Professional Documents
Culture Documents
Flow Meter Lab Report Annagracekeel
Flow Meter Lab Report Annagracekeel
Team 3
Abstract ........................................................................................................................................... 3
Introduction ..................................................................................................................................... 4
Objectives ....................................................................................................................................... 4
Methodology ................................................................................................................................... 5
Results and Discussion ................................................................................................................... 7
Initial Data Collection ................................................................................................................. 7
Volumetric Flow Rate Comparative Analysis ............................................................................. 7
Flow Meter Device Analysis ....................................................................................................... 9
Energy Loss Assessment ......................................................................................................... 9
Pressure Drop Assessment .................................................................................................... 10
Discharge Coefficient Assessment .........................................................................................11
Conclusion .................................................................................................................................... 12
References ..................................................................................................................................... 14
2
Abstract
A flow metrology apparatus allows for the investigation of unique flow meter characteristics
affecting the quantified energy loss and efficiency for a device. The methodology of the EDIBON
FNE18 system includes analyzing the venturi, rotameter, and orifice device by performing trials
of six set rotameter volumetric flow rates. The rotameter flow rates were compared to the eight
manometer taps and the dump valve system flow rates. A comparison analysis for the volumetric
flow rates between the rotameter and the dump valve system revealed a large percentage error
between the respective devices. The Bernoulli Equation provided a basis for the energy loss and
pressure drop assessment. The venturi meter is the most effective flow meter device and has
minimal energy losses in comparison to the orifice and rotameter. For example, at a flow rate of
3.7 x 10-4 m3/s the venturi had an energy loss of .137 J/kg, the rotameter had an energy loss of
2.335 J/kg, and the orifice had an energy loss of 1.138 J/kg. The pressure drop assessment confirms
the relationship between pressure drop and flow meter efficiency. For example, at a flow rate of
2.1 x 10-4 m3/s, venturi had a pressure drop of 58.86 Pa, orifice had a pressure drop of 470.88 Pa,
and rotameter had a pressure drop of 2295.54 Pa. The experimental data confirmed that a higher
pressure drop provided greater inefficiency for the flow meter. A coefficient discharge analysis
was performed on venturi and orifice. Venturi provides a theoretical value of 0.98 and estimated
value of 1.97 while orifice provides a theoretical value of 0.61 and estimated value of 0.76.
Through the analysis, it was confirmed that larger data outliers provide a greater percent error
for the venturi and smaller data outliers provide a smaller percent error for orifice. It was proven
that the order of efficiency in flow meters from least to greatest are rotameter, orifice, and venturi.
The experiment highlights how the volumetric flow rates, energy loss, pressure drop, and
3
Introduction
The experimental lab highlights the EDIBON FN18 flow metrology apparatus. It analyzes the
relationship between the volumetric flow rates in the rotameter and dump valve system, quantifies
the energy losses for each flow meter device, assesses the pressure drop for each flow meter device,
and compares the estimated and theoretical discharge coefficient for both venturi and orifice flow
meters. The flow meter devices analyzed in the experiment include venturi, rotameter, and orifice.
The venturi meter was invented by Clemens Herschel and is the most costly and accurate flow
measurement device, however, it is often only utilized for challenges containing small pressure
drops (Cengel & Cimbala, 2018). The oil industry, for example, relies heavily on flow meters for
the accuracy of product flow. High oil prices cause companies to drill for highly viscous oils with
significantly lower Reynolds numbers, resulting in problematic accuracy for oil flow measurement
(Hollingshead et al., 2011). Each flow meter device is unique to its own advantages and
disadvantages. For example, the venturi meter is preferred for larger pipe diameters while the
orifice meter is preferred for higher pressure drops (Meribout et al., 2020). The venturi, rotameter,
and orifice flow metrology devices are analyzed to determine their respective energy losses,
pressure drops, and discharge coefficients. The experiment identifies how the rate of pipe flow
Objectives
The focus of the flow meter lab is to highlight the EDIBON FN18 flow metrology apparatus which
includes the venturi flow meter, rotameter flow meter, orifice flow meter, manometer taps, and a
dump valve system. The experimental objective is to determine the discharge coefficients for the
4
venturi and orifice flow meter and assess the energy losses from the pipe flow in the venturi,
Methodology
The EDIBON FNE18 utilized in the experiment contains three flow metrology devices: venturi,
rotameter, and orifice. The FNE18 device consist of eight manometer taps which measure the head
loss for each respective flowmeter. Tap 1 through 3 correlated with venturi, tap 4 through 5 with
rotameter, and tap 6 through 8 with orifice. Tap 1 minus tap 3 calculated venturi head loss, tap 4
minus tap 5 calculated rotameter head loss, and tap 6 minus tap 8 orifice head loss. A dump valve
system attached to the device measured the volumetric flow rate through the system. The flow rate
was initially set to 1315 L/h for an increment of 30 seconds. For each set flow rate, the initial and
final volume of the water was observed at the tank. Simultaneously, measurements of manometer
taps 1-8 were recorded. The measurements for the rotameter volumetric flow rates, dump system
volumes, and manometer taps were observed and recorded for each of the remaining five trials.
