and head were evaluated in this research.
In the wear protocol, in spite of there being three wear groups, only some pos-
sible clinical situations were simulated, and no simulations were performed on
clinically randomized conditions, which should be explored in future work.
57
Discussion
The experiment was repeated five times, which was evaluated in terms of intra-
observer variability except for interobserver variability. However, this has al-
ready been validated to be excellent by laboratory colleagues.
Seven different pelvic positions, including neutral, tilted ±5 deg, rotated ±5 deg,
lateral tilted ±5 deg, were simulated in our experiment. However, in clinical
practice, the pelvic position is more complicated, which is more a combination
of multiple directions rather than a single direction.
The direction of the wear vector was not addressed in this research, which was
inaccurate for the extent of wear. Wear patterns were multidirectional, most of
which occurred in the coronal plane. A study indicated that a single direction
of wear might underestimate the overall amount of true wear (Yamaguchi et al.
1997). Wear directions were not calculated because the position and direction
of the coordinate system were different in the two software settings, which
needed to be readjusted and reanalyzed. However, the wear measurement of
x, y, and z axes might provide an estimate of the corresponding direction of
wear.
On the other hand, the research will shift from phantom experiments to patients,
which requires more funding and more complicated procedures. If a 3D wear
58
Discussion
4.6 Conclusion
In summary, the accuracy and precision of two computerized wear measure-
ment programs have been analyzed according to different wear groups and
different pelvic orientations.
• For two methods, precision and accuracy are better in RSA, but 2Dwear
is an alternative to 3Dwear, especially in medium wear group, despite
being less accurate and precise.
• Most results of 3Dwear are not susceptible to lower extremity positions.
• The low and high wear groups in 2Dwear and the medium and high wear
groups in 3Dwear are susceptible to lower extremity positions.
59
Reference list
5 Reference list
Abdel MP, von Roth P, Jennings MT, Hanssen AD, Pagnano MW (2016) What Safe
Zone? The Vast Majority of Dislocated THAs Are Within the Lewinnek Safe
Zone for Acetabular Component Position. Clin Orthop Relat Res. 474:386-91.
Affatato S, De Mattia JS, Bracco P, Pavoni E, Taddei P (2016) Wear performance of
neat and vitamin E blended highly cross-linked PE under severe conditions:
The combined effect of accelerated ageing and third body particles during
wear test. J Mech Behav Biomed Mater. 64:240-52.
Affatato S, Spinelli M, Zavalloni M, Traina F, Carmignato S, Toni A (2010) Ceramic‐
On‐Metal for Total Hip Replacement: Mixing and Matching Can Lead to High
Wear. Artif Organs. 34:319-323.
Alexander-Malahias M, De Martino I, Gu A, Baral E, Wright TM, Sculco PK (2019)
Complete wear-through of a metal-backed acetabular cup in an ambulatory
patient. Arthroplast Today. 5:394-400.
Atesok K, Galos D, Jazrawi LM, Egol KA (2015) Preoperative planning in orthopaedic
surgery: current practice and evolving applications. NYU HJD. 73:257-257.
Baldursson H, Egund N, Hansson LI, Selvik G (1979) Instability and wear of total hip
prostheses determined with roentgen stereophotogrammetry. Arch Orthop
Trauma Surg. 95:257-263.
Barrack RL, Lavernia C, Szuszczewicz ES, Sawhney J (2001) Radiographic wear
measurements in a cementless metal-backed modular cobalt-chromium
acetabular component. J Arthroplasty. 16:820-8.
Bayraktar V, Weber M, von Kunow F, Zeman F, Craiovan B, Renkawitz T, Grifka J,
Woerner M (2017) Accuracy of measuring acetabular cup position after total
hip arthroplasty: comparison between a radiographic planning software and
three-dimensional computed tomography. Int Orthop. 41:731-738.
Bellare A, Spector M (2005) The polyethylene history. In: Total Knee Arthroplasty, J.
Bellemans, M. D. Ries, J. M. K. Victor (ed.) Springer, Heidelberg, 45-50.
