Professional Documents
Culture Documents
Up 890 Installation
Up 890 Installation
Up 890 Installation
SONOLINE Elegra
Ultrasound Imaging System
48 45 801 REV 00
Wiedergabe sowie Vervielfältigung dieser Unterlage, Verwertung und Mitteilung Copyright 2000 Siemens Medical Systems, Inc.
ihres Inhalts nicht gestattet, soweit nicht ausdrücklich zugestanden. Zuwieder- Proprietary data, company confidential. All rights reserved.
handlungen verpflichten zu Schadenersatz. Alle Rechte für den Fall der Patent- Confié à titre de secret d´entreprise. Tous droits réservés.
erteilung oder GM-Eintragung vorbehalten. Confiado como secreto industrial. Nos reservamos todos los derechos.
Go to Main TOC
SONOLINE Elegra Ultrasound Imaging System - B/W Video Printer UP-890 Installation
Table of Contents
SONOLINE Elegra Ultrasound Imaging System - B/W Video Printer UP-890 Installation
1.1 Introduction
NOTICE If you are installing this device onto a 15 A SONOLINE Elegra ultrasound imaging system, you
must first upgrade the system to a 20 A system.
This procedure describes the installation of the black-and-white (B/W) video printer into the
upper OEM bay of the SONOLINE Elegra system.
1.1.3 Preparation
! WARNING To avoid damage to the system and to the accessories caused by short circuit or injuries to
yourself caused by electric shock during installation, ensure that you have disconnected
the system power cable from the external main power source before you begin with the
installation.
2. Move the system if necessary to provide enough space for easy access during the
installation.
3. Lock the wheels to prevent the system from rolling during the installation.
SONOLINE Elegra Ultrasound Imaging System - B/W Video Printer UP-890 Installation
AC IN
Figure 1 Front (on the left) and rear (on the right) panel of the B/W printer UP-890.
After setting the switches on the front panel set the switches on the rear panel of the printer:
6. Set the GAMMA switch to II.
7. Set the PAPER TYPE switch to IV.
8. Set the DIP switches 2, 3, 4, 5, 6, 7, 10, 11 and 12 to ON, all others to OFF.
NOTICE If there is a VCR installed in the upper OEM bay you must unbundle and disconnect those OEM
cables in order to pull the tray out to the fully extended position.
Nylon Strap
Strap Buckle
OEM
Mounting
Pads Ground
Wire
Mounting Bracket
SONOLINE Elegra Ultrasound Imaging System - B/W Video Printer UP-890 Installation
6. Peel off the backing and position the OEM mounting pads on the mounting bracket so
that the hole in each mounting pad is centered around one of the holes in the mounting
bracket.
7. Place the printer on the tray with the printer feet positioned in the mounting pad holes.
8. Secure the printer to the OEM tray using one nylon strap in the front slots of the OEM
tray, as shown in Figure 2. Loop the end of the strap through the slots in the sides of
the tray and tighten so that the printer is held securely to the tray. Position the buckle
on the right side of the printer.
NOTICE If a VCR or medium drawer is also installed in the upper OEM bay, use one nylon strap to fasten
both the printer and the VCR or drawer to the OEM tray.
SONOLINE Elegra Ultrasound Imaging System - B/W Video Printer UP-890 Installation
1 0 -3
2 0 -4
3 +1 -3
4 +1 -4
3. Use these values as a starting point. Adjust the CONTR and BRIGHT knobs as nec-
essary to suit the customer’s preference. Refer to the operating instructions of the
printer to optimize the print settings if necessary.