Professional Documents
Culture Documents
Measurement 14
Measurement 14
Measurement 14
14.1 CLASSIFICATION OF RESISTANCES conditions in the circuit are not disturbed. However,
The classification of resistances, from the point of in practice this is not possible and hence both the
view of measurement, is as follows : methods give inaccurate results.
(i) Low resistances. All resistances of the order of Consider circuit of Fig. 14.1(a). In this circuit the
1 Q and under may be classified as low ammeter measures the true value of the current
resistances. through the resistance but the voltmeter does not
(ii) Medium resistances. This class includes resis measure the true voltage across the resistance. The
tances from 1 Q upwards to about 0.1 MQ. voltmeter indicates the sum of the voltages across the
(iii) High resistances. Resistances of the order of ammeter and the measured resistance.
0.1 MQ and upwards are classified as high
resistances.
The classification outlined above is not rigid, but
forms a basis for techniques, followed or measure
ment, which may be different for different classes.
(421)
422 Electrical and Electronic Measurements and Instrumentation
It is clear from Eqn. 14.5 that the error in measure Fig. 14.1(b). If the two instruments are of 0.5%
ments would be small if the value of resistance under accuracy and are read near full scale, the instrumental
measurement is large as compared to the internal error in the result may be any thing from 0 to 1%. If
resistance of the ammeter. Therefore the circuit of read near half scale, the percentage error may be twice
Fig. 14.1(a) should be used when measuring high as great and for lower readings may mount
resistance values. considerably higher. With less accurate instruments
Consider the circuit of Fig. 14.1(b). In this circuit
the possible error, of course is increased. It is difficult
the voltmeter measures the true value of voltage but to obtain accuracy much better than 1% in a resistance
the ammeter measures the sum of currents through value under usual conditions and the error sometimes
the resistance and the voltmeter. may be considerably higher. However, the method is
useful in some laboratory work in which high
Let Rp be the resistance of the voltmeter.
accuracy is not required.
Current through the voltmeter, I = VI Iv
The suitability of a particular method for
Measured value of resistance,
resistance measurement depends upon the relative
r = V- V _ V R values of resistance under measurement and the
m2 1 Ir + Iv~V/R + V/Rv~ 1+R/Rv resistance of the meters.
The division point between the two methods is at
True value of resistance,
the resistance for which the relative errors given by
R=^n2Rv =R f----- 1------- ...(14.6) the two methods are equal.
Rv~Rm2 The relative errors for the two cases are equal
From Eqn. 14.6 it is clear that the true value of when :
resistance is equal to the measured value only if the
=— (See Eqns. 14.5 and 14.9)
resistance of voltmeter, Rv, is infinite. However, if the
resistance of voltmeter is very large as compared to
R Rv
the resistance under measurement: or when true value of resistance
the position of r are unaltered, the two resistances Hence, unknown resistance
must be equal. Thus the value of unknown resistance
R = (S+G)-1-G
R is equal to the dial settings of resistance S. 02
= (0.5 x 106 +10 x 103)x (41 / 51) -10 x 103
= 0.4xl06Q =0.4 MQ.
For the galvanometer current to be zero, the 14.2.4 Sensitivity of Wheatstone Bridge
following conditions also exist: It is frequently desirable to know the galvano
meter response to be expected in a bridge which is
...(14.12)
1 3 P+Q slightly unbalanced so that a current flows in the
galvanometer branch of the bridge network. This may
and / -j =_L_ ...(14.13) be used for :
2 4 R+S
(z) selecting a galvanometer with which a given
where E - emf of the battery. unbalance may be observed in a specified
Combining Eqns. 14.11, 14.12 and 14.13 and bridge arrangement,
simplifying, we obtain : (zz) determining the minimum unbalance which
can be observed with a given galvanometer
in the specified bridge arrangement, and
(z'zz) determining the deflection to be expected for
from which QR = PS ...(14.15)
a given unbalance.
Equation 14.15 is the well known expression for
The sensitivity to unbalance can be computed by
the balance of Wheatstone bridge. If three of the
solving the bridge circuit for a small unbalance. The
resistances are known, the fourth may be determined
solution is approached by converting the Wheatstone
from Eqn. 14.15 and we obtain :
bridge of Fig. 14.3 to its "Thevenin Equivalent" circuit.
P
R = S± ...(14.16) Assume that the bridge is balanced when the branch
resistances are P, Q R, S so that P / Q = R / S. Suppose
where R is the unknown resistance, S is called the the resistance R is changed to R +\R creating an
'standard arm' of the bridge and P and Q are called the unbalance. This will to cause an emf e to appear across
'ratio arms'. the galvanometer branch. With galvanometer branch
In the industrial and laboratory form of the open, the voltage drop between points a and b is :
bridge, the resistors which make up P, Q and S are p =1p= _
mounted together in a box, the appropriate values ab 1 P+Q
being selected by dial switches. Battery and galvano E(R + AR)
Similarly, Efld=/2(R + AR) =
meter switches are also included together with a R+AR+S
galvanometer and a dry battery in the portable sets. P
Therefore voltage difference between points d and
and Q normally consist of four resistors each, the
b is :
values being 10, 100, 1000 and 10,000 Q respectively S
consists of a 4 dial or 5 dial decade arrangement of
resistors. Figure 14.4 shows the commercial form of ab R+AR+S P+Q
Wheatstone bridge. ...(14.17)
P _ R
Battery © Galvanometer terminals P+Q-R+S
terminals © ©GO
Terminals © %
R + AR P
e= E
for connecting R + AR + S ~P+S
unknown © X2 ESAR
resistance
(R + S)2 + AR(R + S)
Ratio selector
ESAR
~ (R + S)2 ...(14.18)
R + AR R
The bridge sensitivity SB is defined as the
deflection of the galvanometer per unit fractional R + AR + R R+R
change in unknown resistance. 'R + AR 1
Bridge sensitivity 2R + AR 2
...(14.20)
B AR/R as AR « R. ...(14.25)
_ SVESR
...(14.21) The resistance of the Thevenin equivalent circuit
“(R + S)2
is found by looking back into terminals c and d
From Eqn. 14.21, it is clear that the sensitivity of (Fig. 14.3) and replacing the battery by its internal
the bridge is dependent upon bridge voltage, bridge resistance. In most cases, however, the extremely low
parameters and the voltage sensitivity of the galvano resistance of the battery can be neglected and this
meter. Rearranging the terms in the expression for simplifies the solution as we can assume that
sensitivity, terminals a and b are shorted. The Thevenin equivalent
S - S»E - S» E resistance can be calculated by referring to Fig. 14.5.
