Professional Documents
Culture Documents
Concrete Tests
Concrete Tests
Concrete Mechanics
The specific gravity of cement value will be determined by using the Le-
Chatelier Flask Method. It is a lab test experiment to identify the value.
Required Apparatus
● Cement
● Kerosene
● OPC Cement
● Le-Chatelier Flask of 250ml Or Pycnometer
● Weighing Balance (Accuracy of 0.1 gm)
● Ensure that the flask is free from dust and impurities. It should be
thoroughly dried.
● Now weigh the empty flask at weighing balance and note down the
reading as W1.
● Next, fill the cement into the flask about 50 gm and note down the
weight as W2.
● Now fill the flask with kerosene up to the top of the bottle.
● Mix the contents of the flask well and remove any air bubbles. Note
down the weight as W3.
● Now empty the flask and fill the kerosene fully (up to the top) and
weigh it as W4.
Notes:
Principle of air permeability method is in observing the time taken for a fixed
quantity of air to flow through a compacted cement bed of specified dimension
and porosity. Under standardized conditions the specific surface of cement is
proportional to where t is the time for a given quantity of air to flow through the
compacted cement bed. The number and size range of individual pores in the
specified bed are determined by the cement particle size distribution which
also determines the time for the specified air flow.
Required Apparatus
● Blaine apparatus
Test Procedure
● Place the perforated disc on the ledge at the bottom of the cell and
place on it a new filter paper disc. Place the weighed quantity of
cement, m1, in the cell. Place a second new filter paper disc on the
leveled cement. Insert the plunger and press it gently but firmly until
the lower face of the cap is in contact with the cell. Slowly withdraw
the plunger, rotate it through 90 ° and press once again. The bed is
now compacted and ready for the permeability test.
● Test is performed on the Blaine apparatus. It is practically a
manometer in the U-tube form. One arm of the manometer is
provided at the top with a conical socket to form an airtight fit with
the conical surface of the cell. The same arm has four etched lines
M1 to M4 and T-joint, which lead to an airtight stopcock beyond
which is attached an aspiration rubber bulb.
● Manometer is filled to the level of the lowest etched line with non-
volatile, non-hygroscopic liquid of low viscosity and density (such as
dibutyl phthalate or light mineral oil).
● Insert the conical surface of the cell into the socket at the top of the
manometer.
● Open the stopcock and with gentle aspiration raise the level of the
manometer liquid to that of the highest etched line. Close the
stopcock and the manometer liquid will begins to flow. Start the
timer as the liquid reaches the second etched line and stop it when
the liquid reaches the third etched line. Record the time, t, and the
temperature, T.
● The procedure repeats three times
● Calculate three values of the specific surface and the mean of
them.
Apparatus Required
Test Procedure
● Fix and place the Vicat apparatus on a flat surface and make
necessary adjustments.
● Now take 400gm of cement specimen and add 28% of water by the
weight of cement and mix gently for up to 3 to 5 minutes.
● Then fill the cement paste into the Vicat mould and remove the
excess cement paste by the trowel.
● Now place the Vicat mould on the Vicat apparatus and release the
plunger gently by contacting the cement paste surface.
● Then release the plunger and allow it to penetrate the cement paste
and note down the reading from the gauge scale from the bottom of
the Vicat mould.
● Now again, add water with cement paste at a different water ratio
until the reading lies between 5mm and 7mm.
Materials required
● Cement (say OPC)
● Sand (standard grade)
● Water (for having a standard consistency)
● The ratio of cement and sand that should be taken to carry out the
test should be 1:3.
● Let us take 200 gm of cement, then 600 gm of sand should be
taken.
● They should be mixed thoroughly for 1 minute.
● Then the calculation of water required to make a standard
consistency is calculated using the formula:
Then, we have P = 30 for OPC cement. Putting the value of P in the above
equation, we get the amount of water = 84 gm.
● The three cubes each for third, seventh and 28th days are then
tested for compressive strength.
● The weight of cubes is measured and data is recorded.
● The cubes are placed in the compression testing machine and
load is applied in the rate of 35 N/mm²/min.
● The load at which the cube is crushed is noted.
Then compressive strength for each cube on their respective test day is
calculated using the formula:
● The average compressive strength of three cubes is the final
compressive strength of the cement.
Apparatus required
● Vicat apparatus
● Weighing machine (with the accuracy of ±1 gm in every 1000gm)
● Trowel
● Stopwatch
Procedure
Procedure
1. The Le- chatlier mould and the glass plates are lightly oiled before
conducting the test
2. Prepare a cement paste as in consistency test with 0. 78 times the
water required to give a paste of standard consistency
3. Fill the cement paste in the Le- chatelier mould taking care to keep the
edges of the mould gently together during the operation.
4. Cover the mould with another piece of a glass plate and place a small
weight over the cover plate.
5. Submerge the whole assembly immediately in water at a temperature of
27o+-2oC and keep it there for 24 hours.
6. Take out the assembly again in water at 27+/- 2 deg C. The distance
between the indicator points are measured as A.
7. Submerge assembly again in water at 27 +/- 2 deg C
8. Bring the water to boiling in 25 to 30 minutes and keep at boiling for 3
hours. The assembly should be immersed in water during this process.
9. Remove the mould from water and allow it to cool to 27 +/- 2 deg C
10. Measure the distance between the indicator points as B.
Apparatus required
● Calorimeter
● Mortar and pestle
● Glass/Plastic vials
● Stopwatch or timer
● Sieve
● Muffle furnace
● Weighing balance
Heat of hydration = Heat released from hydrated cement – Heat released from
unhydrated cement. For low heat cement, The heat of hydration for 7 days
should not be greater than 66 cal/g and for 28 days should not be greater than
75 cal/gm.
