Professional Documents
Culture Documents
86502-4A Diamond Service Manual Jun 2001 ENG
86502-4A Diamond Service Manual Jun 2001 ENG
6HUYLFH0DQXDO
June 2001 86502-4A
COPYRIGHT 2001 by Instrumentarium 1. PLACE FOR OPERATION
Imaging
INSTRUCTIONS
Documentation, trademark and the software
are copyrighted with all rights reserved.
Under the copyright laws the documentation 2. PLACE FOR INSTALLATION
may not be copied, photocopied, INSTRUCTIONS
reproduced, translated, or reduced to any
electronic medium or machine readable
form in whole or part, without the prior 3. SERVICE PROGRAMS AND
written permission of Instrumentarium
Imaging. SOFTWARE LOAD
This document is the English-language
original. In case of interpretation disputes 4. ELECTRICAL SYSTEM
the English text applies. ARCHITECTURE
Instrumentarium Imaging reserves the right
to revise this publication from time to time 5. MECHANICAL SYSTEM
and to make changes in the content of it
without obligation to notify any person of PREVENTIVE MAINTENANCE
such revision or changes.
127(6&$87,216$1':$51,1*6
:$51,1* $/:$<6,167$//7+(3523(5)86(6,1727+(81,7)$,/85(
72'262-(23$5',=(63$7,(1723(5$725$1'(48,30(17
6$)(7<
:$51,1* 7+,681,7+$6)/2$7,1**5281'6$/:$<60$.($//6&23(
0($685(0(176',))(5(17,$//<$1'$/:$<6%(&$5()8/
:+(16(59,&,1*7+(81,7
&$87,21 :+(1+$1'/,1*&,5&8,7%2$5'6$/:$<687,/,=(*5281'
:5,67675$3625*5281',1*.,767235(9(17'$0$*(72
7+(0267<3(&+,36$/:$<6.((3&,5&8,7%2$5'6,1$
67$7,&3522)%$*
:$51,1* 7+,6 ;5$< 81,7 0$< %( '$1*(5286 72 3$7,(17 $1'
23(5$725 81/(66 6$)( (;32685( )$&7256 $1'
,16758&7,216$5(2%6(59('
5HYLVLRQKLVWRU\WDEOH
5HYLVLRQOHYHO 'DWHRIDPHQGPHQW 1RWHV
- June 11, 2001 New manual
7$%/(2)&217(176
23(5$7,21,16758&7,216
,167$//$7,21,16758&7,216
',$021'6(59,&(352*5$06$1'62)7:$5(/2$'
29(59,(:)81&7,21$/'(6&5,37,21
6(59,&(02'(29(59,(:
3.2.1 A: System setup ...................................................................................... 3
3.2.2 B: Defaults.............................................................................................. 4
3.2.3 C: Diagnostics - Not in use at the moment............................................. 9
3.2.4 D: AEC ................................................................................................. 10
3.2.5 E: C-arm Rotation................................................................................. 20
3.2.6 F: Collimator ........................................................................................ 22
3.2.7 G: Compression .................................................................................... 24
3.2.8 H: Filter ................................................................................................ 27
3.2.9 I: Generator........................................................................................... 28
3.2.10 J: Tube Park.......................................................................................... 30
3.2.11 K: Stereo Movement............................................................................. 31
3.2.12 Exiting the service mode ...................................................................... 32
/2$',1*1(:62)7:$5(
+<3(57(50,1$/,17(5)$&()25',$021'
3.4.1 Configuration of the HyperTerminal: ................................................... 38
3.4.2 Capturing the data................................................................................. 39
3.4.3 Saving the settings data ........................................................................ 40
',$021'7(67',63/$<6
-7$*/2$'(5
(/(&75,&$/6<67(0$5&+,7(&785(7+(25<2)23(5$7,21
29(59,(:
,138732:(5
4.2.1 Supply Voltage Regulation..................................................................... 3
35,17('&,5&8,7%2$5' 3&% &20081,&$7,21
4.3.1 Internal Serial Bus .................................................................................. 4
4.3.2 External Serial Bus ................................................................................. 4
6(16256
4.12.1 Cassette Holder Sensor......................................................................... 18
4.12.2 Cassette Sensors ................................................................................... 19
4.12.3 Grid Sensor........................................................................................... 20
4.12.4 Collimator Sensors ............................................................................... 20
4.12.5 Filter Sensor.......................................................................................... 20
4.12.6 C-arm Rotation Sensors........................................................................ 21
4.12.7 Compression Sensors............................................................................ 22
4.12.8 Tube Park Movement Sensors.............................................................. 24
4.12.9 Stereo Sensors ...................................................................................... 24
)86(692/7$*(6$1'/(',1',&$7,216
0(&+$1,&$/6<67(0&+(&.6$1'35(9(17,9(0$,17(1$1&(
;5$<78%(3(5)250$1&(&+(&.
5.1.1 Half-value layer (HVL) measurement .................................................... 1
5.1.2 Tube efficiency check on the cassette holder ......................................... 3
5.1.3 Focal spot measurement procedure ........................................................ 4
5.1.4 Image quality evaluation procedure........................................................ 6
5.1.5 Measurement of the feedback circuitry .................................................. 8
5.1.6 kV, mAs and exposure time test............................................................. 9
$8720$7,&(;32685(&21752/(9$/8$7,21
5.2.1 Quick AEC test..................................................................................... 10
5.2.2 Exporting and storing the system settings data..................................... 10
5.2.3 Automatic AEC detector gain adjustment ............................................ 11
5.2.4 Adjustment of an individual detector ................................................... 11
0(&+$1,&$/&+(&.6
5.3.1 Compression system ............................................................................. 12
5.3.2 C-arm rotation mechanism ................................................................... 12
5.3.3 Column and carriage............................................................................. 12
5.3.4 Compression force checking ................................................................ 12
5.3.5 Testing the x-ray field illumination ...................................................... 13
5.3.6 Adjustment tube head movement and torque ....................................... 14
5(*8/$50$,17(1$1&(3/$16
5(3$,5,16758&7,216
&2/8015(3$,5
6.1.1 Preparing the carriage and counterweight............................................... 1
6.1.2 Replacement of the Z-motor................................................................... 2
6.1.3 Replacement of the column cap.............................................................. 3
6.1.4 Replacement of the counterweight wires................................................ 4
/2&$7,212)(/(&75,&$/%2$5'6$1'0$,16
75$16)250(5
6.2.1 Replacement of Power, Supervisor and Generator boards ..................... 9
6.2.2 Replacement of the AMD board............................................................. 9
6.2.3 Replacement of Control Panel Board ..................................................... 9
6.2.4 Replacement of the Compression Board or Compression Display Board10
6.2.5 Replacement of the mains transformer ................................................. 11
&$505(3$,5
6.3.1 Removal and assembly of the Diamond covers.................................... 12
6.3.2 Attaching the tube head covers............................................................. 13
6.3.3 Replacement of handles........................................................................ 20
6.3.4 Replacement of the vertical drive control switch ................................. 21
6.3.5 Replacement of the x-ray tube .............................................................. 21
6.3.6 Replacement of the high voltage unit ................................................... 22
6.3.7 Replacement of the compression system .............................................. 23
6.3.8 Replacement of the compression motor ............................................... 25
6.3.9 Replacement of the compression thickness measurement system........ 28
6.3.10 Adjustment of the compression mechanism belt .................................. 30
6.3.11 Adjustment of the mechanical compression limit ................................ 31
6.3.12 Replacement of the C-arm motor ......................................................... 32
6.3.13 Replacement of ECS mechanism/ECS motor ...................................... 33
27+(5$'-8670(176255(3/$&(0(176
6.4.1 Replacement of the light field bulb ...................................................... 38
6.4.2 Grid alignment ...................................................................................... 38
6.4.3 Grid movement mechanism check........................................................ 40
6.4.4 Replacement of Bucky carbon fiber ..................................................... 42
6.4.5 Replacement of c-arm membrane buttons ............................................ 42
7528%/(6+227,1*
,1)250$7,21)257+(6(59,&(&$//
',$021'(5525&2'(6
7.2.1 Diamond error code chart ....................................................................... 2
1232:(5,1',&$7,216%2$5'/('6
$(&,123(5$7,9(
6833/<92/7$*(6
*5,'/,1(6
(&67528%/(6+227,1*
6(5,$/%86&20081,&$7,216&+(&.
127(62162)7:$5(83*5$',1*
48(67,216 $16:(56
63(&,),&$7,216
6&+(0$7,&6/$<2876
&$%/('5$:,1*6
$33(1',&(6
',$021'6(59,&(352*5$06$1'62)7:$5(/2$'
29(59,(:)81&7,21$/'(6&5,37,21
Main Diamond operation modes are Auto kV, Auto Time and Manual mode. In
Auto-kV mode, Diamond determines the required amount of kV. In Auto Time
mode, Diamond determines the required exposure time. In Manual mode, the user
sets up all the required values for the examination.
Many programmable parameters that affect these modes can be set and adjusted
in Diamond service mode. Generation of radiation may be switched off during the
service work. This enables the service technician to make test exposures to verify
the correct voltages on various boards. Chapter 7.5 in Diamond Installation
Instructions (86501) describe how to disable the emission of x-rays (for example
at exhibition).
127( 'LDPRQGWXEHKHDGDVVHPEO\LQFRUSRUDWHVDKHDWLQWHJUDWRUWKDWFRROLQJWLPH
RIWRVHFEHWZHHQH[SRVXUHLVLPSOHPHQWHGDOVRLQWKHVHUYLFHPRGHDV
LQWKHQRUPDOPRGH7KLVZLOOSURWHFWWKHXQLWIURPEHLQJRYHUORDGHGGXULQJ
WHVWLQJ
This manual has been divided into separate software, mechanical, service,
maintenance and troubleshooting sections. Schematics, layouts and cables have
been placed at the end of this manual.
The software sections helps to calibrate and to set up correct parameters for the
user or change the existing ones. The mechanical section explains about tests and
other mechanical issues not covered in software and electronics section. The
electronics section explains about the operation of the PCB-boards, various
connections and measurements. The maintenance section covers the regular
maintenance procedures. The troubleshooting sections include error codes and
solutions to some problems that might come about.
6(59,&(02'(29(59,(:
Service program menu contains special programs
for system setups, calibration and diagnostics. You
can enter into the service mode by pressing
and buttons of the control panel down
simultaneously.
Diamond service mode interface have been divided into several topic groups in
which the system setting and parameters can be changed:
Scroll with or -keys until the required group is between the brackets
as illustrated in the example . Press to enter
the group. The following sections describe each group and available parameter
choices. In parameter sub-screens, accept the changes with -key or
cancel the changes with the -key. Then Diamond returns to the group screen.
You can exit the service mode by pressing and -buttons down
simultaneously.
$6\VWHPVHWXS
The system and communication parameters can be set and changed in 6<67(0
6(783 screen:
6\VWHP3DUDPHWHUV3U$
3DUDPHWHU 5DQJH 1RWHV
6HULDOQR 12345 Serial number of the unit
&RXQWU\FRGH 123 Not in use at the moment.
(IILFLHQF\0R At 28 kV R/100mAs
(IILFLHQF\5K At 28 kV R/100mAs
(IILFLHQF\$O At 39 kV R/100mAs
+9/0R At 28 kV mmAl value
+9/5K At 28 kV mmAl value
+9/$O At 39 kV mmAl value
&RPPXQLFDWLRQ6HWXS3U$1RWLQXVHDWWKHPRPHQW
3DUDPHWHU 5DQJH 1RWHV
/DEHORXW Com1 Prints out the patient data labels
for films (ID camera). Not in use
at the moment.
/DEHOW\SH 1 to 3 The number of rows. Not in use at
the moment.
'DWDRXW Com2 The port that sends out the
exposure data.
&RP 19200 The speed for Com1 port. Used in
software updates and diagnostics.
&RP 1200 The speed for Com2 port. Used
by ID II film flasher.
([LWLQJWKH6<67(06(783VFUHHQ
If you accept the changes, press -key. If you do not want changes to take place,
press -key. Diamond returns to *52836 screen.
%'HIDXOWV
The default parameters can be set and changed in '()$8/76 screen:
([LWLQJWKH'()$8/76VFUHHQ
If you accept the changes, press set -key. If you do not want changes to take
place, press cancel-key . Diamond returns to *52836 screen.
&'LDJQRVWLFV1RWLQXVHDWWKHPRPHQW
Various system components can be checked in ',$*1267,&6 screen:
([LWLQJWKH',$*1267,&6VFUHHQ
If you accept the changes, press set -key. If you do not want changes to take
place, press cancel-key . Diamond returns to *52836 screen.
'$(&
The AEC default parameters can be set and changed in service program D00:
127( 3ULRUWRDQ\$(&ZRUNDQGWHVWLQJVHWWDUJHWWLPHVDQGPDNHVXUHWKDWN9V
DUH2.
)XQFWLRQDOGHVFULSWLRQRI$(&IXOODXWRPDWLFPRGH
The Auto kV performance can be optimized and configured according to the user
preferences. The configurable items are the LQLWLDON9 (initial kV versus breast
thickness table), amount of DOORZDEOH N9FKDQJH(Max kV change in service
program '$(&2SWLFDO 'HQVLWLHV, PrD03) and desired exposure time range
(WDUJHWWLPH).
7KLFNQHVVFRPSHQVDWLRQDGMXVWPHQW
5RWDWLRQSRLQW3
Rotation point is the exposure time where the O.D is adjusted correctly. Thickness
compensation is achieved by altering the exposure time around this point. Default
value is 1.00.
$PRXQWRIURWDWLRQUDQGU
The values of r1 and r2 can vary from 0.70 to 1.30. If the parameter is 1.30 it
means that exposure times are increased by 30%. If the parameter is 0.80 it means
that the exposure times are decreased by 20% etc. Default values are 1.0 for both.
&RPSHQVDWLRQFXUYHVORSHSDUDPHWHUV6DQG6
Parameters S1 and S2 determine the curve slope for exposure times shorter or
longer than the URWDWLRQSRLQW respectively. S1 determines the slope at the point
where exposure time is lower than the URWDWLRQSRLQW. S2 determines the amount
of compensation at exposure time higher than the URWDWLRQSRLQW. Refer to the
curves below for effects of the S parameters. The rotation point is 1.0 seconds and
the amount of rotation (r1 and r2) is 0.8, 0.9, 1.1 or 1.2 (= 80%, 90%, 110%,
120%). Default values are 200 for both.
&RPSHQVDWLRQFXUYHVORSHSDUDPHWHUV6DQG6
Parameters S1 and S2 determine the curve slope for exposure times shorter or
longer than the URWDWLRQSRLQW respectively. S1 determines the slope at the point
where exposure time is lower than the URWDWLRQSRLQW. S2 determines the amount
of compensation at exposure time higher than the URWDWLRQSRLQW. Refer to the
curves below for effects of the S parameters. The rotation point is 1.0 seconds and
the amount of rotation (r1 and r2) is 0.8, 0.9, 1.1 or 1.2 (= 80%, 90%, 110%,
120%). Default values are 200 for both.
