LCD3.3
Detection
Smiths
Detection
Equipment Part No. 15133
Smiths
Detection OPERATIONAL INSTRUCTIONS/USER MANUAL
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Prepared by: Smiths Detection - Watford Ltd.
Smiths
Detection
Park Avenue
Bushey
Smiths Watford
Detection Hertfordshire
Smiths
WD23 2BW
Detection U.K.
Smiths
Detection Part No. 15459-7
Smiths Date of Issue December 2012
Detection
Smiths
Detection
Smiths
Preliminary Pages
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Chemical Agent Detector
Smiths
Detection
Smiths
LCD3.3
Detection
Smiths
Detection Equipment Part No. 15133
Smiths
Detection
OPERATIONAL INSTRUCTIONS/USER MANUAL
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Smiths
Detection
Page i
Preliminary Pages
Page ii
Preliminary Pages
TABLE OF CONTENTS
FOREWORD .......................................................................................................................................... v
Page iii
Preliminary Pages
Page iv
Preliminary Pages
FOREWORD
© Smiths Detection – Watford Limited
All rights reserved. No part of this publication may be reproduced or utilised in any form or by any
means, electronically or mechanically, including photocopying and microfilm, without permission in
writing from Smiths Detection – Watford Limited.
It shall be the responsibility of the purchaser solely to establish, to its entire satisfaction, the suitability of
the product for particular applications. It is also the purchaser's responsibility to use and maintain the
product in accordance with these procedures and recommendations.
The purchaser is advised that the equipment is subject to End User Licence Agreements (EULA) and
the technology used in the equipment may be subject to further military and/or commercial restrictions
regarding Intellectual Property Rights (IPR) and the use or transfer of information.
Smiths Detection – Watford Limited have made every effort to ensure the accuracy of this manual but in
accordance with our policy of continuous improvement we reserve the right to make changes, without
prior notice, to the equipment which is described. Smiths Detection Limited accepts no liability for any
consequential loss, injury or damage resulting from the use/misuse of the supplied information or from
any errors or omissions to this manual.
For further information or any other queries regarding equipment maintenance, servicing etc. contact:
Customer Services
Smiths Detection – Watford Limited
459 Park Avenue
Bushey, Watford
Hertfordshire
England
WD23 2BW
Tel: +44 (0) 1923 658170
Fax: +44 (0) 1923 240285
Email: UK Support@smithsdetection.com
Page v
Preliminary Pages
Page vi
Preliminary Pages
SAFETY SUMMARY
The following are general safety precautions and instructions that must be understood and applied
during phases of operation and maintenance to ensure personal health and safety and the protection of
the equipment. Sections may be repeated elsewhere in this publication for emphasis.
EQUIPMENT LIMITATIONS
The LCD3.3 is a chemical point detector and samples only in the immediate vicinity of the rain cap. The
operator must always be aware of changing wind direction as this could quickly bring a hazardous level
of agent vapour to a previously safe area.
The detector is designed to detect chemical vapours. If chemicals are encountered in their liquid state
these may not be detected. Physical contact by the detector with the liquid will cause contamination of
the equipment.
The equipment does not present any significant hazard to personal safety during normal operations.
However, the WARNINGS and Cautions and general safety information notes that follow have been
identified as potential hazards if the equipment is damaged, mishandled or used incorrectly.
WARNINGS
WARNING – RISK OF EXPLOSION
LCD3.3 MAY PROVIDE AN IGNITION SOURCE IN AN EXPLOSIVE ATMOSPHERE. DO NOT USE
LCD3.3 IN AN ENVIRONMENT WHERE EXPLOSIVE VAPOURS ARE PRESENT.
Page vii
Preliminary Pages
WARNING – BATTERIES
THE LCD3.3 CAN USE LITHIUM NON-RECHARGEABLE BATTERIES OR ALKALINE NON-
RECHARGEABLE BATTERIES OR NIMH RECHARGEABLE BATTERIES. THESE BATTERIES
ARE A FLAMMABLE, CORROSIVE AND VAPOUR HAZARD. LITHIUM BATTERIES CONTAIN,
LITHIUM IRON DI-SULPHIDE, AND AN ELECTROLYTE. THE LITHIUM REACTS VIGOROUSLY
WHEN IMMERSED IN WATER. THE ELECTROLYTE IS FLAMMABLE AND HIGHLY CORROSIVE
DO NOT IMMERSE BATTERIES IN WATER, DECONTAMINANT SOLUTION OR OTHER LIQUIDS.
DO NOT CRUSH OR BURN BATTERIES.
DO NOT ATTEMPT TO RECHARGE NON-RECHARGEABLE BATTERIES.
DO NOT STORE BATTERIES AT TEMPERATURES ABOVE 158°F (70°C)
IN THE EVENT THAT A BATTERY WITHIN THE LCD3.3 IS SHORTED OR STARTS TO HEAT UP,
IT SHOULD BE REMOVED FROM THE UNIT AND TAKEN TO A WELL VENTILATED AREA TO
COOL DOWN. ONCE THE TEMPERATURE HAS DROPPED THE BATTERY CAN BE DISPOSED
OF ACCORDING TO LOCAL PROCEDURES FOR HAZARDOUS WASTE. SUITABLE
PROTECTIVE CLOTHING SHOULD BE WORN WHEN HANDLING SUSPECTED BATTERIES. IF
THE SKIN OR EYES COME INTO CONTACT WITH THE ELECTROLYTE, WASH THOROUGHLY
WITH WATER AND SEEK MEDICAL ATTENTION.
Page viii
Preliminary Pages
Page ix
Preliminary Pages
WARNING – BURNS
IF THE OUTER COVER OF A BATTERY INSTALLED IN THE BATTERY CASSETTE IS DAMAGED,
A SHORT CIRCUIT CAN OCCUR AND THE BATTERIES MAY OVERHEAT. ALWAYS CHECK THE
OUTER COVER OF EACH BATTERY IS UNDAMAGED BEFORE INSTALLING IN THE BATTERY
CASSETTE
CAUTIONS
Page x
Preliminary Pages
Page xi
Preliminary Pages
Page xii
Preliminary Pages
REVISION RECORD
Page xiii
Preliminary Pages
Page xiv
Chapter 1
CHAPTER 1 INTRODUCTION
1.1 INTRODUCTION
In operation LCD3.3 continually samples the air for gases and vapours and uses non-radioactive
Corona Discharge Ion Mobility Spectrometry (IMS) techniques for identification of chemical vapours.
When operated as a personal detector the LCD3.3 has a sampling cycle time of five seconds i.e.
samples the environment once every 5 seconds. This detection cycle provides for a longer battery life.
When the LCD3.3 is used as a survey instrument the sampling cycle is altered to sample every two
seconds allowing for quicker and more thorough sampling but at the expense of battery life. The
LCD3.3 issues visual and audible alarms in case of detection and the visual displays inform the user of
the type and concentration of the hazard detected.
