GENERAL TECHNICAL MANUAL

MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
1 / 40

To copy this manual, even if partially, is forbidden.

Geenneerraall Teecchhnniiccaall Maannuuaall


ffo
orr

B,, S &
&N C
Clla
assss

Auuttooccllaavveess

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
2 / 40

To copy this manual, even if partially, is forbidden.

INDEX

CHAP SUBJECT

DISPLAY & CONTROL BOARD


HOW TO USE THE KEYBOARD
1
ICONS
PROGRAMMING MENU & “EMERGENCY CYCLE”

DEVELOPMENT OF THE CYCLE


SWITCHING ON
PRE-HEATING
2
START OF THE STERILIZATION CYCLE
AUTOMATIC PROCEEDING OF THE CYCLE
GRAPHICS OF THE CYCLES

3 MESSAGES ON THE DISPLAY


MEANING OF THE MESSAGES ON THE DISPLAY

4 LIST OF THE ALARMS


UNDERSTANDING AND SOLUTION OF THE ALARMS

SAFETY SYSTEMS
5 LIST OF THE SAFETY SYSTEMS
INTERVENTION OF THE SAFETY SYSTEMS

VARIOUS ANOMALIES
LEAKAGE OF WATER FROM THE JOINT/BACK HOSE
6 THE STERILISED MATERIAL COMES OUT VERY WET
THE GEAR REDUCTION UNIT OF THE DOOR DOES NOT WORK
THE TANK DOES NOT DRAIN OFF THE DIRTY WATER

7 SETTING OF TEMPERATURE PROBES


SETTING PROCEDURE

MAINTENANCE
CLEANING OF THE FILTERS
CLEANING OF THE FRESH WATER TANK
CONTROL OF THE CLEAN WATER LEVEL SENSOR
CONTROL OF THE DIRTY WATER LEVEL SENSOR
REGULATION OF THE DOOR SEAL
8
CONTROL OF THE DOOR MICROSWITCHES
CONTROL OF SOLENOID VALVES
CONTROL OF THE WATER INJECTION PUMP
CONTROL OF THE WATER CHARGE PUMP
FUSES
BLUE GASKET CHANGE INSTRUCTIONS

9 WATER DIAGRAM

10 WIRING DIAGRAMS

11 SPARE PARTS

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
3 / 40

To copy this manual, even if partially, is forbidden.

1. DISPLAY & CONTROL BOARD


HOW TO USE THE KEYBOARD:

STERILIZATION STERILIZATION
CHAMBER CHAMBER
TEMPERATURE PRESSURE
085.9° 0.12bar

REMAINING TIME 121°C solids wrapped SELECTED CYCLE


OF THE CYCLE
(FROM THE 22:37min c/n°:0142
BEGINNING OF THE CYCLES COUNTER
STERILIZATION (it can be read by
PRE-HEATING
PHASE INCLUDING keeping pressed the
THE DRYING TIME) button “DOOR”)

ACTIVE PHASE OF ICONS


THE RUNNING (see following par.)
CYCLE

PUMP
DOOR PRESS this button
DOOR SELECT PUMP
Press this button to to load the water in
open the door at the reservoir. If it is
the end of the pressed for 5 sec. it
cycle. makes the “counter
If you keep this resins reset”.
pressed for 10”
while keeping the
door closed, you SELECT
activate the closure This button is to
START/STOP
of the door without select a different
starting a new cycle.
cycle.
START / STOP
Press to start the cycle: if you press for
more than 1 second, it interrupts the cycle.

NOTE: when the cycle is NOT in progress, on the second line of the display it is possible to
read the length of the sterilization time (str) and drying time (dry), respectively.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
4 / 40

To copy this manual, even if partially, is forbidden.

ICONS:

This icon appears immediately after the start of the cycle and it indicates that the door is
locked.

This icon indicates that the cycle is running: this can be seen by the rotating movement of
the arrows.

This icon warns that the paper is missing in the printer or, that the door of the printer is not
correctly closed. The machine, however, will continue to cycle normally.

B This icon indicates the type of cycle selected (B, S or N).

The autoclaves are equipped with software which is able to communicate and to write data
onto a USB key. Some sterilisers have a USB connector already on board the machine; others
require the STS-Datalogger system which must be connected to the RS232 serial port: both
systems are OPTIONAL.
When the USB key is inserted in the connector (or on the Datalogger), the USB icon appears
on the display. With the USB, on board the machine or with the Datalogger system, it is
always necessary to read the specific enclosed manual carefully to prevent data saving
errors.
DATA SAVING BEGINS AND ENDS ONLY AND EXCLUSIVELY IF THE USB KEY IS ALREADY
INSERTED IN THE CONNECTOR BEFORE THE CYCLE IS STARTED.
This icon warns that the fresh water has reached the minimum level and cannot start a new
cycle. Recharge the machine with water to carry on with a new cycle.
Only the VACUUM TEST can be started even if you see this icon.

