Development of An Active Intravascular MR Device With An Optical Transmission System

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 5

IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 27, NO.

12, DECEMBER 2008 1723

Development of an Active Intravascular MR Device


With an Optical Transmission System
Stephan Fandrey*, Steffen Weiss, and Jörg Müller

Abstract—Magnetic resonance imaging (MRI) is a safe and reli- tracking of the catheter tip is needed. The images produced by
able medical imaging method providing good soft tissue contrast an MRI scanner can then be used to guide the catheter inter-
while avoiding harmful ionizing radiation. It is highly desirable actively during surgery. Active MR devices enable very fast
to use the MRI technology for interventional procedures. How-
ever, due to resonance effects that can result in tissue heating, and robust catheter tracking and also intravascular imaging. By
long conducting cables must be avoided. Motivated by the need using local receiver coils at the catheter tip, high resolution and
for more radio-frequency (RF) safety, we developed an optical high signal-to-noise ratio (SNR) images of the surrounding local
transmission system for active intravascular MRI devices. An tissue can be obtained. For active intravascular devices, a signal
optical transmitter sends the MR signal via an optical fiber. A transmission is needed that transmits the received MR signal
miniature optical modulator was designed to be integrated into
a catheter tip. Furthermore, power is supplied optically to the along the catheter.
transmitter. This system can target new medical applications, due Resonance effects in long conducting cables can cause tissue
to safe catheter tracking and safe intravascular imaging. heating [4], which prevents the clinical use of active devices
Index Terms—Catheter, intravascular imaging, magnetic reso- equipped with standard cables. A transformer-based cable has
nance imaging (MRI), optical power supply, optical transmission, been proposed that reduces RF heating for active devices [5]
radio-frequency (RF) safety. and [6]. The transformers suppress common mode currents as-
sociated with RF heating, but they do not completely avoid such
currents and also introduce some signal loss. In this work, we
I. INTRODUCTION present an active MR device with optical signal transmission
and an optical power supply, in which optical fibers are used in-
stead of conducting cables. Optical fibers are inherently RF safe,
T HE USE of minimally invasive procedures has expanded
in the last years. In many of these procedures, a catheter is
used to treat vascular diseases like coronary heart disease. Cur-
so that there is no risk of RF heating. Other optical concepts
have been presented in [7]–[10]. In [7], a coil at the catheter tip
is tracked by intravoxel dephasing caused by a photo current. In
rently, surgeries are performed under X-ray fluoroscopy guid-
[8] and [9], an optically tunable resonance circuit is used at the
ance, which exposes the patient and the medical staff to ionizing
catheter tip. By tuning and detuning the circuit via an optically
radiation. Investigations were made to apply minimally invasive
varied MNOS-capacitance, the catheter tip can be localized au-
therapy under guidance of magnetic resonance imaging (MRI).
tomatically. The method in [10] uses a Faraday effect position
MRI avoids ionizing radiation and has good soft tissue contrast
sensor, which measures the local magnetic field for localization
with the ability of high image resolution and might provide an
of the catheter.
alternative to X-ray imaging in interventional therapy. High-res-
In contrast to these techniques, the potential advantages of the
olution intravascular MRI can be used to obtain diagnostic in-
concept presented here are direct high positive contrast com-
formation about atherosclerotic plaque in vessels by using micro
pared to a low negative contrast from a difference measure-
radio-frequency (RF) coils at the catheter tip [1]–[3].
ment technique in [7] and direct signal transmission instead of
Because real-time MRI is based on the frequent update of
an inductive signal coupling depending on tip orientation in [8]
one or more slices, unlike projection-based X-ray fluoroscopy,
and [9]. Because of the direct signal transmission, the presented
system has the ability to carry image information and is, there-
fore, not limited to tracking as in [10].
Manuscript received February 7, 2008; revised April 16, 2008. Current ver-
sion published November 21, 2008. This work was supported by the German The presented system consists of a miniature resonant re-
Research Foundation (DFG) within in the context of the project “Active intravas- ceiver circuit and an optical transmitter, which transforms the
cular MR-probe in micro system technology for locally high resoluted charac- electrical signal into an optical signal. One objective of this work
terisation of tissue based on a optical signal transmission” in collaboration with
Philips Research Europe-Hamburg and the supply of the photonic power con- was to design all catheter borne components as the transmitter
verter by JDSU.Asterisk indicates corresponding author. with as few miniature components as possible to allow the inte-
*S. Fandrey is with the Hamburg University of Technology, Institute of Micro gration into the minimally invasive device.
Systems Technology, 21073 Hamburg, Germany (e-mail: stephan.fandrey@tu-
harburg.de).
J. Müller is with the Hamburg University of Technology, Institute of II. MATERIALS AND METHODS
Micro Systems Technology, 21073 Hamburg, Germany (e-mail: j.mueller@
tu-harburg.de). In this section, firstly the concept, the fabrication methods,
S. Weiss is with the Philips Research Europe—Hamburg, Sector Medical and the characterization of all components are described. Sec-
Imaging Systems, 21073 Hamburg, Germany (e-mail: steffen.weiss@philips.
com). ondly, the methods to evaluate the complete system for MR ac-
Digital Object Identifier 10.1109/TMI.2008.925079 tive tracking and MRI are described.
0278-0062/$25.00 © IEEE