S=0
1.4 1.4
S=0
S = 700
1.3 1.3
r1 = 1.2
1.2 1.2
S = 700 r1 = 1.1
1.1 1.1
P = 1.0sec P = 1.0sec
1 1
0.9 0.9
S = 700
r1 = 0.9
0.8 0.8
r1 = 0.8
0.7 0.7
S=0 S = 700
S=0
0.6 0.6
0 0.2 0.4 0.6 0.8 1 0 0.2 0.4 0.6 0.8 1
Exposure time / sec. Exposure time / sec.
0.9 0.9
r2 = 0.9 S=0
S = 50
0.8 S = 100 0.8
S = 200
S = 350 r2 = 0.8 S=0
0.7 S = 500 0.7
S = 600
S = 700 S = 700
0.6 0.6
1 1.5 2 2.5 3 3.5 4 4.5 5 1 1.5 2 2.5 3 3.5 4 4.5 5
Exposure time / sec. Exposure time / sec.
$(&6HWXS3U'
3DUDPHWHU 5DQJH 1RWHV
7DUJHW2' 1.00 to 2.00 Adjustable O.D. range. Not in use
at the moment.
7DUJHWWLPHOLPLWORZ 0.2 to 1.00 Lower limits for O.D.
7DUJHWWLPHOLPLWKLJK 1.00 to 3.00 Higher limits for O.D.
$XWR3RLQWOLPLW 0 to 2500 Default value 1500.
If images are too dark when exposing at medio-lateral projection, a detector under
muscle might be in use (solid ellipses in the image below). Increase the AutoPoint
limit value.
$(&JDLQV3U'
3DUDPHWHU 5DQJH 1RWHV
$(&JDLQ J 0.5 to 2.00 Software gain to obtain the
required optical density. This gain
affects all detectors.
'HWHFWRU/ 'HW 0.7 to 1.30 Detector specific gain factor that
Diamond adjusts in relation to
detector C1 performance. Values
less than 1.0 increase gain and
produce darker images. Values
over 1.0 decrease gain and
produce lighter images.
'HWHFWRU/ 'HW 0.7 to 1.30 As above.
'HWHFWRU5 'HW 0.7 to 1.30 As above.
'HWHFWRU5 'HW 0.7 to 1.30 As above.
'HWHFWRU& 'HW 0.7 to 1.30 Benchmark detector.
'HWHFWRU& 'HW 0.7 to 1.30 As above.
'HWHFWRU& 'HW 0.7 to 1.30 As above.
'HWHFWRU& 'HW 0.7 to 1.30 As above.
4=C4 80%
5=C3 90%
7=
98
L2
%
R2
6=C2 102%
3=
98
The graph above describes how various detectors receive radiation. In order to
equalize them, detector specific factors are used for adjustment. Detector C1 is
used as a benchmark.
2SWLFDO'HQVLWLHV3U'
3DUDPHWHUVXEVFUHHQV 1RWHV
$XWRN9 Optical density settings for Auto kV.
$XWR7LPH Optical density settings for Auto Time.
7KLFNQHVV&RPSHQVDWLRQ Optical density settings for Thickness
compensation.
N9&RPSHQVDWLRQ Optical density settings for kV compensation.
7DUJHW2' ,4& Not in use at the moment.
2SWLFDO'HQVLWLHV$XWRN9VXEVFUHHQ3U'
3DUDPHWHU 5DQJH 1RWHV
$XWRN9%XFN\ Adjustment when using the Bucky.
Mo -7.0 to +7.0 Increases or decreases the optical
density of the image at all kV
values.
Rh -7.0 to +7.0 As above.
Al -7.0 to +7.0 As above.
$XWRN90DJQLILFDWLRQ Adjustment for the use of Magnification tunnel.
Mo -7.0 to +7.0 Increases or decreases the optical
density of the image at all kV
values.
Rh -7.0 to +7.0 As above.
Al -7.0 to +7.0 As above.
$XWRN9&RQWDFW Adjustment using the Contact tunnel.
Mo -7.0 to +7.0 Increases or decreases the optical
density of the image at all kV
values.
Rh -7.0 to +7.0 As above.
Al -7.0 to +7.0 As above.
2SWLFDO'HQVLWLHV$XWR7LPHVXEVFUHHQ3U'
3DUDPHWHU 5DQJH 1RWHV
$XWR7LPH%XFN\ Adjustment when using the Bucky.
Mo -7.0 to +7.0 Increases or decreases the overall
optical density of the image.
Rh -7.0 to +7.0 As above.
Al -7.0 to +7.0 As above.
$XWR7LPH Adjustment when using the Magnification
0DJQLILFDWLRQ tunnel.
Mo -7.0 to +7.0 Increases or decreases the overall
optical density of the image.
Rh -7.0 to +7.0 As above.
Al -7.0 to +7.0 As above.
$XWR7LPH&RQWDFW Adjustment when using the Contact.
Mo -7.0 to +7.0 Increases or decreases the overall
optical density of the image.
Rh -7.0 to +7.0 As above.
Al -7.0 to +7.0 As above.
2SWLFDO'HQVLWLHV7KLFNQHVV&RPSHQVDWLRQVXEVFUHHQ3U'
3DUDPHWHU 5DQJH 1RWHV
%XFN\ Adjustment when using the Bucky.
’P’ Rotation point of compensation
curve (Exposure time in seconds
where required optical density
was received).
’r1’ Compensation factor for exposure
times shorter than P.
’r2’ Compensation factor for exposure
times longer than P.
’S1’ Compensation parameter for very
short exposure times (fine-
tuning).
’S2’ Compensation parameter for very
long exposure times (fine-tuning).
0DJQLILFDWLRQ Adjustment when using the Magnification tunnel.
’P’ Rotation point of compensation
curve (Exposure time in seconds
where required optical density
was received).
’r1’ Compensation factor for exposure
times shorter than P.
’r2’ Compensation factor for exposure
times longer than P.
’S1’ Compensation parameter for very
short exposure times (fine-
tuning).
’S2’ Compensation parameter for very
long exposure times (fine-tuning).
&RQWDFW Adjustment when using the Contact.
’P’ Rotation point of compensation
curve (Exposure time in seconds
where required optical density
was received).
’r1’ Compensation factor for exposure
times shorter than P.
’r2’ Compensation factor for exposure
times longer than P.
’S1’ Compensation parameter for very
short exposure times (fine-
tuning).
’S2’ Compensation parameter for very
long exposure times (fine-tuning).
2SWLFDO'HQVLWLHVN9&RPSHQVDWLRQVXEVFUHHQ3U'
3DUDPHWHU 5DQJH 1RWHV
N9&RPSHQVDWLRQ%XFN\ Adjustment when using the Bucky.
Mo 0.70 - 1.30 If image at certain kV is too light
or dark, adjust this kV-specific
value.
Rh 0.70 - 1.30 As above.
Al 0.70 - 1.30 As above.
N9 &RPSHQVDWLRQ Adjustment when using the Magnification tunnel.
0DJQLILFDWLRQ
Mo 0.70 - 1.30 If image at certain kV is too light
or dark, adjust this kV-specific
value.
Rh 0.70 - 1.30 As above.
Al 0.70 - 1.30 As above.
N9 &RPSHQVDWLRQ Adjustment when using the Contact.
&RQWDFW
Mo 0.70 - 1.30 If image at certain kV is too light
or dark, adjust this kV-specific
value.
Rh 0.70 - 1.30 As above.
Al 0.70 - 1.30 As above.
2SWLFDO'HQVLWLHV'HWHFWRU&RPSHQVDWLRQVXEVFUHHQ3U'
1RWLQXVHDWWKHPRPHQW
3DUDPHWHU 5DQJH 1RWHV
'HWHFWRUV 0.5 - 2.0 Adjustment of the individual
detector signal output level.
Values below 1 amplify the signal
and value over 1 diminish it
7DUJHW7LPHVSURJUDPPLQJ7DUJHW7LPHVXEVFUHHQ3U'
3DUDPHWHU 5DQJH 1RWHV
7DUJHW7LPH7 Settings for Target Time 1 curve
%XFN\ Bucky settings specified
0R Molybdenum filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
5K Rhodium filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
0DJQLILFDWLRQ Magnification settings specified
0R Molybdenum filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
5K Rhodium filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
&RQWDFW Contact settings specified
0R Molybdenum filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
5K Rhodium filter settings specified
Thickness 1 2.0 cm Thickness at curve point 1
Target 1 0.2 - 1.0 sec Target time at curve point 1
Thickness 2 6.0 cm Thickness at curve point 2
Target 2 1.0 - 3.0 sec Target time at curve point 2
,QLWLDON9WDEOHV3U'
3DUDPHWHU 5DQJH 1RWHV
N9YDOXHVIRU%XFN\
Thickness range 23-25 kV Scroll through thicknesses and
(at 0.5 cm intervals) change initial kV values for each
thickness.
N9 YDOXHV IRU
0DJQLILFDWLRQ
Thickness range 23-25 kV Scroll through thicknesses and
(at 0.5 cm intervals) change initial kV values for each
thickness.
N9YDOXHVIRU&RQWDFW
Thickness range 23-25 kV Scroll through thicknesses and
(at 0.5 cm intervals) change initial kV values for each
thickness.
'HWHFWRU$GMXVWPHQW3U'
3DUDPHWHU 5DQJH 1RWHV
1HZ JDLQV EDVHG RQ No/Yes Adjusts the gains for individual
UHFHQWH[SRVXUH detectors in relation to the C1
detector being 1.0.
([LWLQJWKH$(&VFUHHQ
If you accept the changes, press set -key. If you do not want changes to take
place, press cancel-key . Diamond returns to *52836 screen.
(&DUP5RWDWLRQ
The c-arm rotation default parameters can be set and changed in &$50
527$7,21 screen:
$QJOH&DOLEUDWLRQ3U(
3DUDPHWHU 1RWHV
FDOLEUDWLRQ Calibration of the c-arm angle
1) Drive the C-arm manually to 180°-angle (real angles) with manual rotation
keys
2) Press -key
If the calibration is successfully performed, the LCD screen then indicates '3DVV
In case the calibration was not successful, the LCD screen then indicates )DLO
2) If the calibration still was not successful, drive the c-arm to from 180° to
0° and then to 180° on the other side. When c-arm passes the 0° it helps
to calibrate.
3URMHFWLRQV3U(
3DUDPHWHU 5DQJH 1RWHV
2EOLTXH 135°, 225° Angle setting for the left OBL
(software button. If set at 135°, it operates
degrees, at 0° as in image A. If set at 225°, it
the tube is operates as in image B. The right
upside down) OBL drives to mirror angle.
/DWHUDO 0 - 360° Angle setting for the left LAT
(software button. Set at 90°. The right LAT
degrees) drives to mirror angle.
0LUURUDQJOH On/Off The C-arm runs to an equivalent
angle on the opposite side of the
unit.
$XWR'ULYH3U(
3DUDPHWHU 5DQJH 1RWHV
$XWRPDWLFGULYH On/Off If selected, CC, OBL or LAT-keys
and the c-arm will move to the
required angle. If not selected, the
key must be pressed on until the
required angle is reached.
([LWLQJWKH&$50527$7,21VFUHHQ
If you accept the changes, press set -key. If you do not want changes to take
place, press cancel-key . Diamond returns to *52836 screen.
)&ROOLPDWRU
The collimator default parameters can be set and changed in &2//,0$725
screen:
5HDU3ODWHV3U)
3DUDPHWHU 5DQJH 1RWHV
Front plate 12345 The number of steps that the
motor moves the plate. A 200-
step change equals to about 1mm
change on the film. Adjust using
1-step interval.
Bucky 18x24 12345 As above.
Bucky 24x30 12345 As above.
Contact 18x24 12345 As above.
Contact 24x30 12345 As above.
Magnification 1.6 12345 As above.
Magnification 1.8 12345 As above.
Magnification 2.0 12345 As above.
Spot Magnification 1.6 12345 As above.
Stereo digital 12345 As above.
Stereo film 12345 As above.
reserved 12345 Additional 4 plates may be added.
6LGH3ODWHV3U)
3DUDPHWHU 5DQJH 1RWHV
Front plate 12345 The number of steps that the
motor moves the plate. A 200-
step change equals to about 1mm
change on the film. Adjust using
1-step interval. If the value is too
big, motor back up time is
exceeded and an error message is
displayed.
Bucky 18x24 12345 As above.
Bucky 24x30 12345 As above.
Contact 18x24 12345 As above.
Contact 24x30 12345 As above.
Magnification 1.6 12345 As above. These values should be
Magnification 1.8 12345 the same in order for collimator
Magnification 2.0 12345 not to start driving to home
position.
Spot Magnification 1.6 12345 As above.
Stereo digital 12345 As above.
Stereo film 12345 As above.
reserved 12345 Additional 4 plates may be added.
&ROOLPDWRU/LJKW3U)
3DUDPHWHU 5DQJH 1RWHV
Intensity 1-5 Adjustment of the light intensity
Time normal 20 - 60 sec Time that the lamp is lighted up
during a normal procedure.
Time Stereo/Biopsy 20 - 60 sec Time that the lamp is lighted up
during a Stereo/Biopsy procedure.
([LWLQJWKH&$50527$7,21VFUHHQ
If you accept the changes, press set -key. If you do not want changes to take
place, press cancel-key . Diamond returns to *52836 screen.
*&RPSUHVVLRQ
The compression default parameters can be set and changed in &2035(66,21
screen:
&RPSUHVVLRQ6HWXS3U*
3DUDPHWHU 5DQJH 1RWHV
Slow speed 1-7 Slow compression arm speed. Use
no higher than 5. If the speed
exceeds 5, the synchronization
with ECS movement may be
disturbed.
High speed 2-7 High compression arm speed.
Release speed 1-7 The speed at which Diamond
release the compression arm in
the event of an emergency or a
power loss.
AutoRelease travel 0 - 20 cm The distance the Auto Release
function moves the compression
moves after an exposure.
Speed Threshold 0 - 150N The compression force at which
the compression arm starts to run
at slow speed.
3DGGOH&DOLEUDWLRQ3U*
3DUDPHWHU 5DQJH 1RWHV
Base point -9.5 … +9.5 Base point adjustment that affects
all paddles.
Standard small -9.5 … +9.5 Compensation for paddle
flexibility. Check for country-
specific requirements.
Standard large -9.5 … +9.5 As above.
Magnification spot -9.5 … +9.5 As above.
Standard magnification -9.5 … +9.5 As above.
Implant -9.5 … +9.5 As above.
Magnification implant -9.5 … +9.5 As above.
Delta 32 -9.5 … +9.5 As above.
Spot -9.5 … +9.5 As above.
Mag spot enhanced -9.5 … +9.5 As above.
Biopsy 1 -9.5 … +9.5 As above.
Biopsy 2 -9.5 … +9.5 As above.
Spot enhanced -9.5 … +9.5 As above.
reserved -9.5 … +9.5 As above.