The fresh water tank is full. If you press again PUMP the display will warn, by displaying
“FULL WATER RESERVOIR”, that it is not possible to charge with more water.

When you see this icon you cannot start any cycle: you must empty the waste water tank.
Let the water drain completely before closing the tap.

PROGRAMMING MENU & “EMERGENCY CYCLE”:

You reach the programming menu by switching the machine on by the ON/OFF switch and by keeping
pressed at the same time the SELECT button until you see the pre-set language. When the display shows

ENGLISH
It is possible to select the language by using the DOOR and PUMP buttons. By pressing the START/STOP
button you go to …

HOUR: hh mm ss
In this page you set the time. By pressing the DOOR button you increase the value, by the PUMP button
you decrease it and by the SELECT button you move the indicator under the value to modify. By the
START/STOP button you go to…

DATE: dd mm yy
In this page you set the date. By pressing the DOOR button you increase the value, by the PUMP button
you decrease it and by the button SELECT you move the indicator under the value to modify. By the
START/STOP button you go to…

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
5 / 40

To copy this manual, even if partially, is forbidden.

WATER CHARGE BY PUMP


This page is used to select the type of water charge on the machine. If the user uses an external charge
(demineraliser), by pressing the DOOR and PUMP buttons the chosen value can be set and by keeping
pressed the SELECT button for 5 seconds (until the beep confirmation sound), it records the operation.
By pressing the START/STOP button you go to…

REPRINTING OF THE CYCLES


This function is most important as it allows the operator to re-print the last 10 cycles made by the
autoclave, making it very simple to understand the problems which may have caused the fault and to set
the alarm. It is possible to re-print the cycles by keeping pressed the SELECT button for 5 seconds. By
pressing the START/STOP button you go to …

EMERGENCY ON/OFF
The setting “Emergency ON” eliminates the phases of initial vacuum and it allows the unit to function in a
thermodynamic way. This function becomes very important when the machine is unable to reach the set
vacuum, thus causing an alarm. By the function “EMERGENCY ON” you by-pass the vacuum phase it is
possible to look for the fault by using the pressure of the machine (see relative graphic).
You activate the “EMERGENCY” cycle by selecting the chosen feature by pressing the DOOR and PUMP
buttons and by recording the data by pressing SELECT until the confirmation sound is heard (by
activating the “EMERGENCY” cycle, the selection of all the other types of cycle is inhibited until you set
“EMERGENCY OFF”).
By pressing the START/STOP button you go to…

TO ENTER AND LEAVE LANGUAGE AND USER MENU

By pressing and holding the SELECT key when switching on the autoclave (Green switch) you will enter
the user and language setting, by keeping START/STOP pressed for a long time you can return to the
main menu. You can return to the main menu at any time using this procedure.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
6 / 40

To copy this manual, even if partially, is forbidden.

2. DEVELOPMENT OF THE CYCLE


SWITCHING ON:

When the machine is switched on by the ON/OFF switch, it asks to open the door to allow the
alignment of the sensor with the room pressure: on the display in fact, you will read “OPEN
THE DOOR” and no button apart from DOOR is activated. A double beep informs that the
operation is complete and you can then use the other buttons.

PRE-HEATING:

When the machine is switched on it automatically starts a pre-heating phase of the sterilization
chamber by the heating element: the external probe (TRS) will stop the heating at 80°C.

It is possible that some versions of the software do not include the pre-heating
or that they reach a different temperature (usually 80°C).

IT IS ALWAYS POSSIBLE TO START A STERILIZATION CYCLE, even though the external


temperature has not yet reached the pre-set parameter.

STARTING OF THE STERILIZATION CYCLE:

The commencement of the cycle by pressing the START/STOP button is accepted only if there
are the ideal conditions, made by:
1. sufficient water level:
In case of lack of water, on the display you will see the corresponding icon.
2. door closed:
The door is closed when the DOOR button is activated.
3. dirty water level icon not showing:
In this case, on the display you will see the corresponding icon.

ATTENTION:

NEVER START THE CYCLE BY KEEPING THE MICROSWITCH PRESSED DOWN BY HAND
WITH THE DOOR OPEN

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
7 / 40

To copy this manual, even if partially, is forbidden.