Apparatus required
● Cement
● Balance
● Sieve
● Nylon bristol brush
Apparatus required
● Water – Either free water, water tied into hydrates (water of
crystallization) or hydroxy compounds is quantified by drying at 110C for
120 minutes
● Organics – Any organic matter is either volatilized or oxidized by heating
up to 550C
● Inorganic degradation – Carbonates and iron oxidation as well as some
other species that degrade at high temperatures can be identified by
heating up to 950-1000C
Apparatus required
● A PVC cassette (unweighed)
● PVC cassette (pre-weighed)
● Dust sampler that meets ISO and CEN criteria
Apparatus required
● Eye protection (note 2)
● Safety screens (note 3)
● Filter paper, 11 cm diameter
● Pipeclay triangle
● Tripod
● Plastic beaker or thick-walled glass beaker, 1 dm3 (must fit between
tripod legs)
● Sand (see diagram)
● Heat-resistant mats
● Plastic magnetic retriever or small bar magnet
Chemicals
For the thermite mixture:
● Aluminium powder (medium grade) (HIGHLY FLAMMABLE), 3 g
● Iron(III) oxide, 9 g
For the igniter:
● Domestic sparkler, 16 cm long (sold for indoor use)
Apparatus
● Platinum crucible
● Muffle furnace
● Glass rod with a flattened end
● Heating device
● Weighing balance
Materials
● Cement sample
● Water
● Hydrochloric acid
● Stannous chloride solution
● Saturated mercuric chloride solution
● Manganese sulfate solution
● Standard solution of potassium permanganate
Test Procedure
● Take 1 g of sample.
● Mix 40 ml cold water with the sample and stir vigorously. While stirring,
add 50 ml of hydrochloric acid.
● Heat the solution if required. Ground the cement with the flattened end
of the glass rod till the cement is completely decomposed.
● Heat the solution to the boiling point and stir continuously. While stirring,
add drops of stannous chloride solution till the solution is decolorized.
● Add a few more drops of stannous chloride solution and allow it to cool
at room temperature.
● After cooling, add 15 ml saturated solution of mercuric chloride & 25 ml
solution of manganese sulfate.
● Titrate the above solution with a standard solution of potassium
permanganate till pink color is observed.
● Calculate the iron as ferric oxide.
● Calcium Oxide test (Gravimetric method)
IS: (4032)
Apparatus
● Evaporating dish
● Beaker
● Filter paper
● Muffle furnace
● Weighing balance
Materials
Test Procedure
● Add hydrochloric acid to the combined filtrate reserved from ferric oxide
and alumina test to acidify it.
● Then evaporate it to a volume of 100 ml.
● Add 40 ml of saturated bromine water to the hot solution.
● Introduce ammonium hydroxide till the solution is undoubtedly alkaline.
● Boil the solution for not less than 5 minutes ensuring the solution stays
alkaline all the time.
● Settle the precipitate, filter it, and then wash it with hot water.
● Wash the beaker and filter with nitric acid and then with hot water.
● Remove any manganese dioxide precipitate left.
● Add hydrochloric acid to the filtrate and boil it till the bromine is ousted.
● Whilst the solution is still boiling, add 25 ml of ammonium oxalate
solution.
● Stop boiling once the precipitated calcium oxalate takes on a well-
defined granular form.
● Filter the settled precipitate.
● Wash 1 g of precipitate with 1 liter of ammonium oxalate solution.
● Reserve the filtrate & washings for the chemical test of cement of –
magnesia.
● Heat the precipitated lime at 1100-1200 °C.
● After cooling, weigh the lime.
Calculation
Apparatus
● Beaker
● Stirring rod
● Muffle furnace
● Weighing balance
Materials
● Cement sample
● Ice water
● Hydrochloric acid
● Ammonium hydrogen phosphate solution
● Ammonium hydroxide
● Reagent
Test Procedure
● Acidify the filtrate reserved from the calcium oxide test with the help of
hydrochloric acid to a concentration of about 150 ml.
● Add 10 ml of ammonium hydrogen phosphate solution in the amount of
250 g/l.
● Cool this solution by placing it in ice water.
● After cooling, stir the solution and drop ammonium hydroxide till the
formation of crystalline magnesium ammonium phosphate starts.
● After that, add reagent in a moderate excess amount and carry on
stirring.
● Allow the solution to settle for not less than 16 hours in a cool
atmosphere.
● Then filter the precipitate and wash it with the ammonium nitrate wash
solution.
● Char the paper and then ignite the precipitate at 1100-1200 ° C till
constant weight is obtained. Burn carefully such that pyrophosphate
does not start melting.
● Calculate magnesia content by weighing the obtained magnesium
pyrophosphate.
Calculation
Apparatus
● Glass rod with a flattened end
● Filter paper
● Steam bath
● Muffle furnace
● Desiccator
● Platinum/Porcelain crucible
● Weighing balance
Materials
● Cement sample
● Water
● Hydrochloric acid
● Barium chloride
Test Procedure
● Add 25 ml cold water to 1 g cement sample and stir the mixture
vigorously.
● Add 5 ml hydrochloric acid whilst stirring the mixture.
● Heat the solution if required. Ground the cement with the flattened end
of the glass rod till the cement is completely decomposed.
● Dilute the solution to 50 ml.
● Filter the residue and wash it thoroughly with water.
● Put aside the residue along with filter paper. Take its weight as W4.
● Reserve the residue for the chemical test of cement of – insoluble
residue.
● Dilute the filtrate to 250 ml & heat it till boiling point.
● Drop 10 ml of hot barium chloride solution while boiling till the
precipitate is well formed.
● Keep aside the solution in the steam bath for 4 hours or overnight.
● Filter the precipitate and wash it thoroughly.
● Put the precipitate along with filter paper in a weighed
platinum/porcelain crucible.
● Char the filter paper without igniting it.
● Then ignite the precipitate to 800-900 ° C and cool it in a desiccator.
● Weigh the barium sulfate W5.
● Sulfuric anhydride content can be calculated from this weight of barium
sulfate.
Calculation
Apparatus
● Beaker
● Muffle furnace
● Crucible
● Filter paper
● Desiccator
● Weighing balance
Materials
Test Procedure
● Dissolve the residue from the sulfuric anhydride test with filter paper in
30 ml hot water & 30 ml of 2N sodium carbonate solution while
maintaining the constant volume.
● Heat the solution to a temperature just below boiling point. After
reaching this temperature, continue boiling for 10 minutes.
● Then filter the solution.
● Wash the filtered solution with dilute hydrochloric acid and then with hot
water till it is chloride-free.
● Take the residue in a crucible and ignite it at 900—1000 ° C
temperature and cool it afterward in a desiccator.
● Weigh this residue to obtain the weight of insoluble residue present in
cement.
Field tests of Cement
Points to follow
Cement should be used up within 90 days of its manufacture. The reason for
using the cement as soon as possible is as follows:
● High humidity
● High temperature
● Type of cement (the finer the cement is, the more susceptible it is to
moisture absorption)
In general, it is always recommended to use cement as soon as possible after
its manufacture and to store it properly to maintain its properties.
Point to note
Points to follow
● Smell of Cement
IS 1661 (1972)
Points to follow
● The smell of cement should be more leaning towards the neutral side.