It is important to note that LCD3.3 is a point detector and samples air only in the immediate vicinity of
the Rain Cap and cannot give a realistic assessment of a vapour hazard over an area from one
position. Air sample conditions a short distance away from the detector may be quite different and the
user must therefore be aware that a change in wind direction could quickly bring a hazardous level of
agent vapour to a previously safe area.
Page 1
Chapter 1
The LCD3.3 may be used by personnel in various environmental conditions and can be held in either
hand and operated whilst dressed in protective clothing. The detector can also be operated while inside
the supplied carrying pouch. The carrying pouch can be fastened to clothing, especially armed forces
apparel (Load Bearing Equipment (LBE)), to give a “hands free” detection capability.
The LCD3.3 uses various audible signals for system alarms and alerts, and visual displays are a
combination of a Liquid Crystal Display (LCD) and Light Emitting Diodes (LED). All media are used to
convey information to the user about the status of the detector and about any substances that have
been detected. The LCD visual display is also used to show the menu system that is used to configure
the detector.
The LCD3.3 can be configured by the operator to suit local operating requirements and conditions using
the control buttons to access a series of menus shown on the liquid crystal display (LCD) The menu
system is accessed and the various settings selected using the menu (M), enter (E), up-arrow and
down-arrow buttons on the detector. Using the menu and control buttons the operator can access
system functions, modify the system settings or interrogate or change system information as shown
below.
1. Perform system tests – Provides the operator with the facility to test the display and audio functions,
perform a confidence test to prove correct operation of the system.
2. Configure the system – Provides the operator with the facility to set the display language and date
format to suit operational requirements and apply a ‘lock’ to the control button to prevent accidental
activation.
3. Set the detection mode – Provides the operator with the facility to select the system detection mode.
a) Standard mode. In this mode LCD3.3 provides a rapid visual and audible warning when Chemical
Warfare Agent (CWA) and Toxic Industrial Chemical (TIC) vapours at or above the pre-set alarm
threshold are detected with the display showing substance identification (CWA only) and
concentration of the hazard. While in this mode LCD3.3 samples the environment once every 5
seconds. This is known as a Standard detection cycle. This detection cycle provides a longer life
for a single set of batteries.
b) CWA mode. In this mode the LCD3.3 operates in a similar fashion to survey mode but for CWA
only thereby reducing the possibility of false positive alarms under certain circumstances. While in
this mode LCD3.3 samples the environment once every 5 seconds.
c) Survey mode. In this mode LCD3.3 samples for CWA similar to CWA mode but the sampling
cycle is altered to sample every 2 seconds. This allows quicker and more thorough sampling to
investigate specific areas of cargo, equipment, personnel or facilities. Use of the LCD3.3 in
Survey mode has an adverse effect on battery life. In survey mode the detector Raincap is
replaced by a re-usable Survey Nozzle. For further information on installing the Survey Nozzle
refer to Chapter 2 of this handbook.
4. Set Hazard Dose Monitoring and Hydrogen Cyanide (AC) Sensitivity. In addition to the modes of
operation available to LCD3.3 there are two further settings that affect the operation of the detector.
These are Hazard Dose Monitoring and Hydrogen Cyanide (AC) Sensitivity.
a) Hazard Dose Monitoring. The LCD3.3 functionality includes monitoring of Hazard Dose (also
known as accumulated dose) exposure to selected target substances, providing indication when
pre-determined dosage levels are reached based on exposure time and substance concentration.
Dose information is stored in the instrument log and may be viewed and interrogated after
deployments. Hazard dose alarms may be disabled as required, though the information will
continue to be logged for later use. For further information on about Hazard Dose reports see
Section 2.4.1.8.2. For further information on about Hazard Dose Alarm setting see Section
2.4.1.9.6.
b) Hydrogen Cyanide (AC) Sensitivity. The LCD3.3 has been designed and configured to alarm to
hydrogen cyanide (AC) at less than half of the IDLH value. For operation in environments where
there will be a known source of AC in the vicinity of the detector e.g. firing of munitions, burning of
organic material, etc. a reduced sensitivity function is available to limit the occurrence of nuisance
Page 2
Chapter 1
alarms from AC. Operation in this setting is indicated on the instrument display, and users should
be aware that AC sensitivity is lower than the stated specifications until the sensitivity is returned
to the default level. For further information on setting AC sensitivity see Section 2.4.1.9.7.
5. Set display properties – Provides the operator with the facility to set the display lighting level,
contrast and backlight to suit operational requirements. Night vision goggles (NVG) can be worn
when viewing the display whilst the unit is on the NVG setting.
6. Set audio level – Provides the operator with the facility to set the system audible alarm and alert
levels to suit operational requirements.
7. Monitor status reports – Provides the operator with the facility to access the LCD3.3 system fault log
and other system information, observe the system run time indicator and accumulated dose
indicator.
8. Reset system counters – Provides the operator with the facility to reset the system runtime indicator,
sieve pack life indicator and accumulated hazard dose indicator.
For further information on the configuration and operation of the LCD3.3 refer to CHAPTER 2 of this
handbook.
Visual Indications
The LCD3.3 visual display is a combination of a Liquid Crystal Display (LCD) and Light Emitting Diodes
(LED). Both media are used to convey information to the user about any substances that have been
detected and about the status of the detector.
The detector has two Light Emitting Diode (LED) lamps on the top of the detector. The smaller of the
LEDs is the system status lamp and is yellow/amber in colour. This lamp serves as a health indicator
and gives two basic indications
Detector alert.
With the detector in this condition the lamp will flash at a rate of ½-second on then ½-second off
indicating that the detector requires attention. The user can ascertain the nature of the alert by
observing the display on the front of the unit where an alert message will be shown e.g. “Battery
low”. For further information on alert messages refer to 0 Maintenance Information and Instructions.
Detector fault or major fault.
With the detector in this condition the lamp will be continuously lit. The user can ascertain the
nature of the fault or major fault by observing the front display on the unit where a fault message
will be shown e.g. “MAJOR FAULT Health check” or “FAULT Low temperature”. For further
information on alert messages refer to 0 Maintenance Information and Instructions.
The larger LED is the hazard alarm lamp and is red in colour. This lamp serves to warn the user that the
detector is in an alarm condition. The user can ascertain the nature of the alarm by observing the front
display on the unit where an alarm message will be shown giving the Chemical Warfare Agent identity
(GA, GB, AC, VX etc.) and a bar display with solid or hollow blocks to indicate relative agent
concentration in the sample. Detection of TICs gives an ‘8-bar’ display only; the identity of the
substance detected is not displayed. For further information on alarm messages refer to CHAPTER 2
Operating Instructions.
Page 3
Chapter 1
Audible Indications
1. Alert: The alert tone is used to bring the operator’s attention to equipment status such as when
there is only 10% of battery life remaining, or when there is approximately 72 hours of operational
life left in the sieve pack. The alert tone is different from that of the detection alarm.
2. Alarm: The alarm tone is used to indicate the detection of a recognised hazardous substance at or
above the alarm threshold. The audible alarm is a repeating, swept frequency tone, each sweep
rising from 2 kHz to 3 kHz over a period of 1 second.