AUTOMATIC PROCEEDING OF THE CYCLE:


(see graphics following the description)

After having started the cycle by pressing the START/STOP button, the machine commences
automatically until the end of the process. The following steps are indicated below:

The described
phase is on
the autoclave Description of the operations made by the machine
type…
B S N
1 √ - - Reaching the minimum starting resistor temperature (TRS): ~80°C.
First vacuum.
During the vacuum phase the steam generator (TVP) & the heater band
2 √ √ -
(TRS) reach a correct temperature of work. The vacuum pump
(PM3+EV5, 3 ways valve) removes the air from the chamber.
First pressure.
Injection of steam in the chamber with the injection pump (PM1+EV3)
3 √ √ √ until it reaches the pre-set pressure. For the class N the injection of
steam happens only when the steam generator has reached a correct
temperature of work.
First thermodynamic discharge.
The machine starts a steam discharge phase by means of the EV2: the
4 √ √ √ heater band (TRS) is interrupted while the vaporizer (TVP) goes back to
a correct temperature of work.
5 √ - - Second vacuum: as in point 2.
6 √ √ √ Second pressure: as in point 3.
7 √ √ √ Second thermodynamic discharge: as in point 4.
8 √ - - Third vacuum: as in point 2.
9 - - √ Third thermodynamic discharge: as in point 4.
Final rise.
After having reached the value of the last thermodynamic discharge (or
10 √ √ √ of the last foreseen phase of vacuum) value the EV2 closes and the
machine restarts the heating and the steam injection until it reaches the
parameters of the selected cycle.
Sterilization.
The temperature and pressure parameters are kept in the set range by
11 √ √ √ the help of the heating element (in the first part of the pressure the least
values of the select cycle are visible).
Final discharge.
The final discharge is made by the EV1. The EV is open (charged) until
12 √ √ √ the vacuum pump starts working (B and S) or until the end of the cycle
(N). Sometimes is opened (charged) also the EV2.
Drying vacuum.
13 √ √ -
The vacuum pump aspirates the steam left in the chamber.
END CYCLE.
14 √ √ √ It is possible to open the door by the DOOR button.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
8 / 40

To copy this manual, even if partially, is forbidden.

GRAPHICS OF THE CYCLES:

Example: 134°C cycle (fractionated vacuum - B type)

2,5
2,06
2
1,5
Pressure (Bar)

1
0,50 0,50
0,5
0 0 0

-0,5
-0,80 -0,75 -0,75 dry
-1
Time

Example: 134°C cycle (single pre-vacuum - S type)

2,5
2,06
2
1,5
Pressure (Bar)

1,00 1,00
1
0,5
0 0 0,20 0,20 0

-0,5
- 0,70 dry
-1
Time

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
9 / 40

To copy this manual, even if partially, is forbidden.

Example: 134°C cycle (thermodynamics vacuum - N type)

2,5
2,06
2

1,5
Pressure (Bar)

1,00 1,00 1,00


1

0,5
0,20 0,20 0,20
0 0
dry

-0,5
Time

EMERGENCY cycle (This is possible only on 134°C)

2,5
-DISARMED FAN
2,06
2

1,5
Pressure (Bar)

1,10 1,10
1

0,5 0,50 0,50


dry
0 0

-0,5
Time

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
10 / 40

To copy this manual, even if partially, is forbidden.

3. MESSAGES ON THE DISPLAY


MEANING OF THE MESSAGES ON THE DISPLAY:
Since the machine is switched on to the end of the cycle, on the display there are messages
which help to simplify the use of the steriliser. Each message implies a precise operation:

MESSAGE CAUSE SOLUTION


When the machine is switched on it asks to
open the door to allow the machine to make Open the door and wait for the
OPEN THE DOOR the pressure check-up. It is not possible to double beep to allow the
make any operation apart from opening the automatic pressure check-up.
door by the DOOR button.
It is necessary to keep the door
pressed in order to activate the
door micro switch.
OPEN DOOR A cycle was started with the door open. NEVER START THE CYCLE
BY PRESSING THE MICRO
SWITCH by HAND TO
SIMULATE DOOR CLOSED
1.Press the DOOR button to
Even though the door is perfectly closed, the
reopen the door closure
machine does not “read” the door as “non
DOOR NOT mechanism.
blocked” for mechanic causes (if this error
LOCKED 2.Open and close the door.
code appears again you have to check the
3.Start a cycle by pressing
internal micro switches – see chap. 8).
START.
A cycle was started when the minimum fresh
Fill tank with purified water as
TO LOAD WATER water led was blinking on the display (see
described in the user’s manual.
chap. 1).
A cycle was started when the maximum waste
TO DRAIN Discharge the waste water as
water led was blinking on the display (see
WATER described in the user’s manual.
chap. 1).
The operator is trying to fill the fresh water Disconnect the water filling tube.
FULL WATER
tank when on the display the icon shows the It is then possible to start a new
RESERVOIR
tank as full. (see chap. 1). cycle.
1.Close the door and start a new
At the end of the cycle the door locking system
cycle. After a few seconds,
does not open completely (if the error
DOOR STILL interrupt by START/STOP.
appears again you have to check the
LOCKED 2.Reset the alarm (chap. 4) and
closing door motor and the internal micro
open the door by DOOR to
switches).
check.
The operator is trying to start the VACUUM
Wait, keeping the door open, for
CHAMBER TEMP. TEST cycle with the temperature inside the
the temperature to decrease
>40° chamber above 40°C. Wait for the temperature
below 40°C on the display.
to decrease and start the test cycle.