Authorized licensed use limited to: UNIVERSITY COLLEGE CORK. Downloaded on January 11,2023 at 15:39:22 UTC from IEEE Xplore. Restrictions apply.
1724 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 27, NO. 12, DECEMBER 2008

Fig. 1. Electrical circuit of MR probe: The MR signal is received by a resonant


Fig. 3. (a) Optical transmitter containing the laser diode chip and the transistor
chip mounted on an Al O -SMD-substrate. (b) Schematic drawing of the op-
circuit (L1, C1) and transmitted via an optical transmitter (T1, LD) which is
powered by an optical power supply.
tical transmitter.

Fig. 2. Flexible foil-based micro-Helmholtz coils used as catheter receiver coil Fig. 4. Photovoltaic power converter PPC with a light sensitive area subdivided
on a five French catheter tube. into three diodes connected in series.

A. Optical MR Probe pair of crossed diodes, which also suppresses spin saturation
The principle setup of the MR probe is shown schematically during RF excitation.
in Fig. 1. The MR signal is received by a resonant circuit (L1, The entire transmitter was fabricated on an -SMD-sub-
C1) connected to the gate of a transistor T1. The transistor strate, which was metalized and structured by lithography
modulates the current through a laser diode LD which emits (Fig. 3). For the conductors, a titanium-copper-gold layer was
an optical signal modulated by the MR signal. To power the deposited. The titanium improves the adhesion, the copper
transmitter, an optical power supply is used to avoid conducting serves as a highly conductive material, and the gold layer
cables which are prone to RF induced heating. protects the copper layer from oxidation.
1) Receiver Coils: Dedicated foil-based flexible micro- 3) Optical Power Supply: To avoid any long conducting ca-
Helmholtz-coils were fabricated for optimal integration into bles, the power for the optical transmitter is supplied optically
the catheter tip (Fig. 2). The coils were structured using a by a laser diode (ADL-85501TL, Thorlabs) located outside the
UV-LIGA process and electroplating applied to the planar catheter. The optical power is transmitted via a fiber (GIF 625,
polyimide foil. Subsequently, the foil-based flexible coils were 125 , Thorlabs) to a photovoltaic power converter PPC (PPC-
wrapped around five French catheter tips [11]. Each coil was 3Me, 1 1 0.25 mm , JDSU, Milpitas, CA) in the optical MR
tuned to the Larmor frequency by capacitor C1. The resonance probe at the tip. The PPC has a light sensitive area with a diam-
of this receive circuit was measured with and without being eter of 0.8 mm and is subdivided into three diodes connected in
connected to the optical transmitter using weak inductive cou- series to achieve a voltage sufficient for the VCSEL. For max-
pling of the micro coil to a pickup coil connected to a network imum efficiency it is essential to illuminate the three diodes
analyzer. The Q-factor was determined from the width of the equally, because the weakest illuminated element will limit the
resonance according to current. This is achieved by arranging the diodes in 120 seg-
ments (Fig. 4). Due to the space requirements the PPC with its
(1) width of 1 mm was laid flat in the catheter tube.
A dedicated reflector was developed, to deflect the light beam
2) Optical Transmitter: The optical transmitter consists of a from the fiber end to the light sensitive area of the PPC (Fig. 5).
transistor chip (J-FET 2N4393, 530 460 200 , Central The reflector was anisotropically wet-etched into a 500
Semiconductor, Hauppauge, NY [12]) and a laser diode chip silicon substrate by KOH, which stops etching at the 111 crystal
(VCSEL ULM850, 250 250 0.15 , Ulm Photonics, plane leaving the reflector under an angle . The nu-
Ulm, Germany) with a very low threshold current of 1 mA, merical aperture of NA=0.275 of the fiber corresponds to an
which is coupled to an optical fiber (GIF 625, 125 , Thor- aperture angle of
labs, Newton, NJ). The gate of the transistor is protected from
voltages inducted by the high-power RF excitation pulse by a (2)

