Professional Documents
Culture Documents
BX-3010 Software InstallationVer1.0
BX-3010 Software InstallationVer1.0
Version 1.0
Status
Validator ...
Document history
(Add one line for each new release or review version).
Document distribution
Sysmex Asia Pacific Pte Ltd
Sysmex Corporation
3
SYSMEX ASIA PACIFIC PTE LTD
1. Preparatory setup (in Windows Vista)
1.1 Internet Protocol (TCP/IP) Setup
1. Click [Start] > [Control Panel] > [Network and Sharing Center].
2. Click [Manage network connections] > “Local Area Connection” icon. Right-click the "Local Area
Connection” icon and select “Properties”.
Figure 1.1-1
3. Tick “Internet Protocol Version 4 (TCP/Ipv4) and click the [Properties] button in the [Networking] tab.
4. Tick “Use the following IP address" in the [General] tab and enter the following information.
IP address: 172.17.31.201
Subnet mask: 255.255.0.0.
Default gateway: (blank)
Figure 1.1-2
4
SYSMEX ASIA PACIFIC PTE LTD
1.2 Security setting (Firewall settings off)
6. Click [Start] > [Control Panel] > [Security Center].
Figure 1.2 - 3
7. Click the [Windows Firewall] and display the “Windows Firewall” window.
8. Click the [Change Settings] and display the “Windows Firewall Settings” window.
Figure 1.2 - 4
9. Select [General] tab and tick “Off (not recommended)”. Click the [OK] button.
Figure 1.2 -5
5
SYSMEX ASIA PACIFIC PTE LTD
1.3 Security setting (Automatic updating off)
10. Click [Start] > [Control Panel] > [Security Center].
11. Click “Windows update” and display the “Windows Security Center” window.
12. Click “Automatic updating” and “Check for updates” and display the “Change settings” window.
Figure 1.3 - 6
13. Tick “Never check for update (not recommended)”. Click the [OK] button.
Figure 1.3 -7
6
SYSMEX ASIA PACIFIC PTE LTD
1.4 Security setting (Malware protection off)
14. Click [Start] > [Control Panel] > [Security Center].
15. Click “Windows update” and display the “Windows Security Center” window.
16. Click [Malware protection] > [Show my available protections] and display the “Windows Security
Center” window.
Figure 1.4 - 8
17. Tick “Don’t monitor my antivirus software state (not recommended)”. Click the [OK] button.
Figure 1.4 – 9
7
SYSMEX ASIA PACIFIC PTE LTD
1.5 Screen setting
18. Click [Start] > [Control Panel] > [Personalization].
19. Click the [Screen saver] and display the “Screen Saver Settings” window.
Figure 1.5 - 11
8
SYSMEX ASIA PACIFIC PTE LTD
21. Click the [Change power setting…] and display “Power Options” window.
Figure 1.5 - 12
22. Click the [Change plan settings] at the Preferred plans and display the “Change settings for the plan:
Balanced” window.
Figure 1.5 - 13
9
SYSMEX ASIA PACIFIC PTE LTD
23. Specify the setting in the “Change settings for the plan: Balanced” window.
(a) Turn off the display : Never
(b) Put the computer to sleep: Never
Figure 1.5 - 14
Figure 1.5 – 15
Figure 1.5 – 16
10
SYSMEX ASIA PACIFIC PTE LTD
26. Click the [Personalization] > [Display Settings].
Figure 1.5 - 17
27. Specify the setting in the [Monitor] tab.
(a) Screen resolution: 1024 × 768 pixels
(b) Color quality: Medium (16 bit) and more
Figure 1.5 - 18
11
SYSMEX ASIA PACIFIC PTE LTD
1.6 Regional setting
In multilingual version Windows Vista, regional setting is required.
Note: some regions may not be covered depending on software.
Country Standards and Location Language for non- Decimal
formats Unicode programs symbol
France (FRA) French France French .
Germany (DEU) German Germany German .
Spain (ESN) Spanish Spain Spanish .
Chinese (CHS) Chinese China Chinese .
Russia (RUS) Russian Russia Russian .