The cycle counter of the autoclave invites you IMPORTANT


to check the quality of the resins of the water To define the quality of the
REPLACE RESINS depuration system. If everything is OK, keep water, follow carefully the
pressed the button PUMP for 5 sec. (until the instructions of the supplier of
beep) to return to zero the resins-counter. the type in use.

The machine has run an excessively large Contact your dealer as soon as
SERVICE number of cycles and needs periodic possible and request the services
overhauling by a technician. of a technician.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
11 / 40

To copy this manual, even if partially, is forbidden.

4. LIST OF THE ALARMS


UNDERSTANDING AND SOLUTIONS OF THE ALARMS:

ATTENTION:
SOME OF THE LISTED ALARMS MAY NOT APPEAR IN SOME MODELS
AUTOCLAVE

RESET ALARM =
Keep pressed together START/STOP + SELECT buttons for 5 sec.

In case of alarm, the material inside the chamber should be


considered NON STERILE.

Solution of the problem:


Code and
How it happens RESET ALARM =
meaning
START/STOP + SELECT for 5 sec.
Reset the alarm and repeat the
AL0001 cycle: if the problem continues
This happens if the START/STOP button is
Mistakenly make sure that the START/STOP
pressed for more than 1 sec.
interrupted cycle button does not remain
continuously pressed.
AL0002 Power This is caused by a voltage fall of more than Reset the alarm then repeat the
failure 10% or to its complete interruption. cycle.
Reset the alarm and repeat the
AL0003 This happens if one of the micro switches of the
cycle: if the problem continues
Open door during door control detects “open door during the
check if the micro switches work
the cycle cycle”.
as described in chap.8
Reset the alarm, re-set hour and
date: if the problem continues
This intervenes after 3 seconds from switching
AL0004 you have to replace the battery
on if the values of the clock (date and hour)
Timer out of order on the display board. Before use
are out of range: ex. hour 85h 24m 36s.
leave the autoclave on for at
least one hour.

The alarm appears if the autoclave does not 1. Reset the alarm, clean the
AL0011
reach the 1st vacuum. gasket and adjust the door
1st failed vacuum
It happens after 15 min. (see chap.8) making sure it
The alarm appears if the autoclave does not makes a good contact with
AL0012 the seal.
reach the 2nd vacuum.
2nd failed vacuum 2. Check inside the chamber for
It happens after 15 min.
material or envelopes which
The alarm appears if the autoclave does not may clog the hoses.
AL0013 rd
reach the 3 vacuum. 3. Check if the vacuum pump
3nd failed vacuum
It happens after 15 min. works correctly and if the
AL0014 cooling fan works well.
The alarm appears if the autoclave does not
4th failed vacuum 4. Activate the “Emergency
reach the 4th vacuum.
(USUALLY IT IS cycle” and look for pressure
It happens after 15 min.
NOT USED) leakage (see chap. 1).
This happens if during the drying phase the
AL0015
autoclave does not reach the minimum set Reset the alarm and check the
Failed drying
vacuum. When this alarm appears the machine vacuum pump.
vacuum
has already made the sterilisation.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
12 / 40

To copy this manual, even if partially, is forbidden.

AL0021 The machine does not 1st set Reset the alarm and refill the
reach the
1st failed pressure pressure. clean water tank until it reaches
rise It appears after 25 min. the maximum level, then repeat
the cycle. If the problem
nd continues, make sure there are
AL0022 The machine does not reach the 2 set
no pressure leakages from:
2nd failed pressure pressure.
rise It happens after 15 min.
1.Door seal.
2.Solenoids valves.
AL0023 3.Injection pump (PM1) and
3rd failed pressure rd relative solenoid valve (EV3).
The machine does not reach the 3 set
rise
pressure.
(USUALLY IT IS It is a good thing to check the
It happens after 15 min.
USED ONLY ON mains voltage because should it
CLASS N) be very low (~200V) the cycle
time would be much longer. This
The machine does not reach the working can cause many possible alarm
AL0024
pressure. codes.
Failed final raise
It happens after 35 min. See chap. 8

After having reached the 1st pressure the unit 1.Reset the alarm, remove the
AL0031
cannot discharge the pressure. tray holder and make sure
1st failed discharge
It happens after 4 min. that inside the chamber the
hoses are not clogged. Clean
After having reached the 2nd pressure the unit inside the sterilisation
AL0032
cannot discharge the pressure. chamber and repeat the
2nd failed discharge
It happens after 4 min. cycle.
AL0033 2.Check the EV2, through which
3rd failed discharge After having reached the 3rd pressure the unit the machine makes the
(USUALLY IT IS cannot discharge the pressure. thermodynamic discharges.
USED ONLY IN It happens after 4 min. 3.Make sure the power board
CLASS N) activates the EV2.
4.Check the discharge circuit,
AL0034 In the drying phase the unit does not discharge
from the upper and lower EVS
Failed final the pressure.
onwards.
discharge It happens after 5 min.