When set to the ‘High’ audio level the alarm volume is 65 dB(A) at one metre. LCD3.3 may be used with
an ear-piece for operation in tactical situations. Use of the ear-piece will mute the audio alarm/alert
provided by the instrument’s integral sounder and produce equivalent audible alarm/alert signals at the
user's ear.
The LCD3.3 equipment has initially been packed by the manufacturer for storage in a clean, dry
environment for at least 10 years.
Page 4
Chapter 1
Page 5
Chapter 1
WARNING – BATTERIES
THE LCD3.3 CAN USE LITHIUM NON-RECHARGEABLE BATTERIES OR ALKALINE NON-
RECHARGEABLE BATTERIES OR NIMH RECHARGEABLE BATTERIES. THESE BATTERIES
ARE A FLAMMABLE, CORROSIVE AND VAPOUR HAZARD. LITHIUM BATTERIES CONTAIN,
LITHIUM IRON DI-SULPHIDE, AND AN ELECTROLYTE. THE LITHIUM REACTS VIGOROUSLY
WHEN IMMERSED IN WATER. THE ELECTROLYTE IS FLAMMABLE AND HIGHLY CORROSIVE.
DO NOT IMMERSE BATTERIES IN WATER, DECONTAMINANT SOLUTION OR OTHER LIQUIDS.
DO NOT CRUSH OR BURN BATTERIES.
DO NOT ATTEMPT TO RECHARGE NON-RECHARGEABLE BATTERIES.
DO NOT STORE BATTERIES AT TEMPERATURES ABOVE 158°F (70°C)
IN THE EVENT THAT A BATTERY WITHIN THE LCD3.3 IS SHORTED OR STARTS TO HEAT UP,
IT SHOULD BE REMOVED FROM THE UNIT AND TAKEN TO THE OUTDOORS OR TO A WELL
VENTILATED AREA TO COOL DOWN. ONCE THE TEMPERATURE HAS DROPPED THE
BATTERY CAN BE DISPOSED OF ACCORDING TO LOCAL PROCEDURES FOR HAZARDOUS
WASTE. SUITABLE PROTECTIVE CLOTHING SHOULD BE WORN WHEN HANDLING
SUSPECTED BATTERIES. IF THE SKIN OR EYES COME INTO CONTACT WITH THE
ELECTROLYTE, WASH THOROUGHLY WITH WATER AND SEEK MEDICAL ATTENTION.
Page 6
Chapter 1
The batteries used to operate the LCD3.3 Detector are installed into a cassette prior to insertion into the
detector. The detector uses 4 x AA- size batteries. Lithium Iron Disulphide batteries are recommended
as this battery type gives the best battery life performance, particularly at low temperatures. Alkaline
Manganese Dioxide and NiMH batteries can be used as an alternative but give a reduced performance,
particularly at low temperatures.
3.6v AA Lithium Thionyl Chloride batteries and Nickel-Cadmium (Ni-Cd) batteries are not recommended
and should not be used in the LCD3.3 Detector.
Table 2 ‘Typical Battery Life Times’ gives the expected performance for different operating modes with
Lithium Ion Disulphide batteries fitted.
All batteries used within the detector should be of the same type, any mixing of battery types may cause
damage to the detector or could cause the detector to give misleading battery indications and warnings.
When the battery type is set correctly, the detector will determine the battery low and change battery
warnings at the correct voltage levels for the batteries in use.
When there is approximately 5 hours of battery life remaining a ‘low battery’ indication will be given. The
low battery indication means that all batteries should be replaced.
When there is approximately 1-2 hours of battery life remaining the detector will give a ‘change battery‘
warning. The ‘change battery’ warning is the last warning that will be given before the detector stops
sampling. When a ‘change battery’ indication is given all batteries should be replaced immediately.
When there is less than 1 hour battery life remaining the detector will not sample and will continue to
display the ‘change battery’ warning. The detector will not provide sampling or detection functions until
alternate power is provided. If batteries are completely depleted the detector will not switch on.
Page 7
Chapter 1
Equipment should be stored in a dry environment within the following ambient temperature range:
Equipment placed into storage for any period of time are to have the batteries removed.
If equipment is to be placed into storage for a period up to and including 12 months it is not necessary
to remove sieve packs although it is likely that they will need to be replaced when equipment is reused.
If equipment is to be placed into storage for 12 months or more then the operational Sieve Packs
should be removed and discarded and a Dummy Sieve Pack installed, See storage procedures in Para.
2.9.1.
Page 8
Chapter 2
Page 9
Chapter 2
Page 10
Chapter 2
Page 11
Chapter 2
For further information on the menu system, and use, refer to Section.2.4.
The LCD3.3 Detector has two Light Emitting Diodes (LEDs). The smaller of the LEDs is the System
Status Alert LED and is yellow/amber in colour. This serves as a power and health indicator and gives
two indications:
The larger LED is the Detector alarm and is red in colour. This lamp serves to warn the user that the
detector is in an alarm condition. The user can determine the nature of the alarm by observing the front
display on the Detector Unit. This shows alarm message giving the Chemical Agent identity (GA, GB,
AC, VX etc.) and a bar display with solid or hollow blocks to indicate the Chemical Agent concentration.
Page 12
Chapter 2
Page 13
Chapter 2
2.2 START UP
1. Inspect the LCD3.3 to confirm that it has the correct Sieve Pack installed. Install the correct Sieve
Pack if required. For further information on replacing the Sieve Pack refer to Section 5.4.4 in 0 of
this handbook.
WARNING – BURNS
IF THE OUTER COVER OF A BATTERY INSTALLED IN THE BATTERY CASSETTE IS DAMAGED,
A SHORT CIRCUIT CAN OCCUR AND THE BATTERIES MAY OVERHEAT. ALWAYS CHECK THE
OUTER COVER OF EACH BATTERY IS UNDAMAGED BEFORE INSTALLING IN THE BATTERY
CASSETTE
2. Inspect batteries for damage. Do not use batteries that are damaged.
3. Install batteries (qty 4) to the Battery Cassette of the detector Unit, observing the polarity markings
on the Battery Cassette, then install Battery Cassette into the Detector. For further information on
battery installation refer to the section entitled Battery Replacement in Section 5.4.2 in 0 of this
manual.
4. Switch on the detector by rotating the Rain Cap to the ON position. Observe the display and LEDs
for normal operation and make sure the case of the detector does not become warm to touch. If the
display or the LEDs appear not to operate correctly or the case of the detector becomes warm or
hot to touch the batteries inside may have short circuited. If this should happen switch off the
detector and immediately remove the batteries from the affected detector. Obtain replacement
batteries and restart the detector. If the situation re-occurs obtain a replacement Battery Cassette
for the detector. If the situation again re-occurs obtain a replacement detector. If the detector
appears to be working normally switch off the instrument by rotating the Raincap clockwise to
close.