United States English United States of English .
(ENU) America
Japan (JPN) Japanese Japan Japanese .
Portugal (PTG) Portuguese Portugal Portuguese .
Italy (ITA) Italian Italy Italian .
Norway (NOR) Norwegian Norway Norwegian .
Sweden (SVE) Swedish Sweden Swedish .
Greece (ELL) Greek Greece Greek .
Poland (PLK) Polish Poland Polish .
28. Click [Start] > [Control Panel] > [Regional and Language Options].
29. Select the [Formats] tab in the “Regional and Language Options” window.
(a) Select the country name from drop-down menu.
Figure 1.6 - 19
12
SYSMEX ASIA PACIFIC PTE LTD
30. Click the [Customize this format…] button and display the “Customize Regional Options” window.
Figure 1.6 - 20
(a) Specify in the [Numbers] tab and click the [Apply] button.
Decimal symbol: . (Period)
Figure 1.6 - 21
13
SYSMEX ASIA PACIFIC PTE LTD
(b) Specify in the [Time] tab and click [Apply] button.
Time format: H.mm.ss
AM symbol: AM
PM symbol: PM
Figure 1.6 - 22
(c) Specify in the [Date] tab and click the [Apply] button.
Short date: yy/MM/dd
Long date: yy/MM/dd
Figure 1.6 - 23
14
SYSMEX ASIA PACIFIC PTE LTD
31. Select the [Location] tab in the “Regional and Language Options” window.
(a) Select the country name from drop-down menu.
Figure 1.6 - 24
32. Select the [Keyboards and Languages] tab in the “Regional and Language Options” window.
(a) Click the [Change keyboards…] button and display the “Text Services and Input Languages"
window.
15
SYSMEX ASIA PACIFIC PTE LTD
(c) When not found in the country name list, add the name.
Figure 1.6 - 26
16
SYSMEX ASIA PACIFIC PTE LTD
1.7 Windows setting
35. Click [Start] > [Control Panel] > [Personalization].
17
SYSMEX ASIA PACIFIC PTE LTD
1.8 Daylight saving time setting
Daylight saving time is utilized depending on the countries and regions. Tick the “Automatically
adjust clock for Daylight Saving time” for adopting when it appears in the setting window.
39. Click [Start] > [Control Panel] > [Date and Time].
Figure 1.8 - 31
40. Click the [Change time zone…] button in the [Date and Time] tab and display the “Time zone settings”
window.
Figure 1.8 - 32
18
SYSMEX ASIA PACIFIC PTE LTD
1.9 Printer setting
Follow the printer setting as described in its operator’s manual since the installation procedure varies depending
on the printer type.
43. Click [Start] > [Control Panel] > [Printers].
44. Select and right-click the “Printers” icon and display the “Printer Properties” window.
45. Select the [Advanced] tab.
● Spool the document to be printed to process program printing at high-speed.
○ Spool data for all pages and send it as print data to printer: (remove the checkmark.)
● Send print data to printer immediately.
Figure 1.9 - 34
19
SYSMEX ASIA PACIFIC PTE LTD
1.10 Telnet connection setting
48. Click [Start] > [Control Panel] > [Programs and features] and display “Program and functions" window.
49. Click the [Turn Windows features on or off] and display the “Windows features” window.
Figure 1.10 - 35
Figure 1.10 - 36
20
SYSMEX ASIA PACIFIC PTE LTD
Figure 1.10 - 37
53. To terminate this session, click the [X] button at the upper-right corner.
Figure 1.11 - 38
Figure 1.11 – 39
21
SYSMEX ASIA PACIFIC PTE LTD
(b) Select the [General] tab and specify the following.
Profile Name: ca270
Host Name/Address: 172.17.31.202
Username: ca270
Password/Phrase: qnx
Initial Local Folder: C:\
Initial Host Folder: \home\ca270
Figure 1.11 – 40
Figure 1.11 – 41
22
SYSMEX ASIA PACIFIC PTE LTD
(d) Select the [Kanji Code] tab and specify the following.