Reset the alarm, switch the


AL0100
This alarm comes from the auto-diagnosis of autoclave off and on: if the
T1 probe coding
the power board. problem continues you have to
error (CN8)
replace the display board.
AL0101 Make sure the temperature
The probe T1 is read open.
OPEN T1 (CN8) probe works correctly: if the
AL0102 probe is OK, replace the display
The probe T1 is detected in short circuit. board.
C.C. T1 (CN8)

AL0110
The probe T1 has increased by more than 4°C
High temperature Reset the alarm and make sure
the temperature of the selected cycle.
probe T1 (CN8) there are no pressure leakages
from:
AL0111 1.Door seal.
Low temperature During the sterilization phase the probe T1 is 2.Solenoids valves (EV1, EV2,
probe T1 (CN8) below the allowed limits. EV5 and EV7).
during sterilization

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
13 / 40

To copy this manual, even if partially, is forbidden.

Reset the alarm, switch


AL0200
The alarm comes from the auto-diagnosis of autoclave off and on: if the
T2 probe coding
the power board. problem continues you have to
error (CN9)
replace the power board.
AL0201 Make sure the temperature
The probe T2 is read open.
OPEN T2 (CN9) probe works correctly: if the
AL0202 probe is OK, replace the display
The probe T2 is detected in short-circuit.
C.C. T2 (CN9) board.
AL0210
The probe T2 has increased by more than 4°C Reset the alarm and make sure
High temperature
the temperature of the selected cycle. there are no pressure leakages
probe T2 (CN9)
from:
AL0211
1.Door seal.
Low temperature During the sterilization phase T2 has decreased
2.Solenoids valves (EV1, EV2,
probe T2 (CN9) below the allowed limits.
EV5 and EV7).
during sterilization

Reset the alarm, switch


AL0300
The alarm comes from the auto-diagnosis of autoclave off and on: if the
P probe coding error
the power board. problem continues you have to
(CN10)
replace the power board.
AL0301
The probe P is read open. Make sure the pressure
OPEN P (CN10)
electronic transducer works
correctly: if the sensor is OK,
AL0302
The probe P is read in short-circuit. replace the display board.
C.C. P (CN10)

Check the upper solenoid valve


AL0310 EV2, through which the
High pressure During the sterilization phase the probe P autoclave makes the
during the (CN10) has exceeded the allowed limits. overpressure discharges.
sterilization Make sure the power board
activates the EV2.
Reset the alarm and make sure
there are no pressure leakages
AL0311 from:
During the sterilization phase the probe P
Low pressure during 1.Door seal.
(CN10) has decreased below the allowed limits.
the sterilization 2.Solenoids valves (EV1, EV2,
EV5 and EV7).
3.Various attachments.

Reset the alarm, switch


AL0400
The alarm comes from the auto-diagnosis of autoclave off and on: if the
TVP probe coding
the power board. problem continues you have to
error (CN11)
replace the display board.
AL0401 Make sure the temperature
The probe TVP is read open.
OPEN TVP (CN11) probe works correctly: if the
AL0402 probe is OK, replace the display
The probe TVP is detected in short-circuit.
C.C. TVP (CN11) board.
Make sure the manual
AL0404 thermostat on the vaporizer is
The probe TVP does not reach the working
Low temperature activated, then make sure the
temperature.
probe TVP (CN11) power board activates the
heating element of the vaporizer.
Make sure the power board stops
AL0405
The probe TVP has exceeded the max working the heating element of the
High temperature
limit. vaporizer (TVP) above the
probe TVP (CN11)
180°C.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
14 / 40

To copy this manual, even if partially, is forbidden.

Reset the alarm, switch


AL0500
The alarm comes from the auto-diagnosis of autoclave off and on: if the
TRS probe coding
the power board. problem continues you have to
error (CN12)
replace the display board.
AL0501 Make sure the temperature
The probe TRS is read open.
OPEN TRS (CN12) probe works correctly: if the
AL0502 probe is OK, replace the display
The probe TRS is detected in short circuit.
C.C. TRS (CN12) board.
Make sure the manual
AL0504 thermostat on the heating
The probe TRS does not reach the minimum
Low temperature element (TRS) is activated, then
temperature.
TRS (CN12) make sure the power board
activates the heating element.
Make sure there are no pressure
leakages from:
1.Door seal.
2.Solenoids valves.
Moreover you have to make sure
AL0505 The probe TRS has exceeded the maximum
that the water level sensor and
High temperature limit.
the injection pump (PM1) work
probe TRS (CN12)
correctly.
Make sure that heater band is
not continuously under tension
by the power board.
See chap. 8