5. The LCD3.3 is now ready for use.
Page 14
Chapter 2
1. Inspect the LCD3.3 for damage and serviceability. Make sure that the detector has batteries and a
sieve pack installed as detailed in Section 2.2.1. For further information on installing batteries and a
sieve pack see Chapter 5.
2. Switch on the detector by rotating the Rain Cap counter-clockwise so that the pointer is pointing
directly at the raised mark on the body of the detector adjacent to the earpiece socket. The detector
automatically starts up in the last used mode.
Page 15
Chapter 2
3. Allow the detector to complete its start up routine. This should take approximately 1 minute. If the
detector has not completed its start up routine after 5 minutes refer to Fault Analysis in Chapter 5 of
this handbook.
4. Perform a Confidence Test of the detector. For further information on performing a Confidence Test
see Section 2.3.1 of this handbook.
5. The LCD3.3 is ready for use following successful completion of a Confidence Test. If the detector
fails the confidence test close the Rain Cap to switch off the instrument and refer to Fault Analysis in
Chapter 5 of this handbook.
Page 16
Chapter 2
NOTE. Low temperatures and high winds can affect the performance of the confidence tester. In strong
winds, the user should shield the application of simulants by seeking shelter from the wind or act as a
windbreak by turning their back to the wind.
NOTE. While the confidence tester is kept in its sealed package, it is possible for vapours of the two
simulants to become mixed. If the confidence test is carried out immediately after the confidence tester
is removed from its sealed package, it is possible that the detector may show both G and H alarms
during the test. To prevent this, the confidence tester should be left to “air” for approximately one minute
to allow the mixed vapours to dissipate. Moving the confidence tester gently through the air will help to
dissipate the mixed vapours. In the event of a simultaneous G and H alarm during the confidence test
when one end of the tester is presented to the detector, this is acceptable and does not indicate a fault
with the detector.
NOTE. A confidence test may be performed in any detection mode.
1. Make sure the detector is switched on and has completed its start up routine. For further information
refer to section 2.2.
2. Obtain the Confidence Tester from its stowed location.
3. Select the “Confidence Test / Yes” option in the detector Menu system. See section 2.4.1.2 for
further information.
4. The detection mode indicated on the display will be “TEST”. When the message “Sampling” appears
on the display, the LCD3.3 is ready for the confidence test.
5. Remove the cap from the vapour vent at one end of the tester. The simulant type is indicated by the
raised letter G or H on the tester casing.
6. Place the open end of the confidence tester adjacent to but not touching the detector Raincap or
survey nozzle inlet, and, without pressing the sample button, allow the detector to sample the
simulant for approximately 1 second.
NOTE. The sample button on the confidence tester should only be pressed if the normal diffusion of
the simulant fails to produce an alarm e.g. if low temperature and/or high winds is adversely
affecting the diffusion of simulant. If necessary, briefly depress the sample button. If the detector
fails to alarm repeat depression of the sample button at one second intervals up to a maximum of 5
Page 17
Chapter 2
puffs or until the unit alarms. If the unit fails to alarm refer to Fault Analysis in Chapter 5 of this
handbook.
7. The display will show bars next to the appropriate icon (G or H). When 3 or more bars are displayed
the LCD3.3 will give an audible alarm and the red hazard LED on the top display will flash. When
the detector has responded to the simulant in accordance with the test parameters the bar display
will be replaced with the words “Test pass”.
See previous note regarding simultaneous G and H alarms. The audible alarm may be cancelled
after it has started to sound. For further information refer to the section of the handbook on
cancelling the audible alarm.
8. After the detector has correctly responded to the first simulant repeat the test using the simulant at
the opposite end of the confidence tester. If functioning correctly the detector will initially show bars
followed by the words “Test pass”.
9. After successful completion of the confidence test using both “G” & “H” simulants, the message
“Clearing down” will appear on the display The detector will remain in confidence test mode to allow
the inlet to clear from simulant vapour before returning to the detection mode. If the inlet does not
clear from simulant vapour within 10 minutes, the detector will report a saturation fault. In this fault
state, the detector will continue to attempt to clear down and will only return to the appropriate
detection mode once the inlet is clear from simulant vapour. In the event that the detector does not
clear down it must be treated as contaminated and should be decontaminated in accordance with
the procedure shown in this handbook.
10. Return the Confidence Tester to its stowage. The detector is now ready for use.
Note: If the confidence test is not carried out successfully in both G and H modes within 30 seconds,
the message “WAIT - Clearing down” will appear on the display and the LCD3.3 will return to the
appropriate detection mode.
The operator must be aware that some configurable settings are retained when equipment is switched off while
others revert to a default setting. For further information refer to TABLE 7.
Page 18
Chapter 2
Language
Confidence Test
Detection Mode
Page 19
Chapter 2
AC Detection
Display
Time/Date
Audio
Status
Setup
Keyguard
System Tests
The full range of options cannot be displayed on the LCD screen. To reveal options that are not
displayed the operator must press the ‘Cursor Down ( )’ button. To access the sub menu for the
required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’ buttons and press
the ‘Enter (E)’ button to select and proceed to the next menu.
To return to the System Operational Status Screen press the ‘Menu (M)’ button
2.4.1.1 Language
The Language menu is accessed from the Main Menu and is comprised of the options as follow:
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection.
Page 20
Chapter 2
Yes
No
Select the ‘Yes’ option to continue with the Confidence Test. Select the ‘No’ option to abort the
Confidence Test and return to the Main Menu. If no test is performed, the system will ‘timeout’ and
revert to the current mode. There is no requirement to select a specific Detection Mode when performing a
Confidence Test.
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection.
Page 21
Chapter 2
Default
Low
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection.
The Detection mode selected is indicated on the screen in the status line. If the AC menu has not been
enabled using the Setup Menu (see section 2.4.1.9.7) only the current AC Sensitivity setting will be
displayed.
Lights
Backlight Mode
Contrast
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection and proceed to the next menu. To
return to the Main Menu press the ‘Menu (M)’ button. Select a new option from the Main Menu or press
‘M’ again to return to the System Operational Status Screen.
Page 22
Chapter 2
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection and return to the Display Menu. Select
a new option from the Display Menu or press ’M’ again to return to the Main Menu. Select a new option
from the Main Menu or press ‘M’ again to return to the System Operational Status Screen.
Power save
Continuous
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection, then Press ‘M’ to return to the Display
Menu. Select a new option from the Display Menu or press ‘M’ again to return to the Main Menu. Select
a new option from the Main Menu or press ‘M’ again to return to the System Operational Status Screen.
Page 23
Chapter 2
Select the required level using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’ buttons and press the ‘Enter
(E)’ button to confirm the selection and return to the Display Menu.
Select a new option from the Display Menu or press ‘M’ again to return to the Main Menu. Select a new
option from the Main Menu or press ‘M’ again to return to the System Operational Status Screen.
Time and Date Format can be changed using the Date Format screen in the Setup Menu. The access
code is required.
Page 24
Chapter 2
High
Medium
Low
Off
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection, then Press ‘M’ to return to the main
menu.