The trials gradually decreased the set flow rates to 1000 L/h, 750 L/h, 500 L/h, 350 L/h 200 L/h
respectively.
5
F igure 1. EDIBON F NE18 F low Met er Dem onst rat ion Device
6
Results and Discussion
Given the objectives for the experiment, the initial data was collected over a repeated process of
six trials starting from a flow rate of 1315 L/h and gradually decreasing to 200 L/h. The
measurements recorded included the manometer tap height (mm) for tubes 1-8 (Table 1), rotameter
flow rates (L/h), and the initial and final volumes (L) for the dump valve system (Table 2).
The dump system initial and final observed volumes were converted into a volumetric flow rate
by calculating the change in volume divided by the set time increment of 30 seconds (Table 3).
7
Table 3. Rot am et er and Dump Syst em F low Rat es
The percent error of the rotameter and dump system flow rates were calculated (Figure 3). The
calculated percent error decreased as the observed flow rate form the rotameter increased. It is
notable that the large percent error potentially results from incorrect observations collected for the
rotameter along with the initial and final volumes from the dump system. The rate of pipe flow
could impact the difference between the two volumetric flow rates, resulting in an inconsistent
conclusion for the accuracy of the rotameter. It is advisable to perform more experimental testing
before employing the rotameter and dump valve system into the industry.
F igure 3. Relat ionship Bet ween P ercent Error of t he Rot am eter and Dum p Syst em F low Rat e
8
Flow Meter Device Analysis
The graphical representation between energy loss and increasing flow rate reveals the order of
efficiency for each flow meter device (Figure 4). The most to least effective flow meter device in
respective order is the venturi, orifice, and the rotameter. The venturi’s energy loss increases
slightly as volumetric flow rate increases from a flow rate of 6x10-5 m3/s to 3.7x10-4 m3/s. The
venturi meter includes a converging and diverging section connected by a throat allowing the
pressure drop to determine the flow rate (Mokhatab et al., 2018). The venturi meter design results
in significantly lower pressure drops, providing a system that results in minimizing energy loss.
𝑃 𝑃2
ℎ𝑒𝑎𝑑 𝑙𝑜𝑠𝑠 (𝑚) = ℎ𝐿 = 𝜌𝑔1 − (1)
𝜌𝑔
P = Pressure (Pa)
𝑘𝑔
𝜌 = 𝐷𝑒𝑛𝑠𝑖𝑡𝑦 (1000 )
𝑚3
𝐽
𝑒𝑛𝑒𝑟𝑔𝑦 𝑙𝑜𝑠𝑠 (𝑘𝑔) = ℎ𝐿 ∗ 𝑔 (2)
9
F igure 4. Relat ionship Bet ween Energy Loss and F low Rat e
The graphical representation between flow rate and square root of pressure drop reveals as the
pressure drop increases, the flow rate increases. The graphical intercepts were set to zero. Venturi
and orifice flow meters have a strong linear trendline with an R 2 parameter of 0.9467 and 0.9933
respectively. It is notable that the rotameter trendline is not linear and remains vertical at an
approximate constant pressure between 47 and 48 Pa. Due to the efficient design of the venturi, it
proves to have the lowest pressure drop values. For example, for trial 1 at flow rate of 3.7x10-4 m3
/s, the venturi square root pressure drop is 12.914 Pa while the orifice square root pressure drop is
33.734 Pa. The graphical representation provides further evidence of the correlation between
pressure drop values and flow meter efficiency; higher pressure drop equals greater flow meter
inefficiency. It is notable to consider the relationship between pressure drop and flow rate to
10
determine the size of pump or motor to employ as an engineer. Higher pressure drops across a pipe
will maximize the energy loss and therefore require a more powerful pump or motor (Pomelo,
2023).
