ERAVAP – Vapor Pressure Tester Turning on the analyzer Starting the measurement Position the analyzer properly on the bench and connect Connect the Luer with the tube to the inlet of the tester the outlet tube and the waste container to the left-hand and put the tube into your sample. side of the analyzer. After connecting the power cable to the back panel of the analyzer, press the main power After having entered all settings, the measurement can be switch. started by pressing the “RUN” button (either on the display or via the button below the “RUN” sign on the display). The analyzer will load the firmware. During this process, If additional rinsing of the measuring cell is desired, press the LED on the front will flash. The loading takes the “RINSE” button before starting the measurement with approximately 15 seconds. After loading, the analyzer will “RUN”. automatically display the measurement menu: As soon as the measurement is started, the screen changes from the settings to the measurement screen. This screen provides information about the status of the measurement including information about the actual temperature and pressure inside the measurement cell.
The entire measurement is fully automated and takes
approximately 5 minutes for a vapor pressure measurement of gasoline at 37.8°C (100°F) according to ASTM D5191.
When the measurement
is finished, the result is displayed on the measurement screen. At the same time the result is also saved in the result memory.
Press “OK” to return to
the settings of the measurement screen to start the next measurement. Measurement settings Choose the “Standard”, the “Sample ID”, the “Operator” Results menu name and the number of measurements (“No. Meas”) by The results are stored directly in the “Results” menu and using the drop down lists. can be viewed and compared there. Be aware that for several standard methods the sample has to be prepared (e.g. chilled and air saturated). This The results are shown for one method only. Change the prepared state may change if the sample is kept under method to view the corresponding results. ambient temperature longer than necessary. Print results ASTM D5191 (EN 13016-1) Select the results to be printed by the line number and If you choose the standard method ASTM D5191, the test then press the “Print” button. temperature is automatically set to 37.8°C (100°F) and The selected results are printed (for printer selection and cannot be changed. configuration see the “System” menu). ASTM D6378 For a form feed press the “Page” button. If you choose the standard method ASTM D6378, the test temperature can be changed as well, but the DVPE Save results on USB memory stick formula is valid for test temperatures at 37.8°C (100°F) For saving all the results connect a USB memory stick and only. wait a few seconds. The memory stick is automatically mounted. Pressing the “Save” button all results from all Document Revision: ERAVAP-m-71003-short standard methods are stored.