&DVVHWWH+ROGHU&DOLEUDWLRQ3U*
3DUDPHWHU 5DQJH 1RWHV
No holder -9.5 … +9.5 Base point adjustment that affects
all holders.
Bucky 18x24 -9.5 … +9.5 Adjustment for an individual
cassette holder.
Bucky 24x30 -9.5 … +9.5 As above.
Magnification 1.6 -9.5 … +9.5 As above.
Unknown -9.5 … +9.5 As above.
Contact 18x24 -9.5 … +9.5 As above.
Contact 24x30 -9.5 … +9.5 As above.
Mag 1.8 -9.5 … +9.5 As above.
Spot Magnification 1.6 -9.5 … +9.5 As above.
Stereo digital -9.5 … +9.5 As above.
Stereo film -9.5 … +9.5 As above.
Magnification 2.0 -9.5 … +9.5 As above.
reserved -9.5 … +9.5 As above.
&RPSUHVVLRQ)RUFH&DOLEUDWLRQ3U*
3DUDPHWHU 5DQJH 1RWHV
Force to 0 N 0N Apply 0N force and accept.
Force to 200 N 200 N Apply 200 N with a calibrated
device and accept if the screen
displays correct figure.
&RPSUHVVLRQ)RUFH/LPLWV3U*
3DUDPHWHU 5DQJH 1RWHV
Mode 0/1 0=Force limits are not applied.
1=Force limits applied.
Position 1 0 - 255 N Compression force stops when
this limit is reached.
Position 2 0 - 255 N As above.
Position 3 0 - 255 N As above.
Position 4 0 - 255 N As above.
Position 5 0 - 255 N As above.
Position 6 0 - 255 N As above.
1HJDWLYH&RPSUHVVLRQ)RUFH&DOLEUDWLRQ3U*
3DUDPHWHU 5DQJH 1RWHV
Apply weight 0N Apply 0N force and accept.
Adjust force 0 - 10 N Adjust for negative compression
force required for the stop
reaction.
([LWLQJWKH&2035(66,21VFUHHQ
If you accept the changes, press set -key. If you do not want changes to take
place, press cancel-key . Diamond returns to *52836 screen.
+)LOWHU
The filter default parameters can be set and changed in ),/7(5 screen:
$XWR)LOWHU6HWXS3U+
3DUDPHWHU 5DQJH 1RWHV
Change thickness 0 - 15 cm Above the range, the Rhodium is
used. Below the range, the
Molybdenum filter is used.
)LOWHUFDOLEUDWLRQ3U+
3DUDPHWHU 5DQJH 1RWHV
Molybdenum 12345 The number of steps (from home
position) applied for this filter
selection. A four-step change
moves the filter about 1,3 mm.
Adjust using 4-step intervals
(4,8,12…). Calibrate the filter
from the front edge of the
collimator using small focus
(magnification tunnel).
Rhodium 12345 As above.
Aluminum 12345 As above.
([LWLQJWKH),/7(5VFUHHQ
If you accept the changes, press set -key. If you do not want changes to take
place, press cancel-key . Diamond returns to *52836 screen.
,*HQHUDWRU
The generator default parameters can be set and changed in *(1(5$725
screen:
*HQHUDWRU6HWXS3U,1RWLQXVHDWWKHPRPHQW
3DUDPHWHU 5DQJH 1RWHV
Anode Brake On/Off Anode brake operation regulation.
3UH+HDWDGMXVWPHQWODUJHIRFXV /) 3U,1RWLQXVHDW
WKHPRPHQW
1) Press Exposure button to start automatic PreHeat calibration for large focus.
2) If the new PreHeat value is valid and the LCD screen displays 3DVV ,
Diamond returns to the *(1(5$725 screen.
3UH+HDWDGMXVWPHQWVPDOOIRFXV 6) 3U,1RWLQXVHDW
WKHPRPHQW
1) Press Exposure button to start automatic PreHeat calibration for large focus.
2) If the new PreHeat value is valid and the LCD screen displays 3DVV ,
Diamond returns to the *(1(5$725 screen.
3UH+HDW&RPSHQVDWLRQODUJHIRFXV /) 3U,
3DUDPHWHU 5DQJH 1RWHV
PreHeat LF 0.5 - 1.50 Large focus preheat adjustment
that affects all kVs.
LF 15 - 39 0 - 255 Large focus preheat adjustment
for each kV. Typical voltage steps
are 54-56.
3UH+HDW&RPSHQVDWLRQVPDOOIRFXV /) 3U,
3DUDPHWHU 5DQJH 1RWHV
PreHeat SF 0.5 - 1.50 Small focus preheat adjustment
that affects all kVs.
SF 15 - 39 0 - 255 Small focus preheat adjustment
for each kV.
([LWLQJWKH*(1(5$725VFUHHQ
If you accept the changes, press set -key. If you do not want changes to take
place, press cancel-key . Diamond returns to *52836 screen.
-7XEH3DUN
The tube park default parameters can be set and changed in 78%(3$5. screen:
7XEH3DUN6HWXS3U-1RWLQXVHDWWKHPRPHQW
3DUDPHWHU 5DQJH 1RWHV
Slow speed 1-5 Slow speed adjustment for the
moving tube head.
Fast speed 1-5 Fast speed adjustment for the
moving tube head.
([LWLQJWKH78%(3$5.VFUHHQ
If you accept the changes, press set -key. If you do not want changes to take
place, press cancel-key . Diamond returns to *52836 screen.
.6WHUHR0RYHPHQW
The stereo movement default parameters can be set and changed in 67(5(2
029(0(17 screen:
6WHUHR0RYHPHQW6HWXS3U.
3DUDPHWHU 5DQJH 1RWHV
Low speed limit 1-4 Low speed adjustment.
High speed 4-7 High speed adjustment.
Speed offset -2 … +2 Power boost for c-arm. Do not
change before consulting with the
manufacturer.
Breaking speed 2-4 Slow down speed when required
is being approached. Do not
change before consulting with the
manufacturer.
Ramp in 10ms 5 - 50 Acceleration of ramp speed. Do
not change before consulting with
the manufacturer.
6WHUHR0RYHPHQW6HWXS3U.
3DUDPHWHU 5DQJH 1RWHV
Auto drive On / Off If the function is selected (On),
one press is enough to drive to the
stereo angle. If it is not selected
(Off), the stereo drive buttons
have to be pressed down all the
time until the stereo angle is
reached.
([LWLQJWKH67(5(2029(0(17VFUHHQ
If you accept the changes, press set -key. If you do not want changes to take
place, press cancel-key . Diamond returns to *52836 screen.
([LWLQJWKHVHUYLFHPRGH
You can exit the service mode by pressing kV- and Den- buttons
down simultaneously.
/2$',1*1(:62)7:$5(
Diamond software consists of Flash Memory and Battery RAM. The software
itself is stored in the Flash Memory while the settings are stored in the Battery
RAM.
:$51,1* %HIRUH\RXVWDUWXSWKH)ODVK/RDGHUSURJUDPFORVHDOORWKHUSURJUDPV
HVSHFLDOO\ WKH RQHV XVLQJ &20SRUWV 6LPXOWDQHRXVO\ UXQ SURJUDPV PD\
VHYHUHO\GLVWXUEWKHVRIWZDUHORDGLQJSURFHVV
4) Select the directory where the .bin-files are stored. This is performed by
writing the directory name on the %LQ ILOH SDWK-field or clicking the
browse -button and searching for the correct directory on hard disk.
When you find the correct directory, click any of the .bin-files listed.
Diamond will then set up default path. If the directory has already been
defined, you can proceed to next step.
&RQQHFWLQJ'LDPRQGWRWKHODSWRS
After the connection has been successfully established, the screen displays
a green light above the connect-button.
:$51,1* 'XULQJWKHVRIWZDUHXSORDGSURFHVVGRQRWH[LWWKHSURFHVVRUVZLWFKRIIWKH
XQLW:KHQWKHVRIWZDUHILOHVKDYHEHHQHUDVHGUHVWDUWLQJWKHXQLWZLOOEH
FRPSOLFDWHG,QFDVHEODFNRXWVRFFXULQWKHHQYLURQPHQWZKHUHWKHXQLWDKV
EHHQSODFHGLWLVUHFRPPHQGWRKDYHDQ836RUVLPLODUHOHFWULFDOEDFNXS
V\VWHPZKLOHXSORDGLQJWKHVRIWZDUH
:$51,1* 7KHGHOHWHGVRIWZDUHILOHVFDQQRWEHUHVWRUHG%HIRUH\RXVWDUWORDGLQJDQHZ
VRIWZDUH YHUVLRQ ZULWH GRZQ DOO LPSRUWDQW VHUYLFH PHQX VHWWLQJV 7KHQ
FRQWLQXHWKHORDGLQJSURFHVVWRWKHHQG
At this point, the system has no operational software. Do not switch off
the power. Proceed to next step to install required software files.
In case you get an error message "Error opening file", check that the file path is
correct.
When the software has been loaded, the connection light turns red and the
following information is displayed:
127( ,QFDVHWKHHOHFWULFLW\ZHQWRIIGXULQJWKHVRIWZDUHXSORDGRUIRUVRPHRWKHU
UHDVRQWKHXQLWZLOOQRWVWDUWFDOOPDQXIDFWXUHUIRUIXUWKHULQVWUXFWLRQV
+<3(57(50,1$/,17(5)$&()25',$021'
The system settings can be exported in hexadecimal format and can be viewed in
Windows HyperTerminal.
For this you need a PC, hyperterminal software such Windows HyperTerminal®
by Hilgraeve Inc., serial cable (9 pin male and female; lines 2,3 and 7 directly
connected, cable code 85102).
&RQILJXUDWLRQRIWKH+\SHU7HUPLQDO
1) Select COM-port that you want to use at FILE\PROPERTIES\CONNECT
TO-tab sheet:
&DSWXULQJWKHGDWD
1) Connect the cable between Diamond COM-port and PC serial port. Switch
off Diamond before this.
7) If the file in which you want to store data is correct, click on START. If
not, look for the file with Browse-function.
11) Terminate the HyperTerminal session and disconnect the PC cable from
Diamond.
127( ,IWKHFRQQHFWLRQWR+\SHU7HUPLQDOKDVQRWEHHQWHUPLQDWHGZKHQ'LDPRQG
LVUHVWDUWHG'LDPRQGZLOOWKHQHVWDEOLVKWKHFRQQHFWLRQ
6DYLQJWKHVHWWLQJVGDWD
1) Start capture.
In case restoration of some old system parameters is necessary, send the printout
or a text file to the manufacturer. The data is examined and the instructions will
be sent to you.
',$021'7(67',63/$<6
In service mode, Diamond Control Panel displays various figures that convey
information about the system condition. The printout of the screen items is
included in appendices.
3) Press button 2 (in the middle, below LCD-screen). Screen displays various
test figures.
4) Check out listing of the page number in upper left corner of the LCD-
screen.
6) After you have finished with test displays, exit the screen with -key and
exit the service mode.
-7$*/2$'(5
If flash memory has for some reason become messed up, then JTAG loader can
be used to erase it and prepare it to receive software download.
4) Plug in JTAG cable (code 85105) X15 on Supervisor Board and to LPT1
on laptop/PC.
7) Click on JTAG.
8) Click on Connect.
10) Select path where JTAG files are located. This is performed by writing the
directory name on the %LQILOHSDWK-field or clicking the browse -
button and searching for the correct directory on hard disk.
13) When the loading is complete, deselect JTAG. The program returns to
Normal Programming state.
19) If the software download was successful, attach side cover back on.
3UREOHPZLWK-7$*"
(/(&75,&$/6<67(0$5&+,7(&785(7+(25<2)
23(5$7,21
29(59,(:
This section provides theory of operation from the system level to detailed circuit
level.
The electronics contains both low voltage – low energy circuits and power
electronics circuits which are directly connected to the mains line voltage.
The voltages inside the system are hazardous and therefore great caution should
be exercised when measuring electric signals especially with electronic boards
that contain line referenced circuits.
For the system architecture, refer to the documents in Chapter 11, Appendices.
1. Inverter board
2. Filament Control board
3. High voltage unit
4. Anode motor drive (AMD) board
5. Mains transformer
6. Energy storage capacitor
7. Tube housing
7HFKQLFDOVSHFLILFDWLRQVDUH
,138732:(5
The mains line is applied to the Power Board via mains fuses, line filter and mains
switch. The AC mains voltages are further applied to Generator Board, Stereo
Board, AMD Board and to main transformer that provides all the low voltage
(ground referenced) power.
6XSSO\9ROWDJH5HJXODWLRQ
Supply voltages from the main transformer are rectified at Power Board. The
output voltages are ±24Vac and +34Vac. The voltage rectifiers are located as
follows:
The communication takes place with two serial (RS-485) links and parallel signals
that provide link between the Supervisor Board and other PCBs. The first serial
bus is an external bus and it connects the Supervisor Board to Control Panel(s)
and other external control devices. The second serial bus is internal and it
connects the Supervisor to all PCBs inside the C-arm of the system. The parallel
ports connect to the Generator Board and Power Board that are located in the
carriage.
,QWHUQDO6HULDO%XV
The internal serial bus connects the Supervisor board to PCBs located in the C-
arm. The bus is separated into two segments: the first one connects the AEC
Board and Stereo Board. The other segment connects through Stereo Board and
it connects Compression Board and Collimator Board.
Each of these above-mentioned boards has its own microcontroller that controls
the input/output on that particular board and boards that connect to them (AMD
Board, Drive Board). Each board has its own address and the Supervisor Board
reads and writes alternately to each board every 20ms. If a board does not receive
data for a certain period of time it will reset all its outputs to a non-active state.
The serial bus has two lines: receiving (RXD) and sending line (TXD-from
Supervisors viewpoint). The Supervisor controls the sending line and all other
boards observe that line all the time. The receiving line is occupied for a certain
time by each board alternately after the Supervisor has addressed the board.
The serial cable is a shielded cable and the shield should always be grounded at
both ends of the cable.
([WHUQDO6HULDO%XV
The external serial bus is similar to the internal bus except that it is intended for
one or two control panels and other control devices that are located out side the
unit.
6XSHUYLVRU%RDUG
The Supervisor Board is the center of the system. It is built around Intel’s 386EX
microprocessor. The Supervisor Board controls all the other boards as discussed
in Chapter 4: System Architecture. The software can be programmed into the
memory via JTAG link or COM port. For further details, refer to Chapter 3.3,
Loading new software.
6FKHPDWLFV/D\RXWV
3RZHU%RDUG
The Power Board provides supply voltages and main generator power as well as
controls C-arm rotation and Vertical drive motors.
3RZHUVXSSO\VHFWLRQ
Power supply section of the Power Board produces supply voltages +5 V, ±15 V,
±24 V, 34 V and 310 V. Voltages are rectified from output of main transformer
TF1 and direct from line voltage.