B (R + S)2/SR ^+2 + 4
5 1\
...(14.22)
P +2 +Q
Q P
From Eqn. 14.22, it is apparent that maximum
sensitivity occurs where R / S = 1. As the ratio becomes
either larger or smaller, the sensitivity decreases. Since
Fig. 14.5 Finding resistance of bridge looking into
the accuracy of measurement is dependent upon terminals d and b.
sensitivity a limit can be seen to the usefulness for a
given bridge, battery and galvanometer combination. Thevenin equivalent resistance of bridge
RS PQ
For a bridge with equal arms, R = S - P = Q, R + S+ P+Q ...(14.26)
S E
Bridge sensitivity ...(14.23)
Considering AR « R
As explained above the sensitivity is maximum For a bridge with equal arms,
when the ratio is unity. The sensitivity with ratio p=Q = S = Rf
P / Q = R/ S =1000 would be about 1/250 of that for R0 = R ...(14.27)
unity ratio. The sensitivity with PIQ = R/S = 1000
The Thevenin equivalent of the bridge circuit
would similarly be about 1/250 of that for unity ratio.
therefore reduces to a Thevenin generator with an emf
Thus the sensitivity decreases considerably if the £0 and an internal resistance Rq. This circuit is shown
ratio P/ Q - R/ S is greater or smaller than unity. This in Fig- 14.6.
reduction in sensitivity is accompanied by a reduction
in accuracy with which a bridge can be balanced.
Galvanometer current. The current through the
galvanometer can be found out by finding the
Thevenin equivalent circuit. The Thevenin or open
circuit voltage appearing between terminals b and d
with galvanometer circuit open circuited is,
aa ac
£Un = E . -En = UR
2v + AR)-LP
' 1 Fig. 14.6 Thevenin equivalent circuit of
_ £(R + AR) EP Wheatstone bridge.
~ R + AR + S P+Q
The current in the galvanometer circuit
R + AR P
...(14.24) ...(14.28)
R + AR + S P+Q
For a bridge with equal arms, R = S = P = Q. where G = resistance of the galvanometer circuit.
Measurement of Resistance 427
For a bridge with equal arms, reversal. The results are obtained by averaging the
I _ E(AR/4R) two readings. This way the effect of thermo-electric
...(14.29) emfs is eliminated with the added advantage of
s (R + G)
doubling the sensitivity of the bridge.
The deflection of the galvanometer for a small
3. Temperature effects. The errors caused by
change in resistance in the unknown arm is,
change of resistance due to change of temperature
SVESAR produces serious errors in measurements especially in
0=
(R + S)2 the case of resistances made up of materials having a
Sj large value of resistance temperature coefficient. For
But Sv = example in the case of copper, which has a resistance
temperature coefficient of 0.004/qC, a change in
where S = current sensitivity of the galvanometer. temperature of 1°C will cause an error of 0.4% or 1 part
SESAR in 250.
0 =------ -l------------- j ...(14.30)
(Ro + GXR + S)2 4. Contact resistance. Serious errors may be
caused by contact resistances of switches and binding
For a bridge with equal arms,
posts. A dial may have a contact resistance of about
SEAR 0.003 Q and thus a 4 dial resistance box has a contact
0 = —l----- ----- ...(14.31)
4R(R + G) resistance of about 0.01 Q. This value is quite high
Also bridge sensitivity, especially when low resistance measurements are
being done. Another aspect of the contact resistance is
0 S. ESR
...(14.32) that error caused by it is difficult to account for since
AR/R (Rq + GXR + S)2
its magnitude i.e., magnitude of contact resistance is
For a bridge with equal arms, uncertain.
Bridge sensitivity, In precision resistance measurements, the most
accurate comparisons are made on an equal ratio
S --§£
...(14.33) bridge with a fixed standard resistance nominally
B 4(R+G)
equal to resistance under test. Then with equal leads,
equal currents and equal heating of the ratio arms, the
14.2.5 Precision Measurement of Medium
possible errors are minimized. The problem is then
Resistances with Wheatstone Bridge
reduced to that of determining the exact ratio of the
It is sometimes necessary to measure resistances
unknown resistance, R^ to the standard resistance S or
to a precision of 1 part in 10,000 or even greater by the difference between them. The different methods
comparing them with standard resistances. In such used for this purpose are :
cases more than ordinary precautions are necessary in
(/) Change in ratio arms. Small known changes are
order to secure the required accuracy. The following
made in the ratio arms and the exact balance is
factors should be taken into consideration.
obtained.
1. Resistance of connecting leads. A lead of
(n) Using a high resistance shunt. In this case the
22 SWG wire having a length of 25 cm has a resistance
bridge is balanced by means of an adjustable high
of about 0.012 Q and this represents more than 1 part
resistance put in parallel with one of the bridge arms.
in 1000 for a 10 Q resistance or more than one part in
A decade resistance box is used for this purpose.
10,000 for a 100 Q resistance.
Suppose the test resistance, R, is slightly lower in
2. Thermo-electric effects. Thermoelectric emfs value than the standard resistance, S and let the
are often present in the measuring circuit and they balance be obtained with a resistance xS, put in
must be taken into account since they affect the parallel with S.
galvanometer deflection in the same way as an emf
The resistance of this arm is then
occurring because of unbalance. The effect of
thermoelectric emfs and other parasite emfs on the xS2/(S + xS), i.e., S/(1 + 1/ x).
measurement may be eliminated by reversing the
:. Value of unknown resistance
battery connections to the bridge through a quick
acting switch and adjusting the balance until no R=£._L_.
change in galvanometer deflection can be observed on Q 1+1/x
428 Electrical and Electronic Measurements and Instrumentation
Since x is very large and therefore 1/x is very Now, for first balance,
small and hence we can write, p R + Lr + S+(L-L)r R + S+Lr ...
p — +1 — •••(0
R = ~-S(l-l/x) ...(14.34) Q S + (L-Z1)r S + (L-/1)r
J-02
R _ 7554
—03 -<v2 —03
^1/10
S ~ 5346
Vm/10 ^2/10 v3/io n vouf
—o 2
°2 —0 2 i—01
^1/10
Vin/W V2/IO ^3/10 ||
—o 1 ---O1 —0 1 • 0u
Kn/10 ^1/10 Vz/lO
—o 0 u —0 0
•------------- ------------------- i
* Resistance of each unit in first decade is 10 kQ The second decade has a total resistance of 2 x 10 = 20 kQ. Therefore, each unit
in the second decade has a resistance of 20/10 = 2 kQ In other words, we can say that if the first decade consists of 11 coils of
resistance r ohm each, the next decade has 11 coils of r/5 ohm each and the third decade has 11 coils of 2r/5 ohm each.