Authorized licensed use limited to: UNIVERSITY COLLEGE CORK. Downloaded on January 11,2023 at 15:39:22 UTC from IEEE Xplore. Restrictions apply.
FANDREY et al.: DEVELOPMENT OF AN ACTIVE INTRAVASCULAR MR DEVICE WITH AN OPTICAL TRANSMISSION SYSTEM 1725

TABLE I
Q-FACTOR OF MICRO COIL IN AIR AND IMMERSED INTO PHANTOM LIQUID

C. MR Phantom Measurements
Imaging and tracking experiments were performed in a 1.5-T
MR system (Achieva, Philips Medical Systems, Best, The
Netherlands). The receiving coil was placed in a bowl filled
with standard phantom liquid ( doped with 0.08%
and 0.2% NaCl) and the optical receiver was connected to the
MR preamplifier.
1) MRI Experiments: Microscopic MR images were ac-
quired with the coil axis parallel and perpendicular to the image
plane (FOV 64 mm, resolution 0.5 mm). In order to assess
the transmisson loss of the optical transmitter, images were
Fig. 5. Reflector concept for powering the optical power supply. a) V-groove
used to align the fiber. b) Reflector area covered with a thin gold layer to maxi-
acquired in two modes, either using optical transmission or
mize its reflectance. c) Illumination of light sensitive area defined by the angle using a direct electrical connection of the receive circuit (L1,
of the reflector = 54 7 : and the numerical aperture of the fiber = 15 96: . C1, crossed diodes) to the MR preamplifier. This second mode
bypassed the optical transmission and served as a benchmark.
The signal-to-noise ratio (SNR) was determined in all images.
It was calculated from the maximum signal intensity S at the
location of the receive coil and the mean noise signal of the
zero signal region in the modulus images according to [13]

(3)

2) MR Tracking Experiments: Fast projection-based tip


tracking was performed during real-time MRI. For the tracking
projections, the entire volume was excited, and frequency
Fig. 6. Reflector with V-groove etched in KOH; the reflecting area is covered encoded projections were acquired on the catheter channel for
with a titan gold layer for high performance.
the x, y, and z direction: From the position of the signal peak
in the projections, the 3-D-position of the tip was calculated
automatically by the MR system [14], and the imaging slice was
Under these constraints the distance and the height of the repositioned to show the tip. Additionally, the catheter tip was
fiber relative to the center of the power-device was calculated moved manually in the bowl along a ”P”-shaped path during
to achieve equal illumination of the elements. The reflector was continuous tracking of the tip position. All measured tracking
covered with a thin gold layer to maximize its reflectance. A positions were recorded and displayed in the MR image.
V-groove etched in the same process step as the reflector was
used to align the fiber (Fig. 6). Finally, the illuminated reflector III. RESULTS AND DISCUSSION
was carefully aligned with the PPC to achieve maximal output
power and fixated in that position. A. Optical MR Probe