AL0600
Excessive loss
During the first 5 minutes of the VACUUM TEST
during the Make sure that no pressure
cycle (Stabilization) there has been an
STABILIZATION leakages take place from:
excessive loss of pressure.
phase of “VACUUM 1.Door seal.
TEST” 2.Solenoids valves.
AL0601 To facilitate the search for small
Excessive loss pressure leakages it is possible
During the following 10 minutes of the VACUUM
during the to activate the “Emergency
TEST cycle (UPKEEPING) there has been an
“VACUUM TEST” cycle” (see ch. 1).
excessive loss of pressure.
during the
HOLDING phase

During the sterilization phase the two internal Make sure that no pressure
AL0700 sensory detected discordant temperatures. It leakages are to be found from:
Comparison T1/T2 intervenes if the difference between T1 & T2 is 1.Door seal.
greater of 2.0 °C. 2.Solenoids valves.

ATTENTION:

The manufacturer declines all responsibilities for calibrations or modifications done


on mechanic, pneumatic and electronic parts, carried out by non-authorized
technicians.

The use of NON ORIGINAL spare parts immediately excludes the warranty of the
machine and FREES the manufacturer from any responsibility deriving from it.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
15 / 40

To copy this manual, even if partially, is forbidden.

5. SAFETY SYSTEMS
LIST OF SAFETY SYSTEMS:

The safety systems on the models class B, S and N are of two kinds:

Electric or electronic Pressure transducer


Temperature probes
Solenoids valves
Safety thermostat (MANUAL RESET THERMOSTAT)

Mechanic Safety valve


Door micro-switches

INTERVENTION OF THE SAFETY SYSTEMS:

PRESSURE TRANSDUCER: Intervenes as described in the alarms chart.

TEMPERATURE PROBES: Intervene as described in the alarms chart.

SOLENOIDS VALVES: During the sterilization phase intervene upon command of


the transducer when pressure exceeds the required value.
The EV2 opens for 1 sec. If pressure remains above the
correct value EV2 opens for 3 sec.
If pressure does not reduce, the relative alarm starts.

SAFETY THERMOSTATS: Thermostats intervene if the resistances surfaces (band and


vaporizer) reach 220°C +/- 5° .
The reset of the thermostats are MANUALLY regulated.

SAFETY VALVE: The valve intervenes if the electric/electronic control system


does not work properly. The intervention value is indicated
on the valve itself. In normal conditions the safety valve
MUST NOT INTERVENE.

DOOR MICROSWITCHES: See chap. 8.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
16 / 40

To copy this manual, even if partially, is forbidden.

6. VARIOUS ANOMALIES
LEAKAGE OF WATER FROM THE JOINT/BACK HOSE:

If during normal operation water comes out from the attachment or from the back pipe, it
means that the maximum waste water level sensor does not work properly. It is necessary to
intervene by checking if the sensor works perfectly as described in chap. 8.
If steam comes out from the same attachment it is necessary to use the special silicon pipe
supplied with the accessories kit and put its end into a re-condensation container.

Water can flow out from these attachments.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
17 / 40

To copy this manual, even if partially, is forbidden.

THE STERILIZED MATERIAL COMES OUT VERY WET:

If at the end of the cycle the material comes out too wet, it is possible to intervene in order to
improve the quality of the final drying process.

Suggestion for the technicians:

1. Check the set-up of the autoclave: check the horizontal level of support.
2. To clean the fan and the condenser radiator described as follows:

ATTENTION:
After many cycles it is necessary to perform the cooling system cleaning: if the radiator strips
are too dusty, this could cause a loss of efficiency. The vacuum pump alone would not carry
out an effective drying procedure.

Suggestions for the user:

1. Slightly turn the tray holder: this operation will enable steam condensation discharge from
the trays (see picture).
2. Use the special tray cloth: if you place them on the bottom of the trays they will absorb the
moisture and dry the instruments more efficiently.
3. When the cycle has ended, open the door and leave it slightly open for some minutes: this
operation will enable the steam left inside the chamber to fully evaporate. The heat of the
machine will help the instruments dry better.

Tray holder rotation facilitates water discharge and facilitates envelopes drying.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
18 / 40

To copy this manual, even if partially, is forbidden.

THE GEAR REDUCTION UNIT OF THE DOOR DOES NOT WORK:

If the door not open when the cycle is ended, even though the DOOR button has been pushed
many times.
Wait a few minutes and repeat opening operation.
If the problem persists, remove the main cover of the machine and take off the motor reducer
from its support by unscrewing the screws indicated in the picture.

Motor-screw

Now test the motor reducer by applying direct 12V DC voltage (using the battery included in
the Service Kit code SERKITKRB1).

Battery 12V

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
19 / 40

To copy this manual, even if partially, is forbidden.

THE TANK DOES NOT DRAIN OFF THE DIRTY WATER:

This problem usually occurs after many working cycles of the autoclave: it is usually connected
with a clogged drainage tap or with broken internal seals.
In both cases the following procedure must take place.