Strong earthly smell indicates the presence of excessive clay or silt,
which make the concrete weaker in strength.
● If the smell is too strong or has a chemical odor, it could indicate that
the cement has been contaminated or has gone bad.
● However, it is important to note that smell is not a reliable indicator of
the quality or performance of cement, as the smell of cement can be
affected by a variety of factors such as the manufacturing process,
storage conditions, and humidity.
● Temperature of Cement
IS 1661 (1972)
Points to follow
Insert your hand in the cement bag. The cement should give a cool feeling.
Warmer temperature indicates that the hydration process has been
commenced in the cement.
● Texture of Cement
IS 1661 (1972)
Points to follow
● Take a pinch of cement from the bag and rub it between the forefingers
and thumb. It should be smooth and not rough.
● The texture of cement should be fine and powdery. Cement is made by
grinding and blending limestone, clay, and other minerals, so it should
have a consistent, fine texture that is free of clumps.
● If the cement has a rough or uneven texture, it could indicate that the
cement has not been properly ground or that it has been contaminated
by other materials like sand.
● Floating of Cement
IS 1661 (1972)
Points to follow
● Setting of Cement
IS 1661 (1972)
Points to follow
Points to follow
● Prepare a cement block of 1:6 (cement : sand) mix and 25 * 25 * 200 (W
* B * L) mm dimension.
● Then the cement block is immersed in water for 7 days.
● Put a weight of 34 kg above the cement and observe. The cement
should not show a failure sign.
Tests for checking water quality
● PH value test
IS 3025-11 (1983)
Procedure
● Rinse each test tube with the water sample. Gloves should be worn to
avoid skin contact with the water.
● Fill the tube to the 5mL line with sample water.
● While holding a dropper bottle vertically, add 10 drops of Wide Range
Indicator Solution.
● Cap and invert several times to mix.
● Insert the tube into the Wide Range pH Comparator. Hold the
comparator up to a light source. Match the sample color to a color
standard.
● Record the pH value.
● Wash your hands
● Nitrate Testing
IS 3025-11 (1983)
Procedure
● Fill the sample bottle with sample water. Use gloves if drawing the
sample by hand.
● Rinse and fill one test tube to the 2.5 mL line with water from the sample
bottle.
● Dilute to the 5 mL line with the Mixed Acid Reagent. Cap and mix. Wait
2 minutes.
● Use the 0.1 g spoon to add one level measure (avoid any 50-60 times in
one minute). Wait 10 minutes.
● Insert the test tube into the Nitrate Nitrogen Comparator. Match the
sample color to a color standard. Record the result as mg/L(ppm)
Nitrate Nitrogen (NO3-N). To convert to mg/Nitrate (NO3) multiply by
4.4.
● Place the reacted sample in a clearly marked container. Arrangements
should be made with toxic material handlers for safe disposal. Please
wash your hands after this water test is completed.
● Dissolved Oxygen Testing
IS 3025-11 (1983)
Procedure
Procedure
Procedure
● Place the glass fiber filter on the membrane filter apparatus or insert it
into the bottom of a suitable Gooch crucible with the wrinkled surface
facing up.
● Apply vacuum while washing the disc with three successive 20 mL
volumes of distilled water.
● Continue applying vacuum even after the water has passed through to
remove all traces of water.
● Remove the filter from the membrane filter apparatus or both the
crucible and filter if a Gooch crucible is used. Dry it in an oven at 103-
105°C for one hour.
● Transfer the dried filter to a desiccator and store it until needed. Repeat
the drying cycle until a constant weight is obtained (weight loss is less
than 0.5 mg). Weigh the filter immediately before use.
● When handling the filter or crucible/filter, use forceps or tongs only.
Process of filtration
● Assemble the filtering apparatus and initiate suction.
● Wet the filter with a small volume of distilled water to ensure it is
properly seated against the fritted support.
● Vigorously shake the sample and quantitatively transfer the
predetermined sample volume determined in step 7.2 to the filter, using
a graduated cylinder.
● Continue applying vacuum to remove all traces of water after the
sample has passed through.
● While maintaining suction, wash the graduated cylinder, filter, non-
filterable residue, and the walls of the filter funnel with three portions of
distilled water, ensuring complete drainage between each wash.
Remove all traces of water by continuing to apply vacuum after the
water has passed through.
● Carefully detach the filter from the filter support, or alternatively, remove
the crucible and filter from the crucible adapter.
● Dry the filter for at least one hour at 103-105°C. Allow it to cool in a
desiccator and then weigh it.
● Repeat the drying cycle until a constant weight is obtained (weight loss
is less than 0.5 mg).
● Calculate the non-filterable residue.
● Chloride test
IS 3025 (Part 32)
Apparatus:
● Burette
● conical flask
● Pipette
● measuring cylinder
Reagents:
● Potassium chromate indicator solution
● Standard silver nitrate as a titration agent..
Procedure:
Apparatus required
● UV-Visible Spectrometer
● Sample tubes
● Standard flask
● Beaker
● Spatula
● Measuring Cylinder
● Wash Bottle
● Tissue Paper
Chemicals required
● Isopropyl Alcohol
● Glycerol
● Concentrated Hydrochloric acid
● Sodium Chloride
● Barium chloride
● Sodium sulphate
● Distilled water
Preparations of reagent
Conditioning reagent
Preparation of Blank,
● Take six 50 mL glass stoppered standard flask (four for standards, one
for the sample and one for the blank).
● Add 10 mL of the standard sulphate solution to the first standard flask,
20mL to the second, 30 mL to the third and 40 mL to the fourth.
● To the fifth standard flask add 20 mL of the sample water.
● The sixth standard flask is for the blank, to this standard flask add
distilled water alone.
● Add 5 mL of conditioning reagent to all the standard flasks.
● Then make up the volume to the 100 mL mark using distilled water.
1. Quantitative analysis
2. Spectrum analysis
3. Time course measurement
4. Fixed wavelength measurement
5. Absorbance transmittance meter
6. Environment
To measure the sulphate content select the fixed wavelength measurement
method. Double click fixed wavelength measurements, the system information
are transferred and then the following window will appear. Click parameter
option and select absorbance, fast response and enter starting value of the
wave length in nanometers. For this experiment load 420 nm as the starting
value. Click ‘add’ and enter the sample number, here it is 100.Then load the
number of cycles the analysis has to be made. Enter 1 for this case. Click ok.
The details of the entries will be displayed as shown. Note that the
absorbance is not zero so to reset the reading to zero, take distilled water in
the two sample tubes and place them in the chambers of the spectrometer.