430 Electrical and Electronic Measurements and Instrumentation
certain extent by increasing the emf applied to the Q = 100 £1, R = 2,005 Q and S = 200 Q. The battery has an
bridge but in this case care has to be taken to avoid emf of 5 V and negligible internal resistance. The
overheating of any arm of the bridge. Inaccuracy may galvanometer has a current sensitivity of 10 mm/pA and an
also be introduced on account of leakage over internal resistance of 100 Q. Calculate the deflection of
insulation of the bridge arms when measuring very galvanometer and the sensitivity of the bridge in terms of
high resistances. For measurement of very high deflection per unit change in resistance.
resistances, a megohm bridge (described on page 441) Solution. Resistance of unknown resistor
is preferable. required for balance
The lower limit for measurement is set by the R=(P/Q)S = (1000 /100) x 200 = 2000 Q.
resistance of the connecting leads and by contact In the actual bridge the unknown resistor has a
resistance at the binding posts. The error caused by value of 2005 Q or the deviation from the balance
leads may be corrected fairly well, but contact resis
conditions is A R =2005 -2000 = 5 Q.
tance presents a source of uncertainty that is difficult
Thevenin source generator emf
to overcome. The lower limit for accurate measure
ment is in the neighbourhood of 1 to 5 ohm. For low = -
-------- P
resistance measurements, therefore, a Kelvin bridge R + S P+Q
(described in Art. 14.3.2 on page 434) is generally
r 2005 1000
preferred. —5
L2005+200 1000+100
Example 14.5 A resistance of approximately 3000 Q is
= 1.0307 xlO’3 V.
needed to balance a bridge. It is obtained on a 5 dial
resistance box having steps of 1000, 100, 10, 1 and 0.1 Q. Internal resistance of bridge looking into
The measurement is to be guaranteed to 0.1 percent. For this terminals b and d.
accuracy, how many of these dials would it be worth RS PQ
adjusting ? R + S +P+Q
Solution. Limiting value of 3000 Q resistor 2005x200 + 1000x100 _ 272 8 q
= 3000 ± 3000 x (0.1 /100) = 3000 ± 3Q 2005 + 200 1000 + 100 ~
= 2997 to 3003 Q Hence the current through the galvanometer
Thus it would be worth adjusting the 1000,100, 10 1.0307 x IQ"3
A = 2.77 p A.
and 1 Q dials. R0 + g 272.8 + 100
Example 14.6 Each of the ratio arms of a laboratory type Deflection of the galvanometer
Wheatstone bridge has guaranteed accuracy of ± 0.05%, 0 = SJg =10x2.77 =27.7 mm/fi.
while the standard arm has a guaranteed accuracy of Sensitivity of bridge
± 0.1%. The ratio arms are both set at 1000 Q and the bridge
r 0 27.7 c
is balanced with standard arm adjusted to 3,154 Q. S„ ----- =------ = 5.54 mm/Q.
B AR 5
Determine the upper and the lower limits of the unknown
resistance, based upon the guaranteed accuracies of the Example 14.8 A Wheatstone bridge has ratio arms of
known bridge arms. 1000 £1 and 100 £1 and is being used to measure an
Solution. Value of unknown resistance unknown resistance of 25 £1. Two galvanometers are
available. Galvanometer 'A' has a resistance of 50 £1 and a
R = (p j Q)x S = (1000 / 1000)x 3154
sensitivity of 200 mm/pA and galvanometer 'Bt has values
= 31540. of 600 £1 and 500 mm/pA. Which of the two galvanometers
:. Percentage error in determination of R. more is sensitive to a small unbalance on the above bridge,
5R 8P , 5Q 8S and what is the ratio of sensitivities ? The galvanometer is
— = ± — ± —-- ± —
R P Q S connected from the junction of the ratio arms to the opposite
comers. Comment upon the results.
= ±0j05 ± 0j05 ± 0j01 = ± 0.2 %.
Solution. The arrangement of this bridge is
Limiting values of
shown in Fig. 14.3. Value of standard resistance under
R = 3154 ± 0.2% = 3091 to 3217 Q. balance conditions
Example 14.7 In the Wheatstone bridge of Fig. 14.3, the S = R.® =25x —°°=250Q.
values of resistances of various arms are P=1000£l, P 100
Measurement of Resistance 431
Internal resistance of bridge looking into Example 14.10 A regular Wheatstone bridge is used to
terminals b and d. measure high resistances (in the megohm range). The bridge
has ratio arms of 10,000 Q and 10 Q.
p KS PQ
R+S P+Q The adjustable arm has a maximum value of 10,000 Q. A
battery of 10 V emf and negligible resistance is connected
25x250 100x1000
=---------- +-------------- = 113.6 Q. from the junction of ratio aims to the opposite corner ?
25 + 250 100 + 1000
(a) What is the maximum resistance that can be
From Eqn. 14.30, the deflection for a small change measured by this arrangement ?
in resistance (b) If the galvanometer has a sensitivity of
_ S. ESAR 200 mm/pA and a resistance of 50 Q, how much
“ (R,, + G)(R + S)2 unbalance is needed to give a galvanometer
deflection ofl mm for the maximum resistance of
Ratio of deflections for two galvanometers part (a) ?
0^= (Sl)4ESAR / (Sl)besar (c) If the galvanometer of part (b) is replaced by a
0b (% + Gx)(R + S)2 / (Ro + Gb)(R + S)2 galvanometer of sensitivity of 1000 mm/pA and a
resistance of 1000 Q, calculate the change in
_ (^i)x ^0 + ^B
resistance to cause a deflection of 1 mm.
Solution. The arrangement of the bridge is shown
200 (113.6 + 600) „ „„ in Fig. 14.3.
=----- x ' = 1.75
500 (113.6 + 50) Now we have,
Thus galvanometer A has a sensitivity of unknown resistance, R = (P / Q) x S.