1. Remove the autoclave cover.


2. Pull off the internal pipe going from the tank to the discharge tap, paying attention to
water leaking.
3. Remove the cir-clip from the internal part of the discharge tap.
4. Pull out the tap by pulling it from the front part of the autoclave.

Cir-clip O’ring to be checked (no. 3 identical O’rings) H2O Drain pipe

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
20 / 40

To copy this manual, even if partially, is forbidden.

7. SETTING OF TEMPERATURE PROBES


SETTING PROCEDURE:

The temperature probes calibration should always be carried out while the autoclave is
operating, during the STERILIZATION phase and only by means of a calibrated control
instrument to make a comparison.
You can observe the probes set-up at any moment by keeping the SELECT button pushed
and pushing the DOOR button three times. When, in the fourth line of screen, you will read
value T01… release SELECT button.
It is possible to modify the probe default T01... (1st INTERNAL SENSOR) with DOOR and PUMP
buttons: after the probe read-out has been aligned with the comparison instrument, store the
parameter by pushing PUMP and DOOR buttons at the same time until you hear the confirm
sound. Move on to T02… (2nd INTERNAL SENSOR) with SELECT button. The variation and
storage modes are exactly the same as those indicated above.
It is possible to modify also probes P (PRESSURE), TVP (PROBE VAPORIZER TEMPERATURE)
and TRS (SENSOR RESISTANCE TEMPERATURE) but we recommend not to modify the P or
TVP and TRS settings.
Exit probes set-up by keeping SELECT button pushed.

Index:

SELECT + 3 DOOR = set-up access


DOOR = increases default value
PUMP = decreases default value
DOOR + PUMP until sound = stores variations
SELECT = moves on to the next value
SELECT keep pressed = exits set-up

Probes identification:

T01 = 1st internal sensor (visualized on the display)


T02 = 2nd internal sensor
P = pressure transducer
TVP = probe vaporizer temperature
TRS = sensor resistance temperature

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
21 / 40

To copy this manual, even if partially, is forbidden.

8. MAINTENANCE
CLEANING OF THE FILTERS:

STAINLESS STEEL FILTER (INSIDE THE CHAMBER):

Support for filter (with filter inside) Stainless steel filter

Before unscrewing the filter support, check that there aren’t impurities around the filter. Using
a 13 mm wrench, unscrewing the support of the filter and REMOVE the filter from its seat.
Replace the stainless steel filter with a new one.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
22 / 40

To copy this manual, even if partially, is forbidden.

CLEANING OF THE FRESH WATER TANK:

OPEN TANK

Use a clean cloth to clean the inside of the tank.


Replace the original steel cover seal.

CLOSE TANK

Screw the 4 screws which secure the tank cover.


Do not over-tighten the screws.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
23 / 40

To copy this manual, even if partially, is forbidden.

CONTROL OF THE CLEAN WATER LEVEL SENSOR:

Sensor at MAXIMUM LEVEL Sensor at MIMIMUM LEVEL

The MAX LEV is detected with a 5 second delay to allow a complete filling of the water tank.
During the filling from water mains the level sensor is detected immediately.

The MIN LEV is detected immediately.

CONTROL OF THE DIRTY WATER LEVEL SENSOR:

Sensor on MAXIMUM LEVEL

Message (and relative icon) of dirty water MAX LEV is immediate. The probe is supplied as per
the photo.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
24 / 40

To copy this manual, even if partially, is forbidden.

REGULATION OF THE DOOR SEAL:

ATTENTION:

If the door is not correctly regulated it might create different problems:


• If it is too tight, at the end of the cycle the motor of the door would have problems in
opening the door. You have to wait 10 minutes and press the DOOR button repeatedly
to open the door.
• If it is too loose, there could be either steam leakages during the cycle (ex. AL0024) or
an alarm connected to a vacuum phase (ex. AL0011).

In case of steam leakages from the upper part see Picture 1


In case of steam leakages from the lower part see Picture 2

PICTURE 1 PICTURE 2
Upper hinge Lower hinge

To uniformly adjust both hinges turn one notch only and then repeat the cycle.
----------------------------------------------------------------------------

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
25 / 40

To copy this manual, even if partially, is forbidden.

CONTROL OF THE DOOR MICROSWITCHES:

SQ6:

Adjust the depth of the microswitch SQ6


so that the internal spring has still
~1.5mm free, when the door is closed.

SQ5:

Adjust the microswitch SQ5 so that the rod


which operates the microswitches is at
~1mm from the end of the stroke and that it
is completely free, when the door is open.

SQ4:

Adjust the microswitch SQ4 so that the rod


which operates the microswitches has still
~2mm free, when the door is closed.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
26 / 40

To copy this manual, even if partially, is forbidden.