Voltage level of the main supply can be read by the supervisor from signal
CFORCE provided by the power board. Voltage level is measured and calculated
based on a low-level voltage +24Vdc. CFORCE is a dual-purpose signal and is
enabled for measuring the voltage level from +24Vdc while the C-arm is not run.
If the main supply have been interrupted for about 10 to 15 milliseconds, signal
POWERFAIL is passed on to the supervisor. POWER-ON LAMP is switched on
by relay K6 when 34 Vdc is available.
The full wave rectified 310 Vdc is applied to the Generator through relays K6 and
K7. The Supervisor board controls relays with signals RG1 and RG2. When an
exposure is initiated, K7 applies voltage through current limiting resistors R153
and R154 that allow controlled charging of the energy storage capacitor. K6 is
activated by RG2 about one second later and provides full power to the generator
board.
([WHUQDOLQGLFDWRUV
Connector X16 provides relay contacts for X-ray system status indication. X16
pins 1 and 2 are short-circuited when the unit is on. X12 pins 1 and 2 are short-
circuited when X-rays are being generated. The contacts are rated for 5A, 24VDC
/240VAC.
&DUPURWDWLRQPRWRUFRQWURO
The C-arm motor control is an 8-speed supervisor programmed driver for 24 Vdc
motor that rotates the C-arm. Speed of the movement is stabilized with speed
feedback produced by a tachometer on the C-arm motor. The tachometer provides
a pulsed signal C-PULSE that is transformed to voltage speed feedback by
frequency to voltage converter ICA7.
Direction is controlled with signal C-DIR-LEFT and speed with signals C-MUX1,
C-MUX2 and C-MUX3. Four C-position optoswitches provide signals C-RIGHT-
POS, C-RMIDDLE-POS, C-LMIDDLE-POS and C-LEFT-POS. If signal C-
RIGHT-POS or C-LEFT-POS is active, then C-arm movement to the right or left
is restricted respectively. Signal C-RUN is required to run C-motor. C-motor
current is monitored and restricted to app. 15.7 Amperes.
9HUWLFDOPRYHPHQWRU=PRWRUFRQWURO
Z-motor drives vertical movement of the device. Signal Z-DIR-UP defines
direction of the movement. If signal Z-UPPER-LIMIT or Z-LOWER-LIMIT is
active, then movements up or down are not allowed, respectively. Two
microswitches provide Z-limit signals. Signal Z-RUN is required to run Z-motor.
Z-motor current is monitored and limited to app. 4 Amperes. Z-motor is driven
with two speeds. Once the drive is enabled it is started at a low speed and after a
little delay the speed is increased.
The drive of the motor is possible only if motor-stop buttons are installed and no
button is activated (pressed down).
)XVHV
6FKHPDWLFV/D\RXWV
*(1(5$725%2$5'
Generator Board regulates the kV and mA of the X-ray tube. The Inverter Board
contains a high voltage and a low voltage section. The high voltage section is
isolated from the low voltage section with optocouplers ICD14-ICD17 and current
transformer TF2.
:$51,1* 9ROWDJHVLQWKHKLJKYROWDJHSDUWRIWKH*(1(5$725%RDUGDUHKD]DUGRXV
7KHSHDNWRSHDNYROWDJHH[FHHGV9LQQRUPDORSHUDWLRQ
.9&RQWURO
The MOSFET driving pulses are generated at the low voltage section of the board
and are provided by optocouplers ICD14 - ICD17.
TF2 is a current transformer, which monitors the bridge current. If the bridge
current exceeds the preadjusted current limit the control circuit in the low voltage
section turns the MOSFETs off.
4.5.1.2 Timing
The input voltage is applied through the connector X6. When an exposure is
initiated, relays K1 and K2 on the Power Board provide the line voltage. The
activation of the EXPENA signal (H8) enables the H-bridge drive circuitry.
(;3(1$VLJQDO
An exposure is initiated by the Supervisor Board by activation of the EXPENA
signal. This enables pulsing of the MOSFET drive optocouplers and allows the
kV reference signal to rise providing "soft start".
N9UHIHUHQFH
The level of the used kV is set by the SUPERVISOR BOARD with signals
KVMAREF0 -KVMAREF7. This 8-bit digital word contains the kV reference.
The kV reference is converted into an analog signal in D/A converter ICA6
(TP26).
The analog kV reference (TP26) and the kV feedback (TP20) signals correspond
to actual kV as follows:
kV = DVM read-out * 10 kV
For example, if the digital multimeter (DVM) read-out is 2.5 V, the actual high
voltage is 25kV.
N9DGMXVWPHQW
127( 7KHN9FDOLEUDWLRQLVSHUIRUPHGDWWKHIDFWRU\7KHWULPPHUSRWHQWLRPHWHU
5LVVHDOHG0DQLSXODWLRQRIWKHSRWHQWLRPHWHUFDQEHFDUULHGRXWRQO\E\
,QVWUXPHQWDULXPDXWKRUL]HGVHUYLFHWHFKQLFLDQ
0$&RQWURO
The mA control adjusts the X-ray tube’s filament heating level to provide the
desired tube current (mA). This is done by regulating the filament voltage with
"FLYBACK" type pulse width modulated switch mode power supply (T1, T2,
TF1, D14, C85, and ICA5). The control circuitry uses current mode control, which
provides better stabilization and protects the semiconductor switches against over-
current. The operation consists of two consecutive sequences: 1) the preheat
sequence, and 2) normal mA control sequence. The preheat sequence is used for
heating of the filament to the desired level prior to the exposure, in order to
achieve optimum mA rise at the start up.
During the preheat sequence the control feedback comes from the filament
voltage, and during an exposure the mAFB signal is used, providing the actual
tube current monitoring.
P$JHQHUDWLRQWLPLQJ
*(1(1$ becomes active 20-40 ms after the hand switch is depressed. The relay
K1 is powered up by GENENA and applies power to the filament transformer
TR1. During the time between PREHEAT and EXPENA, the pulse width
modulator (ICA5) uses filament voltage as its feedback voltage. The reference is
software controlled to achieve proper tube current rise at the beginning of an
exposure.
When EXPENA goes active, actual exposure begins and regulation of the tube
current becomes a more important issue than filament voltage. This is performed
by connecting the common of the filaments to ground through one of the mA
feedback resistors (R7 or R8+R9). This signal, called P$)(('%$&. 0$)%
is now fed to the pulse width modulator by an analog switch (ICA10). Its output
was transferred from filament voltage to MAFB with the signal 35(0$&75/.
The PRE/MACTRL signal is activated approximately 7 ms after the EXPENA
signal. The duty cycle of the pulse width will vary depending on what is required
to maintain proper mA.
3UHKUHIHUHQFH
The preheat reference is the reference for filament voltage. It is converted from
8 (KVMAREF0-7) parallel databits into preheat reference information in two-
channel D/A converter (ICA4).
P$UHIHUHQFH
The mA reference is the reference for the actual tube current. It is converted from
8 (KVMAREF0-7) parallel databits into mA reference information in two-channel
D/A converter (ICA4). The selection between preheat and mA channel is made
with signal PRE/MASEL (0 – PREHEAT; 1 – mA).
3UHKHDWDGMXVWPHQW
The preheat adjustment is performed separately for large and small focus by
varying the preheat reference with software.
3URWHFWLRQ&LUFXLWV
The generator board features several circuits to protect the generator or the X-ray
tube.
.92.DQG78%()$,/VLJQDOV
KVOK and TUBEFAIL signals are activated if the KVFB signal does not reach
the kV reference. This typically occurs during a high voltage breakdown or
arching.
As the arching is characteristic to x-ray tubes, and hence may occur during normal
operation, the exposure is continued with a new activation of the EXPENA signal.
The Supervisor Board allows three (3) tube failures per exposure before an error
message is displayed.
0$2.DQG),/$B(5525VLJQDOV
MAOK signal is zeroed if the mA feedback does not reach the mA reference. This
information goes to Supervisor Board and enables software controlled protection.
)%0,66VLJQDO
The FBMISS signal is activated if the feedback cable is not connected or the
connector is not properly seated. It disables the +34V power to the filament as
well as shuts down the filament PWM regulator (ICA5).
%DFN8S7LPHU
The generator has a Back Up Timer (ICA10c) that disables both the +34V for the
filament and the EXPENA signal providing SHUTDOWN approximately fifteen
(15) seconds after the activation of the EXPENA signal. This is only a safety
feature and does not affect normal operation.
*HQHUDWRUERDUGMXPSHUV
&$87,21 7KH7HVW-XPSHU-VKRXOGDOZD\VEHDW1250$/SRVLWLRQZKHQHYHUWKH
KLJKYROWDJHSDUWRIWKH*(1(5$725%2$5'LVHQHUJL]HG)DLOLQJWRGR
WKLVPD\GDPDJHWKH*(1(5$725%2$5'+,*+92/7$*(81,7DQG
WKH;5$<78%(
0RQLWRULQJ
The Supervisor Board monitors the generator performance and measures certain
signals in order to perform calibrations and adjustments.
$QH[SRVXUHZLWKRXW;UD\V
The kV control circuitry is driving the MOSFETS with the minimum frequency
(app. 20kHz) and the filament voltage is regulated to preheat level.
6FKHPDWLFV/D\RXWV
$12'(02725'5,9(%2$5'$0'
The unit features a rotating anode tube. The anode plate is accelerated with 230
VAC before the emission of electrons from cathode. During exposure, stator
needs a certain running voltage (40-100 VAC), and after exposure the anode plate
can be held steady with a brake stop.
All this is performed with the phase control of a triac (V1). The control element
is Telefunken TEA1007 (A7). Acceleration is achieved directly with mains
voltage. During exposure, the firing angle of the triac is set to a level that
corresponds to the required RMS voltage. Brake stop is optional (software) and
is done by half-wave rectifying the mains voltage. Both acceleration and brake
procedures take approximately 1 second. Safety issues are also taken into account
by monitoring the current of the triac. Should errors be detected, the
microprocessor will not allow exposures. The exchange of information between
the microprocessor and this board is optoisolated because the anode motor drive
board is floating on mains voltage.
6FKHPDWLFV/D\RXWV
67(5(2%2$5'
Stereo Board is the central board in the C-arm. Its main tasks are the electronics
bus segment handling, sensor and keyboard signal monitoring, power supply
regulation and motor control.
The Stereo Board handles the internal electronics serial bus communication so
that the two bus segments can communicate. Each time a board is sending data to
Supervisor Board (that is, writing to the bus) it must take over the TXD line. The
segment that contains Collimator and Compression boards connects to the main
segment via Stereo board and therefore Stereo board must take over the TXD line
of the main segment each time someone from the other segment sends data to
Supervisor.
Sensor and keyboard signals on the C-arm (except tube head) are collected in the
Stereo Board.
The ±15Vdc and +5Vdc supply voltages for all the boards in the C-arm are
regulated on the Stereo board. Refer to Chapter 4.2.1, Supply voltage regulation.
Stereo board also drives the compression motor, ECS motor and Stereo motor and
controls cassette and grid motors on the Drive board.
9ROWDJH5HJXODWRUV
The Stereo board contains +5V switch mode power supply (SMPS) that supplies
+5Vdc voltage to the Stereo, Drive, AEC, Collimator and Compression boards.
The +15V linear regulator supplies +15Vdc to the Stereo, Drive, AEC, Collimator
and Compression boards.
The –15V linear regulator is used as preregulator by AEC board’s –12V regulator.
6WHUHRDQG(&6PRWRU'ULYH
Both motors are +24Vdc-motors, which are driven with pulse width modulated
(PWM) power supplies that can provide variable voltage control, thus enabling
speed control. Each drive circuit consists of control circuit and two power
transistors. The output voltage consists of +34 Volt pulses whose width
determines the RMS voltage of the motor. This voltage is proportional to the
reference voltage applied to the control circuitry.
Stereo and ECS motors are driven by the same control and drive circuit (ICA1,
T1, T2). Both drives are speed controlled. The speed feedback for stereo drive
comes from stereo motor tachometer and the speed reference for ECS drive comes
from compression motor tachometer. The active motor is selected with relay K3
and the direction is selected with K4.
The operating speed is software controlled and multiplexed from 7 resistors with
ICD18. ECS speed is either determined by the compression motor (ECS mode)
or fixed speed (C-arm mode).
Both drives also utilize a current limiting circuit that limits the motor current to
approximately 5A.
&RPSUHVVLRQPRWRU'ULYH
The compression motor drive is a software controlled 8 speed PWM drive. The
speed is selected by choosing one reference voltage from multiplexer ICD9.
%DFNXSUHOHDVH
The unit is equipped with a compression release back up feature. If the line
voltage suddenly disappears, then the compression is released. This is
implemented with relay K2. When this relay is powered up, the compression
motor is operated normally. When the relay is not activated, the motor is driven
upwards with capacitor charge in the +34Vdc power supply line.
6HQVRUVDQG.H\ERDUG
Stereo board receives the signals from stereo angulation sensors (refer to Chapter
4.12, Sensors), ECS limit sensors (which connect directly to the board) and
cassette holder sensors, cassette movement and grid position sensors (which
connect via Drive Board).
6FKHPDWLFV/D\RXWV
Refer to the documents in Chapter 11, Appendices.
'5,9(%2$5'
The Drive Board contains drive and control circuits for grid and cassette
movement stepper motors.
*ULGPRYHPHQW
The grid is moved with a stepper motor that is controlled by ICD3. This
microcontroller contains software that controls the grid speed, which is a function
of the grid position.
Prior to an exposure, the Stereo Board activates the grid movement with signal
GRID_START. When the grid has reached the maximum speed the
microcontroller on Drive Board sends the XRAY signal that allows the Supervisor
Board to start the exposure.
&DVVHWWH7UD\PRYHPHQW
Cassette Tray is moved with a stepper motor that is controlled by ICA4. This
stepper motor controller requires enabling, direction and clock signals from Stereo
Board (CAS_ENA, CAS_DIR, CAS_CLOCK).
Each clock pulse corresponds to a certain movement in distance and the pulse
frequency determines the travel speed. The Stereo Board microcontroller provides
the necessary amount of clock pulses to drive the cassette tray. The actual cassette
tray position is sensed with cassette tray movement sensors.
6HQVRUV
Cassette holder sensors (Sensor Board, refer to Chapter 4.12, Sensors), cassette
tray movement and grid position sensors all connect to the Drive Board. The
signals are further taken to the Stereo Board, which sends them to the Supervisor
Board via the RS-485 serial link.
6FKHPDWLFV/D\RXWV
&2//,0$725%2$5'
The collimator board controls the integrated collimator plates, filter movement,
tube park movement and field light.
&ROOLPDWRU3ODWHDQG)LOWHU'ULYH
The two integrated collimator plates and the filter mechanism are each driven by
a stepper motor. Each motor uses the same control and drive circuit
7XEH3DUN'ULYH
The tube drive circuit consists of pulse width modulated (PWM) IC ICA5, which
run at approximately 30kHz. The drive has 8 different software controlled
reference voltages that correspond to 8 different tube park speeds respectively.