B. Optical Receiver In this section the results of the characterization of the op-
tical MR probe components are presented and discussed. Table I
The optical signal from the fiber is demodulated by an op- shows the measured Q-factors of the MR probe for various con-
tical receiver outside of the catheter. The receiver (HFBR-2416, figurations. The isolated receiver circuit (L1,C1) in air provides
Agilent Technologies, Santa Clara, CA) contains a PIN photo- a high Q of 41.5, which demonstrates the quality of the minia-
diode and a low noise transimpedance preamplifier with a band- ture photolithographic coil design. In liquid, the Q-factors for
width of 125 MHz. The demodulated electrical signal is then all configurations decrease due to dielectric losses in the liquid.
transmitted to the MR preamplifier. The transmission charac- Tuning and matching of the receiver circuit to 50 for direct
teristic of the optical transmission system was measured with a electrical transmission leads to a drop of Q-factors by about
spectrum analyzer. a factor 2 as expected, because the effective series resistances

Authorized licensed use limited to: UNIVERSITY COLLEGE CORK. Downloaded on January 11,2023 at 15:39:22 UTC from IEEE Xplore. Restrictions apply.
1726 IEEE TRANSACTIONS ON MEDICAL IMAGING, VOL. 27, NO. 12, DECEMBER 2008

TABLE II
SNR COMPARISON OF DIRECT ELECTRICAL TRANSMISSION
AND OPTICAL TRANSMISSION

the micro-tube with a wall thickness of 0.3 mm can be seen on


which the micro-coil is mounted.
Table II shows the SNR of three measurements. The direct
Fig. 7. MR micro-imaging acquired with the micro-Helmholtz coil: a and b electrical transmission technique reveals a 20 dB higher SNR
acquired with direct electrical transmission. (a) Coil axis through plane. (b) Coil
axis in plane. (c) and (d) Acquired via the optical transmission system: (c) coil
compared with the present optical transmission system. This
axis through plane and (d) coil axis in plane. SNR loss can be partly explained by the degradation of the
Q-factor caused by the optical transmission system as discussed
in the section above.
double. Interestingly, connecting the resonant circuit to the op- Fig. 8(a) shows an exemplary 1-D projection as used for
tical transmitter caused a much higher decrease of the Q-factor catheter tracking. Fig. 8(b) contains about 50 subsequently
from 29.4 to 6.3. This is explained by a high gate capacitance of measured tracking positions while moving the tip manually
20 pF of the transistor that has a low Q-factor, which causes an along a P-shape. The absence of any false position measure-
additional finite load to the resonant circuit. In a signal transmis- ments demonstrates the robustness of catheter tracking with
sion system it is important to use an amplifier with high gain at optical signal transmission in the phantom.
the input to keep the overall noise figure low. For the MR probe Obviously, the SNR loss of 20 dB in optical transmission with
a degradation of the Q-factor, which in this case is the first am- respect to direct electrical transmission is severe. However, a de-
plification step, causes a degradation of SNR for the complete cent comparison must include the aspect of RF safety, i.e., op-
transmission system. The specific JFET in our setup was chosen tical transmission should be compared to alternative methods of
because it was available as plain chip on short notice. It serves RF safe transmission. To this end, the concept of using minia-