1.75 times that of galvanometer Bas far as this bridge The maximum value of
is concerned, even though on its own galvanometer A R = (maximum value of P / Q)
is less sensitive than galvanometer B. x (maximum value of S)
Example 14-9 A highly sensitive galvanometer can
detect a current as low as 0.1 nA. This galvanometer is used
in a Wheatstone bridge as a detector. The resistance of (b) Internal resistance of bridge
galvanometer is negligible. Each arm of the bridge has a
resistance ofl k£l. The input voltage applied to the bridge is
20 V. Calculate the smallest change in resistance which can
be detected. The resistance of the galvanometer can be 10xl06xl0xl03 10xl03xl0
neglected as compared with the internal resistance of bridge. ~ 10xl06 + 10xl03 10xl03 + 10
Solution. Now P = Q = R = S = lkQ = 1000Q. « 10,000 Q
It is a bridge with equal arms. From Eqn 14.30, change in resistance needed for
.’. The internal resistance of bridge 1 mm deflection
a;; etRo + GMR + S)2
Rq = R = 1000Q (See Eqn. 14.27)
SiES
Let the change in resistance be AR. 1 x (10000+ 50)(10x106 +10 xlO3)2
.'. Output voltage of bridge due to unbalance 200 x 106 x 10 x 103
Eo = EAR/4R « 0.5 MQ.
= 20x AR/(4xl000) = 5xl0“3 AR (c) The change in resistance to cause 1 mm
deflection with galvanometer having a resistance,
Current through the galvanometer
G = 1000 0
En 5x10~3AR _ __9
and current sensitivity of S(. = 1000 mm/g A is
1 x (10/100 +1000)(10 x 106 +10 x 103)2
:. The smallest change in resistance which can be ^R— 7 q
1000xl06xl0xl03
detected
AR=20x10-6Q =20 gQ. «0.1 MQ
432 Electrical and Electronic Measurements and Instrumentation
P _____RR.
« p
1200x0.75 900
—_________________ =-----------------------
Fig. 14.11 Thevenin source generator of 10 RCl + R.+Rr 1200 + 075+ R,E 1200.75 + R,L
U C
Example 14.11.
R.R. 0.75 R, 0.75 R,
= resistance of circuit looking into terminals Ra +Rbh + RrL 1200 + 0.75+ R r 1200.75+R,
l
d & c with terminals a &ab short-circuited.
RS PQ 1x5 lxQ tj
R R,L______
_ _________ Cl
1200 R, 1200 R,
__________________ ________ —______________ —____
" R + S+P+Q’l + 5 + l + Q 30 " Ra + R.+R, 1200 + 0.75+ R,g 1200.75 + R.g
b L
= 0.833 + —— kQ.
1 +Q
Now, Rq + G = 24xlO-3/13.6xl06 = 1.765 kQ
Example 14
*12 A modified form of Wheatstone bridge is
shown in Fig 14.12. Calculate the value of unknown
resistance, Rx, if
r=1200Q, Ra=1600Rb,
/tj =800Rfc), R2 = 1.25 R^ and R3=0.5Rb Fig. 14.13 Delta-star transformation for
are the resistance values under balanced conditions. analysis of Fig- 14.12.
Measurement of Resistance 433
The modified circuit is shown in Fig- 14.14. Current through the series resistor
= 2x50xl0-3 =100xl0-3 A.
Resistance of series resistor
R=- = 100 Q.
5 100 xlO-3
Example 14.14 In a Carey-Foster's bridge a resistance
of 1.0125 Q is compared with a standard resistance of
1.0000 Q, the slide wire has a resistance of 0.250 Q in
100 divisions. The ratio arms nominally each 10 Q, are
actually 10.05 and 9.95 Q respectively.
How far (in scale divisions) are the balance positions from
those which would obtain of ratio arms were true to their
Fig. 14.14 Configuration of bridge of Fig. 14.12 nominal value ? The slide wire is 100 cm long.
after Delta-star transformation. Solution. Balance with ratio arms equal to nominal
value. Let l^ be the distance of balance point on slide
For balance,
wire from the unknown resistance end in cm of slide
wire. Let rbe the resistance per cm length of slide wire.
*1 0.0250
r~ 100
1200 R. D 0-75 Rl
480 +------------ — Rr +----------- \ = 0.00025 Q/cm as length of wire is 100 cm.
1200.75 + Rl x 1200.75 + Rl
or In this case
600 0.375 ~
P = Q = 10Q, S = 1.0000 Q and R = 1.0125 Q.
or R = 0.3 Q.
Under balance conditions :
Example 14.13 A Wheatstone bridge is used for P_ R + Z/
measuring the value of change of resistance of a strain gauge Q~ S + (100-ZJr
which forms one of the arms of the bridge. All the arms of the
bridge including the strain gauge have a resistance of 100 Q 10 1.0125+ 0.00025L ,
or 1----- or Z. = 25 cm.
— =--------------------------- *
each. The maximum allowable power dissipation from the 10 1.0000 + 0.025-0.00025 \ 1
strain gauge is 250 mW. Determine the value of maximum
Thus the balance is obtained at 25 and 75 scale
permissible current through the strain gauge and maximum
divisions.
allowable value of bridge supply voltage. Suppose a source of
Balance arms equal to true values :
20 V is available, find the value of series resistance to be
connected between the source and the bridge to limit the Now in this case, P = 9.05 Q and Q = 10.05 Q.
input voltage of the bridge to permissible level. Under balance conditions :
Solution. The resistance of strain gauge R = 100 Q 9.95 _ 1.0125 + 0.00025 f
10.05 ~ 1.0000 + 0.025 -0JD00025ZJ
Suppose I is the current through each arm under
balanced conditions. 1-0.005 10125 + 0.00025 Z.
or -----------=---------------------- -
.-. I2R = P, 1 + 0.005 10250-0.00025/j
Hence, maximum permisible current Thus the balance is obtained at 5 and 95 cm.
resistance of 100 Q is being measured but the same 14.3.2 Kelvin Double Bridge Method of
contact resistance would cause an error of 10% if a low Measurement of Low Resistances
resistance of the value of 0.02 Q is measured. Hence The Kelvin bridge is a modification of the
special type of construction and techniques have to be Wheatstone bridge and provides greatly increased
used for the measurement of low resistances in order accuracy in measurement of low value resistances. An
to avoid serious errors occurring on account of the understanding of the Kelvin bridge arrangement may
factors mentioned above. PV be obtained by a study of the difficulties that arise
Low resistances are constructed with four in a Wheatstone bridge on account of the resistance of
terminals as shown in Fig. 14.15. One pair of terminals the leads and the contact resistances while measuring
CC' (called the current terminals) is used to lead low valued resistors.
current to and from the resistor. The voltage drop is Consider the bridge circuit shown in Fig. 14.15
measured between the other two terminals PF, called where r represents the resistance of the lead that
the potential terminals. connects the unknown resistance R to standard
resistance S. Two galvanometer connections indicated
by dotted lines, are possible. The connection may he
either to point '»/ or to point ‘F. When the galvano
meter is connected to point in, the resistance, r, of the
connecting leads is added to the standard resistance, S,
resulting in indication of too low an indication for
unknown resistance R. When the connection is made
to point n, the resistance, r, is added to the unknown
resistance resulting in indication of too high a value
for R.