Now click the button ‘auto zero’ the instrument
resets and shows 0.0000 as reading.
Testing of sample
Transfer blank to the sample tubes and place it in the chamber. Now click
‘Blank’, the value of absorbance for blank is displayed as 0.0185. Then take
standard 1in the sample tube, place it in the chamber. Click ‘start’ and
observe the reading is 0.0902. Similarly for standard 2 the absorbance is
0.2377 for standard 3 the reading is 0.4604, for standard 4 the reading is
0.6177, finally for standard 5 the value of the reading is 0.8024. Now transfer
the given sample from standard flask to sample tube and place it in the
chamber. Click ‘start’ the absorbance reading obtained is 0.7824.
Tests on Concrete
● Slump Test
IS: 456-2000
Apparatus required
● Mold for slump test i.e. slump cone, non porous base plate, measuring
scale, temping rod.
● The mold for the test is in the form of the frustum of a cone having
height 30 cm, bottom diameter 20 cm and top diameter 10 cm.
● The tamping rod is of steel 16 mm diameter and 60cm long and
rounded at one end.
● V funnel test
EFNARC 2002
Apparatus
● V-funnel
● Bucket (±12 liter)
● Trowel
● Scoop
● Stopwatch
Procedure
● About 12 liter of concrete is needed to perform the test, sampled
normally. Set the V-funnel on firm ground. Moisten the inside surface of
the funnel. Keep the trapdoor to allow any surplus water to drain.
● Close the trap door and place a bucket underneath. Fill the apparatus
completely with the concrete without compacting or tamping; simply
strike off the concrete level with the top with the trowel.
● Open within 10 sec after filling the trap door and allow the concrete to
flow out under gravity.
● Start the stopwatch when the trap door is opened, and record the time
for the complete discharge (the flow time).
● This is taken to be when light is seen from above through the funnel.
● The whole test has to be performed within 5 minutes.
● L box test
EFNARC 2002
Equipment
● L box of a stiff non absorbing material
● Trowel
● Scoop
● Stopwatch
Procedure
● U box test
EFNARC 2002
Equipment
Procedure
Procedure
Remove the specimen from the water after specified curing time and wipe out
excess water from the surface.
Mixing Concrete
● Mix the concrete in batches of 0.6 ft3 or larger.
● Bring the materials used in the concrete to room temperature before
mixing. After mixing, the concrete shall have a temperature of 73°F ±
3°F.
● Prepare the molds by coating their interior surfaces with a release
agent. After coating the molds with the release agent, place the
measuring studs in the end plates, using care to avoid getting any
release agent on the studs.
● All aggregates and about two-thirds of the mixing water shall be mixed
briefly in an open-tub type mixer and allowed to stand for 5 min before
the addition of the cement.
● After the soaking period specified above, the cement and an air
entraining agent, added as a diluted solution made with part of the
remaining mixing water, shall be added and the timer started. The
remaining water required to produce a slump of 3 in. ± ½ in. shall be
added in the first minute of mixing. Mixing shall be for 3 min followed by
a rest period of 3 min, after which mixing shall continue for 2 min. Any
adjustments in amount of mixing water to produce the desired slump
shall be made during the first ½ min of the second mixing period.
● After mixing, determine the slump, density and air content. Determine
the slump in accordance with ASTM C 143/C 143M. Determine the
density in accordance with California Test 518. Determine the air
content in accordance with ASTM C 173/C 173M.
● Discard the concrete used in the determination of the air content. Return
concrete used to determine slump and density to the mixer and remix
briefly before fabrication of test specimens.
● Mold at least three test specimens from each batch in accordance with
the procedure described in ASTM C 157/C 157M, except that in lieu of
rodding and spading the mix into the molds, an external vibrator, such
as a vibratory packer, may be used to compact the mix in two layers.
Use only enough vibration to consolidate the concrete. Large aggregate
protruding at the surface is an indication of over vibration.
● The first layer of concrete should just cover the top of the gage studs.
While vibrating briefly, work the concrete into the corners and around
the gage studs with the fingers, then fill the mold to slightly overflowing
and vibrate again.
● When compaction is complete, strike off the top surface and finish with
a steel straightedge.
● Immediately after molding, release the plates holding the gage pins.
● After molding, cover the specimens with wet mats. (Keep mats wet at all times.)
● When specimens are 24 hr ± 4 hr old, remove them from the molds and place them in
the moist curing room.
● At 7 d after molding, measure the specimens for length (initial length).
● Store on racks in the drying room maintained at a temperature of 73°F ± 3°F and a
relative humidity of 50 % ± 4 %. Store them with at least 1 in. clearance on all sides
except for the necessary supports.
● At 14, 21, and 35 d after molding (7, 14, 28 d of drying), measure the specimens again
for length.
Calculations
Compute the drying shrinkage for each specimen to the nearest 0.001
percentage point, based on the gage length of 10 in. The reported computed
shrinkage must be the average of at least 7 specimens, the individual values
of which do not depart from the average by more than 0.004 percentage point.
Results
Apparatus required
● Compaction factor apparatus consists of trowels, hand scoop (15.2 cm
long), a rod of steel or other suitable material (1.6 cm diameter, 61 cm
long rounded at one end ) and a balance.
Procedure
Place the concrete sample gently in the upper hopper to its brim using the
hand scoop and level it.
Apparatus required
Procedure
Apparatus required
Fix and place the Vicat apparatus on a flat surface and make necessary
adjustments.
● Now take 400gm of cement specimen and add 28% of water by the
weight of cement and mix gently for up to 3 to 5 minutes.
● Then fill the cement paste into the Vicat mould and remove the excess
cement paste by the trowel.
● Now place the Vicat mould on the Vicat apparatus and release the
plunger gently by contacting the cement paste surface.
● Then release the plunger and allow it to penetrate the cement paste and
note down the reading from the gauge scale from the bottom of the
Vicat mould.
● Now again, add water with cement paste at a different water ratio until
the reading lies between 5mm and 7mm.
● Heat of Hydration Test
IS 4031-3 (1988)
Apparatus required
The heat of hydration test is conducted at a standard room temperature of 27
degrees celsius. The apparatus for this test is
1. Calorimeter
2. Mortar and pestle
3. Glass/Plastic vials
4. Stopwatch or timer
5. Sieve
6. Muffle furnace
7. Weighing balance
The calorimeter determines the heat of the paste. Glass vial has a dimension
of 80×20 mm. Muffle furnace maintains the temperature. Before performing
the test, the calorimeter temperature is to be made uniform. The heat of
hydration should be determined at the 7th and 28 days.