To return to the Main Menu without making changes press the ‘Menu (M)’ button. Select a new option
from the Main Menu or press ‘M’ again to return to the System Operational Status Screen.
Page 25
Chapter 2
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection and proceed to the next menu.
To return to the Main Menu press the ‘Menu (M)’ button. Select a new option from the Main Menu or
press ‘M’ again to return to the System Operational Status Screen.
To return to the Status menu press the ‘Menu (M)’ button. Select a new option from the Status menu or
press ‘M’ again to return to the Main Menu. Select a new option from the Main Menu or press ‘M’ again
to return to the System Operational Status Screen.
View
Reset
Dose alarms
To select the required option position the cursor using the ‘Cursor Up’ ( ) and ‘Cursor Down’ ( )
buttons and press the ‘Enter (E)’ button to confirm the selection and proceed to the next screen or
menu. This menu is not available in Survey mode.
Page 26
Chapter 2
Selection of the View option from the Hazard Dose menu will display calculated information to indicate
the accumulated dose over a period of time. This menu is not available in Survey mode.
To return to the Hazard Dose menu press the ‘Menu (M)’ button where the Reset option can be
selected if required. To return to the Status menu press the ‘Menu (M)’ button. Select a new option from
the Status menu or press ‘M’ again to return to the Main Menu. Select a new option from the Main Menu
or press ‘M’ again to return to the System Operational Status Screen.
Selection of the Reset option from the Hazard Dose menu will display the following further options.
Yes
No
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm selection and return to the Hazard Dose menu.
To return to the Status menu press the ‘Menu (M)’ button. Select a new option from the Status menu or
press ‘M’ again to return to the Main Menu. Select a new option from the Main Menu or press ‘M’ again
to return to the System Operational Status Screen.
Selection of the dose alarm option from the Hazard dose menu displays the following:
ON
OFF
If the Dose Alarm has not been enabled using the Dose Alarm menu (see section 2.4.1.9.6), which is
accessed via the Setup Menu, only the current Dose Alarm setting will be displayed. This screen is not
available in Survey mode.
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm selection and return to the Hazard Dose menu.
To return to the Status menu press the ‘Menu (M)’ button. Select a new option from the Status menu or
press ‘M’ again to return to the Main Menu. Select a new option from the Main Menu or press ‘M’ again
to return to the System Operational Status Screen.
Page 27
Chapter 2
To return to the Status menu press the ‘Menu (M)’ button. Select a new option from the Status menu or
press ‘M’ again to return to the Main Menu. Select a new option from the Main Menu or press ‘M’ again
to return to the System Operational Status Screen.
This screen will remain on the display until either it is changed by the operator or the detector enters an
alarm or alert condition.
Page 28
Chapter 2
To return to the Status menu press the ‘Menu (M)’ button. Select a new option from the Status menu or
press ‘M’ again to return to the Main Menu. Select a new option from the Main Menu or press ‘M’ again
to return to the System Operational Status Screen.
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection and proceed to the next menu.
Page 29
Chapter 2
To return to the Main Menu press the ‘Menu (M)’ button. Select a new option from the Main Menu or
press ‘M’ again to return to the System Operational Status Screen.
5 code digits Must be set to the correct code to enable access to extended Setup Menu.
Set Enter the code currently displayed.
Exit Exit the screen without changing access level.
To exit without changing entering an access code use the Menu (M) button to select ‘Exit’ and press the
‘Enter (E)’ button to confirm selection and return to the Setup menu.
Yes
No
Page 30
Chapter 2
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection and return to the Setup menu.
To return to the Status menu press the ‘Menu (M)’ button. Select a new option from the Status menu or
press ‘M’ again to return to the Main Menu. Select a new option from the Main Menu or press ‘M’ again
to return to the System Operational Status Screen.
Once accessed, the Reset Runtime menu will display the following further options.
Yes
No
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm selection and return to the Setup menu.
To return to the Status menu press the ‘Menu (M)’ button. Select a new option from the Status menu or
press ‘M’ again to return to the Main Menu. Select a new option from the Main Menu or press ‘M’ again
to return to the System Operational Status Screen.
The ‘Set Clock’ menu is accessed from the Setup Menu. Once accessed, the Set Clock menu will display the
further options as follows:
In this screen the M button rather than the cursor buttons is used to move the cursor between fields. Use the
procedures as follow to enter the access code or exit without change:
Page 31
Chapter 2
To exit the screen without changing the date and time whether or not the settings have been revised.
UK (dd/mm/yy)
US (mm/dd/yy)
ZULU (dd/hhhh/Z)
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection then press ‘M’ to return to the Setup
menu.
Page 32
Chapter 2
To return to the Status menu press the ‘Menu (M)’ button. Select a new option from the Status menu or
press ‘M’ again to return to the Main Menu. Select a new option from the Main Menu or press ‘M’ again
to return to the System Operational Status Screen.
The Dose Alarm Menu allows the user to enable or disable the Dose Alarm Menu. Selection of the
Dose Alarms option from the Setup Menu will display the following options:
Enable Allows the operator to use a menu to switch dose alarms on or off.
Disable Prevents the operator from changing whether dose alarms are on or off.
If the Dose Alarm menu ‘Disable’ option is selected then the ‘Dose Alarms’ selection menu (see section
2.4.1.8.2.3) available from the Hazard Dose Menu will show only the current alarm setting.
2.4.1.9.7 AC Menu
This option is a restricted access option and is not visible as an option on the Set Up menu until the
access code has been entered via the Access Level Menu. For further information refer to ‘Access
Level Menu’ above.
The AC Menu allows the user to enable or disable the AC Detection sensitivity selection menu.
Selection of the Dose Alarms option from the Setup Menu will display the following options:
Enable Allows the operator to use the AC detection menu to select the
AC sensitivity.
Disable Prevents the operator from changing the AC sensitivity.
Page 33
Chapter 2
Figure 36 AC Menu
If the AC Menu ‘Disable’ option is selected then the ‘AC Detection’ selection menu (see section 2.4.1.4)
available from the Main Menu will show only the current alarm setting.
Modem The detector will communicate with a Radio Modem connected to it.
GPS The detector will communicate with a GPS unit connected to it.
ATP-45 In the event of an alarm the detector will output an ATP-45 (NBC4) report.
The purpose of the Reset Position is to delete the coordinates of the last saved position of the Detector
Unit, should it be moved and not connected to a GPS receiver or a Radio Modem to generate an
updated position. If the Reset Position function is not used, the system will default to the last previously
used position. If the detector is connected to the Radio Modem there is no need to perform a position
reset; as the last position will be automatically overwritten.
Page 34
Chapter 2
Once accessed, the Reset Position menu will display the further options as follow:
Yes Deletes position data – dashes are indicated on the Position screen,
accessed from the Status Reports menu.
No Exits the screen without resetting position
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection.
By specifying the type of battery installed to the detector the unit can more accurately calculate battery
life and the remaining time before the detector would become inoperative.