6HQVRUVDQG.H\ERDUG
Collimator board receives the signals from tube park movement sensors, tube
movement safety clutch sensor, collimator plate and filter home sensors, external
collimation plate sensor board and face shield sensor. Refer to Chapter 4.12,
Sensors.
6FKHPDWLFV/D\RXWV
&2035(66,21%2$5'
The Compression board controls the compression thickness and force
measurement and display. It also controls the displays and keyboard on
compression arm control panel. The microcontroller ICD7 controls the board
operations and sends the data to the Supervisor Board for further processing.
&RPSUHVVLRQ7KLFNQHVVPHDVXUHPHQW
Compression thickness is measured with optocouplers that read the dotted tape.
The dotted tape provides information about distance, direction of movement and
compression and home position. This information is then transferred to processor
87C51. The calibration is performed with the software.
&RPSUHVVLRQ)RUFHPHDVXUHPHQW
Compression force is measured with strain gauges in the compression mechanism.
The signal is brought to the amplifier ICA1. The offset voltage of the amplifier
must be set to 450mV/5mV when there is no strain on the compression paddle.
The voltage is measured between test points TP1 and TP2.
'LVSOD\V
Compression force, compression thickness and the c-arm angle is displayed by
Compression Display Board that contains a driver ICD8 (ICM7218BIQI) to
control the 7-segment display. The Display Board also contains Hall sensors to
recognize various compression paddles.
.H\ERDUG
Buttons for AEC screen, C-arm and Z-movement are connected to 3x8 matrix.
The matrix number is regulated with a micro-controller ICD7 of the Compression
Board.
3X8 MATRIX
SW-A SW-B SW-C
DETEC 0 AUTOPOINT PADDLE 1
DETEC 1 FILTER PADDLE 2
DETEC 2 COMPRESSION + PADDLE 4
DETEC 3 COMPRESSION - PADDLE 8
DETEC 4 FIELD LIGHT VERT-UP
DETEC 5 C-ARM-LEFT VERT-DOWN
DETEC 6 C-ARM-RIGHT N.C.
DETEC 7 N.C. N.C.
6FKHPDWLFV/D\RXWV
$(&%2$5'
The AEC Board measures the radiation that goes through the breast from eight
different ellipsoidal-shaped detectors. It also collects sensor data to Supervisor
Board.
$XWRPDWLF([SRVXUH&RQWURO
Each detector consists of several photodiodes that provide current that is
proportional to the radiation. This current is converted into voltage for each
separate detector. The signals from eight detectors are then multiplexed in ICA5
to one variable gain and variable offset amplifier and taken further to 12-bit
analog to digital converter ICD5. The data from the detectors is then taken to the
microcontroller, which sends the it to the Supervisor Board for further processing.
The second amplifier stage, which is common to all eight detectors, features
software controlled offset circuit, which provides correct offsets for each detector.
Also the gain can be selected to provide optimum point of operation for each
exposure. The principal AEC gain is accomplished with software.
6HQVRUV
The AEC Board receives signals from cassette sensor and cassette lock sensors.
Refer to Chapter 4.12, Sensors.
6FKHPDWLFV/D\RXWV
6(16256
&DVVHWWH+ROGHU6HQVRU
&DVVHWWH+ROGHUGHWHFWLRQ
The Sensor Board consists of four Hall-sensors (image A) that decode the holders
according to table below. Each cassette holder is equipped with one to four
magnets (image B) that encode the cassette holder discussed before.
The signals from the Sensor Board are transferred to Drive board and further to
Stereo Board, which sends the data to the Supervisor Board via RS-485 serial link.
'LDPRQGVHQVRUV &RUUHVSRQGLQJFDVVHWWHKROGHU
6 6 6 6
0 0 0 1 Bucky 18x24
0 0 1 0 Bucky 24x30
0 0 1 1 Mag 1.6
0 1 1 1 Mag 1.8
1 0 1 1 Mag 2.0
1 0 0 1 Digital Stereo - Delta 32
0 1 0 1 Contact 18x24
0 1 1 0 Contact 24x30
1 0 0 0 Spot Mag 1.6
&DVVHWWH6HQVRUV
&DVVHWWH'HWHFWLRQ
The photo sensor connects to the AEC board 85001 and is then taken to the
Supervisor Board via RS 485 serial link by AEC board microcontroller.
&DVVHWWH/RFNV
The cassette locks are used for activation of the cassette tray movement. The
position of the cassette lock is sensed with two Hall-sensors on the Cassette Lock
Sensor Board that detect the small magnet inserted in the cassette lock lever.
The Cassette Lock Sensor Board connects to the AEC board X1 and is then taken
to the Supervisor Board via RS 485 serial link by AEC board microcontroller.
&DVVHWWH7UD\PRYHPHQW
The Cassette Tray movement is detected with three slotted photosensors of which
two of them are connected in parallel providing one signal at both ends of the
movement area.
6HQVRUV &RUUHVSRQGLQJSRVLWLRQ
6 6 6
1 0 0 Home
0 0 0 In between
0 0 1 Large cassette
0 1 0 Small cassette
*ULG6HQVRU
The actual location of the grid is calculated from stepper pulses by the grid
microcontroller on the Drive Board.
+RPHSRVLWLRQ
The home position of the grid is sensed with a slotted photointerrupter that is
activated by a flag in the grid mechanism. The grid home signal is taken to both
Drive Board grid controller and the microcontroller on the Stereo Board, which
further sends the data to Supervisor Board via RS-485 serial link.
&ROOLPDWRU6HQVRUV
,QWHJUDWHG&ROOLPDWLRQ3ODWHV
The actual location of the collimator plates is calculated from stepper pulses by
the microcontroller on the Collimator Board. The position of the front lid is
derived from the drive order of the back and side plates.
+RPH3RVLWLRQ
The actual location of the collimator plate is sensed with a slotted photointerrupter
that is activated with a flag in the collimator plate. The home signal is taken to the
microcontroller on the Collimator Board, which further sends the data to
Supervisor Board via RS-485 serial link.
([WHUQDO&ROOLPDWLRQSODWHGHWHFWLRQ
)LOWHU6HQVRU
The actual location of the filter mechanism is calculated from stepper pulses by
the grid microcontroller on the Collimator Board.
+RPHSRVLWLRQ
The home position of the filter is sensed with a slotted photointerrupter that is
activated by a flag in the filter mechanism. The filter home signal is taken to the
microcontroller on the Collimator Board, which further sends the data to
Supervisor Board via RS-485 serial link.
&DUP5RWDWLRQ6HQVRUV
5RWDWLRQ/LPLWV
There are three limit sensors for the C-arm rotation. The center limit consists of
slotted photointerrupters, which are activated by a flag in the mechanism. The two
other photointerrupter limits prevent the mechanism from colliding to the
mechanical limit at maximum C-arm angles.
The signals from the limits are taken via the Power Board to the Supervisor
Board.
5RWDWLRQ$QJOH
The Rotation angle is calculated from the pulses of the Tacho Board, which is
attached to the C-arm motor. The software calibration is based on fixed rotation
limits that is the center position, where the zero reading is updated each time the
C-arm is at CC-position.
The tacho pulses are taken to the Supervisor Board via the Power Board.
When the power is turned on the current angle value is stored in Supervisor Board
memory where it is fetched from next time the unit is powered up.
5RWDWLRQ6SHHG
The Rotation Speed is measured from the pulses of the Tacho Board, which is
attached to the C-arm motor. The pulses are taken to the Power Board, which uses
a circuit that controls the speed according to a speed reference from Supervisor
Board.
&RPSUHVVLRQ6HQVRUV
&RPSUHVVLRQ7UDYHO/LPLWV
The compression travel limits are measured with the WZR VORWWHG
SKRWRLQWHUUXSWHUV from the bar code ribbon located in the compression
mechanism.
The compression features two limits. The upper is actual travel limit that prevents
collision. The other one is used for compression thickness calibration.
&RPSUHVVLRQ7KLFNQHVV
The calibration is carried out at the two fixed positions that act as compression
travel limits. Each time the power is turned on, the system locates one of the fixed
positions and determines the actual compression arm location.
&RPSUHVVLRQIRUFH
The compression force is measured with two strain gages located in the force
measuring block of the compression arm.
The strain gages connect to the Compression Board, which features a voltage to
frequency converter that is controlled by the on board microcontroller, which
further sends the data to the Supervisor Board via the RS-485 serial link.
&RPSUHVVLRQ6SHHG
The tacho pulses are taken to the Stereo Board that converts the pulses into a
speed reference voltage for the ECS drive to be used in the ECS mode.
&RPSUHVVLRQ3DGGOHGHWHFWLRQ
6HQVRUV &RUUHVSRQGLQJFRPSUHVVLRQ
6 6 6 6 SDGGOH
1 0 1 0 Biopsy paddle 9 and 18mm holes
1 0 1 1 Biopsy paddle 50x90mm square
1 1 0 0 Magnification paddle
0 1 1 0 Magnification paddle implant
0 1 0 1 Paddle implant
0 0 1 0 Paddle standard 18x24
0 0 0 1 Paddle standard 24x30
0 0 1 1 Spot magnification paddle
1 0 0 1 Spot magnification paddle
enhanced circle and square
12x12cm
1 0 0 0 Spot Paddle
1 1 0 0 Spot magnification paddle
enhanced circle and square
0 1 1 1 Delta 32 paddle
(&6/LPLWV
The ECS movement limits are accomplished with a flag that enters the two slotted
photointerrupters. Their purpose is to prevent the mechanical collision at both
ends of the ECS movement.
The signals connect to the Stereo Board, which prevents the ECS from driving to
the corresponding direction. The data from the switches are also taken to the
Supervisor Board via the RS-485 serial link.
7XEH3DUN0RYHPHQW6HQVRUV
7XEH3DUNPRYHPHQWOLPLWV
The Tube Park movement limits are accomplished with a flag that enters the three
slotted photointerrupters.
Two of the switches are located at end points of the movement. Their purpose is
to prevent the mechanical collision at both ends of the tube movement.
The purpose of the two other (middle) switches is to determine the position where
the tube speed will be reduced prior to reaching the ends.
The signals from the four switches are taken to the Collimator Board, which
further sends the data to the Supervisor Board via the RS-485 serial link.
6DIHW\&OXWFK
The Tube Park mechanism features a safety clutch in case the tube movement is
hindered. The status of the safety clutch is monitored with a microswitch that
closes when the clutch has been activated.
The microswitch connects to the Collimator Board, which further sends the data
to the Supervisor Board via the RS-485 serial link.
6WHUHR6HQVRUV
6WHUHR$QJOH
The angles for stereotactic operations are accomplished with four slotted
photointerrupters located underneath the Stereo Board.
The ±15 degree angles are determined by two opto boards on the sides. The center
position of the stereo arm is detected with the 6WHUHR 6HQVRU %RDUG which
contains two opto switches.
The exact positioning of the stereo arm is adjusted by moving the corresponding
opto boards.
6WHUHR6SHHG
The Stereo Speed is measured from the pulses of the Tacho Board, which is
attached to the Stereo motor. The pulses are taken to the Stereo Board, which uses
a circuit that controls the speed according to a speed reference.
)86(692/7$*(6$1'/(',1',&$7,216
127( )2//2:,1*92/7$*(0($685(0(176$1'7+(62)767$57
352&('85( ,16758&7,216 $5( )25 7528%/( 6+227,1*
385326(
Verify that the typical voltages listed are present. Refer to repair instructions in
service manual for the location of the test points. Voltages marked with * are
unregulated.
)86(6
3RZHU%RDUG
$0'
',$021')86(686('%<'(/7$
92/7$*(6
3RZHU%RDUG&RGH
TP1 to TP2 GND Unreg. +34 Vdc*
TP6 GND to TP3 Unreg +24Vdc*
TP6 GND to TP5 Unreg. -24 Vdc*
6WHUHR%RDUG&RGH
TP15 GND to TP14 Reg. +5 Vdc
TP7 GND to TP10 Reg. +15 Vdc
TP7 GND to TP13 Reg. -15 Vdc
$(&%RDUG&RGH
TP9 GND to TP10 Reg. +12 Vdc
TP9 GND to TP11 Reg. -12 Vdc
&RQWURO3DQHO%RDUG&RGH
TP6 GND to TP7 Reg. +5Vdc
6XSHUYLVRU%RDUG&RGH
TP6 GND to TP13 Reg. +5Vdc
:$51,1* %HIRUHFRQQHFWLQJWHVWHTXLSPHQWWR*HQHUDWRU%RDUG73*1'PDNHVXUH
WKDWWKHUHLVQRFRQQHFWLRQWRRWKHUSUHYLRXVO\WHVWHGWHVWSRLQWV,I73
*1'LVFRQQHFWHGWRVRPHRWKHUWHVWSRLQWWKDQ73RU73RQ*HQHUDWRU
%RDUG VKRUW FLUFXLW DQG GDPDJH WR HOHFWURQLF ERDUGV DUH OLNHO\ $IWHU
FKHFNLQJ*HQHUDWRU%RDUGPDNHVXUHWKDWWKHUHLVQRFRQQHFWLRQ73*1'
DQGWKHWHVWHTXLSPHQW
*HQHUDWRU%RDUG&RGH
TP43 GND to TP42 Reg. +20Vdc
TP43 GND to TP44 Reg. +310Vdc, during exposure
/(',1',&$7,216
When Diamond is switched on, the following LEDs should light up.
3RZHU%RDUG
&RGH &RORUGLVSOD\HG
H2, H10, H13, H15, H16, H22 Green
H9 Red
LA2 Orange
6XSHUYLVRU%RDUG
&RGH &RORUGLVSOD\HG
H3, H4, H5, H6, H7, H8, RST Green
*HQHUDWRU%RDUG
&RGH &RORUGLVSOD\HG
H1, H2, H3, H14 Green
6WHUHR%RDUG
&RGH &RORUGLVSOD\HG
H1, H2, H6, H7, H8, H9 Green
'ULYH%RDUG
&RGH &RORUGLVSOD\HG
H1, H2 Green
H3 Red
&ROOLPDWRU%RDUG
&RGH &RORUGLVSOD\HG
H18, H19, H20 Green
0(&+$1,&$/6<67(0&+(&.6$1'35(9(17,9(
0$,17(1$1&(
;5$<78%(3(5)250$1&(&+(&.
Enter the service mode by pressing kV- and Den- buttons down
simultaneously. The following screen appears:
Collimate the X-ray beam by a lead plate with a hole placed close the tube and
restrict the beam to an area slightly larger than the ionization chamber. Use 25 kV
and measure the radiation using 0 - 0.1 - 0.2 - 0.3 - 0.4 mmAl.
Plot your data on the chart and connect data points with a line. Take the 0.0 mmAl
reading and divide it by two. Find where that reading would intersect your graph
and determine the half-value layer from the aluminum axis. Use mAs that gives
you about 400 mR with 0 mmAl.
0.0
0.1
0.2
0.3
0.4
127( +9/#N90867EHPRUHWKDQPP$ODQGVKRXOGEHOHVVWKDQ
PP$O
Exit the service mode by pressing kV- and Den- buttons down
simultaneously.