for the first proof of concept but will be replaced by a different ture transformers to improve RF safety [5], [6] is the most rel-
device with a high Q gate capacitance in future. evant directly analogous method, since it also provides general
The measured dynamic range of the transmitter was 50 dB, purpose MR signal transmission. Each transformer contributes
with the noise floor level at . This transmitter was used an SNR loss of about 2.5 dB, and for an efficient reduction of
at a very low electrical power level of 2.21 mW at a bias current RF induced heating of a catheter in a 1.5-T MR system four
of 1.3 mA provided by the optical power converter. In this ap- transformers are used, resulting in a loss of 10 dB. The optical
plication it is important to keep the power level low, because transmission reveals yet a higher loss, but provides some clear
any power will also generate heat, which has to be avoided. For advantages. The losses do not scale with the field strength as
this reason the VCSEL with the very low threshold current of in the transformer concept, where the required number of trans-
1 mA was chosen. The dedicated reflector for the PPC reduced formers increases with field strength [15]. Most importantly, the
the efficiency by only 4% from 33% to 29%. When fixating the optical transmission is inherently RF safe, while the safe use of a
reflector to the PPC with optical adhesive, the adhesive uninten- transformer cable has to be proven under clinical circumstances
tionally penetrated into the gap of the reflector due to capillary which include a number of variables. Finally, the SNR of the
forces, which decreased the efficiency to 21% due to the change optical transmission is expected to improve as discussed above.
in the acceptance angle. In the future, this will be compensated
by adjusting the distance of the fiber to the reflector. IV. SUMMARY AND CONCLUSION
The safe application of active intravascular MR devices re-
B. MR Phantom Measurements
quires special means to avoid RF induced heating. Therefore,
Fig. 7(a) and (b) depicts the images taken with the micro- long conducting cables have to be avoided. In this paper, an
Helmholtz coil transmitted directly by a cable for a horizontal all-optical system for a safe transmission of MR signals for in-
slice with the coil axis through plane [Fig. 7(a)] and vertical travascular imaging and device tracking is presented. The op-
slices with the coil axis in plane [Fig. 7(b)]. Fig. 7(c) and (d) tical transmission system is designed with small scale compo-
shows the results for the same measurement with the signal nents and examined in MR and transmission experiments. The
transmitted via the optical transmission system. The resulting optical transmitter is based on a small laser diode and a transistor
images show a high spatial resolution of the micro coil; even integrated in chip form. The power supply of this transmitter

Authorized licensed use limited to: UNIVERSITY COLLEGE CORK. Downloaded on January 11,2023 at 15:39:22 UTC from IEEE Xplore. Restrictions apply.
FANDREY et al.: DEVELOPMENT OF AN ACTIVE INTRAVASCULAR MR DEVICE WITH AN OPTICAL TRANSMISSION SYSTEM 1727

with a direct transmission. Tracking is already feasible with this


all optical system, but the noise figure should be improved for
an intravascular imaging application, which is currently being
pursued.
In the future, all components will be integrated into the lumen
of the catheter next to the micro-coil. With the current size of
the optical power converter, this system can be integrated into a
seven French catheter.