Procedure
For low heat cement, The heat of hydration for 7 days should not be greater
than 66 cal/g and for 28 days should not be greater than 75 cal/gm.
● Soundness Test
IS:4031-Part 3-1988
Apparatus required
Procedure
Place the lightly oiled mould on a lightly oiled glass sheet and fill it with
cement paste formed by gauging cement with 0.78 times the water
required to give a paste of standard consistency. [refer IS : 4031 (Part
4 )-1988].
● Cover the mould with another piece of lightly oiled glass sheet, place a
small weight on this covering glass sheet and immediately submerge
the whole assembly in water at a temperature of 27 ± 2°C and keep
there for 24 hours.
● Measure the distance separating the indicator points to the nearest 0.5
mm. Submerge the mould again in water at the temperature prescribed
above.
● Bring the water to boiling, with the mould kept submerged, in 25 to 30
minutes, and keep it boiling for three hours. Remove the mould from the
water, allow it to cool and measure the distance between the indicator
points.
● The difference between these two measurements indicates the
expansion of the cement. This must not exceed 10 mm for ordinary,
rapid hardening and low heat Portland cements. If in case the
expansion is more than 10 mm as tested above, the cement is said to
be unsound.
● Temperature Test
IS 7861-1 (1975)
Apparatus required
● Container: The container shall be made of non-absorptive material
large enough to provide at least 3 in. of concrete in all directions
around the sensor of the temperature measuring device. Concrete
cover must also be at least three times the nominal maximum size of
the coarse aggregate.
● Temperature measuring device: The temperature measuring device
shall be capable of accurately measuring the temperature of the
freshly mixed concrete to ± 1°F throughout a range of 30 to 120°F.
The temperature measuring device shall be of sufficient length that it
can be immersed at least 3 in. during the test.
● Reference Temperature Measuring Device: The reference
temperature measuring device shall be readable and accurate to ±
1°F at the verification points specified below. A certificate or report
that verifies the accuracy shall be available for review. The certificate
or report shall provide documentation that the reference standard
used in the verification is traceable to the National Institute of
Standards and Technology (NIST).
Procedure
Procedure
● The cone is placed in the center of the flow table and filled with fresh
concrete in two equal layers. Each layer is tamped 10 times with a
tamping rod.
● Wait 30 seconds before lifting the cone.
● The cone is lifted, allowing the concrete to flow.
● The flow table is then lifted up 40mm and then dropped 15 times,
causing the concrete to flow.
● After this the diameter of flow of the concrete is measured.
Test specimens
● Four specimens are required per test. Each specimen shall consist of a
70 ± 2 mm diameter concrete disc with a thickness of 30 ± 2 mm, cored
and cut in accordance with Concrete Durability Index Testing Procedure
Manual, Part 1.
● Mark the specimen(s) with a reference number (e.g. 1, 2, 3 and 4) on
the interior face with a permanent marker.
Conditions of specimens
● Place the specimens in the oven, maintained at 50 ± 2°C, for not less
than 7 days and not more than 8 days. Allow sufficient space between
and around specimens to ensure that they can dry uniformly.
● After the drying period, remove the specimens from the oven and
immediately place them in the desiccator.
● Cool the specimens to 23 ± 2 °C in the desiccator. Allow the specimens
to cool for a minimum of 2 hours and a maximum of 4 hours. Remove
the specimens from the desiccator and start the testing within 30 min.
● Measure and record, to an accuracy of 0.02 mm, the thickness and
diameter of each specimen with the Vernier caliper at four points equally
spaced around the perimeter of the specimen. Calculate and record, to
the nearest 0.02 mm, the average of each set of four readings.
Testing of specimens
● Place the specimen in the compressible collar within the rigid sleeve
with the exterior (test) face at the bottom. No gaps should be visible
between the sides of the test specimen and the collar. The specimens
shall be placed so that the interior face rests against the lip of the collar.
Note: The terms “exterior” and “interior” are used with respect to the
position of the face of the specimen in the structure or the cube from
which they were extracted. The provision of a “lip” in the rubber collar is
to ensure that the specimen is held in place tightly when it is subjected
to pressure on the test face during the test.
● Place the sample, collar and rigid sleeve on top of the permeability cell
so that they cover the hole. Place the solid ring (the solid ring is optional
depending on the permeability cell arrangement) on top of the collar,
ensuring that no gaps are visible between the collar and the sleeve.
Place the cover plate on top of the solid ring.
● Partially tighten the top screw on the cover plate to ensure that it is
centred. Once the specimen has been centred, tighten the apparatus
adequately to ensure no leakage of gas. 10
● Open the oxygen inlet and outlet valves of the permeability cell. Open
the valve of the oxygen supply tank to between 100 and 120 kPa, and
allow oxygen to flow through the permeameter cell for 5 seconds. This
will purge the test chamber of gases other than oxygen.
● Close the outlet valve of the permeability cell, ensuring that there are no
leaks.
● Increase the pressure in the permeability cell to 100 ± 5 kPa and close
the inlet valve.
● After 5 min, record the time, t0, to the nearest minute, as the initial time,
and initial exact pressure P0, to the nearest 0.5 kPa. Use t0 and P0 as
such in the calculations. Thereafter, take at least eight readings at
intervals corresponding to a pressure drop rate of 5 ± 1 kPa. A pressure
drop of more than 5 kPa/min might be an indication of leakage. In such
a case, release the pressure in the chamber, check that the sample fits
tightly in the collar, and restart the test immediately,
● Terminate the testing when the pressure has dropped to 50 ± 2.5 kPa or
after 6 hours ± 15 min, whichever occurs first. A minimum of 8 readings
is required.
Apparatus
Test specimens
● Four specimens are required per test. Each specimen shall consist of a
70 ± 2 mm diameter concrete disc with a thickness of 30 ± 2 mm, cored
and cut in accordance with Concrete Durability Index Testing Procedure
Manual, Part 1. It is permitted to use specimens that have previously
been used in the oxygen permeability test, provided they have not been
exposed to moisture. Refer to Part 2: ‘Standard Procedure for Oxygen
Permeability Test’.
● Mark the specimens of the same reference 1, 2, 3 and 4 on the inner
face.