Page 35
Chapter 2
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection and return to the Main Menu.
To return to the Main Menu press the ‘Menu (M)’ button. Select a new option from the Main Menu or
press ‘M’ again to return to the System Operational Status Screen.
To cancel keyguard protection press the ‘M’ and ‘E’ buttons at the same time.
2.4.1.11 System Tests Menu
The System Tests menu is accessed from the Main Menu and is comprised of the following options
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection. The equipment will perform the
selected test.
Page 36
Chapter 2
The Sound Test option is accessed from the System Tests menu and allows the operator to test the
audible tones of the Detector to prove that they are fully functional. Selection of the Sound Test option
from the System Tests menu will display the further options as follows:
To select the required option position the cursor using the ‘Cursor Up ( )’ and ‘Cursor Down ( )’
buttons and press the ‘Enter (E)’ button to confirm the selection. The equipment will perform the
selected test.
The Sound Test does not work when the detection is in the ‘WAIT’ condition.
When the Operator’s Earpiece is used to supply audible alarms to the operator only, the Sound Test
should also be performed prior to use to confirm Earpiece functionality.
Page 37
Chapter 2
The audible alarm can be muted by the operator by pressing any control button, ‘Cursor Up ( )’,
‘Cursor Down ( )’, ‘Enter (E)’ or ‘Menu (M)’. The alarm will restart if one of the trigger conditions
occurs.
1. Remove the Raincap. If the detector is operating this will cause the detector to shut down
automatically. To remove the Raincap from the ‘OFF’ position rotate the Raincap counter-clockwise
to the ‘ON’ position. To remove the Raincap from the ‘ON’ position press down the Raincap and
rotate counter-clockwise until the Raincap comes to a stop and no further rotation is possible. Lift
the Raincap vertically clear of the inlet structure. Take care that the Raincap does not become dirty
or contaminated.
Note. While the Raincap is removed, inspect the inlet for the build up of any dust or dirt. If the inlet is
dusty or dirty blow gently across the inlet to remove any dust or dirt particles. Do not brush or rub
the inlet. If the Raincap itself is dirty it can be rinsed in clean water and dried with a lint free cloth.
Do not use oils, lubricants or solvents.
Page 38
Chapter 2
2. Obtain the survey nozzle from its stowage pot. Note, if a new survey nozzle is required these are
stored in vapour proof packets.
3. Place the Raincap into the stowage pot for safekeeping and stow the stowage pot.
4. Partially close the Survey Nozzle and place into position on the inlet structure. When the Survey
Nozzle is in position fully close the Survey Nozzle making sure that two halves snap together. When
the Survey nozzle is correctly installed the detector will start up automatically
5. Perform a confidence test by applying the confidence tester to the Survey Nozzle inlet to confirm
that the detector is now ready for operation.
Note, when using the detector with the survey nozzle, make sure that the tip of the survey nozzle
does not come into contact with liquid chemical agents or other contaminants.
Removal Procedure (see Figure 44)
Page 39
Chapter 2
The LCD3.3 is designed to be worn by the user on Load Bearing Equipment (LBE). The detector should
be positioned so that it has an uninterrupted airflow and the inlet is not masked or covered.
Page 40
Chapter 2
If it is intended for the LCD3.3 detector to go into storage or to be transported the batteries should be
removed.
2.7 INTERFERENTS
Caution – Impaired Performance.
Due to the chemical composition of munitions, operating the LCD3.3 near artillery fire can
cause nuisance alarms. Operators should be aware of local environmental conditions when
operating the LCD3.3.
There are a few harmless vapours present in the atmosphere that can, under some circumstances, give
a false response in LCD3.3. The situations most likely to give a false response are in enclosed spaces,
or when sampling near strong vapour sources, e.g. in dense smoke. Some of the types of vapours
which have been found to give false readings are as follows:-
Aromatic vapours. Included in this group of materials are some brands of aftershave and perfume
and food flavourings. Some sweets such as peppermints and lozenges can cause a response if the
breath is exhaled directly onto the inlet.
Cleaning compounds. Some cleaning compounds and disinfectants contain additives that give them
a pleasant smell. Some of these additives such as menthol and methyl salicylate (ms) can give false
responses. Cleaning materials are by their nature spread over large surface areas and, therefore,
provide a considerable vapour source, particularly in enclosed spaces.
Chlorinated materials can cause a response especially in confined spaces with high concentrations
of vapour.
Smokes and fumes. The efflux from some rocket motors and breech fumes from some munitions
can give a response.
Page 41
Chapter 2
2.9.1 Storage
WARNING – TOXIC HAZARD
THE SIEVE PACK CONTAINS A MOLECULAR SIEVE. DO NOT DISPOSE OF ANY SIEVE PACK IN
FIRE AS TOXIC SUBSTANCES MAY BE RELEASED. DISPOSE OF A SIEVE PACK AS
HAZARDOUS WASTE IN ACCORDANCE WITH LOCAL RECOMMENDED PROCEDURES.
If the equipment is to be switched off for more than 5 days, one of the following storage procedures should be
used. There is a possibility that LCD3.3 may become contaminated with simulants if stored in close
proximity to the Confidence Tester. Make sure that the detector is stored within its charcoal pouch
(Consumable Part Number 14425) separately from the Confidence Tester.
1. Make sure all equipment and components are as clean and dry as possible prior to stowage. If it is
necessary to stow wet equipment, dry the equipment at the earliest opportunity. Equipment can be
wiped with a clean lint free cloth to remove larger volumes of dirt or fluid then ideally allowed to dry
in air. Do not use excessive direct or indirect heat in an attempt to rapidly dry the equipment.
2. Remove all batteries from the LCD3.3 Detector and dispose of the batteries in accordance with
standard operating procedures or local regulations.
3. Insert the Detector into the charcoal bag.
4. Stow all equipment. Refer to Section 1.2.6 for information on storage environment..
1. Make sure all equipment and components are as clean and dry as possible prior to stowage. If it is
necessary to stow wet equipment, dry the equipment at the earliest opportunity. Equipment can be
wiped with a clean lint free cloth to remove larger volumes of dirt or fluid then ideally allowed to dry
in air. Do not use excessive direct or indirect heat in an attempt to rapidly dry the equipment.
Page 42
Chapter 2
2. Remove all batteries from the LCD3.3 Detector and dispose of the batteries in accordance with
standard operating procedures or local regulations.
3. Remove operational Sieve Pack from the LCD3.3 Detector and dispose of as hazardous waste.
4. Install dummy Sieve Pack. (Ancillary Item Part Number 10720).
5. Insert the Detector into the charcoal bag.
6. Stow all equipment. Refer to Section 1.2.6 for information on storage environment..
2.9.2 Transportation
No equipment preparation is required prior transportation.