REFERENCES
[1] E. Atalar, P. Bottomley, O. Ocali, L. Correia, M. Kelemen, J. Lima, and
E. Zerhouni, “High resolution intravascular MRI and MRS by using a
catheter receiver coil,” Magn. Reson. Med., vol. 53, pp. 226–230, 2005.
[2] C. Farrar, V. Wedeen, and J. Ackerman, “Toward rapid high resolu-
tion in vivo intravascular MRI: Evaluation of vessel wall conspicuity
in a porcine model using multiple imaging protocols,” J. Magn. Reson.
Imag., vol. 23, pp. 135–144, 2006.
[3] C. Hillenbrand, J. Jesberger, E. Wong, S. Zhang, D. Chang, F. Wacker,
J. Lewin, and J. Duerk, “Cylindrical meanderline radiofrequency coil
for intravascular magnetic resonance studies of atherosclerotic plaque,”
Magn. Reson. Med., vol. 53, pp. 226–230, 2005.
[4] M. Ladd, H. Quick, J. Debatin, G. von Schulthess, and G. McKinnon,
“Resonant heating of intravascular RF coils,” in Proc. 6th Annu.
Meeting ISRM, 1998, vol. 35, p. 473.
[5] S. Weiss, P. Vernickel, T. Schaeffter, V. Schulz, and B. Gleich, “Trans-
mission line for improved RF safety of interventional devices,” Magn.
Reson. Med., vol. 54, pp. 182–189, 2005.
[6] P. Vernickel, V. Schulz, S. Weiss, and B. Gleich, “A safe transmis-
sion line for MRI,” IEEE Trans. Biomed. Eng., vol. 52, no. 6, pp.
1094–1102, Jun. 2005.
[7] M. Konings, L. Bartels, C. van Swol, and C. Bakker, “Development
of an MR-safe tracking catheter with a laser-driven tip coil,” J. Magn.
Reson. Imag., vol. 13, pp. 131–135, 2001.
[8] S. Weiss, T. Kuehne, F. Brinkert, G. Krombach, M. Katoh, T. Scha-
effter, R. Guenther, and A. Buecker, “In vivo safe catheter visualiza-
tion and slice tracking using an optically detunable resonant marker,”
Magn. Reson. Med., vol. 52, pp. 860–868, 2004.
[9] T. Uelzen, Development of a fast localisable patient-safe MR-catheter
for MR-guided intravascular interventions based on micro systems
technology (LOMKAT) Teltow, Germany, Info-Börse VDI/VDE-IT
Nr. 55, 2004.
[10] M. Bock, R. Umathum, J. Sikora, S. Brenner, E. Aguor, and W.
Semmler, “A Faraday effect position sensor for interventional mag-
netic resonance imaging,” Phys. Med. Biol., vol. 51, pp. 999–1009,
2006.
Fig. 8. (a) Projection plot of catheter channel signal. From the position of the [11] T. Uelzen, S. Fandrey, and J. Müller, “Mechanical and electrical prop-
signal peak in three projections, the 3-D-position of the tip was calculated auto- erties of electroplated copper for MR-imaging coils,” J. Microsyst.
matically by the MR system. (b) Catheter tip was moved manually in the bowl Technol., vol. 12, no. 4, pp. 343–351, 2006.
along a “P”-shaped path during continuous tracking of the tip position. All sub- [12] PROCESS CP206 Small Signal Transistors N—Channel Switch/
sequent tip positions were marked by the MR system with a cross. Chopper J FET Chip. New York: Central Semiconductor Corp.,
2003.
[13] A. Papoulis and S. Pillai, Probability, Random Variables, and Sto-
chastic Processes, 4th ed. New York: McGraw-Hill, 2002, pp.
consists of a special photovoltaic power converter PPC, which 148–148.
[14] C. L. Dumoulin, S. P. Souza, and R. D. Darrow, “Real-time position
is coupled to an active supply via a low loss reflector to achieve monitoring of invasive devices using magnetic resonance,” Magn.
a maximum efficiency of the opto-electrical power conversion. Reson. Med., vol. 29, pp. 411–415, 1993.
Projection-based tip tracking was achieved with the presented [15] S. Weiss, S. Fandrey, P. Vernickel, T. Schaeffter, and B. Gleich, “An
active tracking catheter for 3T with improved RF safety using a trans-
setup, and high-resolution images were obtained in a phantom. former based transmission line,” in Proc. 6th Interventional MRI Symp.,
The obtained images still have a 20 dB lower SNR compared 2006, p. 12.

Authorized licensed use limited to: UNIVERSITY COLLEGE CORK. Downloaded on January 11,2023 at 15:39:22 UTC from IEEE Xplore. Restrictions apply.

You might also like