Conditioning of specimens
Testing of specimens
● Pour calcium hydroxide solution into the tray. If used, the paper towels
should be saturated with water visible on the top surface. All air bubbles
should be removed by smoothing the paper pad towards the edges,If
rollers or pins are used, they should be arranged so as to support the
specimens and the calcium hydroxide solution should be above the top
of the support, Figure 3.3. The final water level should be such that it
will be slightly above the bottom edge of the specimen and a maximum
2 mm up the side of the specimen as shown in Figure 3.4. Dampen an
additional piece of paper towel for use in removing the excess water
from the specimens (in paragraph (g)) and keep next to the tray to be
used during the test. NOTE: Rollers or pins, if used, should be
sufficiently small so as not materially to affect the area of specimen
surface exposed to the solution.
● Within 30 min after removing the specimen from the desiccator, or
oxygen permeability cell, determine the mass of the specimen to an
accuracy of 0.01 g and record as the dry mass, Ms0. This mass must be
determined after the method used to seal the sides of the specimens.
● Immediately place the specimen with the test face (outer face or
originally exposed face) on the wet paper pad/pins/rollers and start the
stopwatch, at time t0.
● Weigh the specimen at 3, 5, 7, 9, 12, 16, 20 and 25 minutes, after
patting it once on the damp piece of absorbent paper. The specimen
should appear saturated surface dry (SSD) on the exposed face at the
time the mass is determined, i.e. it should look damp, but not have free
water on the test face. During removal of a specimen from the tray for
weighing, care must be taken to prevent dripping from one specimen
onto the top of another.
● Record the mass of the specimen to the nearest 0.01 g within 10 s of
removal of the specimen from the tray and wiping excess water.
Replace the specimens each time with the test face on the wet paper or
rollers/pins if used. The stopwatch shall not be stopped during the
weighing procedure.
● Within a maximum of 1 day after weighing of the specimen is
completed, place the specimen in the vacuum saturation tank. The tape
or sealant must be left in place. The specimens shall be arranged so as
to maximize their exposed surface area. This is typically done by
standing the individual specimens upon their curved edges, rather than
the flat side, as in Figure 3.5. Seal the lid with petroleum jelly and close
it.
● Evacuate the tank to between -75 and -80 kPa and maintain the
specimens under vacuum of between -75 and -80 kPa for 3 hours ± 15
min. The pressure must not be allowed to rise above -75 kPa during this
period.
● After 3 hours ± 15 min isolate the tank and allow calcium hydroxide
saturated water to flow into the chamber until the water level is
approximately 40 mm above the top of the specimens. Air shall not be
allowed to enter the vacuum chamber during this procedure.
● Re-establish the vacuum to between -75 and -80 kPa. This shall be
maintained for 1 hour ± 15 minutes. At no point during this time period
shall the vacuum be permitted to rise above -75 kPa.
● After 1 hour ± 15 min, release the vacuum and allow air to enter. Allow
the specimens to soak for a further 18 ± 1 hours.
● After 18 ± 1 hr soaking, remove the specimens from the solution, dry the
surface to a SSD condition with a paper towel, and immediately weigh
to an accuracy of 0.01 g. Record this as the vacuum saturated mass
Msv of the specimen.
●
● Freeze thaw test
ASTM C666
Equipment
● Dry storage
The concrete specimens are stored in the climate chamber (20°C,
65% RH) for surface drying for 21 days. Monitoring of the weight
is recommended. In the climate chamber the evaporation from a
free water surface shall be 45 ± 15 g/m2h. The evaporation is
measured by weight loss of a water-filled evaporation bowl with a
depth of approximately 40 mm and a crosssectional area of 225 ±
25 cm2. The bowl shall be filled up to 10 ± 1 mm from the brim.
Monitoring of the weight is recommended.
● Preparation of specimens
Between 7 and 2 days before presaturation, the specimens are
sealed on their lateral surfaces either with aluminium foil with butyl
rubber or with a solvent-free epoxy resin.11 The specimens must
be clean and dry. Before sealing the lateral surfaces, it is
recommended to treat them with an adequate primer.
1. Sealing with aluminium foil with butyl rubber: A piece of
aluminium foil with butyl rubber is glued tightly on the lateral
surfaces with an overlap of 20 mm.
2. Sealing with epoxy resin: A solvent-free epoxy resin is
laid on the lateral surfaces, whereas the bottom of the
specimens and the test surface must be kept free.
Test set-up
Test specimens
The tested columns were 3000 mm tall and square or circular cross-
sections. The square cross-section columns had 160 mm x 160 mm and 250
mm x 250 mm and the circular cross-section ones had 250 mm and 300 mm
in diameter. The circular columns had six longitudinal 12 or 20 mm and the
square columns had four 10, 16 or 25 mm rebars. The transversal
reinforcement was performed by 6 or 8 mm diameter stirrups with a spacing
of 100 mm until 700 mm from the supports, and a spacing of 150 mm in the
central part. The 6 mm diameter stirrups were used with the 12 mm diameter
rebars in the circular columns and with the 10 or 16 mm diameter rebars in the
square columns. The 8 mm diameter stirrups were used with the 20 mm
diameter rebars in the circular columns and with the 25 mm diameter rebars in
the square columns. The concrete covering related to the stirrups for all tested
columns was 30 mm. All specimens were fabricated with reinforcing bars of
A500NR steel and calcareous aggregate concrete C20/25. The temperatures
were measured in five cross-sections on the circular columns and in three
cross-sections on the square columns (Fig. 4). Type K thermocouples were
used to measure the temperatures on the specimen while probe
thermocouples were used to measure the
furnace temperatures.
Test Plan
Procedure
● The test set-up for determining temperature distribution in concrete
specimens subjected to freeze-thaw cycles was based on Thermochron
electronic sensors from Maxim integrated. The sensor was used in the
task mode, that is after presenting the start time and time interval
between subsequent readings the results were stored in the internal
memory of the unit. The results were available for viewing only on
completion of testing when the sensors were pulled out from the
concrete specimens and connected to the computer via a special
interface.
● The temperature sensors were embedded at predetermined places
inside 150 mm and 100 mm concrete cubes (Fig. 6). The test settings
included 360 sec. temperature recording interval with measurement
accuracy of 0.5°C, enabling to measure and record the temperatures for
the test period of 34 days.
● The specimens were molded from C35/45 air-entrained concrete,
having the composition of concrete used for road structures. Main mix
constituents consisted of Portland cement CEM I 42.5R (the content of
375 kg/m3), amphibolite crushed aggregate of 2-8 and 8-16 mm,
quarried sand, and water corresponding to w/c= 0.48.