Page 43
Chapter 2
Page 44
Chapter 3
CHAPTER 3 TECHNICAL DESCRIPTION
3.1 GENERAL
The LCD3.3 uses ion mobility phenomena to respond to agent vapours. The technique used is that of
Ion Mobility Spectrometry (IMS). An air sample is drawn into the sampling line by a fan. The sample
then passes two pinhole inlets, one for each of the two ion mobility spectrometers. When the internal
pressure, under the control of a microprocessor, is reduced by the action of a diaphragm, sample is
pumped into the spectrometers through the pinhole inlets. On passing through the inlets, the target
materials enter ionisation sections where ions are generated by corona discharge ionisers.
Ionic clusters of high and low mobilities are then formed as a result of complex interchange reactions.
These clusters are then swept towards a gating grid by an electric field. The gating grids open in turn to
allow the ionic clusters to enter the two drift regions or drift tubes. The two drift tubes operate at
different electrical polarities. One drift tube collects ions of a positive charge to identify Nerve Agents,
whilst the other collects ions of a negative charge to identify Blister Agents.
Within the drift tubes, an electrical field imparts a constant force to the ions, and they separate out
according to their respective mobilities. Ions with the highest mobility travel faster than those with lower
mobilities therefore reaching the further end of the drift tubes first. At the further ends of the drift tubes
the ions collide with collectors giving rise to current pulses.
The actual material identification is conducted by the signal processing software, which sends the
appropriate signal to the detector display.
Page 45
Chapter 3
3.2.1 General
The pneumatic system can be split into two parts. The first is the inlet system, which introduces sample
air into the detector, and the second is the recirculating system, which services the ion mobility system.
The sample air is then passed over the two pinhole inlets (H and G) through which target materials are
drawn in when the diaphragm is activated to reduce the internal pressure. Once within the
spectrometers, and after ionisation is allowed to take place, the target materials are drawn into the
recirculation system. Sample air is expelled from the detector via the exhaust port under the raincap.
Page 46
Chapter 4
CHAPTER 4 INSTALLATION INFORMATION
4.1 GENERAL
No installation information is supplied in this manual for this equipment. Where the detector is used in
conjunction with other equipment where installation is required installation information will be shown in
the manual applicable to that equipment.
Page 47
Chapter 4
Page 48
Chapter 5
5.1 GENERAL
WARNING – ELECTRIC SHOCK
LCD3.3 IS POWERED BY BATTERIES. TO PREVENT ELECTRIC SHOCK MAKE SURE THAT
THE LCD3.3 IS SWITCHED OFF AND THE BATTERIES ARE REMOVED PRIOR TO ANY
MAINTENANCE ACTIVITY.
If the LCD3.3 fails to start up when switched on make sure that serviceable batteries are installed to the
detector. An appropriate display message will indicate other faults that may be remedied by the
operator. A fault condition will trigger an audible alert tone different to that of a CWA/TIC alarm.
The alert tone lasting approximately ½ a second is emitted when the LCD3.3 enters an alert condition
e.g. low battery power. The alert tone will repeat itself at the sample rate for a period of 60 seconds or
until cancelled by the operator or the unit is switched off. If the fault that has caused the alert condition
in the LCD3.3 is not corrected the flashing icon indicating the fault will become constantly lit to indicate
that the unit has stopped sampling. The alert tone will once again be emitted at the sample rate for a
period of 60 seconds or until cancelled by the operator.
A confidence test should always be performed after any corrective maintenance actions or fault
corrections.
Page 49
Chapter 5
The following table shows the corrective maintenance tasks that are applicable to the LCD3.3 and the
maintenance level at which those tasks should be performed.
Page 50
Chapter 5
7. Allow equipment to “weather” for 40 minutes.
8. If required record performance of the decontamination exercise.
9. A confidence test may now be performed to confirm that the detector is ready for operation.
1. If the LCD3.3 is operating, shut down the detector by closing the Raincap and wait for 3 seconds so
that data can be saved and the unit can shut down correctly.
2. Open the Battery Cassette locking cover.
3. Fully withdraw the Battery Cassette from the detector.
WARNING – BATTERIES
THE LCD3.3 CAN USE LITHIUM NON-RECHARGEABLE BATTERIES OR ALKALINE NON-
RECHARGEABLE BATTERIES OR NIMH RECHARGEABLE BATTERIES. THESE BATTERIES
ARE A FLAMMABLE, CORROSIVE AND VAPOUR HAZARD. LITHIUM BATTERIES CONTAIN,
LITHIUM IRON DI-SULPHIDE, AND AN ELECTROLYTE. THE LITHIUM REACTS VIGOROUSLY
WHEN IMMERSED IN WATER. THE ELECTROLYTE IS FLAMMABLE AND HIGHLY CORROSIVE
DO NOT IMMERSE BATTERIES IN WATER, DECONTAMINANT SOLUTION OR OTHER LIQUIDS.
DO NOT CRUSH OR BURN BATTERIES.
DO NOT ATTEMPT TO RECHARGE NON-RECHARGEABLE BATTERIES.
DO NOT STORE BATTERIES AT TEMPERATURES ABOVE 158°F (70°C)
IN THE EVENT THAT A BATTERY WITHIN THE LCD3.3 IS SHORTED OR STARTS TO HEAT UP,
IT SHOULD BE REMOVED FROM THE UNIT AND TAKEN TO A WELL VENTILATED AREA TO
COOL DOWN. ONCE THE TEMPERATURE HAS DROPPED THE BATTERY CAN BE DISPOSED
OF ACCORDING TO LOCAL PROCEDURES FOR HAZARDOUS WASTE. SUITABLE
PROTECTIVE CLOTHING SHOULD BE WORN WHEN HANDLING SUSPECTED BATTERIES. IF
THE SKIN OR EYES COME INTO CONTACT WITH THE ELECTROLYTE, WASH THOROUGHLY
WITH WATER AND SEEK MEDICAL ATTENTION.
WARNING – BURNS
IF THE OUTER COVER OF A BATTERY INSTALLED IN THE BATTERY CASSETTE IS DAMAGED,
A SHORT CIRCUIT CAN OCCUR AND THE BATTERIES MAY OVERHEAT. ALWAYS CHECK THE
OUTER COVER OF EACH BATTERY IS UNDAMAGED BEFORE INSTALLING IN THE BATTERY
CASSETTE
Page 51
Chapter 5
4. Remove the four batteries from the Battery Cassette and discard in accordance with local
regulations.
5. Install four fresh batteries into the Battery Cassette in the correct orientation.
6. Insert the loaded Battery Cassette fully into the battery compartment and close the Battery Cassette
locking cover.
7. A confidence test may now be performed to confirm that the detector is ready for operation.
Page 52
Chapter 5
NOTE, The sieve pack locking cover forms a part of the LCD3.3 unit and is not a part of the sieve pack
assembly. The locking cover must be kept for re-use.
1. Remove the vapour proof packet containing the sieve pack replacement kit from its storage location
and tear open the ‘Used Sieve Pack’ compartment, labelled A, and remove the resealable bag and
cleaning cloth. Open the resealable bag.
2. If the LCD3.3 is operating shut down the detector by closing the Raincap and wait for 3 seconds so
that data can be saved and the unit can shut down correctly.