Fig. 14.15 Ammeter voltmeter method for Suppose that instead of using point in, which
measuring 4 terminal resistances. gives a low result, or n, which makes the result high,
we, make the galvanometer connection to any
The voltage V, indicated in Fig. 14.15, is thus 1R intermediate point 'd' as shown by full line in
times the resistance R between terminals PF and does Fig. 14.16. If at point 'd' the resistance r is divided into
not include any contact resistance drop that may be two parts, r, and such that
present at the current terminals CC'. r, P
-=- ...(14.36)
Resistors of low values are thus measured in
terms of resistance, between potential terminals,
which becomes perfectly and precisely definite in
value and is independent of the contact resistance
drop at the current terminals. Contact resistance drop
at the potential terminals need not be a source of error,
as current crossing at these terminals is usually
extremely small or even zero for null methods. Also
this contact resistance now becomes a part of the
potential circuit and is, therefore, a negligible part of
the total resistance of the potential circuit since
potential circuits have a high value of resistance.
14.3.3 Kelvin Bridge Ohmmeter The circuit for the Thevenin equivalent resistance
as seen from galvanometer terminals c and d is shown
This is a modified form of the Kelvin bridge and is
in Fig. 14.19(a) where Rb is the resistance in the battery
intended for the rapid measurement of the winding
resistances of machines and transformers, and for the circuit.
measurement of contact and earth conductor resistances.
The accuracy is of the order of ± 0.2%. This instrument
is direct reading and the balance is obtained by rotating
a single dial.
Figure 14.18 gives the circuit diagram of a typical
Kelvin bridge ohmmeter. The ratio arms P/Q and p/q
of Fig. 14.17 are replaced by a combination of fixed
resistors P and p and a double slide wire enabling the
bridge ratio to be varied continuously between values
(n) Due to electrostatic effect, stray changes can 14.4.2 Use of Guard Circuit
appear in the measuring circuit causing errors. Some form of guard circuits are generally used to
Alternating fields can also effect the measurements eliminate the errors caused by leakage currents over
considerably. Therefore, critical points of the insulation. Figure 14.20 illustrates the operation of a
measuring circuit must be carefully screened. guard circuit. In Fig. 14.20(a), a high resistance
(iii) In order to obtain definite ratios in the mounted on a piece of insulating material is measured
potential distribution with respect to surroundings, by the ammeter-voltmeter method. The micro
one point of the circuit may be connected to earth for ammeter measures the sum of the current through the
accuracy in measurements. resistor (IR) and the current through the leakage path
(iv) In measurement of insulation resistance the around the resistor (IL). The measured value of
specimen often has considerable capacitance. On resistance computed from the readings indicated on
application of a direct voltage a large charging current the voltmeter and the micro-ammeter, will not be true
flows initially which gradually decays down after a value but will be in error. In Fig. 14.20(F) guard terminal
short interval. Further, insulating materials possess has been added to resistance terminal block. The
the property of dielectric absorption, i.e., after the guard terminal surrounds the resistance terminal
main charging current has decayed down, further entirely and is connected to the battery side of the
charge is slowly absorbed over a considerable period micro-ammeter. The leakage current IL, now bypasses
of time, perhaps for minutes or even hours. Thus the micro-ammeter which then indicates the current IR
measurement of true conduction current should be through the resistor and thus allows the correct
delayed until after the cessation of the charging and determination of the resistance value from the
absorbing currents. But since the absorbing currents readings of voltmeter and micro-ammeter. The guard
take a considerably long time to decay, it is usually terminal and resistance terminal are almost at the
inevitable that the conduction current measured same potential and thus there will be no flow of
includes some absorption in current. The testing current between them.
conditions, including the time between the application
of voltage and observation of the current, must be
specified.
(v) When measuring the resistance of low
conductivity conductors, insulating materials and
products, the effect of various factors upon their
resistance should be taken into account. Thus, a
change in the temperature of cardboard from 20° to
40°C is accompanied by a 13 fold change in its
resistance, changes in humidity form 10 to 60 percent
cause a 30 fold change in resistance of porcelain. Fig. 14.20 Application of guard circuit for
Besides temperature and humidity, the kind of current measurement of high resistance.
employed for measurement, the magnitude and
duration of the applied voltage, and other factors also 14.4.3 Methods for Measurement of
effect the resistance being measured. High Resistance
(vi) Fairly high voltages are used in tests in order The different methods employed are :
to raise the currents to reasonable values in order to be 1. Direct deflection method.
measured. So normally a sensitive galvanometer or 2. Loss of charge method:
micro-ammeter is required and adequate steps have to 3. Megohm bridge.
be taken to prevent damage to these delicate
4. Meggar (This is explained in Art. 9.8, page 256).
instruments.
A voltage supply of 100 V upto a few kV is often Direct deflection method. The direct deflection
used depending upon the nature and breakdown method is basically that of Fig. 14.20(F)- For high
voltage of the test object. The powe supply unit is d.c. resistances, such as insulation resistance of cables, a
transistorized source. Proper smoothing and sensitive galvanometer of 'd' Arsonval type (usually
stabilization circuits are used to ensure constancy of having a current sensitivity of at 1000 mm/pA at a
voltage with time. scale distance of 1 metre) is used in place of the
Measurement of Resistance 439
microammeter. In fact many sensitive type of replacing the insulation by a standard high resistance
galvanometers can detect currents from 0.1 — 1 nA. (usually 1 Mil), the galvanometer shunt being varied,
Therefore, with an applied voltage of 1 kV, resistances as required to give a deflection of the same order as
as high as 1012 to 10 x 1012 Q can be measured. before.
An illustration of the direct deflection method In tests on cables the galvanometer should be
used for measuring insulation resistance of a cable is short-circuited before applying the voltage. The short
shown in Fig. 14.21. The galvanometer G, measures the circuiting connection is removed only after sufficient
current IR between the conductor and the metal time has elapsed so that charging and absorption
Sheath. The leakage current IL, over the insulating currents cease to flow. The galvanometer should be
material is carried by the guard wire wound on the well shunted during the early stages of measurement,
insulation and therefore does not flow through the and it is normally desirable to include a protective
galvanometer. series resistance (of several megohm) in the
galvanometer circuit. The value of this resistance
should be subtracted from the observed resistance
value in order to determine the true resistance. A high
voltage battery of 500 V emf is required and its emf
should remain constant throughout the test.