0.00040
0.00035 y = 3E-05x
R² = 0.9467
0.00030
Flow Rate (m^3/s)
0.00025
0.00020
0.00015
0.00010 y = 1E-05x
R² = 0.9933
0.00005
0.00000
0 10 20 30 40 50 60
Square Root of Pressure Drop (Pa)
F igure 5. Relat ionship Bet ween F low Rat e and Square Root of t he Pressure Drop
Coefficient of discharge analysis was calculated from the slope provided in (Figure 4). The
theoretical discharge coefficient for venturi is 0.98 (Cengel & Cimbala, 2018). The estimated
venturi discharge coefficient of 1.97 was determined by the given equation (4). The percent error
between the two values is 100.68%, this could be linked to a calculated pressure drop of 0 Pa for
trial 5. It is possible that manometer taps 1 and 3 were not the same value but instead incorrectly
read due to human error during the experimental observation. The orifice theoretical discharge
11
coefficient was calculated to be 0.61 for high Reynolds numbers from the developed theoretically
determined equation on page 399 (Cengel & Cimbala, 2018). The estimated discharge coefficient
was determined by equation (4) and was determined to be 0.76. The orifice meter data contain less
(4)
Cd = Coefficient of Discharge
Table 4. P ropert ies of Vent uri and Orif ice F low Met ers
Table 5. Coef f icient of Discharge Com parat ive Asse ssm ent
Conclusion
In conclusion, the flow meter lab highlights the EDIBON FN18 flow metrology apparatus which
includes the venturi flow meter, rotameter flow meter, orifice flow meter, manometer taps, and a
12
dump valve system. A volumetric flow rate comparative analysis was performed between the
rotameter and dump valve system. A percentage error analysis between the two values provided
evidence that percent error decreased as the observed flow rate form the rotameter increased. It
also alluded that the large percentage error values signified inconsistency in devices and team
observations. The graphical representation between energy loss and increasing flow rate revealed
the efficiency order of the flow meter devices from least to greatest: rotameter, orifice, and venturi.
The venturi meter design results in significantly lower pressure drops, providing a system that
results in minimizing energy loss. The graphical representation between flow rate and square root
of pressure drop provide significant evidence of the correlation between pressure drop values and
flow meter efficiency. It was revealed that higher pressure drop provide greater inefficiency in the
flow meter device, supporting the theme that the venturi flow meter is the most accurate and
efficient device due to its’ design. The theoretical discharge coefficient for venturi is 0.98 and the
estimated discharge coefficient is 1.97. The theoretical discharge coefficient for orifice is 0.61 and
the estimated discharge coefficient is 0.76. The large percentage error for venturi is from a
calculated pressure drop of 0 Pa for trial 5, likely due to misreading the manometer height taps
during team observations. The experiment determined the discharge coefficients for the venturi
and orifice flow meter along with assessing the energy losses from the pipe flow in the venturi,
13
References
Cengel, Y. A., & Cimbala, J. M. (2018). Fluid Mechanics: Fundamentals and Applications.
McGraw-Hill Education.
Hollingshead, C. L., Johnson, M. C., Barfuss, S. L., & Spall, R. E. (2011). Discharge coefficient
performance of Venturi, standard concentric orifice plate, V-cone and wedge flow meters
at low Reynolds Numbers. Journal of Petroleum Science and Engineering, 78(3–4), 559–
566. https://doi.org/10.1016/j.petrol.2011.08.008
Meribout, M., Azzi, A., Ghendour, N., Kharoua, N., Khezzar, L., & AlHosani, E. (2020).
Multiphase flow meters targeting oil & gas industries. Measurement, 165, 108111.
https://doi.org/10.1016/j.measurement.2020.108111
Mokhatab, S., Poe, W. A., & Mak, J. Y. (2018, October 26). Gas Processing Plant Automation.
https://www.sciencedirect.com/science/article/pii/B9780128158173000204
Pomelo. (2023b, September 7). Understanding the relationship between flow and pressure. Atlas
Scientific. https://atlas-scientific.com/blog/relationship-between-flow-and-pressure
14