Bills P, Blunt L, Jiang X (2007) Development of a technique for accurately determining
clinical wear in explanted total hip replacements. Wear. 263:1133-1137.
Bland JM, Altman DG (1986) Statistical methods for assessing agreement between
two methods of clinical measurement. Lancet. 327:307-310.
Börlin N, Rohrl SM, Bragdon CR (2006) RSA wear measurements with or without
markers in total hip arthroplasty. J Biomech. 39:1641-50.
Bozic KJ, Kurtz S, Lau E, Ong K, Chiu V, Vail TP, Rubash HE, Berry DJ (2009) The
epidemiology of bearing surface usage in total hip arthroplasty in the United
States. JBJS. 91:1614-1620.
Bragdon CR, Martell JM, Greene ME, Estok DM, 2nd, Thanner J, Karrholm J, Harris
WH, Malchau H (2006a) Comparison of femoral head penetration using RSA
and the Martell method. Clin Orthop Relat Res. 448:52-7.
Bragdon CR, Thanner J, Greene ME, Malchau H, Digas G, Harris WH, Karrholm J
(2006b) Standing versus supine radiographs in RSA evaluation of femoral
head penetration. Clin Orthop Relat Res. 448:46-51.
Callary SA, Solomon LB, Holubowycz OT, Campbell DG, Munn Z, Howie DW (2015)
Wear of highly crosslinked polyethylene acetabular components. Acta Orthop.
86:159-68.
Charnley J, Halley DK (1975) Rate of wear in total hip replacement. Clin Orthop Relat
Res. 112:170-9.
Cheung A, Yan CH, Fu H, Cheung MH, Chan PK, Chiu KY (2019) Ten-to Sixteen-
Year Follow-Up of Highly Cross-Linked Polyethylene in Total Hip Arthroplasty:
What Factors Affect Wear? J Arthroplasty. 34:2016-2021.
Clarke MT, Lee PTH, Arora A, Villar RN (2003) Levels of metal ions after small-and
large-diameter metal-on-metal hip arthroplasty. J Bone Joint Surg Br. 85:913-
917.
60
Reference list
Cooper RA, McAllister CM, Borden LS, Bauer TW (1992) Polyethylene debris-induced
osteolysis and loosening in uncemented total hip arthroplasty: a cause of late
failure. J Arthroplasty. 7:285-290.
Crockarell Jr JR, Snearly CM (2012) Accuracy and precision of two computer-assisted
methods of radiographic wear measurement in total hip arthroplasty. J
Arthroplasty. 27:37-40.
Davila JA, Kransdorf MJ, Duffy GP (2006) Surgical planning of total hip arthroplasty:
accuracy of computer-assisted EndoMap software in predicting component
size. Skeletal Radiol. 35:390-3.
Derbyshire B (2018) Correction of radiographic measurements of acetabular cup wear
for variations in pelvis orientation. Proc Inst Mech Eng H. 232:299-309.
Derbyshire B, Barkatali B (2017) Validation of a new 2-D technique for radiographic
wear measurement of cemented, highly cross-linked polyethylene acetabular
cups. Med Eng Phys. 47:159-166.
Deshmukh AJ, Rathod PA, Rodgers WH, Rodriguez JA (2012) Early failure of a
ceramic-on-metal total hip arthroplasty: a case report. JBJS Case Connect.
2:e25.
Devane PA, Bourne RB, Rorabeck CH, MacDonald S, Robinson EJ (1995)
Measurement of polyethylene wear in metal-backed acetabular cups. II.
Clinical application. Clin Orthop Relat Res. 319:317-26.
Devane PA, Horne JG (1999) Assessment of polyethylene wear in total hip
replacement. Clin Orthop Relat Res. 369:59-72.
Devane PA, Horne JG, Martin K, Coldham G, Krause B (1997) Three-dimensional
polyethylene wear of a press-fit titanium prosthesis: Factors influencing
generation of polyethylene debris. J Arthroplasty. 12:256-266.
Di Puccio F, Mattei L (2015) Biotribology of artificial hip joints. World J Orthop. 6:77-
94.