Japanese Kanji Code: No conversion [S-JIS]
Filename Kanji Code: SHIFT-JIS
Figure 1.11 – 42
Figure 1.11 – 43
23
SYSMEX ASIA PACIFIC PTE LTD
(f) Select the [Special] tab and specify the following.
Tick “Use NLST-R for high-speed recursive search”.
Permissions Change Command: SITE CHMOD
Host Type: Auto
NLST Filename/Option: -aL
24
SYSMEX ASIA PACIFIC PTE LTD
2.0 Analyzer software update
1. Terminate all programs running in the operational PC. Terminate the operational
software in the following procedure.
(a) Press and hold the [Ctrl] [Shift] and [.] keys. The following message appears.
(b) Click the [Power Off] button to shut down the operational software.
2. Double click the “FFFTP” icon on desktop to activate [FFFTP].
Figure 2.0 – 45
3. Select “ca270” in the “Host List” window. Click the [Connect] button.
4. [FFFTP] becomes activated. The following window appears. It indicates the contents [C:] in the
operational PC at the left side of the screen and [/home/ca270] in the analyzer software at the right side.
Figure 2.0 - 46
5. Double-click the [update] folder at the right side of the screen and display the [/home/ca270/update]
contents.
6. Double click the folder that contains the analyzer update software (Need the Unit Main software from
the CD) at the left side and display them.
7. Select the analyzer update software (ca270.tar.F) and click the "Upload" icon on the tool bar. The data
(CA270.tar) is transferred to the [update] folder in the analyzer software.
8. Double-click the shortcut icon for “172.17.31.202ht” on desktop and activate “Hyper Terminal”.
Figure 2.0 – 47
25
SYSMEX ASIA PACIFIC PTE LTD
9. When [Login:_] appears, enter “ca270”.
10. When [password:_] appears, enter “qnx”.
Figure 2.0 - 48
13. The data in the analyzer software is automatically updated. Wait until the processing completed.
14. When “ca270>_" appears after the update completed, enter “logout” to terminate [Hyper Terminal].
26
SYSMEX ASIA PACIFIC PTE LTD
3.0 Installation of IPU software
1. Terminate all programs running in the operational PC. Terminate the operational
software in the following procedure.
(a) Press and hold the [Ctrl] [Shift] and [.] keys. The following message appears.
(b) Click the [Power Off] button to shut down the operational software.
2. Double click the “ca_setup” icon in the software CD.
Figure 3.0 – 49
27
SYSMEX ASIA PACIFIC PTE LTD
4. Wait till the below screen appear and then click “Yes”
Figure 3.0 - 51
Figure 3.0 - 52
7. Wait for the below screen and key in the Engineer no. at the key icon and then press “login” button.
Figure 3.0 - 53
28
SYSMEX ASIA PACIFIC PTE LTD
4.0 User-interface software update
A. System Parameters Backup
1. Select [System (F9) > Backup Procedures]. Click the […] button on the “Directory” for system
parameters backup.
2. Click on the [Save] button.
B. User-interface software termination
3. Terminate all programs running in the operational PC. Terminate the operational software in the
following procedure.
(a) Press and hold the [Ctrl] [Shift] and [.] keys. The following message appears.
(b) Click the [Power Off] button to shut down the operational software.
C. Deletion of the current (before update) operational software
4. Click [Start] > [Control Panel] in Windows XP to display the “Control Panel” Window.
5. Click [Add or Remove Programs] and select [Analyzer]. Click the [Remove] button.
6. The [Analyzer-InstallAware Wizard] window appears. Click the [Next>] button to execute the deletion.
7. After completing the deletion, the [Analyzer-InstallAware Wizard] window automatically disappears.
D. Installation of new operational software
8. Save the update program [ca_setup.exe] on desktop from the external memory medium (such as USB
memory).
9. Double-click the “ca_setup.exe” icon. The [Install Aware Wizard] window appears.
10. Click on the [Next>] button. The installation automatically starts..
11. After installation, the following window appears to restart the operational PC. Click the [Yes] button.
Click [Start] > [Turn Off Computer…] > [Restart].