Measurement of volume and surface
resistivity. The direct deflection method is often used
for measurement of insulation resistance of insulating
material samples available in sheet form. In such cases
we are interested in the measurement of volume
Fig. 14.21 Measurement of insulation resistance
resistivity and the surface resistivity of the material.
of cable having sheath.
Figure 14.23 shows the schematic diagram for
Cables without metal sheaths can be tested in a measurement of volume and surface resistivities of a
similar way if cable, except the end or ends on which specimen of insulating material. The specimen is
corrections are made, is immersed in water in a tank. provided with tin foil or colloidal graphite electrodes;
The water and the tank then form the return path for the upper electrode having a guard ring. For
the current. The cable is immersed in slightly saline measurement of volume resistivity (which in fact is
water for about 24 hours and the temperature is kept the specific resistance) readings of voltage applied and
constant (at about 20°C) and then the measurement is the current through the galvanometer are taken.
taken as in Fig. 14.22. Leakage currents over the edge of the specimen will
The insulation resistance of the cable R = V / ID. flow between the guard ring and the lower electrode
and hence will not introduce error into the
measurement. The volume resistivity, p, can be
calculated as follows :
Let t/j = diameter of the upper electrode d^,
r = thickness of the specimen sheet,
V1 = reading of voltmeter,
and Ij = current through galvanometer G,.
In some cases, the deflection of the galvanometer Fig.14.23 Measurement of volume and surface resis
is observed and its scale is afterwards calibrated by tivities of insulating material specimen.
440 Electrical and Electronic Measurements and Instrumentation
Fig. 14.27 Loss of charge method considering effects Fig. 14.29 Use of guard circuit for measurement
of leakage resistance of capacitor. of high resistances.
The test is then repeated with the unknown resis However if the same resistance is measured by a
tance R, disconnected and the capacitor discharging modified Wheatstone bridge as shown in Fig. 14.29(b)
through Rr The value of R1 obtained from this second with the guard connection G connected as indicated,
test and substituted into the expression the error in measurement is considerably reduced. For
RRa the arrangement shown in Fig. 14.29(b) resistance RBG
...(14.52)
R + R^ is put in parallel with the galvanometer and thus it has
no effect on the balance and only effects the sensitivity
in order to get value of R.
of the galvanometer slightly. The resistance
The leakage resistance of the voltmeter, unless Rag =100MQ is put in parallel with a resistance
very high should also be taken into consideration. P = 100 kQ and therefore for the arrangement shown
442 Electrical and Electronic Measurements and Instrumentation
the measured value has an error of only 0.01 percent Let us consider an annular ring of width dx at a
and this error is entirely negligible for measurements radius x from the centre as shown in Fig- 14.31.
of this type. Insulation resistance of this annular ring is :
The arrangement of Fig. 14.30 illustrates the
operation of a Megohm bridge.
Figure 14.30 shows the circuit of a completely
self-contained Megohm bridge which includes power
supplies, bridge a members, amplifiers, and
indicating, instrument. It has a range from 0.1 MQ to
106 MQ. The accuracy is within 3% for the lower part
of the range to possible 10% above 10,000 MQ.
Dr2 pdX
d/2 2 nxL
p , D 2.303p D
2 tcL d 2 tiL a
0.367 , D
Fig. 14.30 Megohm bridge. = —pl0S>»7
Sensitivity for balancing against high resistance is Now in this example we have,
obtained by use of adjustable high voltage supplies of D = 25xl0-3m ; d = 5xl0~3 m ;
500 V or 1000 V and the use of a sensitive null L = 1 km = 1000 m and R = 16,000 Q.
indicating arrangement such as a high gain amplifier
Specific resistance
with an electronic voltmeter or a C.R.O. The dial on Q
is calibrated 1 -10 - 100 -1000 MQ, with main decade RL 16000x1000
p - ---------------------
1-10 occupying greater part of the dial space. Since 03671og10D/d 0367(logw2.5 /0.5)
unknown resistance R = PS/ Q, the arm Q is made, = 62.5xl0-6 Qm.
tapered, so that the dial calibration is approximately
logarithmic in the main decade, 1-10. Arm S gives Example 14.18 A test voltage is applied for several
five multipliers, 0.1, 1, 10, 100 and 1000. minutes between the conductor of a 400 metre length of
cable and earth. The galvanometer connected in series reads
The junction of ratio arms P and Q is brought on
250 divisions, the value of universal shunt being 2.5 with a
the main panel and is designated as 'Guard' terminal.
standard resistance ofl MQ in circuit, the scale reading is
Example 14^7 Derive an expression for insulation 350, the value of shunt being 1000. Calculate the insulation
resistance of single core cable. The conductor of a cable has a resistance of the cable. What would be the insulation
diameter of 5 mm and the overall diameter of the cable is resistance of the same cable of length 100 metre ?
25 mm. If the insulation resistance of the cable is 16,000
Solution. Since we have a universal shunt with
Tl/km, calculate the specific resistance of insulating material.
the galvanometer and therefore the current through
Solution. Let the circuit is proportional to the deflection of the
d = diameter of conductor, galvanometer times universal shunt multiplier,
D = diameter of insulated cable, Deflection of galvanometer with unknown resistance
L= length of cable, and ‘R’ is circuit 350 divisions and the shunt multiplier
p = resistivity of insulating material. is 2.5.
Measurement of Resistance 443
then becomes constant, and when the auxiliary the other as the commutator rotates. The second pair
electrode A approaches the auxiliary electrode B, the of each of set of brushes is positioned on the commu
resistance rises again. The placing of electrodes is thus tator so that continuous contact is made with one
very important and serious error may be caused by segment whatever the position of the commutator.
incorrect placing of the electrodes. The correct value of The earth tester has four terminals Pv P2. Two
resistance of earth, RE, is when the auxiliary electrode terminals P] and CT are shorted to form a common
Ais at such a distance that the resistance lies on the flat point to be connected to the earth electrode. The other
part of curve of Fig. 14.30. two terminals P2 and C2 are connected to auxiliary
The spacing between the earth electrode E and the electrodes P and C respectively.
auxiliary electrodes A, B should be large so as to get The indication of the earth tester depends upon
proper results. The distance may be a few hundred the ratio of the voltage across the pressure coil and the
metres in case the earth resistance is low. current through the coil. The deflection of its pointer
2. Earth tester. The resistance of earth can be indicates the resistance of earth directly. Although the
measured by an earth tester shown in Fig. 14.35. The "Earth Tester", which is a permanent magnet moving
"Earth Tester" is a special type of Meggar (See Art. 9.8 coil instrument and can operate on d.c. only, yet by
page 256) and it has some additional constructional including the reverser and the rectifying device it is
features additional constructional features and they possible to make measurements with a.c. flowing in
are : the soil.
