GeneXpert® System
(l, II, IV, XVI)
Virtual training for the COVID-19 pandemic
2 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Learning Objectives
The general objective of this module is to guide you through the installation of
the GeneXpert® System
At the end of the training, you will be able to know more about:
• What is included in the package, and what additional items you would need
• Environmental conditions to consider for proper installation
• Setting up the System equipped either with a desktop or a laptop
• Setting up the Windows Operating System
• GeneXpert® Dx Software setup
• What information to communicate to Cepheid
3 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
The Package
4 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
The Package
GeneXpert® System* Cartridge
• Thermal and optical module(s) (l, II, IV, XVI) • Self-contained
• Disposable
• Computer System and GeneXpert® Dx Software
• Assay Specific Protocol
• Barcode scanner
6 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Define the best setting
Before installation, make sure that the GeneXpert® system will be:
In a well-ventilated room
Not directly under an air conditioning vent or a window and in no direct sunlight
5-10 cm of clearance surrounding the instrument from the walls or other instruments
Near a power source (maximum 3 required; one when connection to the UPS)
7 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
System Power Consumption Requirements
• Power Supply: Auto Rating
• GeneXpert® Energy Consumption information
On Power Mode Annual Energy Standby Power
System Size
Consumption (W) Consumption (KWh) Consumption
GX-I 61 263 58
GX-II 85 372 71
GX-IV 100 489 83
GX-XVI 270 1168 170
Infinity-48 2426 5840 1248
Infinity-80 2426 5840 1248
9 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
GeneXpert® System Setup
Desktop Configuration
10 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Installation of a GeneXpert® System with Desktop + UPS
11
301-0181, Rev. C
© 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device. US-IVD For in vitro diagnostic use.
Installing APC UPS
12 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device. US-IVD For in vitro diagnostic use.
GeneXpert® System Setup
Laptop Configuration
13 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Installation of a GeneXpert® System with Laptop + UPS
14
301-0180, Rev. C
© 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Installing APC UPS
15 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Powering Up the System
16 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Start the GeneXpert® System
1 Switch on the GeneXpert®. 2 Switch on the computer. 3 Operating System:
A blue light appears at the Cepheid-Admin account
front of the System. Password: cphd
17 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Verify successful start-up of the GeneXpert® System
4 The GeneXpert® Dx 5 Verify modules availability. 6 Configure your Windows
software starts Operating System
automatically.
Click No.
18 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Computer Setup
Windows Operating System Configuration
19 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Setting Local Date and Time
1 Access the Date 2 Set the correct local date 3 Select the correct local
and Time menu. and time. time zone.
20 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
GeneXpert® Dx Software Setup
21 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Software Compatibility
• Verify the software compatibility with the assay version that will be used: New
Assay may require an updated software version.
‒ If you do not have the correct software version, contact Cepheid technical Support
• Assay Definition File (protocol) should be loaded for any new assay added to
the System (slide 27).
22 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
GeneXpert® Dx Software Configuration
System Name
Modify the System Name
(if applicable). Do not use
characters such as
/ \ : * ? “ < > |.
Patient Data
Select the relevant
options as per your need.
Login Option
Decide if the Login request
should come up anytime or
after a defined period.
Additional Options
Set according to your needs
Example: If you need hard copies, put a
check mark on the print report.
23 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Create an Administrator Account (“Admin”)
Define the person in the laboratory that will have full access to the software
There can be more than one Administrator account per System
1 Credentials
Fill in the requested
information and choose
“Admin” as User Type
2 3
Additional users may be defined as Basic or Detail that will have limited access to specific software features.
Refer to the User Type Configuration menu for more details.
24 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device. US-IVD For in vitro diagnostic use.
Import the Assay Definition File (ADF)
Appropriate Assay Definition File needs to be imported prior beginning to work with a new Xpert Assay.
It only needs to be imported once except if notified otherwise.
Insert the CD provided in the cartridge kit in the CD drive of your computer. 3 Locate the CD
1 Choose Define Assays
Xpert Assay.gxa
Xpert Assay.gxa
2 Click on Import
25 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Data to Collect
26 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.
Documents to save
After the installation is complete:
• FAS will complete D6607 document
• The customer will generate an Installation Qualification Report in the
GeneXpert® GX Software, print, sign and send to the FAS before the virtual
training is complete.
28 © 2020-2022 Cepheid. All rights reserved. US-IVD. In Vitro Diagnostic Medical Device.