CONTROL OF SOLENOID VALVES:

1 SV body with main vent 2 Seal 3 SV external part

After opening the solenoids valves check the following conditions:

1. Check the main vent of the SV (1): if it is damaged, immediately replace the whole SV;

2. Check that the Viton face of the seal (2) is in good condition: if it is damaged, replace
the whole SV.

3. Tightly lock the external part of the SV again (3) to avoid preassure leakages and
maintain watertight integrity.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
27 / 40

To copy this manual, even if partially, is forbidden.

CONTROL OF THE WATER INJECTION PUMP:

To effectively check the injection pump and its relative solenoid valve we suggest to
disassemble the whole group from the machine and test it separately.
1. Fix a silicone pipe onto the pump;
2. Directly supply 230V ac voltage both to the pump and the SV. We recommend to
separate the pump from the SV and check them separately.

ATTENTION:

The “pump injection group” is supplied with vaporizer but without temperature sensor.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
28 / 40

To copy this manual, even if partially, is forbidden.

CONTROL OF THE WATER CHARGE PUMP:

To check the water charge pump, Fix a silicone pipe onto the pump itself and connect it
directly to the voltage (230V ac). If no water suction takes place, follow the instructions below:
1. To supply voltage to the pump;
2. Blow compressed air for 3-4 seconds while the pump is operating;
3. Check operation again with water;
4. In case of faulty operation replace the whole pump.

FUSES:

Fuses features are indicated in the user manual (it is also possible to find the electrical
capacity on the fuse body itself). Test the efficiency of fuses with a common tester by checking
the continuity between the two poles of the fuse.

MINIMUM POWER FUSE : T12Amp


MAXIMUM POWER FUSE : F16Amp

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
29 / 40

To copy this manual, even if partially, is forbidden.

BLUE GASKET CHANGE INSTRUCTIONS:

1 – Open the autoclave door and unscrew the four screws highlighted in the picture to
remove external aluminium / plastic cover

2 – Slightly press the shell on internal 3 – At the same time, loosen the bolt
side of the door as displayed on external side of the door, but
in the picture not completely

4 – Completely remove the blue gasket pulling it slightly out

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
30 / 40

To copy this manual, even if partially, is forbidden.

5 – Insert new gasket replacing the old one. The edge of the steel shell has to perfectly fit
into the gasket housing.

6 – As displayed in the picture, slightly press the shell preventing it from rotating, then
tighten the bolt again. During this operations check the gasket to fit properly into the door
housing.

7 – Tighten the bolt until it perfectly aligns with internal screw (see
the picture). Lock the bolt putting siliconic glue on its external side.

8 – The shell must have a slight mobility around vertical and


horizontal axes. Check this mobility pushing it on its edges.
The mobility is useful for the shell to fit the edge of the
boiler once the door is closed.

9 – Replace the external cover and check the door can


properly close

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
31 / 40

To copy this manual, even if partially, is forbidden.

9. WATER DIAGRAM

LEGEND

PRESENT ON
AT TEST attachment All models
C Chamber All models
SV Safety valve All models
UV Unidirectional valve All models
V Vaporizer All models
EV1 Water discharge EV nc All models
EV2 Over-pressure EV nc All models
EV3 EV nc on injection pump All models
EV4 EV nc water mains sys. B
EV5 EV 3 ways B–S
EV8 EV no bacterial filter B-S
F1 Bacterial filter B-S
F2 Plastic filter All models
PM1 Injection pump All models
PM2 Water charge pump All models
PM3 Vacuum pump B-S
R Radiator All models
S1 Clean water tank All models
S2 Dirty water tank All models
T Pressure transducer All models

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
32 / 40

To copy this manual, even if partially, is forbidden.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
33 / 40

To copy this manual, even if partially, is forbidden.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
34 / 40

To copy this manual, even if partially, is forbidden.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
35 / 40

To copy this manual, even if partially, is forbidden.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
36 / 40

To copy this manual, even if partially, is forbidden.

10. WIRING DIAGRAM

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
37 / 40

To copy this manual, even if partially, is forbidden.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
38 / 40

To copy this manual, even if partially, is forbidden.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
39 / 40

To copy this manual, even if partially, is forbidden.

EV7 was present in the models built before 2006/08/01.

ENGLISH
GENERAL TECHNICAL MANUAL
MT_BSN EN
Rev. 2.3.1

for B, S & N Class AUTOCLAVES (18 & 23 Lts.)


2012.03.26
40 / 40

To copy this manual, even if partially, is forbidden.

EV6 was present in the models built before 2006/08/01.

The manufacturer reserves the right to make technical modifications without any notice.
The manufacturer declines all responsibilities for calibrations or modifications done on mechanic, pneumatic and electronic parts,
carried out by non-authorized technicians.
The use of NON ORIGINAL spare parts immediately excludes the warranty of the machine and FREES the manufacturer
from any responsibility deriving from it.