(i) a rotating current reverser, and The sending of a.c. current through the soil has
(n) a rectifier many advantages and therefore this system is used.
The use of a.c. passing through the soil eliminates
unwanted effects due to production of a back emf in
the soil on account of electrolytic action. Also the
instrument is free from effects of alternating or direct
currents presents in the soil.
Example 14.21 On a 250 V supply a fault having a
resistance of 200 develops between the unearthed end of the
winding of an electric heater and the frame. If the resistance
of the substation earth electrode is 4 O that of human body
2000 0, and the safe maximum current through the body is
25 mA, what is the safe maximum resistance of consumer's
earth electrode ?
Solution. Figure 14.36 shows the diagrammatic
representation of the problem.
Both these additional features consist of simple Fig. 14.36 Diagram of Example 14.20.
commutators made up of 'L' shaped segments. They
Let R be the resistance of earth electrode at the
are mounted on the shaft of the hand driven generator.
consumer end.
Each commutator has four fixed brushes. One pair of
each set of brushes is so positioned that they make Voltage drop across the human body
contact alternately with one segment and then with = 25x10“3x2000 = 50 V.
446 Electrical and Electronic Measurements and Instrumentation
Voltage between points a and b In both cases, the loop circuit formed by the cable
= 250-50=200 V. conductors is essentially a Wheatstone bridge consis
r 200 200 . ting of resistances P, Q, R and X. G is a galvanometer
Current I =------- =----- A. for indication of balance.
20 + 4 24
Current through human body
= I------ ---- =25xl0-3 A.
R+2000
200 R = 25xl0~3
or
24 R + 2000
or 200 R = 0.6 + 1200
or resistance of earth electrode on consumer end R « 6Q.
The methods used localizing ground and short The resistors P, Q forming the ratio arms may be
circuit faults differ from those used for localizing open decade resistance boxes or slide wires.
circuit faults.
In the case of multicore cables it is advisable, first Under balance conditions :
of all, to measure insulation resistance of each core to
earth and also between cores. This enables us to sort R P R+X P+Q
out the core that is earthed in case of ground fault ;
and to sort out the cores that are shorted in case of a X = -^(R + X) ...(14.53)
short circuit fault. Loop tests are used for location of
ground and short circuit faults. These tests can only be where (R + X) is total loop resistance formed by the
used if a sound cable runs along with the faulty cable sound cable and the faulty cable. When the conductors
or cables. The loop tests work on the principle of a have the same cross-sectional area and the same
Wheatstone bridge. The advantage of these tests is resistivity, the resistances are proportional to lengths.
that their set up is such that the resistance of fault is If I, represents the length of the fault from the test end
connected in the battery circuit and therefore does not and I the length of each cable.
effect the result. However, if the fault resistance is
high, the sensitivity is adversely affected. In this Then
section only two types of tests viz., Murray and Varley
loop tests are being described.
'^TTq'2' -<14-54)
Murray loop test. The connections for this test The above relation shows that the position of the
are shown in Fig. 14.37. Figure 14.37(a) relates to the fault may be located when the length of the cable is
ground fault and Fig. 14.37(b) relates to the short known. Also, the fault resistance does not alter the
circuit fault. balance condition because its resistance enters the
Measurement of Resistance 447
Solution. Let X be the resistance of cable from the Example 14.24 A short circuit fault is located by Varley
test end to place of fault and R + X be the total loop test. The circuit of Fig. 14.38 (b) is usedfor the purpose.
resistance of loop. P and Q are the ratio arms. The ratio arms are set at P = 5£l and Q=10Q and the
X/R = Q/P values of variable resistance S are 16 Q for position 1 of
switch Kand 7Q for position 2. The sound and faulty cables
” x = 77q(X+r) are identical and have a resistance of 0.4 Cl/km. Determine
the length of each cable and the distance offault from the test
41 2
= ^(X+R) = 0.292 (X+R) end.
Solution. Let Sj be the value of resistance S2 for
Let /1 be the distance of fault from test end / be
the length of each cable and r resistance per unit position 1 and S2 for position 2.
length. R+X _ P
X = rf and X + R = 2rl S. ~Q
:. Distance of fault from test end : Hence resistance of loop
f =0.292x2x5.2 =3.03 km. P 5
R + X = —S. = —xl6 = 8Q.
Limiting fractional error of unknown resistance Q 1 10
8X_ fSP? ( 5P + 6Q^_ If ±0-5 Y f±0.5?
Resistance of each cable = 8/2=40
x " p J \ p+q J “ vvloo-) A wo~J
.'.Length of each cable = 4/0.4 = 10 km.
= 0.707%. At positron 2 we have,
Therefore limiting error of distance P _ P
= 3.03 x 0.7071100 x 1000 = 21.4 m. X+S2 " Q
and X + R = rj + r2l
1.1 ! 2.92
1000 + 1000 J x 520 = 1.76 Q.
Review Questions
1. (a) Classify the resistances from the point of view 8. Draw the circuit of a Kelvin-Varley slide and
of measurements. explain its working and advantages.
(b) Describe in brief the different methods used for
9. What are the different problems associated with
measurement of medium resistances.
measurement of low resistances ? Explain the
2. Describe the ammeter-voltmeter method of
principle of working a Kelvin's Double Bridge and
measurement of resistances. There are two ways in
explain how the effect of contact resistance and
which the circuit of ammeter voltmeter method
resistance of leads is eliminated.
can be used
10. Draw the circuit of a Kelvin's Double Bridge used
(i) ammeter connected to the side of unknown
for measurement of low resistances. Derive the
resistance and
condition for balance.
(ii) voltmeter connected to the side of unknown
11. What are the different difficulties encountered in
resistance.
the measurement of high resistances ? Explain how
Derive the condition which decides which circuit is
these difficulties are overcome.
to be used for a particular set of ammeter,
12. Explain the loss of charge method for
voltmeter and unknown resistance. Assume equal
measurements of insulation resistance of cables.
relative error in both the cases.
13. Differentiate between volume and surface
3. Describe the substitution method of measurement
resistivities. Explain the method of measuring
of medium resistances. List the factors on which
them.
the accuracy of the method depends.