● The specimens were demoulded after two days of curing and then they
were conditioned for another 28 days in water. After that time they were
placed in the thermostatic chamber and subjected to freeze-thaw cycles
according to the procedure of PN-B-06250 (Fig. 7). An additional
temperature sensor was attached to the top surface of each specimen.
● The temperature was measured for the period of one week in one
laboratory and then the specimens were transferred to the second
laboratory for continuing the test in a different thermostatic chamber. On
completion of the test procedure the sensors were removed from the
cubes to read and analyze the recorded results
● Abrasion Resistance Test
IS 9284 (1979)
3. The nozzle's length, including the tip, must be 15 cm, and it must be
mounted on the cabinet so that the distance between the nozzle's tip
and the surface of the concrete sample is 50.02 cm.
4. An annular copper or brass air tube of 0.48 cm in diameter shall be
the nozzle.
7. Four holes, each having a diameter of 0.48 cm, shall be provided on the
side of the nozzle immediately below the collar for passing sand into the
nozzle.
8. Sand is drawn into the nozzle by the suction caused by high air velocity
placed at the top of the cabinet to store the charge of sand pushed by
air pressure.
2. The entire nozzle assembly is secured to the cabinet with a lock nut and
washer combination.
3. The copper or brass air tube must be connected with a pressure gauge
● Abrasive Charge
Test Procedure
with the surface to be examined facing the nozzle tip (any of the four
smoother vertical surfaces of the cube that came in contact with the
mold).
● The nozzle tip shall be placed at the middle of the half side of the cube
weighed to determine the mass loss in grams for one surface of the
sample.
● The above procedure shall be repeated on the other three vertical
Concrete Structures
IS 1199 (1959)
The specimens selected for this experiment are from a river-crossing tube
sheet in service. The design strength grade is C50, the 28-day compressive
strength is 61.6 MPa and the water–cement ratio is 0.4, according to the
relevant research [42,43] and the ratio parameters provided on site; see Table
1.
To prepare the specimens, the poured pipe pieces were first broken using a
crushing hammer to extract the reinforcement inside the pipe pieces. A
concrete core sample with a diameter of 50 mm was extracted using core-
drilling equipment. Then, the core sample was processed into a 50 mm × 100
mm cylindrical specimen. The specimens were prepared for testing, with an
age of 1150 d and an original compressive strength of 75 MPa.
The original tunnel used methyltrimethoxysilane silane to spray a protective
coating on the surface of the tube sheet before it was put into use. The
experimental specimens used in this paper restore the use of silane materials
on site, and the designed amount of silane for concrete in the reference
project is 450 g/m2. The surface area of the specimens used in this paper is
about 0.01963 m2, and the corresponding design amount is about 8.8357 g.
Considering the accuracy required for actual operation and measurement, the
actual amount of silane used in this paper is 8.9 g for each specimen. The
specimens were divided into a protected group, exposed group and controlled
group. The protected group was coated with the designated amount of silane
before the experiment, and the amount used per single specimen was 8.9 g.
● Specimen Testing
In this paper, three types of methods are combined to analyze the changes in
the properties of concrete specimens during rapid wetting–drying cycles of
erosion. Ultrasonic testing was used to analyze the overall condition of the
concrete. The structural changes to the specimens were analyzed using CT
scanning and NMR testing. Strength tests were used to compare the strength
changes of the specimens under different conditions.
● Ultrasonic Testing
A non-metallic ultrasonic tester was used to test the longitudinal wave velocity
on the side of concrete specimens at a sampling rate of 0.05 μs. After every
ten wetting–drying cycles, a test was conducted on the specimens of the
protected group, the exposed group and the controlled group. The test
specimens were divided into four test planes at equal intervals of 20 mm from
top to bottom. Nine measurement lines are laid out at equal intervals at 20°
angle for each measurement surface. The longitudinal wave velocity of each
specimen was tested at 36 points in total, and the test situation is shown in
Figure 2. We used an HS-CS1H ultrasonic parameter tester made by
Xiangtan Tianhong Electronics Research Institute; the frequency of the arc
probe is 1 MHz.
● CT Test
The specimens’ structure was analyzed using an industrial CT system. We
use the model Y.CT.Modular X-CT scanning equipment, with continuous
spiral scanning mode, tube voltage 225/450 KV, imaging plate size 400 mm ×
400 mm, imaging plate pixel dot size 0.2 mm, detectable diameter range
0~800 mm and maximum penetration thickness 60 mm Fe. The specimens
were scanned in high precision slices for the protected and exposed groups at
the end of 60, 120 and 180 wetting–drying cycles. The minimum scanning
resolution was 5.3 μm, and the number of slices scanned for a single
specimen was 1600.
● NMR Test
After the CT tests were completed for the protected and exposed groups, we
used a Suzhou Newmark AniMR-150 NMR instrument to perform NMR tests
on the specimens before and after the erosion experiment to obtain pore data.
The magnetic field strength of the test system is 0.3 ± 0.05 T, the peak output
is not less than 300 W and the linear distortion is less than 0.5%. The
specimens were soaked in water for 48 h to saturate the specimens before
the test was started.
● Strength Test
The specimens of the exposed group and the specimens of the protected
group that completed the NMR test were dried at 60 °C to the state before
immersion. The specimens of the exposed group, protected group and
controlled group were uniaxially loaded at 0.5 MPa/s using a servo material-
testing machine, and the peak load when the specimen diameter d is
damaged was recorded. The uniaxial compressive strength can be calculated
according to the following formula.
● CARBONATION TEST
IS-516(Part 5/Sec 3):2021 “Part 5 Non-Destructive Testing of Concrete-
Section 3 Carbonation Depth Test"
● Preparation Of Sample
The test shall be performed on freshly exposed concrete surface. This may be
either freshly broken surface of concrete or extracted concrete core sample
which may preferably be split and the test may be conducted on the split face.
If facility for splitting is not available, then the core may be surface dried and
sealed to prevent further carbonation. After breaking, the concrete surface
shall immediately be cleared of any dust or loose particles.
● Test Procedure
● When the carbonation front runs as a straight line parallel to the surface,
the depth of carbonation dk is determined as shown in Fig. 1A.
NOTES :
1 The measured depth of carbonation may be influenced by the time of
measuring after application of the indicator solution. A stable reading
may be obtained after about 5 to 10 min after spraying.
2 If only a weak colouration or none at all appears, it is better to repeat
the spray after the surface has dried.