Digas G, Karrholm J, Thanner J, Malchau H, Herberts P (2004) The Otto Aufranc
Award. Highly cross-linked polyethylene in total hip arthroplasty: randomized
evaluation of penetration rate in cemented and uncemented sockets using
radiostereometric analysis. Clin Orthop Relat Res. 429:6-16.
Eggli S, Pisan M, Müller ME (1998) The value of preoperative planning for total hip
arthroplasty. J Bone Joint Surg Br. 80:382-390.
Floerkemeier T, Schwarze M, Hurschler C, Gronewold J, Windhagen H, von Lewinski
G, Budde S (2017) The Influence of Tribological Pairings and Other Factors
on Migration Patterns of Short Stems in Total Hip Arthroplasty. Biomed Res
Int. 2017.
Foss OA, Klaksvik J, Benum P, Anda S (2008) Polyethylene acetabular wear in hip
prostheses: computer-simulated quantification of error caused by changes in
pelvic orientation and direction of wear. Acta Orthop. 79:618-23.
Galvin AL, Jennings LM, Tipper JL, Ingham E, Fisher J (2010) Wear and creep of
highly crosslinked polyethylene against cobalt chrome and ceramic femoral
heads. Part H: J Engineering in Medicine. 224:1175-1183.
Geerdink CH, Grimm B, Vencken W, Heyligers IC, Tonino AJ (2008) The
determination of linear and angular penetration of the femoral head into the
acetabular component as an assessment of wear in total hip replacement: a
comparison of four computer-assisted methods. J Bone Joint Surg Br. 90:839-
846.
Glyn-Jones S, McLardy-Smith P, Gill HS, Murray DW (2008) The creep and wear of
highly cross-linked polyethylene: a three-year randomised, controlled trial
using radiostereometric analysis. J Bone Joint Surg Br. 90:556-561.
Goldvasser D, Hansen VJ, Noz ME, Maguire Jr GQ, Zeleznik MP, Olivecrona H,
Bragdon CR, Weidenhielm L, Malchau H (2014) In vivo and ex vivo
measurement of polyethylene wear in total hip arthroplasty: comparison of
measurements using a CT algorithm, a coordinate-measuring machine, and a
micrometer. Acta Orthop. 85:271-5.
61
Reference list
Gonzalez Della Valle A, Comba F, Taveras N, Salvati EA (2008) The utility and
precision of analogue and digital preoperative planning for total hip
arthroplasty. Int Orthop. 32:289-94.
Grillini L, Affatato S (2013) How to measure wear following total hip arthroplasty. Hip
Int. 23:233-42.
Grimberg A, Jansson V, Melsheimer O (2019) Jahresbericht 2019, EPRD Deutsche
Endoprothesenregister.
Guermazi A, Hunter DJ, Roemer FW (2009) Plain radiography and magnetic
resonance imaging diagnostics in osteoarthritis: validated staging and scoring.
J Bone Joint Surg Am. 91 Suppl 1:54-62.
Higgins JE, Conn KS, Britton JM, Pesola M, Manninen M, Stranks GJ (2020) Early
Results of Our International, Multicenter, Multisurgeon, Double-Blinded,
Prospective, Randomized, Controlled Trial Comparing Metal-on-Metal With
Ceramic-on-Metal in Total Hip Arthroplasty. J Arthroplasty. 35:193-197. e2.
Hill J, Diamond O, O’Brien S, Boldt J, Stevenson M, Beverland D (2015) Early
surveillance of ceramic-on-metal total hip arthroplasty. Bone Joint J. 97:300-
305.
Hohle P, Schroder SM, Pfeil J (2015) Comparison between preoperative digital
planning and postoperative outcomes in 197 hip endoprosthesis cases using
short stem prostheses. Clin Biomech (Bristol, Avon). 30:46-52.
Hui AJ, Mccalden RW, Martell JM, Macdonald SJ, Bourne RB, Rorabeck CH (2003)
Validation of two and three-dimensional radiographic techniques for
measuring polyethylene wear after total hip arthroplasty. J Bone Joint Surg
Am. 85:505-511.