7XEHHIILFLHQF\FKHFNRQWKHFDVVHWWHKROGHU
Test equipment required: Radiation meter
1) Center the radiation meter on the image receptor so that its ion chamber
is as close as possible to the chest wall edge of the image receptor.
3) For small focal spot testing, remove the image receptor. Make the
exposures in the service mode. For large focus, insert a cassette holder and
make the measurements on the cassette holder top.
mR/mAs
kV SMALL FOCUS LARGE FOCUS
DESIRED ACTUAL CHECK DESIRED ACTUAL CHECK
25 3-9 3-9
29 8 - 15 7 - 14
35 16 - 23 10 - 18
+ If the output of the tube consistently falls below the desired level, check kV,
mA, preheat.
Exit the service mode by pressing kV- and Den- buttons down
simultaneously.
)RFDOVSRWPHDVXUHPHQWSURFHGXUH
Test equipment required: Star phantom
Slit focal
Special focal spot measurement diaphragm
0HDVXUHPHQWRI/DUJH)RFXV
The default setting of the Large Focus is 28 kV, 50 mA. Large focus measurement
parameters for other inserts can be scrolled by pressing the mAs+ button.
3) Install the star pattern phantom to the special focal spot measurement
diaphragm to have the phantom at a correct angle. The diaphragm for
Diamond can be ordered from Instrumentarium Imaging.
4) Place a normal cassette in line with the front edge of the cassette holder
support. Put a guide under the tube side of the cassette so that the cassette
raises 12 mm (0.5 inc.) from the surface.
5) Make an exposure.
6) Develop the film and fill in the form for large focal spot to calculate the
focal spot size. Measure also the magnified diameter of star.
0HDVXUHPHQWRI6PDOO)RFXV
Enter the service mode by pressing kV- and Den- buttons down
simultaneously.
Remove the magnification table and place the cassette on the cassette holder
support as described in previously.
Make an exposure
Fill out the form for small focal spot to calculate the small focal spot.
Fw = C * BW/(M - 1) , FL = C * BL / (M - 1)
Fl = ________________ mm Fl
= _________________mm
Exit the service mode by pressing kV- and Den- buttons down
simultaneously.
,PDJHTXDOLW\HYDOXDWLRQSURFHGXUH
1. For the film processor evaluation procedure, refer to Diamond Installation
Instructions (86501), Chapter 7.2.4, Film processor evaluation.
2. Place a loaded cassette into the Bucky and select AEC operation using
settings (kV) as you would use for an average 4.5 cm compressed breast.
3. Place the RMI 156 phantom on the Bucky in the same position as a breast.
The nipple indent marker should be positioned away from the chest wall.
Use a standard 18 x 24-cm or 24 x 30-cm compression paddle.
4. Set the AEC detector to an inward position under the wax portion of the
phantom.
5. Make an exposure and process the exposed film. This image will represent
the imaging abilities of your machine using these clinical factors.
6. If the image is over or under exposed, make a AEC adjustment and repeat
the exposure.
7. Once an acceptable image has been obtained, have one or more viewers
identify and record which objects are seen. Use a magnifying glass to look
for the structures as you would on a patient’s mammogram. Use the
phantom scoring procedure on the following pages to score system
performance.
8. Make another exposure using magnification table and score that image.
9. Mark both films with the date, kV, mAs and their scores and file them for
future reference.
,0$*(6&25,1*:,7+7+(50,3+$1720
1) 4 Fibers - 1, 2, 3, 4
2) 2 Specs - 7, 8, 9
3) 3 Masses - 12, 13, 14
To meet the ACR criteria the minimum acceptable score would the yield a
minimum score of 10 for fibers, 8 for specs and 3 for masses. Total score would
the be 21.
If your mammography unit is significantly worse, you may wish to check the
accuracy of the kVp, the focal spot size, the half value layer, and the image
processor, You could also consult a medical physicist or engineer who specializes
in diagnostic radiology.
0HDVXUHPHQWRIWKHIHHGEDFNFLUFXLWU\
&$87,21 7KH UHDGLQJV IURP WKH IROORZLQJ UHVLVWDQFH PHDVXUHPHQWV GHSHQG RQ WKH
SUREHSRODULW\
Remove the low voltage connector ;from the HV-unit and measure on the
cable connector end the resistance between ground and pin ; (This is the kV-
feedback resistance on the INVERTER Board). It should beNΩ. (If the
power is ON, the value is 45 kohms).
If 0Ω check the cabling between the connector and the FILAMENT CONTROL
Board, and the cabling between FILAMENT CONTROL Board and the
,19(57(5%RDUG,IWKHUHDGLQJLVDSSN FKDQJHWKHSUREHSRODULW\. Plug
in connector ;.
&$87,21 ;LVKRWZLWKSRZHURQ
Remove the connector ; from the tube and measure on the cable connector end
the resistance between ground and pin ; (this is the mA feedback resistor on
the FILAMENT CONTROL Board). It should be about Ω when small focus
on and Ω when large focus on. Plug in connector ;.
N9P$VDQGH[SRVXUHWLPHWHVW
The measurement must be performed with a calibrated Unfors, RTI, PMX-I/M or
equivalent. The test is performed in the VHUYLFHPRGH. Select PDQXDOPRGH.
Use QRFDVVHWWHKROGHU at all when measuring the /) The Synchro Bucky feature
interrupts the x-rays for 50ms at grid turnaround points. The interruption may
interfere with the measurement.
When measuring the 6), PDJQLILFDWLRQ WXQQHO should be used. Set the
parameters as shown in the table below and make the exposures.
Record the kV, mAs and time reading after each exposure. The time reading is to
be compared with the Diamond control panel time display.
The maximum tolerances for kV, mAs and exposure time are:
For kV
± ( 1.5 kV + internal acceptance limit ±1kV)
For mAs
± 4 mAs for mAs values less than or equal to 40 mAs
± 10 % for mAs values greater than 40 mAs
Measure with scope the voltage from the MAFB TP21 and divide it by
resistance of 19.6R1 for large focus and by 78.4R for small focus.
$8720$7,&(;32685(&21752/(9$/8$7,21
Test equipment needed: Sensitometer, Densitometer, Acrylic blocks, Film &
Cassette. ATTACH THE BUCKY DEVICE.
Enter into the service mode by pressing kV- and Den- buttons down
simultaneously for 3 seconds.
4XLFN$(&WHVW
Testing procedure:
127( &DUU\RXWWKLVWHVWRQFHDZHHN
([SRUWLQJDQGVWRULQJWKHV\VWHPVHWWLQJVGDWD
The system settings can be exported in hexadecimal format and can be viewed and
stored in Windows HyperTerminal. For HyperTerminal properties setting and
about how to capture, refer to Chapter 3.4, Using HyperTerminal interface with
Diamond.
$XWRPDWLF$(&GHWHFWRUJDLQDGMXVWPHQW
In order to adjust the gain of each detector, follow these instructions:
1. Place about 40mm acrylic on the cassette holder. Use 28 kV and Auto
Time mode and make an exposure.
2. Enter the service mode and "AEC" from the program groups list. Then
select "8:Detector Adjustment" from the AEC list. Diamond compares
readings from other detectors and then calculates new factors for them in
relation to C1.
6. Finally, the program asks for confirmation on the system setting changes.
$GMXVWPHQWRIDQLQGLYLGXDOGHWHFWRU
Individual detectors of the Diamond AEC can be adjusted separately in service
program PrD02.
For example, if Det1 reading is 1600 and Det2 reading is 1400, then Det2
amplification must be changed: 1600 times 0.98 divided by 1400 = 1.12, which
after second exposure gives a Det2 reading of about 1570.
Note that above-mentioned detector factors are experimental results for most
film/screen/processor combinations. In order to verify these, measure O.D. and
adjust the gain for each detector separately. Especially detector C4 may change
because of the heel effect.
0(&+$1,&$/&+(&.6
&RPSUHVVLRQV\VWHP
The part of the equipment, which undergoes the most use is the compression
system, so it needs regular attention and treatment when necessary. Spread some
grease (MBDE or other high-quality grease) on the compression driving axle and
support joints. If the driving wheels are noisy or the hand wheels difficult to
rotate, we strongly recommend that you remove the C-arm and grease the inside
of the compression system. Check also the digital compression display according
to the test described in the Installation Instruction.
&DUPURWDWLRQPHFKDQLVP
Reserved text place.
&ROXPQDQGFDUULDJH
Check all cables, especially the cables coming from the C-arm. Check also the
counter weight wire condition.
&RPSUHVVLRQIRUFHFKHFNLQJ
Place a scale under the paddle and compress it like a breast using hand wheels.
The mechanical clutch should stop the compression movement before the pressure
reaches 200 N (45 lbs.). If the compression exceeds this, adjust the clutch by
rotating screw.
7HVWLQJWKH[UD\ILHOGLOOXPLQDWLRQ
In order to test the x-ray field, perform the following procedures:
1) Select the desired film size, load the film into the cassette and put the
cassette and diaphragm in their places.
4) Place for example some coins in the corners of the area illuminated on the
cassette holder and make an exposure.
The coins on the film should not be more than ± 5 mm out of the exposed field.
The light field can be adjusted with two screws at the rear end of the collimator.
The voltage for the field light can be measured at conductor stands (marked by
arrows in the image below).
$GMXVWPHQWWXEHKHDGPRYHPHQWDQGWRUTXH
Exact exposure position of the tube head (tube park stop) may be adjusted with
the nut (A) and the torque of the movement with bolt (B). Set the torque at about
50-60N.
5(*8/$50$,17(1$1&(3/$16
Refer to service agreement what kind of maintenance plan has been ordered.
(YHU\GD\PDLQWHQDQFHSODQ
• QA check.
:HHNO\PDLQWHQDQFHSODQ
0RQWKO\PDLQWHQDQFHSODQ
• Confirm that the exposure indicator light stays on for the length of the
exposure.
• Confirm that the buzzer stays on for the length of the exposure.
• Inspect and confirm that all warning labels and embossed, painted or other
wearable labels have not been defaced or worn and that they are legible.
• Confirm that the ready light is on when various modes are selected. A
daily QA check is recommended. Software monitors the status of each
detector and prevents an exposure with a faulty detector.
• Check that the C-arm moves smoothly in all positions. Check that the
displayed angle corresponds to the selected angle. Also check that ± 15°
motorized stereo angulation movement is functioning correctly.
• Check that all buttons on the control panel are functioning and check that
all lights and display segments function immediately after turning the
main power on.
• Switch on the x-ray field illumination light (46) and check that it covers
the imaging area.
• Pay attention to any strange sounds that your unit may make inform your
service technician about these immediately. if you have any doubts about
the correct safe operation of the unit, switch it off and do not use it until
you can it checked by an Instrumentarium approved Service Technician.
• Check that the emergency stop buttons function correctly. Rotate the C-
arm or move the compression paddle using a foot pedal and press one of
the emergency stop buttons (18). All motorized movements should stop.
Release the emergency stop button, switch on the unit once again and then
repeat the test using the emergency stop button on the opposite side of the
unit.
$QQXDOPDLQWHQDQFHSODQ
• Safety ground Verify that the metal parts of the unit are
connected to protective ground.
• Oil leaks Check that the high voltage unit and the tube
housing do not leak oil.
• Clean the High Voltage Connections and grease with clean grease for
example DowCorning #4.
• Check for the operation of the C-arm gear and lubricate it.
• Inspect the floor fixing points to ensure the secure and safe fixing of the
support base plate of the unit.
• Remove debris from the cassette loading mechanism and lubricate it.
• Focal Spot
• If the unit is been used very much, apply grease (for example, Optimal
HT2EP Optitemp®) to lubrication tip of the C-arm rotation system. If the
unit is been used less, then lubricate every two years.
7+,63$*(+$6,17(17,21$//<%((1/()7%/$1.
5(3$,5,16758&7,216
&2/8015(3$,5
3UHSDULQJWKHFDUULDJHDQGFRXQWHUZHLJKW
127( 7KHFRXQWHUZHLJKWLVKHDY\%HFDUHIXOZKHQKDQGOLQJLW
&$87,21 'XULQJZRUNDURXQGWKHFROXPQEHFDUHIXOQRWWRGURSDQ\SDUWVLQWRLW
:$51,1* 0DNHVXUHWKDWWKHVXSSRUWKROGVWKHFDUULDJHVWHDGLO\DQGWKDWWKHPDWHULDO
XVHGWRWLHLWXSZLWKWKHFROXPQGRHVQRWEUHDNULSRUDQ\KRZJHWGDPDJHG
VR WKDW WKDW LW DOORZV WKH VXSSRUW WR IDLO ,Q VXFK FDVH VHULRXV GDPDJH WR
VHUYLFHWHFKQLFLDQ V RUWRWKHXQLWLVOLNHO\WRUHVXOW
5HSODFHPHQWRIWKH=PRWRU
1) Prepare the carriage as instructed Chapter 6.1.1, Preparing the carriage and
counterweight.
4) Place the column cap back. Check the alignment by running the motor
with slack wires before installing the cap assembly.
5) Secure the screws. In order to achieve correct shaft alignment the motor
assembly bolt (A) must be tightened (max. 15Nm) holder screws.
5HSODFHPHQWRIWKHFROXPQFDS
1) Repeat instructions in Chapter 6.1.1, Replacement of the Z-motor.
2) Open up a bolt and a nut indicated by arrows. Then pull out the wheels and
the counterweight wire comes off the cap.
3) Now the counterweight wire must be placed underneath the wheels of the
new column cap. Open up the same bolt and nut as pictured above. Keep
all parts associated with the bolt separate from the parts associated with
the nut. Do not mix any of these with the parts of the old column cap.
4) Place the counterweight wire lock piece between the wheels and drag it
down towards the counterweight when you attach the column cap back on.
5HSODFHPHQWRIWKHFRXQWHUZHLJKWZLUHV
1) Prepare the carriage as instructed Chapter 6.1.1, Preparing the carriage and
counterweight.
&$87,21 5HOHDVHFDUHIXOO\VRWKDWSDUWVZRXOGQRWGURSGRZQWRWKHFROXPQRSHQLQJ
EHKLQGWKHZLUH3LHFHVLQWKHRSHQLQJPD\VHULRXVO\DIIHFWWKHPRYHPHQWRI
WKHFDUULDJH
3) Open up a bolt and a nut indicated by arrows. Then pull out the wheels and
the counterweight wire comes off the cap.
5) Loosen the plastic piece that locks the wire with the metal pin. Remove the
wire
6) Remove the plastic piece and pull the warning pin out.
7) Pull the wire out of the metal lock piece. Slide a new wire in through it.
8) Attach the column cap back on. Place the counterweight wire lock piece
between the wheels and drag it down towards the counterweight .