3. Using the cloth wipe clean the detector and the sieve pack locking cover.
4. Remove the Battery Cassette from the LCD3.3 as detailed in the section concerning battery
replacement.
Page 53
Chapter 5
Figure 49 Sieve Pack Removal
5. Holding the LCD3.3 firmly, rotate the sieve pack locking cover in an anti-clockwise direction until it is
at 90º (at right angles) to the base of the unit and using the locking cover as a grip, pull the sieve
pack assembly free from the LCD3.3 unit.
6. Transfer the used sieve pack from the LCD3.3 into the resealable bag.
7. Hold the used sieve pack through the resealable bag, remove the locking cover from the used sieve
pack by rotating the locking cover until it is aligned with the sieve pack end face, then pull the
locking cover free from the locating tab on the sieve pack. Keep the locking cover for installation
with the new sieve pack.
8. Open the ‘New Sieve Pack’ compartment labelled B of the vapour proof packet. Remove the new
sieve pack from the packaging. Take care not to disturb the plastic cover.
9. Align the locking cover with the base of the new sieve pack and press into place on the locating tab.
Rotate the locking cover in an anti-clockwise direction until it is at 90º (at right angles) to the base of
the new sieve pack to fix the locking cover onto the sieve pack.
10. Taking care not to touch or contaminate the sealing face slide the sieve pack from the plastic cover.
11. Insert the new sieve pack into the sieve pack compartment and push gently into the detector making
sure that it is fully inserted. The sieve pack can only be inserted into the compartment one way,
make sure the sieve pack is the correct way round before insertion.
12. Rotate the locking cover in a clockwise direction until the locking cover is aligned with the base of
the detector and is engaged with the slots on the detector. The locking cover is now locked into
position making the unit fully watertight.
13. Install the Battery Cassette as detailed in the section concerning battery replacement
14. Reset the sieve pack life timer cover as detailed in the section concerning resetting the sieve pack
life timer.
15. A confidence test may now be performed to confirm that the detector is ready for operation.
Page 54
Chapter 5
5.4.6 Sieve Pack Locking Cover Replacement
1. If the LCD3.3 is operating shut down the detector by closing the Raincap and wait for 3 seconds so
that data can be saved and the unit can shut down correctly.
2. Remove the Battery Cassette from the LCD3.3 as detailed in the section concerning battery
replacement.
3. Remove the Sieve Pack from the LCD3.3 and remove the locking cover from the sieve pack as
detailed in the section concerning sieve pack replacement. If serviceable, keep the sieve pack for
reinstallation, if not serviceable replace the sieve pack with a fresh one.
4. Exchange the existing locking cover for a serviceable unit.
5. Install the locking cover to the serviceable Sieve Pack and install the Sieve Pack fully into the Sieve
Pack compartment and close the locking cover as detailed in the section concerning Sieve Pack
replacement.
6. If a new Sieve Pack has been fitted reset the Sieve Pack Life Timer as detailed in the section
concerning resetting the Sieve Pack Life Timer.
7. A confidence test may now be performed to confirm that the detector is ready for operation.
1. If the LCD3.3 is operating shut down the detector by closing the Raincap and wait for 3 seconds so
that data can be saved and the unit can shut down correctly.
2. Remove the Battery Cassette from the LCD3.3 as detailed in the section concerning battery
replacement.
3. To remove the Raincap rotate the Raincap counter-clockwise to the ‘ON’ position. Press down the
Raincap again and rotate anti clockwise until the Raincap comes to a stop and no further rotation is
possible. Lift the Raincap vertically clear of the inlet.
Note. While the Raincap is removed, inspect the inlet for the build up of any dust or dirt. If the inlet is
dusty or dirty gently blow across the inlet to remove any dust or dirt particles. Do not brush or rub
the inlet. If the Raincap itself is dirty it can be rinsed in clean water and dried with a lint free cloth.
Do not use oils, lubricants or solvents.
Page 55
Chapter 5
4. To install the replacement Raincap, position the replacement Raincap onto the Inlet structure with
the Raincap pointer pointing towards the earpiece socket. Make sure that the pins on the inside of
the Raincap are in the locating slots.
5. Press down on top of the Raincap and rotate clockwise until the Raincap is in the ‘ON’ position and
release.
6. Press down again on top of the Raincap and rotate clockwise until the Raincap is in the ‘OFF’
locked position.
7. Install the Battery Cassette as detailed in the section concerning battery replacement.
8. Switch on the detector and wait for completion of the start up sequence.
9. A confidence test may now be performed to confirm that the detector is ready for operation.
1. If the LCD3.3 is operating shut down the detector by closing the Raincap and wait for 3 seconds so
that data can be saved and the unit can shut down correctly.
2. Remove the Battery Cassette from the LCD3.3 as detailed in the section concerning battery
replacement.
3. Cut the existing lanyard and remove the existing earpiece socket protective cap.
4. Dispose of the existing earpiece socket protective cap and lanyard.
5. Tie the lanyard of the replacement earpiece socket protective cap to the securing point on the body
of the detector.
6. Install the earpiece socket protective cap firmly to the earpiece socket.
7. Install the Battery Cassette as detailed in the section concerning battery replacement.
8. Switch on the detector and wait for completion of the start up sequence.
9. A confidence test may now be performed to confirm that the detector is ready for operation.
Page 56
Chapter 5
1. If the LCD3.3 is operating shut down the detector by closing the Raincap and wait for 3 seconds so
that data can be saved and the unit can shut down correctly.
2. Remove the Battery Cassette from the LCD3.3 as detailed in the section concerning battery
replacement.
3. Using a suitable tool in the slots of the breather rotate the breather anti clockwise until loose and
remove.
4. Dispose of the existing breather as non hazardous waste.
5. Install the replacement breather and tighten in a clockwise direction by hand until finger tight. Do not
use Loctite or any other adhesives when installing the replacement breather.
6. Using a suitable tool in the slots tighten the breather a further ¼ turn clockwise.
7. Install the Battery Cassette as detailed in the section concerning battery replacement.
8. Switch on the detector and wait for completion of the start up sequence.
9. A confidence test may now be performed to confirm that the detector is ready for operation.
Page 57
Chapter 5
Low Battery Battery power low. Detector Replace Batteries. See corrective
message. operational. Battery replacement maintenance procedures.
required.
Low Sieve Pack Sieve Pack remaining life low. Replace Sieve Pack. See corrective
message Detector operational. Sieve Pack maintenance procedures.
replacement required.
Sieve Pack Sieve Pack exhausted. Detector Replace Sieve Pack. See corrective
exhausted not operational. Sieve Pack maintenance procedures.
message replacement required.
Failed Confidence Detector does not give alarm in 1. Replace Confidence Tester and repeat
Test confidence test mode. confidence test.
Page 58
Chapter 6
CHAPTER 6 SPARES INFORMATION
6.1 GENERAL
TABLE 12 details the recommended spare parts, consumables and optional ancillaries for LCD3.3.
Page 59
Chapter 6
Page 60