Ilchmann T, Kesteris U, Wingstrand H (1998) Effect of pelvic tilt on radiographic
migration and wear measurements after total hip arthroplasty. Hip Int. 8:16-
23.
Ilchmann T, Lüem M, Pannhorst S, Clauss M (2012) Acetabular polyethylene wear
volume after hip replacement: Reliability of volume calculations from plain
radiographs. Wear. 282-283:69-75.
Ilchmann T, Mjöberg B, Wingstrand H (1995) Measurement accuracy in acetabular
cup wear: three retrospective methods compared with Roentgen
stereophotogrammetry. J Arthroplasty 10:636-642.
Inoue D, Kabata T, Maeda T, Kajino Y, Fujita K, Hasegawa K, Yamamoto T, Tsuchiya
H (2015) Value of computed tomography-based three-dimensional surgical
preoperative planning software in total hip arthroplasty with developmental
dysplasia of the hip. J Orthop Sci. 20:340-6.
Jacobsen A, Seehaus F, Hong Y, Cao H, Schuh A, Forst R, Sesselmann S (2018)
Model-based roentgen stereophotogrammetric analysis using elementary
geometrical shape models: 10 years results of an uncemented acetabular cup
component. BMC Musculoskelet Disord. 19:335.
Jeffers JRT, Walter WL (2012) Ceramic-on-ceramic bearings in hip arthroplasty: state
of the art and the future. J Bone Joint Surg Br. 94:735-745.
Johanson PE, Furnes O, Ivar Havelin L, Fenstad AM, Pedersen AB, Overgaard S,
Garellick G, Makela K, Karrholm J (2017) Outcome in design-specific
comparisons between highly crosslinked and conventional polyethylene in
total hip arthroplasty. Acta Orthop. 88:363-369.
Johanson PE, Shareghi B, Eriksson M, Kärrholm J (2020) Wear measurements with
use of radiostereometric analysis in total hip arthroplasty with obscured
femoral head. Journal of Orthopaedic Research®.
Kang JS, Park SR, Ebramzadeh E, Dorr LD (2003) Measurement of Polyethylene
Wear in Total Hip Arthroplasty Accuracy Versus Ease of Use. Yonsei Med J.
44:473-478.
Karachalios T, Komnos G, Koutalos A (2018) Total hip arthroplasty: Survival and
modes of failure. EFORT Open Rev. 3:232-239.
Kärrholm J, Gill RHS, Valstar ER (2006) The history and future of radiostereometric
analysis. Clin Orthop Relat Res 448:10-21.
62
Reference list
63
Reference list
64
Reference list
65
Reference list
66
List of abbreviations
6 List of abbreviations
2D two dimensional
3D three dimensional
AP/a.p. anteroposterior
c.c. craniocaudal
CT computed tomography
m.l. mediolateral
67
List of abbreviations
OA osteoarthritis
PE polyethylene
SD standard deviation
68
List of publications
7 List of publications
➢ Junzhe Wu, Yutong Hong, Zhuang Kang, Jing Xu, Raimund Forst, Frank
Seehaus. Assessment of interchangeable applicability of two wear
detection methods for total hip arthroplasty within an experimental
phantom-model setup. (in the process of submission)
➢ Effects of pelvic orientations on two wear detection methods for total hip
arthroplasty. 26th Congress of the European Society of Biomechanics
(Conference submission). In 2020
69
Appendix
8 Appendix
70
Acknowledgements
9 Acknowledgements
This research project was completed under the guidance of my supervisor,
Prof. Raimund Forst. After the dissertation, I would like to express my gratitude
to him. He instructed me to do scientific research with the normative methods
and meticulous attitude, which benefited me a lot.
I also would like to thank my second supervisor, Dr. Frank Seehaus, for his
enthusiastic guidance and help in project design, experimental operation, and
revision of the thesis. When I encountered difficulties in the experiment, he
provided me with many useful suggestions.
On the other hand, I would like to show appreciation to my colleagues for their
help in the operation process of the experiment. Without them, I would not be
able to complete my experiment.
Finally, I will give special thanks to my family for everything they did for me.
71