12. Restart the operational PC. Double-click the shortcut “CAMAIN” and start the operational software.
E. Load the system parameters.
9. Select [System (F9) > [Backup Procedures]. Click the {…} button on the “Directory” for system
parameters backup.
10. Click the [Load] button to load the system parameters.
11. Press and hold the [Ctrl] and [.] keys. Click the [Power Off] button and terminate the operational
software.
12. Double click the shortcut “CAMAIN” and restart the operational software.
F. Final Confirmation
13. Select [System (F9)] > [Versions]. Check the version number of the "User interface” is updated.
14. Select [System (F9)] > [Sequence]. Click the [Initialization] button and check the initialization is
performed properly.
29
SYSMEX ASIA PACIFIC PTE LTD
4.1 Changing correction parameters
A. Changing temperature and quantitative parameter
1. Terminate all programs running in the operational PC. Terminate the operational software in the
following procedure.
(a) Press and hold the [Ctrl] [Shift] and [.] keys. The following message appears.
(b) Click the [Power Off] button and shut down the operational software.
2. Double click the “FFFTP” icon on desktop to activate [FFFTP].
Figure 4.0 - 1
3. Select “ca270” in the “Host List” window. Click the [Connect] button.
4. Open the [/home/ca270/sysboot/system] folder at the “FFFTP” window (right side of screen). Select
the “system.txt”.
5. Click the “download” icon from the toolbar and save the “sytem.txt” file into c: drive in the operational
PC.
Figure 4.0 - 2
6. Use text editor software (such as Notepad) to open the “system.txt” file. Modify and save the file.
Figure 4.0 - 3
9. Double-click the modified “system. txt” uploaded in the “FFFTP” window (left side of screen) and
check that the file is the modified version. After confirmation, click the [X] button on the “FFFTP”
window and exit the [FFFTP].
30
SYSMEX ASIA PACIFIC PTE LTD
B. Changing Position Map (mapoffset.txt)
After replacing or repairing the SPT, the RPT, the MIX1, the MIX2, or the WU, the offset file needs to be
updated only when adjustment for the height direction is required.
1. Terminate all programs running in the operational PC. Terminate the operational software in the
following procedure.
(a) Press and hold the [Ctrl] and [.] keys. The following message appears.
(b) Click the [Power Off] button and shut down the operational software.
Figure 4.0-4
3. Select “ca270” in the “Host List” window. Click the [Connect] button.
4. Open the [/home/ca270/sysboot/pmap] folder at the “FFFTP” window (right side of screen). Select the
file to be modified.
5. Click the “download” icon from the toolbar and save the file into c: drive in the operational PC.
Figure 4.0-5
6. Use text editor software (such as Notepad) to open the file. Modify and save the file.
R Contact our service division when needs to modify the file.
7. Select the modified file at the “FFFTP” window (left side of screen).
8. Click the “upload” icon from the toolbar and save the modified file into the [/home/ca270/sysboot/pmap]
folder in the operational software.
9. Double-click the uploaded file in the “FFFTP” window (left side of screen) and check that the file is the
modified version. After confirmation, click the [X] button on the “FFFTP” window and exit the
[FFFTP].
1. Terminate all programs running in the operational PC. Terminate the operational software in the
following procedure.
(a) Press and hold the [Ctrl] and [.] keys. The following message appears.
(b) Click the [Power Off] button and shut down the operational software.
31
SYSMEX ASIA PACIFIC PTE LTD
2. Double click the “FFFTP” icon on desktop to activate [FFFTP].
Figure 4.0 - 6
3. Select “ca270” in the “Host List” window. Click the [Connect] button.
4. Open the [/home/ca270/sysboot/kokyaku] folder at the “FFFTP” window (right side of screen). Select the
file to be modified.
5. Click the “download” icon from the toolbar and save the file into c: drive in the operational PC.
Figure 4.0 – 7
6. Use text editor software (such as Notepad) to open the file. Modify and save the file.
R Contact our service division when needs to modify the file.
7. Select the modified file at the “FFFTP” window (left side of screen).
End of document
32
SYSMEX ASIA PACIFIC PTE LTD