14. Describe the working of a Megohm bridge.
4. Draw the circuit of a Wheatstone bridge and derive
the conditions of balance. 15. What is the importance of the value of Earth's
5. Derive the expression for bridge sensitivity for a resistance. What are the factors which influence its
Wheatstone bridge with equal arms. Find also the value ? Describe the fall of potential method for
expression for current through the galvanometer measurement of earth resistance.
for a small unbalance. 16. Describe the construction and working of an Earth
6. What are the different factors which affect the Tester. Explain how it can be used for measure-
I
Unsolved Problems
1. A voltmeter of resistance 500 Q and a milliammeter The galvanometer has a resistance of 100 Q, a
of 1.0 Q resistance are used to measure a resistance sensitivity of 100 mm/pA and is connected across
by ammeter voltmeter method. If the voltmeter AC. A source of 4 V d.c. is connected across BD.
reads 20 V and milliammeter 100 mA. Calculate the current through the galvanometer
and its deflection if the resistance of ami DA is
Calculate the value of measured resistance
changed from 400 Q to 401 Q.
(i) if the voltmeter is put across the resistance and
the milliammeter connected in series with the [Ans. 1.63 pA ; 16.3 mm]
unknown resistance 6. In a Wheatstone bridge, the ratio arms AB = 10 Q
(if) if the voltmeter is put across the unknown and BC = 100Q ; standard-resistance across
resistance with ammeter connected on the CD = 10 Q. The shunt across 10 Q ratio arm has to
be changed from 22310 to 27670 Q, when the
supply side.
resistor was changed for Rj in the arm DA.
[Ans. 199 Q ; 333 Q]
Calculate in magnitude the difference between the
2. In a laboratory a voltmeter of 200 Q resistance and resistances of Rj and R^. The bridge is balanced in
an ammeter of 0.02 Q resistance are available. both the cases. [Ans. 86.8 pQ]
Calculate the value of resistance that can be
measured by the ammeter voltmeter method for 7. A modified Wheatstone bridge network is
which the two different circuit measurements give constituted as follows :
equal errors. [Ans. 2 Q] AB is a resistance P in parallel with resistance p;
3. In a measurement of resistance by the substitution BC is a resistance Q in parallel with a resistance q;
method a standard resistance of 100 kQ is used. CD and DA are resistances R and S respectively.
The galvanometer has a resistance of 2000 Q, and
The nominal values of P, Q and S are each 10 Q.
gives the following deflections :
With resistance R in circuit, balance is obtained
(i) With unknown resistance : 46 divisions, with p = 30,000 Q and q = 25,000. With R replaced
(n) With standard resistance : 40 divisions. by a standard resistance of 10 Q, balance is
obtained when p = 15,000 Q and q = 40,000 Q.
Find the value of unknown resistance
Calculate the value of R. [Ans. 9.99952 Q]
[Ans. 86.7 kQ]
8. A Wheatstone bridge is used to measure the
4. The four arms of a Wheatstone bridge are as resistance of a resistor. The bridge was balanced
follows : with the values shown in Fig. 14.40. It is found that,
AB = 100Q; BC= 10Q; due to presence of chemical impurities emfs are set
CD = 4Qand DA = 50Q.
The galvanometer has a resistance of 20 Q and is
connected across BD. A source of 10 V d.c. is
connected across AC Find the current through the
galvanometer. What should be the resistance in the
arm DA for no current through the galvanometer ?
[Ans. 5.15 mA ; 40 Q]
5. The four arms of a Wheatstone bridge are as
follows :
AB= 100Q;
BC=1000Q;
CD = 4000 Q and
Fig. 14.40 Diagram of Problem 14.8.
DA = 400 Q.
Measurement of Resistance 451
Outer ratio arm 100 Q and 1000 Q ; The further set of readings was taken with resistor
Inner arms, ratio 99.92 Q and 1000.6 Q ; R removed from circuit, as follows :
voltmeter is given as 50 MQ. 1.80 Q. Find the resistance of fault and its distance
from test end.
13. The following observations were made for a loss of
17. In a Murray loop test for ground fault on a
charge method for the determination of a high
resistance R. The charged capacitor of capacitance 500 metre long cable having a resistance of
1.6 Q/km, the faulty cable is looped with a sound
12.5 pF was connected across an electrostatic
! cable of same length and cross-section. If
voltmeter and R in parallel and the voltage
resistances of ratio arms are 3 : 1, calculate the
measured after intervals of time.
distance of the fault from the test end. [Ans. 250 m]
452 Electrical and Electronic Measurements and Instrumentation
6. A Kelvin-Varley Slide consists of 4 decade (c) cables have a low value of capacitance which
dividers. The first decade is constructed by having draws a high value of charging current
11 coils of 10 kQ resistance each. The subsequent (d) none of the above.
decades will have coils of :
11. A circular piece of specimen has a surface
(fl) 11 coils of 20 kQ each, 11 coils of 40 kQ each, 11 resistance Rg. Its diameter is d and the thickness is
coils of 80 kQ each
t. The surface resistivity ps of the specimen is given
(b) 11 coils of 10 kQ each, 11 coils of 5 kQ each, 10 by:
coils of 1 kQ each
7td2Ks
(c) 11 coils of 2 kQ each, 11 coils of 400 Q each, 11 («)
t
coils of 80 Q each
*dRs
(d) 11 coils of 2 kQ each, 11 coils of 400 Q each, 10
t
coils of 80 Q each.
(d) to eliminate the effect of leads. (b) depth to which electrode is driven into earth.
8. In a Kelvin's Double Bridge two sets of readings (c) specific resistance of soil
are taken when measuring a low resistance, one (d) all the above.
with the current in one direction and the other with 13. From the point of view of safety, the resistance of
direction of current reversed. earthing electrode should be :
This is done to (a) low
(a) eliminate the effect of contact resistance (b) high
(b) eliminate the effect of resistance of leads (c) medium
(c) correct for changes in battery voltage (d) the value of resistance of earth electrodes does
(d) eliminate the effect of thermo-electric emfs. not affect the safety.
14. The advantage of Varley loop tests over Murray
9. High resistances are provided with a guard
loop tests is
terminal.
(a) they can be used for localizing of short circuit
This guard terminal is used to :
fauits
(fl) bypass the leakage current
(b) they can be used for localizing of earth fault
454 Electrical and Electronic Measurements and Instrumentation
(c) the loop resistance can be experimentally deter (a) affects the balance conditions
mined
(fe) affects the value of cable resistance
(d) their accuracy is higher.
(c) affects the sensitivity of the bridge
15. When localizing ground fault with the help of loop
(d) all the above.
tests, the resistance of the fault:
\ 5* \ 3 ’C sf ? M •
l.(c) X(t) 3. (a) 4. (c) (S.(d) 6- (rf) 7. (fe) 8. (d) 9. (a) 10. (fe)
11. (!>) 12. W 13. (a) 14. (c) 15. (c)