&$87,21 $WWDFKFDUHIXOO\VRWKDWSDUWVZRXOGQRWGURSGRZQWRWKHFROXPQRSHQLQJ
EHKLQGWKHZLUH3LHFHVLQWKHRSHQLQJPD\VHULRXVO\DIIHFWWKHPRYHPHQWRI
WKHFDUULDJH
10) Insert the warning pin and the plastic lock piece. In case one of
counterweight wires breaks up, the remaining wire pulls the warning pin
to the other side and it starts to scratch the column. This is a warning
signal for wire replacement.
11) Lay out the wire evenly on the top part of the counterweight cage. Tighten
up the plastic lock piece. Add cover plate.
12) Tighten three bolts a bit. Pull the wire out a bit to remove slack. Secure the
wire with three bolts.
/2&$7,212)(/(&75,&$/%2$5'6$1'0$,16
75$16)250(5
The location of Diamond boards is indicated in the image below. If you replace
EPROM chips, use microchip extractor tool.
5HSODFHPHQWRI3RZHU6XSHUYLVRUDQG*HQHUDWRUERDUGV
1) Unplug the mains cord.
4) Remove the plastic nuts from the corners of each board, and replace the
board.
5) Check kV, mA and preheat. Refer to Chapter 5.1.6, kV, mAs and exposure
time test.
5HSODFHPHQWRIWKH$0'ERDUG
1) Unplug the mains cord.
2) Remove the upper cover of the C-arm and disconnect all cables leading to
the anode motor drive board.
3) Replace the board, reconnect all cables, and check the assembly.
5HSODFHPHQWRI&RQWURO3DQHO%RDUG
1) Open up the screws at the back of the Diamond Control Panel.
2) Gently separate the front and rear pieces of the Control Panel
5) Attach the front and rear parts back together and secure with screws.
5HSODFHPHQWRIWKH&RPSUHVVLRQ%RDUGRU&RPSUHVVLRQ'LVSOD\%RDUG
1) Open up four screws at the bottom part of the Position Control Center
4) Attach the top and bottom parts back together and secure with screws.
5HSODFHPHQWRIWKHPDLQVWUDQVIRUPHU
1) Remove the power supply plug.
3) Remove all cables connected to the transformer, the nut and cover plate
on the transformer.
&$505(3$,5
5HPRYDODQGDVVHPEO\RIWKH'LDPRQGFRYHUV
For removal of the c-arm covers, refer to Diamond Installation Instructions
(86501), Chapter 6.6.3, Left side cover and Chapter 6.6.4, Right side cover.
Perform the assembly process in reverse order. Start the removal process with the
right cover, because the emergency stop cable is attached to the right side of the
C-arm carriage.
$WWDFKLQJWKHWXEHKHDGFRYHUV
For protection of the user’s fingers, Diamond has sliding sheets on both sides of
the unit to close up the opening caused by the ParkBack movement. The plastic
sheet slides along the groove of the plastic rails.
127( :KHQLQVWDOOLQJWKHWXEHFRYHUPDNHVXUHWKHILQJHUSURWHFWLRQVKHHWVVOLGH
LQVPRRWKO\DQGHYHQO\VRWKDWWKH\DUHQRWWDQJOHGXS
Figure 6.1 Finger protection sheet Figure 6.2 Groove in the plastic
rail
2) In order to facilitate the process, drive the tube head forward to attach the
front-end finger protection sheets and backward to attach the rear end
sheets.
2) Push the sheets forward and pay attention that the sheets slide smoothly
along the plastic rails.
&$87,21 %HYHU\FDUHIXOZLWKWKHRSWRV:LWKURXJKKDQGOLQJWKHUHLVDKLJKULVNRI
GDPDJLQJWKHRSWRV
4) Secure side screws. Do not use any tools or the paint might scrub off.
&$87,21 ,IWKHFRQQHFWRULVDWWDFKHGWKHZURQJZD\QRWRQO\WKHEXWWRQVZLOOQRW
ZRUNEXWDOVRWKHV\VWHPLVDWULVNRIJHWWLQJGDPDJHG
2) Lower the top cover (A). Check that cables are evenly spread out inside
the cover. Then, lay down the front part (B). Check that lower tube head
cover slides in smoothly into the top cover groove. Then lower the rear
end of top cover down so that the top cover lies all over the lower top
cover (C). Check that the resin covers smoothly the opening between top
and lower tube head cover.
3) Secure with fingers the front end of the top cover. Do not use any tools or
the paint might scrub off.
5HSODFHPHQWRIKDQGOHV
1) Open up screws to remove the handle in use.
5HSODFHPHQWRIWKHYHUWLFDOGULYHFRQWUROVZLWFK
1) Remove the Position Control Center cover as instructed in Chapter
6.4.12, Adjustment of the mechanical compression limit.
2) Gently separate cover and bottom part of the Position Control Center.
3) Inside the cover, open the screws that hold the vertical switch secured.
5) Attach cover and bottom part back together. Secure with screws.
5HSODFHPHQWRIWKH[UD\WXEH
1) Remove the tube head covers.
3) Unplug these cables and ground wire (attached to the high voltage unit).
5) Replace the tube and secure it with the four bolts. Then attach cables and
ground wire.
5HSODFHPHQWRIWKHKLJKYROWDJHXQLW
1) Disconnect the ground wires and the plugs of the high voltage unit, and
loosen the four screws, two on each side of the unit.
2) After replacement, check the assembly and the ground cables carefully.
4) Readjust the generator kVs again. Refer to Chapter 5.1.6, kV, mAs and
exposure time test.
5HSODFHPHQWRIWKHFRPSUHVVLRQV\VWHP
1) Drive the stereo arm to approximately 10° angle to either side
2) Release three bolts connecting the compression system to the stereo arm
In order to reach all screws, use a ball-headed Allen key for this procedure.
4) Open up six bolts that connect the compression system to the c-arm
5) Disconnect cables and slide the compression system out and perform
necessary service procedures
5HLQVWDOOLQJWKHFRPSUHVVLRQV\VWHP
5) When you want to reinstall the compression system, align it with the c-arm
and screw in the six bolts that were released in step 4
5HSODFHPHQWRIWKHFRPSUHVVLRQPRWRU
In order to replace the compression motor, the compression system has to
detached as instructed in section 6.4.8, Replacement of the compression system.
Then proceed to following steps:
2) Release three bolts that attach the motor to the compression system.
4) Loosen up the worm drive clip and slide it towards the motor so that the
parts around tacho meter come off
5) The tachometer is fixed to the motor axle with a screw. Open the screw
and detach the tachometer disc.
,QVHUWLQJDQHZPRWRU
8) Assemble the rest of the tachometer parts so that it will look like this:
10) Fix the motor with three bolts that were released in step 2
11) Reattach the compression system cap that was detached in step 1
12) Attach the compression system back to the stereo arm as instructed in
section 6.4.8, Replacement of the compression system
5HSODFHPHQWRIWKHFRPSUHVVLRQWKLFNQHVVPHDVXUHPHQWV\VWHP
The compression thickness is measured with a special tape that is read by two
sensors (front and rear sensor).
5HSODFHPHQWRIWKHIURQWVHQVRU
5HSODFHPHQWRIWKHUHDUVHQVRU
1) Drive the stereo arm so that an opening marked in the picture becomes
visible
$GMXVWPHQWRIWKHFRPSUHVVLRQPHFKDQLVPEHOW
The compression belt should
bend about 1cm, when pressed
with finger. If not, adjust the
belt.
$GMXVWPHQWRIWKHPHFKDQLFDOFRPSUHVVLRQOLPLW
1) Open the four screws to remove the compression arm cover.
2) If the mechanical limit releases when too few kilograms are applied,
tighten up the mechanical clutch adjustment nut with a 13mm wrench. If
the compression allows too many kilograms to be applied, loosen the nut.
5HSODFHPHQWRIWKH&DUPPRWRU
1) Remove Generator Board.
2) Open up screws that hold C-arm motor. If more space is necessary, remove
Supervisor Board.
3) Loosen up the worm drive clip and slide it towards the motor so that the
parts around tacho meter come off
4) The tachometer is fixed to the motor axle with a screw. Open the screw
and detach the tachometer disc.
7) Assemble the rest of the tachometer parts so that it will look like this:
10) Check kV, mA and preheat. Refer to Chapter 5.1.6, kV, mAs and exposure
time test.
5HSODFHPHQWRI(&6PHFKDQLVP(&6PRWRU
In order to replace the ECS system, perform steps 1-11. In order to change the
ECS motor, perform all steps.
1) Drive ECS half way and then lock up the counterweight with the C-arm
assembly pin.
3) Then open panduit and push the cable tubes (3) upward so that you get
more maneuverability for the C-arm rest.
4) Prepare a steady place right next to the unit where the C-arm can rest while
you change the ECS-motor.
5) Unplug X2 and X15 from Stereo Board. Then, gently pull the X15 cable
out a bit.
6) Remove brake lock. Rotate the brake lock upwards so that pieces inside
would not fall off.
7) Lay the brake lock somewhere in upright position so that the pieces would
stay inside.
8) Remove topscrew (A). Loosen up others (B, C). Perform this on both
sides.
9) Lift the C-arm a bit upwards and pull it out very gently because the Delta
32 cable is located inside the C-arm tube.
&$87,21 %HFDUHIXOQRWWRH[HUWVXGGHQRUVWURQJSXOOLQJIRUFHRQ'HOWDFDEOH
10) Lay the C-arm on some steady surface that is very close to the unit.
13) Hold steady the wheel marked with dotted line and lift up the motor.
15) Replace the old motor with a new one. Secure the motor to the plate.
16) Repeat steps 4-13 in reverse order. At step 12, make sure that the belt is
tight enough to bear stress without jumping over. Apply some grease to
the trapezoidal shaft (for example, Optimal HT2EP Optitemp®).
17) Check kV, mA and preheat. Refer to Chapter 5.1.6, kV, mAs and exposure
time test.
27+(5$'-8670(176255(3/$&(0(176
5HSODFHPHQWRIWKHOLJKWILHOGEXOE
For instructions to change the light field bulb, refer to Diamond Operation
Instructions (86500), Chapter, 16.6, replacement of the light field bulb.
*ULGDOLJQPHQW
The grid is aligned and the speed is adjusted at the factory and should normally
not be adjusted at the site. If adjustment and/or alignment is necessary, however,
follow the instructions below.
1) Remove the carbon fiber cover on the Bucky. In the small Bucky the grid
should be at the bottom of the grid drive block groove and fastened with
four setscrews. In the large Bucky the grid should be fastened only with
two setscrews and aligned by the two rear set screws.
2) Check that the grid is correctly centered. The line on the grid is aligned
with imaginary dashed line (A).
3) Install the Bucky and check that the grid does not touch the side lips of the
Bucky (B). Check also that the grid is aligned parallel to the side lips (C).
4) If the intermittent movement of the grid reaches one of the side lips,
remove the Diamond c-arm bottom cover.
6) Slide the grid system to the opposite direction from the side lip.
7) After the adjustment, tighten up the screws and check the movement.
Before initiating the exposure, attach the service diaphragm. For exposure,
use manual mode and set for example 20kV and 200mAs.
8) When the Bucky is installed, the grid system drives to home position.
Check that the grid system movement is not blocked by the edges of the
opening.
*ULGPRYHPHQWPHFKDQLVPFKHFN
The first grid movement from left to right takes up about 670-820ms. The rest of
the movements take up about 3070-3260ms. Exposure stops for 150ms when grid
switches the direction of movement. If grid lines appear, check that the wheels
that connect the motor to the grid are correctly aligned and secured.
1) Remove Bucky and plastic cover (held by eight screws). Then open up
four screws that hold the cassette holder table.
2) Gently pull the cassette holder table out a bit. You may need to unplug
cables CX92, CX93, CX96-98, CASX1, CAS, DRX5, DRVX1, GRID,
AECX2 (near AEC detectors) and AECX3 (near AEC detectors). In order
to reach some of the cables, remove the bottom cover.
4) Press inwards the pin that moves the grid. It should slide in and out
smoothly. The pin should protrude the frame by 4mm (C). Check also that
the pin is tightened up all the way (D). Measure slack by holding the pin
between fingers and slide the mechanism to left and right.
5) Drive the mechanism to extreme positions and check that the opto flag
sensor does not disturb the movement of the opto flag. Check that flag is
correctly aligned and does not touch grid mechanism frame.
6) If the problem still persists or there was significant slack at step 6, replace
the movement mechanism assembly. Release motor screws (A). Loosen
up screws that fix connection wheel package to motor axis (B). Slide out
the motor (C). Open two screws that fix the mechanism to the cassette
table (D).
5HSODFHPHQWRI%XFN\FDUERQILEHU
If the carbon fiber of the Bucky has become scratched or otherwise damaged, it
must be replaced. The cover is attached with 6 screws (two on topside, two on
sides and two at the bottom). Open up screws and slide out the cover out. Then
slide a new cover on and secure with screws.
5HSODFHPHQWRIFDUPPHPEUDQHEXWWRQV
Membrane buttons can be replaced by unplugging flat cables inside the c-arm.
Then remove the button and residues, if any. Then attach a new one in place and
plug it in.
7528%/(6+227,1*
,1)250$7,21)257+(6(59,&(&$//
Should it happen that a problem arises, the following pieces of information help
service technician find out and solve the problem faster.
D *HQHUDO UHTXLUHGLQIRUPDWLRQ
E )RUSKRWRWLPLQJTXHVWLRQV $(&
F 2WKHUSUREOHPV
',$021'(5525&2'(6
'LDPRQGHUURUFRGHFKDUW
Time-out limit in cases of communication is set at 40 sec. For all other situations
it is set at 20 sec.
(UURUFRGHLQ
WLPHUGLVSOD\ 5HDVRQIRUHUURU &RUUHFWLYHDFWLRQ
(U Program check sum error Call service.
(BattRAM).
(U Tube Park safety clutch acts - Clear and try again. If problem
tube obstacled, microswitches persists, check microswitches and
misaligned or a collimator Board collimator.
problem.
(U Tube movement back up time Check all four flags (rear/front, 2
reached - mechanical friction, for slow speed).
opto, flag or Collimator Board
problem.
(U Stereo not in 0 degrees when Drive the stereo arm to 0 degrees.
exposing or Stereo Board
problem.
(U Exposure button released during Hold the exposure button pressed
exposure. down until the audio-signal
terminates.
(U Device not ready for exposure. Check rear plate, it might be
Rear plate error. moving.
(U Error in stereo motor pulse Check motors and encoder optos.
measuring or missing pulses.
(U Tube head heat limit exceeded. Wait, the remaining cooling time
is displayed on Control Panel
LCD-screen.
(U Left side collimator back up time Check collimator for any
reached. blocking objects.
(U Tube head is not in front. Drive tube head to the front.
There is no error message on the battery reaching the end of its lifetime.
Five quick beeps ((UURUPHVVDJH) indicate that the Control Panel is missing.
Attach the Control Panel to the Diamond unit.