● TEST FOR AIR CONTENT OF FRESHLY
MIXED CONCRETE BY THE PRESSURE
METHOD
ASTM C231
Apparatus required
● Air Meter: Commercially available air meter of the modified Washington
type consisting of a base and top section, and furnished with all
accessories necessary for calibration. The directions outlined below
under Operation of Apparatus apply to the brands used by the California
Department of Transportation. These meters operate on the principle of
equalizing a known volume of air at a known pressure in a sealed
chamber with an unknown volume of air in the concrete sample.
Calibration and operation of these meters is not affected by barometric
pressures.
● Tamping Rod: A round, straight, steel rod 9/16 in. to 5/8 in. in diameter
and approximately 24 in. in length. The tamping end shall be rounded to
a hemispherical tip with a diameter equal to the rod.
● Strike-off Bar: A flat, straight bar of steel or other suitable material.
● Scoop: A small metal scoop.
● Syringe: A small rubber ear syringe.
Procedure
Use the same sampling and rodding procedure for all types of air
meters. Obtain the sample of concrete for the air content determination
in accordance with the applicable provisions of ASTM Designation: C
172.
● Rodding and Tapping: Using the scoop, fill the base with freshly mixed
concrete in three layers of equal depth. Rod each layer 25 times with
the tamping rod, distributing the strokes evenly over the surface of the
layer. After rodding each layer, tap the sides of the base with the hand
or a wood stake to remove entrapped air along the sides and to close
any holes left by the tamping rod. In rodding the first layer, penetrate
nearly full depth into the layer, but avoid striking the bottom of the base.
In rodding the second and third layers, penetrate slightly into the layer
below with each stroke.
● Striking Off: Strike off the base, level full, with the straightedge and
clean the top edge of the base to ensure an airtight seal with the cover.
● Completion of the Testing Using a Press-Ur-Meter:
a. Clamp the lid on with both petcocks open.
b. Using the syringe, inject water through one petcock until all air is
expelled through the opposite petcock. Jar meter gently to remove
entrapped air. Leave petcock open.
c. With built-in pump, pump air in sealed chamber to “initial pressure”
line on gauge. Wait a few seconds for the compressed air to cool, then
stabilize gauge hand at initial pressure line by pumping or bleeding off
air as needed by “cracking” cap over air valve.
d. Close both petcocks and press down on thumb lever to release the
air into the base. Hold lever down a few seconds and tap gauge lightly
with fingers until gauge hand stabilizes.
e. Read the percent of entrained air on the dial.
f. Open petcocks slowly to release pressure, then remove the cover.
Clean the base, cover, and petcock opening thoroughly. Take care not
to damage the top edge of the base.
Drop Hammer
○ The drop hammer is used to drive the split spoon sampler into the
borehole.
○ The weight of the hammer in the SPT test is 63.5 kg.
Guiding Rod
The Guiding rod serves the purpose of guiding the hammer to the anvil.
Procedure
● Set up the tripod over the test hole and place the unit together.
● Let the spoon rest on the bottom of the hole.
● Hit the spoon with a hammer falling 75 cm to drive it 45 cm into the
ground. Alternatively, stop after hitting the sampler 100 times.
● Note down the number of blows required to penetrate the sampler to a
depth of 15 cm. It indicates the seating load and the number of blows
required for the first 15 cm penetration are to be discarded.
● The penetration resistance value N of the soil is the number of blows
needed for the second and third 15 cm penetration. In other words, the
number of blows required to drive the sampler to a depth of 30cm is
called the penetration resistance value.
Procedure
Apparatus required
● A steel cone
● A friction jacket
● Sounding rod
● Mantle tube
● A driving mechanism
● Measuring equipment
Procedure
● The cone and friction jacket assembly is put vertically touching the
ground.
● The cone is then pushed down at a constant speed of penetration
ranging from 1.5 to 2.5cm/sec to a depth a (say 40 mm). Then, cone tip
resistance (qc) is recorded.
● Now the sounding rod is pushed further to a depth b (say 40 mm). Then,
sleeve friction (qf) is recorded.
● Then, the mantle tube is pushed down to a+b=80mm depth. This brings
the cone and jacket assembly to step 1.
● Then the friction ratio is calculated using the formula:
Apparatus required
● Pan
● Support device, rods, pins, or other devices
● Top-pan balance, accurate to at least ±0.01 g
● Timing device
● Paper towel or cloth
● Environmental chamber
● Polyethylene storage containers
● Caliper
● Sealing materials like duct tape or aluminum tape
● Plastic bag or sheeting
Procedure
● Take out specimens from the storage container and weigh them to the
nearest of 0.01 g.
● Measure a minimum of four diameters of the specimen at the surface to
be exposed to water and calculate the specimen's average diameter.
● Seal the side surface of specimens with suitable seal material, and seal
one end that is not exposed to water with a plastic sheet that can be
secured using an elastic band or other equivalent systems.
● Weigh the sealed specimen and record it as the initial mass.
● Place specimen support at the pan's bottom and pour tap water into the
pan until it rises nearly 3 mm above the specimen supports, see Figure-
1. This level of water in the pan needs to be maintained during the test.
● Start the timing device and put the unsealed surface of the specimen on
the supports in the pan. Write the time and date of the initial contact of
the sample with the water.
● Record the mass of the specimen after first contact according to the
time interval provided in Table-1.
● For each record of a mass, remove the concrete specimen from the pan.
Then, stop the timing device if the contact time is less than 10 minutes,
and wipe out any surface water with dampened paper or cloth. Invert the
specimen so that the wet surface does not contact the balance pan.
Measure the mass within 15 seconds of specimen removal from the
pan. Finally, place the specimen on the support in the pan and restart
the timing device.
Tests on Admixtures
IS 9103 (1999)
Procedure:
● Make the cement slurry.
● Find out the flowability of the slurry (this is because flowability is
ultimately influenced by cement paste).
● Take 2 kg cement to be used and prepare cement slurry by adding 1-
liter water.
● Add 0.1% superplasticizer.
● Mix them in a mechanical mixer.
● Close the aperture of the marsh cone with a finger and pour 1-liter slurry
in it.
● Remove the finger and start stop watch at that moment.
● Record the time taken for the slurry to completely flow out from the
marsh cone. This time is called “marsh cone time”.
● Repeat the procedure altering the dosage of the superplasticizer.
● Plot a graph taking superplasticizer dosage as X axis and marsh cone
time as Y axis.
Procedure
● Oven(3000c)
● Bottle
● Balance(0-10kg)
Procedure
Calculation
Reports
The result of the test should be reported to the nearest whole number.