1232:(5,1',&$7,216%2$5'/('6
A. Check the mains switch.
B. Check that the power cable and the display cable are connected.
C. Check the circuit breaker in the rear bottom.
D. Check that Power Board fuses are OK.
$(&,123(5$7,9(
Check grounding
- Check Cassettes/films/screens
- Check detector positioning
- Check HV-grease
- Check spot diaphragm
6833/<92/7$*(6
If you suspect changes in Diamond electrical system, check the voltages according
to the list in Diamond Service Manual (86502), Chapter 4.13, Fuses and voltages.
*5,'/,1(6
The grid in the Bucky is stuck or noisy.
(&67528%/(6+227,1*
3UREOHP ,WHPVWRFKHFN
The ECS does not move. - The signal from the switches reaches the X2 on
Stereo Board
- Motor connection to X15 on the Stereo Board
- Movement limits are OK (X2 on Stereo Board)
- The motor works (use external voltage supply)
The ECS does not stop when - The connector X2 is not properly connected
the compression reaches the
current limit.
The ECS stops before the - The compression motor exhausts before the
compression reaches the limit. Increase the speed in service program
current limit. PrG01.
Return function does not work. - The signals from the switches do not reach
connector X2 on Stereo Board
The ECS moves too slow. - Adjust the speed at trimmer R42 on Stereo Board
The ECS moves too fast. - Adjust the speed at trimmer R42 on Stereo Board
The ECS does not move from - Check pulses from compression motor encoder
ECS pedal. X17/Pin 5 on Stereo Board
If the motor or whole ECS mechanism has to be replaced, instructions are located
in Chapter 6.3.14, Replacement of ECS system/ECS motor.
6(5,$/%86&20081,&$7,216&+(&.
Check that the following boards have semi-dim LEDs. Non-existent or totally
bright LED indicates a communication problem.
127(62162)7:$5(83*5$',1*
9
%HIRUHXSJUDGH
$IWHUXSJUDGH
127( :KHQ\RXDUHORDGLQJGHIDXOWVPDNHVXUHWKDW\RXDUHLQ6SHHGVPHQX
A: Check that the counterweight tiles are laid out evenly completely inside
counterweight cage. If the noise appears only in lateral position, check and
adjust four rollers between the carriage and the column. Instructions in
Diamond Installation Instructions (86501), Chapter 5, Testing the control
functions.
63(&,),&$7,216
6<67(063(&,),&$7,21602'(/0*;
Manufacturer’s
quality system ISO9001
Manufacturer’s environmental
management system ISO14001
Apparent resistance of
supply mains
Enclosure IP 20
(19,5210(17$/63(&,),&$7,216
*(1(5$725
Type B
kV ripple < 5%
$8720$7,&(;32685(&21752/ $(&
Solid state 15 step density selector
Operating range: for large focus 23 - 35 kV
for small focus 25 - 35 kV
),/75$7,21
DIAMOND 0.025 mm Molybdenum
0.025 mm Rhodium
0.50 mm Aluminum
326,7,21,1*
movement rotation ±185°
vertical 760 mm (30")
5(&25',1*6<67(06
Contact for 18 x 24 and 24 x 30cm film
Bucky for 18 x 24 and 24 x 30cm film
Magnification for 18 x 24cm film 1.6, 1.8 & 2.0
exposed size 10 x 24
Stereotactic device for 18 x 24cm film or digital ccd-camera
&255(6321'(1&(%(7:((1
;5$<),(/'$1',0$*(5(&(37,21$5($
Chest wall 0...+3 mm
Other walls +1...+5 mm
Correspondence between collimator light field and X-ray field PPDQ\ZDOO
&2//,0$725
Type MGB 600
Filtration 0mm Al
Electrical input 34Vac (for half-step motors)
9-14Vdc (for 12Vdc lamp)
Lamp type HLX (halogen), 12V, 50W
5$',$7,21),(/'
Maximum radiation field 24x30cm
$77(18$7,21(48,9$/(17
Bucky, cassette tunnel
or magnification tunnel 0.25 mm Al (Total of all layers)
$17,6&$77(5*5,'
52&*5,'
',63/$<$&&85$&<
and
0($685(0(17&5,7(5,$)25(;32685(7,0(
6&+(0$7,&6/$<2876
The name Alpha X in the schematics/lay-outs is the project code name of Diamond.
'RFXPHQW1DPH 'RFXPHQWFRGH
AMD Board 32305-3S_, 32305-4K_
AEC Board 85001-3S_, 85001-3K_
Generator Board 85003-3S_, 85003-3K_
Collimator Board 85009-3S_, 85009-3K_
Drive Board 85011-3S_, 85011-4K_
Stereo Board 85013-3S_, 85013-3K_
Power Board 85020-3S_, 85020-3K_
Compression Board 85022-3S_, 85022-3K_
Compression Display Board 85024-4S_, 85024-4K_
Caslock Sensor Board 85026-4S_, 85026-4K_
Tacho Board 85028-4S_, 85028-4K_
Sensor2 Board 85030-4S_, 85030-4K_
Collimator Sense Board 85032-4S_, 85032-4K_
Csense Board 85065-4S_, 85065-4K_
Control Panel 85069-3S_, 85069-3K_
Supervisor Board 85078-3S_, 85078-3K_
Casopto Board 85090-4S_, 85090-4K_
Flatcable Adapter Board 85094-4S_, 85094-4K_
HV-plus 85108-4K_
HV-minus 85109-4K_
HV-unit 85112-4S_
Flat cable for membrane switches of Control panel 86000-4S_
AEC Membrane Switch 86001-3S_
Vertical Membrane Switch 86002-4S_
Tube Park Switch 86003-4S_
Projection Switches 86004-4S_
Stereo Switch 86005-4S_
Tube Cover Switches 86006-4S_
&$%/('5$:,1*6
The name Alpha X in the cable drawings is the project code name of Diamond. The
location of the cables can be checked in the Appendices section.
'RFXPHQW1DPH 'RFXPHQWFRGH
Main cable CB10 85016-3J_
Cable SB1 85017-4J_
Cable SB31 85018-4J_
Cable PW20 85019-4J_
Flat cable CS1 85035-4J_
Flat cable CS2 85036-4J_
Cable CS3 85037-3J_
Cable CS4 85038-4J_
Cable CS20 85039-4J_
Cable CS31 85040-4J_
Cable CS32 85041-4J_
Cable CS33 85042-4J_
Cable CS36 85043-3J_
Cable CS35 85044-4J_
Cable CS21 85045-4J_
Cable CS22 85046-3J_
Cable CS23 85047-4J_
Cable CS24 85048-4J_
Cable CS25 85049-3J_
Cable CS26 85050-3J_
Flat cable CS27 85051-4J_
Cable CS28 85052-4J_
Cable CS29 85053-4J_
Cable CS40 85054-3J_
Cable CS11 85055-4J_
Cable PW1 85056-4J_
Cable PW2 85057-4J_
Cable PW3 85058-4J_
Cable PW21 85059-4J_
Cable PW30 85060-4J_
Cable PW31 85061-4J_
$33(1',&(6
This chapter includes a Diamond hardware block diagram and test display point
lists. For software block diagrams as well as a list of factory default settings, AEC
calibration sheet, AEC detector pattern, refer to Diamond Installation Instructions
(86501), appendices. Note that the name Alpha X in the diagram is a project code
name of Diamond.
09.05.2001
SDJH LQWKHXSSHUFRUQHU
test1 Rear collimator stepmotor home limit (opto switch behind collimator) test1 test2 test3 test4
test2 Rear collimator stepmotor steps ready (from microprocessor) test5 test6 test7 test8 test9
test3 Side collimator stepmotor home limit (opto switch) test10 test11 test12 test13 test14
test4 Side collimator stepmotor steps ready (from microprocessor) test15 test16 test17 test18 test19
test5 Filter stepmotor home limit (opto switch)
test6 Filter stepmotor steps ready (from microprocessor)
test7 Tube park position limit (opto switch)
test8 Tube almost in park position (opto switch)
test9 Tube almost in front position (opto switch)
SDJH LQWKHXSSHUFRUQHU
test41 Filter selection button in control panel (membrane switch) test41 test42 test43 test44
test42 Filter selection button in compression panel (membrane switch) test45 test46 test47 test48 test49
test43 Compression force plus button (membrane switch) test50 test51 test52 test53 test54
test44 Compression force minus button (membrane switch) test55 test56 test57 test58 test59
test45 Light button (membrane switch)
test46 Cassette sense switch 1 (hall sensor on lower right)
test47 Cassette sense switch 2 (hall sensor on lower left)
test48 Cassette sense switch 4 (hall sensor on upper left)
test49 Cassette sense switch 8 (hall sensor on upper right)
SDJH LQWKHXSSHUFRUQHU
test61 Stereo left button (membrane switch) test61 test62 test63 test64
test62 Stereo right button (membrane switch) test65 test66 test67 test68 test69
test63 Stereo center button (membrane switch) test70 test71 test72 test73 test74
test64 C-arm left lateral pos button (membrane switch) test75 test76 test77 test78 test79
test65 C-arm left oblique pos button (membrane switch)
test66 C-arm center pos button (membrane switch)
test67 C-arm right oblique pos button (membrane switch)
test68 C-arm right lateral pos button (membrane switch)
test69 Cassette sense (opto switch)
test70 Paddle foot switch
MAINS
POWER CABLE
85083-4J
20354 23388 MAINS SWITCH
NETWORK CABLE
85088-4J Stop
button C-arm Tube Head Tube motor
Field Light
Coll motors
Filter Motor
Tube Head pos.
Moment
MAINS FUSES LINE FILTER
85084-4J
CX13 Motor
Tubehead panel
Mirror Solen. Sensors
Stop
CS5 Terminal block button 86006-4PJ CX3 CX4 CX5 CX6 CX7 CX8
CS34
X5
Footpedals Generator Board Adapter
85082-4J CB1 CS9 85092
Board 85076-3J
85086-4J 85087-4J
X4
CB30 CS31 CS32 CS33
85062-4J 85041-4J 85042-4J
X-ray Tube 85040-4J
X2
+310Vdc +20Vline X1,2,7 X15 X4
X3
PW1 Projection Projection X1 X9 X10 X4 X3
CS6 panel panel AMD Board
COM1 85056-4J Collimator Board
Molex
CS35
+20Vline PW33
X6
85080-4J +310Vdc 90136
X6
X5
86004-4P Molex
X1
85096-4J
X2
+34Vdc 90136
CS0 PW2 230Vac Tube Tube
87C51FA 85044-4J
COM2 Park +34Vdc
X10
Park
panel
X1
85077-4J HV Unit
X4
panel
85057-4J
-15Vdc +15Vdc 86003-4P +5Vdc +15Vdc 230Vac
Carriage X3
+15Vdc -15Vdc
Molex Molex PW32 FAN
X6
X2
Minifit X7 X3
X8
Adapter Adapter Minifit
Board Board
X13
C-arm CS3 -24Vdc +24Vdc 85063-4J
X8
Sensors 85037-4J CX10 CS11
85035-4J
86000-4P 85055-4J
+34Vdc
CS4 CB30
85038-4J
Control Panel
X11
C-arm
Motor X9 Power Board 85062-4J CX9
CS1 CX2
CX1 CS2 PW31
85036-4J
87C51FA
CS7 CX11 85061-4J
Z-motor (Optional)
85085-4J
X8 X5 X6
X20
-24Vdc CX12
SB31
PW3 CB10 85018-4J
X9 X7
X10
85016-3J
+5Vdc 87C51FA
X12
DS75176 (RS485)
85058-4J Compression Block
X6
DS75176 (RS485)
X3 MUX Molex
Compression Board
SB40 &
X14
90136 +5Vdc
X11 85064-4J Cmpr. Display Brd
X3
CS8
X10
85079-4J MAX232 (RS232)
DS75176 (RS485) DS75176 (RS485)
X1
TL16C451
X22
87C51FA
X2
X18
-15Vdc +34Vdc 85060-4J
MAX232 (RS232)
X2
ECS limits
CX24 X2
SB2
MAX232 (RS232)
Stereo Board CS23
85047-4J
X17
+5Vdc
X15
85081-4J i386EX +15Vdc Compression Motor
ECS Motor
& Tacho CS40
X1
85054-4J
CS22 Comp. sensors Comp. panel
15 pin D-connector (rear box) CS21 Stereo Sensors
X3
X4 X16 X21 X11 X19 85046-4J
MAX232 (RS232) 85045-4J Strain Gages 86001-4P
DS75176 (RS485) (measure comp force)
SB1 PW20
PW21
85059-4J
85051-4J
CS27
+5Vdc 85019-4J
85017-4J -24Vdc
X12
X1
Cass. Tray sensor
X1
Stereo panel 85049-4J
Legend:
85050-4J
Stereo Motor
Ctrl/ Data signal +XVdc
Voltage regulator 86005-4P CX22 Molex
Power signal Stereo Tacho +5Vdc 90130- +12Vdc
CS24 Cassette Motor 3210 87C51FA
X4
Connector +XVdc 85048-4J & Grid Motor Cass. Locks &
Supply voltage Cass. Sensor AEC Board
CXxx Cable Connector Drive Board CX23
SBxx Signal Bus Cable Image Receptor Support
PWxx Power Supply Cable
CSxx Signal & Drive Cable
CBxx Cable Assembly Dokumentti luotu Micrografx Designer ohjelmalla
,167580(17$5,80,0$*,1*2)),&(62876,'(2)),1/$1'
,QVWUXPHQWDULXP,PDJLQJ,QF
300 West Edgerton Avenue, Milwaukee, Wisconsin 53207, USA
Tel. (800) 558 6120 • +1 414 747 1030 • Fax +1 414 481 8665
info@usa.instrumentarium.com • usa.instrumentarium.com
,QVWUXPHQWDULXP,PDJLQJ)UDQFH
4, Avenue des Roses, 94386 Bonneuil Sur Marne Cedex, France
Tel. +33 1 43 39 51 51 • Fax +33 1 43 39 75 75
instrumentarium.imaging@wanadoo.fr
,QVWUXPHQWDULXP,PDJLQJ,WDOLDVUO
20041 Agrate Brianza (MI) Centro Direzionale Colleoni,
Viale Colleoni, 11, Interno 3, Italia
Tel. +39 039 689 0408 • Fax +39 039 609 1299
instrumentarium@it.instrumentarium.com • www.instrumentarium.it
,QVWUXPHQWDULXP,PDJLQJ'HQWDO*PE+
P.O.Box 2044, 77680 KEHL AM RHEIN, Germany
Tel. +49 7851 932 90 • Fax +49 7851 932 930
kontakt@instrumentarium-imaging.de
,QVWUXPHQWDULXP,PDJLQJ=,(+0*PE+
Isarstrasse 40, D-90451 Nuremberg, Germany
Tel. +49 911 642 070 • Fax +49 911 642 0739
info@ziehm-eu.com • www.ziehm-eu.com
=,(+0,QWHUQDWLRQDO0HGLFDO6\VWHPV
4181 Latham Street Riverside, CA 92501, (USA)
Tel. + 1 909 781 6660 • Fax +1 909 781 6457
mail@ziehm.com. • www.ziehm.com